HP Agilent-Technologies E5070b E5071b Network Analyzer
HP Agilent-Technologies E5070b E5071b Network Analyzer
Service Guide
Eighth Edition
2
Safety Summary
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or with
specific WARNINGS elsewhere in this manual may impair the protection provided by the
equipment. In addition it violates safety standards of design, manufacture, and intended use
of the instrument.
Agilent Technologies assumes no liability for the customer’s failure to comply with these
requirements.
NOTE This equipment is MEASUREMENT CATEGORY I (CAT I). Do not use for CAT II, III,
or IV.
NOTE This equipment is tested with stand-alone condition or with the combination with the
accessories supplied by Agilent Technologies against the requirement of the standards
described in the Declaration of Conformity. If it is used as a system component,
compliance of related regulations and safety requirements are to be confirmed by the
builder of the system.
3
safety features are maintained.
• Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous procedures
throughout this manual. Instructions contained in the warnings must be followed.
WARNING Dangerous voltages, capable of causing death, are presenting this instrument. Use
extreme caution when handling, testing, and adjusting this instrument.
Safety Symbol
General definitions of safety symbols used on the instrument or in manuals are listed
below.
Instruction Manual symbol: the product is marked with this symbol when it is necessary for
the user to refer to the instrument manual.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
In position of push-button switch.
Out position of push-button switch.
Frame (or chassis) terminal. A connection to the frame (chassis) of the equipment which
normally include all exposed metal structure.
Stand-by.
WARNING This warning sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could result in
injury or death to personnel.
CAUTION This Caution sign denotes a hazard. It calls attention to a procedure, practice, condition or
the like, which, if not correctly performed or adhered to, could result in damage to or
destruction of part or all of the product.
NOTE Note denotes important information. It calls attention to a procedure, practice, condition or
the like, which is essential to highlight.
4
Certification
Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent Technologies further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institution’s calibration facility, or to the
calibration facilities of other International Standards Organization members.
Documentation Warranty
The material contained in this document is provided "as is," and is subject to being
changed, without notice, in future editions. Further, to the maximum extent permitted by
applicable law, Agilent disclaims all warranties, either express or implied with regard to
this manual and any information contained herein, including but not limited to the implied
warranties of merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or consequential damages in connection with the
furnishing, use, or performance of this document or any information contained herein.
Should Agilent and the user have a separate written agreement with warranty terms
covering the material in this document that conflict with these terms, the warranty terms in
the separate agreement will control.
Exclusive Remedies
The remedies provided herein are buyer’s sole and exclusive remedies. Agilent
Technologies shall not be liable for any direct, indirect, special, incidental, or
consequential damages, whether based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
Addresses are provided at the back of this manual.
5
Typeface Conventions
Sample (bold) Boldface type is used when a term is defined or
emphasised.
Sample (Italic) Italic type is used for emphasis.
6
Contents
1. General Information
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Software Installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Organization of Service Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Instrument Covered by This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Power Meter Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Power Meters That Can Be Tested Using This Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Equipment Used for the Power Meter Accuracy Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Description of the Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
TEST RECORD FOR POWER METER ACCURACY TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2. Performance Tests
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Test Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Softkey Selection Procedure for Performance Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
1. FREQUENCY ACCURACY TEST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
2. RF OUTPUT LEVEL ACCURACY AND FLATNESS TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3. RF OUTPUT LEVEL LINEARITY TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4. TRACE NOISE CW TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5. CROSSTALK TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Crosstalk test program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6. SYSTEM DYNAMIC RANGE TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
System Dynamic Range test program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
7. DYNAMIC ACCURACY TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
7
Contents
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
8. UNCORRECTED SYSTEM PERFORMANCE TEST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
E5070B/E5071B Performance Test Calculation Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
2. RF Output Level Accuracy and Flatness Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
3. RF Output Level Linearity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
4. Trace Noise CW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7. Dynamic Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK . . . . . . . . . . . . . . . . . . . . . . . . . 130
Frequency Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
RF Output Level Accuracy and Flatness Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
RF Output Level Linearity Test (@ Port1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Trace Noise CW Test (Magnitude) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Crosstalk Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
System Dynamic Range Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Dynamic Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Uncorrected System Performance Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
E5070B Performance Test Record for Serial Prefix: MY422, and above . . . . . . . . . . . . . . . . . . . . . . . . . 142
Frequency Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
RF Output Level Accuracy and Flatness Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
RF Output Level Linearity Test (@ Port1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Trace Noise CW Test (Magnitude) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Crosstalk Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
System Dynamic Range Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Dynamic Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Uncorrected System Performance Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK . . . . . . . . . . . . . . . . . . . . . . . . . 154
Frequency Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
RF Output Level Accuracy and Flatness Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
RF Output Level Linearity Test (@ Port1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Trace Noise CW Test (Magnitude) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Crosstalk Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
System Dynamic Range Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Dynamic Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Uncorrected System Performance Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
E5071B Performance Test Record for Serial Prefix: MY422, and above . . . . . . . . . . . . . . . . . . . . . . . . . 173
Frequency Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
RF Output Level Accuracy and Flatness Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
RF Output Level Linearity Test (@ Port1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Trace Noise CW Test (Magnitude) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
Crosstalk Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
System Dynamic Range Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
Dynamic Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
8
Contents
3. Adjustment
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Warm-up for Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Required Adjustment after Replacing Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
Writing ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
Frequency Reference Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
Oven Reference Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
IF Range Gain/Phase Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
Source Power Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
Test Ports Characteristics Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
4. Troubleshooting
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
How to exit from the E5070B/E5071B Measurement View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
To Troubleshoot the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Primary Trouble Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
No Display troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and below, E5071B :
JP1KKxxxxx, MY42301396 and below . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207
Boot Process Troubleshooting for E5070B : MY42300633 and above, E5071B : MY42301397 and above .
214
Troubleshooting Using Internal Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
Power On Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
PLL unlock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
External reference signal phase unlock (Opt. 1E5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
To Execute the Internal Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219
Internal Test Failure Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Function Specific Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
To Check the Device Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
To Check the Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
To Check the Touch Panel (Option 016 only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
To Check the LCD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
To Check the External Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232
To Check the Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232
To Check the FDD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232
To Check the Video output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
To Check the External Trigger Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
To Check the LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
To Check the Printer Parallel Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
To Check the GPIB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238
To Check the USB (USBTMC) Interface Card. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238
Troubleshooting Using External Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 239
Contents of the external test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 239
Test equipment required for external test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 239
To Execute the External Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240
9
Contents
5. Replaceable Parts
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288
Direct Mail Order System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288
Exchange Assemblies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 289
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632 and below, E5071B :
JP1KKxxxxx, MY42301396 and below. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290
Top View (Major Assemblies) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290
Top View (Cables) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 292
Top View (Miscellaneous Parts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294
Top View (Under Power Supply) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 296
Top View (Motherboard and Other Parts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 298
Replaceable Parts List (Top View) for E5070B : MY42300633 and above, E5071B : MY42301397 and above
(A27 mass storage disk drive model) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300
Top View (Major Assemblies) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300
Top View (Cables) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302
Top View (Miscellaneous Parts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 304
Top View (Under Power Supply) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 306
Top View (Motherboard and Other Parts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 308
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above, E5071B S/N MY42402455 and
above. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 310
Top View (Removable Hard Disk Assemblies) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 310
Top View (Major Assemblies) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 312
Top View (Cables) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 314
Top View (Miscellaneous Parts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 316
Top View (Under Power Supply) for E5070B S/N from MY42401101 to MY42401986, E5071B S/N from
MY42402455 to MY42404509 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 318
Top View (Under Power Supply) for E5070B S/N MY42401987 and above, E5071B S/N MY42404510
and above . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 320
Top View (Motherboard and Other Parts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 322
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and MY421, E5071B serial prefix
JP1KK, and MY421 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 324
Front View (Analog) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 324
10
Contents
11
Contents
Removable Hard Disk Assemblies for E5070B S/N MY424 1101and above, E5071B S/N MY42402455
and above . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 372
Cover Assembly (1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 374
Cover Assembly (2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 375
Label on Rear Face . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 376
Other Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 377
Power Cables and Plug Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 378
6. Replacement Procedure
Replacing an Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 380
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 381
Outer Cover Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 382
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 382
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 382
Front Panel Removal for E5070B : JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx,
MY42301396 and below . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 384
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 384
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 384
Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B : MY42301397 thru
MY4232454 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 386
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 386
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 386
Front Panel Removal for E5070B S/N MY42401101 and above, E5071B S/N MY42402455 and above 388
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 388
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 388
A27 Mass Storage Disk Drive Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 390
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 390
Removal Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 390
Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 390
Restore Backup Files onto the New Mass Storage Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 391
Calibration of the Touch Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 391
Removable Hard Disk Drive Replacement for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 392
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 392
Removal procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 392
Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 393
Restore Backup Files onto the New Removable Hard Disk Drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 394
Calibration of the Touch Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 395
Removable Hard Disk Assembly Removal for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 396
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 396
Removal procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 396
Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 396
A28 FDD (Floppy Disk Drive) Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 398
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 398
Removal procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 398
Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 398
A50 ATX Power Supply Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400
12
Contents
13
Contents
14
Contents
7. Post-Repair Procedures
Post-Repair Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 466
A. Manual Changes
Manual Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 472
B. System Recovery
System Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 474
Types of system recoveries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 474
Notes on executing the factory recovery function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 474
Procedure to execute the factory recovery for E5070B : JP1KKxxxxx, MY42300632 and below, E5071B
: JP1KKxxxxx, MY42301396 and below . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 475
Procedure to execute the factory recovery for E5070B : MY42300633, and above, E5071B : MY42301397
and above. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 478
Procedure to create the user backup image for E5070B : JP1KKxxxxx, MY42300632 and below, E5071B
: JP1KKxxxxx, MY42301396 and below . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 481
15
Contents
Procedure to create the user backup image for E5070B : MY42300633, and above, E5071B : MY42301397
and above . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 484
Procedure to execute the user recovery function for E5070B : JP1KKxxxxx, MY42300632 and below,
E5071B : JP1KKxxxxx, MY42301396 and below. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 487
Procedure to execute the user recovery function for E5070B : MY42300633, and above, E5071B :
MY42301397 and above. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 490
Initial Registration of E5070B/E5071B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 493
Calibration of the Touch Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 496
C. Firmware Update
Firmware Update for revision A.06.51 or lower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 498
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 498
How to make E5070B/E5071B Firmware Installation Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 498
How to update the E5070B/E5071B firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 499
Firmware Update for revision A.08.01 and A.08.11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 501
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 501
How to make E5070B/E5071B Firmware Installation Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 501
How to update the E5070B/E5071B firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 501
Firmware Update for revision A.09.10 or later . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 503
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 503
Location of the latest E5070B/71B Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 503
How to update the E5070B/E5071B firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 503
E. Power Requirement
Preparation for Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 518
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 518
Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 518
Turning the Power ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 520
Turning the power ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 520
Turning the power OFF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 520
F. Messages
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 523
A............................................................................... 523
B ............................................................................... 523
C ............................................................................... 523
D............................................................................... 524
E ............................................................................... 524
F ............................................................................... 525
G............................................................................... 526
H............................................................................... 526
I. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 526
L ............................................................................... 528
16
Contents
M . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 528
N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 528
O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 529
P. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 529
Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 531
R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 532
S. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 532
T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 533
U . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 534
V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 534
Warning Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 536
17
Contents
18
1. General Information
1 General Information
The Service Manual is a guide to servicing the E5070B/E5071B ENA Series Network
Analyzer. The manual contains information requisite to do performance tests, adjustments,
troubleshooting, and repairs.
19
General Information
Precautions
Precautions
This section describes cautions that must be observed in operating the E5070B/E5071B.
Software Installed
The Windows operating system installed in this machine is customized for more effective
operation, and has different functions that are not part of the Windows operating system for
ordinary PCs (personal computers).
Therefore, do not attempt to use the system in ways other than those described in this
manual or to install Windows-based software (including anti-virus software) for ordinary
PCs as doing so may cause malfunctions.
Also note the followings.
• Do not update the Windows operating system installed in this machine to the Windows
operating system for ordinary PCs. Doing so will cause malfunctions.
• Do not attempt to update VBA (Visual Basic for Applications) software installed in this
machine to its equivalent developed for ordinary PCs. Doing so will cause
malfunctions.
• Do not allow any computer virus to infect the system. This machine has no virus check
function nor anti-virus software installed.
Agilent Technologies will not be held liable for any failure or damage arising from
negligence regarding these prohibitions and warnings.
NOTE If the pre-installed software is damaged somehow, resulting in errant behavior by the
machine, perform a system recovery. For further details of system recovery, refer to
Appendix B.
20 Chapter 1
General Information
1. General Information
Organization of Service Manual
Chapter 3 , “Adjustment,”
This chapter provides the adjustment information for the E5070B/E5071B ENA
Series Network Analyzer to ensure that the it is within its specifications. The
adjustment must be performed Agilent’s qualified service personnel. If you need the
adjustment for your E5070B/E5071B, it should be sent to the nearest Agilent
Technologies service office.
Chapter 4 , “Troubleshooting,”
This chapter contains information for ordering replacement parts for the
E5070B/E5071B ENA Series RF Network Analyzers.
This chapter provides procedure for removing and replacing the major assemblies in
the E5070B/E5071B ENA Series Network Analyzer.
This chapter lists the procedures required to verify the E5070B/E5071B operation
after an assembly is replaced with a new one.
This appendix contains the information required to adapt this manual to versions or
configurations of the E5070B/E5071B manufactured earlier than the current printing
date of this manual. The information in this manual applies directly to
E5070B/E5071B units with the serial number that is printed on the title page of this
manual.
This appendix describes how to recover the operating system (Windows 2000) when
the operating system has been damaged.
Chapter 1 21
General Information
Organization of Service Manual
This appendix describes how to update the E5070B/E5071B firmware. When you
want to update the E5070B/E5071B firmware, refer to this appendix.
Appendix F, “Messages,”
The E5070B/5071B can display error messages as well as messages that indicate the
internal operating status of the equipment. This appendix explains what these
messages mean by listing them in alphabetical order.
22 Chapter 1
General Information
1. General Information
Instrument Covered by This Manual
An instrument manufactured after the printing date of this manual may have serial number
prefix that is not listed on the title page. This unlisted serial number prefix indicates the
instrument is different from those described in this manual. The manual for this new
instrument may be accompanied by a yellow Manual Changes supplement or have a
different manual part number. This sheet contains “change information” that explains how
to adapt the manual to the newer instrument.
In addition to change information, the supplement may contain information for correcting
errors (Errata) in the manual. To keep this manual as current and accurate as possible,
Agilent Technologies recommends that you periodically request the latest Manual Changes
supplement. The supplement for this manual is identified by this manual's printing data and
is available from Agilent Technologies. If the serial prefix or number of an instrument is
lower than that on the title page of this manual, see Appendix A, Manual Changes. For
information concerning, a serial number prefix that is not listed on the title page or in the
Manual change supplement, contact the nearest Agilent Technologies office.
Chapter 1 23
General Information
Required Equipment
Required Equipment
Table 1-1 lists the recommended equipment for performing maintenance on the
E5070B/E5071B.
Error: ≤ ± 1× 10–10/year
24 Chapter 1
General Information
1. General Information
Required Equipment
Frequency Counter Frequency: 50 MHz to 3 GHz Agilent 53131/2A with Opt.010 1 P,A
Accuracy: < 2.5 ppm and 030*2
Chapter 1 25
General Information
Power Meter Accuracy Test
NOTE It is recommended that the revision number for the power meter “Main Firmware” be
Ax.03.00 or higher.
26 Chapter 1
General Information
1. General Information
Power Meter Accuracy Test
NOTE It is recommended that a copy of the test record on page 30 be made, and the values be
recorded on the copy, thus preserving the original for future use.
Test Procedure
NOTE This procedure assumes the use of the recommended equipment model numbers. The
actual steps required, therefore, may differ for other model numbers of equipment used.
a. Connect the DC REFERENCE OUTPUT connector on the rear panel of the range
calibrator to the DVM voltage input.
b. Connect the POWER METER output of the range calibrator to the input of the power
meter being tested.
c. Switch on the power to the power meter, the range calibrator, and the digital voltmeter.
Step 3. Perform the following steps for each channel on the power meter:
Chapter 1 27
General Information
Power Meter Accuracy Test
• POLARITY: NORMAL
• RANGE: 1 mW
• FUNCTION: CALIBRATE
Step 6. Allow the equipment to warm up for approximately 30 minutes. Do not change any
connections or control setting during this time.
Step 7. Zero and calibrate the power meter channel to which the range calibrator is connected:
Step 8. Monitor the drift rate of the power meter reading: Five minutes following calibration, the
meter must read 0.001, 0.000, or -0.001 dBm. If the power meter reading is not one of these
values, allow additional warm up time, then check the drift rate again. The range calibrator
must remain connected to the power meter during this warm up time.
Step 9. Zero and calibrate the power meter channel to which the range calibrator is connected:
NOTE After a channel on the power meter is calibrated, do not allow more than 5 minutes to
elapse before completing the remaining measurement steps for that channel.
Step 10. Record the DVM voltage reading as value A in the test record on page 30.
NOTE All DVM readings in this procedure should be recorded showing five significant digits.
Step 11. The reading on the power meter should be 0.000±0.001 dBm.
Step 13. Record the DVM voltage reading as value B in the test record.
Step 14. Wait for the power meter reading to settle (no settling drift within 20 seconds).
Step 15. Record the power meter reading as value C in the test record.
28 Chapter 1
General Information
1. General Information
Power Meter Accuracy Test
Step 17. Record the DVM voltage reading as value D in the test record.
Step 18. Wait for the power meter reading to settle (no settling drift within 20 seconds).
Step 19. Record the power meter reading as value E in the test record.
Step 20. If testing a dual-channel power meter, perform Step 7 through Step 19 for the other
channel.
Step 21. Perform the pass/fail calculations indicated on the test record.
NOTE If a channel of the power meter does not pass this test, the power meter cannot be used in
applications that require Option G12 or H12. There are no adjustments that can be
performed to improve the performance of the power meter. Typically, replacing the A6
measurement assembly associated with the failed channel will correct the problem.
Chapter 1 29
General Information
Power Meter Accuracy Test
Test Results
1 mW A= 0.00±0.001 A= 0.00±0.001
300 μW B= C= B= C=
100 μW D= E= D= E=
Pass/Fail Calculation
S=10(C/10) = S=10(C/10) =
% ERROR=((R-S)/R)x100 = % ERROR=((R-S)/R)x100 =
Limits: ±0.13% Limits: ±0.13%
Pass __ Fail __ Pass __ Fail __
U=10(E/10) = U=10(E/10) =
% ERROR=((T-U)/T)x100 = % ERROR=((T-U)/T)x100 =
Limits: ±0.10% Limits: ±0.10%
Pass __ Fail __ Pass __ Fail __
30 Chapter 1
2. Performance Test
2 Performance Tests
31
Performance Tests
Introduction
Introduction
This literature provides the performance test procedures for the Agilent E5070B/E5071B
RF Network Analyzers. The performance test names are listed in Table 2-1. The test
procedures are described sequentially in the following pages.
The test name indicates the tested performance and to which performance group the tested
performance belongs.
Each procedure consists of the following parts:
Description: describes the test procedure.
Specification: describes the performance verified in the test.
Test Equipment: describes test equipment required in the test.
Procedure: describes test procedure step by step.
NOTE Allow the analyzer to warm up for at least 30 minutes before you execute any of the
performance tests.
Perform all performance tests in an ambient temperature of 23 ± 5 °C
NOTE The performance tests should be performed periodically. The recommended test interval is
12 months.
The test interval depends on maintenance of use and the environmental conditions under
which the instrument is used. You may find that the interval could be shortened or
lengthened; however, such a decision should be based on substantial quantitative data.
NOTE Before performing any tests, make extra copies of the calculation sheets and the
performance test record pertaining to the test procedure. These are required in the test
procedure. For explanation of how to use these records, see the calculation sheet and
performance test record at the end of this literature.
NOTE All the test procedures are described without using the optional touch screen LCD features
(option 016) of the E5070B/E5071B. For the option 016 instruments, the touch screen
operating procedures corresponding to the described procedures may be used.
Para. Title
5 Crosstalk test
32 Chapter 2
Performance Tests
Introduction
Para. Title
2. Performance Test
The required equipment for the performance test is listed on Table 1-1and Table 1-2. Use
only calibrated equipment when doing the performance test.
NOTE The procedure of the softkey selection depends on the firmware revision. Please note that
some softkey selection procedures for your E5070B/E5071B might be a little bit different
from the procedure described in this chapter.
Chapter 2 33
Performance Tests
1. FREQUENCY ACCURACY TEST
Description
This test checks the frequency accuracy of the E5070B/E5071B test port output signal. The
frequency accuracy is checked at 50 MHz and 3 GHz with a frequency counter. Since the
E5070B/E5071B employs a PLL frequency synthesizer for the signal source, the frequency
accuracy test at these two frequency points can verify the accuracy for the entire frequency
range.
Specification
@23 ± 5 °C, referenced to 23 °C
Test equipment
Procedure
Step 1. Connect the test equipment as shown in Figure 2-1. For testing the E5070B/E5071B
equipped with option 1E5, connect a BNC(m)-BNC(m) adaptor between the Ref In
connector and the Ref Oven (10 MHz) connector on the E5070B/E5071B rear panel.
NOTE For testing without option 1E5, disconnect the above BNC(m)-BNC(m) adaptor.
NOTE Figure 2-1 shows the test setup for the E5070B/E5071B with 4-port option (Opt. 413/414).
For 2-port and 3-port options (Opt. 213/214/313/314), connect the frequency counter input
cable to Port 1 as well.
34 Chapter 2
Performance Tests
1. FREQUENCY ACCURACY TEST
2. Performance Test
Step 2. Initialize the frequency counter. Set the frequency counter as follows:
Controls Setting
50 Ω/1 MΩ 50 Ω
Frequency span: 0 Hz - -
Step 4. Press - and to make a Single sweep measurement. Wait for frequency
counter reading to settle.
Step 5. Record the frequency counter reading to a 1 Hz resolution in the performance test record
("Test result" column for the Frequency accuracy test.)
Chapter 2 35
Performance Tests
1. FREQUENCY ACCURACY TEST
Step 6. Connect the BNC cable to the frequency counter input channel measurable for a 3 GHz
input signal. See Figure 2-2.
36 Chapter 2
Performance Tests
2. RF OUTPUT LEVEL ACCURACY AND FLATNESS TEST
Description
This test checks the level accuracy and frequency flatness of the E5070B/E5071B test port
output signal. The level accuracy is checked for an output power level setting of 0 dBm at
2. Performance Test
50 MHz using a power meter. The frequency flatness is tested by measuring the power
level at 12 frequency points from 10 MHz to 8.5 GHz and calculating the differences of the
power meter readings from the level at 50 MHz.
Specification
Level accuracy
Flatness
E5070B
E5071B
NOTE The level accuracy and flatness specifications apply to Port 1 only. The levels for other
ports are given as supplemental performance characteristic.
Test equipment
Procedure
Step 1. Connect the power sensor to the power meter. Calibrate the power meter for the power
sensor used.
Chapter 2 37
Performance Tests
2. RF OUTPUT LEVEL ACCURACY AND FLATNESS TEST
NOTE Figure 2-3 shows the test setup for the E5070B/E5071B with 4-port option (Opt. 413/414).
For 2-port and 3-port options (Opt. 213/214/313/314), connect the power sensor to Port 1
as well.
Frequency span: 0 Hz - -
a. Record the power meter reading (same as that in Step 5-b) in the calculation sheet
("Power meter reading [Ref]" column for the level flatness test).
38 Chapter 2
Performance Tests
2. RF OUTPUT LEVEL ACCURACY AND FLATNESS TEST
E5070B/E5071B 10 MHz
550 MHz
2. Performance Test
1.05 GHz
1.55 GHz
2.05 GHz
3 GHz
NOTE The RF output level is measured at the minimum, maximum and proper frequencies other
than the source adjustment frequencies within the specified frequency range.
Chapter 2 39
Performance Tests
3. RF OUTPUT LEVEL LINEARITY TEST
Description
This test checks the level accuracy of the E5070B/E5071B test port output signal across the
specified level range. The RF output level is measured for power level settings of -15 dBm
to 10 dBm, 8 dBm, or 6 dBm in 1 dB step increments at 10MHz, 3 GHz, 4.25 GHz, 6 GHz
and 8.5 GHz as serial number prefix is MY421, or JP1KK. The RF output level is
measured for power level settings of -15 dBm to 10 dBm, 9 dBm, 7 dBm, or 5 dBm in 1 dB
step increments at 10MHz, 3 GHz, 4.25 GHz, 6 GHz and 8.5 GHz as serial number prefix
is MY422, and above.
Specification
E5071B
E5071B
NOTE The level linearity specification applies to Port 1 only. The levels for other ports are given
as supplemental performance characteristic.
40 Chapter 2
Performance Tests
3. RF OUTPUT LEVEL LINEARITY TEST
Test equipment
Procedure
Step 1. Connect the power sensor to the power meter. Calibrate the power meter for the power
2. Performance Test
sensor used.
NOTE Figure 2-4 shows the test setup for the E5070B/E5071B with 4-port option (Opt. 413/414).
For 2-port and 3-port options (Opt. 213/214/313/314), connect the power sensor to Port 1
as well.
Frequency span: 0 Hz - -
Chapter 2 41
Performance Tests
3. RF OUTPUT LEVEL LINEARITY TEST
Step 5. Wait for power meter reading to settle. Record the power meter reading in the calculation
sheet ("Power meter reading [Ref]" column for the level linearity test).
Step 6. Press - - to set the power level to 1 dBm, which is the second value of
the level linearity test levels listed in Table 2-3.
Table 2-3 RF output level linearity test conditions for serial prefix MY421, or JP1KK
Table 2-4 RF output level linearity test conditions for serial prefix MY422, and above
Power level 0 dBm (reference level) 0 dBm (ref) 0 dBm (ref) 0 dBm (ref)
1 dBm 1 dBm 1 dBm 1 dBm
2 dBm 2 dBm 2 dBm 2 dBm
3 dBm 3 dBm 3 dBm 3 dBm
4 dBm 4 dBm 4 dBm 4 dBm
5 dBm 5 dBm 5 dBm 5 dBm
6 dBm 6 dBm 6 dBm
7 dBm 7 dBm 7 dBm
8 dBm 8 dBm
9 dBm 9 dBm
10 dBm
NOTE The RF output level linearity is tested at the minimum, maximum and proper frequencies
other than the source adjustment frequencies within the specified frequency range.
Step 8. Wait for the power meter reading to settle. Then record the reading in the calculation sheet
("Power meter reading [a]" column for the level linearity test).
Step 9. Change the E5070B/E5071B power level setting in accordance with Table 2-3 and perform
42 Chapter 2
Performance Tests
3. RF OUTPUT LEVEL LINEARITY TEST
Step 7 and 8 for each power level to 10, (8, or 6) dBm for serial prefix MY421, or KP1KK.
Change the E5070B/E5071B power level setting in accordance with Table 2-4 and perform
Step 7 and 8 for each power level to 10, (9, or 7, or 5) dBm for serial prefix MY422, and
above.
Step 10. Press - - - to set the power level to -1 dBm, which is the second
value of the level linearity test levels listed in Table 2-5.
Table 2-5 RF output level linearity test conditions
2. Performance Test
Test E5070B/E5071B E5071B only
frequency
10 MHz 3 GHz 4.25 GHz 6 GHz 8.5 GHz
Step 12. Wait for the power meter reading to settle. Then record the reading in the calculation sheet
("Power meter reading [a]" column for the level linearity test).
Step 13. Change the E5070B/E5071B power level setting in accordance with Table 2-5 and perform
Step 7 and 8 for each power level to -15 dBm.
Step 14. Set the power level to 0 dBm and change the E5070B/E5071B center frequency to the next
level-linearity test frequency in accordance with Table 2-5.
Step 16. Calculate test results using the equation given in the calculation sheet. Record the
calculated test results in the performance test record ("Test result" column for the level
linearity test).
Chapter 2 43
Performance Tests
4. TRACE NOISE CW TEST
Description
This test checks the trace noise level for each test port of the E5070B/E5071B. The trace
noise level is quantified by performing a "through" measurement 32 times at 3 MHz, 1.3
GHz, 2.1 GHz, 3 GHz, 4.25 GHz, 6 GHz, 7.5 GHz and 8.5 GHz, with a cable connected
between two test ports. Standard deviation of the measured values at each frequency is
calculated and, then translated into a noise level expressed in dB rms.
Specification
E5071B
≤ 1 mdB rms @ 23 ± 5 °C, 3 MHz to 4.25 GHz, IFBW 3 kHz, 10 dBm, through
≤ 3 mdB rms @ 23 ± 5 °C, 4.25 GHz to 6 GHz, IFBW 3 kHz, 8 dBm, through
≤ 3 mdB rms @ 23 ± 5 °C, 6 GHz to 7.5 GHz, IFBW 3 kHz, 6 dBm, through
≤ 5 mdB rms @ 23 ± 5 °C, 7.5 GHz to 8.5 GHz, IFBW 3 kHz, 6 dBm, through
E5071B
≤ 1.2 mdB rms @ 23 ± 5 °C, 3 GHz to 4.25 GHz, IFBW 3 kHz, 9 dBm, through
≤ 3.6 mdB rms @ 23 ± 5 °C, 4.25 GHz to 6 GHz, IFBW 3 kHz, 7 dBm, through
≤ 3.6 mdB rms @ 23 ± 5 °C, 6 GHz to 7.5 GHz, IFBW 3 kHz, 5 dBm, through
≤ 6 mdB rms @ 23 ± 5 °C, 7.5 GHz to 8.5 GHz, IFBW 3 kHz, 5 dBm, through
Test equipment
44 Chapter 2
Performance Tests
4. TRACE NOISE CW TEST
Procedure
Step 1. Connect test equipment (N-N cable) as shown in the following figures:
Opt. 213/214: Figure 2-5
Opt. 313/314: Figure 2-6
Opt. 413/414: Figure 2-7
2. Performance Test
Figure 2-6 Trace Noise CW Test setup (Opt. 313/314)
Chapter 2 45
Performance Tests
4. TRACE NOISE CW TEST
NOTE Do not touch or move the N-N cable during measurements. To do so will cause a variance
in measured values.
