VGi - User Manual
VGi - User Manual
www.qrverona.it
COD. 97050272
NOTE
This document is provided for the own use of the operator of the equipment.
QR s.r.l. reserves the right to change the contents of this manual without notice.
This document may not, in whole or in part, be modified, copied, reproduced, distributed, translated, stored on
magnetic or optical media and published, over networks, electronic bulletin boards, web sites or other on-line
services, without the express written permission of QR s.r.l.
The original version of this manual has been written in English language.
Any change, modification, update on either hardware1, or software2, on the device and operating settings as
released form factory and as installed (and in any case different from what has been specified on the device's
attached documentation), can cause the partial or total device impossibility to operate as expected. Moreover its
safety features can be changed with increased hazard for the patient, operator and surrounding environment.
So, in case that the final user need to modify or change any of the above specified items, he/she has to be allowed in
written by QR s.r.l.
Any act outside of what has been stipulated on this notice will cancel the warranty. Moreover the person and/or
organization will be liable for any damage and/or incident and/or deterioration in the state of health of a patient,
user or other person and/or the surrounding environment.
1 Adding a memory expansion, a new board on the connection bus, a new printer, the replacement of the graphic adapter is a
significant change.
2 Including the Operating System an applications already installed when the device has been delivered and/or installed. Operating
system automatic updates, changes on the network connection parameters, modification and/or addition and/or removal of hardware
interface software (device driver) and/or system services (example: file sharing) and/or applications, constitute a significant change.
SOMMARIO
1 INTRODUCTION TO THIS MANUAL ......................................................................................................... 1-1
1.1 CONTENTS ................................................................................................................................................ 1-1
1.2 STRUCTURE ............................................................................................................................................... 1-1
1.3 STYLE CONVENTIONS ................................................................................................................................... 1-2
1.4 ABOUT USING THIS MANUAL .......................................................................................................................... 1-2
2 ABOUT SAFETY ....................................................................................................................................... 2-1
2.1 GENERAL SAFETY RULES ............................................................................................................................... 2-1
2.2 DEVICE SWITCHING ON AND TURNING OFF ........................................................................................................ 2-3
2.3 EMERGENCY STOP ....................................................................................................................................... 2-3
2.4 PATIENT'S AND USER'S SAFETY GUIDELINES ....................................................................................................... 2-4
2.4.1 Patient positioning ........................................................................................................................ 2-4
2.4.2 Patient's exit ................................................................................................................................... 2-5
2.4.3 Protection from ionizing radiation ............................................................................................... 2-5
2.5 LASER EXPOSURE ........................................................................................................................................ 2-6
2.6 SAFETY NOTES ........................................................................................................................................... 2-7
3 SAFETY AND MAINTENANCE OF THE DEVICE ........................................................................................ 3-1
3.1 INSTALLATION REQUIREMENTS ....................................................................................................................... 3-1
3.2 GUIDELINES FOR SAFETY .............................................................................................................................. 3-2
3.3 CHANGES IN THE DEVICE .............................................................................................................................. 3-2
3.4 DEVICE MAINTENANCE.................................................................................................................................. 3-4
3.5 CLEANING AND DISINFECTION ........................................................................................................................ 3-5
3.5.1 Chemical agents to avoid .............................................................................................................. 3-5
3.5.2 Cleaning ......................................................................................................................................... 3-6
3.5.3 Disinfection .................................................................................................................................... 3-6
3.5.4 Sterilization ................................................................................................................................... 3-6
3.6 TRANSPORT AND STORAGE............................................................................................................................ 3-6
3.7 DEVICE DISPOSAL ........................................................................................................................................ 3-7
3.7.1 Information for the device owner................................................................................................. 3-7
3.7.2 Information for the Disposing/Recycle centers ............................................................................ 3-7
4 GETTING STARTED… ............................................................................................................................... 4-1
4.1 INTRODUCTION TO THE SYSTEM ..................................................................................................................... 4-1
4.2 WORKING PRINCIPLE ................................................................................................................................... 4-1
4.3 OVERALL VIEW ........................................................................................................................................... 4-2
4.4 SCANNER UNIT (SCANNER) ........................................................................................................................... 4-3
