0% found this document useful (0 votes)
85 views32 pages

BA DT 045 07 14 EN Dulcotest Sensor CGE 3 EN

The document provides assembly and operating instructions for the DULCOTEST® Sensor CGE, detailing safety information, user qualifications, and intended use. It includes a comprehensive table of contents covering topics such as installation, calibration, troubleshooting, and maintenance. The sensor is designed to measure total available chlorine in swimming pool water and requires careful handling and adherence to safety protocols.

Uploaded by

damianstanev
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
85 views32 pages

BA DT 045 07 14 EN Dulcotest Sensor CGE 3 EN

The document provides assembly and operating instructions for the DULCOTEST® Sensor CGE, detailing safety information, user qualifications, and intended use. It includes a comprehensive table of contents covering topics such as installation, calibration, troubleshooting, and maintenance. The sensor is designed to measure total available chlorine in swimming pool water and requires careful handling and adherence to safety protocols.

Uploaded by

damianstanev
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Assembly and operating instructions

DULCOTEST® Sensor CGE, Type:


CGE 3-mA-2 ppm; CGE 3-mA-10 ppm

EN

A1953

Please carefully read these operating instructions before use! · Do not discard!
The operator shall be liable for any damage caused by installation or operating errors!
Technical changes reserved.

984482 BA DT 045 07/14 EN


Supplemental instructions

General non-discriminatory approach


In order to make it easier to read, this
document uses the male form in grammat‐
ical structures but with an implied neutral
sense. It is aimed equally at both men and
women. We kindly ask female readers for
their understanding in this simplification of
the text.

Supplementary information
Read the following supplementary infor‐
mation in its entirety!
The following are highlighted separately in
the document:
n Enumerated lists
Instructions
ð Results of the instructions

Information

This provides important information


relating to the correct operation of the
system or is intended to make your
work easier.

Safety information
Safety information are provided with
detailed descriptions of the endangering
situation, see Ä Chapter 1.1 ‘Explanation
of the safety information’ on page 4

2
Table of contents

Table of contents
1 Introduction............................................................................................................. 4
1.1 Explanation of the safety information............................................................. 4
1.2 Users' qualifications........................................................................................ 6
1.3 General safety information ............................................................................ 7
1.4 Intended Use.................................................................................................. 8
1.5 Information in the event of an emergency...................................................... 8
2 Brief Description of the Functions........................................................................... 9
2.1 Measured variable.......................................................................................... 9
2.2 Construction of the sensor........................................................................... 10
3 Transport and Storage.......................................................................................... 11
3.1 Storage ........................................................................................................ 11
3.2 Transport...................................................................................................... 11
4 Assembly.............................................................................................................. 12
5 Installation............................................................................................................. 16
6 Putting the Sensor into Operation......................................................................... 18
6.1 Calibration.................................................................................................... 19
7 Information on Troubleshooting and Fault Elimination......................................... 21
8 Maintenance and Repair Work on the Sensor...................................................... 23
9 Decommissioning and Disposal............................................................................ 24
10 Ordering Information............................................................................................. 25
11 Technical Data...................................................................................................... 27
12 Directives / standards adhered to......................................................................... 28
13 Index..................................................................................................................... 29

3
Introduction

1 Introduction
These operating instructions provide infor‐
mation on the technical data and functions
of the DULCOTEST ® Sensor for Organi‐ DANGER!
cally Bound Chlorine, Type CGE 3. Nature and source of the danger
Consequence: Fatal or very serious
injuries.
Standard scope of supply
Measure to be taken to avoid this
n Sensor complete with sensor cap, danger
protective cap and clamp ring
Danger!
n Bottle of electrolyte
n Spare membrane cap – Denotes an immediate threat‐
n Operating instructions ening danger. If this is disre‐
garded, it will result in fatal or
n Screw driver very serious injuries.