Frequency span: 0 Hz - -
IF Bandwidth: 3 kHz - - -
46 Chapter 2
Performance Tests
4. TRACE NOISE CW TEST
Step 7. Press - - - to set the power level to 10 dBm, which is the value of
the trace noise test levels listed in Table 2-6 or Table 2-7.
Step 9. Record the s.dev value in the calculation sheet ("s.dev [μU]" column for the CW trace
noise test).
Step 10. Calculate the dB value of the trace noise level using the following equation and record the
2. Performance Test
calculated value in the calculation sheet ("Trace noise level [dB rms]" column).
Step 11. Change the E5070B/E5071B center frequency in accordance with Table 2-6 or Table 2-7
and perform Step 8 through 10 for each frequency.
Table 2-6 Trace noise CW Test test frequencies/power level for serial prefix MY421, or
JP1KK
Table 2-7 Trace noise CW Test test frequencies/power level for serial prefix MY422, and
above
NOTE The CW trace noise level is tested at the minimum and maximum frequencies of the
specified frequency range, the frequencies where an internal frequency divider works
(between 1.3 GHz and 2.1 GHz) and, the frequency where the signal source frequency
band is switched (4.25 GHz).
Step 12. Record the calculated values of the trace noise level in the performance test record ("Test
result [dB rms]" column for the CW trace noise test).
Chapter 2 47
Performance Tests
4. TRACE NOISE CW TEST
NOTE Perform Step 16 through 19 for the option 313, 314, 413 and 414 instruments only.
Step 16. Connect the test equipment (N-N cable) as shown in Figure 2-8 (Opt. 313/314) and Figure
2-9 (Opt. 413/414).
Step 17. Select S-parameter from menu in accordance with Table 2-8.
48 Chapter 2
Performance Tests
4. TRACE NOISE CW TEST
Option
313/314 413/414
1 S23 S43
2 S32 S34
2. Performance Test
Step 18. Set the center frequency to 3 MHz.
Step 19. Perform Step 8 through 12 for each of the S-parameters shown in the first and second rows
in Table 2-8.
Chapter 2 49
Performance Tests
5. CROSSTALK TEST
5. CROSSTALK TEST
Description
This test checks the crosstalks between test ports of the E5070B/E5071B. The crosstalk is
tested by performing "through" measurements with two test ports connected together and,
short-ended "isolation" measurements with the test ports terminated with N-type "Short"
devices. A "through" calibration is performed to have the through measurement data as the
reference to which the isolation measurement data is compared. With segment sweep
points for a specified frequency range, a swept measurement with the short-ended test ports
is repeated 16 times and the measurement data is averaged. The worst crosstalk value is
determined from the peak value of the average data.
Specification
E5071B
E5071B
50 Chapter 2
Performance Tests
5. CROSSTALK TEST
Test equipment
2. Performance Test
Procedure
Step 1. Connect the N-N cable as shown in the following figures:
Opt. 213/214: Figure 2-10
Opt. 313/314: Figure 2-11
Opt. 413/414: Figure 2-12
Chapter 2 51
Performance Tests
5. CROSSTALK TEST
NOTE Do not touch or move the N-N cable during measurements. To do so will cause a variance
in measured values.
52 Chapter 2
Performance Tests
5. CROSSTALK TEST
Step 3. Press and select Lin Mag from display format menu.
IF Bandwidth: 10 Hz - IF Bandwidth - - -
2. Performance Test
Averaging: ON - - (select Averaging) -
Step 5. Set the test frequencies using the Segment Table as follows:
b. Press in the ENTRY block. The softkey menu title area is displayed in blue.
c. Press - (select List Power) - to set the list Power function to ON.
The row (Power) for setting appears in the segment table.
d. Press in the ENTRY block. The window frame of the segment table is
displayed as bright.
e. For the E5070B, proceed to the next Step f. For the E5071B, skip to Step j or Step x.
E5070B
NOTE See Table 2-9 for the entire Segment Table setting for the E5070B.
Chapter 2 53
Performance Tests
5. CROSSTALK TEST
l. Press - - to set the number of sweep points to 50. "Power" will be focused.
m. Press - - to set the power level to 10 dBm.
o. Press - - - - - - - - - - - to enter
4.250000001 GHz as the start frequency. "Stop" frequency will be focused.
p. Press - to enter 6 GHz as the stop frequency. The "Points" will be focused.
q. Press - - to set the number of sweep points to 20. "Power" will be focused.
r. Press - to set the power level to 8 dBm.
t. Press - - - - - - - - - - - to enter
6.000000001 GHz as the start frequency. "Stop" frequency will be focused.
u. Press - - - to enter 8.5 GHz as the stop frequency. "Points" will be
focused.
v. Press - - to set the number of sweep points to 30. "Power" will be focused.
w. Press - to set the power level to 6 dBm.
NOTE See Table 2-10 for the entire Segment Table setting for the E5071B (for serial prefix
MY421, or JP1KK).
Table 2-10 E5071B Segment Table setting (for serial prefix MY421, or JP1KK)
z. Press - - to set the number of sweep points to 35. "Power" will be focused.
aa. Press - - to set the power level to 10 dBm.
54 Chapter 2
Performance Tests
5. CROSSTALK TEST
will be focused.
ae. Press - - to set the number of sweep points to 15. "Power" will be focused.
af. Press - to set the power level to 9 dBm.
2. Performance Test
ai. Press - to enter 6 GHz as the stop frequency. The "Points" will be focused.
aj. Press - - to set the number of sweep points to 20. "Power" will be focused.
ak. Press - to set the power level to 7 dBm.
am.Press - - - - - - - - - - - to enter
6.000000001 GHz as the start frequency. "Stop" frequency will be focused.
an. Press - - - to enter 8.5 GHz as the stop frequency. "Points" will be
focused.
ao. Press - - to set the number of sweep points to 30. "Power" will be focused.
ap. Press - to set the power level to 5 dBm.
NOTE See Table 2-11 for the entire Segment Table setting for the E5071B (for serial prefix
MY422, and above).
Table 2-11 E5071B Segment Table setting (for serial prefix MY422, and above)
Response calibration
Chapter 2 55
Performance Tests
5. CROSSTALK TEST
Step 12. Select 2-1 (S21) for the option 213/214/313/314 instruments and 3-2 (S32) for the option
413/414 instruments as shown in the first row in Table 2-12.
Table 2-12 Select Ports settings for response calibration
Option
Step 13. Press and to perform Thru calibration. Wait until the calibration is completed.
Step 15. Change Select Ports setting as shown in the second row in Table 2-12.
Step 17. For the option 213/214 instruments, skip to Step 20. For the option 313/314/413/414,
proceed to Step 18.
Response calibration for options 313, 314, 413 and 414 only
Step 18. Connect the test equipment (N-N cable) as shown in Figure 2-13 (Opt. 313/314) and Figure
2-14 (Opt. 413/414).
56 Chapter 2
Performance Tests
5. CROSSTALK TEST
2. Performance Test
Figure 2-14 Response calibration setup for crosstalk test (Opt.413/414)
Step 19. Perform the response calibration for each of the Select Ports settings shown in the third
and fourth rows in Table 2-12.
Crosstalk measurement
Step 21. Disconnect the N-N cable and connect an N-type coaxial Short termination to each test port
as shown in the following figures:
Chapter 2 57
Performance Tests
5. CROSSTALK TEST
58 Chapter 2
Performance Tests
5. CROSSTALK TEST
2. Performance Test
Step 22. Program or download the VBA test program shown in Program 2-1.
NOTE Also set a UserForm window to enable the test result output data to be displayed. An
example of the UserForm window is shown in Figure 2-18. Refer to the Agilent
E5070B/E5071B VBA Programmer's Guide for the UserForm setup procedure.
Step 23. Press and select the following S-parameter from menu:
Option 213/214/313/314: S21
Option 413/414: S32
Option
Chapter 2 59
Performance Tests
5. CROSSTALK TEST
NOTE The number of averaging times is displayed in the status bar of the measurement channel
window.
Step 25. Press and to set the trigger operation to Hold mode.
Step 26. Press to execute the test program. This test program searches the maximum
point of the trace for each of the following frequency ranges.
E5071B: 3 MHZ to 3 GHz, 3 GHz to 6 GHz, 6 GHz to 7.5 GHz and 7.5 GHz to
8.5 GHz (four ranges)
The test results are represented in dB and displayed in the VBA UserForm window as
shown in the display example below.
Step 27. Read the maximum value(s) and record them in the performance test record ("Test result
[dB]" column for the crosstalk test).
Step 30. Change the S-parameter setting in accordance with Table 2-13 (the second through fourth
rows) and perform Step 24 through Step 29.
Step 3. Click Insert and select Module from pull-down menu. VBA Project - Module1 [Code]
window screen for coding a program will be displayed.
60 Chapter 2
Performance Tests
5. CROSSTALK TEST
Ver = Name
2. Performance Test
If Ver = "E5070B" Then
SCPI.CALCulate(1).PARameter(1).SELect
SCPI.CALCulate(1).SELected.Function.TYPE = "MAX"
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 3000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 3000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValA = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 3000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 6000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValC = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 6000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 7500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
Chapter 2 61
Performance Tests
5. CROSSTALK TEST
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValD = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 7500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 8500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValE = SCPI.CALCulate(1).SELected.Function.DATA
End If
End Sub
Function Log10(X)
End Function
62 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Description
This test checks the system dynamic range for the receiver ports of the E5070B/E5071B.
The system dynamic range is tested by performing an "Isolation" measurement 16 times
with segment sweep points for specified frequency ranges (after the response and isolation
2. Performance Test
calibrations are performed) and calculating the RMS deviation value from the 16
measurement data for each sweep frequency point. The maximum RMS deviation value in
each frequency range is extracted to determine the system dynamic range performance.
Specification
E5071B
Chapter 2 63
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
E5071B
NOTE The specification applies when the response and isolation calibration is performed and
averaging factor is 16.
Test equipment
Procedure
Step 1. Connect a load termination to each test port as shown in the following figures:
Opt. 213/214: Figure 2-19
Opt. 313/314: Figure 2-20
Opt. 413/414: Figure 2-21
64 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Figure 2-19 Isolation calibration setup for system dynamic range test (Opt. 213/214)
2. Performance Test
Figure 2-20 Isolation calibration setup for system dynamic range test (Opt. 313/314)
Chapter 2 65
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Figure 2-21 Isolation calibration setup for system dynamic range test (Opt. 413/414)
Step 3. Press and select Lin Mag from display format menu.
IF Bandwidth: 10 Hz - IF Bandwidth - - -
Step 5. Set the test frequencies using the Segment Table as follows:
b. Press in the ENTRY block. The softkey menu title area is displayed in blue.
c. Press - (select List Power) - to set the List Power function to ON.
The row (Power) for setting appears in the segment table.
d. Press in the ENTRY block. The window frame of the segment table is
66 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
displayed as bright.
e. For the E5070B, proceed to the next Step f. For the E5071B, skip to Step j or Step x.
E5070B
2. Performance Test
i. Press - - to set the power level to 10 dBm.
NOTE See Table 2-14 for the entire Segment Table setting for the E5070B.
o. Press - - - - - - - - - - - to enter
4.250000001 GHz as the start frequency. "Stop" frequency will be focused.
p. Press - to enter 6 GHz as the stop frequency. The "Points" will be focused.
q. Press - - to set the number of sweep points to 20. "Power" will be focused.
r. Press - to set the power level to 8 dBm.
t. Press - - - - - - - - - - - to enter
6.000000001 GHz as the start frequency. "Stop" frequency will be focused.
u. Press - - - to enter 8.5 GHz as the stop frequency. "Points" will be
focused.
v. Press - - to set the number of sweep points to 30. "Power" will be focused.
w. Press - to set the power level to 6 dBm.
Chapter 2 67
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
NOTE See Table 2-15 for the entire Segment Table setting for the E5071B (for serial prefix
MY421, or JP1KK).
Table 2-15 E5071B Segment Table setting (for serial prefix MY421, or JP1KK)
z. Press - - to set the number of sweep points to 35. "Power" will be focused.
aa. Press - - to set the power level to 10 dBm.
am.Press - - - - - - - - - - - to enter
6.000000001 GHz as the start frequency. "Stop" frequency will be focused.
an. Press - - - to enter 8.5 GHz as the stop frequency. "Points" will be
focused.
ao. Press - - to set the number of sweep points to 30. "Power" will be focused.
68 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
NOTE See Table 2-16 for the entire Segment Table setting for the E5071B (for serial prefix
MY422, and above).
Table 2-16 E5071B Segment Table setting (for serial prefix MY422, and above)
2. Performance Test
2 3.000000001 GHz 4.25 GHz 15 9 dBm
Step 12. Select 2-1 (21) for the option 213/214 instruments and 3-2 (S32) for the option
313/314/413/414 instruments as shown in the first row in Table 2-17.
Table 2-17 Select Ports settings for response calibration
Option
Step 13. Press - and to perform Isolation calibration. Wait until the calibration is
completed.
Chapter 2 69
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Step 14. Disconnect the load and connect the N-N cable as shown in the following figures.
Opt. 213/214: Figure 2-22
Opt. 313/314: Figure 2-23
Opt. 413/414: Figure 2-24
Figure 2-22 Thru calibration setup for system dynamic range test setup (Opt. 213/214)
Figure 2-23 Thru calibration setup for system dynamic range test setup (Opt. 313/314)
70 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Figure 2-24 Thru calibration setup for system dynamic range test setup (Opt. 413/414)
2. Performance Test
NOTE To avoid possible variance in measured values, do not touch or move the N-N cable during
measurements.
Step 15. Press and to perform Thru calibration. Wait until the calibration is completed.
Step 17. Press - - to select Select Ports (in reference to the position of Done) and
press .
Step 18. Change Select Ports setting as shown in the second row in Table 2-17.
Step 19. Press - to perform Thru calibration. Wait until the calibration is completed.
Step 20. Disconnect the N-N cable and connect the Load termination to each test port as shown in
the following figures.
Opt. 213/214: Figure 2-19
Opt. 313/314: Figure 2-20
Opt. 413/414: Figure 2-21
System dynamic rang measurement for select ports setting as shown in the first and
second row in Table 2-17
Chapter 2 71
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Step 25. Program or download the test program shown in Program 2-2.
NOTE Also set a UserForm window to enable the test result output data to be displayed. An
example of the UserForm window is shown in Figure 2-25. Refer to the Agilent
E5070B/E5071B VBA Programmer's Guide for the UserForm setup procedure.
Step 26. Press and select the following S-parameter from menu:
Option 213/214: S21
Option 313/314/413/414: S32
Option
Step 27. Press to run the test program. This test program repeats a swept measurement 16
times, calculates the RMS deviation from the 16 measurement values at each sweep
frequency point, and extracts the maximum deviation value for each of the following
frequency ranges:
3 MHz to 1.5 GHz, 1.5 GHz to 4 GHz, 4 GHz to 6 GHz, 6 GHz to 7.5
GHz and 7.5 GHz to 8 GHz (5 ranges)
The maximum deviation value in each frequency range is represented in dB and displayed
in the VBA UserForm window as shown in the display example below.
72 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
2. Performance Test
Step 28. Record the maximum RMS deviation values in the performance test record ("Test result
[dB]" column for the system dynamic range test).
Step 29. Press Exit button on the SYSTEM DYNAMIC RANGE TEST window.
Response (Thru) calibration for options 313, 314, 413 and 414 only
Step 33. Connect the Load termination to each test port as shown in Figure 2-26 (Opt. 313/314) and
Figure 2-27 (Opt. 413/414).
Figure 2-26 Isolation calibration setup for system dynamic range test (Opt. 313/314)
Chapter 2 73
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Figure 2-27 Isolation calibration setup for system dynamic range test (Opt. 413/414)
IF Bandwidth: 10 Hz - IF Bandwidth - - -
Step 38. Select 2-1 (21) for the option 313/314 instruments and 4-1 (S41) for the option 413/414
instruments as shown in the third row in Table 2-19.
Table 2-19 Select Ports settings for response calibration
Option
74 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Option
Step 39. Press - and to perform Isolation calibration. Wait until the calibration is
2. Performance Test
comleted.
Step 40. Disconnect the load and connect the N-N cable as shown in the following figures.
Opt. 313/314: Figure 2-28
Opt. 413/414: Figure 2-29
Figure 2-28 Thru calibration setup for system dynamic range test setup (Opt. 313/314)
Chapter 2 75
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Figure 2-29 Thru calibration setup for system dynamic range test setup (Opt. 413/414)
Step 41. Press and to perform Thru calibration. Wait until the calibration is completed.
Step 43. Press - - to select Select Ports (in reference to the position of Done) and
press .
Step 44. Change Select Ports setting as shown in the fourth row in Table 2-19.
Step 46. Disconnect the N-N cable and connect the Load termination to each test port as shown in
the following figures.
Opt. 313/314: Figure 2-26
Opt. 413/414: Figure 2-27
System dynamic rang measurement for select ports setting as shown in the third and
fourth row in Table 2-19
76 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Step 51. Press and select the following S-parameter from menu:
Option 313/314: S21
Option 413/414: S41
Option
2. Performance Test
213/214 313/314 413/414
Step 53. Record the maximum RMS deviation values in the performance test record ("Test result
[dB]" column for the system dynamic range test).
Step 54. Press Exit button on the SYSTEM DYNAMIC RANGE TEST window.
Chapter 2 77
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
Step 3. Click Insert and select Module from pull-down menu. VBA Project - Module1 [Code]
window screen for coding a program will be displayed.
Step 4. Create the test program. The sample program is show below. (if serial prefix MY422, and
above)
Ver = Name
SCPI.TRIGger.SEQuence.Source = "BUS"
SCPI.INITiate(1).CONTinuous = True
For I = 0 To 15
SCPI.TRIGger.SEQuence.SINGle
Dmy = SCPI.IEEE4882.OPC
SCPI.CALCulate(1).PARameter(1).SELect
Val(I) = SCPI.CALCulate(1).SELected.DATA.FDATa
Next I
78 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
For J = 0 To Count
X(J) = Val(0)(J) ^ 2
Next J
For J = 0 To Count
For I = 1 To 15
2. Performance Test
X(J) = X(J) + Val(I)(J) ^ 2
Next I
Y(J) = Sqr(X(J) / 16)
Next J
SCPI.INITiate(1).CONTinuous = False
Rms = CVar(Y)
SCPI.CALCulate(1).SELected.DATA.FDATa = Rms
SCPI.DISPlay.WINDow(1).TRACe.Y.SCALe.AUTO
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 1500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 3000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValB = SCPI.CALCulate(1).SELected.Function.DATA
Chapter 2 79
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
SCPI.CALCulate(1).SELected.Function.TYPE = "MAX"
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 3000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 1500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValC = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 1500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 3000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValD = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 3000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 4000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValE = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 4000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 6000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValF = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 6000000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 7500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValG = SCPI.CALCulate(1).SELected.Function.DATA
SCPI.CALCulate(1).SELected.Function.DOMain.STARt = 7500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STOP = 8500000000#
SCPI.CALCulate(1).SELected.Function.DOMain.STATe = True
SCPI.CALCulate(1).SELected.Function.EXECute
MaxValH = SCPI.CALCulate(1).SELected.Function.DATA
80 Chapter 2
Performance Tests
6. SYSTEM DYNAMIC RANGE TEST
2. Performance Test
Else
Buff = MsgBox("Error: This program is designed for the
E5070B/E5071B, then please use the corresponding programs.",
vbOKOnly, "SYSTEM DYNAMIC RANGE TEST")
End If
End Sub
Chapter 2 81
Performance Tests
7. DYNAMIC ACCURACY TEST
Description
This test checks the dynamic accuracy of the E5070B/E5071B. The dynamic accuracy is
tested at 1.195 GHz for each receiver port using the Agilent Z5623A (with option H01)
Dynamic Accuracy Test Kit. A block diagram of the Z5623A is shown in Figure 2-31. The
1.195 GHz source signal of the E5070B/E5071B goes through the Z5623A's first
attenuator variable from 0 dB to 11 dB in 1 dB steps and second attenuator variable up to
100 dB in 10 dB steps and is measured with the receiver port. The test procedure checks
the measured receiver-input power for each 5 dB increment in the range from 10 dBm to
100 dBm. The ratios of the measured receiver-input powers to the reference input level of
-10 dBm are calculated and compared to the attenuation values of the Z5623A. The 1.195
GHz signal flow is split with a power splitter present between the two attenuators in the
Z5623A and measured with an external power meter to calibrate the attenuated power
level.
NOTE Since the dynamic accuracy does not have frequency dependence, this test is performed at
1.195 GHz only.
-45 0.078
Test Equipment
82 Chapter 2
Performance Tests
7. DYNAMIC ACCURACY TEST
Coaxial cable with N-type (m) connectors, 61 cm (24 Agilent N6314A (p/n
in), 2 ea. 8120-8862)
2. Performance Test
GPIB controller (or PC with GPIB Interface Card)
Procedure
Step 1. Connect the power sensor to the power meter. Calibrate the power meter for the power
sensor used.
Chapter 2 83
Performance Tests
7. DYNAMIC ACCURACY TEST
84 Chapter 2
Performance Tests
7. DYNAMIC ACCURACY TEST
2. Performance Test
NOTE To avoid possible variance in measured values, do not touch or move the N-N cable during
measurements.
NOTE This adjustment is required to subtract residual noise power of the Z5623A from the
subsequent signal power measurements.
Step 3. Set the first attenuator (A1) of the Z5623A to 11 dB and the second attenuator (A2) to 110
dB by executing the following GPIB control command:
GPIB:
Chapter 2 85
Performance Tests
7. DYNAMIC ACCURACY TEST
Frequency span: 0 Hz - -
IF Bandwidth: 10 Hz - - - -
Step 6. Wait for the power meter reading to settle and perform zero adjustment of the power meter.
Step 7. Press and select Log Mag from display format menu.
Frequency span: 0 Hz - -
Step 10. Set Port Char function to OFF in accordance with the following procedure:
a. Press - - - - - - - - - (select
Service Menu), .
Step 14. Set the first attenuator (A1) of the Z5623A to 0 dB and the second attenuator (A2) to 10 dB
86 Chapter 2
Performance Tests
7. DYNAMIC ACCURACY TEST
GPIB:
Step 15. Adjust the source power for 0 dBm on the power meter display. Wait for power meter
reading to settle when verifying the adjustment result.
2. Performance Test
NOTE Since the source power resolution is 0.05 dB, adjust for the nearest possible value to 0
dBm.
Step 17. Record the value of the source power setting in the calculation sheet ("Power setting [dB]"
column for the dynamic accuracy test). The record starts from DA10 row.
Step 18. Record the power meter reading in the calculation sheet ("Pm [dB]" column).
Step 19. Read the mean value of the T2(S21) displayed in the upper left corner of the graphic
display and record it in the calculation sheet ("Tx [dB]" column).
Step 20. Set the first attenuator (A1) of the Z5623A to 5 dB by executing the following GPIB
control command:
GPIB:
Step 22. Calculate the dynamic accuracy for -15 dBm using the following equation:
NOTE The records for -10 dBm are used as the reference values in the dynamic accuracy
calculation.
Step 23. Record the calculated value in the calculation sheet ("Running D.A. [dB]" column for
DA15).
Step 24. Record the same value in the "True D.A. [dB]" column in the calculation sheet.
Step 25. Set the first attenuator (A1) of the Z5623A to 10 dB by executing the following GPIB
control command:
GPIB:
Chapter 2 87
Performance Tests
7. DYNAMIC ACCURACY TEST
Step 27. Calculate the dynamic accuracy for -20 dBm using the following equation:
Step 28. Record the calculated value in the calculation sheet ("Running D.A. [dB]" and "True D.A.
[dB]" columns for DA20).
Step 29. Set the first attenuator (A1) of the Z5623A to 0 dB and the second attenuator (A2) to 20 dB
by executing the following GPIB control command:
GPIB:
NOTE This attenuator state is named DA20a. The following Steps 30 through 32 are performed to
equalize the receiver port input levels for the DA20 and DA20a.
Step 31. Calculate the difference in Tx [dB] values for the DA20 and DA20a by using the following
equation:
Step 33. Adjust the source power for the following value on the power meter display.
NOTE It will be possible to promptly approximate the source power to the target value by
decreasing the source power setting by Δ ( Tx ) from that for the DA20.
Step 35. Set the first attenuator (A1) of the Z5623A to 5 dB by executing the following GPIB
control command:
GPIB:
Step 37. Calculate the dynamic accuracy for -25 dBm using the following equation:
88 Chapter 2
Performance Tests
7. DYNAMIC ACCURACY TEST
Step 38. Record the calculated D.A. in the calculation sheet ("Running D.A. [dB]" column for
DA25).
Step 39. Calculate the true dynamic accuracy using the following equation:
True D.A. = (Running D.A. [dB] @ DA25) + (True D.A. [dB] @ DA20)
NOTE When the A2 decade attenuator setting is increased from X dB to X+10 dB, calculate the
True D.A. value as follows:
2. Performance Test
True D.A. (X+10) = Running D.A. (X+10) + True D.A. (X)
Where True D.A. (X+10): True D.A. value when A2=X+10 dB,
True D.A. (X): True D.A. value when A1=10 dB and A2=X dB
Step 40. Record the true D.A. in the calculation sheet ("True D.A." column for DA25).
Step 41. Set the first attenuator (A1) of the Z5623A to 10 dB by executing the following GPIB
control command:
GPIB:
Step 43. Calculate the dynamic accuracy for -30 dBm using the following equation:
Step 44. Record the calculated D.A. in the calculation sheet ("Running D.A. [dB]" column for
DA30).
Step 45. Calculate the true dynamic accuracy using the following equation:
True D.A. = (Running D.A. [dB] @ DA30) + (True D.A. [dB] @ DA20)
Step 46. Record the true D.A. in the calculation sheet ("True D.A." column for DA30).
Step 47. Subsequently perform Step 29 through 45 for the attenuator settings of 35 dB to 100 dB
(DA30a to DA100) shown in Table 2-22.
Table 2-22 Z5623A attenuator settings
Chapter 2 89
Performance Tests
7. DYNAMIC ACCURACY TEST
90 Chapter 2
Performance Tests
7. DYNAMIC ACCURACY TEST
2. Performance Test
OUTPUT 712; "at_b=60"
Step 48. Set the first attenuator (A1) of the Z5623A to 10 dB and the second attenuator (A2) to 0 dB
by executing the following GPIB control command:
GPIB:
Step 49. Adjust the source power for -10 dBm on the power meter display. Wait for power meter
Chapter 2 91
Performance Tests
7. DYNAMIC ACCURACY TEST
NOTE Since the source power resolution is 0.05 dB, adjust for the nearest possible value to -10
dBm.
Step 51. Set the first attenuator (A1) of the Z5623A to 5 dB by executing the following GPIB
control command:
GPIB:
Step 53. Calculate the dynamic accuracy for -5 dBm using the following equation:
Step 54. Record the calculated D.A. in the calculation sheet ("Running D.A. [dB]" and "True D.A.
[dB]" columns for DA5).
Step 55. Set the first attenuator (A1) of the Z5623A to 0 dB by executing the following GPIB
control command:
GPIB:
Step 57. Calculate the dynamic accuracy for 0 dBm using the following equation:
Step 58. Record the calculated D.A. in the calculation sheet ("Running D.A. [dB]" and "True D.A.
[dB]" columns for DA0).
Step 60. Adjust the source power for 5 dBm on the power meter display. Wait for power meter
reading to settle when verifying the adjustment result.
NOTE Since the source power resolution is 0.05 dB, adjust for the nearest possible value to 5
dBm.
Step 62. Calculate the dynamic accuracy for 5 dBm using the following equation:
Step 63. Record the true D.A. in the calculation sheet ("True D.A." column for DA+5).
92 Chapter 2
Performance Tests
7. DYNAMIC ACCURACY TEST
Step 65. Adjust the source power for 10 dBm on the power meter display. Wait for power meter
reading to settle when verifying the adjustment result.
NOTE Since the source power resolution is 0.05 dB, adjust for the nearest possible value to 10
dBm.
2. Performance Test
Step 67. Calculate the dynamic accuracy for 10 dBm using the following equation:
Step 68. Record the true D.A. in the calculation sheet ("True D.A." column for DA+10).
Step 69. Connect the test equipment and select S-parameter in accordance with Table 2-23.
Step 70. Perform Step 10 through 67 for each test setup and S-parameter shown in Table 2-23.
