CV-7200 Computerized Vision Tester Manual
Topics covered
CV-7200 Computerized Vision Tester Manual
Topics covered
VISION TESTER
M o d e l : C V- 7 2 0 0
Operation Manual
Please observe the following precautions when using this product:
·This is a precise instrument. Install and store it where the following normal life
conditions are kept:
temperature (10~40℃), humidity (30~85%) and atmospheric pressure
(70~106kPa).
·Don’t expose this instrument directly to the sun.
·This instrument must be installed in a level and stable place to operate
smoothly. Don’t put anything on this instrument.
·Before using this instrument, connect all the cords correctly.
·Use a power supply within the range of ±10% of the rated voltage (50/60 Hz).
·Turn off the power when this instrument is not used.
1
Figure 1
1) Examination window……….The patient’s eyes are observed through the
window, and display lenses are sent here.
2)Near vision illumination……Light up when near vision test, light off when far
vision test.
2
1.2 Accessories
Figure 2
12)Face shield (2 pcs.)………..Attached to the instrument to put the patient’s
face in place.
14)Forehead rest (2 pcs.) ……. The patient’s forehead rests here. Detachable
by a touch, use one while the other is being
sterilized.
3
1.3 Power Unit
Figure 3
WARNING
To avoid potential danger, do not remove the power unit.
2. PREPARATIONS
2.1 Assembly Of The Near Point Distance Meter
Insert the card holder in the top end of the near-point rod. The card holder is
designed to slide along the rod. As shown in the illustration, insert the
near-point card in the spring attached to the card holder.
4
Card holder Near point distance meter
Figure 4
Next, insert the near-point rod into the near-point rod mount and firmly fasten
with the set screw. When not in use, hold it upright as shown in the next
illustration.
Figure 5
5
Figure 6
To remove the forehead rest, see it from a side, turn it up and pull off.
Figure 7
Before using, make sure the main body is level. If the bubble is not between
two lines, make it level by operating the leveling knob.
The forehead rest move back and forth by turning the forehead rest knob.
With the patient’s forehead set to the forehead rest, look through the cornea
aligning window from about 25cm apart.
6
Scale Patient's eye
Illumination
Collimator
Approx. 25cm
13.75mm position
Examiner's eye
12mm position
Figure 8
Look at the patient’s eye so that long lines of collimator and scale are aligned,
as shown in the illustration.
If the eye comes on the long line, the spherical power of eyeglass lens is
measured at 12mm from the cornea surface. Short lines are given at 2mm
intervals from the long line.
The dotted line is 13.75mm from the cornea surface. For example, if the
cornea surface is 4mm away (on the second short line from the long line), the
spherical power of eyeglass lens is measured at a position of 12+4=16mm. If
the eyeglass wearing distance and the measuring distance is different,
correction the Forehead rest knob as show in figure 1
When the patient wears multi-focal lenses, the eyeglass spherical power is
measured at a near distance.
Bring down the near-point rod. If the near-point rod is almost level, the position
is the measuring position.
The scale of the near-point rod is graduated in cm and inch unit.
Value matching both end lines of the card holder indicate the distance between
the eye and the card.
7
40
45
Set the desired target on the near-point card attached to the card holder.
To set the target, turn the rotary part at the bottom part of the card. The target
appears in the sight window. Target number of the near-point card and
contents:
Sight window
8
Button area divided into three parts from the function in control panel: menu
and auxiliary function key area, testing key area, vision diagram key area.
button:open/close keystone.
button:combination key
9
3.2 Testing key area
* + :in the screen of testing screen, close/open right eye, will not
change the mode of testing eye.
* + :in the screen of testing screen, close/open right eye, will not
change the mode of testing eye.
:In the screen of testing, if not insert cross cylinder, it cover the left
eye automatically, and then insert right cross cylinder, set the dial of right eye
testing, and set cross cylinder as side [1] .
10
In the screen of testing, if insert cross cylinder, then set cross cylinder as
side [1] .
