The VACUSAFE operating instructions provide essential information for the installation, operation, and maintenance of the VACUSAFE aspiration systems, which are designed for laboratory use with non-explosive aqueous solutions. The document includes safety notes, descriptions of device components, installation procedures, and operational guidelines. It also outlines compliance with various safety and electromagnetic compatibility standards, ensuring safe usage of the device.

Uploaded by

CarlosReis
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
27 views20 pages

Vacu Safe

The VACUSAFE operating instructions provide essential information for the installation, operation, and maintenance of the VACUSAFE aspiration systems, which are designed for laboratory use with non-explosive aqueous solutions. The document includes safety notes, descriptions of device components, installation procedures, and operational guidelines. It also outlines compliance with various safety and electromagnetic compatibility standards, ensuring safe usage of the device.

Uploaded by

CarlosReis
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 20

VACUSAFE Operating instructions

158340_V04
Declaration of conformity | Konformitätserklärung |
Déclaration de conformité | Declaración de conformidad |
Dichiarazione di conformità

INTEGRA Biosciences AG – 7205 Zizers, Switzerland


declares on its own responsibility that the product | erklärt in alleiniger Verantwortung,
dass das Produkt | déclare sous sa responsabilité exclusive, que le produit |
declara bajo su propia responsabilidad que el producto | dichiara sotto la propria respon-
sabilità che il prodotto

VACUSAFE models: 158300, 158310, 158320, 158610, 158620

in accordance with EC directives | gemäss der EU-Richtlinien | est conforme au terme de


la directives CE | de acuerdo con las directivas CE | in conformità alle direttive CE
2006/95/EC Low voltage equipment
2004/108/EC Electromagnetic compatibility
2002/95/EC Restriction of Hazardous Substances
2002/96/EC Waste Electrical and Electronic Equipment

is in compliance with the following normative documents: | mit den folgenden normativen
Dokumenten übereinstimmt: | aux documents normatifs ci-après: | cumple las
documentos normativos: | soddisfa le normative seguenti:
EN 61010-1:2001-02 Safety requirements for electrical equipment for measurement,
control and laboratory use - General requirements.
EN 61326-1:2006 Electrical equipment for measurement, control and laboratory
use - EMC requirements.

Standards for Canada and USA


CAN/CSA-C22.2 Safety requirements for electrical equipment for measurement,
No.1010.1-92 control and laboratory use - General requirements.
UL Std. No. 3101-1 Safety requirements for electrical equipment for measurement,
control and laboratory use - General requirements.
FCC, Part 15, Class A Emission

Zizers, 1.9.2009

Elmar Morscher Thomas Neher


CEO Quality Manager
VACUSAFE – Operating instructions V04

Table of contents

Chapter 1 Introduction

1.1 Symbols used ...........................................................................5


1.2 Intended use.............................................................................5
1.3 Safety notes..............................................................................5

Chapter 2 Description of the device

2.1 Scope of delivery ......................................................................7


2.2 Overview of the VACUSAFE ....................................................8

Chapter 3 Installation

3.1 Operating environment .............................................................9


3.2 Power supply ............................................................................9
3.3 Set-up .......................................................................................9

Chapter 4 Operation

4.1 Vacuum activation and notes on use......................................12


4.2 Use of the level sensor ...........................................................13
4.3 Troubleshooting......................................................................14

Chapter 5 Maintenance

5.1 Cleaning and servicing ...........................................................15


5.2 Autoclaving and chemical sterilisation....................................15
5.3 Equipment disposal ................................................................16

Chapter 6 Technical data

6.1 Specifications .........................................................................17


6.2 Chemical compatibility............................................................17

Chapter 7 Accessories and consumables

www.integra-biosciences.com 3
VACUSAFE – Operating instructions V04

Imprint

© 2009 INTEGRA Biosciences AG


All rights to this documentation are reserved. In particular the rights of
reproduction, processing, translation and the form of presentation lie with
INTEGRA Biosciences AG. Neither the complete documentation nor parts
thereof may be reproduced in any way, or stored and processed using
electronic media or distributed in any other way without the written consent
of INTEGRA Biosciences AG.
This operating instruction manual has the part number 158340, version
number V04.

