7.5 Riken Oxygen Analyser
7.5 Riken Oxygen Analyser
7.4.3
1. Summary
Riken portable oxygen indicator, Model OX-226 and OX-227 provide a quick, convenient
method for determination of oxygen content of any atmosphere. It is intended primarily as an
indicator of oxygen deficiency, with good readability from 0 – 25%. The instrument is routinely
calibrated on normal atmospheric oxygen concentration (21%). These models are most
suitable and recommended for testing tanks, manholes, vessels and other spaces to determine
safety from the standpoints of oxygen deficiency before entering and while work is in progress.
2. Principle
The oxygen cell operates by an electro-chemical process in which a voltage is set up between
two electrodes. Under a test where one electrode is exposed to the atmosphere, a change in
oxygen concentration on this electrode produces a proportional change in the cell’s output
voltage.
Therefor, an increase in oxygen concentration will "speed up" the electro-chemical process,
producing a higher output voltage, and a decrease in oxygen concentration will "slow down"
the process, lowering the output voltage. The centre electrode is exposed to the atmosphere
by means of a Teflon membrane placed directly in contact with the polished top surface. This
Teflon membrane serves two functions simultaneously. First, it has the ability to pass oxygen
molecules freely, thus placing the electrode in direct contact with the atmosphere and
secondly, it keeps the electrolyte contained in the cavity between the two electrodes.
3. Measurement procedure
a). Preparation - Connect the sampling hose (6) to the gas sampling probe (7) and then
connect it to the gas inlet of the instrument.
b). Voltage checks of battery - Turn the control switch (1) to "Batt" zone and check the meter
needle marks inside of "Batt" zone. If the case of model OX-226, the battery drop can be
heard as a buzzer sound.
c). Span adjustment - Turn the control switch (1) to "25" and make span adjustment by
spanning adjusting knob so as to bring the meter needle to 21%. When making span
adjustment of Model OX-227, try it with 0-25% range.
4. Measurement
After finishing the above procedure items 1, 2 and 3, the instrument is ready to run. Introduce
the sampling probe to the source and start measurement. In the case of Model OX-226, when
the oxygen concentration is less than 18% by volume, alarm light (4) illuminates and it gives
us the warning of oxygen deficiency by buzzer sound.
Caution
1. 1. Check the flow pump by the flow monitor during operation.
2. 2. Operate the instrument in leather case when in use.
3. 3. The replacement of batteries and recharging procedure must be done in non-
hazardous areas.
Maintenance procedure
The replacement of batteries and recharging procedure:
a). Take off the leather case from the instrument and turn the battery box knob (11) to "open"
position.
b). Pull out the whole battery box and replace the batteries with new ones.
c). When the replacement of batteries is finished, put back the battery box in correct position
and turn the battery box knobs (11) to "Lock" position with finger press.
Replacement of sensor.
When the meter needle can not be adjusted to 21% by turning the span adjusting knob and
the indication of meter needle gets unstable, this is the sign to replace the sensor. In this
case, take off the bottom screws of the instrument and remove the cover. The cover comes off
by sliding it sidewise. Turn the sensor to left and adjust the mark to "open". Now the sensor
can be removed. Insert the new sensor and turn it in clockwise direction to the mark "lock".
Place the cover back.
Replacement of filter
The filters are filled in the gas-sampling probe and in instrument. When they appear dirty,
replace them with new ones.
Take off the tip of the sampling probe by turning the metal part of roulette and replace the
cotton filter with a new one.
Pull out the filter holder (10) of the instrument’s flank and take out the filter. Replace it with a
new one.
Zero adjustment
As the zero adjustment is factory set, there is no need of zero adjustment procedure in normal
operation. But, when it is high sensitive type instrument such as Model OX-227A with 0-5 and
0-25% etc., make zero adjustment. Induct 100% clean nitrogen and turn the adjusting screw to
bring the needle to zero.