4.4.1 Indicators panels on the scanner unit ........................................................................................... 4-3
4.4.2 Remote control command ............................................................................................................. 4-4
4.5 CONTROL BOX ........................................................................................................................................... 4-5
4.5.1 Main switch .................................................................................................................................... 4-5
4.5.2 Input panel..................................................................................................................................... 4-5
4.6 STANDARD ACCESSORIES............................................................................................................................... 4-6
4.6.1 Cables............................................................................................................................................. 4-7
4.7 OPTIONAL ACCESSORIES ............................................................................................................................... 4-7
4.8 SYSTEM STARTUP ........................................................................................................................................ 4-8
4.9 SYSTEM SHUTDOWN .................................................................................................................................... 4-8
5 PRELIMINARY PROCEDURES ................................................................................................................... 5-1
5.1 X-RAY SOURCE CONDITIONING....................................................................................................................... 5-1
5.2 DAILY CHECK............................................................................................................................................. 5-3
5.3 BLANK ACQUISITION.................................................................................................................................... 5-4
5.3.1 Invalidate Blank Acquisition .......................................................................................................... 5-5
5.4 BEAM LIMITER TEST .................................................................................................................................... 5-6
5.4.1 Beam Limiter Test (VGi 7 FOV option) ......................................................................................... 5-6
5.4.2 Beam Limiter Test (VGi 3 FOV option) ......................................................................................... 5-7
5.5 COMPLETELY LIFT DOWN THE ROTATING ARM ................................................................................................... 5-7
6 SCANNING ................................................................................................................................................ 6-1
6.1 SCANNING A PATIENT .................................................................................................................................. 6-2
6.1.1 Preparing the patient .................................................................................................................... 6-2
6.1.2 Positioning the patient and starting a new scan .......................................................................... 6-3
6.2 SCANNING A DENTURE ................................................................................................................................. 6-5
6.2.1 Preliminary operations .................................................................................................................. 6-5
6.2.2 Denture positioning ....................................................................................................................... 6-5
7 QUALITY ASSURANCE ............................................................................................................................. 7-1
7.1 PHANTOM POSITIONING................................................................................................................................ 7-1
7.2 SAMPLE IMAGES .......................................................................................................................................... 7-3
7.3 STORAGE OF QA DATA ................................................................................................................................ 7-3
8 TROUBLESHOOTING ............................................................................................................................... 8-1
Dedicated software routines (scan, data processing, reporting, document management) and their user
instructions are described in the “Software Manual” document.
"User Manual" and "Software Manual" should be read and clearly understood in every part before starting
using the device.
It is suggested to store this manual along with any other documentation and to use it whenever a new
staff member has to be trained for using the device.
1.2 Structure
DANGER:
It alerts for a potential danger which may cause injury
or death of a person.
CAUTION:
It signals the presence of a potential danger which may
damage the device.
NOTE:
It gives information not regarding the device and/or the
user's and/or a patient's safety.
To this purpose, chapters regarding people and device safety must be read carefully, getting confident
with the procedures described there.
The table below describes the symbols found on the device labels:
Emergency stop
In case of an emergency, follow the instructions referring to the device emergency stop, refer to Par. 2.3
- Emergency stop".
Artifacts
Please contemplate the repeating of a scan ONLY if suspect or significant artifact are evident in a
patient's images, or if a patient's position appreciably changed during the exam.
CAUTION:
Every operator must have a deep knowledge of the system
operative and environmental features, he/she also must
know the procedures to follow in case of emergency.
CAUTION:
Never use the device without the supervision of an operator.
NOTE:
Please consider the realization of an audio/video
communication system between the operator and the
patient, especially when the operator controls the device
from a protected and remote location.
The turning off of the device through an emergency button causes a prompt break in the x-ray emission
and the interdiction of every motor movement function.
CAUTION:
The emergency button must be used only in a danger
situation, such as:
Emergency
Button
Be sure that a patient is properly placed inside the scan area, with his/her head on the head support, and
that no part of his/her body may impact the device or be pressed during the positioning and the
examination.
Be sure that a patient's garments and hair may not get caught. Similarly regarding catheters and ECG
cables.
Do not perform any movement until the patient's safety is assured and there are no obstacles to the
device movement.
Refer to Par. 6.1.2 - "Positioning the patient and starting a new scan" for a proper positioning for a
patient.
