GEMINI - Assay Programming Manual - Rev 6
GEMINI - Assay Programming Manual - Rev 6
GEMINI COMBO
Compact Microplate Processor
Original Document
STRATEC Biomedical AG
Gewerbestraße 37
75217 Birkenfeld, GERMANY
Neither this manual nor any parts of it may be duplicated or transmitted in any way without the written
approval of STRATEC Biomedical AG.
Table of Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 Typographical Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1.1 Display of Warnings and Notes. . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1.2 Used Warning Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.1.3 Special Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.2 Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.2.1 General Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.2.2 Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.2.3 Laser Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.2.4 Mechanical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1.2.5 Biological Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
1 Introduction
Target of this manual is the explanation of the Gemini and Gemini COMBO
instrument, respectively. After having read the manual, the user should be able to
safely operate the Gemini/Gemini COMBO instrument.
The warnings, notes and symbols described hereafter are used in the current
manual, on the instrument and on its packaging.
DANGER
Danger indicates a hazardous situation that, if not avoided will result in death or
serious injury.
WARNING
Warning indicates a hazardous situation that, if not avoided, could result in death or
serious injury.
CAUTION
Caution indicates a hazardous situation that, if not avoided, could result in minor or
moderate injury.
NOTICE
Notice indicates information considered important, but not hazard-related (e.g.
messages related to property damage). The non-observance of a safety instruction
can result in damage of the instrument or an adverse effect on the instrument
function.
INFO
The non-observance of information can result in an adverse effect on the instrument
function (result deterioration).
Biohazard!
Electrical hazard!
Laser hazard!
Mechanical hazard!
Automatic start-up!
LEDs and LEDs (light emitting diode) and signal lamps are printed in special type.
Signal Lamps Example: Power LED
Menu items and Menu items and buttons are printed in spaced type.
Buttons Example: Open button.
The following safety instructions shall be observed at all times, both before and
during operation and during maintenance.
The Gemini instrument is designed and manufactured in accordance with the safety
requirements for electronic and medical systems. If the law issues regulations
concerning the installation and/or operation of the instrument, then it is the operator's
responsibility to adhere to them.
The manufacturer has done everything possible to guarantee that the equipment
functions safely, both electrically and mechanically. The instruments are tested by the
manufacturer and supplied in a condition that allows safe and reliable operation.
Laboratory equipment
The instrument has been designed and developed as laboratory equipment in
accordance to the requirements of the EC directive 98/79/EC (IVD directive, directive
98/79/EC of the European Parliament and of the Council of 27 October 1998 on in
vitro diagnostic medical devices). In order to assure compliance, applicable
standards recorded in the list of standards harmonized for the IVD directive were
observed. The application of this product for in vitro diagnostics purposes requires a
separate conformity assessment according to EC directive 98/79/EC for the
complete system into which it will be incorporated and/or has to be used in
combination with (e.g. reagent).
Defective instrument
Any defective instrument will result in serious injuries with deadly consequences and
material damage (e.g. fire).
• Immediately disconnect the defective instrument from the mains supply, if a
safe usage is no longer possible.
• Secure the defective instrument against reconnection.
• Label the defective instrument clearly as being defective.
Overheating
Improper placing of the instrument may cause fire or serious instrument damage in
case of overheating.
• Do not block or cover ventilation slots.
• The air shall be able to circulate.
Risk of infection!
The instrument shall be treated as potentially infectious. Improper handling of
infectious parts will cause skin irritations, illnesses and possible death.
• Strictly follow the local and national provisions, legislation and laboratory
regulations.
• Use appropriate gloves!
• Use an appropriate lab coat!
• Use an appropriate eye protection (e.g. protective glasses)!
• Avoid contact between skin/mucous membrane and samples/test reagents or
parts of the instrument.
• Clean, disinfect and decontaminate the instrument immediately if potentially
infectious material has been spilled.
• Do not use broken or chipped tubes or bottles.
• Observe the instructions in the package inserts for correct use of reagents.
• Observe the legal regulations for the handling of infectious material.
• Never use bio-hazardous liquids for testing the instrument!
• The instrument shall be cleaned, disinfected and decontaminated before
servicing!
2 Assay Programming
The Gemini system is an open system. It allows users to process pre-defined assays
but also to edit existing assays or to create their own assays.
Wrong Results
New, modified or assumed assays must be validated to avoid wrong results.
Defining assays with the Gemini instrument software is facilitated by visual dialog
boxes (no actual programming is needed) but because the instrument is very flexible
and because it allows users to define not only the processing steps but also the result
evaluation steps, defining assays remains a process that must be done with
adequate care and reference to the procedures described in this manual.
This introduction covers the basic concepts of assay files (layout, structure,
protection). The second part is a step-by-step description of how to create completely
new assays.
1 Assay tree
2 Separating bar (can be shifted)
3 Display area
The display area shows the plate layout with 96 wells and above some default
settings. The assay explorer tree in the left-hand side of this window shows the
basic steps of an assay. You can work with it in the same manner as you work with
Show Assay Click with the left mouse button on an assay folder to view the respective assay step
Step with the description of the already defined functions in the display area.
Edit Assay Step To edit an assay step, you always have to select the assay step to be edited.
Insert Assay To insert a new assay step, you always have to select the assay step in front of which
Step the new step is to be inserted.
Context Menu The context menu for inserting new assay steps in a new assay can be selected via
for Inserting the menu item Edit on the menu bar, as well as by clicking the right mouse button on
Assay Steps the step before which the new step is to be inserted.
Undo Revokes the last action you have executed in the assay
tree.
Redo Revokes the last U n d o action again.
Cut Cuts out a highlighted assay step and saves it to the
clipboard.
Copy Copies a highlighted assay step to the clipboard.
Paste Inserts an assay step on the clipboard in front of the
highlighted assay step. Keep in mind, however, that the
given assay structure must not be changed: pipetting
steps must always be inserted before Reading,
evaluations steps always after Reading!
Edit To edit an assay step, you always have to select the
assay step to be edited.
Insert Step To insert a new assay step, you always have to select
the assay in front of which the new step is to be inserted.
Delete Deletes a step highlighted in the assay tree.
Select All Selects all assay steps.
Move Up Moves a highlighted assay step up one step.
Move Down Moves a highlighted assay down one step.
Execute Current Execute only the highlighted step of the selected assay
Step... (see notes below).
Check Results
When you are performing the execute functions it is necessary to check the results
carefully!
When you are performing the execute functions during a running worklist, the plate
will be scheduled into this running worklist!
General buttons
which apply on
OK Saves all entries and returns to the main menu.
most dialog
boxes Apply Saves the entries and the dialog box remains displayed.
Canc el Returns to the main menu without saving any entries.
Data that have already been saved by pushing Ap p l y
remain stored.
Help Calls on-line help.
1. Enter the first preparation name in the input field and click on the A d d
button.
The entered name is shown in the list on the right side. Later, the preparation
names are shown in the plate layout.
2. If wished, choose a color for the preparation in the color list below the names
list. The colors are used in the L o a d dialog.
Later the user must load the preparation sample tubes in the same order as the
preparation names have been entered! All preparation sample tubes are assigned to
the sample ID of the first preparation sample tube!
4 Preliminary file name. After saving there is shown the file name of
the assay.
5 Assay steps
Assay Header Double-click on this step to open the respective parameter window (see chapter 2.2 on
page 2-12).
Assay Layout Double-click on this step to open the respective parameter window (see chapter 2.3 on
page 2-18).
Definition of Click with the right mouse button on R e a d to open the I n s e r t S t e p dialog box
possible Assay showing the available steps (see chapter 2.5 on page 2-26).
Process Steps
Figure 2-5: Assay tree with open Read folder - basic setting
Reader Settings Double-click on this step to open the respective parameter window and you can
define the photometer settings (wavelength, etc.) (see chapter 2.6.1 on page 2-64).
Background Double-click on this step to open the respective parameter window (see chapter 2.6.2
Subtraction on page 2-70).
