AS138-005 Rev.
11/15
®
CBC-3D
for MINDRAY/ACON
HEMATOLOGY CONTROLS
CONTROL
ASSAY VALUES AND EXPECTED RANGES
LOT B0525
2025-08-05
Instrument Parameter Low Normal High
LOT B0525L LOT B0525N LOT B0525H
3
Mindray BC-3600 WBC 10 /uL 1.9 ± 0.8 7.4 ± 1.1 21.5 ± 2.9
6
CD version RBC 10 /uL 2.32 ± 0.24 4.44 ± 0.34 5.74 ± 0.44
1.8(FDA) or greater HGB g/dL 5.9 ± 0.7 13.9 ± 1.0 19.7 ± 1.3
HCT % 17.5 ± 1.5 40.1 ± 2.0 54.8 ± 2.4
MCV fL 76.6 ± 5.0 86.9 ± 5.0 95.3 ± 5.0
MCH pg 25.7 ± 2.5 29.3 ± 2.5 33.6 ± 2.5
MCHC g/dL 33.3 ± 3.0 34.6 ± 3.0 35.4 ± 3.0
RDW-CV 15.1 ± 3.0 14.2 ± 3.0 14.1 ± 3.0
3
PLT 10 /uL 64 ± 30 256 ± 60 523 ± 80
MPV fL 7.3 ± 3.0 7.6 ± 3.0 7.5 ± 3.0
LYM % 60.9 ± 9.0 31.8 ± 8.0 15.1 ± 8.0
MONO % 11.0 ± 8.0 5.8 ± 4.0 4.2 ± 3.0
GR % 30.1 ± 9.0 63.1 ± 8.0 81.1 ± 9.0
3
LYM x 10 /uL 1.1 ± 0.2 2.4 ± 0.7 3.0 ± 1.6
3
MONO x 10 /uL 0.2 ± 0.2 0.4 ± 0.3 0.8 ± 0.6
3
GR x 10 /uL 0.6 ± 0.2 4.8 ± 0.7 16.4 ± 1.9
3
MISSION HA-360 WBC 10 /uL 2.3 ± 0.8 7.8 ± 1.1 21.5 ± 2.9
6
RBC 10 /uL 2.43 ± 0.24 4.61 ± 0.34 5.99 ± 0.44
HGB g/dL 6.4 ± 0.9 14.0 ± 1.3 19.8 ± 1.8
HCT % 19.1 ± 1.5 41.6. ± 2.0 58.3 ± 2.4
MCV fL 77.8 ± 5.0 87.9 ± 5.0 96.9 ± 5.0
MCH pg 26.6 ± 2.5 29.2 ± 2.5 32.9 ± 2.5
MCHC g/dL 34.6 ± 3.0 33.6 ± 3.0 33.6 ± 3.0
*RDW-CV 15.4 ± 3.0 14.8 ± 3.0 13.5 ± 3.0
3
PLT 10 /uL 82 ± 30 250 ± 50 520 ± 80
*MPV fL 8.4 ± 3.0 7.8 ± 3.0 7.3 ± 3.0
LYM % 53.5 ± 12.0 29.8 ± 8.0 15.3 ± 7.0
*MONO % 15.0 ± 10.0 7.2 ± 6.0 7.6 ± 6.5
GR % 31.5 ± 9.0 63.0 ± 8.0 77.1 ± 10.0
3
LYM x 10 /uL 1.1 ± 0.3 2.4 ± 0.7 3.4 ± 1.5
3
*MONO x 10 /uL 0.3 ± 0.2 0.6 ± 0.5 1.4 ± 1.4
3
GR x 10 /uL 0.7 ± 0.2 4.8 ± 0.7 16.5 ± 2.2
*For new software version B0285H/01.07.00.
For technical assistance in the USA and Canada call Technical Service at (800) 523-3395.
All brands and products are trademarks or registered trademarks of their respective companies.
R&D Systems, Inc.
614 McKinley Place NE
Minneapolis, MN 55413 USA
AS138-005 Rev. 11/15