1.1 Explanation of the safety


information
WARNING!
Introduction
Nature and source of the danger
These operating instructions provide infor‐ Possible consequence: Fatal or very
mation on the technical data and functions serious injuries.
of the product. These operating instruc‐
tions provide detailed safety information Measure to be taken to avoid this
and are provided as clear step-by-step danger
instructions.
Warning!
The safety information and notes are cate‐
– Denotes a possibly hazardous sit‐
gorised according to the following
uation. If this is disregarded, it
scheme. A number of different symbols
could result in fatal or very
are used to denote different situations.
serious injuries.
The symbols shown here serve only as
examples.

4
Introduction

CAUTION! Type of information


Nature and source of the danger Hints on use and additional informa‐
Possible consequence: Slight or tion
minor injuries, material damage. Source of the information, additional
Measure to be taken to avoid this measures
danger Information!
Caution! – Denotes hints on use and other
– Denotes a possibly hazardous sit‐ useful information. It does not
uation. If this is disregarded, it indicate a hazardous or dam‐
could result in slight or minor inju‐ aging situation.
ries. May also be used as a
warning about material damage.

NOTICE!
Nature and source of the danger
Damage to the product or its sur‐
roundings
Measure to be taken to avoid this
danger
Note!
– Denotes a possibly damaging sit‐
uation. If this is disregarded, the
product or an object in its vicinity
could be damaged.

5
Introduction

1.2 Users' qualifications

WARNING!
Danger of injury with inadequately qualified personnel!
The operator of the plant / device is responsible for ensuring that the qualifications
are fulfilled.
If inadequately qualified personnel work on the unit or loiter in the hazard zone of the
unit, this could result in dangers that could cause serious injuries and material
damage.
– All work on the unit should therefore only be conducted by qualified personnel.
– Unqualified personnel should be kept away from the hazard zone

Training Definition
Instructed personnel An instructed person is deemed to be a person who has been
instructed and, if required, trained in the tasks assigned to him/
her and possible dangers that could result from improper
behaviour, as well as having been instructed in the required
protective equipment and protective measures.
Trained user A trained user is a person who fulfils the requirements made of
an instructed person and who has also received additional
training specific to the system from ProMinent or another
authorised distribution partner.
Trained qualified per‐ A qualified employee is deemed to be a person who is able to
sonnel assess the tasks assigned to him and recognize possible haz‐
ards based on his/her training, knowledge and experience, as
well as knowledge of pertinent regulations. The assessment of
a person's technical training can also be based on several
years of work in the relevant field.

6
Introduction

Training Definition
Electrician Electricians are deemed to be people, who are able to com‐
plete work on electrical systems and recognize and avoid pos‐
sible hazards independently based on his/her technical training
and experience, as well as knowledge of pertinent standards
and regulations.
Electricians should be specifically trained for the working envi‐
ronment in which the are employed and know the relevant
standards and regulations.
Electricians must comply with the provisions of the applicable
statutory directives on accident prevention.
Customer Service Customer Service department refers to service technicians,
department who have received proven training and have been authorised
by ProMinent to work on the system.

Note for the system operator


The pertinent accident prevention regulations, as well as all other generally acknowl‐
edged safety regulations, must be adhered to!

1.3 General safety information


CAUTION!

WARNING! Functional limitations


Possible consequence: Slight or
Unauthorised access! minor injuries, material damage
Possible consequence: Fatal or very
serious injuries – Check the sensor regularly for dirt
and impurities
– Measure: Ensure that there can – Check the membrane cap regu‐
be no unauthorised access to the larly for air bubbles adhering to it
unit
– Observe all applicable national
– The sensor may only be fitted, regulations relating to mainte‐
installed, serviced and operated nance, service and calibration
by personnel trained for this intervals.

7
Introduction

1.4 Intended Use


CAUTION!
Prerequisites for operation NOTICE!
Possible consequence: Slight or
minor injuries, material damage Intended use
– Only use the sensor to determine
– The sensor may only be used in and regulate the concentration of
flow gauges that ensure the cor‐ free chlorine or chlorine bound to
rect flow parameters. cyanuric acid in swimming pool
– There must be a free outlet or at water or water of a similar quality.
most a counter pressure of 1 bar – Any other uses or modifications
at the outlet of the flow gauge. to the system are prohibited
The maximum operating pressure
– The sensor is not a safety com‐
of the respective single compo‐
ponent in the sense of DIN EN
nents must be observed.
ISO 13849-1:2008-12. If there is
– The sensor's voltage supply may a critical process in your meas‐
not be interrupted urement and control system, then
– Following longer interruptions to it is your responsibility to make
the voltage supply (> 2 h), allow sure this process is safe.
the sensor to run-in again and
recalibrate it

1.5 Information in the event of


an emergency
n In the event of an emergency, switch
off the controller
n If liquid escapes from the continuous
flow gauge, close the stopcocks on
the inlet and outlet installed by the
customer.
n Observe the plant operator's safety
information before opening the contin‐
uous flow gauge.

