Unit 2 Chromatography Techniques
Unit 2 Chromatography Techniques
Chromatography terms
1
Chromatogram is the visual output of the chromatograph. In the case of
an optimal separation, different peaks or patterns on the chromatogram
correspond to different components of the separated mixture.
Chromatograph is equipment that enables a sophisticated separation,
e.g. gas chromatographic or liquid chromatographic separation.
Chromatography is a physical method of separation that distributes
components to separate between two phases, one stationary (stationary
phase), the other (the mobile phase) moving in a definite direction.
Eluate is the mobile phase leaving the column. This is also called
effluent.
Eluent is the solvent that carries the analyte.
Eluite is the analyte, the eluted solute.
Immobilized phase is a stationary phase that is immobilized on the
support particles, or on the inner wall of the column tubing.
Mobile phase is the phase that moves in a definite direction. It may be a
liquid (LC and Capillary Electrochromatography (CEC)), a gas (GC), or a
supercritical fluid (supercritical-fluid chromatography, SFC). The mobile
phase consists of the sample being separated/analyzed and the solvent
that moves the sample through the column. In the case of HPLC the
mobile phase consists of a non-polar solvent(s) such as hexane in normal
phase or a polar solvent such as methanol in reverse phase
chromatography and the sample being separated. The mobile phase
moves through the chromatography column (the stationary phase) where
the sample interacts with the stationary phase and is separated.
Preparative chromatography is used to purify sufficient quantities of a
substance for further use, rather than analysis.
Retention time is the characteristic time it takes for a particular analyte
to pass through the system (from the column inlet to the detector) under
set conditions.
Sample is the matter analyzed in chromatography. It may consist of a
single component or it may be a mixture of components. When the
sample is treated in the course of an analysis, the phase or the phases
containing the analytes of interest is/are referred to as the sample whereas
everything out of interest separated from the sample before or in the
course of the analysis is referred to as waste.
Solute refers to the sample components in partition chromatography.
Solvent refers to any substance capable of solubilizing another substance,
and especially the liquid mobile phase in liquid chromatography.
2
Stationary phase is the substance fixed in place for the chromatography
procedure. Examples include the silica layer in thin layer chromatography
Detector refers to the instrument used for qualitative and quantitative
detection of analytes after separation.
1. Paper chromatography
2. Thin layer chromatography
3. Gel chromatography
4. Ion-exchange chromatography
5. Affinity chromatography
6. Hydrophobic interaction chromatography
7. Gas liquid chromatography
8. High pressure/ performance liquid chromatography (HPLC)
1. Paper chromatography
3
separate readily according to how strongly they adsorb onto the stationary phase
versus how readily they dissolve in the mobile phase.
The retardation factor (Rƒ) may be defined as the ratio of the distance
traveled by the substance to the distance traveled by the solvent.
Some compounds in a mixture travel almost as far as the solvent does;
some stay much closer to the base line. The distance travelled relative to the
solvent is a constant for a particular compound as long as you keep everything
else constant - the type of paper and the exact composition of the solvent, for
example.
The distance travelled relative to the solvent is called the R f value. The Rf
value (retention factor) is the distance travelled by a particular component from
the origin (where the sample was originally spotted) as a ratio to the distance
travelled by the solvent front from the origin. For each compound it can be
worked out using the formula:
Procedure
4
ii. The paper is then dipped into a suitable solvent, such as ethanol or water,
taking care that the spot is above the surface of the solvent, and placed in a
sealed container.
iii. The solvent moves up the paper by capillary action, which occurs as a result
of the attraction of the solvent molecules to the paper and to one another. As
the solvent rises through the paper it meets and dissolves the sample mixture,
which will then travel up the paper with the solvent. Different compounds in
the sample mixture travel at different rates due to differences in solubility in
the solvent, and due to differences in their attraction to the fibers in the
paper. Paper chromatography takes anywhere from several minutes to
several hours.
iv. In some cases, paper chromatography does not separate pigments
completely; this occurs when two substances appear to have the same values
in a particular solvent. In these cases, two-way chromatography is used to
separate the multiple-pigment spots. The chromatogram is turned by ninety
degrees, and placed in a different solvent in the same way as before; some
spots separate in the presence of more than one pigment.
v. After development, the spots corresponding to different compounds may be
located by their colour, ultraviolet light, ninhydrin or by treatment with
iodine vapours. The final chromatogram can be compared with other known
mixture chromatograms to identify sample mixture using the Rf value.