After the end of an exam or after an emergency shutdown, you can let the patient go without waiting for
the rotating arm to return to the start position.
CAUTION:
NewTom VGi is a radiological device and therefore it
exposes operators and patients to the risks consequent to
the exposure to ionizing radiation. It must be used
according to safety rules that are contemplated by the
local laws regarding this matter.
Always carefully follow the applicable regulations about radio-protection and the directions from a
Certified Radiation Protection Expert.
Operator
An operator must oversee the examination process from his/her control position according to the laws in
force; NO ONE SHOULD BE BY THE PATIENT DURING THE EXAMINATION PROCESS.
If a patient panics and an operator's intervention is needed during the examination process, the operator
must be equipped with proper protection clothing and devices, according to the laws.
CAUTION:
Never stand close to the device during x-ray emission.
Patient
It is the user's responsibility to protect the patients from needless exposures to x-ray.
CAUTION:
Consider the possibility to use a leaded cloth to protect the
patient from the scattered radiation.
1. A signal on the workstation screen such as the one reported below. It appears on video just after the
START command has been sent via keyboard or mouse (see Chap. 6 - “Scanning”). It is displayed all
along the scan process.
2. Light indicators (LEDs) inside the control panels. They can be found on the sides of the Scanner Unit.
They light on just after the START command has been sent via keyboard or mouse (see Chap. 6 -
“Scanning”). They stay on all along the scan process.
CAUTION:
Do not stare into the laser beams, do not view directly
with optical instruments, and avoid direct exposure to the
beam. The beams can cause permanent eye damage.
CAUTION:
Keep your eyes at least 50 cm from the laser when the
beam is on.
If necessary, use appropriate protective glasses.
CAUTION:
The use of controls, adjustments, or the performance of
procedures other than those specified herein may result in
hazardous radiation exposure.
The device is equipped with a cross shaped positioning laser. The beam is emitted from an aperture in the
front cover.
The vertical line shows the reconstructed volume sagittal central plane. The horizontal line shows the
reconstructed volume axial central plane.
NOTE:
The manufacturer is not accountable for faults and/or
malfunctioning of parts that were not supplied or approved by
QR srl, or not installed by technical staff qualified and
certified by QR srl.
Food and beverage may not be placed or eaten/drunk close to the device and the workstation.
The user is responsible for a proper use of the system, according to the instructions and procedures which
are described inside this manual. Particularly an user must comply with the following points:
The device may be used only by authorized and properly trained (about the system and the
protection from radiation exposure) personnel. He/she may also know the regulations regarding
the use of radiological devices.
The device may never be used in case of any evident electrical mechanical or radiological
malfunctioning. Especially a device can’t be used whenever the x-ray emission displays and/or the
emergency button are not working properly.
A device may not be exposed to acids, corrosive substances, saltiness and wet.
The power supply connection must be implemented according to the laws in force and to the instructions
from the Service Manual.
Don’t use temporary connections such as adapter and extension cords to connect the computer and
peripherals to the mains.
The equipment should be permanently connected to the mains according to the instructions stated in the
Service Manual.
The medical environment for the installation has to be designed by an expert in protection from ionizing
radiation exposure according to the laws in force. Local laws shall also rule the design of the signposting.
CAUTION:
Never move a device after it has been installed. Moving a
device may damage people, the device itself or the
environment.
Connect only approved peripherals, computer and cables
to the equipment as specified by the manufacturer.
Safety systems of the device do not decrease safety measures against the risk of fire in the place in which
the device has been installed.
Electrostatic discharge
Electrostatic discharge may damage electrical parts in the device. Consequently the floor in the
installation room should be made of antistatic materials.
Extinguishers
CO2 extinguishers should be installed in an easily accessible area.
X-Ray lamp
It is possible for the user to install a x-ray lamp in order to signal the ready and emission states of the x-
ray source.
Door switches
The system contemplates the possibility for the installation of an external switch that can stop the
emission (typically installed on the entrance of the device room).
Electromagnetic compatibility
For information regarding the electromagnetic compatibility please refer to APPENDIX A - Technical
Reference.
CAUTION:
It is forbidden to open or manumit the device with any tool.
Whatever change (hardware or software) that was not
previously authorized is forbidden and can inhibit a proper
working in the device. It may also damage the patient, the
user, the device and the environment.