8
Brief Description of the Functions

2 Brief Description of the Functions


Brief description of the functions The measuring signal of the CGE3 sensor
follows the value determined by the
The DULCOTEST® CGE is a membrane- DPD-1.
covered, amperometric two-electrode
sensor. Use the DULCOTEST® CGE to
measure the concentration of total avail‐
able chlorine in swimming pool water or
water of a similar quality.
The chlorine compounds dissolved in the
water pass through the sensor membrane
and are electrochemically transformed on
the working electrode. The primary current
signal resulting from the transformation
can be evaluated to determine the con‐
centration (of chlorine).
The integral sensor signal transformer
converts the primary sensor current into
an output signal of 4 ... 20 mA.
Typical applications:
n Measurement of total available
chlorine in swimming pool water
– from stabilised chlorine (chloro
(iso) cyanuric acid derivatives),
also in combination with free
chlorine from electrolysis pro‐
cesses, with and without dia‐
phragm between the anode and
the cathode.

2.1 Measured variable


The sensor measures the concentration of
total available chlorine in swimming pool
water. This is understood as being the
total chlorine, bound to organic carrier
molecules (e.g. cyanuric acid) and the
resulting free chlorine released (HOCl and
OCl- ). The sensor measures independ‐
ently of the content of combined chlorine
(chloramines)

9
Brief Description of the Functions

2.2 Construction of the sensor

1.

2.

3.

4.

5.

6.
7.

8.
A1954

Fig. 1: Construction of the sensor


1. Opening for M12 threaded connector 5. Electrode shaft
2. Adapter 6. Hose seal
3. 2-wire connection 7. Membrane cap
4. Clamp disc 8. Protective membrane cap
Sensor type CGE3 consists of 2 main parts, the electrode shaft and the membrane cap.
The electrolyte-filled membrane cap constitutes the measuring chamber. Measuring elec‐
trodes are immersed in the measuring chamber. The measuring chamber is sealed from
the measuring medium by a micro-porous membrane. The sensor has integral tempera‐
ture compensation. The amplifier electronics are located in the upper part of the elec‐
trode shaft. The amplifier electronics are cast into a plastic mass. The amplifier elec‐
tronics transform the primary sensor current into an output signal of 4 ... 20 mA.

10
Transport and Storage

3 Transport and Storage


3.2 Transport
NOTICE! The sensor should be transported in its
Original packaging original packaging and in compliance with
the permissible environmental conditions.
Damage to the product No further special conditions have to be
– Only transport, ship and store the observed in relation to transport.
sensor in its original packaging
– Retain the packaging in its
entirety including the polystyrene
inserts

NOTICE!
Maximum storage period
Damage to the product
If the sensor is stored for a long
period of time, return it to ProMinent
for checking or servicing. Otherwise
the safe operation and measuring
accuracy of the sensor can no longer
be guaranteed

3.1 Storage
Permissible ambient temperature: +5 °C
to +50 °C
Humidity: maximum 90 % relative air
humidity, non-condensing
Other: no dust, no direct sunlight
Maximum storage period of the electro‐
lytes in their original packaging: see label
on the bottle
Maximum storage period of the sensor in
its original packaging and normal atmos‐
phere: 3 years

11
Assembly

4 Assembly
n User qualification: trained user, see Filling electrolyte
Ä Chapter 1.2 ‘Users' qualifications’
on page 6
WARNING!
Danger from hazardous substances!
Possible consequence: Fatal or very
serious injuries.
Please ensure when handling haz‐
ardous substances that you have
read the latest safety data sheets pro‐
vided by the manufacture of the haz‐
ardous substance. The actions
required are described in the safety
data sheet. Check the safety data
sheet regularly and replace, if neces‐
sary, as the hazard potential of a sub‐
stance can be re-evaluated at any
time based on new findings.
The system operator is responsible
for ensuring that these safety data
sheets are available and that they are
kept up to date, as well as for pro‐
ducing an associated hazard assess‐
ment for the workstations affected.

