MKDR Xpress
MKDR Xpress
mKDR Xpress
Technical Publication
OM-0481R1_EN_KON
This manual covers the following equipments / Este manual cubre los siguientes equipos
Ce manuel couvre les équipements suivants / Il presente manuale descrive i seguenti dispositivi
SEDECAL
Sociedad Española de Electromedicina y Calidad S.A.
Pelaya, 9 -- 13. Polígono Industrial “Río de Janeiro”
28110 Algete, Madrid -- España (Spain)
Tel: +34 916 280 544 Fax: +34 902 190 385 www.sedecal.com
REVISION HISTORY
1 MAR 08, 2022 Manual Cover and some rendered images have
been updated.
New Section: Appendix C -- Administrator User.
Updated Sections: Safety Symbols, Mains
Connection, Smart ON/OFF System: ON/OFF
Button + RFID System, Infrared Remote Control,
Manual Clutch Screws, Collimator Movements, 3.7
Collimator Controls, Power Reduction, System
Snapshot, System Messages, Factors.
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
OM-0481R1
Mobile X-ray Unit
Operation
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.
OM-0481R1
Mobile X-ray Unit
Operation
TABLE OF CONTENTS
Section Page
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
OM-0481R1 i
Mobile X-ray Unit
Operation
Section Page
ii OM-0481R1
Mobile X-ray Unit
Operation
Section Page
4 CONTROL CONSOLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
OM-0481R1 iii
Mobile X-ray Unit
Operation
Section Page
iv OM-0481R1
Mobile X-ray Unit
Operation
SECTION 1 INTRODUCTION
This manual contains all the information necessary to understand and operate
the Mobile X-ray Units. It provides a general description, safety and regulatory
information, operating instructions and specifications concerning the system.
This Unit is designed for general radiography. It provides all the advantages of
high frequency waveform Generators including lower patient dose, shorter
exposure times as well as greater accuracy and consistency.
All functions, displays and controls are logically arranged, easily accessible and
identified to prevent confusion. Technique factors and functions are selected on
the Control Console.
Illustration 1-1
Mobile X-ray Unit
OM-0481R1 1
Mobile X-ray Unit
Operation
Control Console.
G HV Tank, HV Transformer.
2 OM-0481R1
Mobile X-ray Unit
Operation
Dosimetry (optional).
Manual Collimation.
Heat Unit storage for the X-ray Tube, even after turning ON/OFF the
equipment.
Equipped with closed loop control of X-ray Tube current, kVp and
filaments, which minimize potential errors and the need for
readjustments.
OM-0481R1 3
Mobile X-ray Unit
Operation
Date of manufacture.
Serial number.
Inherent Filtration.
4 OM-0481R1
Mobile X-ray Unit
Operation
Additional description:
The X-Ray image receptors used in this unit are Digital Detectors.
The Normal Use of this equipment is defined as the Intended Use plus the
Maintenance and Service tasks.
1.3.3 CONTRAINDICATIONS
Do not use the equipment for any purposes other than those for which it is
intended. Operation of the equipment for unintended purposes could lead to
fatal or other serious injury.
OM-0481R1 5
Mobile X-ray Unit
Operation
Digital Detectors.
Grids (optional)
Other Accessories.
6 OM-0481R1
Mobile X-ray Unit
Operation
Regulatory information and symbols used in the equipment are detailed in this
section to operate it safely.
2.1 GENERAL
OM-0481R1 7
Mobile X-ray Unit
Operation
8 OM-0481R1
Mobile X-ray Unit
Operation
ITEM
IEC 60601-2-54:2009 , and
21 CFR
IEC60601-2-54:2009+AMD1:2015
Total of all layers composing the front panel of cassette holder 1.2 1.2
Total of all layers composing the front panel of FILM CHANGER 1.2 1.2
Note 1.-- Devices such as RADIATION DETECTORS are not included in the item listed in this table.
Note 2.-- Requirements concerning the ATTENUATION properties of RADIOGRAPHIC CASSETTES and of INTENSIFYING
SCREENS are given in ISO 4090 [3], for ANTI--SCATTER GRIDS in IEC 60627[1].
Note 3.-- ATTENUATION caused by table mattresses and similar accessories is not included in the maximum ATTENUATION
EQUIVALENT for PATIENT SUPPORT.
Note 4.-- Maximum ATTENUATION EQUIVALENT mm Al is only applied to the corresponding item. If several items given in this table
are located in the path of the X-RAY BEAM between the PATIENT and the X-RAY IMAGE RECEPTOR, each corresponding maximum
ATTENUATION EQUIVALENT mm Al is separately applied to each item.
OM-0481R1 9
Mobile X-ray Unit
Operation
2.2 RESPONSIBILITIES
10 OM-0481R1
Mobile X-ray Unit
Operation
OM-0481R1 11
Mobile X-ray Unit
Operation
Because exposure to X-ray radiation can be damaging to the health, use great
care to ensure protection against exposure to the primary beam. Some of the
effects of X-ray radiation are cumulative and may extend over a period of
months or years. The best safety rule for an X-ray operator is “Avoid exposure
to the primary beam at all times”.
Any object in the path of the primary beam produces secondary (scattered)
radiation. The intensity of secondary radiation depends on the energy and
intensity of the primary beam and the atomic number of the object material
struck by the primary beam. Secondary radiation may be of greater intensity
than that of the radiation reaching the receptor. Take protective measures to
safeguard against it.
12 OM-0481R1
Mobile X-ray Unit
Operation
- Use the smallest X-ray field collimation. Make sure that the
area of interest will be completely exposed and the X-ray field
does not exceed the area of interest.
OM-0481R1 13
Mobile X-ray Unit
Operation
The most effective method of determining whether or not the existing protective
measures are adequate is the use of instruments to measure the exposure.
These measurements should be taken at all locations where the operator, or
any portion of the body may be exposed. Exposure must never exceed the
accepted tolerable dose.
14 OM-0481R1
Mobile X-ray Unit
Operation
Manufacturer.
Date of Manufacture.
Medical Device.
Serial Number.
Model Configuration.
OM-0481R1 15
Mobile X-ray Unit
Operation
Ionizing radiation.
Dangerous voltage.
16 OM-0481R1
Mobile X-ray Unit
Operation
Warning: Electricity.
Warning: Do not place fingers between mobile and fixed parts of the
equipment, it may cause serious injuries to patient or operator.
As well, make sure the patient extremities are correctly positioned
into limit areas during operation, movement of parts may cause
serious damages to patient.
No pushing.
No sitting.
No stepping on surface.
Do not handle.
OM-0481R1 17
Mobile X-ray Unit
Operation
Emergency stop.
Stand-By
(Only applies to IEC 60601-1:2005, IEC 60601-1:2005+AMD1:2012)
“ON” power.
“OFF” power.
Alternating current.
18 OM-0481R1
Mobile X-ray Unit
Operation
Direct current.
Earth (Ground).
OM-0481R1 19
Mobile X-ray Unit
Operation
2.7.1 CERTIFICATIONS
The Mobile X-Ray Unit covered by this Operation Manual complies with the
following certifications:
Note . Mobile X-Ray Unit model or type references are stated at the
back cover of this document.
20 OM-0481R1
Mobile X-ray Unit
Operation
OM-0481R1 21
Mobile X-ray Unit
Operation
X-Ray equipment specified for any radiological examination that requires the
operator or staff to be close to the patient during normal use (a.e. some pediatric
examinations or other types of examinations for patients that may require
assistance), shall have at least one “Significant Zone of Occupancy” for the use
of the operator and staff, designated as follows:
Illustration 2-1
Radiographic Examination on the Chest Unit or Front Panel
CHEST UNIT
Focal Spot
S
140 cm
Focal Spot
S
Protective Device
Wall
Dosimeter d Patient Support
X-ray Receptor
d = DISTANCE FROM THE AXIS OF THE X-RAY
BEAM TO THE DOSIMETER CHEST UNIT
Focal Spot
22 OM-0481R1
Mobile X-ray Unit
Operation
Illustration 2-2
Radiographic Examination on any Patient Support or any Table
Focal Spot
SID 100 cm
MOBILE X-RAY UNIT
Phantom
X-ray Receptor
Patient Support
S
RAD TABLE
Focal Spot
S
S4
AT FRONT SIDE OF THE MOBILE UNIT
SIGNIFICANT ZONE OF OCCUPANCY
RAD TABLE
S3
Focal Spot
MOBILE X-RAY UNIT
S5
OM-0481R1 23
Mobile X-ray Unit
Operation
Refer to Illustration 2-1 for position of the X-ray Unit during radiographic
examination on the Chest Unit or Front Panel, and refer to Illustration 2-2 for
position of the X-ray Unit during radiographic examination on any Patient
Support or any Table.
