0% found this document useful (0 votes)
23 views3 pages

Certificate of Calibration: Issued By: Modern Surveying Calibration & Testing Labs

This document is a Certificate of Calibration issued by Modern Surveying Calibration & Testing Labs for a Coriolis Mass Flow Meter on July 23, 2024. The calibration was performed on July 22, 2024, with results indicating an average meter factor of 1.00113 and a deviation of -0.10%. The certificate confirms compliance with ISO/IEC 17025:2017 standards and includes details on the calibration method, environmental conditions, and metrological traceability.

Uploaded by

diyarkarim070
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
23 views3 pages

Certificate of Calibration: Issued By: Modern Surveying Calibration & Testing Labs

This document is a Certificate of Calibration issued by Modern Surveying Calibration & Testing Labs for a Coriolis Mass Flow Meter on July 23, 2024. The calibration was performed on July 22, 2024, with results indicating an average meter factor of 1.00113 and a deviation of -0.10%. The certificate confirms compliance with ISO/IEC 17025:2017 standards and includes details on the calibration method, environmental conditions, and metrological traceability.

Uploaded by

diyarkarim070
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

CERTIFICATE OF CALIBRATION

Issued by: Modern Surveying Calibration & Testing Labs

Date of Issue: July 23, 2024 Certificate #: CC12413813

Head Office:Office No. Q1-03-016/C


Sharjah Airport International Free Zone, UAE.
Tel +971 6 5182008 / +971 506348302
Main Lab: Main Gazna Road, Gazna-6
Erbil, Kurdistan, Iraq.
Tel +964 7510073521 / +964 7515361339
Email:[email protected]
Website :www.modernsurveying.net

CUSTOMER AND JOB DETAILS:


Customer Name : ‫ﻣﻌﻤﻞ ﺣﯿﺎﺓ ﺍﻟﺠﻨﻮﺏ ﻝٲﻧﺘﺎﺝ ﺍﻝٲﺳﻔﻠﺖ ﺍﻟﻤﺆﻛﺴﺪ‬
Address : ‫ﺫﻱ ﻗﺎﺭ ﺍﻟﻨﺎﺻﺮﻳﺔ‬
Request # : QN12410812
Job # : JN12420589

Location of Calibration : Modern Surveying Calibration & Testing Labs

DETAILS OF UNIT UNDER CALIBRATION:


UID No. : UID0724427060
Description : Coriolis Mass Flow Meter Type : Coriolis
Make : ENDRESS + HAUSER Model : PROMASS E 83
Serial # : D4014502000 Asset # : --
Part # : -- Calibration Site: Modern Surveying Calibration & Testing Labs
Calibration Date : July 22, 2024 Calibration Due: January 22, 2025

CHARACTERISTICS OF UNIT UNDER CALIBRATION:


Fluid Used in Calibration : DF2 Process Fluid ** -
Fluid Density @ + 15 °C & Atmospheric pressure : 840.0000 Kg/m3 Cal.Range ( m3/h ) 12
Liquid Temp of Proved Meter as per mfg
: ( -40 to +140 ) °C Proved Meter Size** 1-1/2"
spec.**
As Received K-Factor (Pulse per L)** : 2.100888 As Left K-Factor (Pulse per L) 2.193744
Flow Rate ( m3/h ) ** : 12.00 Accuracy of Proved Meter as per mfg spec.** ± 0.15 %
Average Volume ( Liter ) : 2000.0 As Received Totalizer of Proved Meter ** - Liter
As Received Meter Factor ( Previous ) ** : - As Left Meter Factor ( New ) 1.00113
As Left Repeatability % : 0.05 Certificate of Origin. ** Not Provided.

** Data Provided by Customer

Approved Signatory:
Abdul Hanan (MS)

Doc #: MSL/ET/FL/04a, Issue # 02 , Rev. # 01 , Rev. Date : Jun 05, 2024 Page 1 of 3
CERTIFICATE OF CALIBRATION
Issued by: Modern Surveying Calibration & Testing Labs

Date of Issue: July 23, 2024 Certificate #: CC12413813

CONDITION OF UNIT UNDER CALIBRATION DURING CALIBRATION:


ü Working satisfactory upon receipt
Not working satisfactory upon receipt
Calibration done after adjustment / repair

ENVIRONMENTAL CONDITIONS DURING CALIBRATION:

Start End Uncertainty ( ± )

Temperature (°C) : 48.1 48.3 0.11

Relative Humidity (%RH) : 9.0 9.0 1.1

CALIBRATION METHOD:
API MPMS 4.5 : (Master Meter Provers)

API MPMS 4.8 : (Operation of Proving Systems)

API MPMS 5.6 : (Measurement of Liquid Hydrocarbons by Coriolis Meters)