7.4.4
7.4.5
7.4.6 7.4.11 Detector tubes for health hazardous
gases
Health hazardous gases may be detected through chemical colour reactions, and several
manufacturers make metering pumps and accompanying detector tubes for a great number of
various gases.
Probably the most convenient and suitable equipment to use for measuring very low
concentrations of toxic gases on board tankers are chemical indicator tubes.
These tubes consist of a sealed glass tube containing a proprietary filling which is designed to
react with a specific gas and to give a visible indication of the concentration of that gas. To
use the device, the seals at each end of the glass tube are broken, the tube is inserted in a
bellows-type fixed volume displacement hand-pump, and a prescribed volume of gas mixture
is drawn through the tube at a rate fixed by the bellow’s expansion rate. A colour change
occurs along the tube and the length of the discoloration, which is a measure of the gas
concentration, is read off a scale integrated with the tube. In some versions of these
instruments, a hand operated injection syringe is used instead of a bellow pump.
It is important that all the components used for any measurement should be from the same
manufacturer. It is not permissible to use a tube from one manufacturer with a hand pump
from another manufacturer. It is also important that the manufacturers’ operating instructions
are carefully observed.
Since the measurement depends on passing a fixed volume of gas through the glass tube, if an
extension hose is used it should be placed between the glass tube and the hand pump.
The tubes are designed and intended to measure concentrations of gas in the air. Thus
measurements made in a ventilated tank, in preparation for tank entry, should be reliable.
Under some circumstances errors can occur if several gases are present at the same time, as
one gas can interference with the measurement of another. The manufacturer should be
consulted for guidance.
For each type of tube the manufacturer must guarantee the standards of accuracy laid down
by national standards. Tanker operators should consult the regulatory authority appropriate
for the ship’s flag.
The bellow is pressed together completely and is then released. During the compression the air is
squeezed out of the bellow through an exhaust valve. The suction action of the pump takes place
when the compression springs inside the bellow expand after the compression. The air (to be
measured) flows through the sampling tube and into the bellow while this again expands to its
original volume. The suction movement comes to an end when the distance chain is tight once
again, and at this stage 100cm3 has been sucked through the tube.
The operating instructions, which accompany each packet of tubes, give i.e. the approximate
time for each pump stroke, for example 15 - 25 seconds. The time will depend on how tightly
the powder is packed in the tube.
The specified number of pump strokes, indicated in the operating instructions, should be used
for each sampling tube.
7.4.8 7.4.13 MSA – Detector Tubes and Kwik-
draw Pump
Features
1. 1. Quick and inexpensive to use.
2. 2. A reliable method of testing more than 120 hazardous gases and vapours.
3. 3. Kwik-draw pumps offer accurate one-handed automatic stroke counter and unique
end of stroke indicator on deluxe version.
Description
SA’s Kwik-Draw and Kwik-Draw Deluxe Pumps can be used with an assortment of MSA detector
tubes to spot-test the atmosphere for a wide variety of toxic substances. Kwik-Draw Pumps
are designed for one-hand operation and consistent delivery of a sample draw volume of 100
millilitres (ml). The pumps are constructed with a shaft-guided compression system for a more
consistent and replicable flow rate and volume per stroke than may be available with hand-
guided pumps. MSA offers detector tubes for measuring more than 150 gases and vapours.
Detector tubes.
MSA/Auer detectors are made of glass, have break-off tips and are filled with treated chemical
granules for sampling a variety of substances. Most MSA/Auer detector tubes are packaged 10
in a box.
For ordering information, see the Detector Tube Summary Chart which follows the Detector.
After selecting the appropriate tube, the user would break off the tubes’ end tips and attach
the tube to the sampling pump. After air is drawn through the tube by the pump, the chemical
layer in the tube changes color if the test gas or vapor is present in the air.
The length or shade of the color-change indicates the concentration of the gas or vapor in the
air. A scale is printed on each tube for interpretation of data.