24 OM-0481R1
Mobile X-ray Unit
Operation
Illustration 2-3
Distribution of Stray Radiation on the Chest Unit or Front Panel
200
180
160
140
HEIGHT ABOVE FLOOR (cm)
120
100
80
60
40
20
0
0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000 5500 6000 6500 7000 7500
S11 d = 50 cm
S12 d = 100 cm
S21 d = 50 cm
S22 d = 100 cm
Protective Device
Wall
Patient Support
X-ray Receptor
CHEST UNIT
Focal Spot
OM-0481R1 25
Mobile X-ray Unit
Operation
Illustration 2-4
Distribution of Stray Radiation on any Patient Support or any Table
200
180
160
140
HEIGHT ABOVE FLOOR (cm)
120
100
80
60
40
20
0
0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000 5500 6000 6500 7000 7500 8000 8500 9000 950010000
STRAY RADIATION (μGy/h)
S4
AT FRONT SIDE OF THE MOBILE UNIT
SIGNIFICANT ZONE OF OCCUPANCY
S31 d = 50 cm
RAD TABLE
S32 d = 100 cm
d
S41 d = 50 cm S3 d
S42 d = 100 cm
S51 d = 50 cm
d
Focal Spot
MOBILE X-RAY UNIT
S52 d = 100 cm
S5
26 OM-0481R1
Mobile X-ray Unit
Operation
This equipment generates, uses, and can radiate radio frequency energy.
increase the separation between the equipment and the affected device,
power the equipment from a source different from that of the affected
device,
OM-0481R1 27
Mobile X-ray Unit
Operation
ESSENTIAL PERFORMANCE
The system is designed to use X-rays for diagnostic purposes. The system (e.g.
Tube, Detector, Generator and Patient Support) is designed according to
international standards, to prevent patient, user, and others from electrical and
mechanical hazards by using adequate EMC measures like using filters,
screened cables or housings.
No unintended movement
No unintended X-radiation
28 OM-0481R1
Mobile X-ray Unit
Operation
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or user of this X-ray System should assure that it is used in such an environment.
RF emissions
Class A
CISPR 11
This Mobile Unit is suitable for use in all
establishments other than domestic and those
Harmonic emissions
Class A directly connected to the public low-voltage power
IEC 61000-3-2
supply network that supplies buildings used for
domestic purposes.
Voltage fluctuations/flicker emissions
Complies
IEC 61000-3-3
NOTE - In accordance with Standard IEC 60601-1-2:2014, the emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 Class A). If it is used in a residential environment (for which CISPR 11 Class B is normally required) this equipment might not offer
adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orientating the
equipment.
OM-0481R1 29
Mobile X-ray Unit
Operation
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or user of this X-ray System should assure that it is used in such an environment.
< 5 % UT
(> 95 % dip in UT) > 95 %
for 0.5 cycle for 0.5 periods
40 % UT
Mains power quality should be that of a typical
Voltage dips, short (60 % dip in UT) 60 %
commercial or hospital environment. If the user
interruptions and voltage for 5 cycles for 5 periods
of the Mobile Unit requires continued operation
variations on power supply
during power mains interruptions, it is
input lines.
recommended that the Mobile Unit be powered
70 % UT
from an uninterruptible power supply or a
IEC 61000-4-11 (30 % dip in UT) 30 %
battery.
for 25 cycles for 25 periods
< 5 % UT 100 %
(> 95 % dip in UT) for 250 periods
for 5s
NOTE -- UT is the a.c. mains voltage prior to application of the test level.
30 OM-0481R1
Mobile X-ray Unit
Operation
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or user of this X-ray System should assure that it is used in such an environment.
d = 1.2 P
Conducted RF 3 Vrms 3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz 150 kHz to 80 MHz d = 1.2 P , 80 MHz to 800 MHz
Radiated RF 3 V/m 3 V/m d = 2.3 P , 800 MHz to 2.5 GHz
IEC 61000-4-3 80 MHz to 2.5 GHz 80 MHz to 2.5 GHz
where ’P’ is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and ’d’ is the
recommended separation distance in meters
(m).
NOTE 1 -- At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which this Mobile Unit is used exceeds the applicable RF compliance level above, this Mobile Unit should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating this
Mobile Unit.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
OM-0481R1 31
Mobile X-ray Unit
Operation
This X-ray System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of this X-ray System can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and this X-ray System as recommended below,
according to the maximum output power of the communications equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance ’d’ in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where ’P’ is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 -- At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
NOTE 3 -- RFID chips are typically powered from the electromagnetic field, and therefore only the reader can be regarded as an RF
transmitter.
32 OM-0481R1
Mobile X-ray Unit
Operation
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or user of this X-ray System should assure that it is used in such an environment.
NOTE - UT is the a.c. mains voltage prior to application of the test level.
OM-0481R1 33
Mobile X-ray Unit
Operation
This X-ray System is intended for use in an electromagnetic environment specified below.
The customer or user of this X-ray System should assure that it is used in such an environment.
NOTE -- The ISM (Industrial, Scientific and Medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz
to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz; 3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7 MHz
to 7.3 MHz; 10.1 MHz to 10.15 MHz; 14 MHz to 14.2 MHz; 18.07 MHz to 18.17 MHz; 21.0 MHz to 21.4 MHz; 24.89 MHz to 24.99 MHz;
28.0 MHz to 29.7 MHz; and 50.0 MHz to 54.0 MHz.
34 OM-0481R1
Mobile X-ray Unit
Operation
This X-ray System is intended for use in an electromagnetic environment specified below.
The customer or User of this X-ray System should assure that it is used in such an environment.
Pulse modulation b)
380 -- 390 27
18 Hz
FM c)
430 -- 470 ¦5 kHz deviation 28
1 kHz sine
Pulse modulation b)
704 -- 787 9
217Hz
Pulse modulation b)
800 -- 960 28
18Hz 0.3
Pulse modulation b)
1700 -- 1990 28
217Hz
Pulse modulation b)
2400 -- 2570 28
217Hz
Pulse modulation b)
5100 -- 5800 9
217Hz
OM-0481R1 35
Mobile X-ray Unit
Operation
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user will
be required to correct the interference at his own expense. However, there is
no guarantee that interference will not occur in a particular installation.
36 OM-0481R1
Mobile X-ray Unit
Operation
Equipment:
Instrumentation used:
Dosimeters:
G Unfors Xi R/F
Test details:
OM-0481R1 37
Mobile X-ray Unit
Operation
Quantitative Information
Source-Image Distance
Time Current - mAs
Collimator blades
Output Dose
Voltage - kV
Input Dose
Focal Spot
(μGy/mAs)
Collimator
Selection
(μGy*m2)
Phantom
Phantom
opening
Time (s)
Patient
(μGy/s)
(cm)
(cm)
SID
examination
Grid
(orientative)
38 OM-0481R1
Mobile X-ray Unit
Operation
Deterministic effects may occur when the Radiation dose to a certain organ or
tissue exceeds a specific threshold. Particular organs or tissues of such
concern in diagnostic Radiology are the skin and the eye lens. The numerical
value of the threshold dose is in the range between 1 Gy and 3 Gy.
OM-0481R1 39
Mobile X-ray Unit
Operation
40 OM-0481R1
Mobile X-ray Unit
Operation
Control Panel, with controls to turn ON/OFF the system, System ON/OFF
Indicator, Battery Charge Level Indicator, Emergency Switch OFF.
Controls for the unit motion and controls for Telescopic Column and Arm
movements.
OM-0481R1 41
Mobile X-ray Unit
Operation
Illustration 3-1
Mobile X-ray unit: General Features
1.3 Head Assembly Screen (option) 4.2 Infrared Remote Control (option) 10 Detector Storage/Charge
1.4 Collimator Panel (option) 4.3 Barcode Scanner 11 Parking System
1.5 Motion Controls (on the Handgrips) 5 Detector Storage/Charge 12 Apron Hook
1
1.5
1.1
12
1.2
1.3
11
1.4
2 2.1
3 2.2
10
4.3
4.2
8
4.1
5
7
42 OM-0481R1
Mobile X-ray Unit
Operation
Illustration 3-2
Mobile X-ray unit
OM-0481R1 43
Mobile X-ray Unit
Operation
The Unit should be plugged into a wall socket compliant with local regulations
and electrical requirements of the equipment.
The plug is the device used as a means of disconnecting the Unit from mains.
Position the Unit so that the plug can be easily disconnected.
Plug
For safety reasons and for proper functioning, make sure that
the unit is connected to a standard outlet with GND.
44 OM-0481R1
Mobile X-ray Unit
Operation
Connect the Power Line Cable to the mains in order to allow the Charging
Circuits to charge the Batteries.
Note . When the Unit is ON and plugged into the mains, a “Lightning” icon
appears on the Control Console (next to the Batteries Indicators)
to display that the battery is charging.
This “Lightning” icon will disappear from the Control Console when
the Unit is unplugged from the mains.
Note . The Power Line Cable can only be replaced by the Service
Engineer.
OM-0481R1 45
Mobile X-ray Unit
Operation
1.2.4
1.2
1.3
1.4
46 OM-0481R1
Mobile X-ray Unit
Operation
" ON/OFF
The ON/OFF button icon enables the user to ON/OFF the unit.
The ON/OFF button with the RFID Card Reader symbol (RFID--Radio
Frequency IDentification) enables the user to ON/OFF and access the system
with the RFID Card.
The buttons with the symbols at the bottom of the ON/OFF Keypad, enable the
user:
The symbols at the bottom of the Battery Charge Level Indicators, show:
The Battery Level for the unit Motion (right column symbol).