Do not touch the membrane cap


and electrodes at the bottom of the
electrode shaft
Do not touch or damage the mem‐
brane at the bottom of the membrane
cap and the electrodes at the bottom
of the electrode shaft or bring them
into contact with greasy substances.
Otherwise the sensor will no longer
work accurately. Replace the mem‐
brane cap with a new membrane cap
or send the sensor away to have the
electrodes cleaned.

12
Assembly

2. Press out excess air


3. Fill the membrane cap with electro‐
lyte, avoiding bubbles as far as pos‐
– The electrolyte is sensitive to oxi‐ sible
dation: Always keep electrolyte
bottle sealed after use. Do not Place the electrolyte bottle com‐
decant electrolytes into other con‐ pletely on the membrane cap and
tainers allow the electrolyte to press out
– Do not store electrolyte beyond slowly in a stream of liquid from the
its "Use by" date and note the storage bottle, while at the same
"Use by" date on the label time constantly pulling back the
storage bottle
– Store the electrolyte bottle on its
head so that the electrolyte can ð the cap is completely full if the
be easily poured out, as far as electrolyte can be seen at the
possible free of bubbles bottom thread.
– As far as is possible, fill the elec‐
trolyte so that it is free from bub‐
bles. Small air bubbles are not a
problem, larger air bubbles rise to
the upper edge of the membrane Do not use your fingers to close the
cap vent hole beneath the hose seal
– Only use the membrane cap once
4. Place the electrode shaft vertically
onto the filled membrane cap
1
5.

2 When screwing shut, allow


excess electrolyte to escape
4
unchecked through the air vent
hole underneath the hose seal
3 I

Turn the membrane cap manually


A0289 as far as it will go so that there is no
gap between the membrane cap
Fig. 2: Filling electrolyte and the electrode shaft.
I Electrolyte filling level 6. Wipe up any escaped electrolyte
1 Electrolyte bottle with a soft paper towel or similar
2 Membrane cap
3 Nozzle 7. Rinse the nozzle through thor‐
4 Vent hole oughly and then rinse off with a
clean, hot and powerful water jet,
1. Open the electrolyte bottle and so that there is no longer any
screw on the nozzle adhering electrolyte

13
Assembly

Installing the sensor in the in-line probe


housing

– Do not allow the flow to fall below


Installation instructions the minimum flow rate. Monitor
the flow of the connected con‐
– Only push in or pull out the troller. If the measured value is
sensor slowly to or from the in- used for control, switch off the
line probe housing. Otherwise the control if the flow falls below the
membrane could be damaged. minimum flow rate and/or switch
– Do not allow the membrane to to basic load
touch the in-line probe housing – Only install the sensor in in-line
flow resister probe housing type DLG III A,
– Always keep the sensor moist DLG III B or in DGM (25 mm
after commissioning, that is to module) to guarantee the neces‐
say never allow the flow gauge to sary flow requirements. No guar‐
run dry! antee can be provided if other in-
line probe housings are used
– Avoid installations that allow air
bubbles to form in the sample
water
– Air bubbles, which adhere to
the sensor's membrane, can
result in too low a measured
value and thus lead to incor‐
rect metering in a control cir‐
cuit

Observe the instructions and safety


information contained in the operating
instructions for the in-line probe
housing

1. Push the O-ring from below over


the sensor up to the clamp disc.
2. DLG III: Insert the sensor into the
DLG III and tighten with the
threaded rod

14
Assembly

3. DGM: leave a washer in the DGM.


Insert the sensor into the DGM and
tighten securely with a clamping
screw until the O-ring seals
ð The correct insertion depth of
the sensor is defined by the
clamp disc.