5
Types of Paper Chromatography
=========================================================
6
Retention Factor (Rf)
Apparatus
Plates
7
hydrocarbons. It is also important to note that silica gel is acidic. Therefore,
silica gel offers poor separation of basic samples and can cause a deterioration
of acid-labile molecules. This would be true for alumina plates in acidic
solutions as well. It is important to note that there are differences between silica
gel and alumina. Alumina is basic and it will not separate sample sizes as large
as silica gel would at a given layer thickness. Also, alumina is more
chemically reactive than silica gel and as a result, would require more care of
compounds and compound classes. This care would avoid decomposition and
rearrangement of the sample.
Stationary Chromatographic
Typical Application
Phase Mechanism
steroids, amino acids, alcohols,
Silica Gel adsorption hydrocarbons, lipids, aflaxtoxin, bile,
acids, vitamins, alkaloids
fatty acids, vitamins, steroids, hormones,
Silica Gel RP reversed phase
carotenoids
Cellulose, carbohydrates, sugars, alcohols, amino
partition
kieselguhr acids, carboxylic acids, fatty acids
Aluminum amines, alcohols, steroids, lipids,
adsorption
oxide aflatoxins, bile acids, vitamins, alkaloids
nucleic acids, nucleotides, nucelosides,
PEI cellulose ion exchange
purines, pyrimidines
Magnesium
adsorption steroids, pesticides, lipids, alkaloids
silicate
8
If you know that one component of a mixture is insoluble in a given solvent, but
another component is freely soluble in it, it often gives good separations. How
fast the compounds travel up the plate depends on two things:
If the compound is soluble in the solvent, it will travel further up the TLC
plate
How well the compound likes the stationary phase. If the compound likes
the stationary phase, it will stick to it, which will cause it to not move
very far on the chromatogram.
Acids, bases, and strongly polar compounds often produce streaks rather
than spots in neutral solvents. Streaks make it difficult to calculate an Rf and
may occlude other spots. Adding a few percent of acetic or formic acid to the
solvent can correct streaking with acids. Similarly for bases, adding a few
percent triethylamine can improve results. For polar compounds adding a few
percent methanol can also improve results.
Pipettes
Spots are applied to the plate using very thin glass pipettes. The capillary
should be thin enough to apply a neat spot, but not so thin as to prevent
the uptake of an adequate quantity of analyte. Here is a popular method of
producing TLC pipettes.
9
Developing chamber or vessel
Procedure
1. Cut the plate to the correct size and using a pencil (never ever use a pen),
gently draw a straight line across the plate approximately 1 cm from the
bottom. Do not use excessive forces when writing on a TLC plate as this
will remove the stationary phase. It is important to use a pencil rather
than a pen because inks commonly travel up the plate with the solvent.
An example of how black ink separates is shown in the section labeled
"examples".
2. Using TLC pipettes, apply spots of analyte to the line. Make sure enough
sample is spotted on the plate. This can be done by using the short-wave
UV. A purple spot should be seen. If the spot is not visible, more sample
needs to be applied to the plate. If a standard of the target compound is
available, it is good practice to produce a co-spot by spotting the standard
onto a spot of the unknown mixture. This ensures the identity of the target
compound.
3. Place the plate into the chamber as evenly as possible and lean it against
the side. Never allow the bulk solvent to rise above the line you drew.
Allow capillary action to draw the solvent up the plate until it is
approximately 1 cm from the end. Never allow the solvent to migrate all
the way to the end of the plate.
4. Remove the plate and immediately draw a pencil line across the solvent
front.
5. Use a short-wave UV light and circle the components shown with a
pencil.
10
Visualizing
Iodine is among the most common stains. Plates are placed in a jar
containing iodine crystals, or covered in silica gel with iodine dispersed
throughout, for approximately one minute. Most organic compounds will be
temporarily stained brown. Some popular general use stains are Permanganate,
ceric ammonium molybdate (CAM), and p-anisaldehyde. These can be kept in
jars which plates are dipped into, or in spray bottles.
To develop a plate with permanganate, spray or dip the plate and heat it
with a heat-gun. Hold the plate face up 10 to 20 cm above the heat gun until the
bulk water evaporates. Then move the plate to 5 to 10 cm above the heat gun
and heat it until white/yellow/brown spots appear. Overheating will turn the
entire plate brown, obscuring the spots. If glass plates are used it is often easier
to see spots through the backing because it is harder to overheat.
11
Chromatogram of TLC plate
=========================================================
12
Gel-filtration chromatography
Discovery
The technique was invented by Grant Henry Lathe and Colin R Ruthven,
working at Queen Charlotte’s Hospital, London. While Lathe and Ruthven used
starch gels as the matrix, Jerker Porath and Per Flodin later introduced dextran
gels;[9] other gels with size fractionation properties include agarose and
polyacrylamide
SEC is used primarily for the analysis of large molecules such as proteins
or polymers. SEC works by trapping smaller molecules in the pores of the
adsorbent materials adsorption ("stationary phases"). This process is usually
performed with a column, which consists of a hollow tube tightly packed with
extremely small porous polymer beads designed to have pores of different sizes.