OM-0481R1 47
Mobile X-ray Unit
Operation
3.2.1 SYSTEM ON
The ON/OFF controls are the means to turn the unit ON and OFF.
Note . The unit can operate in Stand-Alone mode, that is, operating
without mains being present or unplugged from mains.
In this case, restart the unit making sure not to touch any part
of the system (switches, touch screens, motion controls, or
any other control).
There are three ways to turn ON the system, depending on the system
configuration:
48 OM-0481R1
Mobile X-ray Unit
Operation
1. Press and hold the ON/OFF button (1) for a few seconds; the System
ON/OFF Indicator (2) will blink white for 5 seconds.
2. Within 5 seconds from turning the unit ON, pass the RFID Card (5) over
RFID Reader symbol (4), there is no need to contact the reader. The
system emits a double beep, the ON/OFF Indicator (2) will light cyan and
the Battery Charge Level Indicator (3) will light according to the current
battery status.
3. When the system is ON the System ON/OFF Indicator (2) will light white.
2
1
Note . The system emits 3-beeps and the System ON/OFF Indicator (2)
blinks orange to indicate a wrong RFID Card. In this case, repeat
the ON/OFF process using the right Card.
OM-0481R1 49
Mobile X-ray Unit
Operation
Depending on the system permissions assigned to each RFID Card, the user
can perform the following actions:
Movements:
When the system is out of the Parking position, it allows the user to
move the Column and the Arm in order to put the system in parking
position.
Panel Out:
It allows the user to unlock the system from the Parking position and
to move the Arm and the Column in order to get access to the
Detectors.
X-Rays:
Admin:
50 OM-0481R1
Mobile X-ray Unit
Operation
G Movements.
G Movements.
G Panel Out.
G Movements.
G Panel Out.
G User Management.
OM-0481R1 51
Mobile X-ray Unit
Operation
1. Press and hold the ON/OFF button (1) for a few seconds; the System
ON/OFF Indicator (2) will blink white for 5 seconds.
2. Enter the four-digit Access Code into the Keypad (3), within 5 seconds
from turning the unit ON, and press the OK button (4); the System
ON/OFF Indicator (2) will light cyan and the Battery Charge Level
Indicator (5) will light according to the current battery status.
3. When the system is ON, the System ON/OFF Indicator (2) will light white.
2 3
Note . The system emits 3-beeps and the System ON/OFF Indicator
Indicator (2) blinks orange to indicate that a wrong code has been
entered; in this case, repeat the ON/OFF process.
The system in the ON status switches ON the Generator, allowing the Mobile
motion and then, Konica-Minolta “Ultra” Software Application for Radiographic
operation will be launched on the Control Console.
52 OM-0481R1
Mobile X-ray Unit
Operation
Place the unit in the Parking position right after turning the
system OFF since positioning controls will remain enabled
for 15 seconds approximately.
Note . After turning OFF the unit, wait at least 10 seconds before turning
it ON again. This action assures a safely start-up of the system.
1. Turn OFF the system from the Control Panel, by holding pressed the
ON/OFF button for 3 seconds approximately.
2. The system ON/OFF Indicator will light cyan for a few seconds and then,
the system will be turned OFF.
Note . If the system does not turn OFF after pressing the ON/OFF button
for 3 seconds, try it again by holding pressed the ON/OFF button
for 12 seconds approximately; it will turn OFF the system safely.
OM-0481R1 53
Mobile X-ray Unit
Operation
In the event of an emergency, the unit is turned OFF by forcibly pressing this
switch (red mushroom-shaped switch). The switch is protected by the Cover,
in order to prevent it from being accidentally pressed.
The Emergency Stop must not be used to switch OFF the unit, in order to avoid
damaging the software.
Note . The user provided with RFID Card for Smart ON/OFF system, with
permission for Radiographic operation (refer to Section 3.2.1.1),
has direct access to the radiographic operation.
54 OM-0481R1
Mobile X-ray Unit
Operation
The system ON/OFF Indicator (1) is on the Control Panel, over the Battery
Charge Level Indicators.
SYSTEM OFF
The ON/OFF Indicator is not illuminated when the system is OFF.
OFF
USER REJECTED
The ON/OFF Indicator blinks orange indicating that the user has been
rejected by the system; e. g. when a wrong RFID Card has been passed,
ORANGE or when a wrong Code has been entered in the Keypad.
(Refer to Sections 3.2.1.1 and 3.2.1.2.).
SYSTEM ON/STANDBY
The ON/OFF Indicator lights white when the initialization process is
completed and the Positioner is operational, in Standby status.
WHITE
OM-0481R1 55
Mobile X-ray Unit
Operation
The Control Console houses the Main Screen which consists of a 19” capacitive
Touchscreen Monitor, with pinch and zoom functionalities.
The system can comprise a second Touchscreen Monitor (9”) located in the
Head-Assembly useful for radiographic parameters selection, technical
settings and patient information.
1 CONTROL CONSOLE
2 MAIN SCREEN
3 HEAD-ASSEMBLYSCREEN
1 2
56 OM-0481R1
Mobile X-ray Unit
Operation
The column with the Exposure symbol indicates the charge level of the Batteries
used for radiographic operations (X-ray exposures) and the column with the
Mobile X-Ray unit symbol indicates the charge level of the Batteries used for
the Mobile motion (motors).
When plugged into the mains (with the Emergency Switch-OFF deactivated),
the Batteries automatically charge. The color Indicators on both columns
illuminate and scroll from the current Generator battery charge level to 100%,
until the Batteries are fully charged. During the charging process, both columns
scroll up from the same level.
Note . Upon disconnecting the unit from the mains if the unit has been
connected for a short period of time, or after several exposures or
after one heavy duty exposure, the Batteries need at least
30 seconds to stabilize the charge, after which the correct charge
level is shown on the Indicator.
OM-0481R1 57
Mobile X-ray Unit
Operation
Full Charge:
Medium Charge:
When the system is at Medium Charge status, the lower
The lower 4 indicators on both columns (2+2) light steady indicators light green and normal operation is allowed.
green and the upper 2 indicators on both columns (1+1)
light blinking green and scrolling upwards, indicating that The batteries charge level is between 31% and 59%.
the system Batteries are at Medium Charge status.
Low Charge: When the system is at Low Charge status, the lower
indicator lights green steady, on one or both columns;
The lower 2 indicators on both columns (1+1) light steady normal operation is allowed although it is recommended
green and the upper 4 indicators light blinking green and to charge the system.
scrolling upwards, indicating that the system Batteries are
at Low Charge status. The batteries charge level is between 20% and 30%.
58 OM-0481R1
Mobile X-ray Unit
Operation
The Following Battery State Alerts can be displayed on the RAD screen:
OM-0481R1 59
Mobile X-ray Unit
Operation
If the indicator on the Motion column, indicates that motion is enabled for
30 minutes or 1 km after which, all the indicators will be OFF.
If the system is still unplugged when the countdown expires, the unit will
automatically shut down.
60 OM-0481R1
Mobile X-ray Unit
Operation
" Time Out Switch Off Pressed. Time set by default:: 3 seconds.
Press and hold the Switch OFF button for 3 seconds; the system will start the
turning OFF process.
" Time Out Delay Switch Off. Time set by default:: 50 seconds.
After holding pressed for 3 seconds the Switch OFF button, the system will be
completely turned OFF in 50 seconds; during this time only movements are
enabled (e. g. it is useful for the user to put the system in the Parking position,
if necessary).
When the system is disconnected from mains and there is not any Generator
activity during 2 minutes, the Generator Tube Filaments turn off, unless the user
performs any action on the RAD Screen, presses any control of the Handswitch
or Infrared Remote Handswitch, or connects the system to the mains; it will
delay the Time Out Inactivity Generator for another 2 minutes without any
Generator activity.
When the system is disconnected from the mains and there is not any
Generator activity for 15 minutes, a Warning message is displayed on the
screens to alert the user of an automatic shutdown after finishing a countdown
timer for another 15 minutes.
If the Operator does not respond within the countdown time of 15 minutes, the
system will automatically shut down; that is, the system will be completely
turned OFF after 30 minutes without activity.
OM-0481R1 61
Mobile X-ray Unit
Operation
The system can be provided with the Status Light Indicator (1), located at the
Head Assembly, which indicates different system status by colors, as described
in the following table.
62 OM-0481R1
Mobile X-ray Unit
Operation
COLORS
SYSTEM STATUS
NOT MOVING MOVING
SYSTEM STANDBY
X-rays are disabled.
The Status Light Indicator is OFF when the system is in Standby.
LIGHT OFF BLINKS WHITE
It blinks white when the system is moving.
EXPOSURE ON
The Status Light Indicator lights solid yellow during the X-ray Exposure: the Handswitch
--
has been fully pressed to the “Exp” status.
YELLOW
Movements are not allowed.
SYSTEM ERROR
The Status Light Indicator lights solid orange indicating that there is a system Error and it
blinks orange when it is moving.
ORANGE BLINKS ORANGE The Status Light Indicator blinks orange too when the system has been at a Battery Critical
Low status for 25 minutes, without being connected to mains.
SERVICE MODE
Only in Service Mode.