Manufac- INTEGRA Biosciences AG


turer CH-7205 Zizers
Switzerland
www.integra-biosciences.com
[email protected]

Customer Please contact your local INTEGRA Biosciences AG representative.


service To find out name and address of your local representative go to
www.integra-biosciences.com.
Further information is available from [email protected].

4 INTEGRA Biosciences
VACUSAFE – Operating instructions V04 Introduction

1 Introduction

These operating instructions contain all information required for installation,


operation and maintenance of the VACUSAFE aspiration systems.
Purpose This chapter informs about the symbols used in these operating instructions,
the intended use of the VACUSAFE and the general safety instructions.

1.1 Symbols used


These operating instructions specifically advise of residual risks with the
following symbols:

WARNING
This safety symbol warns against hazards that could result in injury or death.
It also indicates hazards for machinery, materials and the environment.
It is essential that you follow the corresponding precautions.

CAUTION
This symbol cautions against potential material damage or the loss of data in
a microprocessor controller. Follow the instructions.

NOTE
This symbol identifies important notes regarding the correct operation of the
device and labour-saving features.

1.2 Intended use


The VACUSAFE is a vacuum-based system for aspiration of nonexplosive
aqueous solutions. It is intended for in vitro use only and it shall not be used
for applications other than those specified.

1.3 Safety notes


The VACUSAFE corresponds to the state of the art, complies to the
recognized safety regulations and is safe to operate. The VACUSAFE can be
operated only when it is in perfect conditions and while observing these
operating instructions.
The device may be associated with residual risks if it is used or operated
improperly by untrained personnel.

www.integra-biosciences.com 5
VACUSAFE – Operating instructions V04 Introduction

Any person charged with the operation of the VACUSAFE must have read
and understood these operating instructions, and particularly, the safety
notes, or must have been instructed by supervisors so that safe operation of
the device is guaranteed.
Regardless of the listed safety notes, additionally applicable regulations and
guidelines of trade associations, health authorities, trade supervisory offices,
etc. must be observed.
Do not open, convert or alter the VACUSAFE in any way. Repairs may only
be performed by INTEGRA Biosciences AG or by an authorised after-sales
service member.
Parts may be replaced with original INTEGRA Biosciences parts only.
Use an original INTEGRA Biosciences mains adapter only.

6 INTEGRA Biosciences
VACUSAFE – Operating instructions V04 Description of the device

2 Description of the device

Purpose This chapter describes the most important components of the VACUSAFE
and defines the specific terminology. Different models of the VACUSAFE are
available and specifically described in these operating instructions where
necessary.

NOTE
Please determine which VACUSAFE model you have purchased as it will
simplify the reading of these operating instructions. Find the part number of
your model on the main packaging label and note it here: 158 .

2.1 Scope of delivery


• Base
- Bottle stand (model 158620)
or
- Housing with integrated bottle stand (all other models)
• Bottle
• Lid with openings for tubing
• Tubing and filter
• Cable for level sensor (models 158300, 158310 and 158610)
• Mains adapter (all models except 158620)
• VACUBOY hand operator

CAUTION
Verify the scope of delivery when unpacking the device and check for
potential transportation damage. Do not operate a device that is damaged,
instead contact your local dealer.

www.integra-biosciences.com 7
VACUSAFE – Operating instructions V04 Description of the device

2.2 Overview of the VACUSAFE

7 6 1 Housing (bottle stand in model 158620)

10
2 Bottle for liquid collection
4 3 Lid with openings for tubing
3
4 Bottle handle (except model 158300)
5 VACUBOY hand operator

5
6 Tubing for connection of hand operator to
bottle
2
7 Tubing for connection of base to bottle
9 8 On/off switch with indicator lamp
9 Knob for vacuum regulation
(models 158300, 158310 and 158320)
10 Level sensor
8
1

Rear view (models 158300, 158310, 158320)


1 Button for calibration of level sensor
2 Socket for level sensor cable
3 Mains connection socket
4 Filter port (air inlet)
1 2 3 4 5 5 Air outlet

Rear view (model 158610)


1 Button for calibration of level sensor
2 Socket for level sensor cable
3 Mains connection socket
4 Filter port (air inlet)
5 1 2 3 4 5 Vacuum port for connection to external
vacuum source

8 INTEGRA Biosciences
VACUSAFE – Operating instructions V04 Installation

3 Installation

Purpose This chapter defines the operating environment and describes the installation
of the VACUSAFE.