3. 3. The manufacturer of tubes and pumps must operate under a certified quality
assurance program.
Description - The Kwik-Draw Pump is a one-handed, manually operated bellow pump of 100cc
capacity.
Tube Holder - This rubber part permits mounting of detector tubes, remote sampling lines or
other detectors.
Filter Disc - This porous plastic disc mounted in the rubber tube holder protects the pump
from dirt and dust particles, which may alter the flow or damage the pump.
Exhaust Valve - Located under the valve cover, this valve closes as the bellow re-inflates, and
readily opens on the exhaust stroke so that blow-back through the tube holder is negligible.
Stroke counter - For convenience, a stroke counter is incorporated into the pump handle.
End-of-stroke indicator - As the bellow begins to re-inflate, and after the knob is released,
the indicator eyeball turns high visibility green. As the vacuum decreases, the eye begins to
roll back to black. The stroke is over when the eye is all black.
Note! - Kwik-Draw Pump (part no. 488543) does not have an end-of-stroke indicator.
Operation
1. 1. Using the breaker on the pump, break off both tips of the detector tube.
2. 2. Using a twisting motion, insert the tube into the rubber tube holder. The arrow on
the tube should point toward the pump.
Note! If your pump does not have the end-of-stroke indicator, wait 30 seconds after full
bellow inflation to ensure that all 100cc of the sample has been drawn through the tube. The
detector tube must be held in the sampling area during this period.
7. 7. To evaluate the stain, follow the instructions provided with the detector tubes.
Remote sampling
Remote sampling is accomplished by putting the pump, connecting tube, remote sampling line
and detector tube together, in this order.
Maintenance
Under conditions of normal use, this pump should require little maintenance. Depending on
the frequency of use, periodic cleaning and checks for correct performance as recommended.
Tube holder - Replace tube holder when it shows signs of wear or loss of elasticity. If filter is
not clogged or cracked, save the filter disc for re-use in new tube holder.
Filter disc - Periodically remove the filter disc for cleaning or replacement.
Remove filter disc from tube holder by rolling flange part of tube holder down and
away from the disc.
Gently tap or blow on the surface to remove any foreign matter.
Replace disc so previously exposed surface is once again facing away from pump.
Shaft
If shaft becomes dirty or if bellow inflation is jerky, remove shaft by unscrewing, then clean
with auto wax.
Valves
With the valve cover removed, check the valves for dirt or debris.
Remove dirt with a gentle puff of air or by using a soft brush.
Replace valve(s) if necessary.
Warning! Use of a pump that leaks may result in the under-estimation of a hazard
and could result in property damage, injury or death.
ELECTRICAL CHARACTERISTICS
Accuracy Power supply life Power supply Range
PHYSICAL CHARACTERISTICS
Sample flow rate Construction Dimensions Weight
0.03 to o.o5 cfm Plastic case with 6-1/2 x 7-1/4 x 4 in. 5 lb 2 oz.
(0.8 to 1.4 Lpm). stainless steel
hardware.
Operation:
Introduction
This section provides instructions on unpacking and inspecting the Gascope, making initial
checks, operating the instrument and using the Gascope for various applications.
Initial inspection
Remove the instrument from its shipping container and examine the unit carefully. If damage
or shortage is noted, advise the carrier promptly. Make the proper claim with the carrier and, if
necessary, reorder from MSA.
Initial checks
To verify that the instrument is operating properly and retains factory calibration, perform the
procedure, which follows in an atmosphere free of combustible gases. If the proper indications
cannot be obtained, contact the local sales office or return the instrument to Mine Safety
Appliances Co., Repair and Customer Service Department, Walden Road, Mars, PA 16046.
Open cover and set RANGE switch to LEL.
Set ON/OFF switch to ON, Ready indicator should turn on within approximately 4 seconds.
BATT indicator pointer should be at least halfway into white zone.
Squeeze aspirator bulb eight to ten times to purge instrument with fresh air. Permit bulb to
inflate completely after each squeeze.
