The document outlines a bid proposal for gas chromatographs and related equipment, including the ABB Model NGC8209 and COSA XENTAUR water dew-point analyzer. It specifies various components, configurations, and calibration requirements, along with details on performance metrics and hazardous area classifications. Additionally, it requests confirmation on the inclusion of C9+ GC in the offer and clarification on applicable panels for GC.

Uploaded by

Samaresh Nandi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
2 views3 pages

Rmgas - Chromatograph and Analyser - Fi Obs

The document outlines a bid proposal for gas chromatographs and related equipment, including the ABB Model NGC8209 and COSA XENTAUR water dew-point analyzer. It specifies various components, configurations, and calibration requirements, along with details on performance metrics and hazardous area classifications. Additionally, it requests confirmation on the inclusion of C9+ GC in the offer and clarification on applicable panels for GC.

Uploaded by

Samaresh Nandi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Bidder to confirm if C9+ GC is considered in offer.

Bidder to clarify if any panel is applicable for GC.

MİKTAR
SIRA AÇIKLAMA
(Adet)
ABB Model NGC8209 Gas Chromatograph Dual Head
1.1. 1
Configuration and Hydrocarbon Dewpoint Calculation
ABB TotalFlow Natural Gas Chromatograph Model: 8209 BTU, compact design,

Calculation Method: ISO 6976, GPA 2172


Analysis Components: Methane, Ethane, Propane, IsoButane,
N-Butane, NeoPentane, IsoPentane,
N-Pentane, C6+, CO2, Nitrogen -Air (Contains N2, Ar, CO and
O2)
Hexanes, Heptanes, Octanes, C9+
Analysis Time: 315 sec., 360 sec for Hexanes, Heptanes,
Octanes, C9+
Standart 4 ‘stream’ measurement; Up to two streams can be
assigned as the calibration / validation stream
Sensor: Micro TCD detector
Automatic Calibration, configurable, (one stream will be used for
auto calibration)
Normalization function,
PC monitoring and on-line diagnostics PCCU 32; graphic user
interface software
Repeatability:
C1 – C6+ ±0.0125%CV ( at 15°C ), ±0.025%CV (between -18°C ile 55°C working range)
Hexanes, Heptanes, Octanes, C9+ ±0.025%CV ( at 15°C ), ±0.050%CV (between -18°C ile 55°C working
range)
Supply Voltage: 12 or 24 VDC
Hazardous Area Classification: IECEx Exd IIB + H2 T6; Class 1 Zone1; ATEX II 2G; Ex d IIB + H2 T6, Class 1
Zone 1
NEMA 4X, IP 66 equivalent enclosure
Digital ¼ VGA interactive display, multiple screen input,
Analysis section; Includes stream selection solenoids, pressure regulation, 32 bit digital detector electronics,
"dual train" chromatograph
1 GB SD memory card,
Advanced Data Monitoring and Retention Functions
Two serial digital ports, software selectable for RS232, RS485 or RS422 Modbus (master/slave) A frontend
interface, USB
Ethernet port, Modbus (master/slave)
w/ Manufacturing Test Certificate
Calculated values: Gas compressibility per AGA 8, Real Relative Density (Specific Gravity), Btu/CV Value,
Wobbe Index
Sample system, single stream,
With liquid removal, liquid drain, liquid shut off and coalescing filter,
1x Heated Pressure regulator
MİKTAR
SIRA AÇIKLAMA
(Adet)

1.2. Analytical Calculation: Hydrocarbon Dewpoint in NGC8209 1

1.3. 10m Heated Line 1

1.4. Heated Jacket with Thermostat for Calibration Gas Mixture 1

1.4. Calibration Gas Mixture and Carrier Gas Set 1


1 x 50 L Steel Cylinder BS4 Valve Calibration Gas Mixture (C9+) c/w regulator
Gas mixture w/17025 certificate, analyzed according to ISO 6143
Filling tolerance accuracy ±5%, certification accuracy ± 1%

2x 50 L Steel Cylinder Helium Carrier Gas c/w regulator


Change over manifold system for He cylinders
MİKTAR
SIRA AÇIKLAMA
(Adet)
2.1. COSA XENTAUR - WATER DEW-POINT ANALYZER 1

LPDT2™
LPDT2™ is an ultra-compact, loop-powered dew point transmitter available with a
display. The LPDT2 offers users even more control and options for customized dew
point monitoring.
Type: Hyper-Thin-Film high capacitance Al2O3 dew point range, XTR-100 -100°C to
+20°C (-148°F to +68°F), XTR-65 -65°C to +20°C (-85°F to +68°F)
Sensor Range: Model: XTR-100 -100°C to +20°C (-148°F to +68°F)
Accuracy: ±3°C (± 5.5°F) for -100°C to +20°C dew point
Repeatability: ±0.5°C (±0.9°F)
Operating Temperature range: -10°C to +70°C (+14°F to +158°F) Class T5
Sample Flow Range (linear vel. @ 1 atm.): 1-4LPM
Storage Temperature: 40°C to +80°C (-40°F to +176°F)