3.1 Operating environment


The VACUSAFE has been designed for use in a laboratory. It shall be
installed in a level, dry and dust-free location with a temperature of 5 – 40 °C
and a maximal (non-condensing) relative humidity of 80%.

3.2 Power supply


Fit the mains adapter with the suitable plug. Insert the mains adapter cable
into the VACUSAFE mains connection socket (see “Overview of the
VACUSAFE” on page 8) and plug it to the power source.

WARNING
The supply voltage must meet the requirements of the mains adapter:
100 – 240 VAC, 50/60 Hz.

3.3 Set-up

3.3.1 Connection to an external vacuum source and filter protection

Model 158620
The external vacuum source has to be
protected with the supplied hydrophobic filter.
To connect the filter to the vacuum source use
the tubing connector and a piece of tubing as a
joint if necessary. Take care to connect the filter
with the green dot turned towards the vacuum
source.

www.integra-biosciences.com 9
VACUSAFE – Operating instructions V04 Installation

Model 158610
Link your external vacuum source to the
vacuum port on the back of the VACUSAFE
base (use the supplied tubing connector to join
tubing with different diameters if necessary).
Insert the supplied hydrophobic filter into the
filter port with the green dot facing the
instrument. See “Rear view (model 158610)”
on page 8 for the position of the vacuum and
filter ports.

Models 158300, 158310 and 158320


No connection to an external vacuum is
required as a vacuum pump is integrated in the
housing. Simply insert the hydrophobic filter
into the filter port with the green dot facing the
instrument (see “Rear view (models 158300,
158310, 158320)” on page 8).

WARNING
Take care to connect the filter with the green dot turned towards the
instrument or vacuum source in order to guarantee that the hydrophobic side
of the filter is facing the liquid collection bottle.

3.3.2 Connection of the bottle and hand operator

1) Place the bottle without lid into the bottle


stand (model 158620) or in the base of the
instrument (all other models). Insert the
bottle handle into the neck of the bottle by
1
screwing it in place. Close the bottle with
the lid fitted with tubing openings and make
sure that the lid is screwed on tightly.

10 INTEGRA Biosciences
VACUSAFE – Operating instructions V04 Installation

2) Connect one end of the blue tubing to the


filter and the other end either to the barbed
3 2 fitting on the lid marked “PUMP” (models
158320 and 158620) or to the blue tubing
coupling (models 158300, 158310 and
158610).
When installing model 158620, use one of
the long transparent tubing instead of the
short blue tubing.
3) Connect the long transparent tubing to the
barbed fitting on the lid marked “HAND”
(models 158320 and 158620) or to the red
tubing coupling (models 158300, 158310
and 158610). Insert the VACUBOY hand
operator into the other end of the long
tubing.

3.3.3 Connection and calibration of the level sensor

1) Plug the level sensor cable into its socket at


the back of the instrument and connect it to
the bottle lid (models 158300, 158310 and
158610). Make sure that the sensor pins
inside the lid are dry and unsoiled.
2) Switch on the VACUSAFE. A double beep
confirms correct connection of the level
sensor.
3) Press the “CALIBRATION” button on the back
of the VACUSAFE base until a short beep
is emitted.

In case of failed calibration (indicated by continuous beeping and blinking of


the switch indicator lamp), repeat the calibration after cleaning the sensor
pins inside the lid and checking that the level sensor cable is fully inserted in
the lid and base.

www.integra-biosciences.com 11
VACUSAFE – Operating instructions V04 Operation

4 Operation

Purpose This chapter describes operation and troubleshooting of the VACUSAFE.

4.1 Vacuum activation and notes on use


Model 158620
Activate the external vacuum source and wait a few seconds until the air from
the collection bottle is evacuated.
Model 158610
Activate the external vacuum source and switch on the VACUSAFE to open
the integrated valve (the switch indicator lamp lights). Wait a few seconds
until the air from the collection bottle is evacuated. The valve will close
automatically when the VACUSAFE is not in use to avoid unnecessary
evaporation of volatile liquids.
Models 158300, 158310 and 158320
Switch on the VACUSAFE to start the vacuum pump (the switch indicator
lamp lights). Wait a few seconds until the air from the collection bottle is
evacuated. To increase or decrease aspiration speed, turn the knob for
vacuum regulation to the right or left, respectively (see “Overview of the
VACUSAFE” on page 8).