RED BLINKS RED
Note . For further information, refer to Section 3.4 about the Exposure
process and to Section 5 about System Messages.
OM-0481R1 63
Mobile X-ray Unit
Operation
4
2
5 1
9
3
3. USB Ports, to connect USB powered devices; only for Service personnel
(only for passive devices; a. e. Keyboard, Mouse, etc.).
64 OM-0481R1
Mobile X-ray Unit
Operation
Radiographic exposures are controlled with the “Prep” (preparation) and “Exp”
buttons on the X-ray Handswitch.
The status of the exposure is shown by the “Ready” and “X-ray On” indicators
for the duration of the exposure.
PREP: Press the Handswitch button half-way (“Prep” position) to prepare the
X-ray Tube for exposure. The “Ready” indicator on the Console lights when the
1 X-ray Tube is prepared and there are neither interlock failures nor system faults.
2
Note . The unit cannot perform exposures when the Column is in Parking
position; it only can perform exposures out of the Parking position.
EXP: After the “Ready” indicator is illuminated, fully press the Handswitch
button to start an X-ray exposure. If the button is released before the Generator
completes the selected time, the system aborts exposure and the actual mAs
and Exposure Time will be displayed.
The “X-ray On” indicator remains illuminated and a sound is emitted during the
exposure.
OM-0481R1 65
Mobile X-ray Unit
Operation
Note . The unit cannot perform exposures when the Column is in Parking
position; it only can perform exposures out of the Parking position.
When not in use, place the Infrared Remote Control back into the cradle in order
to charge the battery; the Status LED lighting cyan indicates that the Remote
Control has been placed in the correct position and its battery is charging (the
light will remain on as long as the IR Remote Control is stored in its cradle).
A Service Engineer may set the time it would take to display this
informational message, as well as whether it is accompanied by
an acoustic signal (discontinuous beeping).
2 Cradle
5
3 LED: Correct placement and
1 battery charge indicator
4 IR Emitter
5 IR Receiver Sensor
2
66 OM-0481R1
Mobile X-ray Unit
Operation
Take the Remote Exposure Control device out of its cradle. Aim the Remote
Control at the IR Receiver Sensor on the Mobile unit from a maximum
distance of 10 meters.
2
1 OFF / Prep / Exp
1
2 Collimator Light
3 Status LED
EXPOSURE CONTROL:
Press this button halfway to prepare the X-ray Tube for exposure (“Prep”
position). When the X-ray Tube is “Ready” (green light on, at the Control
Console), fully press and hold this button (“Exp” position) until the X-ray unit
completes the exposure (yellow light on, at the Control Console).
The status LED lights cyan in “Prep” and “Exp” positions.
When the exposure is completed the icon on the console turns OFF.
Return the Control Remote device back to its cradle on the Mobile unit.
OM-0481R1 67
Mobile X-ray Unit
Operation
68 OM-0481R1
Mobile X-ray Unit
Operation
Place the unit in the Parking position right after turning the
system OFF since positioning controls will remain enabled
for 15 seconds approximately.
OM-0481R1 69
Mobile X-ray Unit
Operation
3.6.1.1 HANDLEBAR
The unit is driven by holding the Handlebar with the hands, and it stops when
the Handlebar is released by the user.
The Handlebar is provided with internal sensors that control the direction and
speed of each wheel, operating when the Handlebar is touched and released.
70 OM-0481R1
Mobile X-ray Unit
Operation
Fine Positioning Controls permit the fine positioning adjustment of the system
respecting the patient, with the operator positioned opposite the
Head-Assembly.
The buttons with the arrow icons located on each Handgrip, control the motion
of each driving wheel independently, at low speed, backward/forward.
The buttons correspond to each motor and operate with the unit in/out of the
Parking Position.
Once any Fine Positioning button is pressed, the Unit starts to move, but
displacement will stop after 5 seconds. It is necessary to release the button and
press again to resume the movement. This safety measure prevents possible
uncontrolled movements due to an internal failure of the Fine Positioning
Controls.
1 1
1 Backward
2 Forward
2 2
OM-0481R1 71
Mobile X-ray Unit
Operation
The buttons correspond to each motor and do not change when the unit is in
Parking Position.
Illustration 3-3
Fine Positioning Controls
LB RF
RB LF
RB = Right Back
RF = Right Front
LF = Left Front
LB = Left Back
RB LB
RF LF
LB RB
LF RF
RB LF RF LB
72 OM-0481R1
Mobile X-ray Unit
Operation
When the unit has to be moved manually, the (x2) Clutch Screws (Allen type)
located on each wheel must be removed using the tools provided with the unit.
4. Magnetic Hubcap 1 2
4 3
Dismount the Separating Plate of the Back Holder, in order to gain access
to the Tools Holder mounted inside of the Unit.
Sticker
Separating
Plate
Tools
OM-0481R1 73
Mobile X-ray Unit
Operation
Use the Allen Wrench to remove the Clutch Screws (x2) of each wheel.
This will uncouple the wheels from the motors (releasing the brakes)
allowing the free motion of the unit.
The Proximity Sensors located under the Front Bumper, reduce the system
speed when detecting an obstacle, at a configured distance.
The system alerts the user with the Status Light Indicator blinking white.
Proximity Sensors
74 OM-0481R1
Mobile X-ray Unit
Operation
3.6.1.5 BUMPERS
The Front Bumper is provided with several sensors that stop the motor
movement in the event of a frontal collision.
The Status Light Indicator lights magenta when the Front Bumper is activated.
To do this, press and hold the release button while using the Handlebar to move
the Unit.
Illustration 3-4
Manual Driving Brake Release Push-Button
OM-0481R1 75
Mobile X-ray Unit
Operation
The unit is in Parking Position when the Parking Detent is secure in the Catch.
Fully retract the Telescopic Arm and turn the Column until the Parking
Detent is aligned with the Catch.
Lower the Arm and fully insert the Parking Detent into the Catch, until a
“click” is heard; it indicates that it has been properly placed in Parking
Position.
To release the Arm from Parking Position, press on the Brake Control at the
Head-Assembly.
Parking Detent
Catch
76 OM-0481R1
Mobile X-ray Unit
Operation
1 Handgrips
Hold the Handgrips to activate the Brake Control to move the Column and Arm
until the Head-Assembly is positioned. Release the control to lock in place.
OM-0481R1 77
Mobile X-ray Unit
Operation
Press the black buttons located on the Handgrips to rotate the Head-Assembly
from its vertical position:
¦180o on its transversal axis, i. e. roll rotation (see Illustration 3-5, A).
This movement has detents at --90o, 0o and +90o, and at --180o and 180o.
2 120_ on its horizontal axis, i. e. pitch rotation (see to Illustration 3-5, B).
This movement has detents at 0o , and it has stops at +90_, 0o and -- 30_.
Illustration 3-5
Head-Assembly and Collimator Movements
120o
A
¦180o
¦90o
C
78 OM-0481R1
Mobile X-ray Unit
Operation
The Collimator can rotate ¦90_ on its vertical axis (refer to Illustration 3-5,C),
while the Head-Assembly remains in the same position.
Required SID
X-ray Tube with Collimator rotated at:
Receptor Size
Anode Angle
0° or ¦90° ¦45°
24X30
SID 70 cm SID 70 cm
30X24
16° 35X43
43X35
SID 100 cm SID 100 cm
43X43
OM-0481R1 79
Mobile X-ray Unit
Operation
3
4
4 1 4
3 3
2
4
1
DAP Chamber
Front Knobs
Rear Knobs
Dual Laser
LED Light
Collimator Type
Manual Collimator with front and rear side Knobs, double Laser, Motorized Filters and ready for
p p p p p p p p
both Estimated Dose (EDAP) option and DAP Chamber option
80 OM-0481R1
Mobile X-ray Unit
Operation
Press once the Collimator Light push-button to turn the light ON; it will
remain ON for a few seconds before automatically switching OFF.
Press two times within 5 seconds to turn ON the Double Laser in order
to adjust the SID.
Press three times to turn OFF the light (if it remains ON).
2. Measuring tape.
Table 3-1
Image Size according to the SID and Collimator Opening
SID
COLLIMATOR
OPENING
90 cm (35.4”) 100 cm (39.4”) 180 cm (70.9”)
OM-0481R1 81
Mobile X-ray Unit
Operation
0 mm AL 1 mm Al + 0.1 mm Cu 1 mm Al + 0.2 mm Cu 2 mm Al
6. External Filtration.
There is a rail system at the rear side of the Collimator, with two guides,
in order to install the external additional filters used for pediatric
examinations (0.1 mm Cu or 3.5 mm Al) in the upper guide.
Press two times the Collimator Light push-button within 5 seconds to turn
ON the Double Laser in order to adjust the SID.
If the DAP Chamber is present, it is placed in the lower guide of the rail
system, at the rear side of the Collimator.
The external DAP Chamber (Radiation Meter) is installed in the lower guide of
the Collimator rails. It reads the radiation as Dose Area Product (DAP) in
mGy*cm2.
Note . Do not install any accessories between the DAP Chamber and the
patient. This will disturb the radiation reading.
82 OM-0481R1
Mobile X-ray Unit
Operation
The FSD (Focal-Skin Distance) sensor is installed at the front of the underside
of the Collimator. It reads the distance from the focus of the tube to the surface
of incidence on a patient, measured along the beam axis.
Illustration 3-6
FSD Sensors Location
FSD Sensor
Measuring Tape
OM-0481R1 83
Mobile X-ray Unit
Operation
Depending on the option, the Digital Detector can be stored and/or charged in
the Holders at the Back and/or at the Front Cover. The behavior regarding the
power supply of the Holders for Detectors charging is as follows:
Whenever one or two Detectors are being charged in the Unit’s Holders,
power is continuously supplied to the charging slots.
G Charging will be always continuous in the slot of the Back Holder (1).
Illustration 3-7
Wireless Digital Detectors, Options and Accessories
2
1
2
3
84 OM-0481R1
Mobile X-ray Unit
Operation
Wireless Digital Detectors communicate with the Mobile unit through the
internal Access Point.
OM-0481R1 85
Mobile X-ray Unit
Operation
Connect the Cable to the Detector and to the connector at the Front Cover (2).
When not in use, the Back-up Cable can be stored at the Detector Holders.
The Mobile unit can be equipped with a Barcode Scanner (1), for registering
patient information.
86 OM-0481R1
Mobile X-ray Unit
Operation
3.10.4 GENERAL USE AND MAINTENANCE OF WIRELESS DIGITAL DETECTORS, OPTIONS AND
ACCESSORIES
Note . If at any time the Detector does not connect properly after starting,
try again by turning on the Detector once the unit and the Image
Acquisition Software are operative.
During exposure, do not use the Detector near devices generating a strong
magnetic field.
After every examination, wipe with a cloth slightly damped the patient contact
surfaces as well as the Handle and Grid with disinfectants such as ethanol. For
cleaning, wipe with a cloth damped in neutral detergent.
Before using the Grid, clean the front and back side with a dry cloth to remove
dust and dirt.
Wireless Digital Detectors are prepared to fit into a Frame with a Removable
Grid. Consult the corresponding installation instructions and check that the Grid
is correctly mounted. Usually, a click sound means that the Grid is in place.
OM-0481R1 87
Mobile X-ray Unit
Operation
88 OM-0481R1
Mobile X-ray Unit
Operation
All controls, indicators and displays located on the Control Console are
positioned depending upon their functions.
Note . Use the operating controls as described in this manual. Any other
non-indicated combination may cause an incorrect operation.
For further details, refer to the User Manual of the Konica-Minolta “Ultra”
Software Application.
Illustration 4-1
Ultra Software Application
OM-0481R1 89
Mobile X-ray Unit
Operation
Illustration 4-2
Head-Assembly Console
Rotation and
angulation angles
of the X-ray Tube Workstation
Mute
User actions
Focal-Skin Distance
(Optional)
Collimator
Filter Options
Radiographic
Parameters
& Focal Spot
Message Bar
The “Status” icon of the Head-Assembly Console can vary according to the
operating status, as described below.
NORMAL STATUS: The detector is ready, the RAD technique is correctly set
and there is not Error or Interlock condition in the system.
READY: Indicates that the technique selected is properly set, there are no
interlock failures nor system faults, the anode is rotating and the X-ray Tube is
ready for exposure.
90 OM-0481R1
Mobile X-ray Unit
Operation
The “Heat Units” shows the percentage of utilized thermal capacity of the Tube.
For example, “25%” would indicate that 25% of Heat Units capacity is used
(it can be configured by the service engineer).
OM-0481R1 91
Mobile X-ray Unit
Operation
Full Charge (Stand--alone mode): The Battery icons are displayed in steady
green when the battery charge levels for X-ray and Motion are between 60%
and 100%, and the unit is unplugged from mains.
Full Charge (Charging mode): The Battery icons are displayed in steady
green when the battery charge levels for X-ray and Motion are between 60%
and 100%, and the unit is plugged to mains.
The “Lightning” icon indicates that the batteries are charging.
Medium Charge (Charging mode): The Battery icons are displayed in steady
green when the battery charge levels for X-ray and Motion are between 31%
and 59%, and the unit is plugged to mains.
The “Lightning” icon indicates that the batteries are charging.
Low Charge (Stand--alone mode): The Battery icons are displayed in steady
green when the battery charge levels for X-ray and Motion are between 20%
and 30%, and the unit is unplugged from mains.
Low Charge (Charging mode): The Battery icons are displayed in steady
green when the battery charge levels for X-ray and Motion are between 20%
and 30%, and the unit is plugged to mains.
The “Lightning” icon indicates that the batteries are charging.
Very Low Charge (Stand--alone mode): The Battery icons are displayed in
steady orange when the battery charge levels for X-ray and Motion are between
1% and 19%, and the unit is unplugged from mains.
The icons change the color from steady orange to steady green when the unit
is plugged to mains.
92 OM-0481R1
Mobile X-ray Unit
Operation
Very Low Charge (Charging mode): The Battery icons are displayed in
steady green when the battery charge levels for X-ray and Motion are between
1% and 19%, and the unit is plugged to mains.
The “Lightning” icon indicates that the batteries are charging.
Critically Low Charge (Stand--alone mode): The Battery icons are displayed
in blinking orange when the battery charge level for X-ray and Motion is 0%, and
the unit is unplugged from mains.
The icons change the color from blinking orange to steady green when the unit
is plugged to mains.
Critically Low Charge (Charging mode): The Battery icons are displayed in
steady green when the battery charge level for X-ray and Motion is 0%, and the
unit is plugged to mains.
The “Lightning” icon indicates that the batteries are charging.
Free Motion Speed: “Rabbit” icon appears when the Mobile Unit is in
movement and there is no obstacle.
OM-0481R1 93
Mobile X-ray Unit
Operation
Active when manual adjustments from the operator are required before making
the exposure (e.g. if the Grid is not inserted). If more than one action is required,
the number of actions to perform is shown in the icon.
Although the operator does not need to select any workstation as they are
always associated to an APR technique, a specific workstation may be selected
if needed.
Press on the desired Workstation to select it and return to the main menu by
pressing again on this icon or on the “Home” icon.
Illustration 4-3
Workstation Selection Window
94 OM-0481R1
Mobile X-ray Unit
Operation
Radiographic Parameters are divided into kV, mAs, mA, and Time
(seconds “s”).
Illustration 4-4
Radiographic Parameters Selection
kV shows:
mA shows:
The Time value (in seconds) selected for the radiographic technique.
OM-0481R1 95
Mobile X-ray Unit
Operation
When the “Slider” is positioned over a value not allowed, its pointer comes back
to the nearest allowed value, according to the limit of the Tube and the Unit.
Illustration 4-5
Slider for Value Selection
mA: Selects the X-ray Tube current, changing the mAs value and
keeping constant the select Exposure Time, whenever possible.
96 OM-0481R1
Mobile X-ray Unit
Operation
This indicator displays the selected Focal Spot of the X-ray Tube: “Small” or
“Large”. The Focal Spot is changed by pressing on this indicator. It keeps kVp
and constant mAs, whenever it is possible.
Small and Large Focal Spots can overlap each other, refer to the graphic below
to view an example of Focal Spot change.
Illustration 4-6
Small and Large Focus Overlap
Note . The maximum mA station for the Small Focal Spot and the
minimum mA station for the Large Focal Spot can be configured
by the field engineer during the installation.
OM-0481R1 97
Mobile X-ray Unit
Operation
This option displays the value of the current distance between focus and patient
measured by the Focal-Skin Distance sensor (refer to Section 3.9).
None.
2 mm Al
1 mm Al + 0.1 mm Cu
1 mm Al + 0.2 mm Cu
98 OM-0481R1
Mobile X-ray Unit
Operation
The Message Bar is located in the bottom area of the Main Menu of the
Head-Assembly Console.
Illustration 4-7
Message Bar
To check the message history, press on the Message bar. A pop-up window
(titled “Message List”) will be displayed. To close it, tap on the upper arrow of
the Message List window and go back to the previous screen.
Illustration 4-8
Message List
Tap to return to the Main Menu
Note . For information about the different message windows refer to the
Section 4.3.1.
OM-0481R1 99
Mobile X-ray Unit
Operation
The main message window is the Message List, available from the Message
Bar of the Main Menu, which contains all system messages.
Press on any message to display the date, code and a brief description.
Illustration 4-9
Message Displayed
Press the “Start” button in any of these windows to return to the Main Menu.
NOTIFICATIONS
100 OM-0481R1
Mobile X-ray Unit
Operation
Illustration 4-10
Notifications Pop-up Window
Illustration 4-11
Inhibit Conditions Messages
OM-0481R1 101
Mobile X-ray Unit
Operation
The “System Snapshot” button, located in the upper right corner of the
“Message List” screen, allows the user to generate event logs files, collecting
general information about the system status, which could be useful for the
Service personnel.
1. Press the “System Snapshot” button and confirm the pop-up message.
2. The “Logs Capture” window will be displayed indicating that the process
is in progress.
3. Tap on the arrow in order to expand this window and display the export
progress for each system component and the results. Tap it again in
order to hide this view.
4. To cancel the System Snapshot in progress, touch the “Close” button and
then confirm on the a pop-up window in order to return to the Main Menu.