Responsibility limits
The manufacturer is not in charge for safety, reliability and performance features in the following cases:
Installation, maintenance, changes, repairs and/or updates were not performed by personnel
that was not directly authorized by the manufacturer or distributor.
Environmental conditions were not conform according to the requirements of the laws in
force and to the suggestions of a qualified expert.
The device is used in a way that is not corresponding to what is described inside this manual.
CAUTION:
Always turn the power off the device before any
maintenance act !
CAUTION:
The device has no repairable part inside. Never remove the
device covers.
CAUTION:
The only part that can be repaired by the user is the input
fuse. It is placed in the input panel, on the control box
side. New fuses must comply with the manufacturer
specification.
CAUTION:
In order to guarantee protection against the risk of fire,
replace only with fuses of the same type and range.
Regular maintenance
Regular maintenance is required to guarantee a proper working of the device and safety for the patient,
the user and other people.
The device maintenance and repair must be performed only by personnel that was directly authorized by
the manufacturer or the distributor. Every part of the system can only be tested and if needed substituted
by qualified personnel.
CAUTION:
If the x-ray source has not been working for longer than
three months, please perform a beam forming procedure
(see "Software Manual" - Appendix A).
Preventive maintenance
Please periodically check the workstation-scanner, workstation-control box and control box power supply
cables. Also verify the connections between the workstation parts (display, keyboard, mouse, printer)
according to the manufacturer instructions.
Malfunctioning
If the system is not working as described inside this manual, please immediately contact the technical
service.
CAUTION:
Power the device off before any cleaning.
Such products concentration in the air should never exceed the local laws' limits. In this aim always follow
the manufacturer instructions and be sure that the room is well airy.
Do not use spray or solvent which may enter the device, causing damage electrical parts or forming
inflammable mixtures. This could put in danger the patient, the user and the environment.
Do not use abrasive liquids such as acetone, gas oil or alcohol (exception: alcohol can be used for the
cleaning of the workstation display). These substances can deteriorate surfaces and materials that are
responsible for the device proper working.
Do not use substances that may release ammonia due to dissociation or decomposition. Ammonia can
corrode.
Do not use cleaning products containing silicone: it tends to accumulate and possibly damage
electrical contacts.
CAUTION:
Do not apply cleaning solutions directly on the device
surface.
NOTE:
For further information about safety and maintenance of
the device please contact your local distributor.
3.5.3 Disinfection
Despite the device do not require any disinfection, it is suggested to use standard disinfection product for
the cleansing of the head support.
3.5.4 Sterilization
This symbol, on a device label, shows that the product can’t be disposed along with other urban
garbage.
The separate collection of this equipment at the time of disposal is managed by the dealer.
When disposing the equipment the user should contact the dealer and follow the suggested
procedure to allow the separate collection and recycling of this equipment at the time of
disposal.
The separate collection and recycling of your waste equipment at the time of disposal will help
to conserve natural resources and ensure that it is recycled in a manner that protects human
health and the environment.
In case of unauthorized disposal of the device there are sanctions that vary according to the
local and/or regional regulations.
For workstations and every other peripheral disposal it is necessary to refer to the attached instructions
provided by their manufacturers.
Divide the x-ray source, the mechanical and electrical parts, plastic covers and workstation with its
peripherals.
X-ray sources contain oil that must be extracted for its proper disposal.
For other parts that have no specific instructions by its manufacturer, please refer to the local laws and
guidelines about health, safety and environment protection.
A patient is made to lean on the head support and centered by means of 2 laser modules and "scout view"
images.
The scanning system performs a completely rotation around the patient's head. Radiological images are
acquired, that are then automatically processed by the system. The result is the slices set that forms the
reconstructed volume. At the end of this process the axial slices set composes the Volumetric Data.
Through these data it is possible to display coronal and sagittal views of the reconstructed volume in real
time.
After defining a Region Of Interest (ROI), from the volumetric data the user can start the creation of a
study. The ROI can be inclined from the volumetric data both to obtain perpendicular images and to
correct positioning errors.
Working on the study data it is possible to create panoramic, transaxial and 3D images. You can also work
on these kinds of images measuring distances, angles, putting comment etc.
Finally new images can be saved inside the study.