Aspirate liquid by pressing the green button of the VACUBOY hand operator
fitted with a suitable tip. Check the level of liquid in the bottle frequently,
especially when working without level sensor, and make sure that there is
enough space left for the liquid that needs to be aspirated.

NOTE
To open the bottle, first release the vacuum by opening the venting port on
the lid (open the sealing screw or remove the sealing cap). Alternatively you
can aspirate air with the VACUBOY.
When closing the bottle for use, screw the lid on tightly. The seal in the lid
needs to be adequately compressed in order to function. Also make sure the
venting port is closed.

CAUTION
Prevent liquid or foam overflowing from the vacuum bottle into the filter
thereby blocking it and thus limiting its safety function. In case the filter gets
accidentally wetted or soiled it has to be changed immediately, otherwise

12 INTEGRA Biosciences
VACUSAFE – Operating instructions V04 Operation

retention of liquids can not be guaranteed. Note that the hydrophobic filter
holds back aqueous solutions, whereas retention of apolar liquids is limited.
To prevent the build-up of foam, we recommend the use of an anti-foaming
agent (e.g. Antifoam A from Sigma).

WARNING
When working with dangerous substances, you must comply with the
material safety data sheet (MSDS) and the additionally applicable
regulations of trade associations, health authorities, supervisory offices, etc.
In addition, when working with hazardous volatile substances or
biohazardous agents, either place the VACUSAFE inside a biosafety cabinet
or connect the instrument's air outlet to a safety ventilation by means of a
tube and of a connector (Part No. 158427).

4.2 Use of the level sensor


When the VACUSAFE bottle is full, the level sensor prevents foam or liquid
from overflowing and accidentally soaking filter and instrument. As soon as
foam or liquid is detected, the pump switches off (models 158300, 158310)
or the valve closes (model 158610) to avoid overfilling of the bottle.
Additionally, an audible alarm is activated and the switch indicator lamp starts
blinking.
It is possible to deactivate the level sensor by simply disconnecting the
sensor cable (confirmed by a double beep).

NOTE
Instrument, sensor cable and bottle lid constitute an assembly and they must
be calibrated to match one another (see “Connection and calibration of the
level sensor” on page 11).
If you have several lids, sensor cables and instruments and swap them, you
have to recalibrate the newly formed assemblies. Depending on the degree
of soiling of the sensor pins inside the lid, it may be necessary to clean them
and to calibrate the level sensor periodically.

www.integra-biosciences.com 13
4.3 Troubleshooting

14

VACUSAFE – Operating instructions V04


Problem Probable cause Remedy
Device not running. Level sensor has been activated. Empty the bottle.
No electrical power. Plug the mains adapter in the power source and in the
device.
Device is switched off. Switch on the device (switch indicator lamp must light).
Aspiration is very low Aspiration speed is set to mini- Turn the knob for vacuum control clockwise to increase
and/or pump never mum. speed.
stops running. External vacuum source is too Increase strength of your external vacuum source.
weak.
Leak in the vacuum system. Close the venting port by closing the sealing screw or
sealing cap. Close the bottle lid tightly.
Check tubing and filter for leaks.
Reduce the residual flow of the VACUBOY hand opera-
tor by closing the screw opposite the operator button
(see the VACUBOY operating instructions).
Liquid overflowed thereby blocking Empty the bottle and exchange the filter
filter.
Tubing is blocked. Clean or exchange the tubing.
Level sensor activated Sensor cable is connected incor- Fully insert the sensor cable with the correct orientation
INTEGRA Biosciences

erroneously (bottle is rectly. in the lid and base sockets.


not full). Foam activates the level sensor. Use an anti-foaming agent (e.g. Antifoam A from Sigma)
No or incorrect calibration. Calibrate following the instructions (see “Connection
and calibration of the level sensor” on page 11).