15
Installation

5 Installation
n User qualification: trained qualified
personnel or electrical technician, see
Ä Chapter 1.2 ‘Users' qualifications’ CAUTION!
on page 6 Incorrect metering
Possible consequence: Slight or
minor injuries. Material damage.
WARNING!
– Do not switch off the measuring
Connection to external devices system during intermittent opera‐
Possible consequence: Fatal or very tion
serious injuries – Switch on feeder assembly
with a time delay if necessary
– Ensure that the measuring
device/control device connected – Ensure that the water to be
is galvanically isolated from the measured always contains a suf‐
sensor ficient quantity of the appropriate
feed chemical
– Do not allow the supply voltage to
– Otherwise you will have to
fall below 16 V DC, even for short
factor in longer run-in periods
periods of time
– Ensure that the current
source can be loaded with a
minimum of 35 mA at a min‐
imum of 16 V DC
– Too low a supply voltage
may result in an incorrect
measured value
The interface requirements are auto‐
matically met when connecting to
ProMinent control devices.

16
Installation

Electrical Installation
1. Turn the top section of the sensor a
quarter turn anticlockwise and
remove it
2. Loosen the clamping screw on the
M12 threaded connector and feed
through the measuring line from the
control device

A0102

Fig. 3: 2-wire connection


3. Strip the cable ends, fit the cable
ends with cable end sleeves
(⌀ max = 0.5 mm2) and connect the
cable ends to the 2-wire connec‐
tion: 1 = plus, 2 = minus
4. Bring approx. 5 cm of the meas‐
uring line into the sensor.
5. Tighten the clamping screw on the
threaded connector
6. Push the top section of the sensor
right into the sensor shaft and
tighten it clockwise up to the stop

17
Putting the Sensor into Operation

6 Putting the Sensor into Operation


n User qualification: trained user, see Ä Chapter 1.2 ‘Users' qualifications’ on page 6

WARNING!
Danger from hazardous substances!
Possible consequence: Fatal or very serious injuries.
Please ensure when handling hazardous substances that you have read the latest
safety data sheets provided by the manufacture of the hazardous substance. The
actions required are described in the safety data sheet. Check the safety data sheet
regularly and replace, if necessary, as the hazard potential of a substance can be re-
evaluated at any time based on new findings.
The system operator is responsible for ensuring that these safety data sheets are
available and that they are kept up to date, as well as for producing an associated
hazard assessment for the workstations affected.

CAUTION!
Incorrect metering due to sensor failure
Possible consequence: Slight or minor injuries. Material damage.
– If a sensor fails then there may be an incorrect measured value at the input of
the controller/measuring instrument
– This may result in uncontrolled metering
– The operator should therefore ensure that no subsequent damage results from
this

18
Putting the Sensor into Operation

CAUTION!
Incorrect metering due to premature ageing of the sensor
Possible consequence: Slight or minor injuries. Material damage.
Measure: Do not electrically disconnect the sensor during measuring breaks.
Exception: If there is a break between measurements lasting more than a week and
the proportion of disinfectant in the sample water falls during this period to 0 ppm
then electrically disconnect the sensor.
– An extra run-in period will be necessary following operation without disinfectant.
Switch on feeder assembly with a time delay if necessary.

Run-in period
The sensor requires a specific run-in period to display a steady display value.
Initial commissioning: 1 - 24 h (typically 6 h)*
Recommissioning: 1 - 24 h (typically 3 h)*
Electrolyte or membrane change: 3h
* the exact run-in time depends on the application.

6.1 Calibration

CAUTION!
– Perform a slope calibration every
time if the sensor has been tam‐
pered with (e.g. electrolyte
replacement etc.)!
– Repeat slope calibration at reg‐
ular intervals to ensure the per‐
fect operation of the sensor! If
there are no other regulations,
then only calibrate the sensor
every 3-4 weeks for use in
drinking water and swimming
pool water.