Table 2-23 Dynamic accuracy test setups and S-parameter settings
Chapter 2 93
Performance Tests
7. DYNAMIC ACCURACY TEST
94 Chapter 2
Performance Tests
7. DYNAMIC ACCURACY TEST
2. Performance Test
Figure 2-38 Dynamic accuracy test setup (Opt. 413/414)
Chapter 2 95
Performance Tests
7. DYNAMIC ACCURACY TEST
96 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Description
This test checks the directivity, source match, load match, transmission tracking, and
reflection tracking which are the key S-parameter measurement hardware characteristics.
These characteristics are tested using the Agilent 85032F Calibration Kit at 3MHz, 10
2. Performance Test
MHz, as well as 50 MHz increments from 50 MHz to 3 GHz (up to 8.5 GHz for the
E5071B). After a full two-port (SOLT) calibration is performed for each test port, VBA
test programs are executed to output the directivity, source match, load match, transmission
tracking, and reflection tracking data to the display.
NOTE The VBA test programs for the Uncorrected System Performance Test are not available in
the current version of the E5070B/E5071B test procedure. The test procedure is made
effective when the test programs become available.
Specification
Table 2-24 Uncorrected System Performance (Correction: Off, Sys-
tem Correction: On)
Directivity 25 dB 20 dB 15 dB
Source Match 25 dB 20 dB 15 dB
Load Match 17 dB 12 dB 10 dB
Test Equipment
Coaxial cable with N-type (m) connectors, 61 cm (24 in) Agilent N6314A (p/n
8120-8862)
Procedure
Step 1. Connect the N-N cable as shown in the following figure:
Option 213/214: Figure 2-41.
Option 313/314: Figure 2-42.
Option 413/414: Figure 2-43.
For the connection of the Open, Short and Load terminations, obey the instruction in the
Chapter 2 97
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
following procedure.
Figure 2-41 O/S/L calibration setup for uncorrected system performance test (Opt. 213/214)
Figure 2-42 O/S/L calibration setup for uncorrected system performance test (Opt. 313/314)
98 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Figure 2-43 O/S/L calibration setup for uncorrected system performance test (Opt. 413/414)
2. Performance Test
NOTE To avoid possible variance in measured values, do not touch or move the N-N cable during
measurements.
Step 3. Set System Correction function to ON in accordance with the following procedure:
a. Press - - - - - - - - - (select
Service Menu), .
b. Confirm that the System Correction is ON. If it is OFF, proceed to the next c and d.
Step 4. Press and select Log Mag from display format menu.
Step 5. Set the test frequencies using the Segment Table as follows:
Chapter 2 99
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
automatically set to 2.
NOTE The sweep frequency points for the second segment will be aligned in 50 MHz increments
from 50 MHz to 3 GHz.
NOTE See Table 2-25 for the entire Segment Table setting for the E5070B.
1 3 MHz 10 MHz 2
3 50 MHz 3 GHz 60
E5071B
i. Press - - - to enter 8.5 GHz as the stop frequency. "Points" will be
focused.
j. Press - - - to set the number of sweep points to 170.
NOTE The sweep frequency points for the second segment will be aligned in 50 MHz increments
from 50 MHz to 8.5 GHz.
NOTE See Table 2-26 for the entire Segment Table setting for the E5071B.
1 3 MHz 10 MHz 2
100 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Step 9. Set Offset Delay and Offset Loss of the N-N cable in accordance with the following
procedure:
2. Performance Test
b. Press to select Define STDs function.
c. Press - - - - - - - - - and
to select 11. Thru function.
d. Press - - - - - - - - - and ,
then press - - - - to enter 2.68 nsec as the Offset Delay.
Step 14. Press and to select Reflection calibration and to go down to its menu.
Step 15. Connect Open termination to the Port 1 and perform Port 1 Open calibration.
Step 16. Perform Port 1 Short calibration with Short termination connected to the Port 1.
Step 17. Perform Port 1 Load calibration with Load termination Connected to the Port 1.
Step 18. Connect Open termination to the tip of the N-N cable connected to the Port 2 and perform
Port 2 open calibration.
Step 19. Perform Port 2 Short calibration with Short termination connected to the tip of the N-N
cable.
Step 20. Perform Port 2 Load calibration with Load termination connected to the tip of the N-N
cable.
Step 22. Connect the N-N cable between the selected test ports (Port 1 and Port 2) as shown in the
following figures:
Option 213/214: Figure 2-44.
Option 313/314: Figure 2-45.
Option 413/414: Figure 2-46.
Chapter 2 101
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Figure 2-44 Thru calibration setup for uncorrected system performance test (Opt. 213/214)
Figure 2-45 Thru calibration setup for uncorrected system performance test (Opt. 313/314)
102 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Figure 2-46 Thru calibration setup for uncorrected system performance test (Opt. 413/414)
2. Performance Test
Step 23. Press and to select Transmission calibration and to go down to its menu.
Step 27. Press and verify that the Correction function is set to ON.
Directivity, source match, load match, transmission tracking, and reflection tracking
test
Step 29. Press to run the test program. This test program displays the values of the
directivity, source match and load match at each sweep frequency point, and extracts the
maximum value for each of the following frequency ranges:
The maximum value in each frequency range is represented in dB and displayed in the
VBA UserForm window as shown in the display example below.
a. Set Parameter, Response Port and Stimulus Port as shown in Figure 2-47.
Chapter 2 103
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
b. Press OK button. The maximum value in each frequency range is displayed as shown in
Figure 2-48 or Figure 2-49.
c. Record the displayed values in the performance test record (“Test results [dB]” column
in “Port 1” table for the uncorrected system performance test).
Source match test for port 1
d. Set Parameter, Response Port and Stimulus Port as shown in Figure 2-50.
e. Press OK button. The maximum value in each frequency range is displayed as shown in
Figure 2-51 or Figure 2-52.
104 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
2. Performance Test
Figure 2-52 Source match for E5071B
f. Record the displayed values in the performance test record (“Test results [dB]” column
in “Port 1” table for the uncorrected system performance test).
Load match test for port 1
g. Set Parameter, Response Port and Stimulus Port as shown in Figure 2-53.
h. Press OK button. The maximum value in each frequency range is displayed as shown in
Figure 2-54 or Figure 2-55.
Chapter 2 105
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
i. Record the displayed values in the performance test record (“Test results [dB]” column
in “Port 1” table for the uncorrected system performance test).
Transimission Tracking test for port 1
j. Set Parameter, Response Port and Stimulus Port as shown in Figure 2-56.
k. Press OK button. The maximum value in each frequency range is displayed as shown in
Figure 2-57 or Figure 2-58.
l. Record the displayed values in the performance test record (“Test results [dB]” column
in “Port 1” table for the uncorrected system performance test).
106 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
2. Performance Test
n. Press OK button. The maximum value in each frequency range is displayed as shown in
Figure 2-60 or Figure 2-61.
o. Record the displayed values in the performance test record (“Test results [dB]” column
in “Port 1” table for the uncorrected system performance test).
Step 30. Connect the N-N cable as shown in the following figures:
Option 213/214: Figure 2-62.
Option 313/314: Figure 2-63.
Option 413/414: Figure 2-64.
NOTE The test setups and Select Ports settings for each option configuration are described in
Table 2-27.
Chapter 2 107
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Figure 2-62 O/S/L calibration setup for uncorrected system performance test (Opt. 213/214)
Figure 2-63 O/S/L calibration setup for uncorrected system performance test (Opt. 313/314)
108 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Figure 2-64 O/S/L calibration setup for uncorrected system performance test (Opt. 413/414)
2. Performance Test
Step 31. Perform the full 2-port calibration (Step 10 through 27) in accordance with Table 2-27.
Step 32. Perform Step 29 and 30. Record the directivity, source match and load match values in the
performance test record ("Port 2" table).
NOTE Perform Step 33 and 34 for the option 313, 314, 413 and 414 instruments only.
Step 33. Perform the full 2-port calibration (Step 6 through 27) in accordance with Table 2-27.
Step 34. Perform Step 29 and 30. Record the directivity, source match and load match values in the
performance test record ("Port 1" through "Port 3" tables for Option 313/314 and "Port 1"
through "Port 4" tables for Option 413/414).
Chapter 2 109
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Table 2-27 Test setup, Select Ports settings and calibration sequence
1-2 Figure 2-62 Port 1 Open "Open" to the tip of N-N cable
110 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Table 2-27 Test setup, Select Ports settings and calibration sequence
313/314 1-2 Figure 2-42 Port 1 Open/Short/Load and Same as Opt. 213/214 (N-N cable
Port 2 Open/Short/Load connected to Port 2)
Refer to Step 10 through 27 for the
details.
2. Performance Test
Figure 2-45 Thru N-N cable between Port 1 and Port 2
1-2 Figure 2-63 Port 1 Open/Short/Load and Same as Opt. 213/214 (N-N cable
Port 2 Open/Short/Load connected to Port 1)
1-3 Figure 2-64 Port 1 Open "Open" to the tip of N-N cable
Chapter 2 111
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Table 2-27 Test setup, Select Ports settings and calibration sequence
413/414 1-2 Figure 2-43 Port 1 Open/Short/Load and Same as Opt. 213/214 (N-N cable
Port 2 Open/Short/Load connected to Port 2)
Refer to Step 10 through 27 for the
details.
1-2 Figure 2-64 Port 1 Open/Short/Load and Same as Opt. 213/214 (N-N cable
Port 2 Open/Short/Load connected to Port 1)
112 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Table 2-27 Test setup, Select Ports settings and calibration sequence
3-4 Figure 2-69 Port 3 Open "Open" to the tip of N-N cable
2. Performance Test
Port 4 Open "Open" to Port 4
Perform Step 29 and 30. Record the test results in the performance test record ("port 4")
Response/Stimulus Port of Directivity/Source match/Load match/Transmission
tracking/Reflection tracking test are as follows;
Response Port: 4, Stimulus Port : 3
Record the test results in the performance test record ("port 4").
Figure 2-65 O/S/L calibration setup for uncorrected system performance test (Opt. 313/314)
Chapter 2 113
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Figure 2-66 Thru calibration setup for uncorrected system performance test (Opt. 313/314)
Figure 2-67 O/S/L calibration setup for uncorrected system performance test (Opt. 413/414)
114 Chapter 2
Performance Tests
8. UNCORRECTED SYSTEM PERFORMANCE TEST
Figure 2-68 Thru calibration setup for uncorrected system performance test (Opt. 413/414)
2. Performance Test
Figure 2-69 O/S/L calibration setup for uncorrected system performance test (Opt. 413/414)
Chapter 2 115
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
Introduction
This section contains calculation sheets for each performance test that requires additional
calculations to determine the final test result.
Use the calculation sheet in this section as an aid for recording raw measurement data and
calculating the performance test results.
Calculation sheet entries are provided only for performance tests in which calculations are
required to obtain the test results.
[a] [REF]
50 M ——————
10 M —————— a - REF
(E5071B only)
[a] [REF]
116 Chapter 2
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
NOTE Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
E5070B: 10 MHz and 3 GHz
E5071B: 10 MHz, 3 GHz, and 4.25 GHz
2. Performance Test
for serial prefix: MY421, or JP1KK
CW frequency: ______________ Hz
0 ——————
10 —————— a-REF
9 —————— a-REF
8 —————— a-REF
7 —————— a-REF
6 —————— a-REF
5 —————— a-REF
4 —————— a-REF
3 —————— a-REF
2 —————— a-REF
1 —————— a-REF
-1 —————— a-REF
-2 —————— a-REF
-3 —————— a-REF
-4 —————— a-REF
-5 —————— a-REF
-6 —————— a-REF
-7 —————— a-REF
-8 —————— a-REF
-9 —————— a-REF
Chapter 2 117
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
NOTE Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
E5071B: 6 GHz
for serial prefix: MY421, or JP1KK
CW frequency: ______________ Hz
0 ——————
8 —————— a-REF
7 —————— a-REF
6 —————— a-REF
5 —————— a-REF
4 —————— a-REF
3 —————— a-REF
2 —————— a-REF
1 —————— a-REF
-1 —————— a-REF
-2 —————— a-REF
-3 —————— a-REF
-4 —————— a-REF
-5 —————— a-REF
-6 —————— a-REF
-7 —————— a-REF
-8 —————— a-REF
118 Chapter 2
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
-9 —————— a-REF
2. Performance Test
-13 —————— a-REF
NOTE Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
E5071B: 8.5 GHz
for serial prefix: MY421, or JP1KK
CW frequency: ______________ Hz
0 ——————
6 —————— a-REF
5 —————— a-REF
4 —————— a-REF
3 —————— a-REF
2 —————— a-REF
1 —————— a-REF
-1 —————— a-REF
-2 —————— a-REF
-3 —————— a-REF
-4 —————— a-REF
-5 —————— a-REF
-6 —————— a-REF
-7 —————— a-REF
-8 —————— a-REF
Chapter 2 119
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
-9 —————— a-REF
NOTE Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
E5070B: 10 MHz and 3 GHz
E5071B: 10 MHz and 3 GHz
for serial prefix: MY422, and above
CW frequency: ______________ Hz
0 ——————
10 —————— a-REF
9 —————— a-REF
8 —————— a-REF
7 —————— a-REF
6 —————— a-REF
5 —————— a-REF
4 —————— a-REF
3 —————— a-REF
2 —————— a-REF
1 —————— a-REF
-1 —————— a-REF
120 Chapter 2
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
-2 —————— a-REF
-3 —————— a-REF
-4 —————— a-REF
-5 —————— a-REF
2. Performance Test
-6 —————— a-REF
-7 —————— a-REF
-8 —————— a-REF
-9 —————— a-REF
NOTE Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
E5071B: 4.25 GHz
for serial prefix: MY422, and above
CW frequency: ______________ Hz
0 ——————
9 —————— a-REF
8 —————— a-REF
7 —————— a-REF
6 —————— a-REF
5 —————— a-REF
4 —————— a-REF
3 —————— a-REF
Chapter 2 121
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
2 —————— a-REF
1 —————— a-REF
-1 —————— a-REF
-2 —————— a-REF
-3 —————— a-REF
-4 —————— a-REF
-5 —————— a-REF
-6 —————— a-REF
-7 —————— a-REF
-8 —————— a-REF
-9 —————— a-REF
NOTE Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
E5071B: 6 GHz
for serial prefix: MY422, and above
CW frequency: ______________ Hz
0 ——————
7 —————— a-REF
6 —————— a-REF
5 —————— a-REF
4 —————— a-REF
122 Chapter 2
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
3 —————— a-REF
2 —————— a-REF
1 —————— a-REF
-1 —————— a-REF
2. Performance Test
-2 —————— a-REF
-3 —————— a-REF
-4 —————— a-REF
-5 —————— a-REF
-6 —————— a-REF
-7 —————— a-REF
-8 —————— a-REF
-9 —————— a-REF
NOTE Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
E5071B: 8.5 GHz
for serial prefix: MY422, and above
CW frequency: ______________ Hz
0 ——————
5 —————— a-REF
4 —————— a-REF
3 —————— a-REF
Chapter 2 123
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
2 —————— a-REF
1 —————— a-REF
-1 —————— a-REF
-2 —————— a-REF
-3 —————— a-REF
-4 —————— a-REF
-5 —————— a-REF
-6 —————— a-REF
-7 —————— a-REF
-8 —————— a-REF
-9 —————— a-REF
124 Chapter 2
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
NOTE Make a copy of the following calculation sheet for each combination of the S-parameters
Listed below:
Option 212/213: S21and S12
Option 313/314: S21and S12, S23 and S32
Option 413/414: S21and S12, S43 and S34
2. Performance Test
(One calculation sheet can cover 2 parameters.)
CW frequency [Hz] s.dev [μU] Trace noise level [dB rms] Test result equation
3 MHz
3 GHz
(E5071B only)
CW frequency [Hz] s.dev [μU] Trace noise level [dB rms] Test result equation
4.25 GHz
8.5 GHz
CW frequency [Hz] s.dev [μU] Trace noise level [dB rms] Test result equation
3 MHz
3 GHz
Chapter 2 125
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
(E5071B only)
CW frequency [Hz] s.dev [μU] Trace noise level [dB rms] Test result equation
4.25 GHz
8.5 GHz
126 Chapter 2
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
NOTE A dynamic accuracy calculation example is shown at the end of the calculation sheets.
2. Performance Test
S-Parameter: _____________ (10 dBm to -100 dBm)
Adjust source power for (Pm [db] @ DA20a) - ( Tx [dB] @ DA20a) + ( Tx [dB] @ DA20)
Adjust source power for (Pm [db] @ DA30a) - ( Tx [dB] @ DA30a) + ( Tx [dB] @ DA30)
Adjust source power for (Pm [db] @ DA40a) - ( Tx [dB] @ DA40a) + ( Tx [dB] @ DA40)
Chapter 2 127
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
Adjust source power for (Pm [db] @ DA50a) - (Tx [dB] @ DA50a) + ( Tx [dB] @ DA50)
Adjust source power for (Pm [db] @ DA60a) - ( Tx [dB] @ DA60a) + ( Tx [dB] @ DA60)
Adjust source power for (Pm [db] @ DA70a) - ( Tx [dB] @ DA70a) + ( Tx [dB] @ DA70)
128 Chapter 2
Performance Tests
E5070B/E5071B Performance Test Calculation Sheet
Adjust source power for (Pm [db] @ DA80a) - ( Tx [dB] @ DA80a) + ( Tx [dB] @ DA80)
2. Performance Test
(Tx @ DA85) - ( Tx @ DA80b) - (Pm @ DA85) + (Pm @ DA80b) →
True D.A. = (Running D.A. [dB] @ DA85) + (True D.A. [dB] @ DA80)
Adjust source power for (Pm [db] @ DA90a) - ( Tx [dB] @ DA90a) + ( Tx [dB] @ DA90)
Chapter 2 129
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
50 M ± 250 ± 11
3G ± 15 k k ± 0.64 k
50 M ± 50 ± 11
3G ± 3.00 k k ± 0.64 k
0 50 M ± 0.65 ± 0.16
10 M ± 1.0 ± 0.37
130 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
2. Performance Test
3.00 G ± 1.0 ± 0.20
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
Chapter 2 131
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
132 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
-15 ± 0.75 ± 0.17
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
Direction: S12
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
Chapter 2 133
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
Crosstalk Test
Direction: S21 (Options 213, 214, 313 and 314 only)
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
134 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
2. Performance Test
3M-3G 10 < -120
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Chapter 2 135
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
136 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Direction: S21
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
2. Performance Test
-5 ± 0.031 ± 0.012
Direction: S12
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Chapter 2 137
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
138 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
-55 ± 0.106 ± 0.022
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Chapter 2 139
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Transmission 3 M - 3G ± 1.0
Tracking
Port 2
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Transmission 3 M - 3G ± 1.0
Tracking
140 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY421, or JP1KK
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
2. Performance Test
Transmission 3 M - 3G ± 1.0
Tracking
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Transmission 3 M - 3G ± 1.0
Tracking
Chapter 2 141
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
50 M ± 250 ± 11
3G ± 15 k k ± 0.64 k
50 M ± 50 ± 11
3G ± 3.00 k k ± 0.64 k
0 50 M ± 0.65 ± 0.16
10 M ± 1.0 ± 0.37
142 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
2. Performance Test
3.00 G ± 1.0 ± 0.20
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
Chapter 2 143
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
144 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
-15 ± 0.75 ± 0.17
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
Direction: S12
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
Chapter 2 145
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
Crosstalk Test
Direction: S21 (Options 213, 214, 313 and 314 only)
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
146 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
2. Performance Test
3M-3G 10 < -120
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Chapter 2 147
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
148 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Direction: S21
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
2. Performance Test
-5 ± 0.031 ± 0.012
Direction: S12
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Chapter 2 149
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
150 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
-55 ± 0.106 ± 0.022
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Chapter 2 151
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Transmission 3 M - 3G ± 1.0
Tracking
Port 2
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Transmission 3 M - 3G ± 1.0
Tracking
152 Chapter 2
Performance Tests
E5070B Performance Test Record for Serial Prefix: MY422, and above
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
2. Performance Test
Transmission 3 M - 3G ± 1.0
Tracking
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Transmission 3 M - 3G ± 1.0
Tracking
Chapter 2 153
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
50 M ± 250 ± 11
3G ± 15 k k ± 0.64 k
50 M ± 50 ± 11
3G ± 3.00 k k ± 0.64 k
0 50 M ± 0.65 ± 0.16
10 M ± 1.0 ± 0.37
154 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
2. Performance Test
4.25 G ± 1.0 ± 0.36
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
Chapter 2 155
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
156 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
2. Performance Test
-10 ± 0.75 ± 0.17
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
Chapter 2 157
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
158 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
-14 ± 0.75 ± 0.17
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
Chapter 2 159
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.003
Direction: S12
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.003
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
160 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3G 3k < 0.001
2. Performance Test
6G 3k < 0.003
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.003
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.003
Chapter 2 161
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.003
Crosstalk Test
Direction: S21 (Options 213, 214, 313 and 314 only)
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
162 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
2. Performance Test
3G-6G 10 < -110
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Chapter 2 163
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
164 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
2. Performance Test
7.5 G - 8.5 G 10 < -106
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Chapter 2 165
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
166 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Direction: S12
2. Performance Test
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Chapter 2 167
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
168 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
-35 ± 0.056 ± 0.016
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Chapter 2 169
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Port 2
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
170 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
2. Performance Test
Directivity 3M-3G < -25
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Chapter 2 171
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY421, or JP1KK
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
172 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
2. Performance Test
Humidity: % R.H. Tested by:
50 M ± 250 ± 11
3G ± 15 k k ± 0.64 k
50 M ± 50 ± 11
3G ± 3.00 k k ± 0.64 k
0 50 M ± 0.65 ± 0.16
10 M ± 1.0 ± 0.37
Chapter 2 173
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
174 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
2. Performance Test
-9 ± 0.75 ± 0.17
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.75 ± 0.27
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
Chapter 2 175
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
9 ± 0.75 ± 0.26
8 ± 0.75 ± 0.24
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
176 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
2. Performance Test
-12 ± 0.75 ± 0.17
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
7 ± 0.75 ± 0.23
6 ± 0.75 ± 0.21
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
Chapter 2 177
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
5 ± 0.75 ± 0.20
4 ± 0.75 ± 0.19
3 ± 0.75 ± 0.18
2 ± 0.75 ± 0.18
1 ± 0.75 ± 0.17
-1 ± 0.75 ± 0.17
-2 ± 0.75 ± 0.17
-3 ± 0.75 ± 0.17
-4 ± 0.75 ± 0.17
-5 ± 0.75 ± 0.17
-6 ± 0.75 ± 0.17
-7 ± 0.75 ± 0.17
-8 ± 0.75 ± 0.17
-9 ± 0.75 ± 0.17
178 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
2. Performance Test
2.1 G 3k < 0.001
3G 3k < 0.001
6G 3k < 0.0036
Direction: S12
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.0036
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
Chapter 2 179
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
6G 3k < 0.0036
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.0036
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
3G 3k < 0.001
6G 3k < 0.0036
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3M 3k < 0.001
180 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
CW frequency IF bandwidth Test limit [dB rms] Test result [dB rms]
[Hz] [Hz]
3G 3k < 0.001
2. Performance Test
6G 3k < 0.0036
Crosstalk Test
Direction: S21 (Options 213, 214, 313 and 314 only)
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Chapter 2 181
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
182 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
2. Performance Test
Direction: S12 (Options 213, 214, 313 and 314 only)
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Chapter 2 183
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
184 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
2. Performance Test
3 M - 1.5 G 10 < -120
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Chapter 2 185
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Direction: S12
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
186 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
-75 ± 0.248 ± 0.028
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
Chapter 2 187
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.060
5 ± 0.075 ± 0.022
0 ± 0.042 ± 0.012
-5 ± 0.031 ± 0.012
188 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
2. Performance Test
@ Correction: Off, System Correction: On
Port 1
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Port 2
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Chapter 2 189
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
190 Chapter 2
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
2. Performance Test
Reflection Tracking 3G-6G ± 1.0
System Frequency range [Hz] Test limit [dB] Test result [dB]
performance
Chapter 2 191
Performance Tests
E5071B Performance Test Record for Serial Prefix: MY422, and above
192 Chapter 2
1. Chapter Title
2. Chapter Title
3. Adjustment
4. Chapter Title
3 Adjustment
This chapter provides the adjustment information for the E5070B/E5071B ENA Series
5. Chapter Title
Network Analyzer to ensure that the it is within its specifications. The adjustment must be
performed Agilent’s qualified service personnel. If you need the adjustment for your
E5070B/E5071B, it should be sent to the nearest Agilent Technologies service office.
193
Adjustment
Safety Considerations
Safety Considerations
This manual contains NOTEs, CAUTIONs, and WARNINGs that must be followed to
ensure the safety of the operator and to keep the instrument in a safe and serviceable
condition. The adjustment must be performed by Agilent’s qualified service personnel.
WARNING Any interruption of the protective ground conductor (inside or outside the
equipment) or disconnection of the protective ground terminal can make the
instrument dangerous. Intentional interruption of the protective ground system for
any reason is prohibited.
Required Equipment
Table 1-1 on page 24 lists the equipment required to perform the Adjustment procedures
described in this chapter. Use only calibrated test equipment when adjusting the
E5070B/E5071B.
194 Chapter 3
Adjustment
1. Chapter Title
Required Adjustment after Replacing Assembly
2. Chapter Title
Replaced Assembly Adjustment Item
3. Adjustment
Source Board √ √ √
Receiver (RF) Board √ √
Receiver (IF) Board √ √ √
RF Switch √ √
Crystal Oven (Opt. 1E5) √
Attenuator (Opt.214, 314 and 414) √ √
Hard Disk Drive
4. Chapter Title
Analog Interface Board √
Level Vernier √ √
Power Amp Module √ √
Writing ID
This item writes the serial number and the option structure into the E5070B/E5071B.
Chapter 3 195
Adjustment
Required Adjustment after Replacing Assembly
196 Chapter 3
Adjustment
1. Chapter Title
Required Adjustment after Replacing Assembly
2. Chapter Title
Test Ports Characteristics Adjustment
The purpose of this procedure is to adjust source match, directivity and tracking.
3. Adjustment
4. Chapter Title
5. Chapter Title
Chapter 3 197
Adjustment
Required Adjustment after Replacing Assembly
198 Chapter 3
4. Troubleshooting
4 Troubleshooting
199
Troubleshooting
Introduction
Introduction
WARNING These servicing instructions are for use by qualified personnel only. To avoid possible
electrical shock, do not perform any servicing unless you are qualified to do so.
WARNING The opening of covers or removal of parts is likely to expose dangerous voltages.
Disconnect the instrument from its power supply beforehand.
CAUTION Many of the assemblies in the instrument are very susceptible to damage from ESD
(electrostatic discharge). Perform the following procedures only at a static-safe workstation
and wear a grounding strap.
CAUTION DO NOT operate without following instructions. Programs or files in the instrument may
be broken.
200 Chapter 4
Troubleshooting
How to exit from the E5070B/E5071B Measurement View
Step 1. Connect the mouse and external keyboard to the connectors on the E5070B/E5071B rear
panel.
Step 4. Click Service Menu - Service Function. Password dialog box opens as shown in Figure
4-1.
Step 5. Enter the password kid in the Password box and then click "OK" button.
Step 6. Click Exit in Service Functions Menu. Then the E5070B/E5071B exit the Measurement
View, then.windows desktop screen appears with 4 icons (My Computer, Recycle Bin, My
Network Places, and Network Analyzer.)
NOTE If you wish to return to the Measurement View, double-click "Network Analyzer" icon.
NOTE If you need to shut down the E5070B/E5071B and again turn on, perform in accordance
with the following procedure.
4. Troubleshooting
a. To get "Start" menu bar displayed, move the pointer to the bottom of the screen with
mouse.
b. Click "Start" and "Shut Down..." in the pull down menu. "Shut Down Windows" dialog
box opens.
c. Select "Shut down" button in the pull down menu.
d. Click "OK" button in the dialog box.
Chapter 4 201
Troubleshooting
To Troubleshoot the Instrument
About a few minutes after the E5070B/E5071B is turned on, the measurement view is
displayed on the screen. The display on the screen should be similar to Figure 4-12,
“Measurement view,” on page 213.
• If no display appears on the LCD after the E5070B/E5071B is turned on, go to “No
Display troubleshooting” on page 204.
• The power-on self test is performed once automatically after the E5070B/E5071B
measurement view is displayed. If the power-on self test fails, go to “Troubleshooting
Using Internal Test” on page 219.
If the measurement function does not work correctly, perform the internal test and external
test provided in the E5070B/E5071B's service function. When the internal test fails, go to
“Internal Test Failure Troubleshooting” on page 222. When the external test fails, go to
“External Test Failure Troubleshooting” on page 254.
NOTE The internal test includes some unique measurement function tests in addition to the tests
that are common to the power-on self test. Thus, it is necessary to perform the internal test
even if the power-on self test passed.
202 Chapter 4
Troubleshooting
To Troubleshoot the Instrument
4. Troubleshooting
Chapter 4 203
Troubleshooting
No Display troubleshooting
No Display troubleshooting
If the E5070B/E5071B displays nothing despite it is powered from proper ac power line,
isolate the failure in accordance with the procedure shown in Figure 4-3.
Connect the keyboard to the E5070B/E5071B rear panel connector, turn the power on and
start trouble isolation. The methods of trouble isolation are described in the procedural step
1 to 5.
204 Chapter 4
Troubleshooting
No Display troubleshooting
If the rear panel fan (blower) doesn't run, a failure in A50 ATX power supply is assumed.
Remove the E5070B/E5071B outer cover and check if the following LEDs light:
NOTE To check all the outputs of the A50, measure the dc voltages at the output lead connectors
with a DMM. The dc output voltages and lead color information is provided in the module
cover label of the A50.
