The XN-2000-L Automated Hematology Analyzer was successfully calibrated on 10/2/2025, with all adjustments and background verifications passing the acceptable ranges. Key parameters such as RBC clog, HGB zero, and aspiration sensor adjustments were all within specifications. The report certifies the analyzer's compliance with the manufacturer's recommendations.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
8 views9 pages

Popular, Savar, XN-2000 L (41954)

The XN-2000-L Automated Hematology Analyzer was successfully calibrated on 10/2/2025, with all adjustments and background verifications passing the acceptable ranges. Key parameters such as RBC clog, HGB zero, and aspiration sensor adjustments were all within specifications. The report certifies the analyzer's compliance with the manufacturer's recommendations.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 9

XN - Series Calibration Report

Customer :
Popular, Savar Date: 10/2/2025

Model: XN-2000-L Serial No: 41954

1 Adjustments

Positions Perform
Piercer Position OK
Tube Holder Position OK
Hand Position OK
Sampler Position OK
BCR Position OK

2 BACKGROUND VERIFICATION

Parameters Results Acceptable Range Status


WBC-N 0 ≤ 0.10 x 10*3/uL PASS
WBC-D 0 ≤ 0.10 x 10*3/uL PASS
WBC-P* 1 ≤ 0.10 x 10*3/uL N/A
RBC 0 ≤ 0.02 x 10*6/uL PASS
HGB 0 ≤ 0.1 g/dL PASS
PLT-I 0 ≤ 10 x 10*3/μL PASS
PLT-O* 2 ≤ 10 x 10*3/μL N/A
PLT-F* 2 ≤ 3 x 10*3/μL N/A
* 1 NOTE: Only Applicable to XN-20
* 2 NOTE: Only Applicable when corresponding license has been activated
3 RBC CLOG LEVEL ADJUSTMENT

3.1 The RBC clog adjustment is monitored after the background check. The result is as below :

Parameters Results Acceptable Range Status


RBC CLOG 100 100 ± 1 PASS
NOTE: RBC Clog adjustment IPU screenshot is attached.

4 HGB Zero Adjustment

Parameters Results Acceptable Range Status


HB Blank Gain 5026 5000 ± 200 PASS
NOTE: HGB Blank adjustment IPU screenshot is attached.
5 Blood Aspiration Sensor Adjustment

Parameters Results Acceptable Range Status


Aspiration Sensor 5098 5000 ± 200 PASS

NOTE: Aspiration sensor adjustment IPU screenshot is attached

6 Aspiration Sensor Span

6.1 QC Material :XN-CHECK


50151102
Expiry : 06-04-2025

Parameters Results Acceptable Range Status


Aspiration Sensor Span 6583 6500 ± 500 PASS
NOTE: Aspiration Sensor Span Adjustment IPU screenshot is attached.
7 Blood Sensor

Adjustments Perform
Blood Sensor OK

8 Sensitivity Adjustment

8.1 QC Material :XN-CAL


Lot : 50202101
Expiry : 23-02-2025

WNR

NOTE: WNR sensitivity adjustment IPU


screenshot is attached

Parameters Results Assay Value Acceptable Range Status


WNR-X 182.7 181.9 ± 2 ch PASS
WNR-Y 172.1 172.5 ± 2 ch PASS
WNR-SSC/Z 116.5 116.4 ± 2 ch PASS
GAIN Confirmation

Parameters Results Acceptable Range Status


WNR-X 106 82 - 210 PASS
WNR-Y 157 100 - 255 PASS
WNR-SSC/Z 249 100 - 255 PASS

8.2

WDF

NOTE: WDF sensitivity adjustment IPU


screenshot is attached

Parameters Results Assay Value Acceptable Range Status


WDF-X 168 168.6 ± 2 ch PASS
WDF-Y 96.9 97.4 ± 3 ch PASS
WDF-SSC/Z 125 124.9 ± 2 ch PASS
GAIN Confirmation

Parameters Results Acceptable Range Status


WDF-X 197 100 - 255 PASS
WDF-Y 129 74 - 194 PASS
WDF-SSC/Z 221 100 - 255 PASS

8.3

RBC/PLT

Parameters Results Assay Value Acceptable Range Status


MCV 79.1 79.3 ± 2 ch PASS
MPV 9.5 9.5 ± 1 ch PASS
NOTE: MCV and MPV sensitivity adjustment IPU screenshot is attached.
8.4
HGB

Parameters Results Assay Value Acceptable Range Status


HGB 12 11.95 ± 0.2g/dl PASS
NOTE: HGB sensitivity adjustment IPU screenshot is attached.

9 Calibration

9.1 Whole Blood Calibration Print out is attached

10 QC Verification

10.1 Radar Chart print outs for XN CHECK Level 2 for XN CHECK runs is attached
after the calibration
11 We certify that the XN-2000-L Automated Hematology Analyzer S/N: 41954 has been successfully
calibrated in accordance to the manufacturer's recommendations.

Report and Commissioning Performed By :

Signature (Application) Signature (Engineer)

Name: S. M. Shafiul Alam Name: Md. Faisal Kabir


Date: 10/2/2025 Date: 10/2/2025

Report Reviewed and Accepted By :

Signature (Customer)

Name:
Date: