mt87 Digital Clamp Meter Manual
mt87 Digital Clamp Meter Manual
Contents
1 everwell MT87 Digital Clamp Meter
2 INTRODUCTION
3 GENERAL CHARACTERISTICS
4 ELECTRICAL SYMBOLS
5 PANEL DESCRIPTION
6 SPECIFICATIONS
7 CONTINUITY
8 OPERATION INSTRUCTION
9 BATTERY REPLACEMENT& AUTO POWER
OFF
10 ACCESSORIES
11 Documents / Resources
11.1 References
12 Related Posts
This manual provides all safety information, operation instruction, specifications and maintenance for the meter,
which is compact, handheld, and battery operated meter. This series instrument performs AC/DC voltage, AC
current, resistance, audible continuity, diode measurements. It is a 3 1/2 digits, 2000 counts digital clamp
multimeter.
It has the functions of polarity indication, data hold, maximum value hold, over range indication and automatic
power-off. It can be operated easily and is an ideal instrument tool.
The MT87 series digital clamp multimeter has been designed according to EN61010-1 Standard of oncoming
electronic measuring instruments with an over voltage category (CAT II 600V) and pollution degree
Warning
To avoid possible electric shock or personal injury, and to avoid possible damage to the meter or to the equipment
under test, adhere to the following rules:
Before using the meter inspect the case. Do not use the meter if it is damaged or the case (or part of the case)
is removed. Inspect for cracks or missing plastic pieces. Pay attention to the insulation around the connectors.
Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity.
Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any
terminal and grounding.
The rotary switch should be placed in the right position and no any changeover of range shall be made during
measurement is conducted to prevent damage of the meter.
When the meter is working at an effective voltage over 60V in DC or 36V rms in AC, special care should be
taken for there is danger of electric shock.
Use the proper terminals, function, and range for your measurements.
Do not use or store the meter in an environment of high temperature, humidity, explosive, inflammable and
strong magnetic field. The performance of the meter may deteriorate after dampened.
When using the test leads, keep your fingers behind the finger guards.
Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity or
diodes.
Replace the battery as soon as the battery indicator □ appears. With a low battery, the meter might produce
false readings that can lead to electric shock and personal injury.
Remove the connection between the testing leads and the circuit being tested, and turn the meter power off
before opening the Meter case.
When servicing the meter, use only the same model number or identical electrical specifications replacement
parts.
The internal circuit of the meter shall not be altered at will to avoid damage of the meter and any accident.
Soft cloth and mild detergent should be used to clean the surface of the meter when servicing. No abrasive and
solvent should be used to prevent the surface of the meter from corrosion, damage and accident.
The meter is suitable for indoor use.
Turn the meter power off when it is not in use and take out the battery when not using for a long time.
Constantly check the battery as it may leak when it has been using for some time, replace the battery as soon
as leaking appears.
GENERAL CHARACTERISTICS
ELECTRICAL SYMBOLS
PANEL DESCRIPTION
SPECIFICATIONS
Accuracy is guarantied for 1 year 23±5°C (73.4±41 °F) less than 80%RH
DC VOLTAGE
AC CURRENT
Range Resolution Accuracy
20A 1mA
200A 10mA
±(2.5% of rgd + 10dgts)
400A 1A
AC VOLTAGE
RESISTANCE
Open Circuit Voltage: about 0.5V Overload Protection: 250V DC/AC rms
CONTINUITY
The built-in buzzer will sound if the resistance is less Open circuit
OPERATION INSTRUCTION
Measuring Voltage
1. Connect the BLACK test lead to the “COM” jack and the RED to the “VO “jack.
2. Set Function/Range Switch to the desired V or V range.
3. Connect the test leads across the source or load to be measured.
4. Read LCD display. The polarity of the RED lead connection will be indicated when making a DC measurement.
Note: To avoid damage to the meter, do not measure a voltage that exceeds 600Vdc or 600Vac.
Measuring AC Current
1. Set Function/Range Switch to the 600A range. If the display indicates one or more leading zeros. Shift to the
below small range to improve the resolution of the measurement.
2. Press the trigger to open the transformer jaws and clamp one conductor only it is impossible to make
measurements when two or three conductors are clamped at the same time.
3. Display reading is the flow of the conductor AC current.
Measure Resistance
1. Connect the BLACK test lead to the “COM” jack and the RED to the “ ” jack (Note: The polarity of the
red test lead is positive “+”).
2. Set Function/Range Switch to 200KO range.
3. Connect the test leads across the load to be measured.
4. Read the reading on the display.
Note:
a. When the input is not connected. i.e. at open circuit, the symbol ‘OL” will be displayed as an over range
indicator.
b. Before measuring in-circuit resistance, be sure that the circuit under test has all power removed and all
capacitors are fully discharged.
Continuity Test
1. Connect the BLACK test lead to the “COM” jack and the RED to the “ ” jack (Note: The polarity of the red
test lead is positive “+”).
2. Set Function/Range Switch to range, LCD shows “OL”
3. Press the “HOLD” Button to select continuity test mode, and the symbol “ ” will appear as an indicator.
4. Connect the test leads across the load to be measured.
5. If the circuit resistance is lower than about 500, the built-in buzzer will sound.
Diode Test
1. Connect the BLACK test lead to the “COM” jack and the RED to the “ ” jack (Note: The polarity of the
red test lead is positive “+”).
2. Set Function/Range Switch to range, LCD shows “OL”, Meter displays measurement diode function by
default, the symbol “ ” will appear as an indicator.
3. Connect the red test lead to the anode of the diode to be measured and the black test leads to the cathode.
4. The forward voltage drop in mV will be displayed. If the diode is reversed, figure “OL” will be shown.
Continuity Test
1. Push the “HOLD” button when measuring, data will be held, and the “H” signal will appear. Data won’t be
refreshed until this button is pushed again. Then “H” will disappear, and quit the state of data hold.
2. The data hold and continuity test function use the same button, So the diode and continuity test data cannot be
held.
1. Power off the instrument within 15 minutes without operation, There will be an audible warning before the
instrument automatically shuts down.
2. If the sign “ “appears on the LCD, it indicates battery should be replaced. Remove screws and open the
back case, replace the exhausted battery with new batteries (Size AAA, 1.5V x2 or equivalent).
ACCESSORIES
WARRANTY
This Instrument is warranted to be free from defects in material and workmanship for a period of one year. Any
instrument found defective within one year from the delivery date and returned to the factory with transportation
charges prepaid, will be repaired, adjusted, or replaced at no charge to the original purchaser. This warranty does
not cover expendable items such as batteries & fuses. If the defect has been caused by a misuse or abnormal
operating conditions, the repair will be billed at a nominal cost.
Documents / Resources
References
User Manual
This website is an independent publication and is neither affiliated with nor endorsed by any of the trademark owners. The "Bluetooth®" word mark and logos are registered trademarks owned by Bluetooth
SIG, Inc. The "Wi-Fi®" word mark and logos are registered trademarks owned by the Wi-Fi Alliance. Any use of these marks on this website does not imply any affiliation with or endorsement.