The study images can be used to compile a report, that can be then printed and/or saved on a physical
support.
direct reconstruction of any set of the scanned object points without passing through axial
reconstruction and data re-formatting;
total scan time related to the acquisition electronics, rather than to the x-ray tube power and the
mechanics, usually shorter;
under same conditions of total scan time, less requirements in regards to the source/tube assembly
power and scan mechanics, with constructive and maintenance advantages.
SCANNER
CONTROL
WORKSTATION BOX
NOTE:
The system may not be expanded with parts or accessories
other than the ones described inside this manual.
The scanner unit is the main part of the system, it performs the scan. Two indicators panels are placed on
the columns side. They have the same appearance and function: the indication of the state of the "Power
ON" and the state of x-ray emission.
The light display the "Power On" device state and the light display of the x-ray source emitting state and
the control of the positioning laser beam.
Power ON indicator:
A green LED turn on after the push of the main switch located on
the control box.
R1 R2
V1 V2
EB L
H1 H4
H3 H2
H5 H6
Now it comes described the function of every button placed on the remote control:
EB Enable button:
it enables the functioning of the buttons for rotating arm / vertical
movements.
L Laser button:
press to turn the positioning laser beam on/off. The laser will stay
light up for about 60 seconds
On the external surface of the control box you can find the main switch to turn the system on/off and the
fuse support.
Main Switch
Emergency Button
On the same side of the main switch You can find the panel from which cables depart to the power
supply, to the scanner and to the CAN bus interface on the workstation.
Optional output
for external lamp
Head support:
it is fixed to the scanner unit. It makes possible for correct
patient's positioning and prevents excessive movement of
the patient during the scan.
QA phantom:
used for the execution of the quality assurance procedure.
Calibration support:
it allows for placing the QA phantom on the scanner unit.
Denture support:
it allows for placing the denture on the scanner unit.
The device also includes linking cables between Workstation - Scanner Unit and Workstation - Control Box.
They are:
Furthermore, a bundle of cables exits the Control Box and connects to the Scanner Unit.
The manufacturer also provides the power supply cable with an end connected to the Control Box, it is in
charge of the user to connect it to the mains during installation.
CAUTION:
Accessories, transducer and cables different from the ones
here specified may change the electromagnetic
compatibility characteristics of the device.
1. Power the Scanner Unit on with the main switch (on the Control Box).
4. Log in the operating system with user name and password as described in the “Software Manual”.
NOTE:
Always turn the scanner unit on first. If You try to use the
NNT software before the appliance has done its
initialization successfully, an error will occur.
2. Shutdown the Operating System and attend the turning off of the workstation.
3. Turn off the device by using the main switch placed on the input panel of the Control Box.
CAUTION:
Turn off the device when it has not been used for longer
than 3 hours.
CAUTION:
Always turn the device off at the end of the working time.
X-Ray source conditioning and Blank Acquisition must be performed every two weeks (14 days), while
Daily check is mandatory every day, before patients acquisitions.
If these procedures are not performed the software will disable the scan functions.
Operating modes are described in detail in the same named chapter from the "Software Manual".
NOTE:
In case of mobile device (NewTom™ VGi Flex), if the
ambient temperature is too high or low, it is recommended
to restore it into the device operating range (+10 ÷ +35 °C).
After that, attend at least 2 hours in order that the thermal
equilibrium is restored.
The software asks to perform it every two weeks. It is suggested to perform the conditioning also if the
device does not operate for days.
At the end of conditioning process, the software will automatically starts Daily check and Blank
Acquisition procedures.
CAUTION:
Before starting the procedure, please verify that the scan
area is completely free from objects.
REMOVE
This functionality is valid only for the main workstation. To invalidate the Blank Acquisition please follow
the steps below:
From the window bar select Tools Beam Limiter Test. Select the desired FOV.
Set the radiological parameters depending on the selected FOV (Standard: SFS, 1 mA, 10 msec, KV = 110;
HiRes standard: SFS, 6 mA, 10 msec, KV = 110)
Perform an acquisition.
From the window bar select Tools Beam Limiter Test. Select the desired FOV.
In order to verify the position of specific beam limiter for HiRes Zoom and Zoom fields:
CAUTION:
Prima di usare questa funzione accertarsi che non vi siano
oggetti che ostacolino il movimento verso il basso del
braccio!