Lift and adjust LEL ZERO control to obtain zero indication on meter. (To make zero
adjustments, lift and turn outer sleeve on Zero controls).
Set Range switches to GAS. The READY indicator should momentary turn off and then turn on
within approximately 4 seconds.
Lift and adjust GAS ZERO control to obtain zero indication on meter.
Connect source of 100% methane to inlet fitting of instrument. Pass gas through
instrument and then shut off flow. Meter should indicate 78 to 86 on an instrument
factory calibrated on pentane and at least 95 if calibrated on methane.
WARNING
Do not introduce 100% methane gas to an instrument located near a source of ignition,
otherwise, an explosion may occur.
Repeat step 3 and then set RANGE switch to LEL.
Connect flow control of Calibration Check Kit, Model R, to Calibration Check Gas Cylinder
(0.75% pentane) and connect adapter hose between flow control and inlet fitting of
instrument. For hose instruments calibrated on methane, use 2 % methane cylinder.
Turn flow control valve counter-clockwise to pass gas through instrument. Turn valve clockwise
and, when needle stabilises, meter should indicate between 45 and 55 on an instrument
factory calibrated on pentane and between 37 and 43 if calibrated on methane.
Turn flow control valve clockwise, disconnect hose from inlet fitting and remove flow control
from gas cylinder. Squeeze aspirator bulb eight to ten times to purge instrument with fresh air.
CAUTION
When sampling over liquids, take care that end of sampling line does not touch surface of
liquid. A closed-end probe tube or line trap should be used in this type of test to prevent liquid
from being drawn into instrument. Sampling dusty or smoky atmospheres may clog the flow
system. If either the speed of response of the instrument decreases or the aspirator bulb does
not inflate within 2 seconds, check the flow system as described under Sampling System
Checks in the instruction book (Section 4).
If the meter indicates less than 5, set RANGE switch to LEL and draw sample into instrument by
squeezing aspirator bulb. The meter now indicates the percentage of the lower explosive limit
(LEL) of the gas for which the instrument is calibrated. One of the following conditions may
exist, depending on the deflection of the meter needle:
In atmosphere free of combustible gases, squeeze aspirator bulb eight to ten times to
purge instrument with fresh air. Set ON/OFF switch to OFF, or simply close cover.
WARNING
Do not use the instrument on the LEL range in oxygen deficient (less than 10% oxygen) areas;
the meter may not indicate the actual gas concentration.
The instrument is not designed to test mixtures of hydrogen, acetylene or other combustible
gases, in which the oxygen content exceeds that of normal air (oxygen – enriched
atmospheres). The Gascope is unsafe for use in measuring concentrations of acetylene or
hydrogen in pure oxygen unless specially modified by MSA; however, it can be used to
measure hydrogen or acetylene in air mixtures. It is not suitable to test high boiling point
hydrocarbons which have vaporised in ovens and which will condense in the instrument flow
system at room temperature. In addition, it will not indicate the presence of combustible mists
or sprays such as lubricating oil, or dusts as grain or coal dusts.
Even though the instrument responds to such combustible gases as propane, acetylene and/or
gasoline or solvent vapours, it provides accurate measurements of the specific gas for which it
has been calibrated at the factory. The indications on both ranges are affected by all
combustibles and also may be affected by non-combustibles, due to their thermal conductivity
effects.
CAUTION
Do not use the instrument at locations where compounds such as tetraethyl lead or hydraulic
fluids or lubricants, which contain silicates or silicones, may be present in the atmosphere.
These chemicals contaminate the sensor and thus reduce measurement accuracy. If it is
suspected that such compounds may be present in the test area, check the calibration of the
instrument after making a maximum of five measurements.
When testing atmospheres contaminated with leaded gasoline, replace the standard cotton
filter in the inlet of the instrument with an inhibitor filter; otherwise, the catalytic sensor may
become coated with lead after several measurements, causing sensitivity to be lost.
