Varian Specs

The Varian Trilogy installation specifications outline the necessary room dimensions, radiation shielding, power and cooling requirements, and IT infrastructure for effective operation. Key features include energy modes of 6 MV, 10 MV, and 15 MV, as well as high-definition MLC and 360° gantry rotation. The installation process involves site planning, construction, equipment assembly, commissioning, and regulatory approval.

Uploaded by

Dimi Dimo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
37 views2 pages

Varian Specs

The Varian Trilogy installation specifications outline the necessary room dimensions, radiation shielding, power and cooling requirements, and IT infrastructure for effective operation. Key features include energy modes of 6 MV, 10 MV, and 15 MV, as well as high-definition MLC and 360° gantry rotation. The installation process involves site planning, construction, equipment assembly, commissioning, and regulatory approval.

Uploaded by

Dimi Dimo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 2

Varian Trilogy Installation Specifications

1. Room & Space Requirements


Treatment Room Size: Minimum ~5.5m x 5.5m (18 ft x 18 ft) (larger
recommended for workflow efficiency).
Ceiling Height: Minimum ~3.05m (10 ft) (higher may be needed for certain
configurations).
Floor Load Capacity: Must support ≥ 6,000 kg (13,200 lbs) due to the
weight of the gantry, couch, and shielding.
Doorway Access: Large enough for equipment delivery (often requires
temporary removal of doors or walls).

2. Radiation Shielding Requirements


Primary Barriers: Typically concrete walls ≥ 1.5m (5 ft) thick (varies
based on energy and workload).
Secondary Barriers: ≥ 1m (3.3 ft) concrete for scatter radiation.
Maze Design: Required for neutron shielding if operating at ≥ 10 MV.
Door Shielding: Lead-lined or borated polyethylene for neutron protection.

3. Power & Electrical Requirements


Voltage: 480V, 3-phase (varies by region).
Power Consumption: ~80–100 kVA (depends on configuration).
Stable Power Supply: Requires voltage regulation (±1%) to prevent beam
instability.
Grounding: Low-impedance grounding to prevent interference.

4. Cooling & HVAC Requirements


Water Cooling: Chilled water system (if applicable) with ~5–10 GPM flow
rate.
Room Temperature: Maintained at 20–24°C (68–75°F) for system stability.
Humidity Control: 30–70% RH (non-condensing).

5. Networking & IT Infrastructure


DICOM Compliance: Required for integration with ARIA® Oncology
Information System (OIS).
Network Ports: Gigabit Ethernet for imaging and data transfer.
Data Storage: High-capacity storage for CBCT, kV/MV imaging, and
treatment plans.

6. Additional Considerations
Laser Alignment System: Required for patient positioning.
Floor Markings: For isocenter alignment.
Ventilation: Proper exhaust for ozone (if high-energy modes used).
Varian Trilogy Technical Specifications

Feature Specification

Energy Modes 6 MV, 10 MV, 15 MV (X-rays)

Electron Beams 6–22 MeV (optional)

Dose Rate Up to 600 MU/min (varies by mode)

OBI (On-Board Imager), CBCT, kV/MV


Imaging
imaging

MLC High-definition (HD) MLC (120 leaves)

Gantry Rotation 360° continuous

Table Motion 6-DOF (HexaPOD compatible for SRS/SBRT)

Installation Process Overview

1.Site Planning (Varian engineers assess shielding, power, and space).


2.Foundation & Shielding Construction (concrete, door installation).
3.Equipment Delivery & Assembly (gantry, couch, control systems).
4.Commissioning & Calibration (beam tuning, safety checks).
5.Regulatory Approval (FDA/CE compliance, radiation safety checks).