Illustration 4-12
System Snapshot Confirmation and Logs Capture Window
3
“Close” Button
102 OM-0481R1
Mobile X-ray Unit
Operation
The following icons indicate a different status of the Logs Capture, for
each system component:
Note . When the Logs Capture process is not completed for one or more
components, the system continues with the processing.
5. Once the logs Capture process has finished successfully, touch the
“Close” button and the system will return to the Main Menu.
Illustration 4-13
Exiting the Logs Capture Window
Note . Only for Service purposes: the resulting system logs files are
generated in folder C:\OEM\Snapshots.
OM-0481R1 103
Mobile X-ray Unit
Operation
104 OM-0481R1
Mobile X-ray Unit
Operation
Inhibit condition. These messages are displayed when the Unit itself
or the Image Acquisition System request the Generator not to expose or
inhibit a movement because of multiple reasons: the tube is not pointing
to the detector, acquisition workstation is not ready, etc. But also, there
are originated by the same X-ray Generator. There are different kind of
Interlocks: those that cause an exposure inhibition, movements inhibition
or both at the same time.
OM-0481R1 105
Mobile X-ray Unit
Operation
Correct the error cause and keep touched the “Accept” button until the
Console indication disappears and exposure or procedure can be
accomplished.
All these System Messages are reported in the Message Bar, which is located
in the bottom area of the Console Main Menu. Active messages, i.e. those that
require action by the operator or report an error or warning, will be displayed
consecutively in this area.
Illustration 5-1
Message Bar
Note . For additional information about the Message Area refer to the
Section 4.3.
106 OM-0481R1
Mobile X-ray Unit
Operation
There are also different Message Windows, which are accessed depending on
the origin of the messages, in which detailed information about them can be
consulted (refer to Section 4.3.1).
The main message window is the Message List. To enter it, press on the
Message Bar. A pop-up window will be displayed. To close it, tap on the
message area again to go back to the previous screen.
Illustration 5-2
Message List
OM-0481R1 107
Mobile X-ray Unit
Operation
Error --
Current Image Receptor Synchronization Type is
100941
not supported. Please, select other workstation.
100944 Maximum Preparation Time Reached. Information Remove preparation and exposure order.
Inhibit
101008 INTERLOCK_2
Exposure
Inhibit
101018 Exposure does not allowed in this Desktop
Exposure
101019 The system has to be rebooted for being upgraded Information Press Accept and wait for the System to restart
Inhibit
101021 Inhibit Xray due to system communication lost --
Exposure
108 OM-0481R1
Mobile X-ray Unit
Operation
101022 Generator inactivity state, preheating filament Information Wait for 10 seconds.
Information --
102005 mAs requested out of range
102023 Number of exposure reached. Exposure finished. Information No user action required.
102024 Exposure finished due to limit mAs reached. Information No user action required.
OM-0481R1 109
Mobile X-ray Unit
Operation
Press Accept.
140301 Generator DSP Control. Register configuration Error If the error persists, restart the System.
If the error persists, call Service.
140302
140303
140305
140306
140313
Generator DSP Control.
Control Internal HW protection
140314
110 OM-0481R1
Mobile X-ray Unit
Operation
OM-0481R1 111
Mobile X-ray Unit
Operation
140342
Generator DSP Control. Overvoltage in super small
filament.
140343
Press Accept.
140350 Generator DSP Control. Master Heartbeat error.
Error If the error persists, restart the System.
If the error persists, call Service.
Generator DSP Control. Emergency signal
140351
activated by master.
112 OM-0481R1
Mobile X-ray Unit
Operation
140360
140361
140362
140363
140364
140365
140366
140367
140368
140369
140372
140373
140374
140375
140376
140377
140378
140379
140380
140381
OM-0481R1 113
Mobile X-ray Unit
Operation
140382
140383
140384
140385
140386
140387
140388
140389
140390
140391
Press Accept.
Accept
Generator DSP Control. Power Control Calibration
140392 Error If the error persists, restart the System.
data error.
Iff the error persists, call Service.
S
140393
140394
140395
140396
140397
140398
140399
140400
140401
140402
114 OM-0481R1
Mobile X-ray Unit
Operation
Press Accept.
Generator DSP Control. ROTOR_UDC
140420 Error If the error persists, restart the System.
undervoltage
If the error persists, call Service.
140501
140502
140503
140506
140509
140511
140512 Generator Starter powered Off Information Information - Generator Starter. System Powered Off.
Reset error.
140544 Wrong Dropout Station selection Error If the error persists, restart the System.
If the error persists, call Service.
OM-0481R1 115
Mobile X-ray Unit
Operation
Reboot is required.
140557 Generator Chopper Time Out. If the error persists, restart the System.
If the error persists, call Service.
140558
140559
140560
140561
140562
140563
140564
140565
140566
140567
Error
140568 Press Accept.
Generator Starter.
Starter Current Error.
Error If the error persists,
persists restart the System.
System
140569 If the error persists, call Service.
140570
140571
140572
140573
140574
140575
140576
140577
140578
140579
116 OM-0481R1
Mobile X-ray Unit
Operation
140580
140581
140583
140589
140590
Generator Starter.
Starter Current Error.
Error Press Accept.
Accept
140591 Error If the error persists, restart the System.
Iff the error persists, call Service.
S
140592
140593
140594
140595
Generator Starter. Tube 1 Thermal Switch Tube 1 Temperature. Interlock. - Generator Starter.
140604 Error
Interlock. Tube 1 Thermal Switch
OM-0481R1 117
Mobile X-ray Unit
Operation
Generator Starter. Tube Impedance Read Design Information - Generator Starter. Tube
140611 Information
Information. Impedance Read Value.
140612
Generator Starter Interlock.
Interlock Error Wait until the message dissapears.
dissapears
140613
290005
290006
Component Identification Mismatch Error Call Service.
Service
290007
290008
118 OM-0481R1
Mobile X-ray Unit
Operation
290104 System will power off: Incorrect system in license Error Call Service.
--
Inhibit
290200 Collision detected
Movement
290205 Collimator Apertures have not been calibrated Collimator Blades calibration is needed. Call Service.
The system has not been used for long time and
290400 will shutdown in 00:{0}:{1} unless you connect it to Warning
power source or click here.
--
OM-0481R1 119
Mobile X-ray Unit
Operation
Remote Handswitch is out of the placement longer Return the remote Handswitch to its placement or click
290402 Information
than configuration time on “Accept button” to disable the warning.
290454 Internal error located in the head Error Reboot the system and try to repeat the operation.
Inhibit
290705 Inhibit X-ray due to board configuration pending Call Service.
Exposure
Inhibit
290706 Inhbit X-ray due to System in Critical Battery Level If after charging the unit the error persists, call Service
Exposure
Inhibit
290707 Inhibit X-ray due to Dosimeter is not ready Call Service.
Exposure
Inhibit
290708 Inhibit X-ray due to Collimator filters are moving --
Exposure
Inhibit
290710 Inhibit X-ray due to tube fan is released Call Service.
Exposure
120 OM-0481R1
Mobile X-ray Unit
Operation
Dosemeter values have not been received from Try to repeat the operation.
290750
GPIO Head If the error persists, call Service.
Information
ExposureSoundOn Acknowledge Message has not
290751 If the problem persists, call Service
been received from RFID Board
291001
291002
291014
291015
System communication error located in the battery
monitor.
291016
291017
291018
291019
Peripheral communication error located in the
Error -
battery monitor.
291020
291021
291040
Peripheral communication error located in the Restart the System
y and tryy to repeat
p the operation.
p
E
Error
battery monitor. If the error persists, call Service.
291041
OM-0481R1 121
Mobile X-ray Unit
Operation
291050
291051
Restart the System
y and tryy to repeat
p the operation.
p
Internal error located in the battery monitor.
monitor Error
If the error persists, call Service.
291052
291053
291101 Download not performed due to a hardware failure Information Download not performed due to a hardware failure.
292001
292002
292004
292005
292014
292015
System communication error located in the RFID
RFID.
292016
292017
Restart the System
y and tryy to repeat
p the operation.
p
Error
If the error persists, call Service.
292023 Internal error located in the RFID.
292040
RFID Peripheral Communication Error
292041
122 OM-0481R1
Mobile X-ray Unit
Operation
293014 Motion Control: Right 1 Force Sensor error Restart the System.
Error
If error persists, call Service.
293015 Motion Control: Right 2 Force Sensor error
OM-0481R1 123
Mobile X-ray Unit
Operation
293051 Motion Control: Header key not working properly Warning Restart System. If error persists call Service.
294001
294002
294004
294014
294015
S t
System communication
i ti error llocated
t d iin th
the h
head.
d
294016
294017
124 OM-0481R1
Mobile X-ray Unit
Operation
294018
294019
Peripheral communication error located in the Restart the System and try to repeat the operation.
head. If the error persists, call Service.
294020
294021
294045
Internal error located in the head.
head
294046
294051
Internal error located in the head.
head
294052
294054
I t
Internal
l error located
l t d in
i the
th head.
h d
294059
OM-0481R1 125
Mobile X-ray Unit
Operation
300008 Image receptors configuration could not be loaded Error Call Service.
300012 Workstation mismatch for Generator and Positioner Try to select another workstation or another
radiographic procedure.