Report Settings Double-click on this step to open the respective parameter window (see chapter 2.6.3
on page 2-73).
Saving the As soon as you have defined all steps, select the item F i l e > S a v e A s and save
Assay File the assay under a new name (extension *.asy).
Assay Header *
The assay header includes general information such as: assay name, comments (if any),
password protection, written by, creation date, last edited, plate type used, maximum ambient
temperature.
Assay Layout *
Table representation of a test plate showing the respective positions of the samples, standards
and controls for that assay.
Assay steps before Read (Processing steps)
Pipette Inserts a pipetting step (for samples, controls and
standards).
Wash Inserts a washing step.
Incubate Inserts an incubation step.
Dispense Inserts a dispense step (for reagents).
Verify dispense Inserts the verify dispense step (colorimetric verification
for colored reagents).
Background read Inserts a background reading. The baseline is stored and
later subtracted from the measured plate.
Conditional delay Inserts an incubation time; the length can be set relative to
an event on the plate.
Shake Shake on the plate transport unit.
Incubate Verification Incubator performance verification.
Clean Washer Washer manifold maintenance.
Read/Evaluation Settings
Reader settings * Definition of the colorimeter settings.
Background subtraction * Definition of the background calculation method.
Validation criteria Definition of validation criteria.
Qualitative settings Qualitative evaluation.
Quantitative settings Quantitative evaluation.
Spreadsheet Spreadsheet for user-defined calculation operations.
Report Settings *
Definition of the contents of the result printout, of the combined report and of the format and
contents of the result export file.
End of Protocol *
For users who create their own assays, it is strongly recommended that they use the
assay password protection (once they have finished defining an assay). This will
ensure that no accidental change can be made by operators who use the assays for
test runs.
Password Option to enter a password. This rules out that the assay cannot be modified by
anyone who does not know the password. However, no password is required to run
the assay.
List of plate types Selection box listing the plate types pre-defined for the
instrument.
Name The name of plate type highlighted in the menu is
displayed. Click on Add to define a new plate type and
enter a new name. This will create a new plate type
which has to be combined with a *.mpc file, i.e. you have
to choose a *.mpc file in the C o o r d i n a t e s drop-
down menu.
Coordinates Drop-down menu showing the available *.mpc files.
These files contain the plate coordinates and are
created using the teacher software.
Rows / C o l u m n s Select the number of rows and columns of the plate
type.
X Offset Possibility to change pipette position from centre of well.
X offset of 0 is recommended.
Add Click this button to create a new plate type. The
previous entry in the N a m e box is deleted. The next
running number is entered as default name, which can
be overwritten. The generated plate type is entered in
the menu to the left and it has to be combined with an
*.mpc file.
Delete Deletes the plate type highlighted in the selection box.
Close Closes the P l a t e T y p e Ed i t o r dialog. Changes
are saved. The program returns to the A s s ay
P r o t o c o l H e a d e r dialog.
If you use the multi-preparation function, it is not possible to use replicates! See also
Multi-preparation Assays (see chapter 2.1.1.1 on page 2-5).
Only standards can be taken into account in the quantitative analysis. But in some kit
inserts, standard O is deemed to be a "negative control". In this case, users should
redefine this "control" as a standard in order to maximize the number of data used to
build the signal/concentration curve.
Individual wells:
1. Select well type in the menu and define the respective parameters.
2. Click consecutively on the wells to be filled with this well type. The label is
entered with consecutive numbering.
Well ranges:
1. Select well type in the menu and define the respective parameters.
2. With the mouse pointer, highlight the desired area: Highlight a row from left to
right or a column from top to bottom or a rectangular range from the top left
corner to the bottom right corner. The well type is entered in the highlighted
range with a consecutive number.
If you use the multi-preparation function, then you must select at least as many wells
as you have different preparations (see chapter 2.1.1.1 on page 2-5).
New Label
The new label has not the same functions as the original label!
All assay steps from all assays of the worklist programmed before the schedule
break will be performed in advance.
The S c h e d u l i n g b r e a k step can be changed by the M o v e u p and M o v e
d o wn functions in the Context Menu for Inserting Assay Steps.
Example:
The assay has at first the pipette step, followed by the schedule break and a
dispense step.
The scheduling shows, that at first the sample pipetting of both plates will be
performed; after this step, the two plates are scheduled according the remaining
assay steps.
After selection of a step the respective dialog box is displayed for parameter entry.
Once you have entered your parameters and saved your entries, the new step is
inserted in the assay tree.
Order of Assay You can arrange the individual steps in any order. The order is defined during new
Steps definition of a step by clicking with the right mouse button on the folder before the
one you the want to place the new step.
But when defining your steps, you can freely add them in any order to start with
(some are easier to define than others) and then rearrange them in the correct order
of the assay with the M o v e U p / M o v e D o w n items of the E d i t menu (or of
the context menu).
Editing Assay Double-click on the respective symbol in the assay tree to open the respective
Steps parameter window.
Replicate steps If your assay includes steps that are repeated twice identically or almost identically
(this is often the case for W a s h steps or I n c u b a t i o n steps), you can speed up
the definition process by using the C o p y / P a s t e function of the E d i t menu (or
of the context menu). Edit the parameters if necessary.
The P i p e t t e dialog is also used to define predilution steps. In this case, it is used
not only to define the aspiration and dispense of samples and controls but also of
reagents (e.g. diluents) into the dilution plates (or tubes).
Here you define the wells to pipette into and the pipetting steps to be performed.
Several sub-steps can be performed in one pipetting step (step1, step2...).
Plate layout The plate layout defined in assay layout definition (see
chapter 2.3.2 on page 2-20) is displayed. Default setting is
that all wells have a checkmark. A checkmark in the
Plate Layout means that pipetting is performed into the
respective well. Click on a well to remove the respective
checkmark (= deselect the pipetting function); click
again to add the checkmark again.
S t e p s list Menu displaying the pipetting steps that have been
defined. Default setting if no step has yet been defined
is St ep 1 . You can edit the step name in the E d i t
O p e r a t i o n s dialog.
New Step Adds a new pipetting step to the S t e p s menu (default
is Step1, Step2...).
Delete Step Deletes the highlighted pipetting step in the S t e p s
menu. The button is not available, when only one step
exists because this step cannot be deleted.
Incompatible functions
Note that the function E l i m i n a t e a s s a y d r i f t c a u s e d b y t h i s
o p e r a t i o n is strictly incompatible with the worklist option m u l t i p i p e t t i n g (see
"Instructions for use Manual")!
Proceed as follows:
The plate layout defined in assay layout definition (see chapter 2.3 on page 2-18) is
displayed in this window. Default setting is that all wells have a check mark and step
1 (1st pipetting step) is displayed in the S t e p s menu. A check mark in the plate
layout means that pipetting is performed into the respective well. Click on a well to
remove the respective check mark (= deselect the pipetting function); click again to
add the check mark again.
Now you can define the various pipetting steps in accordance with your assay:
1. Deselect wells for S t e p 1 which should not be pipetted into.
2. Click on E d i t O p e r a t i o n s . . . and define the individual pipette functions
and the respective parameters for S t e p 1 in the Ed i t O p e r a t i o n s
dialog (aspirate, dispense). You can change the name St ep 1 as you wish
(see chapter 2.5.1.1 on page 2-31). The default setting W a s h always has to be
enabled to ensure that the pipette tip is replaced after each pipetting step.
3. Click O K to confirm the entries in the E d i t O p e r a t i o n s dialog. The
program returns to the P i p e t t e dialog.
4. Click on N e w S t e p to define a second pipetting step. In the St ep s
menu, S t e p 2 is displayed and highlighted. In the plate layout, all wells
have a check mark. Define the wells for this step to which the pipetting
functions will apply; then click on E d i t O p e r a t i o n s . . . to define the
individual pipetting functions in the E d i t O p e r a t i o n s dialog.