Operation
Soiled sensor pins. Clean pins inside the lid and recalibrate.
Electromagnetic interferences dis- Eliminate or inactivate the source of electromagnetic
turb level sensor. interference.
VACUSAFE – Operating instructions V04 Maintenance

5 Maintenance

5.1 Cleaning and servicing

WARNING
Always switch off the VACUSAFE and disconnect it from the mains when
carrying out maintenance work.
If the VACUSAFE gets soiled, it can be cleaned with a cloth moistened with
soapy water or with a 70% ethanol solution.
It is recommended to change the hydrophobic filter regularly, at least once a
year. Should the filter get wetted or soiled accidentally, it has to be changed
immediately.
Regular servicing of the VACUSAFE by a qualified representative of
INTEGRA Biosciences AG is recommended.

5.2 Autoclaving and chemical sterilisation


To inactivate potential biohazards that contaminate the aspirated liquid and
consequently also some parts of the VACUSAFE, read the information
described below.
See also the warnings in “Vacuum activation and notes on use” on page 12.

5.2.1 Autoclaving
It is recommended to autoclave the full polypropylene bottle for at least 60
min. and the full glass bottle for at least 20 min. at 121°C. To determine the
sterilisation time precisely, measure the temperature in the autoclave using
as reference an identical filled bottle and by following the autoclave
manufacturer’s instructions.

CAUTION
The lid with black tubing couplings made of POM (delivered with model
158610) is not autoclavable. It is not recommended to autoclave the
VACUBOY regularly, however, it can be autoclaved sporadically for
decontamination. Do not re-use an autoclaved filter.

CAUTION
The bottle lid has to be always unscrewed and kept loosely on the bottle
during autoclaving. The tubing couplings made of PVDF (delivered with

www.integra-biosciences.com 15
VACUSAFE – Operating instructions V04 Maintenance

models 158300 and 158310) have to be always disconnected during


autoclaving.

5.2.2 Chemical sterilisation


Two methods for chemical inactivation of biohazardous agents are
commonly used. A suitable disinfectant is either added to the full collection
bottle and left to react for the recommended time or the disinfectant is added
to the empty bottle so that the aspirated liquid gets sterilised as it
accumulates in the bottle.
When using the first method, any disinfectant can be used, provided that it is
compatible with the bottle material (glass or polypropylene).

CAUTION
When using the second method of chemical sterilisation, use disinfectants
that are free of chlorine or other corrosive agents, especially when corrosive
vapours are generated and aspirated. A suitable disinfectant is for example
TriGene manufactured by Medichem International.

5.3 Equipment disposal


The VACUSAFE is labelled with the “crossed-out wheeled bin” symbol to
indicate that this device must not be disposed of with unsorted municipal
waste. Instead, it is your responsibility to correctly dispose of your waste
equipment by handing it over to an authorised facility for separate collection
and recycling. It is also your responsibility to decontaminate the device in
case of biological, chemical, and/or radiological contamination so as to
protect from health hazards the persons involved in the disposal and
recycling of equipment.
For more information about where you can drop off your waste equipment for
recycling, please contact your local dealer from whom you originally
purchased the product or your local council.
By doing so, you will help conserve natural resources and you will ensure
that your waste equipment is recycled in a manner that protects human
health and the environment. Thank you!

16 INTEGRA Biosciences
VACUSAFE – Operating instructions V04 Technical data

6 Technical data

6.1 Specifications

Vacuum range –300 to –600 mbar, continuously adjustable


(models 158300, 158310, 158320)
–600 mbar, fix (model 158610)
Flow rate Pump: 8 L/min (air); Aspiration: 15 ml/s (liquid)
(models 158300, 158310, 158320)
Dimensions 420 x 210 x 210 mm (model 158620)
(H x W x D) 530 x 180 x 320 mm (all other models)
Weight 2.0 kg (model 158620)
2.7 kg (model 158610)
3.4 kg (models 158300, 158310, 158320)
Electrical data Mains adapter input: 100–240 VAC, 50/60 Hz
Mains adapter output: 18 VDC, 0.84 A
Device:
18 VDC, 0.16 A (model 158610)
18 VDC, 0.66 A (models 158300, 158310, 158320)
Degree of protection IP21
Ambient conditions Operation:
5 to 40 °C, max. 80% RH (non-condensing)
Storage:
–10 to 40 °C, max. 95% RH (non-condensing)