Error If the error persists, restart the System and try to
Active Procedure mismatch for Generator and repeat the operation.
300013
Positioner
If the error persists, call Service
300019 Generator working in Service mode Warning Please restart the System
300020 Positioner working in Service mode Warning Please restart the System
126 OM-0481R1
Mobile X-ray Unit
Operation
Inhibit
300032 Error loading Fluoro technique. Restart fluoro calibration.
Exposure
Waiting for generator operation mode to be If message does not disappear after the start up is
300037 Information
initialized... finished, call Service.
Waiting for positioner operation mode to be If message does not disappear after the start up is
300038 Information
initialized... finished, call Service.
132 OM-0483RD
Mobile X-ray Unit
Operation
128 OM-0481R1
Mobile X-ray Unit
Operation
OM-0481R1 129
Mobile X-ray Unit
Operation
4. Instruct patient to maintain the required position. Prepare the X-ray Tube
by pressing the Handswitch button to the “Prep” position and maintain it
until the “Ready” indicator is illuminated.
5. Instruct patient to remain still and to hold his breath as required, then
make the X-ray exposure by pressing the Handswitch button fully to the
“Exp” position and maintain it throughout the exposure. The “X-ray On”
indicator will light and an audible signal will sound during the exposure.
2. Center the Collimator light, which corresponds to the X-Ray beam, with
respect to receptor. For that, use the Collimator Light centering marks
and the laser line on the receptor handle if applicable.
130 OM-0481R1
Mobile X-ray Unit
Operation
4. Turn ON the Collimator Lamp and adjust the field size with the Collimator
controls.
Illustration 6-1
Patient Positioning
Laser Line
OM-0481R1 131
Mobile X-ray Unit
Operation
132 OM-0481R1
Mobile X-ray Unit
Operation
There are two levels of maintenance, the first consists of tasks which are
performed by the user/operator, and the second are those tasks to be
performed by qualified X-ray service personnel.
The first periodic maintenance service should be performed six (6) months after
installation, and the subsequent services at twelve (12) month intervals.
If the unit has not been used or it has been stored for two
months, it should be energized to prevent deep discharge of
the batteries. A deep discharge will cause permanent damage
to the batteries.
Recharge the batteries for at least 30 minutes at the beginning of the day
before using the unit.
Recharge the batteries for at least 30 minutes at the end of the day after
using the unit.
Fully recharge the batteries when the unit is going to be disconnected for
more than 3 weeks.
Fully recharge the batteries when the unit has been disconnected for
more than 3 weeks.
Keep the unit connected to the mains whenever possible to maintain the
batteries at the floating maintenance level, this increases their lifetime.
OM-0481R1 133
Mobile X-ray Unit
Operation
Do not allow the batteries to be deeply discharged because they will lose
storage capacity and will never be able to recover the 100% of their
original capacity.
The first periodic maintenance service should be performed six (6) months after
installation, and the subsequent services at twelve (12) month intervals.
1. With the Unit OFF, plug it in and leave it sufficient time to completely
charge. The recommended time is approximately 8 hours, until the
Battery Charge Level Indicators on both columns stop scrolling and all
the Indicators light Green.
2. Once fully charged, unplug the Unit from the mains power. Wait a few
minutes and reconnect the Unit to the mains. The upper Green Indicators
should scroll up for approximately one minute.
If the Battery charge level Indicators begin to scroll up from any other
Indicator below, contact the Service Department.
3. Switch the equipment OFF by shutting down the computer. Turn OFF the
switch ON/OFF control and unplug the unit from mains.
134 OM-0481R1
Mobile X-ray Unit
Operation
Clean the equipment frequently. Clean external covers and surfaces, especially
parts which might be in contact with patients, with a cloth moistened in warm
water with neutral soap. Wipe with a cloth moistened in clean water.
OM-0481R1 135
Mobile X-ray Unit
Operation
136 OM-0481R1
Mobile X-ray Unit
Operation
8.1 FACTORS
Maximum Power kW
(Refer to 20 kW 32 kW 40 kW 50 kW
Identification Label)
40 to 150
kVp Range
From 40 kVp to 150 kVp in 1 kVp steps
mAs Range Product of mA x Time values from 0.1 mAs to 500 mAs
10 to 400 10 to 500
Milliseconds: 1, 1.25, 1.6, 2, 2.5, 3.2, 4, 5, 6.3/6.4, 8, 10, 12.5, 16, 20, 25, 32, 40,
Exposure Time Range 50, 63/64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 630/640, 800.
Seconds: 1, 1.25, 1.6, 2, 2.5, 3.2, 4, 5, 6.3/6.4, 8, 10.
Maximum leakage radiation depends on the type of X-ray Tube (<0.88 mGy/h)
Manual Collimator with front and rear side Knobs, double Laser, Motorized Filters
Collimator
and ready for both Estimated Dose (EDAP) option and DAP Chamber option
OM-0481R1 137
Mobile X-ray Unit
Operation
Maximum Power kW
(Refer to 20 kW 32 kW 40 kW 50 kW
Identification Label)
Power Line Impedance must be less than the maximum indicated value:
300 m for 100 V~, 1 for 110 V~, 2.5 for 230 V~, 2.6 for 240 V~
Operation independent
from mains supply Standard
(Stand-Alone)
The required time for the Batteries to be 100% charged is approximately: 8 hours,
(80% of the total charge is available after 4 hours charging; approximately 20% per hour).
Battery Capacity Total Energy storage Capacity: 5760 Wh
With the Batteries fully charged and disconnected from the mains,
the Mobile system can be in continuous movement during approximately 25 km, at 5.5 km/h
Once the exposure capacity is exhausted, the system motion is enabled for 1 km.
The Mobile Unit in Stand-Alone (disconnected from the mains) will be 100% discharged from full charge
in approximately: 11 hours.
Radiation Output
Accuracy (Reproducibility C.V. (Coefficient of variation) ≤ 0.05
related to loading factors)
138 OM-0481R1
Mobile X-ray Unit
Operation
Maximum Power kW
20 kW 32 kW 40 kW 50 kW
(Refer to Identification Label)
X-ray Tube Inserts E7886 Low Speed / High Speed E7886 High Speed
OM-0481R1 139
Mobile X-ray Unit
Operation
Dimensions in mm.
Tolerance in Dimensions ±1%
140 OM-0481R1
Mobile X-ray Unit
Operation
120o
¦180o
¦90o
Dimensions in mm.
Tolerance in Dimensions ±1%
OM-0481R1 141
Mobile X-ray Unit
Operation
142 OM-0481R1
Mobile X-ray Unit
Operation
Please review the following link and reduce pediatric technique factors
accordingly: http://www.pedrad.org/associations/5364/ig/
As the use of Grids require higher doses, never use Grids in pediatric
exams. Remove the Grid from the receptor assembly and select the
lower possible doses. If the Grid cannot be detached, pediatric exams
cannot be performed using this device.
Appendix A A-1
Mobile X-ray Unit
Operation
Pediatric patients are not as likely as adults to understand the need to remain
still during the procedure. Therefore, it makes sense to provide aids to
maintaining stable positioning. It is strongly recommended the use of
immobilizing devices such as bean bags and restraint systems (foam
wedges, adhesive tapes, etc.) to avoid the need of repeating exposures due to
the movement of the pediatric patients. Whenever possible use techniques
based on the lowest exposure times.
Shielding:
Technique factors:
You should take steps to reduce technique factors to the lowest possible levels
consistent with good image acquisition.
For example, if your adult abdomen settings are: 70--85 kVp, 200--400 mA,
15--80 mAs, consider starting at 65--75 kVp, 100--160 mA, 2.5--10 mAs for a
pediatric patient. Whenever possible use high kVp techniques and large SID
(Source Image Distance).
Summary:
Use the lowest amount of radiation for adequate imaging based on size
of the child (reducing tube output -- kVp and mAs).
Try to use always short exposure times, large SID values and
immobilizing devices.
A-2 Appendix A
Mobile X-ray Unit
Operation
Antivirus protection:
Total AV
Norton by Symantec
PC Protect
Limit access to devices through the authentication of users (e.g. user ID and
password or smart card).
Appendix B B-1
Mobile X-ray Unit
Operation
We affirm our commitment to providing you with validated software updates and
patches as needed throughout the life cycle of the medical device to continue
to assure its continued safety and effectiveness.
Please promptly apply software updates and patches provided by us and never
use image management software supplied by anyone else. Our development
process utilizes the CISCO AMP protection. We are constantly scanning our
development computers for malware. We hope you are doing the same.
We perform daily backups to our external hard drives. The backups are
in another place.
CONCLUSION
It is our JOINT responsibility to ensure your medical image software and image
collection is safe and secure. We must both do our parts.
B-2 Appendix B
Mobile X-ray Unit
Operation
TABLE OF CONTENTS
Section Page
Appendix C C-1
Mobile X-ray Unit
Operation
C-2 Appendix C
Mobile X-ray Unit
Operation
C.1 INTRODUCTION
Note . This Appendix only applies to Units equipped with RFID ON/OFF
System.