If the system fans on the chassis inside the E5070B/E5071B don't run, problem seems in
the A21 analog interface board or the flat cable between the A21 and the A22 front panel
keyboard. In this case, remove the E5070B/E5071B outer cover and make sure whether the
fans run or not.
If a beep and a power shutdown occur immediately after power is turned on, there is a
possibility that either of the fans won’t run. There are three system fans, which are
independently controlled by A6 RF switch/temperature controller board. These fans
initially run fast after the power is turned on, and usually stop after a while. One or some of
them restart running when the instrument warms up. The power shutdown occurs the
moment the system fan stops by any anomaly. In this case, check the A6 board and the fan
that doesn't run.
NOTE If the power shutdown occurs without a beep, the problem seems in the A21 or the A20
board. Check BIOS status of A20 is correct as described in “Configure the Motherboard”
on page 262.
Before replacing the board, check if the jumper setting on the A20 is correct as described in
“Configure the Motherboard” on page 262.
4. Troubleshooting
Press "Num Lock" key on the keyboard. If the LED in the key doesn't light as shown in
Figure 4-4, a problem seems in the A20 digital motherboard.
Chapter 4 205
Troubleshooting
No Display troubleshooting
• Whether all the connections to the A20 are normal or not. Check if there is any
disconnection or connection working loose.
• Whether the jumper setting on the A20 is correct or not as described in “Configure the
Motherboard” on page 262.
• Whether BIOS options are correct or not as described in “To Confirm or Set the BIOS
Options” on page 264.
Connect an external VGA monitor to the VIDEO output on the E5070B/E5071B rear
panel.
• If something is displayed on the external monitor, the problem is present around the
LCD. Also check the A21 and A22 because the ON/OFF setting of the LCD backlight
is controlled by the A21 through the A22 front interface board.
• If nothing is displayed even on the external monitor, the problem seems in the A20
digital motherboard or A26 LCD interface card.
NOTE Check if the A26 board is securely connected to the A20 board.
Check a flat cable between the A26 and A31 LCD connector.
Check A52 inverter board and a cable between the A52 and A26. Also check the cables
between the A51 LCD and A52. If the cables are normal, check the A51 LCD.
206 Chapter 4
Troubleshooting
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
4. Troubleshooting
Chapter 4 207
Troubleshooting
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
Step 1. Video BIOS booting
Video BIOS title screen shown in Figure 4-6 is displayed first, when the E5070B/E5071B
is turned on. If this screen doesn't appear (nothing appears), go to “No Display
troubleshooting” on page 204.
NOTE If the power shutdown occurs without a beep, the problem seems in the A21 or the A20
board. Check BIOS status of A20 is correct as described in “Configure the Motherboard”
on page 262.
Before replacing the board, check if the jumper setting on the A20 is correct as described in
“Configure the Motherboard” on page 262.
The splash screen is displayed with Agilent logo as shown in Figure 4-7.
If the splash screen is displayed, you can assume that the A20 digital motherboard is
functioning correctly.
NOTE While the splash screen is displayed. If you want to run the BIOS setup utility, push Delete
key as soon as the message of "Press TAB to show POST screen, DEL to enter SETUP, F12
to select BOOT DEVICE" in the splash screen.
Key Action
208 Chapter 4
Troubleshooting
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
Key Action
DEL Run the BIOS Setup Utility. For details of how to run the
BIOS setup utility, refer to “Run the BIOS setup utility”
on page 264.
F12 Select the boot device. This menu is used for system
recovery. For the detail of the recovery, refer to Appendix
B, “System Recovery,” on page 473.
4. Troubleshooting
Step 3. Searching for boot record
E5070B/E5071B starts booting from A27 mass storage (IDE-0) after the splash screen
appears. However, if the E5070B/E5071B cannot boot from the A27, a message of "Boot
Failure" is displayed as shown in Figure 4-8. In this case, the E5070B/E5071B may have a
problem around the A27.
Chapter 4 209
Troubleshooting
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
Figure 4-8 Message of Boot Failure
Check whether the A27 has a problem or is not using BIOS setup utility as shown in Figure
4-9. For details of how to run the BIOS setup utility, refer to “Run the BIOS setup utility”
on page 264.
210 Chapter 4
Troubleshooting
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
When the mass storage is detected, the HDD model name is displayed as shown in Figure
4-9 in the Standard CMOS Features window under the BIOS setup utility.If the mass
storage is not detected, "None" is displayed The A27 mass storage is connected to the IDE
connector on the A20 digital motherboard via A30 3.5"-2.5" adapter and a flat cable.
Check the flat cable first. If it has no problem, replace the A27 mass storage.
The Windows boot screens are displayed. The Windows boot screens consists of three
screens. Each screens is display in the order as shown in Figure 4-10. If the Windows boot
screens are displayed, it is assumed that the A27 works. While the Window screens are
displayed, Windows operating system is starting up.
4. Troubleshooting
If you encounter the following problems, try to reinstall the operating system before
replacing the A27.
• The Window boot screen is not displayed after the splash screen is displayed.
NOTE If the E5070B/E5071B was turned off without shutdown process, Microsoft Scandisk runs
while the windows boot screens are displayed. If a serious problem is found in the
scandisk, reinstall the operating system. For details of the operating system installation,
refer to Appendix B, “System Recovery,” on page 473. If the operating system still doesn't
boot up properly after reinstallation, replace the A27.
Chapter 4 211
Troubleshooting
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
NOTE The operating system automatically checks the device drivers, which are necessary to use
the E5070B/E5071B functions and are installed in the system before the E5070B/E5071B
is shipped from Agilent factory. If the operating system doesn't detect them, a message box
is displayed. In this case, install the device driver.
The firmware revision and hardware option information along with copyright declaration is
displayed as shown in Figure 4-11. The E5070B/E5071B firmware quickly starts up just
before this display appears. While the revision and option information is displayed, the
applications of various devices in the system are initialized.
If the display whited out, entirely blued or appeared with a dialog box, a mass storage
problem is suspected. Try to perform the mass storage recovery procedure.
NOTE If a message of "Will Shut Down in Five Seconds" is displayed in place of "Initializing.."
and the shutdown occurs, the A21 board fails in starting up. The following message may be
displayed before the shutdown occurs:
This message indicates that the A21 board doesn't work or is not properly connected to the
A20.
212 Chapter 4
Troubleshooting
Boot Process Troubleshooting for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
If this happened, the A21 board failed in writing DSP program into flash ROM when the
firmware was installed first or updated to the newest version. A problem in the A21 or A20
is suspected.
The measurement view as shown in Figure 4-12 is displayed after the system initialization
is completed without problem.
4. Troubleshooting
The power-on self test is executed once automatically before the measurement starts.
While the power-on self test is in progress, "Power on test" is displayed at the left in the
instrument status bar. If the power-on test fails, an error message is displayed there. For
more details, refer to “Troubleshooting Using Internal Test” on page 219.
Chapter 4 213
Troubleshooting
Boot Process Troubleshooting for E5070B : MY42300633 and above, E5071B
: MY42301397 and above
214 Chapter 4
Troubleshooting
Boot Process Troubleshooting for E5070B : MY42300633 and above, E5071B
: MY42301397 and above
The splash screen is displayed with Agilent logo as shown in Figure 4-14.
If the splash screen is displayed, you can assume that the A20 digital motherboard is
functioning correctly.
NOTE While the splash screen is displayed, if you want to run the BIOS setup utility, push F2 key
as soon as in the screen.
The Windows boot screens are displayed. The Windows boot screens consists of two
screens. Each screen is displayed in the order as shown in Figure 4-15. If the Windows
boot screens are displayed, it is assumed that the A27 HDD works. While the Window
screens are displayed, Windows operating system is starting up.
4. Troubleshooting
Chapter 4 215
Troubleshooting
Boot Process Troubleshooting for E5070B : MY42300633 and above, E5071B
: MY42301397 and above
Figure 4-15 Windows boot screens
If you encounter the following problems, try to reinstall the operating system before
replacing the A27.
• The Window boot screen is not displayed after the splash screen is displayed.
NOTE If the E5070B/E5071B was turned off without shutdown process, Microsoft Scandisk runs
while the windows boot screens are displayed. If a serious problem is found in the
scandisk, reinstall the operating system. For details of the operating system installation,
refer to Appendix B, “System Recovery,” on page 473. If the operating system still doesn't
boot up properly after reinstallation, replace the A27.
NOTE The operating system automatically checks the device drivers, which are necessary to use
the E5070B/E5071B functions and are installed in the system before the E5070B/E5071B
is shipped from Agilent factory. If the operating system doesn't detect them, a message box
is displayed. In this case, install the device driver.
The firmware revision and hardware option information along with copyright declaration is
displayed as shown in Figure 4-16. The E5070B/E5071B firmware quickly starts up just
before this display appears. While the revision and option information is displayed, the
applications of various devices in the system are initialized.
216 Chapter 4
Troubleshooting
Boot Process Troubleshooting for E5070B : MY42300633 and above, E5071B
: MY42301397 and above
Figure 4-16 Firmware revision and option information
If the display whited out, entirely blued or appeared with a dialog box, a mass storage
problem is suspected. Try to perform the mass storage recovery procedure.
NOTE If a message of "Will Shut Down in Five Seconds" is displayed in place of "Initializing.."
and the shutdown occurs, the A21 board fails in starting up. The following message may be
displayed before the shutdown occurs:
This message indicates that the A21 board doesn't work or is not properly connected to the
4. Troubleshooting
A20.
If this happened, the A21 board failed in writing DSP program into flash ROM when the
firmware was installed first or updated to the newest version. A problem in the A21 or A20
is suspected.
The measurement view as shown in Figure 4-17 is displayed after the system initialization
is completed without problem.
Chapter 4 217
Troubleshooting
Boot Process Troubleshooting for E5070B : MY42300633 and above, E5071B
: MY42301397 and above
Figure 4-17 Measurement view
The power-on self test is executed once automatically before the measurement starts.
While the power-on self test is in progress, "Power on test" is displayed at the left in the
instrument status bar. If the power-on test fails, an error message is displayed there. For
more details, refer to “Troubleshooting Using Internal Test” on page 219.
218 Chapter 4
Troubleshooting
Troubleshooting Using Internal Test
NOTE There are cases where the internal test is ineffective (insufficient) for trouble isolation
because the test objects are limited to the circuit blocks and functions that can be tested
with nothing connected to the test ports. In such cases, refer to external test to perform the
diagnosis for the range uncovered with the internal test.
NOTE The following procedure can be used to restart the power-on self test as required.
a. Press key.
b. Click Service Menu and, then, Test Menu in the softkeys.
c. Click Power On Test to restart the test. Wait until the power-on test ends.
d. The test result (OK or Failed) is displayed in the Power On Test key.
PLL unlock
When a PLL of the frequency synthesizers is unlocked, not the "Power on self test failed"
4. Troubleshooting
but "Phase lock loop unlocked" message is displayed. If it occurs, A1 Source board may be
faulty. (There is also a low possibility that A17 ADC part of A2 Receiver board is faulty).
Chapter 4 219
Troubleshooting
Troubleshooting Using Internal Test
NOTE To perform the internal test properly, the following conditions must be met:
NOTE Do not operate front panel keys, keyboard and mouse during the internal test. Changing the
instrument settings while the internal test is in progress will cause incorrect test results.
Step 3. Press - - keys (or click Load Project) to select Load Project
function."Open" dialog box will be displayed as shown in Figure 4-18.
Step 4. Select "User [D]" (preset state) from menu in the "Lock in:" box.
Step 7. Click "Internal Test. VBA" program file to select it from program menu.
Step 9. Press - - - (or click Select Macro) to select Select Macro function.
Step 10. Press (or click Module1 main) to open the Module1 main program file. "Internal
Test" dialog box will appear as shown in Figure 4-19. The dialog box shows the menu of
test groups to choose from.
220 Chapter 4
Troubleshooting
Troubleshooting Using Internal Test
Step 11. All the test groups are preset to take effect. If it is necessary to perform a specific test group
or some test groups only, check in (click) the check boxes for the desired test group(s) and
clear the other check boxes.
NOTE Clicking "Select None" button clears all the check boxes. Clicking "Detail" button displays
the names of individual tests included in each test group. Click "Exit" to return to the
Internal Test dialog box.
Step 12. Click "Start Test" button to run the internal test program. The dialog box displays "Now
testing ..." while the internal test is in progress.
Step 13. When the internal test is completed, the dialog box displays Pass or Fail for each test group
as an example shown in Figure 4-20.
4. Troubleshooting
If any of the test groups failed, click "Detail" button of that test group to look into the
breakdown of the test results. Figure 4-21 shows a breakdown test result example.
Chapter 4 221
Troubleshooting
Troubleshooting Using Internal Test
NOTE Clicking "Select None" button clears all the check boxes, but it does not clear the test
results.
Test Test First failed test Test point / objective Probable faulty board assembly
No. group
A1 A3 A4 A6 A2 Receiver A7
Src Lvl Amp Sw/T Att
Vern Mod ctrl A15 A17
222 Chapter 4
Troubleshooting
Troubleshooting Using Internal Test
Test Test First failed test Test point / objective Probable faulty board assembly
No. group
A1 A3 A4 A6 A2 Receiver A7
Src Lvl Amp Sw/T Att
Vern Mod ctrl A15 A17
4. Troubleshooting
27 SVT(2.11GHz) Source swept synthe on ### #
A1
Chapter 4 223
Troubleshooting
Troubleshooting Using Internal Test
Test Test First failed test Test point / objective Probable faulty board assembly
No. group
A1 A3 A4 A6 A2 Receiver A7
Src Lvl Amp Sw/T Att
Vern Mod ctrl A15 A17
224 Chapter 4
Troubleshooting
Troubleshooting Using Internal Test
Test Test First failed test Test point / objective Probable faulty board assembly
No. group
A1 A3 A4 A6 A2 Receiver A7
Src Lvl Amp Sw/T Att
Vern Mod ctrl A15 A17
4. Troubleshooting
59 S33(6GHz) Source power and Ref ## ## ## ## ##
channel receiver gain for
Port 3 (E5071B
Opt.313/314/413/44)
Chapter 4 225
Troubleshooting
Troubleshooting Using Internal Test
Test Test First failed test Test point / objective Probable faulty board assembly
No. group
A1 A3 A4 A6 A2 Receiver A7
Src Lvl Amp Sw/T Att
Vern Mod ctrl A15 A17
NOTE Internal tests from item number 1 to 50 are common to the power on self test.
NOTE When the internal test for a specific receiver port fails, identify the faulty A2 receiver
board as shown in Figure 4-22.
226 Chapter 4
Troubleshooting
Function Specific Troubleshooting
Front panel keys All the E5070B/E5071B functions except for VBA Refer to “To Check the Front
and service functions can be set and controlled via Panel” on page 230.
the front panel keys.
Touch panel The E5070B/E5071B equipped with option 016 Refer to “To Check the
has a touch screen display that allows all the Touch Panel (Option 016
functions in the menu bars, setup windows and only)” on page 231.
dialog boxes to be set by a touch to the screen
panel.
LCD display Almost all the information including the Refer to “To Check the
measurement value, setup state, result data LCD” on page 231.
processing, menu bar, softkey label and others are
indicated on the 10.4-inch color LCD display.
External keyboard The external keyboard can be used for the entry of Refer to “To Check the
numerical and character data when it is connected External Keyboard” on
to the keyboard interface connector (PS-2) on the page 232.
rear panel.
Mouse The mouse can be used to move the pointer on the Refer to “To Check the
LCD display, select a function and change a Mouse” on page 232.
setting, when it is connected to the mouse interface
4. Troubleshooting
connector (PS-2) on the rear panel.
Floppy disk drive The 3.5-inch floppy disk drive on the front panel is Refer to “To Check the
used to save the E5070B/E5071B setup state and FDD” on page 232.
measurement data in a 3.5-inch, 1.4 MB floppy
disk in MS-DOS compatible format.
Video output An external color monitor can be used to display Refer to “To Check the
the same information as the E5070B/E5071B LCD Video output” on page 233.
display, when it is connected to the Video output
connector (24-pin D-Sub) on the rear panel.
External trigger input The external trigger input terminal (BNC) on the Refer to “To Check the
rear panel allows an external trigger source to be External Trigger Input” on
used for measurement trigger. page 233.
Chapter 4 227
Troubleshooting
Function Specific Troubleshooting
LAN port The LAN interface port on the rear panel allows Refer to “To Check the
the E5070B/E5071B to be connected to 10/100 LAN” on page 233.
Base-T Ethernet.
Printer parallel port A specified printer can be used to print the Refer to “To Check the
E5070B/E5071B measurement display, setup Printer Parallel Port” on
display and others, when it is connected to the page 237.
Printer port (36-pin D-Sub) on the rear panel.
GPIB interface The GPIB compatibility allows the Refer to “To Check the
E5070B/E5071B to be operated as a talker/listener GPIB” on page 238.
or system controller on IEEE 488 interface bus.
Handler I/O port The Handler I/O port can be used to transfer a Perform “[7] Handler I/O
comparator decision output data to and perform Board tests” in
timing synchronization with an external handler. “Troubleshooting Using
External Test” on page 239 .
Step 1. Exit from the E5070B/E5071B measurement view in accordance with the procedure
described in “How to exit from the E5070B/E5071B Measurement View” on page 201.
Then, Windows desktop screen is displayed.
Step 2. Click "My Computer" with the right button and select "Properties" as shown in Figure
4-23. Then, the System Properties(Figure 4-24) will appear.
228 Chapter 4
Troubleshooting
Function Specific Troubleshooting
Step 3. Click Hardware tab and Device Manager button. The operating system detects all the
necessary device drivers and displays the device names as shown in Figure 4-25.
4. Troubleshooting
Click the icon with the right button and click Property to show the detail of the status. as
shown in Figure 4-26.
Chapter 4 229
Troubleshooting
Function Specific Troubleshooting
Procedure
Randomly press the front panel keys and rotate the knob to verify that they work normally.
Step 2. Click Service Menu and, then, Test Menu in the softkeys.
Step 3. Click Front Panel in the test menu. This opens "Front Panel Test" dialog box as shown in
Figure 4-27.
Step 4. Randomly press the front panel keys. The key code along with the name of the pressed key
are displayed in the dialog box as shown in Figure 4-28. Turn the rotary knob clockwise or
counterclockwise. The dialog box indicates the direction of the turned knob and a count of
RPG output.
230 Chapter 4
Troubleshooting
Function Specific Troubleshooting
Step 5. To exit the front panel test, press key three times.
• If multiple keys fail to work, a problem in A22 front interface board or A21 analog
interface board is suspected. Also check the flat cable between the A21 and A22.
• If only a specific key doesn't work, check first if the key is subsided in the panel.
• If the rotary knob doesn't work, check the A22 board involving the RPG.
Procedure
By touching the LCD display panel, select or change the setting of a function in the softkey
menu and, then, perform the same operation with hardkeys.
• If the touch panel doesn't work correctly whereas the hardkeys function normally, a
failure seems in the touch screen controller assembly (5183-4184) or touch-panel LCD
assembly (E5070-60102). (The touch panel is not replaceable independently of the
LCD.)
• Check the cable between the touch screen controller and the serial interface connector
on the A20 digital motherboard.
• If no problem is found in the above checks, a failure in the A20 digital motherboard is
suspected.
Procedure
4. Troubleshooting
Step 1. Press key.
Step 2. Click Service Menu and, then, Test Menu in the softkeys menu.
Step 3. Click Display in the test menu. The whole of the LCD screen turns Red, Green, Blue,
White and Black every 2 seconds and returns to the measurement view. If the color test
screen doesn't appear correctly, perform step 4.
Step 4. Connect an external VGA monitor to the VIDEO output port on the E5070B/E5071B rear
panel.
• If the monitor screen view is the same as the LCD display, the problem seems in the
A26 LCD interface card. Check first if the A26 board is securely connected to the A20
board.
• If only the LCD display has a problem, check the flat cable between the A26 and A31
LCD connector.
Chapter 4 231
Troubleshooting
Function Specific Troubleshooting
• If the LCD is not illuminated with backlight, check A52 inverter board and the cable
between the A52 and A26. Also check the cables between the A51 LCD and A52.
• If the cables are normal, check the A51 LCD.
Procedure
Step 1. Connect the external keyboard to the E5070B/E5071B rear panel connector.
Step 4. Press and keys on the external keyboard, and verify that the cursor on the menu
bar moves up and down. If it doesn't work, the external keyboard or the A20 digital
motherboard may be faulty.
Procedure
Step 3. Move the mouse and verify that the mouse pointer moves smoothly. If it doesn't move
smoothly, check first whether a foreign substance (dust, lint, etc.) is in the track ball hole of
the mouse or not.
Step 4. Verify that the mouse buttons work normally. If any button doesn't work or the mouse
pointer doesn't move, a failure in the mouse or the A20 digital motherboard is suspected.
Procedure
Step 1. Connect the external keyboard to the E5070B/E5071B rear panel connector.
Step 3. Insert a 1.44 MB floppy disk, formatted in DOS format, into the FDD slot.
Step 4. Press key, and click Save State in the menu bar.
Step 5. Click File Dialog... to open the "Save As" dialog box.
Step 6. Select 3 1/2 Floppy [A:] from "Save in" pull-down menu.
Step 7. Enter e5070b in the file name box from the keyboard.
Step 9. Press key, and click Recall State in the menu bar.
232 Chapter 4
Troubleshooting
Function Specific Troubleshooting
Step 10. Click File Dialog... to open the "Open" dialog box.
Step 11. Select 3 1/2 Floppy [A:] from "Look in" pull-down menu.
Step 12. Select e5070b file from the file menu or enter e5070b in the file name box.
If the file save or recall operation fails, a failure in the A28 FDD or the flat cable between
the A28 and A20 digital motherboard is suspected.
Procedure
Step 1. Connect an external VGA color monitor to the Video output port on the E5070B/E5071B
rear panel.
Step 3. Verify that the monitor screen view is the same as the display on the LCD. If the monitor
screen view is abnormal, a failure seems in the A26 LCD interface card.
Procedure
Step 3. Click Trigger Source and, then, External in the menu bar to set the trigger mode to
"External".
Step 4. Connect a BNC Short or 50 Ω termination to the Ext Trig connector on the rear panel and
disconnect it. Thereby a measurement trigger should be generated and a measurement
result (trace) should be refreshed.
4. Troubleshooting
Step 5. If no trigger occurs, a failure in the A21 analog interface board is suspected.
Procedure
Step 1. Connect a LAN cable between the LAN port on the rear panel and an external computer
(PC).
NOTE Use a crossed LAN cable to enable the peer-to-peer communication between the
E5070B/E5071B and the PC. If the E5070B/E5071B needs to be connected to the PC via a
multi port Hub, use a straight LAN cable.
Step 3. Click - Misc Setup - Network Setup - Network Configuration in the softkey menu to
Chapter 4 233
Troubleshooting
Function Specific Troubleshooting
show the Network and Dial-up Connections window as shown in the left of Figure 4-29.
Step 4. Check the Local Area Connection icon to confirm whether the network connection is
enabled or disabled. The meaning of the icon is shown in the right of Figure 4-29. If the
connection is disabled, Double-click the Local Area Connection icon.
a. Click the Local Area Connection icon with the right button and click “Property”. This
opens the "Local Area Connection Property" dialog box as shown in Figure 4-30.
234 Chapter 4
Troubleshooting
Function Specific Troubleshooting
d. If the IP Address and the Subnet Mask are already assigned, go to Step 6. If they are
not assigned yet, enter them as follows:
IP Address 192.168.0.1
Subnet Mask 255.255.255.0
Default Gateway blank
To enter a specific IP Address and Subnet Mask, click "Use the following IP address"
option button to select it. Input your IP address in the "IP Address" box and your subnet
mask in the Subnet Mask box (overwrite the initial values).
e. Double-click “My Computer” icon on Windows desktop of the external PC. And
double-click Control Panel. The Control Panel window will appear.
4. Troubleshooting
f. In the Control Panel window, double-click Network. The Network dialog box shown in
fig will appear. In the Configuration tab, select TCP/IP (display it in revers video) and
press the Properties button.
g. The dialog box appears. If Specify an IP address has been selected, record the IP
address and subnet mask.
h. If the Obtain an IP address automatically has been selected, select the Speccify an IP
address. Enter the IP address xxx.xxx.xxx.y and subnet mask xxx.xxx.xxx.xxx as
shown in Figure 4-32. The letters x represent the IP address and subnet mask of the
E5070B/E5071B. The letter y is different from the IP address of the E5070B/E5071B.
Press the OK button.
Chapter 4 235
Troubleshooting
Function Specific Troubleshooting
Step 6. By clicking “OK” button, the network settings take effect. Verify the E5070B/E5071B
LAN settings and response to a command from the external computer as follows:
a. On the external computer, click "Start" button in the Windows Start menu bar using a
mouse.
b. Click "Programs" and "MS-DOS Prompt" (or "Command Prompt") to open the DOS
(Command) Prompt window.
c. Type a ping command followed by the E5070B/E5071B IP address as "ping
xxx.xxx.xxx.xxx" (where, xxx.xxx.xxx.xxx is the IP address such as 192.168.0.1
assigned in step 3-d.) Press key on the keyboard.
d. The external computer displays a message of ping command response from the
E5070B/E5071B. Figure 4-33 shows an example of normal result that indicates the IP
address, packet size, and a response time counted by the computer. The ping command
is repeated 4 times.
If the E5070B/E5071B doesn't return the response, the external computer displays a
236 Chapter 4
Troubleshooting
Function Specific Troubleshooting
message of "Request timed out" as shown in Figure 4-34. In this case, a failure in the
A20 digital motherboard is suspected.
Test equipment
Procedure
Step 1. Connect the printer cable between the printer and the E5070B/E5071B printer parallel port.
Turn the printer on.
4. Troubleshooting
Step 3. Press key.
Step 5. Depending on the printer used, Click proper icon with the right button in the Property
window. Then, click "Set as Default Printer" button.
Step 6. Click Print in the softkey menu. Confirm the printout of the display image.
Step 7. If the printer fails to work, the problem seems in the A20 digital motherboard, printer
driver or connection of the printer cable.
Chapter 4 237
Troubleshooting
Function Specific Troubleshooting
Procedure
Perform the E5070B/E5071B performance test program. If the controller cannot detect the
E5070B/E5071B, the problem seems in the A24 GPIB card or the connection of the GPIB
cable.
Procedure
Connect the USB cable between the controller PC and the E5070B/E5071B. Turn the
controller PC on. If the E5070B/E5071B cannot detect the controller PC, the problem
seems in the USB (USBTMC) Interface Card or the A20 digital motherboard is suspected.
238 Chapter 4
Troubleshooting
Troubleshooting Using External Test
2 IF Ranging
3 Receiver Linearity
4 Dynamic Range
5 Trace Noise
6 Error Term
4. Troubleshooting
Required test equipment Qty Recommended model
NOTE Required quantity of Short termination is 3 in addition to one included in the calibration
kit.
Chapter 4 239
Troubleshooting
Troubleshooting Using External Test
NOTE To perform the external test properly, the following conditions must be met:
NOTE Do not operate front panel keys, keyboard and mouse during the external test. Changing
the instrument settings while the external test is in progress will cause incorrect test results.
Step 3. Press - - keys (or click Load Project) to select Load Project function.
"Open" dialog box will be displayed as shown in Figure 4-35.
Step 4. Select "User [D]" (preset state) from menu in the "Lock in:" box.
Step 7. Click "External Test. VBA" program file to select it from program menu.
Step 9. Press - - - (or click Select Macro) to select Select Macro function.
Step 10. Press (or click Module1 main) to open the Module1 main program file. "External
Test" window will be displayed as shown in Figure 4-36. This window shows the test items
and their results.
240 Chapter 4
Troubleshooting
Troubleshooting Using External Test
Step 11. To perform the individual tests in each test group, refer to the procedure described below.
Step 12. To end the external test, click "Exit" button in the dialog box.
Required equipment:
Procedure:
Perform the following procedure after Step 11.
a. Click "Test" button in the right of “Tch Power Gain” in the window to open the "Tch
Power Gain" window. The window displays the S parameters and frequencies as the
test points. Their results are displayed in the matrix.(Figure 4-37)
4. Troubleshooting
Figure 4-37 Tch Power Gain tests dialog box example (E5071B Opt. 413/414)
b. Select the S parameter for the test by clicking the check boxes. Click the "Start Test"
button.The instruction dialog box opens as shown in Figure 4-38. Follow the
instruction.
Chapter 4 241
Troubleshooting
Troubleshooting Using External Test
NOTE Clicking “Select All” button checks all the check box. Clicking "Select None" button
clears all the check boxes. Click "Exit" to return to the External Test dialog box
c. Connect an type-N (m) coaxial cable to the test ports in accordance with Table 4-5.
Table 4-5 Tch Power Gain tests setups
Test name Objective of the test Connect an N Freq and IFBW settings
cable between (Automatic)
S12 Source power for port 2 Port 1 and Port2 50 MHz/3 GHz/6GHz, 10
and test channel kHz BW
receiver gain for port 1
S21 Source power for port 1 Port 1 and Port 2 50 MHz/3 GHz/6GHz, 10
and test channel kHz BW
receiver gain for port 2
S31 Source power for port 1 Port 1 and Port 3 50 MHz/3 GHz/6GHz, 10
and test channel kHz BW
receiver gain for port 3
(Opt. 313/314/413/414)
S41 Source power for port 1 Port 1 and Port 4 50 MHz/3 GHz/6GHz, 10
and test channel kHz BW
receiver gain for port 4
(Opt. 413/414)
NOTE RF output power is 5 dBm and Nop is 101 for all the tests shown in the table.
d. Click "OK" button to run the test program. The dialog box displays "Now testing..."
while the Tch Power Gain tests are in progress.
e. When the tests are completed, the dialog box displays Pass or Fail for each test as
shown in Figure 4-39.
