CTM 2000 KN

The document is a calibration certificate for a Digital Compression Testing Machine belonging to Jindal Steel & Power Ltd, issued on August 1, 2024. The calibration was performed on July 30, 2024, with results showing various applied forces and their corresponding observed readings, indicating the machine's accuracy. The certificate also includes details about environmental conditions, standard equipment used, and the next due date for calibration, which is set for July 30, 2025.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
30 views1 page

CTM 2000 KN

The document is a calibration certificate for a Digital Compression Testing Machine belonging to Jindal Steel & Power Ltd, issued on August 1, 2024. The calibration was performed on July 30, 2024, with results showing various applied forces and their corresponding observed readings, indicating the machine's accuracy. The certificate also includes details about environmental conditions, standard equipment used, and the next due date for calibration, which is set for July 30, 2025.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

F. No.

:ATSPL/QPM-32F Page 1 of 1
ULR No. : CC337624000003588F Certificate No. : ATSPL24000003588
CUSTOMER NAME & ADDRESS:- Service Request No. &Date : 0532 & 29-07-2024
M/s. Jindal Steel & Power Ltd Date of Receipt : 30-07-2024
Kharsia Road, Raigarh, (C.G) Pin Code -496001 Date of Calibration : 30-07-2024
Sug. Cal. Due Date : 30-07-2025
Issue Date : 01-08-2024
ITEM'S DESCRIPTION & IDENTIFICATION
Instrument Name : Dig. Compression Testing Machine Cal. Mode : Compression
Make : Aimil Ltd Range / Size : 0 to 2000 kN
Model No. : 317E-DG-1 Least Count : 0.01 kN
Serial No. : 232753 Visual Inspection : Ok
Party ID Mark : ------- Location : ------
ENVIRONMENTAL CONDITIONS Calibration Procedure : ATSPL/CP/M-58
0
Temperature : (25 ± 10) C Reference Standard : IS: 1828
Relative Humidity : (50 ± 20)% Calibration Performed at : Site
STANDARD EQUIPMENT USED
Standard Instrument Name Serial/Ident. No. ULR/ Certificate No. Valid upto
Load Cell 134764401 23090795/DI.05/C-373 23-12-2025
CALIBRATION RESULT
Division at Division at Observed Reading (Div.) Rel.
AVG. of AVG. of Rel.
Applied Force Std Temp. Room Reptitab Rel. Resol. Rel. Zero Error
24 °C Temp. 27.2 Reading Reading Acccuracy
(KN) 0° 120° 240° lity Error Error (a)% (F0)%
°C Division KN Error (q)%
(b) %
200 13319 N/A 13273 13246 13289 13269 199.3 0.37 0.13 0.01 0.00
400 26676 N/A 26770 26589 26569 26643 399.5 0.13 0.34 0.00 0.00
600 40072 N/A 40018 40026 39996 40013 599.1 0.15 0.03 0.00 0.00
800 53393 N/A 53325 53289 53339 53318 798.9 0.14 0.04 0.00 0.00
1000 66751 N/A 66668 66712 66701 66694 999.1 0.09 0.03 0.00 0.00
1200 80133 N/A 79849 79989 79959 79932 1197.0 0.25 0.08 0.00 0.00
1400 93471 N/A 93247 93166 92953 93122 1394.8 0.37 0.13 0.00 0.00
1600 106832 N/A 106755 106572 105965 106431 1594.0 0.38 0.32 0.00 0.00
1800 120200 N/A 120169 119882 119892 119981 1796.7 0.18 0.11 0.00 0.00
2000 133557 N/A 132989 132891 133530 133137 1993.7 0.32 0.21 0.00 0.00
Removal Force 0 0 0 0.0 - - - - -
Class of Machine : I Expanded Uncertainty of Measurement: ± 0.61 %
NOTE :
1. DUC = Device Under Calibration, UUC=Unit Under Calibration
2. Expanded uncertainty of measurement at approx 95% confidence level for coverage factor (k) = 2
3. Suggested next due date of calibration is given as requested by customer.
4. DUC has been calibrated againest lab standards are traceable to national & international standards.
5. This result of calibration refer to only to the particular item submitted for calibration.
6. Certificate issued is only scientific or industrial purposes only.
7. The result produced in this certificate are valid under the stated conditions at the time of calibration.
No report should be reproduced, except in full, unless a written permission for the publication of an approved abstract has been
8.
obtained from the CEO, Accutech Technical Services Pvt. Ltd., New Delhi.
Issued & Authorised By

Sanuj Kumar Harshikesh kumar


(Calibrated By) *End Of Report* ( TM / QH )