In under-defined screen, set current auxiliary lens as cross cylinder, testing
eye mode is right left, and the testing parameter is axis.
:In the screen of testing, if not insert cross cylinder, it cover the right
eye automatically, and then insert right cross cylinder, set the current right eye
testing screen, and then set[1].
In the screen of testing, if insert cross cylinder, then set cross cylinder as[2].
In under-defined screen, set current auxiliary lens as cross cylinder, testing
eye mode is right left, and the testing parameter is axis.
[DIAL]:In the screen of testing, alter the value of current testing item.
Clockwise rotation: decrease the value of the current item.
Counterclockwise rotation: increase the value of the current item.
In under-defined screen, set as the alteration of the data.
[SET]:
In the screen of testing, if it is under the standard program or under-defined
* +[SET]: save the editing content, and return to the previous step
program editor.
11
* + :When external vision chart display is LCD, it can show the
reverse color.
:
If the current vision figure in a single line/single display mode, it will show the
corresponding icon above the chart. If icon is on the first line of the current
vision chart, then switch to the previous chart in the same type, and keep the
eye cover mode the same.
If the current vision figure in a single row/ without eye cover/red/green mode,
the switch to the previous chart in the same type, and keep the eye cover
mode the same.
:
If the current vision figure in a single line/single display mode, it will show the
corresponding icon under the chart. If icon is on the bottom line of the current
vision chart, then switch to the next chart in the same type, and keep the eye
cover mode the same.
If the current vision figure in a single row/ without eye cover/red/green mode,
the switch to the next chart in the same type, and keep the eye cover mode the
same.
:
If the current vision figure in a single display mode, it will show the
corresponding icon left the chart. If icon is the first row first line of the current
vision chart, then switch to the previous chart in the same type, and keep the
eye cover mode the same.
If the current vision figure in a single line mode, position shift to the left row,
, and keep the eye cover mode the same.
If the current vision figure in a without eye cover/red/green cover mode, the
switch to the single-row mode.
12
:
If the current vision figure in a single display mode, it will show the
corresponding icon right the chart. If ICONS is in the bottom right corner of the
current vision chart, then switch to the next chart in the same type, and keep
the eye cover mode the same.
If the current vision figure in a single row and eye cover mode, position
shift to the right and keep the eye cover mode the same.
If the current vision figure in a without eye cover/red/green cover mode, the
switch to the single-row mode.
13
Figure4.1 introduction of main testing interface block
The key to realize two functions: display the current eyes PD and enter, exit
pupil distance measurement mode. When entering the PD measurement mode,
the automatic test program is for binoculars PD mode.
Under PD measurement mode, the main area will display PD testing project.
Blue highlight can input the value to change. Right and left eyes auxiliary
lenses will be automatically inserted the PD measure lenses, VD observation
lamp will automatically open .Auxiliary button area display the option of VD
observation lamp. It can turn the driving plate, or switch the button+, - to adjust
the PD.
15
Figure 4.2 main test interface into PD measurement mode
The key to realize two functions: switch distance measuring mode and
display the current far and near point detection.
16
Near point measurement mode display:
*If the current mode is far vision detection, switch to the near vision detection
mode, if not taken near vision detection before, it will automatically copy S +
ADD, C, A degree in far point vision to the S, C, A degree in near vision
detection mode; If the near point detection mode has been completed, will not
copy.
*If the current mode is far point detection, switch to the near vision detection
mode, if not adjusted PD value in the near vision detection mode, it will
calculate corresponding PD value with PD value according to the far vision
mode; If completed the PD adjustment in the near point detection, which it's
irrelevant.
17
2) Detection mode selection button of right and left sphere eye
3) Display area of the sphere of left eye and detection mode selection button
for sphere of left eye
4) Display area of the cylinder of right eye and detection mode selection button
for cylinder of right eye
5) Detection mode selection button of right and left cylinder eye
6) Display area of the cylinder of left eye and detection mode selection button
for cylinder of left eye
7) Display area of the right eye axis degree and detection mode selection
button of axis of right eye
8) Detection mode selection button of axis
9) Display area of the left eye axis degree and detection mode selection button
of axis of left eye.