6.2 Chemical compatibility


The table below lists VACUSAFE parts that come into contact with the
aspirated liquid or its aerosols and vapours, and rates the compatibility of
these parts to a few of the chemicals commonly used in laboratories. To
determine the compatibility of a component to a chemical not listed in the
table, please consult one of the several tables available on the internet.
Note that the rating refers to soaking in the concentrated chemical; however,
more relevant here is the attenuated effect resulting from indirect contact with
the diluted chemical. It is recommended to test the compatibility of relevant
components to a specific chemical prior to extensive use.
INTEGRA Biosciences AG does not warrant that the information in the table
is accurate or complete and that any material is suitable for any purpose.

www.integra-biosciences.com 17
Chemical compatibility chart

18

VACUSAFE – Operating instructions V04


Bleach (e.g. NaClO)

Sodium hydroxide
Hydrochloric acid

Trichlorethylene

Chloroform
Acetone
Ethanol

Phenol
DMSO
Parts Materials
VACUBOY Plastic parts POM P G F G G G P F F
O-rings FPM (Viton) G G G P P F G G G
Valve spring Stainless steel F G P F G G G G G
Tubing Silicone G G P P P G F P P
Bottle + Lid Bottle/Lid Polypropylene G G G G G G F F G
Tubing couplings PVDF G G G P P P G G G
POM P G F G G G P F F
Filter Membrane PTFE G G G G G G F F G
Housing Polypropylene G G G G G G F F G
VACUSAFE base Pump housing PPS G F G G G G G G G
Pump membrane FPM (Viton) G G G P P F G G G
INTEGRA Biosciences

Valve housing PPS G F G G G G G G G


Valve membrane PTFE G G G G G G F F G

Technical data
Compatibility ratings:
G = Good: no or minor effects (discolouration or slight corrosion).
F = Fair: moderate effects (softening, swelling and loss of strength), not recommended for continuous use.
P = Poor: not recommended.
VACUSAFE – Operating instructions V04 Accessories and consumables

7 Accessories and consumables

Accessories Part No.


Bottle 4 litres, polypropylene 158370
3 litres, glass 158420
Lid for 4L bottle with quick couplings (PVDF) + level detection 158401
with quick couplings (POM) + level detection 158402
with barbed tubing fittings 158372
Lid for 3L bottle with quick couplings (PVDF) + level detection 158418
4L bottle and lid comprises part no. 158 370 and 158 401 158431
comprises part no. 158 370 and 158 402 158433
comprises part no. 158 370 and 158 372 158432
3L bottle and lid comprises part no. 158 420 and 158 418 158430
Level sensor cable 158403
Bottle handle for carrying polypropylene bottle 158625
(see “Overview of the VACUSAFE” on page 8)
Bottle stand for 4L polypropylene bottle 158630

Stand for PIPETBOY (acu or comfort) and VACUBOY 158065

Mains adapter 100-240 VAC, 50/60 Hz 158395


Y-connector for connection of second VACUBOY hand operator 158354
Air outlet connector for connection of tubing to the air outlet of the pump 158427
VACUBOY hand operator 155510
VACUBOY set (see VACUBOY operating instructions) 155500

www.integra-biosciences.com 19
VACUSAFE – Operating instructions V04 Accessories and consumables

Consumables Part No.


Tubing for connection of hand operator to bottle, 1.8 m, silicone 158332
for connection of base to bottle, 0.7 m, silicone, blue 158331
Filter for protection of vacuum source, 0.45 μm, hydrophobic 158015
Tubing set + filter comprises part no. 158 331, 158 332, 158 015 158342

Quick tubing couplings Material Part No.


Coupling plug self-closing, with barbed fitting, white/blue PVDF 158416
(linked to tubing) self-closing, with barbed fitting, white/red PVDF 158417
open, with barbed fitting, black POM 158437

Coupling socket self-closing, with O-ring, white/blue PVDF 158423


(mounted on lid) self-closing, with O-ring, white/red PVDF 158424
self-closing, with O-ring, black/blue POM 158435
self-closing, with O-ring, black/red POM 158436

20 INTEGRA Biosciences