To completely lift down the rotating arm follow the steps below:
1. From the NNT software main toolbar select Tools Scanner Test. The service window will appear.
2. From the service window toolbar select the Tools Lift down the rotating arm.
Si raccomanda inoltre di fare riferimento ai Chap. 2 – “About safety” and Chap. 3 – “Safety and
maintenance of the device”.
The choice of the modality can be performed selecting the corresponding FOV button from the "Scan
Manager" panel located on the bottom right corner of the NNT main window.
Figura 6: "Scan Manager" panel of the NNT software (VGi 7 FOV option)
Room setting
Be sure that the scanner unit is clean and ready to scan ("Daily Check" and "Blank Acquisition" already
performed).
Particular patients
A special care has to be taken in case of children, elderly people, claustrophobic people, physically or
mentally disabled.
Proper breathing
Ask the patient to breathe slowly during the scanning (a slow and continuous breath helps from
swallowing).
Relaxation
Ask the patient to keep the mouth closed without grinding the teeth together.
Avoiding delays
Good examination times can be obtained also by carrying out all the preliminary procedures before
starting the scan itself.
Voice instructions
Introduce the patient possible voice instructions that the operator may give during the scan.
Hereafter is a description of how to position and center a patient inside the scan area. Perform these
action just when the software signals them.
CAUTION:
The scan area (the place in which a patient is positioned
during the scan) must be free from any objects, except
from the head support. They may damage the patient or
invalidate the scan results.
NOTE:
Ask the patient to close the eyes before turn on the
positioning lasers.
NOTE:
If the patient ends up in a forced position, you may
need to re-adjust the height of the gantry via the
remote control.
Data about the denture are to be inserted, if they are not already in the software database. For details
refer to the “Software Manual”.
Hereafter are the operations to perform to properly position and center the denture inside the scanning
area. Perform such steps according to the software instructions.
Remove from the head support the movable lower part for
the patient's chin.
Before starting the Quality Assurance (QA) procedure the FOV must be selected.
REMOVE
1a scout view
2a scout view
Side view.
Axial view.
Panoramic view.
It is possible to save a copy of the QA report analysis in PDF format: select menu File Save as PDF.
Here are reported specific instructions to perform such tests (please refer to norm IEC61223-3-5) along
with the corresponding paragraph from the norm.
Not applicable.
5.2 Patient positioning 5.2.1 Axial patient positioning 10mm from isocenter
accuracy accuracy
Standard Voxel
[15x15] 18 s / 3.6 s
[15x12] 18 s / 3.6 s
[12x8] 18 s / 3.6 s
[8x8] 18 s / 3.6 s
[12x8] HiRes 26.7 s / 5.4 s
[8x8] HiRes 26.7 s / 5.4 s
[6x6] HiRes 26.7 s / 5.4 s
[15x15]:
Ø max 15 cm, h max 15 cm
[15x12]:
Ø max 15 cm, h max 12 cm
[6x6] HiRes:
Ø max 6 cm, h max 6 cm
Height 1500 mm
Width 272 Kg
Weight 925 mm
S/N 90 dB
Full:
Image pixels 768 x 960 Pixels
Pixel depth 14 n.
HiRes Zoom:
Image pixels 1344 x 1088 Pixels
Pixel depth 14 n.
Zoom:
Image pixels 672 x 544 Pixels
Pixel depth 14 n.
[15x15]:
Image pixels 768 x 960 Pixels
Pixel depth 14 n.
[15x12]:
Image pixels 768 x 848 Pixels
Pixel depth 14 n.
[12x8]:
Image pixels 672 x 544 Pixels
Pixel depth 14 n.
Pixel depth 14 n.
[12x8] HiRes:
Image pixels 1344 x 1088 Pixels
Pixel depth 14 n.
[8x8] HiRes:
Image pixels 1088 x 1088 Pixels
Pixel depth 14 n.
[6x6] HiRes:
Image pixels 800 x 864 Pixels
Pixel depth 14 n.