METROLOGICAL TRACEABILITY:
The Measurements in this certificate are traceable through National Metrology Institutes (National / International) to S.I.
Units.
REFERENCE STANDARD AND MEASUREMENT EQUIPMENT:
The following reference standards and measurement equipment were used in the calibration:

Description : Data Logger Temperature Transmitter Mass Flowmeter

Make : Rotronic Endress+Hauser Emerson

Model : HL-1D TMT82 CMF300M356D2BZEZZZ

Serial # : A19080369 K300861430C 21379182

Certificate # : ICC12330310 ICC12330268 21379182

Due Date : December 14, 2024 August 17, 2024 August 11, 2024

Traceability : MSL (IQAS Accredited) MSL(IQAS Accredited) Micro Motion

CALIBRATION DATA :
Flow Rates ( m3/h ) : 12
Volume ( Liter ) : 2000

Number of Repeats : 5

Results Indicated as : Meter Factor (MF)

Actual Volume passed through the meter


MF =
Volume indicated by the meter
For calibration results see next page(s)

Doc #: MSL/ET/FL/04a, Issue # 02 , Rev. # 01 , Rev. Date : Jun 05, 2024 Page 2 of 3
CERTIFICATE OF CALIBRATION
Issued by: Modern Surveying Calibration & Testing Labs

Date of Issue: July 23, 2024 Certificate #: CC12413813

CALIBRATION RESULTS :- (FOUND - LEFT)@ 12 m3/h


Master Meter Data
Flow Rate Temperature Pressure Density Volume Corrected Volume
CTL CPL VCFmm Meter Factor
m3/h °C Bar Kg/m3 Liter Liter
12.059 39.70 0.98 840.0000 2001.155 1950.887 0.97479 1.00009 0.97488 1.00000

12.124 39.70 0.99 840.0000 2000.260 1950.014 0.97479 1.00009 0.97488 1.00000

12.097 39.70 0.98 840.0000 2001.245 1950.975 0.97479 1.00009 0.97488 1.00000

12.105 39.80 0.99 840.0000 2001.148 1950.675 0.97469 1.00009 0.97478 1.00000

12.112 39.80 0.99 840.0000 2000.089 1949.642 0.97469 1.00009 0.97478 1.00000

Average Data of Master Meter


m3/h °C Bar Kg/m3 Liter Liter CTL CPL VCFmm Meter Factor

12.099 39.74 0.99 840.0000 2000.779 1950.439 0.97475 1.00009 0.97484 1.00000

Meter Under Calibration

Flow Rate Temperature Pressure Density Volume Corrected Volume Deviation


CTL CPL VCFmm Meter Factor
m3/h °C Bar Kg/m3 Liter Liter %

12.059 39.70 0.98 840.0000 1998 1948 0.97479 1.00015 0.97494 1.00152 -0.15

12.124 39.70 0.99 840.0000 1998 1948 0.97479 1.00015 0.97494 1.00107 -0.10

12.097 39.70 0.98 840.0000 1999 1949 0.97479 1.00015 0.97494 1.00106 -0.10

12.105 39.80 0.99 840.0000 1999 1949 0.97469 1.00015 0.97484 1.00101 -0.09

12.112 39.80 0.99 840.0000 1998 1948 0.97469 1.00015 0.97484 1.00098 -0.08

Average Data of Meter Under Calibration


m3/h °C Bar Kg/m3 Liter Liter CTL CPL VCFmm Meter Factor %

12.099 39.74 0.99 840.0000 1998 1948 0.97475 1.00015 0.97490 1.00113 -0.10

Expanded uncertainty
N.Repeatability % Average Meter Factor
±%
0.077 0.05 1.00113
Note 1: "The reported expanded uncertainty is based on the standard uncertainty multiplied by a coverage factor k=2, which provides a
confidence level of approximately 95%"

Note 2: Deviation = (Correct Volume of MUT - Correct Volume of MM) / (Correct Volume of MM) x 100
Note 3: Used Criteria for the repeatability ± 0.05 %.
REMARKS (TICK WHICHEVER APPLICABLE):
ü Values were rounded in computation.
** Values not covered in accredited scope
*Uncertainty value has been pulled up to CMC value as the uncertainty calculated is below CMC of the lab.
ü No statement of compliance with specifications is made or implied on this certificate.
Other
Performed by : Abdul Hanan / Ismail Shkur

End of calibration results


This certificate is issued in accordance with the requirements of ISO/IEC 17025:2017 Standard, General Requirements for the competence of
testing and calibration laboratories.The results given here in are valid only to the item calibrated.This certificate may not be reproduced other than
in full, except with the prior written approval of Modern Surveying Calibration & Testing Labs.

Doc #: MSL/ET/FL/04a, Issue # 02 , Rev. # 01 , Rev. Date : Jun 05, 2024 Page 3 of 3

You might also like