19
Putting the Sensor into Operation

Zero point calibration


– Avoid air bubbles in the sample
water! Air bubbles, which adhere If the sensor is operated with a ProMinent
to the sensor membrane, can controller, then zero point calibration is not
result in too low a measured generally necessary. Zero point calibration
value and thus lead to dangerous is recommended if you wish to use the
incorrect metering. sensor at the lower threshold of the meas‐
– Please note the pertinent national uring range.
guidelines for calibration inter‐ 1. Place the sensor in a vessel with
vals! clean water free of chlorine and oxi‐
dising agent (e.g. mineral water
without gas).
2. Use the sensor to stir it until the
Prerequisites measured value on the controller
– The sensor is ready for meas‐ remains stable for 5 minutes.
uring (wait for run-in period) 3. Calibrate the controller to zero in
– Constant flow rate at the in-line accordance with its operating
probe housing instructions.
– Constant temperature of the
sample water 4. Refit the sensor into the in-line
probe housing (DGM; DLG III).
– Each time the sensor is removed
or inserted, wait until the run-in
period has expired and a con‐
stant measured value is ach‐ Slope calibration
ieved, a minimum of 15 minutes 1. Use an appropriate reference
to avoid drifts caused by temper‐ method (e.g. DPD 1) to determine
ature compensation. the chlorine content of the sample
– No concentration fluctuations of water.
the feed chemical in the sample
water 2. Set the determined value on the
– Constant pH value within the per‐ controller in line with its operating
missible range instructions.
– Ensure that the sample is taken ð Repeat calibration during com‐
at the place the sensor is missioning, and on the day fol‐
installed lowing any tampering with the
sensor.

20
Information on Troubleshooting and Fault Elimination

7 Information on Troubleshooting and Fault Elimination


User qualification: instructed user, see Ä Chapter 1.2 ‘Users' qualifications’ on page 6

Error Possible cause Remedy


Sensor cannot be cali‐ Run-in period too short Observe the run-in period
brated and measured value
greater than DPD measure‐ Membrane cap damaged Replace membrane cap
ment
Troublesome substances in Check water for trouble‐
the water some substances and
remedy this
Short circuit in the meas‐ Identify short circuit and
uring line eliminate the cause
Distance between mem‐ Screw the membrane cap
brane/electrode is too great up to its stop
DPD chemicals out of date User new chemicals,
repeat calibration
pH value < pH 5.5 Increase pH value (pH
5.5 ... 9.5)
Sensor cannot be cali‐ Run-in period too short Observe the run-in period
brated and sensor meas‐
ured value less than DPD Coating on the membrane Replace membrane cap,
measurement cap allow sensor to run in and
calibrate
Sample water flow rate too Correct flow rate
low
Air bubbles on the outside Increase the flow within the
of the membrane permitted range
pH value > pH 9.5 Lower pH value (pH 5.5 ...
9.5)
No electrolyte in the mem‐ Fill with new electrolyte
brane cap
Electrolyte pushed out by Contact ProMinent
gas bubbles in the sample
water

21
Information on Troubleshooting and Fault Elimination

Error Possible cause Remedy


Measured value display is Chlorine content below the Add chlorine and then
‘0’ lower limit of the measuring repeat calibration and/or
range use a suitable sensor
Measuring line broken Replace measuring line
Sensor connected to the Connect the sensor cor‐
controller with incorrect rectly to the controller
polarity
Run-in period too short Observe the run-in period
Sensor defective Send sensor in for regener‐
ation
Sensor measured value is Membrane damaged Replace membrane cap;
unstable allow sensor to run in and
calibrate
Air bubbles on the outside Remove air bubbles by tap‐
of the membrane ping them and increase
flow rate if necessary
Reason lies with the con‐ Eliminate cause
troller

When you have tried everything: Check whether the reference electrode at the end of the
electrode shaft is silvery-white instead of brownish-grey. The reference electrode is then
worn out and needs to be replaced by ProMinent Dosiertechnik GmbH.