Sampling Lines
Sampling lines permit samples to be taken at remote locations or inaccessible areas
to test the atmosphere in such spaces as manholes, sewers and bar-holes. The
available sampling lines, which range in length from 5 to 50 feet, are made of
synthetic material specially compounded to resist absorption of combustible vapors.
CAUTION
Do not use sampling lines made of ordinary rubber or any synthetic material which absorbs
solvent vapours since the absorption will result in erroneous indications that are usually lower
than the actual value.
The shortest possible length of sampling line should always be used to minimize the number of
times the aspirator bulb must be squeezed to obtain a valid indication.
To test the atmosphere at a remote location, position the male coupling of the sampling line at
the desired sampling point so that dirt particles do not clog the tube or liquids are not drawn
into the instrument. Connect the coupling on the other end of the line to the inlet fitting of the
instrument, (fig. 2-2 in the instruction book). Make sure that the connection is gas-tight. The
concentration of the gas at the sampling point can then be measured by performing the
procedure under OPERATING THE GASCOPE.
Probes:
Probes permit samples to be taken in areas that cannot be reached with a sampling line. Bar-
holes, manholes and sewers behind obstructions or accessible only through narrow openings
can be examined by connecting the probe to a sampling line, (see fig. 2-2 in the instruction
book). Three probes are available for use with the Gascope; a four-foot, solid probe rod; a
three-foot, hollow-brass probe tube; and a three-foot, dielectric probe tube. The probe rod may
be used to prevent liquids from entering the instrument if the open end of a sampling line
inadvertently dips into the liquids in tanks etc.
Warning
Do not use the brass probe where shock hazards exist; that is, where contact may be made
with electrical equipment or power lines. The high dielectric plastic probe, which does not
conduct electrical current, should be used in these areas.
Line Trap:
The line trap is used primarily to prevent liquids from being inadvertently drawn into the
instrument when the atmospheres in tanks, sewers or sumps are sampled. (Refer to the sheet
supplied with the line trap for instructions on how to install and use the trap).
Mixtures in the rich region, which extend from the UEL to 100% gas, contain too much gas in
relation to air to be combustible. However, since the addition of air to these high
concentrations of gas mixture creates mixtures in the flammable region, they must be
considered equally dangerous.
Low-Range Operation.
The Gascope utilizes the flammable property of the combustible gas for which it is calibrated
to measure its concentration in air below the LEL of the mixture.
The instrument contains a catalytic platinum filament which comprises one arm of a bridge
circuit (see the block diagram in the instruction book). This hot wire filament is activated to
permit combustible gas to combine with oxygen at a much lower temperature than would be
required for normal combustion. In the event that the flammable mixture is drawn into the hot-
wire filament well where ignition can occur, the flashback arresters prevent propagation of the
flame.
When the combustible gas in the atmosphere enters the catalytic filament well and makes
contact with the catalyst on the surface of the filament, a combustible gas-oxygen reaction
occurs which increases the temperature of the filament. This increase in temperature causes a
corresponding increase in filament resistance, which unbalances the bridge. The resultant
bridge signal voltage is read out on the meter as the percentage of LEL.
High-Range operation.
To measure the concentration of combustible gas in air above the LEL of the mixture, the
Gascope employs a heated thermal conductivity (TC) filament which comprises one arm of a
second bridge circuit (see the block diagram in the instruction book). This filament, a
catalytically inert wire that has a high temperature coefficient of resistance, is specially treated
to ensure stable measurements. When combustible gas in the atmosphere enters the TC
filament well and passes over the filament, it lowers the temperature of the filament. This
decrease in temperature causes a corresponding decrease in filament resistance, which
unbalances the bridge. The resultant bridge signal is read out on the meter as the percentage
of gas in the atmosphere between 0 and 100% by volume.
