TSI HM685M - SN. 72025001 - Hydronic Manometer

The document is a Certificate of Calibration for a Hydronic Manometer (Model HM685) issued to STED Saudi Co Limited, detailing its calibration results and conditions. The calibration was performed on September 30, 2024, with a recommended due date of September 30, 2025, and all measurements were reported to be within specified accuracy limits. The certificate includes information on the calibration method, environmental conditions, and measurement uncertainty.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
24 views3 pages

TSI HM685M - SN. 72025001 - Hydronic Manometer

The document is a Certificate of Calibration for a Hydronic Manometer (Model HM685) issued to STED Saudi Co Limited, detailing its calibration results and conditions. The calibration was performed on September 30, 2024, with a recommended due date of September 30, 2025, and all measurements were reported to be within specified accuracy limits. The certificate includes information on the calibration method, environmental conditions, and measurement uncertainty.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

EXPERT POWER

Measurement and Calibration EA-JAS


öÆE191.pW.I ülhJl Æ4.5 www.expertpower.com.sa

CERTIFICATE OF CALIBRATION
Certificate Number : EXPR-PRL-23373-01 Job Order Number : 23373
Customer Details :
STED Saudi co Limited
Riyadh, Kingdom of Saudi Arabia
Instrument Details of UUC :
Description : Hydronic Manometer
Manufacturer TSI
Manufacturer Model Number HM685
Manufacturer Serial Number 72025001
ID. Number
Range As per Specification
Resolution/Least Count As per Specification
Accuracy As per Specification
Calibrated at Lab
Visual Inspection 0k
Received Date 30-Sep-24
Issue Date 30-Sep-24

Calibration Date 30-Sep-24


Recommended Calibration Due Date 30-Sep-25
Instrument Conditions of UUC :
As Found : Within Tolerance As Left : Calibrated within Tolerance

Environmental Conditions :

Ambient Temperature 23±2 oc


Relative Humidity 10 0/0

Calibration Method: EXP-CP-PRE-OI


Details of Standard Equipment Used.:

Description Sl. No /
ID. No Certifizate No Validity
Pressure Calibrator 91 S507577 2023C09436 14-Nov-24
Note:
I. The Certificate refers only to the particular item(s) submitted for calibration.
2. All Results Reported are valid at the time of and under the stated conditions of measurements.
3. The equipment/tool has been calibrated using measurement standards with traceability o? measurement to
national or international standards, which realize the International system of Units(SI)
4. This certificate shall not be reproduced, without the written permission of Expert Power Calibration Laboratory.
5. The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by
the coverage factor (k=2) such that the
coverage probability corresponds to approximately 95%. The uncertainty evaluation
has been carried out in accordance with
JCGM 100:2008 "Guide to the Expression of uncertainty in Measurement".
Calibration Technician Authorised Signatory

LIBRA,TED

Calibration Tech. Stamp Sun- ara Moorthi BoomiRaj


RPT-TCC-003v2.O (Technical Manager)
QLF-14-Ol Issue No:02 Issue:02-oct-2022 Page I of 3
Head Office:
Riyadh, 11438, Saudi Arabia

BQCv
Sitteen Street, Malaz P.O. Box,32933,
Tel-Fax: +966 114210481
•USINCss
Branch: Certinc•thn

Daharan-Jubail Express Highway ISO 9001:2015


Dammam-Tel: +966114210481-Ext. 200
CERTIFICATE OF CALIBRATION

Certificate Number : EXPR-PIU-23373-01 Job Order Number : 23373

Calibration Results

Expanded
Standard Reading in
UUC Reading in Mean Value in Error in Measurement
Uncertainty
Up Down

psi psi Psi


Psi psi

Diff.Pressure (+)

0.22
0.000 0.000 0.000 0.000 0.000

0.22
10.000 10.006 10.002 10.00 -0.004
0.23
50.000 50.034 50.02 -0.023
50.011
0.23
100.000 100.043 100.065 100.05 -0.054
0.23
200.000 200.031 200.048 200.04 -0.039

0.24
300.000 300.052 300.052 300.05 -0.052

Diff.Pressure (-)
0.22
0.000 0.000 0.000 0.000 0.000

0.22
-10.000 -10.005 -10.003 -10.004 0.004

0.22
-50.000 -50.015 -50.017 -50.016 0.016

0.23
-100.000 -100.023 -100.027 - 100.025 0.025

0.23
-200.000 -200.042 -200.033 -200.038 0.037

0.24
-300.000 -300.044 -300.044 -300.044 0.044

Gauge Pressure (+)


0.22
0.000 0.000 0.000 0.000 0.000

0.22
10.000 10.007 10.004 10.006 -0.005
0.22
50.000 50.019 50.01 1 50.015 -0.015
0.22
100.000 100.024 100.024 100.024 -0.024
0.23
200.000 200.042 200.032 200.037 -0.037

0.24
300.000 300.055 300.0 300,055 -0.055

QLF-14-Ol Issue No:02 Issue:02-cct-2022 Page 2 of 3


CERTIFICATE OF CALIBRATION

Certificate Number : EXPR-PIU-23373-01 Job Order Number : 23373

Expanded
Standard Reading in
UUC Reading in Mean Value in Error in
Measurement
Uncertainty
Up Down

psi psi Psi Psi Psi

Gauge Pressure (-)


0.22
0.000 0.000 0.000 0.000 0.000

0.22
-10.000 -10.009 -10.002 -10.006 0.006

0.22
-50.000 -50.018 -50.016 -50.017 0.017

0.22
-100.000 -100.022 -100.025 -100.024 0.024

0.23
-200.000 -200.043 -200.041 -200.042 0.042

0.24
-300.000 -300.029 -300.029 -300.029 0.029

Conclusion / Remarks:
I .

Measurement Uncertainty at 95.45


% Confidence level and Coverage Factor k=2.
2. The above Results are with in specified accuracy limits.
3. Calibration points marked "*"
are not within the scope of Accreditation.

QLF-14-01 Issue No:02 Issue:02-oct-2022 Page 3 of 3