5. Further steps can be defined.
6. If you highlight a step in the S t e p s menu, the respective plate layout (the
wells with defined pipetting functions and in the E d i t O p e r a t i o n s dialog
the defined pipetting functions) is displayed automatically. You can view the
respective parameters in the respective dialog boxes and edit them, if
necessary.
Click on one of the buttons to open the respective dialog and enter the respective
parameters. If you confirm the parameters with O K , the step is entered in the
Op e r a t i o n s menu before the highlighted position. The steps W a s h / T i p
eject and < E n d > are always defaulted and must remain in the menu. Double-
click on a row in the menu to edit the values of an already defined step. The
respective parameter window opens.
Wrong results/spillage
It is necessary to validate pipetting profiles with affected assays to avoid wrong
results or spillage.
Wrong results
All assays using multishots must be validated for their respective number!
Incompatible Functions
Do not confuse this Dispense sub-step, which is defined (within the Pipette step) for
samples and controls (or reagents/diluents when they are dispensed into a dilution
plate), with the actual Dispense step which is used for reagents dispensed into test
plates (see chapter 2.5.4 on page 2-46).
Wrong results/spillage
It is necessary to validate pipetting profiles with affected assays to avoid wrong
results or spillage.
Incompatible functions
Do not use the B a c k t o s o u r c e function together with multiaspirate step
(several aspirate steps)!
Incompatible functions
Note that this function is strictly incompatible with the worklist option m u l t i
p i p e t t i n g (see "Instructions for use Manual")!
If assays are combined on one plate, where only one assay has the assay drift for
pipette/wash step enabled, the software applies the drift compensation on all assays
on the plate!
Because pipetting samples (or dispensing reagents) into a whole plate can be fairly
long, the time span between the moment when the first strip and the last strip of a
plate is pipetted (dispensed) can be significant. But then, when these strips get
washed, they get washed almost simultaneously (as the washer operates much
faster than the pipettor). In other words, the waiting time between pipetting
(dispensing) and washing is much longer for the wells of the first strip than for those
on the last strip.
In some assays, this can create problems (inconsistent results from one strip to the
other).
The purpose of the E l i m i n a t e a s s a y d r i f t caused by this operation item
available in the P i p e t t e and in the D i s p e n s e dialog is to correct this.
If you check this item in the P i p e t t e dialog, the system delays the first wash cycle
of each strip in such a way that the time span between pipetting (dispensing) and
washing is the same for each strip (Time needed to pipette all the samples / controls
= Time used to perform the first wash cycle of the following wash step).
If you check this item in the D i s p e n s e dialog, the system monitors the time
needed to dispense the reagent into the whole plate and delays the first wash cycle
accordingly (Time needed to dispense the reagent = Time used to perform the first
wash cycle of the following wash step).
Because the E l i m i n a t e a s s a y d r i f t function results in an increase (for all
strips except the first one) of the waiting time span between pipetting (dispensing)
and washing, the duration of the plates room-temperature incubation between those
two steps has to be reduced accordingly. This reduction would be different for each
strip, and therefore be difficult to define.
Instead, you just need to check the incubation time from the start of the previous
assay step item in the I n c u b a t e dialog (see chapter 2.5.3 on page 2-44). The system
then calculates the incubation time from the start of the pipetting (dispensing) step.
That way, all strips will have an identical incubation time.
This feature works only for room-temperature incubation.
Example:
In case 1, with drift elimination, you can see that each strip gets a precise 30 min.
waiting time (room-temperature incubation) between pipetting and its first wash
cycle. In case 2, without drift elimination, only the last strip gets an adequate waiting
time of 30 min., all the other strips get an extended waiting time.
Wash Buffer To select the wash buffer, click the arrow button to open the drop-down list. If the
desired buffer is not included in the list you may define it by clicking the B u f f e r s
button (see chapter 2.5.11.1 on page 2-59).
Strips Select the strips to be washed. Each box corresponds to a strip. The strips marked
by a check mark are washed. To remove/add a check mark, click on the respective
strip number.
Under default settings, all strips are checked. If, for example, you use plates with
removable strips without including all 12 strips, make sure that you uncheck the
boxes corresponding to the strips not included. On defining different wash
parameters on the same plate or on defining a strip-wise washing mode, see below.
It is useful to mark all strips. The system does not wash unused strips.
Heights Automatic or freely definable height settings for the washer head. In case the
automatic height dialog is enabled, the heights defined in the S y s t e m S e t u p . . .
/ W a s h e r will be used. In this case, the respective edit box is disabled. To freely
define the height setting, uncheck the box and use the edit box on the right to specify
a value. You can visualize the respective height setting through the window on the
front panel of the instrument.
Top wash height The wash head is on the level of the upper edge of the
well, i.e. higher than during dispensing. 0 = highest
position of dispensing needle.
Bottom wash The wash head is directly above the well bottom =
height height for executing the bottom wash function, i.e.
intense washing of the wells.
Aspirate height Height of the wash head during aspiration.
Wash Mode
List Use the drop-down list to select the wash mode:
• Plate wash mode
In a plate-wise washing mode, the washer performs
the first wash cycle on the whole plate, then the
second wash cycle on the whole plate, etc. until the
complete wash step is finished.
• Strip wash mode
In a strip-wise washing mode, the washer performs
all the wash cycles included in the wash step on the
first strip, then all the wash cycles on the second
strip, etc.
Purge The washer's Purge and Clean sequences are normally defined at system level (see
"Instructions for use Manual" - "System set-up - Washer tab") and are identical for all
assays. The Purge and Clean fields in this dialog allow you to define assay-specific
Purge and Clean settings, for example for assays using special reagents that require
a more thorough Purge or Clean function or a different clean fluid.
If purge volume and cleaning parameters are changed, the default values are
overridden.
Dispense Parameters of all the wash cycles before last. These parameters include aspirate
and dispense parameters.
Aspirate Aspiration step after the last wash step to empty the plate.
Assay Drift If you select S t r i p w a s h m o d e , this means that each strip is washed at a
slightly different time than the others. This "waiting time", for example between the
moments the wash step is completed on the first strip and the moment the wash step
is completed on the last strip is equivalent to a longer room-temperature incubation
(for the first strip). In some assays, this "drift" effect can have adverse consequences
on the results.
Using strip wash mode for long wash cycles can lead to assay drift and/or well
dehydration. For such assays use P l a t e wa s h m o d e or skip final aspiration
during strip wash mode and insert second wash step for final aspiration to avoid wells
fall dry.
Well Generally, a wash step includes one or more wash cycles followed by a final
dehydration aspiration of the liquid from the plate wells.
In Pl a t e w a s h m o d e , the final aspiration is performed almost simultaneously
on all strips at the end of the complete wash step.
In S t r i p w a s h m o d e , the complete wash step, including the final aspiration, is
performed individually on each strip. This means that the first strip will be aspirated
a relatively long time before the last strips on the plate (the exact time difference
depending on the duration of the wash step and the number of strips used). The
result of this is a risk of dehydration of the wells of the first strips. To avoid this, you
can, if the assay absolutely requires a S t r i p w a s h m o d e , program two
consecutive wash steps instead of one: the first one in Strip mode without aspiration
and the second one in Plate mode with only the final aspiration.
If you have any doubt about when and how to use the Strip wash mode option,
please call your application engineer for assistance.
Incompatible functions
Avoid combining assays with Plate wash mode and assays with Strip wash mode
on the same plate! This could result in a delayed final aspiration including on the
Plate wash mode assay.
Incompatible functions
This function is strictly incompatible with the worklist option m u l t i p i p e t t i n g (see
"Instructions for use Manual")!
Shaking Enter the shaking parameters for one incubation step. This applies only to heated
incubators. For plates which need shaking but do not need heated incubation, you
can define an independent S h a k e step (see chapter 2.5.8 on page 2-56) or enable
shaking prior to reading (see chapter 2.6.1 on page 2-64).
Wrong results
All assays using multishots must be validated for their respective number!
