www.vaisala.
com
PTU300 Combined Pressure, Humidity and
Temperature Transmitter
Technical data
Performance
Barometric pressure
Pressure range 500 ... 1100 hPa, 50 ... 1100 hPa
Accuracy 500 ... 1100 hPa 500 ... 1100 hPa 50 ... 1100 hPa
Class A Class B
Linearity ±0.05 hPa ±0.10 hPa ±0.20 hPa
Hysteresis* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Repeatability* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Calibration ±0.07 hPa ±0.15 hPa ±0.20 hPa
uncertainty**
The Vaisala PTU300 Combined Pressure, Humidity and
Accuracy at +20 °C*** ±0.10 hPa ±0.20 hPa ±0.30 hPa
Temperature Transmitter is a versatile, multi-purpose
Temperature ±0.1 hPa ±0.1 hPa ±0.3 hPa instrument.
dependence****
Total accuracy ±0.15 hPa ±0.25 hPa ±0.45 hPa
(-40 ... +60 °C/ Response time (90 %) at +20 °C (+68 °F) in still air
-40 ... +140 °F) with grid filter 8 s / 17 s*
Long-term stability/year ±0.1 hPa ±0.1 hPa ±0.2 hPa with grid + steel netting filter 20 s / 50 s*
Response time with sintered filter 40 s / 60 s*
(100 % response) * with HUMICAP® 180R or 180RC sensor
one sensor 2 s• 1 s• 1 s• Temperature
Pressure units hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia Measurement range, all probes -40 ... +60 °C (-40 ... +140 °F)
* Defined as ±2 standard deviation limits of endpoint non-linearity,
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.4 °F)
hysteresis error or repeatability error and calibration.
** Defined as ±2 standard deviation limits of accuracy of the working Temperature units °C, °F
standard including traceability to NIST. Accuracy over temperature range
*** Defined as the root sum of the squares (RSS) of endpoint non-linearity,
hysteresis error, repeatability error and calibration uncertainty at room ∆ °C
temperature. 0.5
**** Defined as ±2 standard deviation limits of temperature dependence over 0.4
the operating temperature range. 0.3
0.2
Relative humidity 0.1
°C
Measurement range 0 ... 100 % RH 0
-0.1
Accuracy (including non-linearity,
-0.2
hysteresis, and repeatability at
-0.3
+15 ... +25 °C ±1 %RH (0 ... 90 % RH)
-0.4
±1.7 %RH (90 ... 100 %RH) -0.5
at -20 ... +40 °C ±(1.0 + 0.008 x reading) %RH -40 -20 0 20 40 60
at -40 ... +60 °C ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C ) Temperature sensor PT100 RTD 1/3 Class B IEC 751
(Defined as ±2 standard deviation ± 0.6 % RH (0 ... 40 %RH)
limits. Small variations possible, ± 1.0 % RH (40 ... 97 %RH) Operating Environment
see also calibration certificate.) Operating temperature -40 ... +60 °C (-40 ... +140 °F)
Sensor with display 0 ... +60 °C (+32 ... +140 °F)
for typical applications Vaisala HUMICAP® 180 or 180R* Humidity range non-condensing
for applications with chemical Electromagnetic compatibility EN61326-1:1997 + Am1:1998
purge/warmed probe Vaisala HUMICAP® 180C or 180RC* +Am2:2001; Industrial Environment
Technical data
Inputs and outputs Menu languages English, Finnish, French, German, Japanese,
Operating voltage 10 ... 35 VDC, 24 VAC Chinese, Spanish, Swedish, Russian
with optional power supply module 100 ... 240 VAC, 50/60 Hz Analog outputs (optional)
Power consumption at +20 °C (Uin 24 VDC) current output 0 ... 20 mA, 4 ... 20 mA
RS-232 max. 28 mA voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Uout 3 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 33 mA Humidity and temperature
Iout 3 x 0 ... 20 mA max. 63 mA accuracy at +20 °C ±0.05% full scale
display and backlight +20 mA temperature dependence ±0.005%/°C full scale
during chemical purge max. +110 mA Pressure 500 ... 1100 hPa 50 ... 1100 hPa
during probe heating +120 mA accuracy at +20 °C ±0.30 hPa ±0.40 hPa
Settling time at power-up (one sensor) accuracy at -40 ... +60 °C ±0.60 hPa ±0.75 hPa
class A 4s
class B 3s Mechanics
External loads Cable bushing M20 x 1.5 for cable diameter
current outputs RL < 500 ohm 8 ... 11 mm/0.31 ... 0.43"
0 ... 1 V output RL > 2 kohm Conduit fitting 1/2" NPT
0 ... 5 V and 0 ... 10 V outputs RL > 10 kohm User cable connector (optional) M12 series 8-pin (male)
Recommended wire size 0.5 mm2 (AWG 20) stranded wires option 1 female plug with 5 m (16.4 ft) black cable
Digital outputs RS-232, RS-485 (optional) option 2 female plug with screw terminals
Service connection RS-232, USB Probe cable diameter
Relay outputs (optional) 0.5 A, 250 VAC PTU303 6.0 mm
Ethernet interface (optional) other probes 5.5 mm
Supported standards 10/100Base-T Housing material G-AlSi 10 Mg (DIN 1725)
Connector RJ45 Housing classification IP 65 (NEMA 4)
Protocols Telnet Weight
Software support Vaisala MI70 link depending on selected probe 1.5 ... 2.0 Kg
WLAN interface (optional)
Supported standards 802.11b
Accessories
PC software and cable 215005
Antenna connector type RP-SMA
USB-RJ45 Serial Connection Cable 219685
Protocols Telnet
Connection cable for HM70 211339
Security WEP 64/128,WPA
Wall mounting plate (plastic) 214829
Software support Vaisala MI70 link
Pole installation kit 215108
Authentication / Encryption (WLAN)
Rain shield 215109
Open / no encryption
DIN rail installation set 211477
Open / WEP
Duct installation kit, PTU303/307 210697
WPA Pre shared key / TKIP
Cable gland and AGRO, PTU303/307 HMP247CG
WPA Pre shared key / CCMP (a.k.a. WPA2)
Solar radiation shield, PTU303/307/30T DTR502B
Optional data logger with real-time clock
Meteorological installation kit HMT330MIK
Logged parameters max. three with trend/min/max values
Duct installation kit (T probe) 215003
Logging interval 10 sec (fixed)
Max. logging period 4 years 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphic trend display
of any parameter
TYPE APPROVED PRODUCT
CERTIFICATE NO.: A-11440
For more information, visit Ref. B210892EN-A ©Vaisala 2009
This material is subject to copyright protection, with all
www.vaisala.com or contact copyrights retained by Vaisala and its individual partners. All
us at
[email protected] rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.