10, 11, 12,13,14,15 is for repeated use area. Under the prism project displays
status, reuse of prism XY model and polar coordinates mode. XY model and
polar coordinates Mode switch
Through auxiliary function buttons.
10) Display area of the right eye prism degree and horizontal prism detecting
mode selection
/ value display area of polar coordinates Mode and detection mode selection
for right eyes
11) Horizontal prism detecting mode selection / detection mode selection in
polar coordinates mode for binoculus.
12) Display area of the left eye prism degree and horizontal prism detecting
mode selection
/ value display area of polar coordinates Mode and detection mode selection
for left eye
13) Display area of the right eye vertical prism degree and vertical prism
detecting mode selection
18
/ value display area of polar coordinates Mode and detection mode selection
for right eyes
14) Binoculus vertical prism degree detection mode selection
/ detecting mode selection of polar coordinates Mode
15) Display area of the left eye vertical prism degree and vertical prism
detecting mode selection
/ value display area of polar coordinates Mode and detection mode selection
for left eyes
16) Right eye ADD display area and ADD detecting mode selection
17) Binoculars ADD detecting mode selection
18) Left eye ADD display area and ADD detecting mode selection
*when enter in to ADD, the machine automatically Run automatically to the
near point detection mode, low beam lights opens.
*If current is ADD detecting project, click any button of S/C/A, it exit ADD
detecting mode, the machine automatically Run automatically to the far point
detection mode.
19) Right eye VA value display area and VA input mode selection
20) Binoculus VA detecting mode selection
21) Left eye VA value display area and VA input mode selection
22)
4.2.5 Auxiliary function button area
19
Figure 4.5 auxiliary function buttons
Under the different detection mode, the auxiliary function button will display
different content and have different functions.
Sphere testing mode:
Function
display Operation content
button
F1 0.25D Set the sphere step as 0.25dptr。
F2 0.5D Set the sphere step as 0.5dptr。
F2 blank
F3 blank
F4 blank
F5 blank
Function
display Operation content
button
F1 0 Set the cylinder step as 0。
20
F2 45 Set the cylinder step as 45。
NC: the naked eye vision detection mode. In this mode, all the lenses value is
withdraw input VA values.
TL: call out the data of lens meter. You can choose the data import to CV, and
switch to the CV test mode.
21
RM: call out the data of auto refractometer. You can choose the data import to
CV, and switch to the CV test mode.
CV: call out CV data.
FC: final data. Copy the CV data to FC, according to the needs and experience,
adjust the final data.
22
2) Confirm the back key in Preview mode. Display In preview state, click the
button, switch the currently selected vision chart 4, and back to the basic
test interface or un-defined program editing interface.
3) Does not change the selected current vision chart, back to the basic test
interface or un-defined program editing interface.
4) Display details of the currently selected vision chart.
5) Change the preview/non- preview state
6) close and open external display/projection device.
7) Near point additional lenses testing. Press the icon button; automatically
enter to the near point, near point crossed cylindrical lens inserted into the
front of the eyes.
8) Near point vision test chart. Press the icon button; automatically enter to
the near point mode.
* Under the vision chart selection interface, if the current external projector is
LCD, press [SET] key to enter LCD's Settings menu, press [SET] key
repeatedly to switch SET , through the direction keys to change the current
Settings.
23
Figure4.8 far point left eye auxiliary lens selecting interface
Figure 4.9 Near point right eye auxiliary lens selecting interface
Figure 4.10 Near point left eye auxiliary lens selecting interface
24
Click the lens key, will change the current lens inserted in the CV - 7200
auxiliary lens state. And back to the basic test interface or un-defined program
editing interface. Click back button, will not make any change to return to the
previous interface.