This D i s p e n s e step is not to be used for dispensing samples and controls. For
these, refer to the P i p e t t i n g step (see chapter 2.5.1 on page 2-28). Refer also to the
P i p e t t i n g step concerning the dispensing of reagents (e. g.: diluents) into dilution
plates.
This window shows the plate layout (see chapter 2.3.2 on page 2-20); in the default
setting all wells have a check mark. This means that all wells will receive the selected
reagent. Click on a well to remove the check mark (= deselect the dispense function);
click again to add the check mark again.
Aspirate Check If this item is checked, the system verifies if a dive out
position is detected when retracting the tip after the
aspirate operation and if the difference between dive in
and dive out position is in accordance with the expected
drop of the liquid surface (corresponding to the
aspirated volume).
Pressure If this item is checked and pressure monitoring is
Monitoring activated in the assay header (see chapter 2.2 on page 2-
12), the air pressure in the pipetting system is monitored
during the aspirate and the data are compared with
standard values for a liquid type. This function ensures
the pipetting and dramatically reduces the risk of false
pipetting.
Liquid Type If pressure monitoring is used, a liquid type has to be
selected to obtain standard values for pressure
measurements.
Intra-well delay Define delay between individual well dispense steps in
milliseconds. This option can be used to avoid air
bubbles in the wells. However, it lengthens the dispense
time.
Volume Define the dispense volume per well in µl. The aspirate
volume will be calculated by the system:
Asp. vol.= (No. of multishots x Disp. vol.) + Oversoak
vol.
Note that the dispense volume must not be lower than
the specified minimum volume.
Ov ers o ak Define how much oversoak volume in µl is to be
aspirated (see chapter 2.5.4.1 on page 2-48).
N o . o f M u l t i s h o t s For multiple shots (fill one time, dispense several times):
number of dispense steps (see chapter 2.5.4.1 on page 2-
48).
See note above!
Back to source Volume in µl which is to be moved back into the reagent
container (see chapter 2.5.4.1 on page 2-48).
Eject tip/wash If you select this item, the pipette tip is ejected between
t i p b e t we e n each dispense step.
s ho t s Performing a multidispense step this item has not to be
used.
The wash tip option is not available since Gemini
instruments operate with disposable tips only.
Agitation Agitation time for mixing on the plate transport unit after
dispensing in seconds.
In-Well The number of I n - W e l l M u l t i s h o t s defines how
Multishots many dispense steps are executed within one well. The
volume for each in-well dispensing is calculated by the
software. The software distributes evenly the overall
volume to all dispense steps.
Wash No entries required since Gemini instrument is working
parameters with disposable tips.
Reagent Mixing Allows to mix a reagent before the dispense step will be executed.
Parameters
No. of mix Number of mix cycles for optimum mixing. A mix cycle
cycles consists of aspiration and dispensing of the liquid within
a reagent bottle.
Volume to mix Define the volume to be used for mixing.
with The volume is only considered when the N o . o f m i x
c y c l e s is not 0.
Eject tip/Wash Eject the used disposable tip (or wash the needle) after
tip after mixing mixing.
Mix between Mixes the reagent always before it will be aspirated.
shots
In the Result Report, wells for which a dispense failure has been noted are shown at
the top of the Result Report and the corresponding samples are individually flagged
(VDFail) in the Combined report.
The plate layout you have defined with the standard setting is displayed; all wells
have a check mark, meaning that all wells are to be checked. Click on a well to
remove a check mark (= deselect verification); click again to add the check mark
again.
<= Well Value < Define the OD-value range within which the OD value of
= each well should lie. An error message is displayed if
the value determined is outside the defined range. Here
you can enter logical functions.
Insert Function Opens the Insert function dialog and you can select the
logical functions for definition of the value range (see
query < = W e l l V a l u e <= ).
See chapter 2.8 on page 2-108
Select All All wells are selected for verification (i.e. all wells of the
plate that have a check mark). Click on the respective
well again to deselect individual wells.
Clear All Clears all checkmarks on the plate layout.
Wavelengths
Test wavelength Define the wavelength of the filter used to perform the
reading.
Ref. wavelength Define the wavelength of a possibly required reference
filter.
Relative to last If this item is enabled, the last value measured with
verification dispense verification and the current value are
compared. The value must lie within the range defined
above.
Shake
Mode Only standard mode to shake on the plate transport.
Duration Shake time.
QA
Store mean If you select this item, the mean value of the highlighted
value in QA well is saved in the QA (Qualitative Analysis) database
database under the name of the reagent name selected here. To
view the database, select the menu item F i l e | N e w
and then in the N e w dialog the item Q . A . R e p o r t .
The Q. A. Re p or t dialog is displayed any you can
select the respective reagent. Then the course of the
reagent verification is displayed as a curve. Then the
measured value is displayed in a diagram. If this item is
enabled, you have to select the respective reagent. In
the Q . A . R e p o r t for the Reagent the verification
step values can be verified over time.
Reagent Select the reagent used. The mean values are saved in
the QA database under this name.
You can measure blank plates at certain wavelengths. The value determined is then
subtracted from the reading result.
Read
Parameters
Shake Define the shake time before read in seconds.
Shake mode Select the shake mode (there is only one mode
available).
Test wavelength Define the wavelength of the filter used to perform the
reading.
Ref. Wavelength Define the wavelength of a possibly required reference
filter.
Read interval Define the read intervals in seconds (e.g. 180 = the
plate will be read every 3 minutes until the set end-of-
incubation condition is reached).
Timeout End of incubation time, even if the desired measured
value has not been reached (see explanation on W a i t
u n t i l ).
Condition Define the end-of-incubation condition. This condition can be set either in reference
to a set OD-value or in reference to a variation of the OD-value over time (OD/minute
rate). When the set value (or rate) is reached, the incubation time is over.
This shake step lets you define only the duration parameter, not the frequency as in
the incubation shaking function.
Verification
Index Enter the chamber number that you want to use for
incubation. This should be 0 or 1 for incubation
chambers or 0, 1 or 2 for ambient temperature
chambers.
Other Buttons All other buttons or fields of the I n c u b a t e verification dialog are identical with the
or Fields corresponding buttons or fields of the I n c u b a t e dialog (see chapter 2.5.3 on page 2-
44).
The dialog shows only reagents or wash buffers, which can be used in the assay step
(e. g. only wash buffers for the W a s h step).
Wrong results
New, modified or assumed assays must be validated to avoid wrong results.
Several functions are not available for reagents or wash buffers! The affected
functions be disabled!
Reagent Define the desired aspirate and dispense profiles, which are pre-defined in the
Dynamics U t i l i t i e s | S y s t e m S e t u p menu on the P i p e t t e tab.
Aspirate profile Select the aspirate profile. The name of the selected
profile is displayed.
Dispens e pr ofile Select the dispense profile. The name of the selected
profile is displayed.
Wrong results
It is necessary to use only validated assay/profile combinations to avoid wrong
results.
Define the desired aspirate and dispense profiles, which are pre-defined in the
U t i l i t i e s | S y s t e m S e t u p menu on the P i p e t t e tab.
Click on Add or E d i t button the enter/change the profiles for assigned volume
values.
Maximum Shows the maximum volume for the used aspirate and
volume dispense profiles
Aspirate profile Select the aspirate profile. The name of the selected
profile is displayed.
Dispense profile Select the dispense profile. The name of the selected
profile is displayed.
Wrong results
It is necessary to use only validated assay/profile combinations to avoid wrong
results.
Insert It is possible to insert further evaluation steps (see chapter 2.7 on page 2-86).
Evaluation
Steps
Editing Assay Double-click on the respective symbol in the assay tree to open the respective
Steps parameter window.
Test wavelength Define the wavelength of the filter used to perform the
reading. You can set the wavelength via the arrow
buttons.
Ref. wavelength Define the wavelength of a possibly required reference
filter. If this item is checked you can set the wavelength
via the arrow buttons.
Multi- Parameters for 2nd reading using a different wavelength, when the results of the first
Wavelength reading were outside the reading range.