Use the corresponding RFID Card to turn on the unit and access the software
with Administrator permissions. (Refer to System ON Section in the Operation
Manual for detailed information).
Appendix C C-3
Mobile X-ray Unit
Operation
C-4 Appendix C
Mobile X-ray Unit
Operation
Press and hold for a few seconds on the System Status icon, located at the
upper left corner of the Head-Assembly Control Console Screen.
After a few seconds, the Administrator Menu (Service Mode Protection utility)
screen will be displayed.
Appendix C C-5
Mobile X-ray Unit
Operation
OPERATOR SETTINGS
Refer to Section C.2.3.
Press the “Operator Settings” button to display the configurations related to the
system operation.
MAINTENANCE
Refer to Section C.2.4.
The following icons and buttons can be displayed in the Administrator Menu
main screen:
ENABLED FILAMENTS
This button indicates that filaments are enabled; radiation will be produced
during the exposure. Press on this icon to disable filaments.
DISABLED FILAMENTS
This button indicates that filaments are disabled; no radiation will be produced
during the exposure. Press on this icon to enable filaments.
BACK
Press on the “Back” button to leave the current menu. This button can be found
in every screen of the Administrator Menu to turn back to the Main Menu.
Note . Every time the “Back” button is used to leave the Administrator
Settings Menu, after making configuration changes, the System
requires a reboot to transfer the new data.
Press this button from the main menu window to return to the Control Console
in operation mode.
C-6 Appendix C
Mobile X-ray Unit
Operation
SAVE DATA
RESTORE DATA
NEXT STEP
Press to proceed to the next Calibration step or to skip the current one before
starting its Calibration.
PREVIOUS STEP
Note . All the other buttons and indicators on the upper toolbar (common
in the Operation Mode) are described in Section 4 of the Operation
Manual.
Appendix C C-7
Mobile X-ray Unit
Operation
Illustration C-1
Operator Settings
C-8 Appendix C
Mobile X-ray Unit
Operation
" LANGUAGE
Illustration C-2
Operator Settings: Culture Settings, Language Selection
Appendix C C-9
Mobile X-ray Unit
Operation
Use the “Measure Units” button to switch between the Metric system and the
American/Imperial system. Available options are “Centimeters” or “Inches”.
Illustration C-3
Operator Settings: Culture Settings, Measuring units
C-10 Appendix C
Mobile X-ray Unit
Operation
Use the “Time Zones” button to to select the relevant time zone. Press the
“Store” button to save the selected zone.
Illustration C-4
Time Zone Selection
Appendix C C-11
Mobile X-ray Unit
Operation
Use the “Time Units” menu to select “Milliseconds” or “Seconds” as the time unit
in the exposure parameters.
Illustration C-5
Operator Settings: Culture Settings, Time Units
C-12 Appendix C
Mobile X-ray Unit
Operation
The “Usability Settings” screen is used to set the Sound Settings, Visual
Settings, Anticollision Settings and the Power Off settings of the system.
Illustration C-6
Usability Settings
To set the brightness percentage and the colors for Standby and Prepared
status of the Mobile Unit.
Appendix C C-13
Mobile X-ray Unit
Operation
Press and hold the Switch OFF button for 3 seconds; the system will start the
turning OFF process.
After holding pressed for 3 seconds the Switch OFF button, the system will be
completely turned OFF in 50 seconds; during this time only movements are
enabled (e.g., it is useful for the user to put the system in the Parking position,
if necessary).
When the system is disconnected from mains and there is not any
Generator activity during 2 minutes, the Generator Tube Filaments turn
off, unless the user performs any action on the RAD screen, presses the
handswitch/remote control (option) or connects the system to the mains;
it will delay the “Time Out Inactivity Generator” for another 2 minutes
without any Generator activity.
When the system is disconnected from the mains and there is not any
Generator activity for 15 minutes, a Warning message is displayed on the
screens to alert the user of an automatic shutdown after finishing a
countdown timer for another 15 minutes.
C-14 Appendix C
Mobile X-ray Unit
Operation
To configure the time (in minutes) that the system will take to display an
informational message if the IR Remote Control is not placed in its holder.
Disabled by default.
Timeout: Seconds.
Allows to set the time it will take for the Unit to stop automatically when a
displacement is being performed using the Fine Positioning buttons.
Appendix C C-15
Mobile X-ray Unit
Operation
C.2.4 MAINTENANCE
Illustration C-7
Maintenance Options
C-16 Appendix C
Mobile X-ray Unit
Operation
Press the “Software / Hardware Versions” button to display the software and
hardware versions of all the system components.
Illustration C-8
Software / Hardware Versions Screen
Appendix C C-17
Mobile X-ray Unit
Operation
C.2.4.2 LICENSE
Press the “License” button to display all the system functionalities (Standard
and Extended) enabled by license.
Illustration C-9
License Screen
C-18 Appendix C
Mobile X-ray Unit
Operation
The purpose of the “System Logs” menu is to display the logs files generated
by the system, as well as to enable them to be exported.
Illustration C-10
System Logs Menu
Appendix C C-19
Mobile X-ray Unit
Operation
Counters
Illustration C-11
Counters
C-20 Appendix C
Mobile X-ray Unit
Operation
Exposures
Press this button to display the a detailed record of all the exposures
performed with the system.
Illustration C-12
Exposures
Errors
Press this button to display a detailed list with all the system messages
stored during system operation.
Illustration C-13
Errors
Appendix C C-21
Mobile X-ray Unit
Operation
Use this functionality to export the log files to a USB flash drive. (This
option is only available when an operating memory stick is connected to
a USB port).
Illustration C-14
Log Export
Select an stored log file by its date and press the “Export” button. When
the export process is finished, a confirmation message will indicate that
the files have been successfully copied to the USB drive.
C-22 Appendix C
Mobile X-ray Unit
Operation
The “Tag Admin” screen enables the local tags (RFID Cards) management
used to initiate the system and access it with the privileges assigned to them.
The available options in the “Tag Admin” window are “Add tag”, “Delete tag” and
“Show tags”.
Note . The tags (RFID Cards) managed in this menu only apply to the
local database of the Base Unit and are not synchronized with the
Acquisition Software.
Illustration C-15
Tag Admin Window
Appendix C C-23
Mobile X-ray Unit
Operation
1. Press on the “Add tag” button. Swipe the RFID card through the RCC
reader to register the tag, according to the instructions in the new window
displayed.
Illustration C-16
Add Tag Window
If the entered card is already included in the RFID tag record, the
following message will be displayed:
Illustration C-17
Tag Already Registered Message
C-24 Appendix C
Mobile X-ray Unit
Operation
2. In the next window, the tag name and the user’s access permissions, can
be defined.
Illustration C-18
Tag Edition Window
Depending on the system permissions assigned to the tag (RFID Card) in this
window, the user can perform the following actions:
Movements:
When the system is out of the Parking position, it allows the user to
move the Column and the Arm in order to put the system in parking
position.
Panel Out:
It allows the user to unlock the system from the Parking position and
to move the Arm and the Column in order to get access to the
Detectors.
X-Rays:
Appendix C C-25
Mobile X-ray Unit
Operation
Admin:
Illustration C-19
Registration Confirmation Message
C-26 Appendix C
Mobile X-ray Unit
Operation
DELETING A TAG
1. Press on the “Delete tag” button. Swipe the RFID card through the RCC
reader to unregister the tag, according to the instructions in the new
window displayed.
Illustration C-20
Delete Tag Window
If the entered card has not been previously registered in the RFID tag
record, the following message will be displayed:
Illustration C-21
Tag Not Registered Message
Appendix C C-27
Mobile X-ray Unit
Operation
Illustration C-22
Confirmation Window of the Tag Removal
3. Press the “Store” button to finish unregistering the tag. If the tag deletion
is successful, a message will be displayed in the previous window.
Illustration C-23
Tag Deletion Confirmation Message
4. Now it is possible to delete other tag or press the “Previous Step” button
to return to the Tag Admin menu.
C-28 Appendix C
Mobile X-ray Unit
Operation
EDITING TAGS
Press on the “Show tags” button to display a list of the different tags with the
name and permissions with which they were registered.
This window displays, for each tag, a text box with the name registered and the
permissions assigned to it (Movements, Panel out, X-rays, Admin, Service, RFID
Tag ID and Sync). Assigned permissions are represented with a green check mark,
while those that have not been granted are shown with a red ban icon.
Illustration C-24
Show Tags Window
The “Synchronization” icon is used to indicate that the tag is also registered in
the Acquisition Software.
The code displayed in the ”RFID Tag ID” would allow the RFID
Card to be identified in the Acquisition Software.
Appendix C C-29
Mobile X-ray Unit
Operation
Note . The radio button on the first row allows to select all the listed tags.
Illustration C-25
Removing Tag(s) - Selection and Trash Button
Illustration C-26
Removing Tag(s) - Confirmation Message
C-30 Appendix C
Mobile X-ray Unit
Operation
RENAMING TAGS
Illustration C-27
Renaming Tag
Illustration C-28
Renaming Tag(s) - Confirmation Message
Appendix C C-31
Mobile X-ray Unit
Operation
C-32 Appendix C