22
Maintenance and Repair Work on the Sensor

8 Maintenance and Repair Work on the Sensor


User qualification: instructed user, see Cleaning the membrane
Ä Chapter 1.2 ‘Users' qualifications’
on page 6 If the membrane is dirty and the sensor
cannot be calibrated, try to clean the
Maintenance interval membrane carefully. Firstly dismantle the
sensor. Gently remove any adhesive dirt:
1. Rinse the membrane under a gentle
CAUTION! stream of cold water
– Maintain the sensor regularly in 2. Remove deposits (lime scale, rust):
order to avoid over-metering in
the event of a sensor failure n Dismantle the membrane cap
– Please note the applicable n Rinse the sensor with plenty of
national guidelines for mainte‐ water
nance intervals
ð Now fill the sensor with electro‐
– Do not touch the electrodes or lyte, allow it to run in and recali‐
bring them into contact with sub‐ brate.
stances containing grease
– Never clean the membrane with
alkaline or acidic solutions, Replacing the mem‐
cleaning reagents or by mechan‐
ical means (brush or similar) brane
If calibration is no longer
possible even after the
Weekly to monthly, depending on the membrane has been
water quality cleaned or if the membrane
Maintenance work
is damaged, then you need
to replace the membrane
Use a suitable measuring instru‐ cap.
ment (e.g. DPD) to check the meas‐
ured value displayed on the sen‐
sor's controller.
ð If necessary, recalibrate the Repairing the sensor
sensor
The sensor can only be repaired in the
factory. Return it to us in its original pack‐
aging.

23
Decommissioning and Disposal

9 Decommissioning and Disposal


n User qualification: instructed user,
see Ä Chapter 1.2 ‘Users' qualifica‐
tions’ on page 6 NOTICE!
Regulations governing disposal of
used parts
NOTICE! – Note the current national regula‐
Decommissioning the sensor tions and legal standards which
apply in your country
Observe all of the safety information

Dispose of the electrolyte in accordance


1. Disconnect the sensor from the with the electrolyte's safety data sheet
power source
ProMinent GmbH, Heidelberg will take
2. Depressurise the in-line probe back decontaminated used devices pro‐
housing viding that they are covered by adequate
3. Loosen the clamping screw postage.

4. Remove the sensor slowly from the


in-line probe housing
5. Screw open and empty the mem‐
brane cap over a sink or similar
vessel
6. Flush the membrane and electrode
with clean water and dry until free
of dust
7. Loosely screw on the membrane
cap to protect the electrodes
8. Replace the membrane protection
cap to protect the membrane cap

n User qualification: instructed user,


see Ä Chapter 1.2 ‘Users' qualifica‐
tions’ on page 6

24
Ordering Information

10 Ordering Information
Standard scope of supply

Sensor CGE3
Bottle with electrolyte (50 ml)
Nozzle for bottle
Spare membrane cap
Operating instructions
Screw driver

Complete set. Only a complete set of sensors can be ordered:

Part Order number


CGE 3-mA-10 ppm 1047975
CGE 3-mA-2 ppm 1047959

Spare parts and accessories

Part Order number


Set 740048
n 2 membrane caps
– CGE 2-mA-2/10 ppm
n Bottle of electrolyte (50 ml)
Set 792862
n 1 membrane cap
– CGE 2-mA-2/10 ppm
n Bottle of electrolyte (50 ml)
Bottle of electrolyte (50 ml) 792892
Mounting kit for DGM 791818

25
Ordering Information

Part Order number


Mounting kit for DLG III 815079

Two-wire measuring line, mA version (2 x 0.24 mm2, Ø 4 mm) 725122

Photometer DT 1B 1039315

26
Technical Data

11 Technical Data
n Measured variable n Conductivity of sample water
– Free chlorine and organically 30 μS/cm ... 10,000 μS/cm
bound chlorine n Service life of membrane cap
n Area of application – typically 1 year, depending on the
– Drinking water, swimming pool water quality
water and water of a similar n Material
quality
– Membrane cap, PPE
n Measuring ranges
– Electrode holder, PMMA
– CGE 3-mA-2 ppm;
0.02 ... 2.0 mg/l – PVC electrode shaft
– CGE 3-mA-10 ppm; – Electronics, electronic compo‐
0.1 ... 10 mg/l nents
n pH-range n Supply voltage
– 5.5 ... 9.5 – 16 ... 24 V DC
n Temperature range – at least 35 mA at 16 V DC
– 5 ... 45 °C (temperature-compen‐ n Output signal
sated) – 4 ... 20 mA (mA version)
– No sudden changes in tempera‐ n Degree of protection
ture – IP 65
n Storage temperature
– 5 ... 50 °C
n Resolution
– corresponds to the lower limit of
the measuring range
n Maximum operating pressure
– DGM: 3.0 bar (free flow), no neg‐
ative pressure
– DLG III: 1.0 bar (free flow), no
negative pressure
n Flow
– In-line probe housing DLG III A/
B / DGM
– Optimum: 30 l/h
– Minimum: 20 l/h
– Maximum: 100 l/h
n Cross-sensitivity
– Bromine, iodine, ozone and CIO2,
chloramines and other oxidants
lead to measuring errors