If the RANGE switch is now set to the LEL position, the meter and READY indicator circuits are
de-energized for approximately 4 seconds to allow the catalytic filament to warm up and
stabilize. After the stabilization period, the READY indicator turns on to signify that the
instrument cam be used.
If the sample now drawn into the catalytic filament well contains no combustible gas, the
bridge remains balanced and the meter indicates zero. If the sample contains up to the LEL of
the combustible gas, the amplitude of the bridge signal is proportional to its actual
concentration. The resultant signal is read out on the meter as the percentage of the LEL of the
mixture. If the sample contains between the LEL and UEL of the gas, the bridge remains
unbalanced and the resultant signal causes the meter needle to deflect and remain off-scale. If
the concentration of gas exceeds the UEL, the bridge initially becomes unbalanced and the
needle deflects off-scale. The needle then deflects rapidly downscale because the gas absorbs
heat from the filament, increasing its resistance.
The zero adjustments are used to null the bridge signals for both ranges so that the meter
indicates zero when air free of combustible gases is drawn into the instrument. The span
adjustment are used to set the meter indications on both ranges to a given concentration of
calibration gas.
The voltage regulator circuit maintains a constant 0.94 V across the bridges, even as the
battery voltage decreases from its nominal value of 1.5 V to 1.1 V. When the battery voltage
decreases below 1.1 V, the regulator drops out of regulation. When the dropout point is
reached, the READY indicator turns off to automatically signify the end of battery life. The
BATT indicator pointer will either be in or very near the red zone at the end of battery life.
Maintenance
The following maintenance procedures refer to the indicator’s instruction book.
Instrument cleaning.
Troubleshooting.
Corrective maintenance.
Replacing TC filament.
Replacing batteries.
Ordering parts.
Squeeze aspirator bulb eight to ten times to purge instrument with fresh air. Permit bulb to
inflate completely each squeeze.
Adjust LEL ZERO control to obtain zero indication on meter.
Connect flow control of calibration Check Kit., Model R, to Calibration Check Gas Cylinder
(0.75% pentane) and connect adapter hose between flow control and inlet fitting of
instrument. For instruments with factory calibration on methane, use 2% methane cylinder.
Turn flow control valve counter-clockwise. When needle stabilises, meter should indicate 50 for
standard instruments factory calibrated on pentane. For instruments calibrated on methane,
reading should be 40. If it does not, turn LEL span adjustments to obtain 50 (40 on methane)
on meter. If 50 (40 on methane) cannot be obtained by turning LEL span adjustment, perform
procedure under Replacing Catalytic Filament in the instruction book and then repeat this
procedure.
Turn flow controls valve clockwise and disconnect adapter hose from inlet fitting.
Squeeze aspirator bulb eight to ten times to purge the instrument with fresh air. Meter should
indicate 0. If it does not, adjust LEL ZERO control to obtain zero indication on meter; then
connect adapter hose to inlet fitting and repeat steps 7 and 8. Remove flow control from gas
cylinder.
Set RANGE switch to GAS and, when READY indicator turns on, adjust GAS ZERO to obtain zero
indication on meter.
Connect source of 100% methane to inlet fitting. Pass gas through instrument and then shut
off flow. Meter should indicate 82 on an instrument factory calibrated on pentane. If it does
not, turn GAS span adjustment to obtain 82 on meter. If 82 cannot be obtained by turning GAS
span adjustment, refer to Replacing TC Filament in the instruction book, and then repeat
this procedure. For instrument factory calibrated on methane, meter should indicate 100; if it
does not, turn GAS span adjustment to obtain 100 on meter.
WARNING!
Do not introduce 100% methane gas to an instrument located near a source of ignition; otherwise
an explosion may occur.
Disconnect source of methane and squeeze aspirator bulb eight to ten times to purge
instrument with fresh air. Meter should indicate zero. If it does not, adjust GAS ZERO control to
obtain zero indication on meter; then repeat steps 11 and 12.
Place instrument in bottom section of case and clasp top and bottom section together. Close
cover.