242 Chapter 4
Troubleshooting
Troubleshooting Using External Test
Figure 4-39 Power Gain test results display example (E5071B Opt. 413/414)
f. If at least one of the Tch Power Gain tests failed, refer to Table 4-11 External Test
Failure Troubleshooting Information.
g. To return to the External Test dialog box, click "Exit" button.
Required equipment:
Description Recommended model
Figure 4-40 IF Ranging tests dialog box example (E5070B/E5071B Opt. 413/414)
4. Troubleshooting
b. Select the port numbers for the test by clicking the check boxes. Then, click “Start
Test“ button. The instruction dialog box opens as shown in Figure 4-41. Follow the
instruction.
NOTE Clicking “Select All” button checks all the check boxes. Clicking "Select None" button
clears all the check boxes. Click "Exit" to return to the External Test dialog box
Chapter 4 243
Troubleshooting
Troubleshooting Using External Test
c. Connect a 10 dB fixed attenuator and an type-N (m) coaxial cable in series between the
test ports in accordance with Table 4-6. Make the connection as shown in Figure 4-42.
Port 1 IF signal level ranging operation Port 1 and Port 2 50 MHz/0 Hz, 1 kHz BW, 101,
of test channel receiver for Port 1 -5 dBm
Port 2 IF signal level ranging operation Port 1 and Port 2 50 MHz/0 Hz, 1 kHz BW, 101,
of test channel receiver for Port 2 -5 dBm
Port 3 IF signal level ranging operation Port 1 and Port 3 50 MHz/0 Hz, 1 kHz BW, 101,
of test channel receiver for Port 3 -5 dBm
(Opt. 313/314/413/414)
Port 4 IF signal level ranging operation Port 1 and Port 4 50 MHz/0 Hz, 1 kHz BW, 101,
of test channel receiver for Port 4 -5 dBm
(Opt. 413/414)
d. Click "OK" button to run the test program. The dialog box displays "Now testing..."
while the Tch IF Ranging tests are in progress.
244 Chapter 4
Troubleshooting
Troubleshooting Using External Test
e. When the tests are completed, the dialog box displays Pass or Fail for each test as
shown in Figure 4-43.
Figure 4-43 Tch IF Ranging test results display example (E5070B/E5071B Opt. 413/414)
f. If at least one of the Tch IF Ranging tests failed, refer to Table 4-11 External Test
Failure Troubleshooting Information.
g. To return to the External Test dialog box, click "Exit" button.
Required equipment:
Procedure:
4. Troubleshooting
Perform the following procedure after Step 11.
a. Click "Test" button in the right of “Receiver Linearity” in the window to open the
"Receiver Linearity" window. The window displays the port number and receiver
channel as the test points. The result for each S parameter is displayed as the 4×2
matrix.(Figure 4-44).
Chapter 4 245
Troubleshooting
Troubleshooting Using External Test
Figure 4-44 Receiver Linearity tests dialog box example (E5070B/E5071B Opt. 413/414)
b. Select the S parameters for the test by clicking the check boxes. Then, click “Start Test“
button. The instruction dialog box opens as shown in Figure 4-45. Follow the
instruction.
NOTE Clicking “Select All” button checks all the check boxes. Clicking "Select None" button
clears all the check boxes. Click "Exit" to return to the External Test dialog box
c. Connect an type-N (m) coaxial cable between the test ports in accordance with Table
4-7.
Table 4-7 Receiver Linearity test setups
Test name Objective of the test Connect an N Sweep freq span, IFBW,
cable between Nop and RF power
settings (Automatic)
S12 Receiver linearity for Port 1 Port 1 and Port 2 Full span, 10 kHz BW, 201,
max. source power and -10
dBm below max..
S21 Receiver linearity for Port 2 Port 1 and Port 2 Full span, 10 kHz BW, 201,
max. source power and -10
dBm below max..
S31 Receiver linearity for Port 3 Port 1 and Port 3 Full span, 10 kHz BW, 201,
(Opt. 313/314/413/414) max. source power and -10
dBm below max..
S41 Receiver linearity for Port 4 Port 1 and Port 4 Full span, 10 kHz BW, 201,
(Opt. 413/414) max. source power and -10
dBm below max..
246 Chapter 4
Troubleshooting
Troubleshooting Using External Test
d. Click "OK" button to run the test program. The dialog box displays "Now testing..."
while the Receiver Linearity tests are in progress.
e. When the tests are completed, the dialog box displays Pass or Fail for each test as
shown in Figure 4-46.
Figure 4-46 Receiver Linearity test results example (E5070B/E5071B Opt. 413/414)
f. If at least one of the Receiver Linearity tests failed, refer to Table 4-11 External Test
Failure Troubleshooting Information.
g. To return to the External Test dialog box, click "Exit" button.
Required equipment:
4. Troubleshooting
Type-N adapter
Procedure:
Perform the following procedure after Step 11.
a. Click "Test" button in the right of “Dynamic Range” in the window to open the
"Dynamic Range" window. The window displays the S parameters and frequency as
the test points. The results are displayed as matrix.(Figure 4-47).
Chapter 4 247
Troubleshooting
Troubleshooting Using External Test
Figure 4-47 Dynamic Range tests dialog box example (E5071B Opt. 413/414)
b. Select the S parameters for the test by clicking the check boxes. Then, click “Start Test“
button. The instruction dialog box opens as shown in Figure 4-48. Follow the
instruction.
NOTE Clicking “Select All” button checks all the check boxes. Clicking "Select None" button
clears all the check boxes. Click "Exit" to return to the External Test dialog box
NOTE For more information on the Dynamic Range tests setup, see Table 4-8.
Test parameter Connect Short Sweep freq span, IFBW, Nop and
terminations to RF power settings (Automatic)
248 Chapter 4
Troubleshooting
Troubleshooting Using External Test
d. Click "OK" button to run the test program. The dialog box displays "Now testing..."
while the Dynamic Range tests are in progress.
e. When the tests are completed, the dialog box displays Pass or Fail for each test as
shown in Figure 4-49.
Figure 4-49 Dynamic Range test results example (E5071B Opt. 413/414)
f. If at least one of the Dynamic Range tests failed, refer to Table 4-11 External Test
Failure Troubleshooting Information.
g. To return to the External Test dialog box, click "Exit" button.
Required equipment:
4. Troubleshooting
or part of
85033E/85050D with
Type-N adapter
Procedure:
Perform the following procedure after Step 11.
a. Click "Test" button in the right of “Trace Noise” in the window to open the "Trace
Noise" window. The window displays the S parameters and frequencies as the test
points. The result for each S parameter is displayed as the 9×2 matrix.(Figure 4-50).
Chapter 4 249
Troubleshooting
Troubleshooting Using External Test
b. Select the S parameter for the test by clicking the check boxes. Then, click “Start Test“
button. The instruction dialog box opens as shown in Figure 4-51. Follow the
instruction.
NOTE Clicking “Select All” button checks all the check boxes. Clicking "Select None" button
clears all the check boxes. Click "Exit" to return to the External Test dialog box
NOTE For more information on the Trace Noise tests setup, see Table 4-9.
S11 All test ports 3 MHz, 1.3 GHz, 1.31 GHz, 2.1 GHz and 2.11
GHz (E5070B),
S22
3 MHz, 1.3 GHz, 1.31 GHz, 2.1 GHz, 2.11
S33 GHz, 4.25 GHz, 4.26 GHz, 7.5 GHz and 8.5
GHz (E5071B),
S44
3 kHz BW, 101, max. source power, 100 ms
d. Click "OK" button to run the test program. The dialog box displays "Now testing..."
while the Trace Noise (port 1) tests are in progress.
e. When the tests are completed, the dialog box displays Pass or Fail for each test as
shown in Figure 4-52.
250 Chapter 4
Troubleshooting
Troubleshooting Using External Test
f. If at least one of the Trace Noise tests failed, refer to Table 4-11 External Test Failure
Troubleshooting Information.
g. To return to the External Test dialog box, click "Exit" button.
Required equipment:
Procedure:
Perform the following procedure after Step 11.
a. Click "Test" button in the right of “Error Term” in the window to open the "Error Term"
window. The window displays the port number and frequencies as the test points. The
4. Troubleshooting
result for each test port is displayed as the 4×3 matrix.(Figure 4-53).
Chapter 4 251
Troubleshooting
Troubleshooting Using External Test
Figure 4-53 Error Term (port 1) tests dialog box example (E5071B)
b. Select the test ports for the test by clicking the check boxes. Then, click “Start Test“
button. The instruction dialog box opens as shown in Figure 4-54. Follow the
instruction.
NOTE Clicking “Select All” button checks all the check boxes. Clicking "Select None" button
clears all the check boxes. Click "Exit" to return to the External Test dialog box
NOTE For more information on Error Term tests setup, see Table 4-10.
Source match Open Full span, 1 kHz BW, 201, 5 dBm , System
correction ON
Directivity Short
d. Click "OK" button to continue the test. The "Error Term" dialog box displays "Now
calibrating..." for a few seconds. Wait until "SHORT" dialog box opens.
e. Connect an type-N Short termination to the specified test port.
f. Click "OK" button to continue the test and wait until "LOAD" dialog box opens.
252 Chapter 4
Troubleshooting
Troubleshooting Using External Test
k. If at least one of the Error Term tests failed, refer to Table 4-11 External Test Failure
Troubleshooting Information.
l. To return to the External Test dialog box, click "Exit" button.
Required equipment:
4. Troubleshooting
Description Recommended model
Procedure:
Perform the following procedure after Step 11.
a. Click "Handler I/O Board" in the pull down menu of the Test List box.
b. Click "Test" button in the dialog box. Then "Handler I/O Board" tests dialog box opens
as shown in Figure 4-56.
Chapter 4 253
Troubleshooting
Troubleshooting Using External Test
c. Click the "Test" button in the dialog box. A message of "Connect Handler I/O Test Kit"
will be displayed.
d. Connect the handler I/O test kit to the Handler I/O connector on the rear panel.
e. Click "OK" button. The dialog box displays "Now testing..." while the Handler I/O
Board test is in progress.
f. When the tests are completed, the dialog box displays Pass or Fail of test result as
shown in Figure 4-57.
g. If the Handler I/O Board test failed, refer to Table 4-11 External Test Failure
Troubleshooting Information.
h. To return to the External Test dialog box, click "Exit" button.
NOTE To end the external test, click "Exit" button in the "External Test" dialog box.
254 Chapter 4
Troubleshooting
Troubleshooting Using External Test
Test Test group First failed Test objective Probable faulty board assembly
No. test
A1 A3 A4 A6 A2 Receiver A23
Src Lev Pow Sw/T Hnd
Vern Amp cont A15 A17
4. Troubleshooting
Chapter 4 255
Troubleshooting
Troubleshooting Using External Test
Test Test group First failed Test objective Probable faulty board assembly
No. test
A1 A3 A4 A6 A2 Receiver A23
Src Lev Pow Sw/T Hnd
Vern Amp cont A15 A17
256 Chapter 4
Troubleshooting
Troubleshooting Using External Test
Test Test group First failed Test objective Probable faulty board assembly
No. test
A1 A3 A4 A6 A2 Receiver A23
Src Lev Pow Sw/T Hnd
Vern Amp cont A15 A17
4. Troubleshooting
2
Chapter 4 257
Troubleshooting
Troubleshooting Using External Test
Test Test group First failed Test objective Probable faulty board assembly
No. test
A1 A3 A4 A6 A2 Receiver A23
Src Lev Pow Sw/T Hnd
Vern Amp cont A15 A17
258 Chapter 4
Troubleshooting
Troubleshooting Using External Test
NOTE When the external test for a specific receiver port fails, identify the faulty A2 receiver
board as shown in Figure 4-58.
4. Troubleshooting
Chapter 4 259
Troubleshooting
Performance test failure troubleshooting
NOTE There are no adjustment related to the crosstalk, system dynamic range and uncorrected
system performance.
5 Crosstalk test
260 Chapter 4
Troubleshooting
Performance test failure troubleshooting
NOTE When Crosstalk, System dynamic range or Uncorrected system performance test fails,
check first whether the connections of the RF semi-rigid cables between the suspicious
assembly and others are tight or loose. Also check for possible disconnection (impairment)
of the cables and connectors.
4. Troubleshooting
#: Possible faulty assembly
Chapter 4 261
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
NOTE When the power shutdown occurs without a beep, clear A20 CMOS and set BIOS
according to the following procedure.
4. After powering on the system, press to enter the BIOS setup utility.
262 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-59 A20 digital motherboard jumpers location and settings
4. Troubleshooting
When the A20 digital motherboard is replaced with a new one, it is advisable to perform
the DRAM test before proceeding to the BIOS setup utility procedure. Confirm the DRAM
count in accordance with the following procedure.
Step 1. Connect the external keyboard and mouse to the E5070B/E5071B rear panel connectors.
Step 2. Turn the instrument on. Wait until the E5070B/E5071B boots up. Do not press any key
until the E5070B/E5071B measurement view appears.
Step 6. Scroll the function viewer to find "System" icon and double-click "System". This opens
"System Properties" window as shown in Figure 4-60.
Chapter 4 263
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-60 System Properties window
Step 7. Verify that "523,828 KB RAM" is displayed in the Computer profile as shown in Figure
4-60. If it is not correct, the A20 board is faulty.
Step 9. Turn the E5070B/E5071B off before proceeding to the next step “To Confirm or Set the
BIOS Options”.
Procedure Step 1. Connect the external keyboard to the E5070B/E5071B rear panel connector.
Step 3. Press key as soon as the message "Press TAB to show POST screen, DEL to enter
SETUP, F12 to select BOOT DEVICE" is displayed. Wait a few seconds until the message
"Enter CURRENT Password:" appears.
Step 5. Enter password "e507xa" and press . Then BIOS setup utility main menu is
264 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
displayed as shown in Figure 4-61.
Step 6. The BIOS setup utility has 14 selectable windows. Brief descriptions and reference pages
for these windows are provided in Table 4-14. The details of the BIOS option settings in
each window are described on the page listed in the table.
NOTE If you want to quit the BIOS option setting and exit the BIOS setup utility without saving
changes in the setup, select "Exit Without Saving" and press .
4. Troubleshooting
Table 4-14 BIOS setup utility windows and references
Chapter 4 265
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Table 4-14 BIOS setup utility windows and references
Save & Exit Setup The procedure to save the option settings and on
exit from BIOS setup utility. page 277
Exit Without Saving The procedure to exit from BIOS setup utility on
without saving the option settings. page 278
NOTE Perform the following procedure only when the A20 digital motherboard is replaced. Jump
to “Standard CMOS Features” when you check whether the BIOS options are correctly set
or not.
Load Setting 2dialog box is displayed by choosing "Load Setting 2 Defaults" using
keys and by pressing key at the BIOS Setup Utility main menu. The
dialog box prompts you to select Y (Yes) or N (No) about "Load Setting 2 Defaults (Y/N)
?" as shown in Figure 4-62. Press and to load the setting 2 stored in the BIOS
ROM. As a result, almost of all the BIOS options are automatically configured to the
optimal settings for the E5070B/E5071B.
266 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-62 Load Setting 2 dialog box
If you want to exit from this page and return to main menu, press key.
4. Troubleshooting
Chapter 4 267
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-63 Standard CMOS Features window
If you want to exit from this page and return to main menu, press key.
268 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-64 Advanced BIOS Features window
Table 4-15
4. Troubleshooting
Security Option Setup
Chapter 4 269
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Table 4-15
Console Redirection
If you want to exit from this page and return to main menu, press key.
270 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-65 Advanced Chipset Features window
Table 4-16
4. Troubleshooting
*1.The setting is different from that of “Load Setting 2 Defaults” on
page 266
Integrated Peripherals
Integrated Peripherals options are displayed by choosing "Integrated Peripherals " using
keys and by pressing key at the BIOS Setup Utility main menu.
Then the window as shown in Figure 4-66 appears. Compare the actual BIOS option
settings with the default settings shown in Table 4-17. If there is any difference, select it
using and keys, and change the option setting by pressing or keys
so that it becomes the same as the default setting.
If you want to exit from this page and return to main menu, press key.
Chapter 4 271
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-66 Integrated Peripheral window
Table 4-17
272 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Table 4-17
If you want to exit from this page and return to main menu, press key.
CAUTION If the Power Management Setup options are set incorrectly, the E5070B/E5071B may not
be turned on. In this case, perform the following procedure.
4. Troubleshooting
3. Turn the instrument off and turn it on again.
4. Restore jumper setting in Figure 4-59 on page 263.
5. Turn the instrument off and turn it on again.
6. Configure the BIOS from the beginning.
Chapter 4 273
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-67 Power Management Setup window
Table 4-18
274 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Table 4-18
PnP/PCI Configurations
PnP/PCI Configurations options are displayed by choosing "PnP/PCI Configurations"
using keys and by pressing key at the BIOS Setup Utility main
menu. Then the window as shown in Figure 4-68 appears. Compare the actual BIOS option
settings with the default settings shown in Table 4-19. If there is any difference, select it
using and keys, and change the option setting by pressing or keys
so that it becomes the same as the default setting.
If you want to exit from this page and return to main menu, press key.
4. Troubleshooting
Table 4-19
Chapter 4 275
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
PC Health Status
PC Health Status options are displayed by choosing "PC Health Status" using
keys and by pressing key at the BIOS Setup Utility main menu. Then the
window as shown in Figure 4-69 appears. Compare the actual BIOS option settings with
the default settings shown in Table 4-20. If there is any difference, change the option
setting by pressing or keys so that it becomes the same as the default
setting.
If you want to exit from this page and return to main menu, press key.
Table 4-20
Frequency/Voltage Control
Frequency/Voltage Control options are displayed by choosing "Frequency/Voltage
Control" using keys and by pressing key at the BIOS Setup Utility
main menu. Then the window as shown in Figure 4-70 appears. Compare the actual BIOS
option settings with the default settings shown in Table 4-21. If there is any difference,
select it using and keys, and change the option setting by pressing or
keys so that it becomes the same as the default setting.
If you want to exit from this page and return to main menu, press key.
276 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Figure 4-70 Frequency/Voltage Control window
Table 4-21
4. Troubleshooting
Utility main menu and press key.
b. Enter the password of "e507xa". Never set another password.
c. Press key.
d. If a message of "New supervisor password installed. Press any key to continue" is
displayed, press key.
Chapter 4 277
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396
and below
Exit Without Saving
If you cancel the changes in the BIOS option settings, select "Exit Without Saving" using
keys at the BIOS Setup Utility main menu and press key. Then, a
message of "Quit Without Saving (Y/N)?" is displayed. Press and keys.
The BIOS setup utility is closed without saving the current settings.
278 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
4. Troubleshooting
Chapter 4 279
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
the DRAM test before proceeding to the BIOS setup utility procedure. Confirm the DRAM
count in accordance with the following procedure.
Step 1. Connect the external keyboard and mouse to the E5070B/E5071B rear panel connectors.
Step 2. Turn the instrument on. Wait until the E5070B/E5071B boots up. Do not press any key
until the E5070B/E5071B measurement view appears.
Step 6. Scroll the function viewer to find "System" icon and double-click "System". This opens
"System Properties" window as shown in Figure 4-72.
Step 7. Verify that "522,292 KB RAM" is displayed in the Computer profile as shown in Figure
4-72. If it is not correct, the A20 board is faulty.
Step 9. Turn the E5070B/E5071B off before proceeding to the next step “To Confirm or Set the
BIOS Options”.
280 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
Run the BIOS setup utility
Use the external keyboard to select and set various options in the BIOS setup utility. Do not
operate the E5070B/E5071B front panel until you exit the BIOS option setup procedure.
Procedure Step 1. Connect the external keyboard to the E5070B/E5071B rear panel connector.
Step 3. Press [F2] key as soon as the splash screen as shown in Figure 4-14 is displayed. Wait a
few seconds until the message "Enter CURRENT Password:" appears.
Step 5. Enter password "e507xa" and press . Then BIOS setup utility main menu is
displayed as shown in Figure 4-73.
4. Troubleshooting
Step 6. The BIOS setup utility has primary menus, most of which have sub-menus. Brief
descriptions and reference pages for these menus are provided in Table 4-22. The details of
the BIOS option settings in each menu and sub-menus are described on the page listed in
the table.
Step 7. Select a desired menu using keys. Select an item in the menu or sub-menu window
using keys. To perform an operation, use key.
NOTE If you want to exit the BIOS option setting and exit the BIOS setup utility without saving
changes in the setup, select "Exit Discarding Changes" and press key at the Exit
Menu.
Chapter 4 281
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
NOTE Perform the following procedure only when the A20 digital motherboard is replaced. Jump
to “Advanced Menu” when you check whether the BIOS options are correctly set or not.
Exit menu shown below is displayed by choosing "Exit" using keys and by
pressing key at the BIOS Setup Utility main menu. Select “Load Setup Default”
using keys. The dialog box “Setup Comfirmation” prompts you to select Y (Yes)
or N (No). Press and to load the manufacture Setup Defaults into the BIOS
ROM. As a result, almost all of the BIOS options are automatically configured to the
optimal settings for the E5070B/E5071B.
282 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
Figure 4-74 Exit menu (for Load Setup Defaults)
Advanced Menu
If you want to exit from this page and return to main menu, press key.
4. Troubleshooting
Chapter 4 283
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
Select “Advanced Chipset Control” using keys. The current chipset control
states are displayed as shown in Figure 4-76.
Select “Video boot type” using keys, and change the setting to “Disable
Onboard Video” by pressing keys, and then, return to “Advanced Menu” as
shown in Figure 4-75.
NOTE When PCI Video Card has not been installed, do not set Advanced Chipset Control.
• Advanced OS Options
Select “Advanced OS Options” using keys. The current setup states are
displayed as shown in Figure 4-77.
Select “Installed OS” using keys, and change the setting to “Plug and Play” by
pressing keys. Next step as, select “Enable ACPI” using keys, and
change the setting to “No” by pressing keys. And then, return to “Advanced
Menu” as shown in Figure 4-75.
284 Chapter 4
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
Figure 4-77 Advanced OS Options Sub-menu
Boot Menu
The Boot Menu shown below is displayed by choosing "Boot" using keys and by
pressing key at the BIOS Setup Utility main menu. The current setup states are
displayed as shown in Figure 4-78. If the setup is incorrect, select it using keys,
and change the setting by pressing keys.
If you want to exit from this page and return to main menu, press key.
4. Troubleshooting
Chapter 4 285
Troubleshooting
To configure the A20 Digital Motherboard and BIOS for E5070B :
MY42300633 and above, E5071B : MY42301397 and above
Change Supervisor Password
A password needs to be entered to execute the BIOS setup utility. If a password has not
been assigned in the E5070B/E5071B, set the password using this BIOS setup utility as
follows.
a. Select "Security" using keys at the BIOS Setup Utility main menu and press
key.
d. Press key.
e. If a message of "New supervisor password installed. Press any key to continue" is
displayed, press key.
When you want to change and save a BIOS option, Select "Exit" using keys at the
BIOS Setup Utility main menu and press key. Select “Save CMOS to Flash” using
and press key. Then select “Exit Saving Changes” using and press
key. A message of "SAVE to CMOS and EXIT (Y/N)?" is displayed. Press and
keys.
The BIOS setup utility is closed after the current settings are saved.
286 Chapter 4
5 Replaceable Parts
This chapter contains information for ordering replacement parts for the E5070B/E5071B
5. Replaceable Parts
287
Replaceable Parts
Ordering Information
Ordering Information
To order part listed in the replaceable part lists, quote the Agilent part number (with a
check digit), indicate the quantity required, and address the order to the nearest Agilent
office. The check digit will ensure accurate and timely processing of the order.
To order a part not listed in the replaceable part table, include the instrument model
number, the description and function of the part, and the quantity of parts required. Address
the order to the nearest Agilent office.
• Direct ordering and shipping from the Agilent Parts Center in Mountain View,
California.
• No maximum or minimum on any mail order (there is a minimum order amount for
parts ordered through a local Agilent office when the orders require billing and
invoicing)
• Prepaid transportation (there is a small handling charge for each order).
• No invoices.
In order for Agilent to provide these advantages, please send a check or money order with
each order.
Mail order forms and specific ordering information are available through your local
Agilent sales office. Addresses and telephone numbers are located in a separate document
shipped with the manuals.
288 Chapter 5
Replaceable Parts
Exchange Assemblies
Exchange Assemblies
Under the rebuilt-exchange assembly program, certain factory-repaired and tested
assemblies are available on a trade-in basis. These assemblies are offered al lower cost than
a new assembly, but meet all factory specifications required of a new assembly.
The defective assembly must be returned for credit under the terms of the rebuilt-exchange
assembly program. Any spare assembly stock desired should be ordered using the new
assembly part number.