Reconstructed volume
Full:
Shape Cylinder
Image pixels 512 x 512 614 x 614 680 x 680 672 x 672 Pixels
HiRes Zoom:
Shape Cylinder
Image pixels 820 x 820 884 x 884 672 x 672 672 x 672 Pixels
Zoom:
Shape Cylinder
Image pixels 410 x 410 492 x 492 614 x 614 820 x 820 Pixels
Image pixels 512 x 512 614 x 614 680 x 680 672 x 672 Pixels
[15x12]:
Shape Cylinder
Image pixels 512 x 512 614 x 614 764 x 764 672 x 672 Pixels
[12x8]:
Shape Cylinder
Image pixels 410 x 410 492 x 492 614 x 614 820 x 820 Pixels
[8x8]:
Shape Cylinder
Image pixels 272 x 272 326 x 326 408 x 408 544 x 544 Pixels
[12x8] HiRes:
Shape Cylinder
Image pixels 820 x 820 884 x 884 672 x 672 672 x 672 Pixels
Image pixels 544 x 544 652 x 652 672 x 672 672 x 672 Pixels
[6x6] HiRes:
Shape Cylinder
Image pixels 410 x 410 492 x 492 614 x 614 672 x 672 Pixels
Model HF1R
Mechanical data
Weight 15 kg
Electrical Data
Cathode to Ground 60 kV
Anode to Ground 60 kV
Rotor HF1R
INPUT
Maximum current 16 A
OUTPUT
MECHANICAL DATA
Weight 7 kg
2.411 4.512
Full
2.013
HiRes Zoom -
7.314 3.315
Zoom
Date: 2011-01-12
Wavelength 635 nm
Classification Class 1
Other data
Absorbed power 200 V ( 10%) / 215 V ( 10%) / 230 V ( 10%) / 240 V ( 10%)
50/60 Hz ( 1%)
5 A (during x-ray emission)
0.66 A (stand by)
Clause 6.8.3.201 Technical Description - Tab 201 Guidance and manufacturer's declaration
- electromagnetic emissions - for all equipment and systems
TABLE: Guidance and manufacturer's declaration - electromagnetic emissions
The device NewTom VGi is intended for use in the electromagnetic environment specified below. The
customer or the user of the device NewTom VGi should assure that is used in such an environment.
RF emissions Group 1 The equipment NewTom VGi uses RF energy only for its internal function.
CISPR11 Therefore its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions Class A The equipment NewTom VGi is suitable for use in all establishments other
CISPR11 than domestic and those directly connected to the public low-voltage power-
supply network that supplies buildings used for domestic purposes.
Voltage Complies
fluctuation/flicker
emissions
IEC 61000-3-3
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment –
guidance
Electrical fast transient ±2 kV for power supply lines 2 kV Mains power quality should be that of a
/ burst typical commercial and/or hospital
IEC 61000-4-4 ±1 kV for input/output lines 1 kV environment.
<5% Ut
Functions Interruption
(>95% dip in Ut) for 5 sec
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment –
guidance
Notes:
(1) At 80 MHz and 800MHz, the separation distance for the higher frequency range applies
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people
*Field strengths from fixed transmitters, such as base stations for radio (celular/cordless), telephones and land mobile radios,
amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field
strength in the location in which the NewTom VGi is used exceeds the RF compliance level above, the NewTom VGi should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating NewTom VGi
**Over the frequency range 150 kHz to 80 MHz, field strhgths should be less than 3 V/m
For transmitter rated at a maximum output power not listed above, the recommended separation distance in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (w) according to the transmitter manufacturer.
Notes:
(1) At 80 MHz and 800MHz, the separation distance for the higher frequency range applies
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
DANGER:
Use of accessories, transducers, and cables other than those
specified may result in degraded electromagnetic
compatibility performance of this device!
CAUTION:
NewTom VGi should not be stacked directly on top of other
equipment, and other equipment should not be stacked on
top of NewTom VGi. If stacking is necessary, observe
NewTom VGi to verify normal operation in the stacked
configuration in which it will be used!
Essential performance
In case of a scan stopped by a temporary or permanent malfunctioning, the operator will have the
opportunity to store the data acquired until the failure appeared.
The quality of the reconstructed images will vary depending on the size of acquired data but it will be
probably lower compared to the standard performance of the equipment.