These pores may be depressions on the surface or channels through the bead. As
the solution travels down the column some particles enter into the pores. Larger
particles cannot enter into as many pores. The larger the particles, the faster the
elution. The larger molecules simply pass by the pores because those molecules
are too large to enter the pores. Larger molecules therefore flow through the
13
column more quickly than smaller molecules, that is, the smaller the molecule,
the longer the retention time.
One requirement for SEC is that the analyte does not interact with the
surface of the stationary phases, with differences in elution time between
analytes ideally being based solely on the solute volume the analytes can enter,
rather than chemical or electrostatic interactions with the stationary phases.
Thus, a small molecule that can penetrate every region of the stationary phase
pore system can enter a total volume equal to the sum of the entire pore volume
and the interparticle volume. This small molecule elutes late (after the molecule
has penetrated all of the pore- and interparticle volume—approximately 80% of
the column volume). At the other extreme, a very large molecule that cannot
penetrate any the smaller pores can enter only the interparticle volume (~35% of
the column volume) and elutes earlier when this volume of mobile phase has
passed through the column. The underlying principle of SEC is that particles of
different sizes elute (filter) through a stationary phase at different rates. This
results in the separation of a solution of particles based on size. Provided that all
the particles are loaded simultaneously or near-simultaneously, particles of the
same size should elute together.
Each size exclusion column has a range of molecular weights that can be
separated. The exclusion limit defines the molecular weight at the upper end of
the column 'working' range and is where molecules are too large to get trapped
in the stationary phase. The lower end of the range is defined by the permeation
limit, which defines the molecular weight of a molecule that is small enough to
penetrate all pores of the stationary phase. All molecules below this molecular
mass are so small that they elute as a single band.
The filtered solution that is collected at the end is known as the eluate.
The void volume includes any particles too large to enter the medium, and the
solvent volume is known as the column volume.
The basic components of the gel filtration experiment are the matrix,
chromatography column and the elution buffer. The matrix is the material in
the column that is actually the separation medium. It is the stationary phase of
the chromatography.
The column is a tube with a frit and elution spout fitted at the bottom. The
frit is a membrane or porous disk that supports and retains the matrix in the
column but allows water and dissolved solutes to pass. The elution buffer is the
mobile phase of the chromatography and flows through the matrix and out of
the column. The column, with the matrix and applied sample, is ―developed‖ by
the elution buffer. This means that the molecules in the sample are carried by
the flow of buffer into the matrix where they are gradually separated. The
14
separated zones of molecules then flow out of the column where they are
collected for analysis Filling the chromatography column with matrix is referred
to as ―packing‖. The packed matrix is called the ―bed‖ and the volume it
occupies is termed the ―bed volume‖.
The gel filtration matrix consists of microscopic beads that contain pores
and internal channels. The larger the molecule, the more difficult it is for it to
pass through the pores and penetrate the beads. Larger molecules tend to flow
around and in between the beads. The total volume of buffer between the beads
is the ―void volume‖. Smaller molecules tend to spend more time in the maze of
channels and pores in the bed. Consequently, the larger, higher molecular
weight molecules are eluted from the column before smaller molecules. Larger
molecules take the faster, more direct path that involves less time in the beads
(Figure )
Analysis
15
purification), other techniques may be necessary to identify the contents of each
fraction.
=========================================================
Ion-exchange chromatography
16
Ion exchange mechanism
Mobil phases consist an aqueous buffer system into which the mixture to
be resolved. The stationary phase usually made from inert organic matrix
chemically derivative with ionizable functional groups (fixed ions) which carry
displaceable oppositely charged ion. Ions which exist in a state of equilibrium
between the mobile phase and stationary phases giving rise to two possible
formats, anion and cation exchange are referred to as counter ion (Figure).
17
mixture is applied to the top of the column and allowed to pass into the bed of
ion- exchange material.
In this process two types of exchangers i.e., cationic and anionic exchangers
can be used.
18
Stationary phase
19
of an organic solvent are used in which most of the ionic compounds are
dissolved better than in others in.
Sodium chloride is probably the most widely used and mild eluent for
protein separation due to has no important effect on protein structure.
The salt mixture can itself be a buffer or a separate buffer can be added to
the eluent if required. The competing ion which has the function of eluting
sample components through the column within reasonable time is the essential
component of eluting sample.
Commonly used eluent additives which have been successfully used in ion
exchange chromatography can be given as follow;
Buffer
Buffers
PH range
Ammonium acetate 4 to 6
Ammonium formate 3 to 5
Pyridinium formate 3 to 6
Pyridinium acetate 4 to 6
Ammonium carbonate 8 to 10
20
Detection
For example anions that absorb in the UV spectral region can be detected
spectrophotometrically. In this case, an eluent anion is selected that does not
absorb UV.