5. Replaceable Parts
Chapter 5 289
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
290 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
Table 5-1 Top View (Major Assemblies)
5. Replaceable Parts
Chapter 5 291
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
292 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
Table 5-2 Top View (Cables)
5. Replaceable Parts
Chapter 5 293
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
294 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
Table 5-3 Top View (Miscellaneous Parts)
2 E5070-01229 4 1 ANGLE
3 0515-1946 8 4 SCREW
10 1400-1051 4 1 SADDLE-WIRE
11 0403-1143 0 1 GUIDE-PC BD
12 1400-0866 7 1 CLAMP-CABLE
5. Replaceable Parts
Chapter 5 295
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
296 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
5. Replaceable Parts
Chapter 5 297
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
298 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : JP1KKxxxxx, MY42300632
and below, E5071B : JP1KKxxxxx, MY42301396 and below
Table 5-5 Top View (Motherboard and Other Parts)
2 1400-0611 0 1 CLAMP-CABLE
5 1400-1334 6 1 CLAMP-CABLE
6 E5070-01267 7 1 BRACKET PS
5. Replaceable Parts
Chapter 5 299
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
300 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
Table 5-6 Top View (Major Assemblies)
5. Replaceable Parts
Chapter 5 301
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
302 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
Table 5-7 Top View (Cables)
5. Replaceable Parts
Chapter 5 303
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
304 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
Table 5-8 Top View (Miscellaneous Parts)
2 E5070-01229 4 1 ANGLE
3 0515-1946 8 4 SCREW
10 1400-1051 4 1 SADDLE-WIRE
11 0403-1143 0 1 GUIDE-PC BD
12 1400-0866 7 1 CLAMP-CABLE
5. Replaceable Parts
Chapter 5 305
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
306 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
1400-0493 2 CLAMP-CABLE
5. Replaceable Parts
Chapter 5 307
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
308 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B : MY42300633 and above,
E5071B : MY42301397 and above (A27 mass storage disk drive model)
Table 5-10 Top View (Motherboard and Other Parts)
2 1400-0611 0 1 CLAMP-CABLE
5 1400-1391 6 1 CLAMP-CABLE
6 E5070-01267 7 1 BRACKET PS
5. Replaceable Parts
Chapter 5 309
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
310 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Table 5-11 Top View (Removable Hard Disk Assemblies)
5. Replaceable Parts
Chapter 5 311
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
312 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Table 5-12 Top View (Major Assemblies)
5. Replaceable Parts
Chapter 5 313
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
314 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Table 5-13 Top View (Cables)
5. Replaceable Parts
Chapter 5 315
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
316 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Table 5-14 Top View (Miscellaneous Parts)
2 E5070-01256 - 1 ANGLE
3 0515-1946 8 4 SCREW
9 1400-1051 4 1 SADDLE-WIRE
10 0403-1143 0 1 GUIDE-PC BD
11 1400-0866 7 1 CLAMP-CABLE
5. Replaceable Parts
Chapter 5 317
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Top View (Under Power Supply) for E5070B S/N from MY42401101 to
MY42401986, E5071B S/N from MY42402455 to MY42404509
318 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
1400-0493 2 CLAMP-CABLE
5. Replaceable Parts
Chapter 5 319
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Top View (Under Power Supply) for E5070B S/N MY42401987 and
above, E5071B S/N MY42404510 and above
320 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
1400-0493 2 CLAMP-CABLE
12 1400-3361 1 CLAMP-CABLE
5. Replaceable Parts
Chapter 5 321
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
322 Chapter 5
Replaceable Parts
Replaceable Parts List (Top View) for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Table 5-17 Top View (Motherboard and Other Parts)
2 1400-0611 0 1 CLAMP-CABLE
5 1400-1391 6 1 CLAMP-CABLE
6 E5070-01267 7 1 BRACKET PS
5. Replaceable Parts
Chapter 5 323
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
324 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
8 See Table 5-43 - 1 STEP ATTENUATOR ASSEMBLY (Opt. 214, 314, 414)
5. Replaceable Parts
Chapter 5 325
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
8 E5070-01222 5 1 PLATE
326 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
11 1810-0118 1 2 TERMINATION-COAX
Chapter 5 327
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
328 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
11 1810-0118 1 2 TERMINATION-COAX
5. Replaceable Parts
Chapter 5 329
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
330 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
17 1810-0118 1 4 TERMINATION-COAX
5. Replaceable Parts
Chapter 5 331
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
332 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
17 1810-0118 1 4 TERMINATION-COAX
5. Replaceable Parts
Chapter 5 333
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
334 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
19 1810-0118 1 1 TERMINATION-COAX
5. Replaceable Parts
Chapter 5 335
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
336 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix JP1KK, and
MY421, E5071B serial prefix JP1KK, and MY421
19 1810-0118 1 1 TERMINATION-COAX
5. Replaceable Parts
Chapter 5 337
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
338 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
8 See Table 5-44, Table - 1 STEP ATTENUATOR ASSEMBLY (Opt. 214, 314, 414)
5-45
5. Replaceable Parts
Chapter 5 339
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
340 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
12 1810-0118 1 2 TERMINATION-COAX
Chapter 5 341
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
342 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
19 1810-0118 1 4 TERMINATION-COAX
5. Replaceable Parts
Chapter 5 343
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
344 Chapter 5
Replaceable Parts
Replaceable Parts List (Front View) for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above
21 1810-0118 1 1 TERMINATION-COAX
5. Replaceable Parts
Chapter 5 345
Replaceable Parts
Replaceable Parts List (Others)
Front Panel
346 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
3 E5070-01231 8 1 ANGLE-N-CONN
2950-0132 6 1 NUT-HEX-DUB-CHAM
7 5182-7594 2 1 KNOB
15 3100-3711 1 SWITCH-ROTARY
Chapter 5 347
Replaceable Parts
Replaceable Parts List (Others)
Rear View
348 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
4 2950-0054 1 2 NUT-HEX-DBL-CHAM
5. Replaceable Parts
Chapter 5 349
Replaceable Parts
Replaceable Parts List (Others)
Side View
4 E5100-40002 0 4 STANDOFF
5 5041-9167 1 4 FOOT FL
1460-1345 5 2 WIREFORM
350 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
5. Replaceable Parts
Chapter 5 351
Replaceable Parts
Replaceable Parts List (Others)
352 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
5. Replaceable Parts
Chapter 5 353
Replaceable Parts
Replaceable Parts List (Others)
A27 Mass Storage Disk Drive Assembly for E5070B serial prefix
JP1KK, MY423 and below, E5071B serial prefix JP1KK, MY423 and
below
Figure 5-38 A27 Mass Storage Disk Drive Assembly for E5070B serial prefix JP1KK, MY423 and
below, E5071B serial prefix JP1KK, MY423 and below
Table 5-38 A27 Mass Storage Disk Drive Assembly for E5070B serial prefix JP1KK,
MY423 and below, E5071B serial prefix JP1KK, MY423 and below
3 E5070-66530 6 1 DISK DR
354 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Figure 5-39 A27 Mass Storage Disk Drive Mount Assembly for E5070B serial prefix JP1KK,
MY423 and below, E5071B serial prefix JP1KK, MY423 and below
Table 5-39 A27 Mass Storage Disk Drive Mount Assembly for E5070B serial prefix
JP1KK, MY423 and below, E5071B serial prefix JP1KK, MY423 and below
1 1410-1580 4 BUSHING
2 E5070-24011 0 4 SPACER
Chapter 5 355
Replaceable Parts
Replaceable Parts List (Others)
356 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Figure 5-41 A8 RF Switch Assembly for E5070B serial prefix JP1KK or MY421, E5071B serial
prefix JP1KK or MY421
Table 5-41 A8 RF Switch Assembly for E5070B serial prefix or MY421, E5071B serial
prefix JP1KK or MY421
5. Replaceable Parts
Chapter 5 357
Replaceable Parts
Replaceable Parts List (Others)
Figure 5-42 A8 RF Switch Assembly for E5070B serial prefix MY422 and above, E5071B serial
prefix MY422 and above
Table 5-42 A8 RF Switch Assembly for E5070B serial prefix MY422 and above, E5071B
serial prefix MY422 and above
358 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Figure 5-43 A7 Source Step Attenuator Assembly for E5070B serial prefix JP1KK or MY421,
E5071B serial prefix JP1KK or MY421 (Opt. 214, 314, 414)
Table 5-43 A7 Source Step Attenuator Assembly for E5070B serial prefix or MY421,
E5071B serial prefix JP1KK or MY421 (Opt. 214, 314, 414)
2 E5070-01213 6 1 PLATE
5. Replaceable Parts
Chapter 5 359
Replaceable Parts
Replaceable Parts List (Others)
Figure 5-44 A7 Source Step Attenuator Assembly for E5070B serial prefix MY422 and above,
E5071B serial prefix MY422 and above (Opt. 214)
Table 5-44 A7 Source Step Attenuator Assembly for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above (Opt. 214)
2 E5070-01213 6 1 PLATE
360 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Figure 5-45 A7 Source Step Attenuator Assembly for E5070B serial prefix MY422 and above,
E5071B serial prefix MY422 and above (Opt. 314, 414)
Table 5-45 A7 Source Step Attenuator Assembly for E5070B serial prefix MY422 and
above, E5071B serial prefix MY422 and above (Opt. 314, 414)
2 E5070-01214 6 1 PLATE
5. Replaceable Parts
Chapter 5 361
Replaceable Parts
Replaceable Parts List (Others)
2 E5070-01266 9 1 BRACKET
362 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Chapter 5 363
Replaceable Parts
Replaceable Parts List (Others)
1 E5070-20031 6 2 BLOCK
364 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
5. Replaceable Parts
Chapter 5 365
Replaceable Parts
Replaceable Parts List (Others)
Chassis Assembly
1 E5070-60023 8 1 CHASSIS
366 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Keyboard Assembly
5. Replaceable Parts
Chapter 5 367
Replaceable Parts
Replaceable Parts List (Others)
LCD and Inverter Assembly for E5060B S/N MY42401986 and below,
for E5061B S/N MY42404509 and below
368 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
LCD and Inverter Assembly for E5070B S/N MY42401987 and above,
for E5071B S/N MY42404510 and above
5. Replaceable Parts
Chapter 5 369
Replaceable Parts
Replaceable Parts List (Others)
Standby Switch Assembly for E5070B S/N MY42401986 and below, for
E5071B S/N MY42404509 and below
370 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Standby Switch Assembly for E5070B S/N MY42401987 and above, for
E5071B S/N MY42404510 and above
5. Replaceable Parts
Chapter 5 371
Replaceable Parts
Replaceable Parts List (Others)
Figure 5-56 Removable Hard Disk Assemblies for E5070B S/N MY42401101 and above, E5071B
S/N MY42402455 and above
372 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Table 5-56 Removable Hard Disk Assemblies for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
5. Replaceable Parts
Chapter 5 373
Replaceable Parts
Replaceable Parts List (Others)
6 E5070-00621 8 1 COVER
374 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
2 1400-1334 6 2 CLAMP-CABLE
3 1400-1391 1 CLAMP-CABLE
6 1400-0611 0 1 CLAMP-CABLE
Chapter 5 375
Replaceable Parts
Replaceable Parts List (Others)
376 Chapter 5
Replaceable Parts
Replaceable Parts List (Others)
Other Parts
Table 5-60 Other Parts
1150-7846 6 1 MOUSE
1150-7970 7 1 KEYBOARD
*1. The number indicated by “x” in the part number of each manual, sample program disk, or CD-ROM, 0 for the first
edition, is incremented by 1 each time a revision is made. The latest edition comes with the product.
*2.Opt. 0BW only
5. Replaceable Parts
Chapter 5 377
Replaceable Parts
Power Cables and Plug Configurations
378 Chapter 5
6. Replacement Procedure
6 Replacement Procedure
This chapter provides procedure for removing and replacing the major assemblies in
the E5070B/E5071B ENA Series Network Analyzer.
379
Replacement Procedure
Replacing an Assembly
Replacing an Assembly
The following steps show the sequence for replacing an assembly in a E5070B/E5071B RF
Network Analyzer.
WARNING These servicing instructions are for use by qualified personnel only. To avoid
electrical shock, do not perform any servicing unless you are qualified to do so.
WARNING The opening of covers or removal of parts is likely to expose dangerous voltages.
Disconnect the instrument from its power supply.
CAUTION Many of the assemblies in this instrument are very susceptible to damage from
ESD(electrostatic discharge). Perform the following procedures only at a static-safe
workstation and wear a grounding strap.
380 Chapter 6
6. Replacement Procedure
Replacement Procedure
Required Tools
Required Tools
The following tools are required for repair of E5070B/E5071B.
Table 6-1 Required Tools
Assembly TORX screwdriver flat edge torque cutting file hex key open wrench open torque
screwdriver screwdriver plier wrench*1
TORX T10
T8 T10 T15 T20 1.5 mm 9/16 in. 5/8 in. 5/16 in.
Outer Cover √ √ √
Front Panel √ √ √ √
Mass Storage √ √ √
Removable Hard Disk
√ √ √ √
Drive
Removable Hard Disk
√ √ √
Assembly
FDD √ √ √
Power Supply √ √ √ √
Handler I/O √ √ √
CPU √ √ √
DIMM √ √ √ √
Digital Mother Board √ √ √ √
Source Board √ √ √ √ √ √
Receiver Board √ √ √ √ √
Level Vernier √ √ √ √ √
Fan & Switch Control
√ √ √
Board
Fan √ √ √ √ √
Attenuator √ √ √ √ √
Switch √ √ √ √ √
Power Amp. √ √ √ √ √
Analog Motherboard √ √ √ √ √ √
Analog Interface
√ √ √
Board
GPIB Board √ √ √
USB (USBTMC)
√ √ √
Interface Card
Display Board √ √ √
Front Keyboard √ √ √ √ √
Inverter √ √ √ √ √
LCD √ √ √ √
LCD Backlight √ √ √ √
Reference Oven √ √ √
N connector √ √ √ √ √ √ √
*1. It’s recommended to use E5070-65100 if you want to wrench nuts easily.
Chapter 6 381
Replacement Procedure
Outer Cover Removal
Tools Required
• TORX screwdriver, T10, T15, and T20
Procedure
Refer to Figure 6-1 for this procedure.
Step 2. Put a plastic cover (p/n 5959-8096, item 1) on the front panel of the analyzer and place the
analyzer on flat table with its front panel down.
Step 4. Remove the four TORX T10 screws (item 3) from the bottom.
Step 5. Remove the four TORX T20 screws (item 4) fastening the side strap handles.
Step 6. Remove the four TORX T15 screws (item 5) fastening the four rear foot
Step 7. Remove the six TORX T15 screws (item 6) fastening the cover.
382 Chapter 6
6. Replacement Procedure
Replacement Procedure
Outer Cover Removal
Chapter 6 383
Replacement Procedure
Front Panel Removal for E5070B : JP1KKxxxxx, MY42300632 and below,
E5071B : JP1KKxxxxx, MY42301396 and below
Tools required
• TORX screwdriver, T10, T15, and T20
• flat edge screwdriver
Procedure
Refer to Figure 6-2 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Release the cable (item 8) from the cable clamp (item 1)
Step 3. Remove the top trim (item 2) using a flat edge driver.
Step 4. Remove the eight TORX T15 screws (item 3) fastening front panel.
Step 5. Gradually push the front panel assembly towards the outside.
Step 6. Release the mylar cable (item 4) from the cable clamp (item 5).
Step 7. Disconnect the three cables (item 6, 7 and 8) from the front panel.
384 Chapter 6
6. Replacement Procedure
Replacement Procedure
Front Panel Removal for E5070B : JP1KKxxxxx, MY42300632 and below,
E5071B : JP1KKxxxxx, MY42301396 and below
Figure 6-2 Front Panel Removal
Chapter 6 385
Replacement Procedure
Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454
Tools required
• TORX screwdriver, T10, T15, and T20
• flat edge screwdriver
Procedure
Refer to Figure 6-3 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Release the cable (item 8) from the cable clamp (item 1)
Step 3. Remove the top trim (item 2) using a flat edge driver.
Step 4. Remove the eight TORX T15 screws (item 3) fastening front panel.
Step 5. Gradually push the front panel assembly towards the outside.
Step 6. Release the mylar cable (item 4) from the cable clamp (item 5).
Step 7. Disconnect the three cables (item 6, 7 and 8) from the front panel.
386 Chapter 6
6. Replacement Procedure
Replacement Procedure
Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454
Figure 6-3 Front Panel Removal
Chapter 6 387
Replacement Procedure
Front Panel Removal for E5070B S/N MY42401101 and above, E5071B S/N
MY42402455 and above
Tools required
• TORX screwdriver, T10, T15, and T20
• flat edge screwdriver
Procedure
Refer to Figure 6-4 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Release the cable (item 8) from the cable clamp (item 1)
Step 3. Remove the top trim (item 2) using a flat edge driver.
Step 4. Remove the eight TORX T15 screws (item 3) fastening front panel.
Step 5. Gradually push the front panel assembly towards the outside.
Step 6. Release the mylar cable (item 4) from the cable clamp (item 5).
Step 7. Disconnect the three cables (item 6, 7 and 8) from the front panel.
The mylar cable and its connector are very fragile. It’s recommended that you should
replace the mylar cable when the cable is disconnected and reconnected frequently.
388 Chapter 6
6. Replacement Procedure
Replacement Procedure
Front Panel Removal for E5070B S/N MY42401101 and above, E5071B S/N
MY42402455 and above
Figure 6-4 Front Panel Removal
Chapter 6 389
Replacement Procedure
A27 Mass Storage Disk Drive Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-5 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the four TORX T10 screws (item 1) fastening the A27 mass storage disk drive.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
Step 2. Restore backup files onto the new mass storage disk drive as described in “Restore Backup
Files onto the New Mass Storage Disk Drive” on page 391.
Step 3. Calibrate the touch screen as described in “Calibration of the Touch Screen” on page 391.
390 Chapter 6
6. Replacement Procedure
Replacement Procedure
A27 Mass Storage Disk Drive Replacement
Restore Backup Files onto the New Mass Storage Disk Drive
The Backup files is stored in the flash ROM on the analog interface board. The file must be
restored in the new mass storage disk drive with the adjustment program. For detail of the
adjustment program, refer to Chapter 3, “Adjustment,” on page 193.
Step 1. Press .
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
Chapter 6 391
Replacement Procedure
Removable Hard Disk Drive Replacement for E5070B S/N MY42401101 and
above, for E5071B S/N MY42402455 and above
Tools Required
• TORX screwdriver, T8, T10, T15, and T20
Removal procedure
Refer to Figure 6-7 for this procedure.
Step 1. Remove the four TORX T10 screws (item 1), and remove the removable hard disk drive’s
lid (item 2).
Step 2. Remove the four TORX T10 screws (item 3) fastening the HDD bracket.
Step 4. Remove the four TORX T10 screws (item 5), and remove the HDD bracket.
Step 5. Remove the hard disk drive from the connector board.
392 Chapter 6
6. Replacement Procedure
Replacement Procedure
Removable Hard Disk Drive Replacement for E5070B S/N MY42401101 and
above, for E5071B S/N MY42402455 and above
Replacement Procedure
Step 1. Paste the heat transfer sheet to the hard disk drive rear side (Refer to Figure 6-8).
Step 2. Replace the removable hard disk drive by inverse procedure of removal.
Step 3. Restore backup files onto the new removable hard disk drive as described in “Restore
Backup Files onto the New Removable Hard Disk Drive” on page 394.
Step 4. Calibrate the touch screen as described in “Calibration of the Touch Screen” on page 395.
Chapter 6 393
Replacement Procedure
Removable Hard Disk Drive Replacement for E5070B S/N MY42401101 and
above, for E5071B S/N MY42402455 and above
Restore Backup Files onto the New Removable Hard Disk Drive
The Backup files is stored in the flash ROM on the analog interface board. The file must be
restored in the new removable hard disk drive with the adjustment program. For detail of
the adjustment program, refer to Chapter 3, “Adjustment,” on page 193.
394 Chapter 6
6. Replacement Procedure
Replacement Procedure
Removable Hard Disk Drive Replacement for E5070B S/N MY42401101 and
above, for E5071B S/N MY42402455 and above
Step 1. Press .
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
Chapter 6 395
Replacement Procedure
Removable Hard Disk Assembly Removal for E5070B S/N MY42401101 and
above, for E5071B S/N MY42402455 and above
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal procedure
Refer to Figure 6-10 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 3. Remove the four TORX T10 screws (item 2) fastening the angle.
Step 4. Lift the angle with the removable hard disk assembly.
CAUTION Lift the angle with the removable hard disk assembly slowly to prevent tension from
cables.
Step 5. Remove three TORX T10 screws (item 3) and remove the removable hard disk assembly
from the angle.
Step 7. Disconnect the cable (item 5) connected to the crystal oven board.
Step 8. Disconnect the flat cable (item 6) connected to the mother board.
Replacement Procedure
Step 1. Replace the removable hard disk assembly by inverse procedure of removal.
396 Chapter 6
6. Replacement Procedure
Replacement Procedure
Removable Hard Disk Assembly Removal for E5070B S/N MY42401101 and
above, for E5071B S/N MY42402455 and above
Figure 6-10 Removable Hard Disk Assembly Removal
Chapter 6 397
Replacement Procedure
A28 FDD (Floppy Disk Drive) Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal procedure
Refer to Figure 6-11 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 3. Disconnect the cable (item 2) connected from the power supply.
Step 7. Remove the four TORX T10 screws (item 5) fastening the FDD to the mount.
Step 8. Remove the two TORX T10 screws (item 6) fastening the plate.
Replacement Procedure
Step 1. Fasten the plate to the new FDD, and then fasten it to the mount as shown in Figure 6-11.
398 Chapter 6
6. Replacement Procedure
Replacement Procedure
A28 FDD (Floppy Disk Drive) Replacement
Chapter 6 399
Replacement Procedure
A50 ATX Power Supply Assembly Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• torque screwdriver, TORX T10 (set to 7 in-lb)
Removal Procedure
Refer to Figure 6-12 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the removable hard disk assembly as described in “Removable Hard Disk
Assembly Removal for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above” on page 396.
NOTE If it is a version where the E5070B/71B doesn't have the removable hard disk assembly,
skip this step.
Step 3. Release the cables from the clamp (item 1) on the angle.
Step 5. Remove the two TORX T15 screws (item 4) fastening the power supply holder.
Step 6. Remove the four TORX T10 screws (item 5) from the rear panel.
Step 7. Lift the A50 ATX power supply assembly toward front side.
Replacement Procedure
Step 1. Replace the A50 ATX power supply assembly by inverse procedure of removal.
NOTE Fasten the TORX T10 screws (item 5) using a torque screwdriver.
400 Chapter 6
6. Replacement Procedure
Replacement Procedure
A50 ATX Power Supply Assembly Replacement
Chapter 6 401
Replacement Procedure
A23 Handler I/O Board Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-13 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the removable hard disk assembly as described in “Removable Hard Disk
Assembly Removal for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above” on page 396.
NOTE If it is a version where the E5070B/71B doesn't have the removable hard disk assembly,
skip this step.
Step 3. Disconnect the flat cable (item 1) from the A23 handler I/O board.
Step 4. Remove the two TORX T10 screws (item 2) fastening the A23 handler I/O board.
Replacement Procedure
Step 1. Replace the A23 handler I/O board by inverse procedure of removal.
402 Chapter 6
6. Replacement Procedure
Replacement Procedure
A23 Handler I/O Board Replacement
Chapter 6 403
Replacement Procedure
DIMM Module Replacement for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
NOTE It cannot replace only DIMMs, when an E5070B serial number is after MY42300633, or
when an E5071B serial number is after MY42301397. It is necessary to replace an A20
digital motherboard.
Tools Required
• TORX screwdriver, T10, T15, and T20
• Cutting plier or scissors
Removal Procedure
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the removable hard disk assembly as described in “Removable Hard Disk
Assembly Removal for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above” on page 396.
NOTE If it is a version where the E5070B/71B doesn't have the removable hard disk assembly,
skip this step.
Step 3. Remove the A26 LCD Interface as described in “A26 LCD Interface Card Replacement for
E5070B S/N prefix JP1KK, MY4231100 and below, E5071B S/N prefix JP1KK,
MY4232454 and below” on page 444.
Step 5. Unlatch the two clips and remove the DIMM memory.
404 Chapter 6
6. Replacement Procedure
Replacement Procedure
DIMM Module Replacement for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
Figure 6-14 DIMM Module Removal
Replacement Procedure
Step 1. Install the DIMM memory in the socket and latch the clips.
Step 2. Clamp the two clips using the new cable tie as shown in Figure 6-14.
Chapter 6 405
Replacement Procedure
A20 Digital Motherboard Replacement for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below
Tools Required
• TORX screwdriver, T10, T15, and T20
• Cutting plier or scissors
Removal Procedure
Refer to Figure 6-15 for this procedure.
Step 1. Remove the mouse, the external keyboard and the parallel printer cable if they are
connected to the rear panel.
Step 2. Remove the outer cover as described in “Outer Cover Removal” on page 382
Step 3. Remove the removable hard disk assembly as described in “Removable Hard Disk
Assembly Removal for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above” on page 396.
NOTE If it is a version where the E5070B/71B doesn't have the removable hard disk assembly,
skip this step.
• A50 ATX Power Supply Assembly (refer to “A50 ATX Power Supply Assembly
Replacement” on page 400.)
• A23 Handler I/O board (refer to “A23 Handler I/O Board Replacement” on page 402.)
• A21 Analog Interface board (refer to “A21 Analog Interface Board Replacement” on
page 440.)
• A24 GPIB board (refer to “A24 GPIB Card Replacement” on page 442.)
• A26 LCD Interface Card (refer to “A26 LCD Interface Card Replacement for E5070B
S/N prefix JP1KK, MY4231100 and below, E5071B S/N prefix JP1KK, MY4232454
and below” on page 444.)
Step 6. Remove three TORX T10 screws (item 6) fastening the guide plate.
Step 7. Remove eight TORX T10 screws (item 7) fastening the digital motherboard.
406 Chapter 6
6. Replacement Procedure
Replacement Procedure
A20 Digital Motherboard Replacement for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below
Replacement Procedure
Step 1. Remove the replaceable parts on the old board.
a. Remove the DIMM modules as described in “DIMM Module Replacement for E5070B
: JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and
below” on page 404.
b. Disassemble the guide plate from rear side of the mother board.
Figure 6-15 A20 Digital Motherboard Removal for E5070B : JP1KKxxxxx, MY42300632 and
below, E5071B : JP1KKxxxxx, MY42301396 and below
Chapter 6 407
Replacement Procedure
A20 Digital Motherboard Replacement for E5070B : MY42300633 thru
MY4231100, E5071B : MY42301397 thru MY4232454
Tools Required
• TORX screwdriver, T10, T15, and T20
• Cutting plier or scissors
Removal Procedure
Refer to Figure 6-16 for this procedure.
Step 1. Remove the mouse, the external keyboard and the parallel printer cable if they are
connected to the rear panel.
Step 2. Remove the outer cover as described in “Outer Cover Removal” on page 382
Step 3. Remove the removable hard disk assembly as described in “Removable Hard Disk
Assembly Removal for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above” on page 396.
NOTE If it is a version where the E5070B/71B doesn't have the removable hard disk assembly,
skip this step.
• A50 ATX Power Supply Assembly (refer to “A50 ATX Power Supply Assembly
Replacement” on page 400.)
• A23 Handler I/O board (refer to “A23 Handler I/O Board Replacement” on page 402.)
• A21 Analog Interface board (refer to “A21 Analog Interface Board Replacement” on
page 440.)
• A24 GPIB board (refer to “A24 GPIB Card Replacement” on page 442.)
• A26 LCD Interface Card (refer to “A26 LCD Interface Card Replacement for E5070B
S/N prefix JP1KK, MY4231100 and below, E5071B S/N prefix JP1KK, MY4232454
and below” on page 444.)
Step 6. Remove three TORX T10 screws (item 6) fastening the guide plate.
Step 7. Remove eight TORX T10 screws (item 7) fastening the digital motherboard.
408 Chapter 6
6. Replacement Procedure
Replacement Procedure
A20 Digital Motherboard Replacement for E5070B : MY42300633 thru
MY4231100, E5071B : MY42301397 thru MY4232454
Replacement Procedure
Step 1. Disassemble the guide plate from rear side of the mother board.
Step 3. Clamp the two clips using the new cable tie for DIMM memorys.
Figure 6-16 A20 Digital Motherboard Removal for E5070B : MY42300633 and above, E5071B :
MY42301397 and above
Chapter 6 409
Replacement Procedure
A20 Digital Motherboard Replacement for E5070B S/N MY42401101 and
above, E5071B S/N MY42402455 and above
Tools Required
• TORX screwdriver, T10, T15, and T20
• Cutting plier or scissors
Removal Procedure
Refer to Figure 6-17 for this procedure.
Step 1. Remove the mouse, the external keyboard and the parallel printer cable if they are
connected to the rear panel.
Step 2. Remove the outer cover as described in “Outer Cover Removal” on page 382
Step 3. Remove the removable hard disk assembly as described in “Removable Hard Disk
Assembly Removal for E5070B S/N MY42401101 and above, for E5071B S/N
MY42402455 and above” on page 396.
NOTE If it is a version where the E5070B/71B doesn't have the removable hard disk assembly,
skip this step.
• A50 ATX Power Supply Assembly (refer to “A50 ATX Power Supply Assembly
Replacement” on page 400.)
• A23 Handler I/O board (refer to “A23 Handler I/O Board Replacement” on page 402.)
• A21 Analog Interface board (refer to “A21 Analog Interface Board Replacement” on
page 440.)
• A24 GPIB board (refer to “A24 GPIB Card Replacement” on page 442.)
• A26 LCD Interface Card (refer to “A26 LCD Interface Card Replacement for E5070B
S/N prefix JP1KK, MY4231100 and below, E5071B S/N prefix JP1KK, MY4232454
and below” on page 444.)
NOTE The mylar cable (item 6) and its connector are very fragile. It’s recommended to replace
the mylar cable when the cable is disconnedted and connected frequently.
410 Chapter 6
6. Replacement Procedure
Replacement Procedure
A20 Digital Motherboard Replacement for E5070B S/N MY42401101 and
above, E5071B S/N MY42402455 and above
Step 6. Remove three TORX T10 screws (item 7) fastening the guide plate.
Step 7. Remove eight TORX T10 screws (item 8) fastening the digital motherboard.
Replacement Procedure
Step 1. Disassemble the guide plate from rear side of the mother board.
Step 3. Clamp the two clips using the new cable tie for DIMM memorys.
Chapter 6 411
Replacement Procedure
A1 Source Board Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Box wrench, 5/8 inch
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-18 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Remove the two 5/8 inch nuts (item 1) fastening the BNC connector on the rear panel.
Step 3. Disconnect the all semi-rigid cables and the termination connected to the board referring to
the figures as shown in Table 6-2.
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Option Figure
214 Figure 5-21 on page 328 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-28 on page 341 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
314 Figure 5-23 on page 332 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-29 on page 342 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
414 Figure 5-25 on page 336 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-30 on page 344 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
Step 4. Remove the two TORX T10 screws fastening the extractor. (item 2)
412 Chapter 6
6. Replacement Procedure
Replacement Procedure
A1 Source Board Replacement
Replacement Procedure
Step 1. Replace the A1 source board by inverse procedure of removal.
NOTE Fasten the semi rigid cable using a torque wrench for replacement.
Chapter 6 413
Replacement Procedure
A2 Receiver Board Replacement
Tools Required
• TORX screwdriver, T10
• Flat edge screwdriver
• Open-end torque wrench, 9/16 inch
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-19 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Remove the N-connector holder as described in “N Connector Replacement” on page 462.
Step 3. Disconnect the semi-rigid cables and the termination from the receiver module as shown in
Table 6-3.
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Option Figure
214 Figure 5-21 on page 328 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-28 on page 341 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
314 Figure 5-23 on page 332 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-29 on page 342 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
414 Figure 5-25 on page 336 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-30 on page 344 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
Step 4. Remove the all TORX T10 screws fastening the extractor . (item 1)
414 Chapter 6
6. Replacement Procedure
Replacement Procedure
A2 Receiver Board Replacement
NOTE Slide the two board at once if two A2 receiver boards are slotted.
Replacement Procedure
Step 1. Replace the A2 receiver boards by inverse procedure of removal.
NOTE Fasten the semi rigid cable using a torque wrench for replacement.
Chapter 6 415
Replacement Procedure
A2 Receiver Board Replacement
Step 1. Remove the eight TORX T10 screws (item 1) fastening the A2 receiver boards.
Step 1. Remove the seven TORX T15 screws (item 2) fastening the A15 Receiver-RF board (item
3) and the A17 Receiver-IF board (item 4).
416 Chapter 6
6. Replacement Procedure
Replacement Procedure
A2 Receiver Board Replacement
Chapter 6 417
Replacement Procedure
A3 Level Vernier Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Open-end wrench, 5/8 inch
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-21 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Disconnect the all semi-rigid cables and the termination connected to the boards referring
to the figures as shown in Table 6-4.
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Option Figure
214 Figure 5-21 on page 328 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-28 on page 341 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
314 Figure 5-23 on page 332 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-29 on page 342 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
414 Figure 5-25 on page 336 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-30 on page 344 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
Step 3. Remove the two TORX T10 screws fastening the extractor. (item 2)
418 Chapter 6
6. Replacement Procedure
Replacement Procedure
A3 Level Vernier Replacement
Replacement Procedure
NOTE Fasten the semi rigid cable using a torque wrench for replacement.
Chapter 6 419
Replacement Procedure
B1 Fan & A6 Switch Control Board Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-22 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the power supply as described in “A50 ATX Power Supply Assembly
Replacement” on page 400.