27
Directives / standards adhered to

12 Directives / standards adhered to


EC directives:
n EC EMC Directive (2004/108/EC)
International standards:
n EN 61010-1
n EN 60335-1
n EN 60529
n EN 61326-1
For sensors with a CAN interface, the fol‐
lowing specifications also apply:
n CANopen CiA DS 301
n CANopen CiA DSP 305
n CANopen CiA DS 404

You can find the EC Declaration of Con‐


formity as a download under
http://www.prominent.de/Service/
Download-Service.aspx

28
Index

13 Index
A Question: How is the sensor con‐
Accessories . . . . . . . . . . . . . . . . . 26 structed? . . . . . . . . . . . . . . . . . . . . 10
Air bubbles . . . . . . . . . . . . . . . . . . 20 Question: What are the basic
functions of the sensor? . . . . . . . . . 9
Question: What does the sensor
C measure? . . . . . . . . . . . . . . . . . . . . 9
Complete set . . . . . . . . . . . . . . . . . 25 Question: What maintenance,
Cross-sensitivity . . . . . . . . . . . . . . 27 cleaning and repair work can I
carry out on the sensor? . . . . . . . . 23
D Question: What standards were
Declaration of Conformity . . . . . . . 28 applied and what are the technical
Degree of protection . . . . . . . . . . . 27 data? . . . . . . . . . . . . . . . . . . . . . . . 28
Disposal . . . . . . . . . . . . . . . . . . . . 24 Question: Where do I find the
Disposal of used parts . . . . . . . . . 24 Declaration of Conformity? . . . . . . 28
Question: Which standards are
E complied with? . . . . . . . . . . . . . . . 28
EC directives . . . . . . . . . . . . . . . . . 28 Question: Why isn't the sensor a
safety component? . . . . . . . . . . . . . 8
G
General non-discriminatory R
approach . . . . . . . . . . . . . . . . . . . . . 2 Resolution . . . . . . . . . . . . . . . . . . . 27
Run-in period . . . . . . . . . . . . . . . . 19
I
International standards . . . . . . . . . 28 S
Safety information . . . . . . . . . . . . . . 4
M Sensor failure . . . . . . . . . . . . . . . . 19
Slope calibration . . . . . . . . . . . . . . 20
Measured variable . . . . . . . . . . . . 27 Spare parts . . . . . . . . . . . . . . . . . 26
Mounting kit . . . . . . . . . . . . . . . . . 26 Standard scope of supply . . . . . . . 25
Standard slope . . . . . . . . . . . . . . . 27
N Standards complied with . . . . . . . . 28
non-discriminatory approach . . . . . . 2 Storage period . . . . . . . . . . . . . . . 11
O U
Original packaging . . . . . . . . . . . . 11 Users' qualifications . . . . . . . . . . . . 6
Output signal . . . . . . . . . . . . . . . . . 27
Z
Q Zero point calibration . . . . . . . . . . 20
Question: How do I locate and
remedy a fault? . . . . . . . . . . . . . . . 21
Question: How do I put the sensor
into operation? . . . . . . . . . . . . . . . 18

29
30
31
ProMinent GmbH
Im Schuhmachergewann 5 - 11
69123 Heidelberg
Telephone: +49 6221 842-0
Fax: +49 6221 842-419
email: info@prominent.com
Internet: www.prominent.com
The latest version of the operating instructions is available on our homepage.

984482, 1, en_GB

© 2014

You might also like