Click the required procedures, directly back to the main testing interface,
callout the preset test Settings.
The sub-interface will automatically list and select the same type chart
according to the current selected chart.
Under the main test interface, press button on the panel, callout the
Classification vision chart and rapidly choose preset sub-interface, then click
again, withdraw the Classification vision chart and rapidly choose preset
sub-interface.
25
Figure 4.12 Classification vision chart and rapidly choose preset sub-interface and
Click the Classification vision chart, is the same as click the far point vision
chart under the sub-surface directly back to the main testing interface, callout
the preset test Settings.
26
button Operation content
Environment setup Enter to the Environment variable
interface to set
Time setup Enter to time set interface to set
27
environment variable .
*if there is change to the "external projection device selection", when you back
to previsou menu ,will show button of save restart or give up save.
Brief display the every step content , click on the corresponding steps, directly
into the corresponding steps to edit interface. Choose defined program and
standard program, different operating option.
28
Figure 4.16 defined Program step list interface
Standard program steps list display has no "remove" button. "program 1", "
program2”, "program 3", "program 4", " program5” ,the corresponding standard
program steps is copied to the corresponding defined program, and switch the
interface to the defined program step list interface.
29
4.3.10 defined program editing interface
In the program step list interface, click on any step, will enter into a defined
program editing interface
30
content
11 save the current modification, back to the previous step
12 Display the current selected left eye auxiliary lenses, call
out the auxiliary lens selection interface
5.Device operation
5.1 Preparation
5.2 Mode selection
Environment variable Settings the interface, select the start mode setting
31
it through the button in the main testing interface in LCD panel, then
choose the required program. After entering any program, every click [SET],
the program will be automatically performing the next step. Press [SHIFT] +
[SET], can back to the previous step program.
We suggest optometrist can set program in advance according to the different
types of customers, are stored in "program 1" to "5" program separately. It only
need to callout the corresponding program, it can greatly improve efficiency.
About programming content, will be introduced in 5.8
We will exemplify the operation process when open the machine and
Choose start mode of "no".
A customer buy glasses, always complain of poor vision.
Check firstly his eyeglasses value with lens meter
R:-1.00DS/-0.50DC X 90°
L:-1.25DS/0-0.50DC X 180°
PD=65mm
Corrected vision is only around 0.7, but we can find the basic reason may be
because of the original degree is not enough to cause vision to reduce, now is
required a better glasses prescription.
5.3.1、set degree + 3.00 D by hand to Right eye sphere value . Because the
customer the right eye degree is -1.00 D, plus + 3.00 D, become + 2.00 D.
5.3.2、the value should be less than 0.1 at this time, project a vision icon by
hand, then gradually adjust the degree, to reduce as per - 0.25 D, from +2.00D
to + 1.75 D, then to + 1.50 D, until his sight value is about 0.5.
5.3.3、call out radiation vision chart:
32
Figure5.2 radiation vision chart
Ask customer "is there the mostly clear line can you see?"
If the customer answers "look all the same", that means no astigmatism, finish
the measurement.
If the customer answer seems one line is particularly clear , then use
30 °multiply that small size number which of the clearest line, this value
negative correction.
For example:if the third line is the most clear to see,so it is 3X30°=90°
Adjust astigmatism axis, and then adjust the astigmatism value, from 0.00 to
-0.25, then to -0.50, until all the line is the same clear.
5.3.4、as - 0.25 D a unit to adjust the sphere value.以-0.25D
The customer is short sighted, sphere is -1.75 D, after roughly check, his
degree values as shown below figure:
33
5.4 accurate astigmatism axis position and degree
mode is more clearer, need to click the front ,device will automatically
adjust the astigmatism axis. then enter and buttons to repeat the
steps before, until the 2 side is the similar clear, can know the accurate
astigmatism axis value.
customer which side is more clearer, if under mode is more clearer, need
to click the front ,if under mode is more clearer, need to click the front
button ,device will automatically adjust the astigmatism value. Then enter
and buttons to repeat the steps before, until the 2 side is the similar
34
clear, can know the accurate astigmatism value.