Practical procedure:
Ion exchange separations are carried out mainly in columns packed with
an ion-exchanger.
First the column is filled with ion exchanger then the sample is applied
followed by the buffer. The tris-buffer, pyridine buffer, acetate buffer,
citrate and phosphate buffers are widely used.
The particles which have high affinity for ion exchanger will come down
the column along with buffers. In next step using corresponding buffer
separates the tightly bound particles.
Then these particles are analyzed spectroscopically.
21
Applications of Ion Exchange Chromatography:
========================================================
22
Affinity chromatography
Principle
23
The other molecules in the mobile phase will not become trapped as they do not
possess this property. The stationary phase can then be removed from the
mixture, washed and the target molecule released from the entrapment in a
process known as dialysis. Possibly the most common use of affinity
chromatography is for the purification of recombinant proteins.
Molecule Ligand
Antigen Antibody
Enzyme Substrate
Receptor Ligand
Nucleic Acid Binding Nucleic
Protein Acid
Polysaccharide,
Lectin
glycoprotein
Procedure
1. Sample preparation
The sample must be a clear solution free from solid particles. This can be
achieved by centrifugation or filtration. Protein solutions should be centrifuged
at at least 10000 g. Cell lysates should be centrifuged at 40-50000 g. A 0.45-µm
pore size filter can be used for filtration.
24
One must also consider how the solubility and stability of the sample or the
desired protein can be influenced by the pH, the salt concentration, or the
presence of any organic solvent. The factors affecting the interactions between
the desired target protein and the matrix-bound ligand (pH, salt concentration,
temperature) should also be determined. The composition of the initial binding
buffer must be adjusted accordingly.
Sample components interfering with the target protein and/or the ligand (e.g.
metabolites in cell lysates) should be removed before loading onto the column.
Elution via pH and/or ionic strength changes: One possible and simple
means of elution is achieved through decreasing the interaction strength
between the ligand and the target protein. Changes in the pH will change the
ionisation state of charged groups of the ligand and/or the target protein, thereby
changing the strength of the interaction. Similarly, increasing the ionic strength
(usually by raising the NaCl concentration) will generally reduce the interaction
strength. In either case, the solubility and stability of the target protein should
be considered.
Competitive elution: For competitive elution, materials are applied that react
with the target protein or the ligand, competing for the pre-existing interaction.
For instance, His-Tag fusion proteins can be readily displaced from the metal
chelate matrix by imidazole buffer
In all cases, the flow rate of the buffer should be reduced during
elution, thereby avoiding excessive dilution of the target protein.
25
=======================================================
The more hydrophobic the molecule, the less salt needed to promote
binding. To elute the proteins, the salt concentration is gradually decreased,
eluting bound species in order of increasing hydrophobicity. Additionally,
elution can also be achieved through the use of mild organic modifiers or
detergent.
26
The stationary phase is designed to form hydrophobic interactions with other
molecules. These interactions are too weak in water. However, addition of salts
to the buffer result in hydrophobic interactions. The following is a list of salts
that increase hydrophobic interactions in the order of their ability to enhance
interactions:
1. Na2SO4
2. K2SO4
3. (NH4)2SO4
4. NaCl
5. NH4Cl
6. NaBr
7. NaSCN
========================================================
27
Gas-liquid chromatography
The Instrument
The sample to be reacted is injected into the gas stream just before it
enters the column.
The components of the mixture are then carried through the column in a
stream of gas.
Each compound distributes itself between the phases to different extents
and therefore emerges from the column at a different time.
Some of the compounds dissolve in the stationary solvents more readily
than others; these travel through the column slower and so emerge last.
The most volatile compounds usually emerge first.
A detector on the outlet tube monitors compounds emerging from the
column.
Signals from the detector are plotted out by a recorder as a
chromatogram
The chromatogram shows the recorder response against the time which
has elapsed since the sample was injected into the column.
28
Each component of the mixture gives rise to a peak on the chromatogram
========================================================
29
High-performance liquid chromatography (HPLC)
Introduction
Principle
30
Instrumentation / Procedure
31
The velocity of each component depends on its chemical nature, on the
nature of the stationary phase (column) and on the composition of the mobile
phase. The time at which a specific analyte elutes (emerges from the column) is
called its retention time. The retention time measured under particular
conditions is an identifying characteristic of a given analyte.
Mobile phase
The chosen composition of the mobile phase (also called eluent) depends
on the intensity of interactions between various sample components ("analytes")
and stationary phase (e.g. hydrophobic interactions in reversed-phase HPLC).
Depending on their affinity for the stationary and mobile phases analytes
partition between the two during the separation process taking place in the
column.
32
Applications of HPLC
========================================================
33