Reading
Enabled If this item is checked you can define the wavelength
used for the 2nd reading. A 2nd reading takes place only
when the results of the first reading were outside the
reading range.
Secondary test Select a filter wavelength for the 2nd reading. If this item
wavelength is checked you can set the wavelength via the arrow
buttons.
Secondary Select a reference filter wavelength for the 2nd reading.
reference If this item is checked you can set the wavelength via
wavelength the arrow buttons.
Out Of Range The default OVER limit and replacement in the system setup can be overwritten for
Values each individual assay.
Use reader Activate this option to use the reader defaults (see
defaults System-Set-up / Colorimeter tab).
OVER limit Enter the upper limit value (OD) for the photometer (e.g.
2.5). For more details see instrument specifications.
Enabled You can select if another character string is to be output
instead of the measured value when the result is above
or below the limit value, e.g. a digit or a comment. Click
on the respective check box to enable the respective
edit box and you can make your entry.
OVER When exceeding the value: [...] should be printed
replacement instead of the OVER text (2.5).
v al u e
UNDR When falling short of the value: [...] should be printed
replacement instead of the UNDER text.
v al u e
Table Select which results you want to output in a table: this is the same parameter
selection as in the M a t r i x area. Proceed in the same manner: Using the arrow
buttons, copy the desired parameters to the S e l e c t e d F i e l d s list. Click
F o nt . . . in the M a t r i x area to select another font and change the defaulted font
parameters.
Scientific Controls when numbers are displayed in scientific format and when they aren't.
Format
Auto Auto format. You can select the number size (small,
intermediate, large)
• If S m a l l N u m b e r s is checked then any values
less than 1.0 are displayed in scientific format.
• If I n t e r m e d i a t e N u m b e r s is checked then
values between 1.0 and 10000.0 are displayed in
scientific format.
• If L a r g e N u m b e r s is checked then any values
above 10000.0 are displayed in scientific format.
Fixed Exponent Presentation with fixed exponent. Set the value by
clicking on the arrow buttons. For example: 3 means x
103.
Exponent Presentation with multiple exponents. Set the value by
Multiple clicking on the arrow buttons. For example: 3 means x
103.
Decimal Place Define the decimal place. For example: 3 means 1.000.
• Paired
Paired subtraction of the background (B) from the
other wells. The reverse order of the categories on
the plate is used. This means: B1 is subtracted from
the first sample of the last category (e.g. T), B2 from
the 2nd sample of the last category, etc. If all
samples of the last category have been calculated in
this manner, the next B-value is subtracted from the
first samples of the second to last category. The last
B-value is then subtracted from the last sample of the
first category.
Example: Samples are arranged on a plate as
follows:
CO1 B1
CO2 B2
PC1 B3
PC2 B4
T1 B5
T2 B6
In this case, B1 is subtracted from T1, B2 from T2,
B3 from PC1, B4 from PC2, B5 from CO1 and B6
from CO2 (see also result file *.res).
• Plate
A background value which has been saved to a file
can be subtracted for the entire plate. The file is
called via the P l a t e I D browser or chosen before a
run as L i n k e d P l a t e I D .
If the results of the linked or specific plate can not be
found, the software shows an error messages and no
background subtraction will be performed.
• P r e v i o u s Va l u e
The average of the B-Wells of the last processed
Av er age blank mode assay is used for
background subtraction.
Due to the limited traceability of the P r e v i o u s
V a l u e, this function should not be used.
See note below!
• Custom
In this mode you can assign B-wells to the T-wells. If
you select this mode, the B-wells are automatically
listed in the menu next to the well matrix. First select
a B-well in the menu and then the other well in the
matrix from which the B-well is to be subtracted.
Proceed in the same manner with all B-wells.
Wrong result
Do not use the calculation method P r e v i o u s Va l u e when assay layout does not
contain a blank.
Plate This item is enabled only if you have selected P l a t e in the previous query.
Subtraction
Mode
Plat e I D Define the file name of the plate ID.
Browse Opens the Open dialog box for selection of a result
file.
Use Linked Allows the user to choose another plate with the
Plat e I D background values before a run.
Paired/Custom The microtiter plate layout is displayed and in Cu s t om mode those wells from
Subtraction which the background is to be subtracted are identified by a check mark.
Mode
Page Header/ The report page header and page footer can be defined by any entries as well as by
Footer the following place holders:
• & D = date
• & F = name of result file
• & I = lab
• & Q = quality control
• & P = page number
• & S = system serial number
• & V = user software version number
The header/footer is printed on each page of the report.
Report The structure of the report is defined by the individual data fields and their order.
Structure Copy the desired data fields from the A v a i l a b l e F i e l d s list to the S e l e c t e d
F i e l ds list and change their order as needed.
• Spreadsheet Results 1
Data specified in the S p r e a d s h e e t layout menu
1.
• Spreadsheet Results 2
Data specified in the S p r e a d s h e e t layout menu
2.
• Spreadsheet Results 3
Data specified in the S p r e a d s h e e t layout menu
3.
• Qualitative Results
Data specified in the Q u a l i t a t i v e layout menu.
• Q u a nt i t at i ve R es u l t s1
Data specified in the Q u a n t i t a t i v e S e t t i n g s
1 layout menu.
• Q u a nt i t at i ve R es u l t s2
Data specified in the Q u a n t i t a t i v e S e t t i n g s
2 layout menu.
• Combined Report
Data specified in the R e p o r t S e t t i n g s
C o m b i n e d R e p o r t menu.
• <Page Break>
Insert of a new page during printout.
Selected Fields Shows the data fields of the reports selected via the
arrow buttons.
> Copies the data fields highlighted in the A v a i l a b l e
F i e l ds list to the S e l e c t e d F i e l d s list.
>> Copies all data fields to the S e l e c t e d F i e l d s list.
< Removes the data fields highlighted from the
S e l e c t e d F i e l d s list.
<< Removes all data fields from the Se l e c t e d F i e l d s
list.
Move Up Moves the data field highlighted in the list up one row.
Move Down Moves the data field highlighted in the list down one row.
• Reader 2 value
OD value (after background subtraction)
• Reader 2 mean
Average reader value of replicates
• Reader 2 SD
Standard deviation of reader value
• Reader 2 CV
Coefficient of variation of reader value
• Reader 2 SE
Standard error of reader value
• Spread. 1 value
Result of spreadsheet 1 calculation
• Spread. 1 mean
Average of the replicates after spreadsheet 1
calculation
• Spread. 1 SD
Standard deviation of spreadsheet 1 value
• Spread. 1 CV
Coefficient of variation of spreadsheet 1 value
• Spread. 1 SE
Standard error of spreadsheet 1 value
• Spread. 2 value
Result of spreadsheet 2 calculation
• Spread. 2 mean
Average of the replicates after spreadsheet 2
calculation
• Spread. 2 SD
Standard deviation of spreadsheet 2 value
• Spread. 2 CV
Coefficient of variation of spreadsheet 2 value
• Spread. 2 SE
Standard error of spreadsheet 2 value
• Spread. x ...
x: calculation/value - see above
• Quant. 1 value
Result of first quantitative step
• Quant. 1 mean
Average value for the replicates after quantitative
step one
• Q u a n t . 1 SD
Standard deviation of result of quantitative 1
• Q u a n t . 1 CV
Coefficient of variation for result of quantitative 1
• Q u a n t . 1 SE
Standard error of result for quantitative 1
• Qual. value
Qualitative result of replicate of sample.
• Qual. mean
Qualitative result of average of the sample replicates.
• Quant. 2 value
Result of second quantitative step
• Quant. 2 mean
Average value for the replicates after quantitative 2
• Quant. 2 SD
Standard deviation of result of quantitative 2
• Quant. 2 CV
Coefficient of variation for result of quantitative 2
• Quant. 2 SE
Standard error of result for quantitative 2
Copy the respective parameters with the arrow buttons
> or >> to the S e l e c t e d F i e l d s list. You can
change the order via the M o v e U p and M o v e
D o w n buttons according to your needs.