Under the condition of Adjust astigmatism value, device can automatically link
with sphere, it means: cylindrical lens change the 0.50 DC .sphere
automatically change + / - 0.25 DS.
The result:-0.50DC×100°
5.4.4 After accurate astigmatism axis and value, enter any chart button
means accurate not enough, add -0.25D,or enter red button,device will
automatically deduce the value。That means the actual value is between
-1.50Dand -1.75D,According to the requirement and need of the customer to
choose high or low. Screen displays at this time:
35
conclusion:right eye: -1.50DS/-0.50DC×100°
next to test left eye。Close the right eye,not close the left eye。Step is 步 the
same as right eye, the result of left eye: -2.00DS/-0.50DC×170°
In order to give customers a better eyesight condition, we need to do
binoculus balance test.
36
Figure 5.7 result of binoculars balance test
First test horizontal heterophoria, select binocular mode, use the right eye lens
selecting button and add the red Maddox to the right eye.
Projection point icon, customers will see red vertical lines from right eye (figure
26a), a lighting point from left eye (figure 26b). Two eyes blending images as
shown in figure 26c or 26d. If as shown in 26d, then there is no or a blend low
strabismus, no need to adjust a prism. if as shown in 26 c, use left eye prism
button BI or BO around the mobile lighting, remind customer, when he saw the
picture as shown in figure 26d, hit to stop, test for several times and get the
average.
37
Figure 5.8 Heterophoria measurement chart
Next to test the vertical heterophoria measurement. Select binocular mode,
use the left eye lens selecting button and add the red Maddox to the left eye.
Cancel the left eye cover.
Customer will see a horizontal red line with left eye, a lighting point from right
eye. Use right prism button to BU or BD to move up/down the lighting points.
Remind customer, when he saw the horizontal line and lighting point ,hit to
stop, test for several times and get the average.
Until now, comes the result.
If heterophoria degree is too big, need to correct with the glasses lens, in this
case the customer is not very serious, so don't need to correct.
The final result is:
Left eye: -1.50DS/-0.50DC×100°
Right eye: -1.75DS/-0.50DC×170°
First, choose the chart, and press the drive plate button, switch to
38
and or press the scieropia degree button to add Cloud value. Finally,
you will get the interface as below:
During the Use of Maddox to test heterophoria, you can set as this.
Firstly, choose pointolite, and then press “lens”, the display will show
the auxiliary lens interface.
39
Press , it will automatically return to program design interface.
Setting testing item as “horizontal prism”, and Measured eye as “binoculus”.
And final display will as below picture:
For the same reason, optometrist can set every optometry step in the program.
If next time you need use the program, you can choose the corresponding
program from the main testing interface.
During the program setting process, after each optometry step, press [SET]
button, the machine will automatically go to next program (press
[SHIFT]+[SET], it can return to last optometry step). Setting project chart or
choose auxiliary lens can improve
Optometry efficiency
When machine CV7200 doesn’t work, please try to eliminate the troubles
according to the next chat.
40
symptom processing methods
When switch on, the phoropter Make sure the power plug is on the
do not have any response. socket
Make sure the Ophthalmic Unit is switch
on.
Console display disappear The Console enter Screen saver mode.
suddenly Touch the screen or press any button on
the control panel, it will come back to
normal display mode.
Console penal button do not It might work abnormally, pleas turn off
work the power and restart the machine.
Press the print button but without Please change the printer paper.
responce. Make sure the printer paper is installed
properly.
Touch screen click response Please do the calibration for Touch
position deviation
screen
If you can not eliminate troubles with above ways, please contact with the
distributor.
41
6.2.1Press the switch on the printer cover(as left picture) to open the printer
cover.
turn off the electric power, pull the power line, then pull out the
power outlet. At the same time, be care for the type of electric
42
7.Specifications
43