Step 4. Remove two TORX T10 screws (item 1) fastening the board.
Replacement Procedure
Step 1. Replace the B1 fun & switch control board by inverse procedure of removal.
420 Chapter 6
6. Replacement Procedure
Replacement Procedure
B1 Fan & A6 Switch Control Board Replacement
Chapter 6 421
Replacement Procedure
B1 Fan Replacement
B1 Fan Replacement
There are three fans as shown in Figure 6-23.
Figure 6-23 The location and direction of the fan 1, fan 2, and fan 3
422 Chapter 6
6. Replacement Procedure
Replacement Procedure
B1 Fan Replacement
Fan 1 and 2
Tools Required
Removal Procedure
Refer to Figure 6-24 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the front panel as described in“Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 3. Remove the three TORX T10 screws (item 1) fastening the fan plate.
Step 4. Disconnect the fan cable from the fan & switch control board.
Step 6. Remove the eight rivet (plastic) fastening the fan as shown in Figure 6-24.
Replacement Procedure
Chapter 6 423
Replacement Procedure
B1 Fan Replacement
Fan 3
Tools Required
Removal Procedure
Refer to Figure 6-25 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Disconnect the fan cable from the fan & switch control board.
Step 4. Remove the three TORX T10 screws (item 1) fastening the fan assembly.
Step 6. Remove the four TORX T20 screws (item 2) fastening the fan.
Replacement Procedure
424 Chapter 6
6. Replacement Procedure
Replacement Procedure
B1 Fan Replacement
Chapter 6 425
Replacement Procedure
A7 Source Step Attenuator Replacement for E5070B S/N prefix JP1KK or
MY421, E5071B S/N prefix JP1KK or MY421 (Option 214, 314 and 414)
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-26 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 3. Remove the three TORX T10 screws (item 1) fastening the fan plate.
Step 4. Disconnect the semi-rigid cables (item 2 and 3) from the A7 source step attenuator.
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Step 6. Remove the two TORX T10 screws (item 5) fastening the A7 source step attenuator holder
from the front side.
Step 7. Remove the two TORX T10 screws (item 6) fastening the A7 source step attenuator to the
holder.
Replacement Procedure
Step 1. Fasten the new source step attenuator to the holder as shown in Figure 6-26.
Step 2. Replace the A7 source step attenuator holder by inverse procedure of removal.
NOTE Fasten the semi rigid cable using a torque wrench for replacement.
426 Chapter 6
6. Replacement Procedure
Replacement Procedure
A7 Source Step Attenuator Replacement for E5070B S/N prefix JP1KK or
MY421, E5071B S/N prefix JP1KK or MY421 (Option 214, 314 and 414)
Figure 6-26 A7 Source Step Attenuator Removal
Chapter 6 427
Replacement Procedure
A7 Source Step Attenuator Replacement for E5070B S/N prefix MY422 and
above, E5071B S/N prefix MY422 and above (Option 214, 314 and 414)
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-27 for Opt.214, and Figure 6-28 for Opt.314, or 414 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 3. Remove the three TORX T10 screws (item 1) fastening the fan plate.
Step 4. Disconnect the semi-rigid cables (item 2 and 3) from the A7 source step attenuator.
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Step 6. Remove the two TORX T10 screws (item 5) fastening the A7 source step attenuator holder
from the front side.
Step 7. Remove the two TORX T10 screws (item 6) fastening the A7 source step attenuator to the
holder.
Replacement Procedure
Step 1. Fasten the new source step attenuator to the holder as shown in Figure 6-27 or Figure 6-28.
Step 2. Replace the A7 source step attenuator holder by inverse procedure of removal.
NOTE Fasten the semi rigid cable using a torque wrench for replacement.
428 Chapter 6
6. Replacement Procedure
Replacement Procedure
A7 Source Step Attenuator Replacement for E5070B S/N prefix MY422 and
above, E5071B S/N prefix MY422 and above (Option 214, 314 and 414)
Figure 6-27 A7 Source Step Attenuator Removal for Opt.214
Chapter 6 429
Replacement Procedure
A7 Source Step Attenuator Replacement for E5070B S/N prefix MY422 and
above, E5071B S/N prefix MY422 and above (Option 214, 314 and 414)
430 Chapter 6
6. Replacement Procedure
Replacement Procedure
A8 RF Switch Replacement
A8 RF Switch Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-29 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Remove the three TORX T10 screws (item 1) fastening the fan plate.
Step 3. Disconnect the all semi-rigid cables from the switch on the left side. (refer to the figure as
shown in Table 6-5)
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Chapter 6 431
Replacement Procedure
A8 RF Switch Replacement
Option Figure
214 Figure 5-21 on page 328 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-28 on page 341 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
314 Figure 5-23 on page 332 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-29 on page 342 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
414 Figure 5-25 on page 336 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-30 on page 344 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
Step 4. Disconnect the semi-rigid cables (item 1 and 2 in Figure 6-27) from the A7 source step
attenuator if the option 214, 314 or 414 is installed.
Step 6. Remove the two TORX T10 screws fastening the switch. (item 2)
Replacement Procedure
NOTE Fasten the semi-rigid cable using a torque wrench for replacement.
432 Chapter 6
6. Replacement Procedure
Replacement Procedure
A8 RF Switch Replacement
Chapter 6 433
Replacement Procedure
A8 RF Switch Replacement
Removal Procedure
Refer to Figure 6-30 for this procedure.
a. Remove the outer cover as described in “Outer Cover Removal” on page 382.
b. Remove the front panel as described in “Front Panel Removal for E5070B :
JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and
below” on page 384.
c. Remove the floppy disk drive as described in “A28 FDD (Floppy Disk Drive)
Replacement” on page 398.
d. Remove the three TORX T10 screws (item 1 in Figure 6-29) fastening the fan plate.
e. Release the cables from the cable clamp (item 1).
f. Disconnect the all semi-rigid cables from the switches on the left side and the A4
power amp. module. (refer to the figure as shown in Table 6-5 on page 432)
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
g. Remove the two TORX T10 screws (item 2) fastening the A4 power amp. module
assembly.
h. Release the cables from the cable clamp (item 3).
i. Disconnect the cables from the switches and the A4 power amp. module.
Step 2. Remove the TORX T10 screws (item 4) fastening the switches.
Replacement Procedure
Step 1. Fasten the new switches to the A4 power amp. module assembly as shown in Figure 6-30.
NOTE Confirm that the cables are connected as shown in Figure 6-30 on page 435
Fasten the semi rigid cable using a torque wrench for replacement.
434 Chapter 6
6. Replacement Procedure
Replacement Procedure
A8 RF Switch Replacement
Chapter 6 435
Replacement Procedure
A4 Power Amp. Module Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-31 for this procedure.
Step 1. Remove the A4 power amp. module assembly as described in Step 1 in “A8 RF Switch
(receiver board side)” on page 434.
Step 2. Remove the four TORX T10 screws (item 1) fastening the A4 power amp. module.
Replacement Procedure
Step 1. Fasten the new amp. module as shown in Figure 6-31.
Step 2. Replace the A4 power amp. module assembly by inverse procedure of removal.
NOTE Confirm that the cables are connected as shown in Figure 6-30 on page 435.
Fasten the semi rigid cable using a torque wrench for replacement.
436 Chapter 6
6. Replacement Procedure
Replacement Procedure
A4 Power Amp. Module Replacement
Chapter 6 437
Replacement Procedure
A10 Analog Motherboard Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Open-end wrench, 5/16 inch and 5/8 inch
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Removal Procedure
Refer to Figure 6-32 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the front frame as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 3. Remove the source board as described in “A1 Source Board Replacement” on page 412.
Step 4. Remove the receiver board as described in “A2 Receiver Board Replacement” on
page 414.
Step 5. Remove the level vernier as described in “A3 Level Vernier Replacement” on page 418.
Step 6. Remove the power supply as described in “A50 ATX Power Supply Assembly
Replacement” on page 400
Step 8. Remove the nine TORX T10 screws (item 1) fastening the A10 analog mother board.
438 Chapter 6
6. Replacement Procedure
Replacement Procedure
A10 Analog Motherboard Replacement
Replacement Procedure
Step 1. Replace the A10 analog mother board by inverse procedure of removal.
Chapter 6 439
Replacement Procedure
A21 Analog Interface Board Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-33 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove four TORX T10 screw (item 1) fastening the angle to the side panels.
Step 3. Disconnect the cables (item 2, 3 and 4) from the A21 Analog Interface board.
Step 4. Remove a TORX T10 screw (item 5) fastening the board to the rear panel.
Replacement Procedure
Step 1. Replace the A21 analog interface board by inverse procedure of removal.
440 Chapter 6
6. Replacement Procedure
Replacement Procedure
A21 Analog Interface Board Replacement
NOTE Refer to Chapter 5, “Replaceable Parts,” on page 287 for the latest cabling too.
Chapter 6 441
Replacement Procedure
A24 GPIB Card Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-34 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the TORX T10 screw (item 1) fastening the card to the rear panel.
CAUTION Be careful not to injure the cable connecting the display board to the front assembly.
Replacement Procedure
442 Chapter 6
6. Replacement Procedure
Replacement Procedure
A24 GPIB Card Replacement
NOTE Refer to Chapter 5, “Replaceable Parts,” on page 287 for the latest cabling too.
Chapter 6 443
Replacement Procedure
A26 LCD Interface Card Replacement for E5070B S/N prefix JP1KK,
MY4231100 and below, E5071B S/N prefix JP1KK, MY4232454 and below
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-36 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Disconnect the mylar cables (item 1) from the display board.
NOTE Use the following procedure when you handle the mylar ribbon cable. The mylar cable and
their connector are very fragile. It’s recommended to replace the mylar cable when you are
instructed to disconnect and connect the mylar cable frequently.
a. Pry up the retainer slightly at either end of the connector as shown in Figure 6-35, using
a small flat edge screwdriver.
Step 3. Remove the TORX T10 screw (item 2) fastening the card to the rear panel.
Replacement Procedure
Step 1. Replace the A26 LCD interface card by inverse procedure of removal.
444 Chapter 6
6. Replacement Procedure
Replacement Procedure
A26 LCD Interface Card Replacement for E5070B S/N prefix JP1KK,
MY4231100 and below, E5071B S/N prefix JP1KK, MY4232454 and below
Figure 6-36 A26 LCD Interface Card Removal for E5070B S/N prefix JP1KK, MY4231100 and
below, E5071B S/N prefix JP1KK, MY4232454 and below
NOTE Refer to Chapter 5, “Replaceable Parts,” on page 287 for the latest cabling too.
Chapter 6 445
Replacement Procedure
USB (USBTMC) Interface Card Replacement for E5070B S/N MY42401101 and
above, E5071B S/N MY42402455 and above
Tools Required
• TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-37 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the TORX T10 screw fastening the card (item 1) to the rear panel.
Step 3. Remove the flat cable (item 2) from the A21 analog interface board.
CAUTION Be careful not to injure the cable connecting the display board to the front assembly.
Replacement Procedure
Step 1. Replace the USB (USBTMC) interface card by inverse procedure of removal.
446 Chapter 6
6. Replacement Procedure
Replacement Procedure
USB (USBTMC) Interface Card Replacement for E5070B S/N MY42401101 and
above, E5071B S/N MY42402455 and above
Figure 6-37 USB (USBTMC) Interface Card Removal for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
NOTE Refer to Chapter 5, “Replaceable Parts,” on page 287 for the latest cabling too.
Chapter 6 447
Replacement Procedure
A22 Front Panel Keyboard Replacement for E5070B S/N prefix JP1KK,
MY4231100 and below, E5071B S/N prefix JP1KK, MY4232454 and below
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Hex key, 1.5 mm
Removal Procedure
Refer to Figure 6-38 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Release the clamped cables from the clamps (item 1) on the cover.
Step 3. Disconnect the cables (item 2 and 3) from the touch screen controller if the option 016 is
installed.
Step 4. Remove the four TORX T10 screws (item 4) fastening the cover to the front panel.
Step 5. Disconnect the cable (item 5 and 6) connected through the cover.
NOTE The item 6 is not connected in the E5070B/E5071B without the option 016.
Step 6. Disconnect the two cables (item 7 and 8) from the front keyboard.
Step 7. Remove the knob from the front panel with a 1.5 mm hex key.
Step 8. Remove the two TORX T15 screws (item 9) fastening the inverter assembly.
Step 9. Remove the eight TORX T10 screws (item 10) fastening the A22 front panel keyboard.
Step 10. Separate the keyboard into the board and the rubber key.
Replacement Procedure
Step 1. Insert the whole jut of the rubber key into the hole on the new A22 front panel keyboard.
Step 2. Replace the A22 front panel keyboard by inverse procedure of removal.
448 Chapter 6
6. Replacement Procedure
Replacement Procedure
A22 Front Panel Keyboard Replacement for E5070B S/N prefix JP1KK,
MY4231100 and below, E5071B S/N prefix JP1KK, MY4232454 and below
Figure 6-38 A22 Front Panel Keyboard Removal
Chapter 6 449
Replacement Procedure
A22 Front Panel Keyboard Replacement for E5070B S/N MY42401101 and
above, E5071B S/N MY42402455 and above
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Hex key, 1.5 mm
Removal Procedure
Refer to Figure 6-39 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B S/N
MY42401101 and above, E5071B S/N MY42402455 and above” on page 388.
Step 2. Release the clamped cables from the clamps (item 1) on the cover.
Step 3. Disconnect the cables (item 2 and 3) from the touch screen controller if the option 016 is
installed.
Step 4. Remove the four TORX T10 screws (item 4) fastening the cover to the front panel.
NOTE The item 5 is not connected in the E5070B/E5071B without the option 016.
Step 6. Disconnect the four cables (item 6,7,8 and 9) from the front keyboard.
Step 7. Remove the knob from the front panel with a 1.5 mm hex key.
Step 8. Remove the two TORX T15 screws (item 10) fastening the inverter assembly.
Step 9. Remove the eight TORX T10 screws (item 11) fastening the A22 front panel keyboard.
Step 10. Separate the keyboard into the board and the rubber key.
Replacement Procedure
Step 1. Insert the whole jut of the rubber key into the hole on the new A22 front panel keyboard.
Step 2. Replace the A22 front panel keyboard by inverse procedure of removal.
450 Chapter 6
6. Replacement Procedure
Replacement Procedure
A22 Front Panel Keyboard Replacement for E5070B S/N MY42401101 and
above, E5071B S/N MY42402455 and above
Figure 6-39 A22 Front Panel Keyboard Removal
Chapter 6 451
Replacement Procedure
A52 Inverter Board Replacement for E5070B S/N prefix JP1KK, MY4231100
and below, E5071B S/N prefix JP1KK, MY4232454 and below
Tools Required
• TORX screwdriver, T8, T10, T15, and T20
• Flat edge screwdriver
Removal Procedure
Refer to Figure 6-40 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Release the clamped cables from the clamps (item 1) on the cover.
Step 3. Disconnect the cables (item 2 and 3) from the touch screen controller if the option 016 is
installed.
Step 4. Remove the four TORX T10 screws (item 4) fastening the cover to the front panel.
Step 5. Disconnect the cable (item 5 and 6) connected through the cover.
NOTE The item 6 is not connected in the E5070B/E5071B without the option 016.
Step 6. Disconnect the two cables (item 7 and 8) from the inverter.
Step 7. Remove the two TORX T15 screws (item 9) fastening the inverter assembly.
Step 8. Remove the two TORX T8 screws (item 10) fastening the inverter to the holder.
Replacement Procedure
Step 1. Fasten the new A52 inverter board to the holder.
452 Chapter 6
6. Replacement Procedure
Replacement Procedure
A52 Inverter Board Replacement for E5070B S/N prefix JP1KK, MY4231100
and below, E5071B S/N prefix JP1KK, MY4232454 and below
Figure 6-40 A52 Inverter Board Removal
Chapter 6 453
Replacement Procedure
A52 Inverter Board Replacement for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Tools Required
• TORX screwdriver, T8, T10, T15, and T20
• Flat edge screwdriver
Removal Procedure
Refer to Figure 6-41 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B S/N
MY42401101 and above, E5071B S/N MY42402455 and above” on page 388.
Step 2. ase the clamped cables from the clamps (item 1) on the cover.
Step 3. Disconnect the cables (item 2 and 3) from the touch screen controller if the option 016 is
installed.
Step 4. Remove the four TORX T10 screws (item 4) fastening the cover to the front panel.
NOTE The item 5 is not connected in the E5070B/E5071B without the option 016.
Step 6. Disconnect the two cables (item 6 and 7) from the inverter.
Step 7. Remove the two TORX T15 screws (item 8) fastening the inverter assembly.
Step 8. Remove the two TORX T8 screws (item 9) fastening the inverter to the holder.
Replacement Procedure
Step 1. Fasten the new A52 inverter board to the holder.
454 Chapter 6
6. Replacement Procedure
Replacement Procedure
A52 Inverter Board Replacement for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Figure 6-41 A52 Inverter Board Removal
Chapter 6 455
Replacement Procedure
A51 LCD Display Replacement for E5070B S/N prefix JP1KK, MY42300632
thru MY4231100, E5071B S/N prefix JP1KK, MY42301397 thru MY4232454
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
Removal Procedure
Refer to Figure 6-42 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Release the clamped cables from the clamps (item 1) on the cover.
Step 3. Disconnect the cables (item 2 and 3) from the touch screen controller if the option 016 is
installed.
Step 4. Remove the four TORX T10 screws (item 4) fastening the cover to the front panel.
Step 5. Disconnect the two cables (item 5 and 6) connected through the cover.
NOTE The item 6 is not connected in the E5070B/E5071B without the option 016.
Step 6. Release the cable from the clamps (item 7) on the LCD.
Step 8. Remove the four TORX T15 screws (item 9) fastening the A51 LCD display to the front
panel.
Replacement Procedure
Step 1. Replace the A51 LCD display by inverse procedure of removal.
NOTE Fasten the TORX T15 screws (item 9) using a torque screwdriver ( set to 1.02 N-m / 9.0
lb-in).
Step 2. When you have replaced the LCD on a E5070B/E5071B equipped with an Option 016
touch screen, you have to calibrate the touch screen. Follow the procedure described on
“Calibration of the Touch Screen” on page 391.
456 Chapter 6
6. Replacement Procedure
Replacement Procedure
A51 LCD Display Replacement for E5070B S/N prefix JP1KK, MY42300632
thru MY4231100, E5071B S/N prefix JP1KK, MY42301397 thru MY4232454
Chapter 6 457
Replacement Procedure
A51 LCD Display Replacement for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
Removal Procedure
Refer to Figure 6-43 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B S/N
MY42401101 and above, E5071B S/N MY42402455 and above” on page 388.
Step 2. Release the clamped cables from the clamps (item 1) on the cover.
Step 3. Disconnect the cables (item 2 and 3) from the touch screen controller if the option 016 is
installed.
Step 4. Remove the four TORX T10 screws (item 4) fastening the cover to the front panel.
NOTE The item 5 is not connected in the E5070B/E5071B without the option 016.
Step 6. Release the cable from the clamps (item 6) on the LCD.
Step 7. Disconnect the cable (item 7) connected to the front panel keyboard.
Step 9. Remove the four TORX T15 screws (item 9) fastening the A51 LCD display to the front
panel.
Replacement Procedure
Step 1. Replace the A51 LCD display by inverse procedure of removal.
NOTE Fasten the TORX T15 screws (item 9) using a torque screwdriver ( set to 1.02 N-m / 9.0
lb-in).
Step 2. When you have replaced the LCD on a E5070B/E5071B equipped with an Option 016
touch screen, you have to calibrate the touch screen. Follow the procedure described on
“Calibration of the Touch Screen” on page 391.
458 Chapter 6
6. Replacement Procedure
Replacement Procedure
A51 LCD Display Replacement for E5070B S/N MY42401101 and above,
E5071B S/N MY42402455 and above
Chapter 6 459
Replacement Procedure
LCD Backlight Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
Removal Procedure
Refer to Figure 6-44 for this procedure.
Step 1. Remove the A51 LCD display as described in“A51 LCD Display Replacement for E5070B
S/N prefix JP1KK, MY42300632 thru MY4231100, E5071B S/N prefix JP1KK,
MY42301397 thru MY4232454” on page 456.
Step 2. Push the stopper (item 1) and slide the backlight as shown in Figure 6-44.
Replacement Procedure
460 Chapter 6
6. Replacement Procedure
Replacement Procedure
A5 Crystal Oven Board (Opt.1E5) Replacement
Tools Required
• TORX screwdriver, T10, T15 and T20
Removal Procedure
Refer to Figure 6-45 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 382.
Step 2. Remove the two cables (item 1 and 2) connected to the crystal oven board.
Step 3. Remove three TORX T10 screws (item 3) fastening the crystal oven board from the
outside.
Replacement Procedure
Chapter 6 461
Replacement Procedure
N Connector Replacement
N Connector Replacement
Tools Required
• TORX screwdriver, T10, T15, and T20
• Flat edge screwdriver
• Open-end wrench, 9/16
• Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
• File
Removal Procedure
Refer to Figure 6-46 for this procedure.
Step 1. Remove the front panel as described in “Front Panel Removal for E5070B : JP1KKxxxxx,
MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below” on page 384,
or “Front Panel Removal for E5070B : MY42300633 thru MY4231100, E5071B :
MY42301397 thru MY4232454” on page 386.
Step 2. Remove the trim from the right side of the front panel.
Step 3. Remove the TORX T15 screw (item 1) fastening the N-connector holder from the right
side.
Step 4. Remove the two TORX T15 screws (item 2) from the bottom.
Step 5. Disconnect the semi-rigid cables from the N connector. (refer to the figure as shown in
Table 6-6)
NOTE The label on the semi-rigid cable shows the last two digits of the part number.
Option Figure
214 Figure 5-21 on page 328 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-28 on page 341 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
314 Figure 5-23 on page 332 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-29 on page 342 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
462 Chapter 6
6. Replacement Procedure
Replacement Procedure
N Connector Replacement
Option Figure
414 Figure 5-25 on page 336 for E5070B with S/N prefix JP1KK or MY421
, for E5071B with S/N prefix JP1KK or MY421
Figure 5-30 on page 344 for E5070B with S/N prefix MY422 and
above, for E5071B with S/N prefix MY422 and above
Chapter 6 463
Replacement Procedure
N Connector Replacement
Replacement Procedure
Step 1. Fasten the new N connector to the holder.
NOTE Match the dents on the holder and the juts on the connector.
File the hole of the holder smooth if the new N connector cannot be inserted into the hole
easily.
NOTE Fasten the semi rigid cable using a torque wrench for replacement.
464 Chapter 6
7. Post-Repair Procedures
7 Post-Repair Procedures
This chapter lists the procedures required to verify the E5070B/E5071B operation after an
assembly is replaced with a new one.
465
Post-Repair Procedures
Post-Repair Procedures
Post-Repair Procedures
Table 7-1 Post Repair Procedures lists the required procedures that must be performed
after the replacement of an assembly, CPU or DIMM(Memory). These are the
recommended minimum procedures to ensure that the replacement is successfully
completed.
Table 7-1 Post-Repair Procedures
A1 Source Board Perform the following required “To Execute the Internal Test” on page 219
adjustments using “Source Board” in “To Execute the External Test” on page 240
Spot Adjustment of the program.
Frequency Accuracy Test
Frequency Reference Adjustment RF Output Level Accuracy and Flatness Test
Source Power Adjustment RF Output Level Linearity Test
Test Port Characteristics Adjustment Trace Noise CW Test
Crosstalk Test
System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
A3 Level Vernier Perform the following required “To Execute the Internal Test” on page 219
Board adjustments using “Level Vernier” in “To Execute the External Test” on page 240
Spot Adjustment of the program.
RF Output Level Accuracy and Flatness Test
Source Power Adjustment RF Output Level Linearity Test
Test Port Characteristics Adjustment Trace Noise CW Test
Crosstalk Test
System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
A4 Power Amp Perform the following required “To Execute the Internal Test” on page 219
Module adjustments using “Power Amp “To Execute the External Test” on page 240
Module” in Spot Adjustment of the
RF Output Level Accuracy and Flatness Test
program.
RF Output Level Linearity Test
Source Power Adjustment Trace Noise CW Test
Test Port Characteristics Adjustment Crosstalk Test
System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
466 Chapter 7
Post-Repair Procedures
Post-Repair Procedures
A15 Receiver-RF Perform the following required “To Execute the Internal Test” on page 219
Board adjustments using “Receiver (RF) “To Execute the External Test” on page 240
Board” in Spot Adjustment of the
RF Output Level Accuracy and Flatness Test
program.
7. Post-Repair Procedures
RF Output Level Linearity Test
Source Power Adjustment Trace Noise CW Test
Test Port Characteristics Adjustment Crosstalk Test
System Dynamic Range Test
Dynamic Accuracy test
Uncorrected System Performance Test
A17 Receiver-IF Perform the following required “To Execute the Internal Test” on page 219
Board adjustments using “Receiver (IF) “To Execute the External Test” on page 240
Board” in Spot Adjustment of the
Trace Noise CW Test
program.
Crosstalk Test
IF Range Gain/Phase Adjustment System dynamic range Test
Source Power Adjustment Dynamic accuracy test
Test Port Characteristics Adjustment Uncorrected system performance
A5 Crystal Oven Perform the following required Frequency Accuracy Test(with Option 1E5)
(for 1E5) adjustment using “Crystal Oven (for
1E5)” in Spot Adjustment of the
program.
Oven Reference Adjustment (Opt. 1E5)
A8 Source Swich Perform the following required “To Execute the Internal Test” on page 219
adjustments using “RF SW” in Spot “To Execute the External Test” on page 240
Adjustment of the program.
Trace Noise CW Test
Source Power Adjustment Crosstalk Test
Test Ports Characteristics Adjustment System dynamic range Test
Dynamic accuracy test
Uncorrected system performance
A6 RF Switch / No adjustment needed “To Execute the Internal Test” on page 219
Temperature
Controller Board
A7 Source Step Perform the following required “To Execute the Internal Test” on page 219
Attenuator (Opt. adjustments using “Att (for “To Execute the External Test” on page 240
214/314/414) 214/314/414)” in Spot Adjustment of
RF Output Level Accuracy and Flatness Test
the program.
RF Output Level Linearity Test
Source Power Adjustment Trace Noise CW Test
Test Port Characteristics Adjustment Crosstalk Test
System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
Chapter 7 467
Post-Repair Procedures
Post-Repair Procedures
A21 Analog Perform the following required Inspect the Booting Process
Interface Board adjustments using “Analog Interface” in
“To Check the Front Panel” on page 230
Spot Adjustment of the program.
“To Check the External Trigger Input” on
Writing ID page 233
A22 Front Panel No adjustment needed “To Execute the External Test” on page 240
Keyboard “To Check the Front Panel” on page 230
A23 Handler I/O No adjustment needed “To Execute the External Test” on page 240
Board
A24 GPIB Board No adjustment needed “To Check the GPIB” on page 238
USB (USBTMC) No adjustment needed “To Check the USB (USBTMC) Interface
Interface Card Card” on page 238
A26 Display Board No adjustment needed Inspect the Booting Process
“To Execute the External Test” on page 240
“To Check the Video output” on page 233
A27 Mass Storage Perform “HDD” (Hard Disk Drive) in Inspect the Booting Process
Disk Drive Spot Adjustment of the program.
Calibration of the Touch Screen (for opt.
016)
Removable Hard Perform “HDD” (Hard Disk Drive) in Inspect the Booting Process
Disk Drive Spot Adjustment of the program.
Calibration of the Touch Screen (for opt.
016)
A28 FDD No adjustment needed “To Check the FDD” on page 232
A50 Power Supply No adjustment needed Inspect the Booting Process
Assembly
468 Chapter 7
Post-Repair Procedures
Post-Repair Procedures
7. Post-Repair Procedures
Touch panel (for Calibration of the Touch Screen Inspect the Booting Process
Opt. 016)
“To Execute the External Test” on page 240
“To Check the LCD” on page 231
A52 Inverter Board No adjustment needed Inspect the Booting Process
“To Execute the External Test” on page 240
Chapter 7 469
Post-Repair Procedures
Post-Repair Procedures
470 Chapter 7
A. Manual Changes
A Manual Changes
This appendix contains the information required to adapt this manual to versions or
configurations of the E5070B/E5071B manufactured earlier than the current printing date
of this manual. The information in this manual applies directly to E5070B/E5071B units
with the serial number that is printed on the title page of this manual.
471
Manual Changes
Manual Changes
Manual Changes
To adapt this manual to your E5070B/E5071B, refer to Table A-1 and Table A-2.
Table A-1 Manual Changes by Serial Number
The ten-character serial number is stamped on the serial number plate (Figure A-1) on the
rear panel.
472 Appendix A
B. System Recovery
B System Recovery
This appendix describes how to recover the operating system (Windows 2000) when the
operating system has been damaged.
473
System Recovery
System Recovery
System Recovery
By executing system recovery, you can return the system of the E5070B/E5071B (the
Windows operating system and the firmware) to the factory state (at the time of
purchase*1).
R Factory recovery
Returns the contents of the C drive to the factory state.
R User recovery*2
Returns the contents of the C drive to a user-specified state. To use this function, you
must prepare for recovery in advance. For information on the preparation, see
“Procedure to create the user backup image” on page 481 or on page 484 for
information on the execution, see “Procedure to execute the user recovery function” on
page 487 or on page 490.
R In addition to the Windows operating system and the firmware, the following settings
of the E5070B/E5071B are returned to the factory state or a user specified state.