Selected Fields List of selected parameters which are output after
reading in form of a matrix (see also A v a i l a b l e
F i e l ds).
> Copies the highlighted parameters from the
A v a i l a b l e F i e l d s list to the S e l e c t e d F i e l d s
l i s t.
>> Copies all parameters from of the A v a i l a b l e
F i e l ds list to the S e l e c t e d F i e l d s list.
< Removes the highlighted parameters from the
S e l e c t e d F i e l d s list.
<< Removes all parameters from the S e l e c t e d F ie l d s
list.
Move Up Moves the highlighted parameter in the S e l e c t e d
F i e l ds list up one row.
Move Down Moves the highlighted parameter in the S e l e c t e d
F i e l ds list down one row.
Font... Opens the F on t dialog box and you can define font,
size and type for data output. The default font is Arial,
10pt, standard. The selected font is valid for matrix as
well as for table presentation.
Gridlines Select this item to show gridlines.
Table Select which results you want to output in a table: this is the same parameter
selection as in the M a t r i x area. Proceed in the same manner: Using the arrow
buttons, copy the desired parameters to the S e l e c t e d F i e l d s list.
Scientific Controls when numbers are displayed in scientific format and when they aren't.
Format
Auto Auto format. You can select the number size (small,
intermediate, large)
• If S m a l l N u m b e r s is checked then any values
less than 1.0 are displayed in scientific format.
• If I n t e r m e d i a t e N u m b e r s is checked then
values between 1.0 and 10000.0 are displayed in
scientific format.
• If L a r g e N u m b e r s is checked then any values
above 10000.0 are displayed in scientific format.
F i x e d Ex p o n e n t Presentation with fixed exponent. Set the value by
clicking on the arrow buttons. For example: 3 means x
103.
Exponent Presentation with multiple exponents. Set the value by
Multiple clicking on the arrow buttons. For example: 3 means x
103.
Decimal Place Define the decimal place. For example: 3 means 1.000.
In this dialog you define the structure of the ASTM E 1394 export:
Results Shows the selected data fields for the ASTM E 1394
export.
Export Control If this item is checked, the values of the control wells are
Well Values also included in the export file.
Identifier For each result specify an ID which will be appended to
the universal test ID field of the result record.
Result Shows the available data fields for the ASTM export
record:
• Reader mean
Average OD value of replicates.
• Qual. mean
Qualitative result of average of the sample replicates.
• Quant.1 mean
Average quantitative 1 value of replicates.
• Spread.1 mean
Average spreadsheet 1 value of replicates.
• Quant.2 mean
Average quantitative 2 value of replicates.
• Spread.2 mean
Average spreadsheet 2 value of replicates.
• Spread.3 mean
Average spreadsheet 3 value of replicates.
• Spread.4 mean
Average spreadsheet 4 value of replicates.
• Reader 2 mean
Average reader 2 value of replicates.
Wrong results
Because of mathematical roundings it is necessary to validate assays!
After selection of a step the respective dialog is displayed for parameter entry. Once
you have entered your parameters and saved your entries, the new step is inserted
in the assay tree.
Order of You can arrange the individual steps in any order. The order is defined during new
Evaluation definition of a step by clicking with the right mouse button on the symbol before the
Steps one you the want to place the new steps.
Each evaluation step uses the output data of the previously programmed evaluation
step. The evaluation step Q u a n t i t a t i v e S e t t i n g s can be selected twice, all
other steps only once.
Editing Assay Double-click on the respective symbol in the assay tree to open the respective
Steps parameter windows.
Examples:
The mean of the positive controls must be above 0.7 OD:
PC>0.7.
Use Stored If this item is checked, all other items on this tab are
Control Value dimmed. In this case, enter the name of the file including
File the control values.
Browse Select the stored control value file.
Histogram of Select a result histogram.
results
Grey Zones Define grey zones, i.e. reading ranges which are not assigned to any result type.
Result Types Describe the result types, i.e. enter the explanation for the characters +, - and ? .
• Auto
In this column you can define for each of the 4
parameters - by clicking on the respective check box
- if it is to be defined automatically (Auto) or if user
wants to enter a known value. A check mark in the
check box means automatic definition of the
parameters. Deselect the check mark to enable the
respective edit box for entry.
• Max. no. of iterations
Enter the number of iterations for curve calculation to
be performed by the program to fit the curve.
• Report values below A as 0.000
In the Quantitative Settings, the data model 4-
Parameter allows to replace values to “0.000” in
case the calculated value is lower than the
parameter “A”. The replaced value of “0.000” will be
used for further calculations. This replacement
applies to all well types (including standards if the
D i s p l a y p r e d i c t e d s t a n d a r d v a l u e s box
is checked).
Data You can choose between logarithmic and linear presentation of the X or Y axis.
Transformation
Standards In this range you have to define the standard values. In the standard list all standards
defined in the assay are displayed. If the standards are always equal, enter the
respective value for each standard. Select the standard you want from the list and
then enter the value in the entry box above. If the standard values are variable, do
not enter any values here, but select the item Variable Std. Values. In this case you
are prompted to enter the standard values prior to each run.
Stored Model This area are enabled if you select S t o r e d M o d e l in the D a t a Mo d e l menu.
FileName
Browse Click B r o w s e to select a file that includes the desired
evaluation curve (*.mod).
Use the Plate ID Allows the user to choose a plate (linked plate ID) with
as the file name the desired evaluation curve before a run.
User-Defined Here you can enter a user-defined evaluation curve as formula. To enable this field
Model you have to select U s e r - D e f i n e d in the D a t a Mo d e l menu. Enter the curve
equation next to X =, and use for this the mathematical character offered in the
I n s e r t F u n c t i o n dialog box.
Input terms: This formula has to include 'Y ', 'S a m p l e' or 'S a m p l e . n' once (n =
individual value of cavity). Moreover, this edit box may contain: well labels, digits,
mathematical operators and brackets, as well as the functions offered via I n s e r t
Func tion. . . .
T[x] Shows all samples selected in the test list below. In the
entry box you can edit the value (dilution factor) of the
sample selected from the menu. This entry is saved if
you select another test from the list.
Auto Fill... Opens the A u t o F i l l S e t t i n g s dialog box (see
below) and you can enter information for automatic
filling of the dilution table. Automatic filling starts with the
highlighted row. Enter a value for this dilution. For all
following cavities you can then define a multiple.
Auto Fill
Settings dialog
Box
Same as The following dilutions will have the same value as the
previous highlighted one.
Add to previous Add to each previous dilution the value defined in the
F a c t o r field.
Example: The highlighted value is 10, the factor 2, this
results in the series of numbers 10, 12,14, 16...
Subtract from Subtract from each of the previous dilutions the value
previous defined in the Fac t or field.
Example: The highlighted value is 100, the factor is 1,
this results in the series of numbers 100, 99,98, 97...
Multiply by Multiply each of the previous dilutions by the value
previous defined in the Fac t or field.
Example: The highlighted value is 10, the factor is 2, this
results in the series of numbers 10, 20, 40, 80...
Divide by Divide each of the previous dilutions by the value
previous defined in the Fac t or field.
Example: The highlighted value is 10, the factor is 2, this
results in the series of numbers 10, 5, 2.5, 1.25...
F ac t o r Enter the desired factor for the calculations. Default
value is 1.
Auto Automatic L a b e l , M i n i m u m , M a x i m u m or
Major Unit.
Label The label is taken from the assay definition or entered
manually.
Minimum Minimum value: automatic or manual setting.
Maximum Maximum value: automatic or manual setting.
Major Unit Automatic or manual setting.
Gridlines Show gridlines.
Logarithmic Logarithmic scale of the respective axis.
Scale
Conditions
Condition In this field you enter the condition.
For example "Sample>2.5" means: "If the result > 2.5,
then use the character sequence in the R e s u l t
R e p l a c e m e n t field for result output instead of the
measured value."
