CALIBRATION CERTIFICATE
ULR No. : CC103925000000791F
Certificate No. : NTS/0791 Date of Calibration : 2/20/2025
Metarial Receipt Date. : 2/19/2025 Suggested Due Date of Calibration : 2/19/2026
Certificate Issue Date : 2/21/2025
Client Name : M/s. SANDHU ENTERPRISES
Address : PLOT NO.-133/1, FAUZI MATKOTA, BHURARANI, RUDRAPUR (U.S.NAGAR) -
263153 (U.K.)
Instrument Details
Name : HIPOT TESTER Uncertainty of measurement : As per Range
Make : ASSOCIATED RESEARCH (at 95% confidence level, K=1.96)
Model No. : GB-001 Visual Inspection : OK
Inst. Sr. No. : --- Ref. Temp : (25 ± 4)0C
Inst. Id. No. : SE/HPT/01 Ref. Relative Humidity : (55 ± 20)%RH
Range : Ref. to result Calibration Procedure : ET-M-CP-01, 02
Least Count : Ref. to result 03, 04, 05
Location : EXPORT
Standard Equipment used for calibration
Name Sr. No. Make/ Model Calibrated By Certificate No. Validity of Cal.
Multifunction Calibrator 2941101 FLUKE / 5080A AIPL ECPL/24/ET/L0181-02 14.05.2025
NOTE : -
1. The calibration results reported in this certificate are valid at the time of and the stated condition of measurement.
2. This report should not be reproduced except in full without our prior permission in writing.
3. Calibration certificate without signature are not valid.
4. All our certificates are Traceable to National standard..
5. DUC : Device Under Calibration.
Calibrated By (Neeraj Kumar) Issued / Approved By
F02(5.10/01) (H.P. Sharma)
Quality Manager
CALIBRATION CERTIFICATE
ULR No. : CC103925000000791F
Certificate No. : NTS/0791 Date of Calibration : 2/20/2025
Metarial Receipt Date. : 2/19/2025 Suggested Due Date of Calibration : 2/19/2026
Certificate Issue Date : 2/21/2025
HIPOT TESTER
Make : ASSOCIATED RESEARCH
Sr. No.: ---
OBSERVATION :-
SET VALUE MEASURED VALUE % Error of ±Expended Uncertainty at
PARAMETER
(MASTER) (DUC) Reading (±) 95% Confidence Level (k=1.96)
DC Voltage
6000V (L.C-0.1mV) 10.00 V 9.9 mV -1.00 0.011 mV
20.00 V 19.8 mV -1.00 0.011 mV
50.00 V 49.6 mV -0.80 0.0002 V
100.00 V 99.95 V -0.05 0.0002 V
200.00 V 199.994 V 0.00 0.0008 V
400.00 V 399.957 V -0.01 0.0008 V
500.00 V 499.978 V 0.00 0.087 V
600.00 V 599.966 V -0.01 0.087 V
700.00 V 699.956 V -0.01 0.82 V
800.00 V 799.963 0.00 0.82 V
1000.00 V 999.897 -0.01 0.82 V
AC Voltage (At 50Hz)
5000 V (L.C-0.01mV) 10.00 V 9.7 V -3.00 0.12 mV
20.00 V 19.6 V -2.00 0.6 mV
50.00 V 49.5 V -1.00 0.0037 V
100.00 V 99.97 V -0.03 0.0037 V
200.00 V 199.986 V -0.01 0.037 V
400.00 V 399.965 V -0.01 0.037 V
500.00 V 499.975 V 0.00 0.51 V
600.00 V 599.954 V -0.01 0.51 V
700.00 V 699.943 V -0.01 1.96 V
800.00 V 799.955 V -0.01 1.96 V
1000.00 V 999.956 V 0.00 1.96 V
Calibrated By (Neeraj Kumar) Issued / Approved By
F02(5.10/01) (H.P. Sharma)
Quality Manager
CALIBRATION CERTIFICATE
ULR No. : CC103925000000791F
Certificate No. : NTS/0791 Date of Calibration : 2/20/2025
Metarial Receipt Date. : 2/19/2025 Suggested Due Date of Calibration : 2/19/2026
Certificate Issue Date : 2/21/2025
HIPOT TESTER
Make : ASSOCIATED RESEARCH
Sr. No.: ---
OBSERVATION :-
SET VALUE MEASURED VALUE % Error of ±Expended Uncertainty at
PARAMETER
(MASTER) (DUC) Reading (±) 95% Confidence Level (k=1.96)
DC Current
40mA(L.C-0.01mA) 5.00 mA 4.999 mA 0.00 0.011 mA
10.00 mA 9.956 mA -0.44 0.63 mA
15.00 mA 14.954 mA -0.31 0.23 mA
20.00 mA 19.785 mA -1.07 0.23 mA
25.00 mA 24.695 mA -1.22 0.23 mA
30.00 mA 29.954 mA -0.15 0.23 mA
40.00 mA 39.644 mA -0.89 0.23 mA
AC Current (At 50Hz)
40mA(L.C-0.01mA) 5.00 mA 4.999 mA 0.00 0.009 mA
10.00 mA 9.956 mA 0.04 0.05 mA
15.00 mA 14.954 mA 0.05 0.6 mA
20.00 mA 19.785 mA 0.21 0.6 mA
25.00 mA 24.695 mA 0.30 0.6 mA
30.00 mA 29.954 mA 0.05 0.6 mA
40.00 mA 39.644 mA 0.36 0.6 mA
Calibrated By (Neeraj Kumar) Issued / Approved By
F02(5.10/01) (H.P. Sharma)
Quality Manager