- IEC 60601-1: 1988-12 Medical electrical equipment - Part 1: General requirements for safety
- IEC 60601-1-1: Edition 2.0 2000-12 Medical electrical equipment - Part 1-1: General requirements for
safety - Collateral standard: Safety requirements for medical electrical systems
- IEC 60601-1-2: 2001-09 Medical electrical equipment - Part 1-2: General requirements for safety -
Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-3:1994-07 Part 1: General requirements for safety - 3. Collateral standard: General
requirements for radiation protection in diagnostic X-ray equipment
- IEC 60601-1-4: Edition 1.1 2000-04 Medical electrical equipment - Part 1-4: General requirements for
safety - Collateral Standard: Programmable electrical medical systems
- IEC 60601-1-6: Edition 2.0 2006-06: Medical electrical equipment - Part 1-6: General requirements for
basic safety and essential performance - Collateral Standard: Usability
- IEC 60601-2-7: Edition 2 1998-02 Medical electrical equipment - Part 2-7: Particular requirements for
the safety of high-voltage generators of diagnostic X-ray generators
- IEC 60601-2-28: Edition 1.0 1993-03 Medical electrical equipment - Part 2: Particular requirements for
the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
- IEC 60601-2-32: Edition 1.0 1994-03 Medical electrical equipment - Part 2: Particular requirements for
the safety of associated equipment of X-ray equipment
- IEC 601-2-44: Edition 2.1, 2002-11 Medical Electrical Equipment - Part 2.44:Particular requirements
for the safety of X-ray equipment for computed tomography
- IEC 60825-1: 2003-02 Edition 4, Safety of laser products - Part 1: Equipment classification,
requirements and user's guide
- IEC 62304 Edition 1.0b: 2006 Medical device software - Software life cycle processes 21 CFR 1020.33
Computed tomography (CT) equipment April 1, 2006]
Use with flammable anesthetics Have not been evaluated for use in the
presence of a flammable anaesthetic mixture
with air, oxygen, or nitrous oxide.
Sterilization and disinfection methods The device is supplied not sterile and it must
not be subjected to sterilization.
(See Chap. 3.5 - “Cleaning and disinfection”)
A B C
A B C
产品名称 口腔X射线计算机体层摄影设备
产品型号 NewTom VG
系统编号 VGXXXXXX
运行方式 间歇加载连续运行
输入电压 220AC+/-10% 50/60Hz
输入功率 瞬时:5A 稳态: 0.75A
制造商 QR s.r.l. 意大利
生产地址 Via Silvestrini,20 - 37135 维罗纳 意大利
生产日期 YYYY-MM
安全标志
Position: left side of the control box, on the right of the control box plate
Inverter
Position: on the structure of the X-Ray source and on the Inverter structure.
Position: on the cover of the scanning unit in frontal and centered position
(NewTom VGi Flex label: device NewTom VGi Flex only)
QR s.r.l. – Italy
P/N 96600741
P/N 99934493
S/N _____________
QR s.r.l. – Italy
CuZn - P/N 99934502
3.90 mm Al @ 70 kV
QR s.r.l. – Italy
CuZn - P/N 99934491
6.70 mm Al @ 70 kV
QR s.r.l. – Italy
CuZn - P/N 99934334
6.70 mm Al @ 70 kV
QR s.r.l. – Italy
CuZn - P/N 99934547
3.90 mm Al @ 70 kV
QR s.r.l. – Italy
P/N 99934325
S/N _____________
ON-OFF
I
FUSE:
Condition Ratings
100 - 115 V~ F1=T10A 250V
200 - 240 V~ F1=T5A 250V
CAUTION: For continued
protection against risk of fire,
replace only with same type
and rating of fuse.
ATTENTION: Pour une
protection continue contre les
risques d'incendie, remplacez
seulement avec an fusible de
le même type et
caractéristiques
ATTENTION:
RAYONS X ÉMISSION DE RAYONS X QUAND
LE DISPOSITIF EST EN MARCHE
Position: left side of the control box, near/under the control box plate
AUX
CAN-BUS
Wavelenght:630-680 nm Max.Output<1mW CW
IEC 60825-1:1993 + A1:1997 + A2:2001
Position: left side of the control box under the emergency push-button
QR srl
Via Silvestrini, 20
37135 Verona, Italy
Phone: +39 045 8202727
Fax +39 045 8203040
e-mail: [email protected]