The condition is automatically entered in the first row (or
if several rows exist, in the highlighted row) of the list of
conditions. Each row starts with "If". The entry you
make is automatically saved to this list and continues
the "If" condition. For example IF(>2.5).
You can make the entry via keyboard or by clicking the
I n s e r t F u n c t i o n . . . button. Click on New to define
a new condition.
Result Here you enter what should appear instead of the result
Replacement if the above given condition is fulfilled, e.g. "overflow".
The entries made here are automatically inserted in the
highlighted row of the list of conditions: e.g. IF
(Sample>2.5, Result:='overflow'. Process no further).
Result Here you can define a flag (e.g. a !) to annotate the
Annotation result which fulfils the above condition. The flag is
shown in the result reports.
The entries you make here are automatically inserted in
the highlighted row of the list of the conditions. e.g. IF
(Sample>2.5, Result:='overflow'. Flag='!').
Do not process If this item is checked, the results which do not fulfil the
result any criteria are not processed any further.
further
List of conditions Here the conditions are listed which are created via the
N e w C o n d i t i o n button.
Insert Function Opens the I n s e r t F u n c t i o n dialog box and you can
select a logical link for the description of your conditions
(see chapter 2.8 on page 2-108).
New Condition Click on this button to create a new condition in the list.
The new row starts with "IF..." You can define a condition
via keyboard or by clicking the I n s e r t F u n c t i o n . . .
button.
Delete Deletes the selected condition from the list of conditions.
Clear All Clears the entire list of conditions.
Move Up Moves the highlighted row in the list of conditions up one
row.
Move Down Moves the highlighted row in the list of conditions down
one row.
Undo Revokes the last action.
Redo Revokes the undo function.
Legends Here you can create a legend and provide explanations on the results and
abbreviations, etc.
Result Enter the term or the result you want to explain in the
legend. It is automatically entered in the first row of the
legend list or in case several legend rows exist in the
highlighted row.
Legend Enter a description. Separated by a tab mark, this text is
written in the same row as the result.
List of legends Result and legend, which were entered in the two edit
boxes above, are displayed separated by a tab stop.
New Legend Creates a new row for result and legend.
Delete Deletes the selected row from the legend list.
Clear All Clears the legend list.
Move Up Moves the highlighted row in the legend list up one row.
Move Down Moves the highlighted row in the legend list down one
row.
Reference You can type text or a formula (using the format function) into this field. The value of
Range this field will then be displayed in the results, stored in the sample results database
and transmitted to the LIMS. This field would normally be used to indicate what range
a normal result might be expected to lie within.
Wrong results
It is necessary to validate assays to avoid wrong results.
A 1 ... H12 In front of this row you see the identification of the well
highlighted in the plate layout. Enter the desired
calculation formulae, (e.g. A1-A2, when the result of well
A2 is to be subtracted from the result of well A1). The
entries are automatically transferred to the highlighted
field of the plate matrix.
Insert Function Opens the I ns er t F u nc t i o n dialog box and you can
view and select the available logical links (see chapter 2.8
on page 2-108).
Plate layout Shows the entered formulae for calculation of the
reading results of the respective cavity.
Result Units Define the unit of calculated results.
Handling:
• Use the mouse and press on the left mouse key to select a cell.
• Press on the Enter key to select the cell in the next row.
For the assay definition you may enter formulae at various points. Enter the
mathematical signs (:, *, +, - ) via keyboard.
The logical functions are listed in the I n s e r t F u n c t i o n dialog, which appears
when you click the I n s e r t F u n c t i o n button. The logical functions must be
selected from this list. Otherwise, the program does not recognize the character
sequence as logical function.
The Gemini COMBO instrument is an open system. It allows users to process pre-
defined assays but also to edit existing assays or to create their own assays.
Wrong results
New, modified or assumed assays must be validated to avoid wrong results.
Wrong results
If assays from a Gemini instrument will be used on a Gemini COMBO instrument,
they must be validated to avoid wrong results.
The instrument doesn’t generate any end result for slides. The slides will only be
prepared for further processing (e.g. evaluation under an external fluorescence
microscope) by the user.
Most of the processes of IFA and ELISA assays are similar, so in some chapters it
will only be referred to the corresponding chapter 2 of the ELISA description.
The display area (right) shows the slide layout and above some default settings. The
assay explorer tree in the left-hand side of this window shows the basic steps of an
assay. You can work with it in the same manner as you work with the Windows
Explorer: a folder including further information is identified by a plus sign (+). Click
on the plus sign to open the folder. An open folder is represented by a minus sign (-
). Click on the minus sign to close the folder.
The assay explorer tree, henceforth referred to as assay tree, includes the 4 basic
steps of an assay which cannot be deleted:
Show Assay Click with the left mouse button on an assay folder to view the respective assay step
Step with the description of the already defined functions in the display area.
Edit Assay Step To edit an assay step, you always have to select the assay step to be edited.
Insert Assay To insert a new assay step, you always have to select the assay step in front of which
Step the new step is to be inserted.
Context Menu The context menu for inserting new assay steps in a new assay can be selected via
for Inserting the menu item Edit on the menu bar, as well as by clicking the right mouse button on
Assay Steps the step before which the new step is to be inserted.
Undo Revokes the last action you have executed in the assay
tree.
Redo Revokes the last U n d o action again.
Cut Cuts out a highlighted assay step and saves it to the
clipboard.
Copy Copies a highlighted assay step to the clipboard.
Paste Inserts an assay step on the clipboard in front of the
highlighted assay step.
Edit To edit an assay step, you always have to select the
assay step to be edited.
Insert Step To insert a new assay step, you always have to select
the assay in front of which the new step is to be inserted.
Delete Deletes a step highlighted in the assay tree.
Select All Selects all assay steps.
Move Up Moves a highlighted assay step up one step.
Move Down Moves a highlighted assay down one step.
Execute Current Execute only the highlighted step of the selected assay
Step... (see notes below).
Execute Executes all remaining steps included the highlighted
Remaining step of the selected assay (see notes below).
Steps...
When you are performing the execute functions during a running worklist, the slide
will be scheduled into this running worklist!
General buttons
which apply on
OK Saves all entries and returns to the main menu.
most dialog
boxes Apply Saves the entries and the dialog box remains displayed.
Canc el Returns to the main menu without saving any entries.
Data that have already been saved by pushing Ap p l y
remain stored.
Help Calls on-line help.
Assay Header Double-click on this step to open the respective parameter window (see chapter 3.2 on
page 3-6).
Assay Layout Double-click on this step to open the respective parameter window (see chapter 3.3 on
page 3-12).
Scheduling Move this step to perform a schedule break (see chapter 3.4 on page 3-13).
Break
Definition of Click with the right mouse button on an assay step (exception A s s a y H e a d e r
possible Assay and A s s a y L a y o u t) to open the I n s e r t S t e p dialog box showing the
Process Steps available steps (see chapter 3.5 on page 3-15).
Report Settings Double-click on this step to open the respective parameter window (see chapter 3.6 on
page 3-20).
Saving the As soon as you have defined all steps, select the item F i l e > S a v e A s and save
Assay File the assay under a new name (extension *.asy).
Assay Header *
The assay header includes general information such as: assay name, comments (if any),
password protection, written by, creation date, last edited, slide type used, maximum ambient
temperature as well as the possibility to enable the Include Sub-Assay feature and activate the
Dynamic Dilutions function.
Assay Layout *
Table representation of a slide showing the respective positions of the samples, standards and
controls for that assay.
Schedule break
All assay steps from all assays of the worklist programmed before the schedule break will be
performed in advance.
Assay steps (Processing steps)
Pipette Inserts a pipetting step (for samples, controls and
standards).
I FA W a sh Inserts a washing step.
Incubate Inserts an incubation step.
Dispense Inserts a dispense step (for reagents).
IFA dispense Inserts the IFA dispense step.
Report Settings *
Definition of the contents of the result printout, of the combined report and of the format and
contents of the result export file.
End of Protocol *
Password Option to enter a password. This rules out that the assay can not be modified by
anyone who does not know the password. However, no password is required to run
the assay.
List of slide types Selection box listing the slide types pre-defined for the
system.
Slide overview Shows the selected slide type as overview. Click on a
well to select a position.
Slide name The name of slide type highlighted in the menu is
displayed. Click on Ad d to define a new slide type and
enter a new name. This will create a new plate type
which has to be combined with a *.rac file, i.e. you have
to choose a *.rac file in the S l i d e f i l e drop-down
menu.
Slide file Drop-down menu showing the available *.rac files.
These files contain the slide coordinates and are
created using the teacher software.
Colour Drop-down menu to select the color of the slide. The
slide will be shown with this color in the user software.
Note: Use different colors for slides with similar layout.
User will be recognized that slides are different.
Position Selection of the slide wells. If you select a position, the
dialog shows the corresponding shape, offsets and
sizes.
Shape Select the shape of the well (circle or rectangle) in mm.
O f f s e t X /O f f s e t Y Position of the selected well on the slide in mm (see
figure 3-1: on page 3-9).
• X = 0: right side of the slide picture (right side of the
real slide)
• Y = 0: upper side of the slide picture (rear side of the
real slide)
S i z e X /S i z e Y Size of the selected well (see figure 3-1: on page 3-9).
Add Click this button to create a new slide type (see below).
The generated slide type is entered in the menu to the
left and it has to be combined with an *.rac file.
Delete Deletes the slide type highlighted in the selection box.
OK Closes the S l i d e T y p e E d i t o r dialog. Changes
are saved. The program returns to the A s s ay
P r o t o c o l H e a d e r dialog.
Maximum of 12 wells
A maximum of 12 wells is specified but more wells can be added on own risk. Only
validated assays can be run on the Gemini Combo.
Correct parameters
The software uses Size X and Size Y parameters for calculating the dimensions of
the slide well which will be used for the In-Well Multishot feature. Make sure that
these parameters are correct.
Slide identification
To avoid confusion it will be useful to choose similar slide names and rac file names.
A sub-assay contains all steps from the primary assay, but P i p e t t e and
Di s p e n s e steps could be customized.
The sub-assay must not be created after finalization of the primary assay.
Verify assay
The V e r i f y A s s a y function checks either the primary assay or the sub-assay
depending on the currently selected view. It must be assured that the basic steps are
equal especially after changes in primary assay or sub-assay are done. Validation of
assay and correct databases are required.
All assay steps from all assays of the worklist programmed before the schedule
break will be performed in advance.
The Sc h e d u li n g b r e a k step can be changed by the Mo v e u p and M o v e
d o w n functions in the Context Menu for Inserting Assay Steps.
Example:
The assay has at first the pipette step, followed by the schedule break and a second
pipette step.
The scheduling shows, that at first the sample pipetting of all slides will be performed;
after this step, the slides are scheduled according the remaining assay steps.
After selection of a step the respective dialog box is displayed for parameter entry.
Once you have entered your parameters and saved your entries, the new step is
inserted in the assay tree.
Order of Assay You can arrange the individual steps in any order. The order is defined during new
Steps definition of a step by clicking with the right mouse button on the folder before the
one you want to place the new step.
But when defining your steps, you can freely add them in any order to start with
(some are easier to define than others) and then rearrange them in the correct order
of the assay with the M o v e U p / M o v e D o w n items of the E d i t menu (or of
the context menu).
Editing Assay Double-click on the respective symbol in the assay tree to open the respective
Steps parameter window.
Replicate steps If your assay includes steps that are repeated identically or almost identically (this is
often the case for I F A W as h steps or I n c u b a t i o n steps), you can speed up
the definition process by using the C o p y / P a s t e function of the E d i t menu (or
of the context menu). Edit the parameters if necessary.
Additional parameter on the Dis p en se dialog (see chapter 2.5.1.3 on page 2-34):
Incompatible functions
When using the I n - W e l l M u l t i s h o t s feature the I n l i q u i d d i s p e n s e
function must not be activated.The software uses size X and size Y parameters
added in the Sl i d e E d i t o r for calculating the dimensions of the slide well. Make
sure that these parameters are correct.
Insufficient liquid
If the liquid volume of a dilution is not sufficient, it can happen that the liquid volume
for the following dilutions and the transfer step is insufficient.
• Take care that sufficient liquid volume for serial dilution and transfer step is
inside the dilution well.
Sweep The sweep parameters are used for washing wells on a slide.
parameters
Number of Describes how often the pipettor needle moves within a
s we e p s well. The track of these moves depends on size of well.
Z Offset The Z offset parameter describes the distance above
Zmax. Zmax is the lowest position of the pipettor needle
in the slide area. Zmax is taught using the Teacher
software.
Sweep speed Sets the X-/Y-move speed of the pipettor during the
sweeps. The speed parameter is in percentage of
maximum speed of the pipettor.
Delay The delay parameter describes the time between start of
aspiration pump and start of dispense pump.
Cycles Defines how often all used wells of one slide shall be
washed. A new cycle begins if all used wells of a slide
were washed.
Final aspirate The marked checkbox F i n a l a s p i r a t e moves the
needle to the height of Z D o w n O f f s e t . During this
movement only the aspirate pump is active. This feature
allows optimal slide washing and to minimize the liquid
remaining on the slide well for the last wash cycle.
Z Down Offset The Z D o w n O f f s e t parameter defines the offset
above Zmax. The aspirate pump switches off, if
Z D o wn O f f s e t is reached.
Dispense pump
parameters
Speed The dispense pump parameter speed value is in
percentage of maximum dispense pump speed.
Wash Buffer To select the wash buffer, click the arrow button to open the drop-down list. If the
desired buffer is not included in the list you may define it by clicking the B u f f e r s
button (see chapter 2.5.11.1 on page 2-59).
Cover Height for fluid dispensing, to avoid the aspiration of liquid on the slide.
parameter
Z Start Start height for dispensing.
Z Stop Stop height for dispensing.
Wash Buffer To select the wash buffer, click the arrow button to open the drop-down list. If the
desired buffer is not included in the list you may define it by clicking the B u f f e r s
button (see chapter 2.5.11.1 on page 2-59).
Dispense pump
parameters
Speed The dispense pump parameter speed value is in
percentage of maximum dispense pump speed.
The system doesn’t generate any end result for slides. The slides will only be
prepared for further processing (e.g. evaluation under an external fluorescence
microscope) by the user.
4 Index
P V
Password protection ............................. 2-11 Verify dispense steps ............................ 2-50
Pipetting steps ....................................... 2-28
Edit operation ......................................... 2-31
W
• Aspirate parameters .......................... 2-32
• Dispense parameters ........................ 2-34 Warning symbols ..................................... 1-2
• Eliminate assay drift .......................... 2-36 Warnings ................................................... 1-1
Pipetting steps (IFA) .............................. 3-16 Wash steps ............................................. 2-38
Plate types .............................................. 2-15 Wash steps (IFA) ................................... 3-17
Primary-assay ........................................ 3-11
Q
Qualitative evaluation ........................... 2-89
Qualitative settings ................................ 2-89
R
Read definition ....................................... 2-63
Reader settings ..................................... 2-64
Reagent database ................................. 2-59
Create a reagent .................................... 2-60
Edit a reagent ......................................... 2-60
Reagents selection ................................. 2-59
Report settings ....................................... 2-73
Report settings (IFA) ............................. 3-20
S
Safety instructions ................................... 1-4
Scheduling break .......................... 2-24, 3-13
Shake steps ........................................... 2-56
Slide types ................................................ 3-8
Special types ............................................ 1-3
Spreadsheet parameters .................... 2-106
Structure of assay files ......................... 2-10
Structure of assay files (IFA) .................. 3-5
Sub-assays ............................................ 3-11
Symbols
Warning .................................................... 1-2
T
Temperature inside the instrument ...... 2-44
Text file export ........................................ 2-82
Typographical conventions .................... 1-1