• Network setting
• GPIB setting
• Printer setting
R The driver for the supported printer installed after purchase is deleted.
R You need to execute initial registration again.
Files you created using the save function (files in the D drive) are not affected, but we
recommend backing them up before executing system recovery for precautionary
purposes. For more information on backup, refer to “Making Backup Files” as described in
E5070B/E5071B User’s Guide.
*1.If the hard disk failed and has been replaced after purchase, the state when the replacement
was performed is recovered.
*2.This function is available when the volume label on the hard disk is CP600 or higher.
474 Appendix B
System Recovery
System Recovery
Step 3. Insert the disk for the system recovery which is furnished with the E5070B/E5071B into
the floppy disk drive of the E5070B/E5071B.
Step 5. When the screen as shown in the figure below appears, press and hold of the
keyboard until this screen disappears.
B. System Recovery
NOTE After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
Appendix B 475
System Recovery
System Recovery
Step 6. The following screen appears. Check that Floppy is selected (highlighted) (if Floppy is not
selected, select it with of the keyboard), and press of the keyboard.
Step 7. The message as shown below appears. Press of the keyboard. If you want to cancel the
factory recovery, press here.
Agilent Technologies System Utilities
Recovery & Backup Options (for the E5070/71B)
NOTE If the above message does not appear, the instrument or the disk for the system recovery is
at fault; contact your local Agilent customer center listed at the end of this manual or
distributor.
476 Appendix B
System Recovery
System Recovery
Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the
factory recovery, press here.
You chose to Restore your system by installing the original factory
installed OS and system software.
WARNING: Press C to Continue only if you are sure that you want
to proceed. The C: Drive will be completely overwritten with no
chance of recovering any data. Use Option 1 to recover the system
from a serious malfunction caused by corrupted or inadvertently
deleted files on the system's primary C: partition.
Step 9. The message as shown below appears. Press of the keyboard to start the factory
recovery. If you want to cancel the factory recovery, press here.
CAUTION! Interrupting this process may leave the system in an
unstable state. Allow the software to complete the
backup and recovery process. This may take up to 20
minutes depending on the system configuration.
CAUTION Never turn off the power during the system recovery because doing so may cause
serious damage to the E5070B/E5071B.
Step 10. The factory recovery will be complete in about 5 minutes. When the factory recovery is
complete, the message as shown below appears. Press , , and of the
keyboard at the same time to restart.
Remove the disk and Press CLT+ALT+DEL to restart your
system.
NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
B. System Recovery
Step 11. After restart, the screen for initial registration appears. Execute initial registration. For
information on the execution procedure, refer to “Initial Registration of E5070B/E5071B”
on page 493.
Step 12. For the E5070B/E5071B equipped with the Option 016 touch screen, execute the
calibration of the touch screen. For information on the execution procedure, refer to
“Calibration of the Touch Screen” on page 496.
Appendix B 477
System Recovery
System Recovery
Step 3. Insert the disk for the system recovery which is furnished with the E5070B/E5071B into
the floppy disk drive of the E5070B/E5071B.
Step 5. When the screen as shown in the figure below appears, press and hold of the keyboard
until this screen disappears.
NOTE After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
478 Appendix B
System Recovery
System Recovery
Step 6. The following screen appears. Select “+Removable Devices” with of the
keyboard, and press of the keyboard.
Step 7. The message as shown below appears. Press of the keyboard. If you want to cancel the
factory recovery, press here.
Agilent Technologies System Utilities
Recovery & Backup Options (for the E5070B/E5071B)
B. System Recovery
3. Recover User Backup Image
4. Exit
_________________________________________________________________
Enter a Choice: _
NOTE If the above message does not appear, the instrument or the disk for the system recovery is
at fault; contact your local Agilent customer center listed at the end of this manual or
distributor.
Appendix B 479
System Recovery
System Recovery
Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the
factory recovery, press here.
You chose to Restore your system by installing the original factory
installed OS and system software.
WARNING: Press C to Continue only if you are sure that you want
to proceed. The C: Drive will be completely overwritten with no
chance of recovering any data. Use Option 1 to recover the system
from a serious malfunction caused by corrupted or inadvertently
deleted files on the system's primary C: partition.
Step 9. The message as shown below appears. Press of the keyboard to start the factory
recovery. If you want to cancel the factory recovery, press here.
CAUTION! Interrupting this process may leave the system in an
unstable state. Allow the software to complete the
backup and recovery process. This may take up to 20
minutes depending on the system configuration.
CAUTION Never turn off the power during the system recovery because doing so may cause
serious damage to the E5070B/E5071B.
Step 10. The factory recovery will be complete in about 5 minutes. When the factory recovery is
complete, the message as shown below appears. Press , , and of the
keyboard at the same time to restart.
Remove the disk and Press CLT+ALT+DEL to restart your
system.
NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
Step 11. After restart, the screen for initial registration appears. Execute initial registration. For
information on the execution procedure, refer to “Initial Registration of E5070B/E5071B”
on page 493.
Step 12. For the E5070B/E5071B equipped with the Option 016 touch screen, execute the
calibration of the touch screen. For information on the execution procedure, refer to
“Calibration of the Touch Screen” on page 496.
480 Appendix B
System Recovery
System Recovery
NOTE This function is available when the volume label on the hard disk is CP600 or higher.
Step 3. Insert the disk for the system recovery into the floppy disk drive of the E5070B/E5071B.
Step 5. When the screen as shown in the figure below appears, press and hold of the
keyboard until this screen disappears.
B. System Recovery
NOTE After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
Appendix B 481
System Recovery
System Recovery
Step 6. The following screen appears. Check that Floppy is selected (highlighted) (if Floppy is not
selected, select it with of the keyboard), and press of the keyboard.
Step 7. The message as shown below appears. Press of the keyboard. If you want to cancel the
create user backup image, press here.
Agilent Technologies System Utilities
Recovery & Backup Options (for the E5070/71B)
NOTE If the above message does not appear, the instrument or the disk for the system recovery is
at fault; contact your local Agilent customer center listed at the end of this manual or
distributor.
482 Appendix B
System Recovery
System Recovery
Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the
create user backup image, press here.
You chose to create a backup image file of your system.
CAUTION Never turn off the power during the create user backup image because doing so may
cause serious damage to the E5070B/E5071B.
Step 10. The create user backup image will be complete in about 5 minutes. When the create user
backup image is complete, the message as shown below appears. Press , , and
of the keyboard at the same time to restart.
Remove the disk and Press CLT+ALT+DEL to restart your
system.
NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
B. System Recovery
Appendix B 483
System Recovery
System Recovery
NOTE This function is available when the volume label on the hard disk is CP600 or higher.
Step 3. Insert the disk for the system recovery into the floppy disk drive of the E5070B/E5071B.
Step 5. When the screen as shown in the figure below appears, press and hold of the keyboard
until this screen disappears.
NOTE After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
484 Appendix B
System Recovery
System Recovery
Step 6. The following screen appears. Select “+Removable Devices” with of the
keyboard, and press of the keyboard.
Step 7. The message as shown below appears. Press of the keyboard. If you want to cancel the
create user backup image, press here.
Agilent Technologies System Utilities
Recovery & Backup Options (for the E5070/71B)
B. System Recovery
3. Recover User Backup Image
4. Exit
_________________________________________________________________
Enter a Choice: _
NOTE If the above message does not appear, the instrument or the disk for the system recovery is
at fault; contact your local Agilent customer center listed at the end of this manual or
distributor.
Appendix B 485
System Recovery
System Recovery
Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the
create user backup image, press here.
You chose to create a backup image file of your system.
CAUTION Never turn off the power during the create user backup image because doing so may
cause serious damage to the E5070B/E5071B.
Step 10. The create user backup image will be complete in about 5 minutes. When the create user
backup image is complete, the message as shown below appears. Press , , and
of the keyboard at the same time to restart.
Remove the disk and Press CLT+ALT+DEL to restart your
system.
NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
486 Appendix B
System Recovery
System Recovery
NOTE This function is available when the volume label on the hard disk is CP600 or higher.
Step 3. Insert the disk for the system recovery into the floppy disk drive of the E5070B/E5071B.
Step 5. When the screen as shown in the figure below appears, press and hold of the
keyboard until this screen disappears.
B. System Recovery
NOTE After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
Appendix B 487
System Recovery
System Recovery
Step 6. The following screen appears. Check that Floppy is selected (highlighted) (if Floppy is not
selected, select it with of the keyboard), and press of the keyboard.
Step 7. The message as shown below appears. Press of the keyboard. If you want to cancel the
user recovery, press here.
Agilent Technologies System Utilities
Recovery & Backup Options (for the E5070/71B)
NOTE If the above message does not appear, the instrument or the disk for the system recovery is
at fault; contact your local Agilent customer center listed at the end of this manual or
distributor.
488 Appendix B
System Recovery
System Recovery
Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the
user recovery, press here.
You chose to recover your own system backup image file.
WARNING: Press C to Continue only if you are sure that you want
to proceed. The C: partition will be completely overwritten with
no
chance of recovering any data. Use Option 3 to recover the system
from a serious malfunction caused by corrupted or inadvertently
deleted files on the system's primary C: partition.
CAUTION Never turn off the power during the system recovery because doing so may cause
serious damage to the E5070B/E5071B.
Step 10. The user recovery will be complete in about 5 minutes. When the user recovery is
complete, the message as shown below appears. Press , , and of the
keyboard at the same time to restart.
Remove the disk and Press CLT+ALT+DEL to restart your
system.
NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
B. System Recovery
Appendix B 489
System Recovery
System Recovery
NOTE This function is available when the volume label on the hard disk is CP600 or higher.
Step 3. Insert the disk for the system recovery into the floppy disk drive of the E5070B/E5071B.
Step 5. When the screen as shown in the figure below appears, press and hold of the keyboard
until this screen disappears.
NOTE After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
490 Appendix B
System Recovery
System Recovery
Step 6. The following screen appears. Select “+Removable Devices” with of the
keyboard, and press of the keyboard.
Step 7. The message as shown below appears. Press of the keyboard. If you want to cancel the
user recovery, press here.
Agilent Technologies System Utilities
Recovery & Backup Options (for the E5070/71B)
B. System Recovery
1. Recover Factory Backup Image
2. Create User Backup Image
3. Recover User Backup Image
4. Exit
_________________________________________________________________
Enter a Choice: _
NOTE If the above message does not appear, the instrument or the disk for the system recovery is
at fault; contact your local Agilent customer center listed at the end of this manual or
distributor.
Appendix B 491
System Recovery
System Recovery
Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the
user recovery, press here.
You chose to recover your own system backup image file.
WARNING: Press C to Continue only if you are sure that you want
to proceed. The C: partition will be completely overwritten with no
chance of recovering any data. Use Option 3 to recover the system
from a serious malfunction caused by corrupted or inadvertently
deleted files on the system's primary C: partition.
CAUTION Never turn off the power during the system recovery because doing so may cause
serious damage to the E5070B/E5071B.
Step 10. The user recovery will be complete in about 5 minutes. When the user recovery is
complete, the message as shown below appears. Press , , and of the
keyboard at the same time to restart.
Remove the disk and Press CLT+ALT+DEL to restart your
system.
NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
492 Appendix B
System Recovery
Initial Registration of E5070B/E5071B
NOTE You cannot use the front panel keys during the initial registration of the E5070B/E5071B
therefore connect the mouse and the keyboard before turning on the power.
NOTE If you perform the following procedure incorrectly, a message asking you whether to
return to the previous registration screen and perform the registration appears. In this case,
follow the instruction to return to the previous registration screen.
Step 2. The Windows 2000 Professional Setup wizard appears. Click the Next > button (Figure
B-1).
B. System Recovery
Appendix B 493
System Recovery
Initial Registration of E5070B/E5071B
Step 3. In the Windows 2000 Professional Setup dialog box, select the I accept this agreement
box and click the Next >button (Figure B-2). Then, the Windows 2000 operating system is
restarted automatically.
Step 4. The Network Identification wizard appears. Click the Next > button (Figure B-3).
494 Appendix B
System Recovery
Initial Registration of E5070B/E5071B
Step 5. In the Network Identification Wizard dialog box (1/2), select the Windows always
assumes the following user has logged on to this computer box and check that agena
is in the User Name box. If not, type in agena. Finally, click the Next> button (Figure
B-4).
Step 6. In the Network Identification Wizard dialog box (2/2), click the Finish button to finish the
initial registration of the E5070B/E5071B (Figure B-4). Then, the measurement display of
theE5070B/E5071B appears.
B. System Recovery
Appendix B 495
System Recovery
Calibration of the Touch Screen
Step 1. Press .
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
496 Appendix B
C Firmware Update
This appendix describes how to update the E5070B/E5071B firmware. When you want to
C. Firmware Update
497
Firmware Update
Firmware Update for revision A.06.51 or lower
Required Equipment
• E5070B or E5071B (with keyboard and mouse).
• Several blank floppy disks (DOS format).
Step 1. Copy the following files from Agilent Technologies web site to the several floppy disks
using your computer’s FDD.
NOTE You need the same number of floppy disks as that of cab files. Where, “n” indicates the
maximum number of floppy disks.
• :
Step 2. Assign the Label name as follows using the properties of your computer’s Windows
Explorer as shown in Figure C-1.
• :
498 Appendix C
Firmware Update
Firmware Update for revision A.06.51 or lower
2. On the Windows Explorer, click 3 1/2 Floppy [A:] and confirm the files.
3. Right click on 3 1/2 Floppy [A:], then click Properties. The Properties dialog box
appears.
4. Enter the Label name in the Label box of properties dialog box (General) for the disk1
as shown in Figure C-2, and press the OK button.
NOTE If the label name is not assigned, the firmware update for the E5070B/E5071B will not be
done correctly.
• OS (Operating System)
• VBA (Visual Basic for Application)
• Calibration Constant Data
The following is the procedure to update the E5070B/E5071B firmware.
Step 2. Connect the mouse and the keyboard to the E5070B/E5071B rear panel. Then turn the
E5070B/E5071B on.
C. Firmware Update
Step 3. Insert the E5070B/E5071B firmware installation disk1 into the FDD of the
E5070B/E5071B.
Appendix C 499
Firmware Update
Firmware Update for revision A.06.51 or lower
1. Press key.
2. Press Service Menu - Update Firmware on the menu bar, and the Password dialog box
will appear.
NOTE If you do not see “Update Firmware” soft key, execute log-off, and re-login as “agena”.
3. Enter the password e507xa into the Password box, then click OK.
4. Open dialog box apperas. Confirm that E5070.msi is displayed at the File name
column. Press the Open button, and the Windows instoller will appear.
5. After the disk 1 installation is finished, insert the firmware installation disk 2 (2 of n)
into the FDD of the E5070B/E5071B, then click OK.
Step 5. After the disk n installation is finished, the instrument will be restarted and the
measurement view will appear.
Step 6. Press -Firmware Revision. Confirm the firmware revision you have just installed.
500 Appendix C
Firmware Update
Firmware Update for revision A.08.01 and A.08.11
Required Equipment
• E5070B or E5071B (with keyboard and mouse).
• Several blank floppy disks (DOS format).
Step 1. Copy the following files from Agilent Technologies web site to the several floppy disks
using your computer’s FDD.
NOTE You need the same number of floppy disks as that of cab files. Where, “n” indicates the
maximum number of floppy disks.
• :
• OS (Operating System)
• VBA (Visual Basic for Application)
• Calibration Constant Data
The following is the procedure to update the E5070B/E5071B firmware.
Step 2. Connect the mouse and the keyboard to the E5070B/E5071B rear panel. Then turn the
E5070B/E5071B on.
Step 3. Insert the E5070B/E5071B firmware installation disk1 into the FDD of the
E5070B/E5071B.
C. Firmware Update
1. Press key.
2. Press Service Menu - Update Firmware on the menu bar, and the Password dialog box
Appendix C 501
Firmware Update
Firmware Update for revision A.08.01 and A.08.11
will appear.
NOTE If you do not see “Update Firmware” soft key, execute log-off, and re-login as “agena”.
3. Enter the password e507xa into the Password box, then click OK.
4. Open dialog box apperas. Confirm that E5070.msi is displayed at the File name
column. Press the Open button, and the Windows instoller will appear.
5. After the disk 1 installation is finished, insert the firmware installation disk 2 (2 of n)
into the FDD of the E5070B/E5071B, then click OK.
Step 5. After the disk n installation is fininshed, the instrument will be restarted and the
measurement view will appear.
Step 6. Press -Firmware Revision. Confirm the firmware revision you have just installed.
502 Appendix C
Firmware Update
Firmware Update for revision A.09.10 or later
Required Equipment
• E5070B or E5071B (with keyboard and mouse).
• USB mass storage device.
• http:\\www.agilent.com/find/ena_support
• OS (Operating System)
• VBA (Visual Basic for Application)
• Calibration Constant Data
The following is the procedure to update the E5070B/E5071B firmware.
NOTE User should log in as “agena“, and user should have administrator authority when
performing firmware update.
Step 1. Download the latest firmware from the donload site. It is prepared as execusion file
(E507071B_xx.yy.exe).
Step 2. Run the “E507071B_xx.yy.exe” to extract the E5070.msi. Then, save the E5070.msi to
USB mass storage device using PC.
NOTE For E5070B/71B firmware revision 9.10 or later, there is only 1 .msi file and no .cab files.
Step 3. Connect the mouse and the keyboard to the E5070B/E5071B rear panel. Then turn the
E5070B/E5071B on.
Step 4. Connect the USB mass storage device into the front USB port of the E5070B/71B.
Step 5. Copy e5070.msi from the USB mass storage device to the directory under D:\ drive.
1. Press key.
2. Press Service Menu - Update Firmware on the menu bar, and the Password dialog box
will appear.
Appendix C 503
Firmware Update
Firmware Update for revision A.09.10 or later
NOTE If you do not see “Update Firmware” soft key, execute log-off, and re-login as “agena”.
3. When prompted for a password, enter the password e507xa into the Password box,
then click OK.
504 Appendix C
Firmware Update
Firmware Update for revision A.09.10 or later
4. Press Cancel if you are asked to insert a disk into drive A and select the D:\ drive from
Look in.
5. Confirm that E5070.msi is displayed at the File name column. Select it, then press the
Open button and the Windows installer will appear.
Step 7. After installation is fininshed, the instrument will be restarted and the measurement view
will appear.
Step 8. Press -Firmware Revision. Confirm the firmware revision you have just installed.
Step 9. To unplug USB mass storage device, press Ctrl & ESC on the keyboard. The Windows
toolbar should appear. There should be the “Unplug/Eject Hardware“ (a green arrow) icon
in the system tray.
Step 10. Left click on the icon to stop USB mass storage device, or right click on the icon then
“Unplug or eject “ to show “Unplug or Eject Hardware“ screen. C. Firmware Update
Appendix C 505
Firmware Update
Firmware Update for revision A.09.10 or later
506 Appendix C
D. Computer Virus Check
This appendix describes an example of how to check the system of E5070B/E5071B for
computer viruses using computer viruses check software.
507
Computer Virus Check
Computer Virus Check
Required equipment
• E5070B or E5071B (with keyboard and mouse).
• An external PC installed anti-virus software such as “Norton AntiVirus®”.
• A crossed LAN cable (It enables the peer-to-peer communication between the
E5070B/E5071B and the PC.)
2. Press key.
3. Press Misc Setup - Network Setup - Network Configuration. The Network and
Dial-up Connections window will appear.
4. Double-click the Local Area Connection icon (1 in Figure D-1) in the Network and
dial-up connections window to enable the network connection function. The Local
Area Connection Status dialog box appears. Click the Properties button.
508 Appendix D
D. Computer Virus Check
Computer Virus Check
Computer Virus Check
5. The Local Area Connection Properties dialog box (Figure D-3) appears. Select
(highlight) Internet protocol (TCP/IP) (1 in Figure D-3) and then click the Properties
button (2 in Figure D-3).
6. The Internet Protocol (TCP/IP) Properties dialog box (Figure D-4 ) appears. If Obtain
an IP address automatically has been selected, you select Use the following IP
address (1 in Figure D-4) and then enter the IP address 192.168.0.1 (2 in Figure D-4)
and subnet mask 255.255.255.0 (3 in Figure D-4).
If a specific IP address and subnet mask have been assigned, press the OK button.
Appendix D 509
Computer Virus Check
Computer Virus Check
7. In the Internet Protocol (TCP/IP) Properties dialog vbox, click the OK button (6 in
Figure D-4 ).
8. In the Local Area Connection Properties dialog box, click the OK button (3 in Figure
D-3).
9. In the Local Area Connection Status dialog box, click the Close button (3 in Figure
D-2).
10. Click the x button (2 in Figure D-1) in the upper right of the Network and Dial-up
Connection window.
1. Press key.
3. The System Properties dialog box as shown in Figure D-5 will appear. Click the
Properties button (1 in Figure D-5).
510 Appendix D
D. Computer Virus Check
Computer Virus Check
Computer Virus Check
4. The Identification Change dialog box (Figure D-6) appears. Confirm the computer
name in the Computer Name box (1 in Figure D-6).
Step 3. Set for sharing the hard disk (C or D drive) in the E5070B/E5071B.
2. The Windows Explorer opens. Right click on (C:) or USER (D:) and then click
Properties in the File menu.
3. The (C:) or USERS(D:) Properties dialog box (Figure D-7) appears. Select the Sharing
tab.
Appendix D 511
Computer Virus Check
Computer Virus Check
4. Select the Share this folder if (C:) is selected. Click the New Share button (1 in Figure
D-8).
5. The New Share dialog box (Figure D-9) appears. Enter the share name (name used
when accessed from the external PC) in the Share Name box (1 in Figure D-9) and
click the OK button (2 in Figure D-9).
512 Appendix D
D. Computer Virus Check
Computer Virus Check
Computer Virus Check
6. In the (C:) or USERS(D:) Properties dialog box, click the OK button (2 in Figure D-8).
1. Double-click “My Computer” icon on Windows desktop of the external PC. And
double-click Control Panel. The Control Panel window will appear.
2. In the Control Panel window, double-click Network. The Network dialog box shown in
fig will appear. In the Configuration tab, select TCP/IP (display it in revers video) and
press the Properties button.
3. The dialog box appears. If Specify an IP address has been selected, record the IP
address and sbnet mask.
4. If the Obtain an IP address automatically has been selected, select the Speccify an IP
address. Enter the IP address xxx.xxx.xxx.y and subnet mask xxx.xxx.xxx.xxx as
shown in Figure D-10. The letters x represent the IP address and subnet mask of the
E5070B/E5071B. The letter y is different from the IP address of the E5070B/E5071B.
Press the OK button.
Step 5. Connect the drive of E5070B/E5071B to the PC, and check the drive using anti-virus
software.
1. Click Programs -Window Explorer on the Start Menu to start up the PC’s Explorer.
Appendix D 513
Computer Virus Check
Computer Virus Check
2. Select a suitable drive name, enter the network path as shown in Figure D-11, and then
press the OK button. For example, the network path is \\C_NAME\S_NAME.
C_NAME in the network path is the computer name of the E5070B/E5071B you
confirmed on 4 of Step2.
S_NAME in the share name of the C or D drive of the E5070B/E5071B you enterd on 5
of Step 3.
3. The dialog box to enter the user name and the password appears. Enter the user name
agena and then click the OK button.
4. The network path will appear on the PC’s Explorer. Confirm that the network path you
enter is connected to the PC.
5. Execute the anti-virus software on the PC, and the menu for the anti-virus software will
appear. Check the drive you entered on 2 of Step 5. For the operation of anti-virus
software, refer to its manual.
1. Click Programs -Window Explorer on the Start Menu to start up the PC’s Explorer.
Click Tools - Disconnect Network Drive on the menu of the Explorer. Disconnect
Network dialog box will appear.
2. Select Network Drive you checked, and press the OK button. The drive will disappear
from the Explorer.
If the drive C or D has been infected, replace the A27 mass storage of the E5070B/E5071B.
Step 7. Change the IP address and subnet mask of the E5070B/E5071B if you select Use the
following IP address and entered the IP address and subnet mask on the 6 of Step 1.
1. Press key.
3. Double-click the Local Area Connection icon (1 in Figure D-1) in the Network and
Dial-up Connections window. The Local Area Connection Status dialog box (Figure
D-2) appears. Click the Properties button (2 in Figure D-2).
4. The dialog box (Figure D-12) appears. Select Obtain an IP address automatically and
514 Appendix D
D. Computer Virus Check
Computer Virus Check
Computer Virus Check
5. In the Internet Protocol (TCP/IP) Properties dialog vbox, click the OK button (6 in
Figure D-4 ).
6. In the Local Area Connection Properties dialog box, click the OK button (3 in Figure
D-3).
7. The Local Network dialog box will appear, press the Yes button to restart the
instrument.
Step 8. Change the IP address and subnet mask of the external PC.
1. Double-click “My Computer” icon on Windows desktop of the external PC. And
double-click Control Panel. The Control Panel window will appear.
2. In the Control Panel window, double-click Network. The Network dialog box appears.
In the Configuration tab, select TCP/IP (display it in revers video) and press the
Properties button.
3. The dialog box appears. If Specify an IP address has been selected on Step 4, enter
the IP address and sbnet mask you recorded on 3 of Step 4.
4. If the Obtain an IP address automatically has been selected on Step 4, select the
Obtain an IP address automatically.
Step 9. Disable the C drive of the E5070B/E5071B when you checked the C drive.
2. The Windows Explorer opens. Select (highlight) (C:) and then click Properties in the
File menu.
3. The (C:) Properties dialog box (Figure D-7) appears. Select the Sharing tab.
Appendix D 515
Computer Virus Check
Computer Virus Check
516 Appendix D
E. Power Requirement
E Power Requirement
517
Power Requirement
Preparation for Power Supply
Power Requirements
The E5070B/E5071B requires the following power source:
Requirements
Frequency 47 to 63 Hz
Power Cable
In accordance with international safety standards, this instrument is equipped with a
three-wire power cable. When connected to an appropriate ac power outlet, this cable
grounds the instrument frame. The type of power cable shipped with each instrument
depends on country of destination. Refer to Figure E-1 for the part numbers of the power
cables available.
WARNING For protection from electrical shock, the power cable ground must not be defeated.
The power plug must be plugged into an outlet that provides a protective earth
ground connection.
518 Appendix E
Power Requirement
Preparation for Power Supply
E. Power Requirement
Appendix E 519
Power Requirement
Turning the Power ON and OFF
Step 1. If the standby switch ( ) in the lower-left part of the front panel is in the pressed down
This operation turns ON the power, and the E5070B/E5071B starts the self-test.
• Press the standby switch ( ) in the lower-left part of the front panel (now in the
These operations will start the E5070B/E5071B shutdown process (required software and
hardware processes for turning the power off), and the power will turn OFF after a few
seconds.
NOTE Under normal circumstances, always press the standby switch ( ), or send the shutdown
command from the external controller, to actuate the E5070B/E5071B shutdown process.
Never cut off the power supply directly by disconnecting the power cable plug from
the rear panel of the unit.
If the power supply is cut off directly by disconnecting the power cable plug or by
disconnecting the power supply to the AC outlet, the shutdown process will not be carried
out, and there is a risk of damage to the software or hardware of the E5070B/E5071B.
520 Appendix E
F. Messages
F Messages
The E5070B/5071B can display error messages as well as messages that indicate the
internal operating status of the equipment. This appendix explains what these messages
mean by listing them in alphabetical order.
Messages showing the status of the E5070B/5071B are displayed in the lower-left area of
the E5070B/5071B LCD screen. These messages include error messages that occur during
521
Messages
the execution of GPIB commands and others that indicate the internal status of the
equipment.
Error messages are indicated following the character string “[Err]” and can be read out by a
GPIB command. Other kinds of messages are indicated without the “[Err]” character string
and cannot be read out by a GPIB command. This section explains the meaning of each
message and how to resolve the problem it indicates.
522 Appendix F
Messages
Additional standard needed
Error Messages
An error message is displayed against a red background in the instrument message/warning
area in the lower left part of the screen. Pushing a front panel key or executing :DISP:CCL
command clears the error message. Errors caused by the operation of a front panel key
simply appear on the display. They are not stored in the error queue with some exceptions.
An error with a positive error number is one uniquely defined for this instrument. On the
other hand, an error with a negative error number is basically one defined for common
GPIB devices in IEEE488.2
A
20 Additional standard needed
The GPIB command that turns ON the calibration function has been sent before all of the
data measurements needed to calculate the calibration factor have been completed. In
1-port calibration, for example, when measurements completed for OPEN and SHORT
standards but not yet for LOAD standard,
You have tried to turn on the calibration function by calculating 1-port calibration
coefficient using SENS:CORR:COLL:SAVE or SENS:CORR:OFFS:COLL:SAVE.Be
sure to measure all necessary calibration data before sending commands.This error is not
F. Messages
generated by front key operations.
B
-168 Block data not allowed
An block-data element has been received at a position where this instrument does not
accept one.
C
240 Calibration data lost
This error occurs when a file containing the system calibration data is not found or in a
damaged state at time of the startup of this instrument, indicating a failure of this
instrument. Contact an Agilent Technology sales office or the company from which you
bought the instrument.
22 Calibration method not selected
This error occurs when the command for validating the calibration data,
SENS:CORR:COLL:SAVE or SENS:CORR:OFFS:COLL:SAVE is executed before the
command for selecting a calibration type, SENS:CORR:COLL:METH:xxxx or
SENS:CORR:OFFS:COLL:METH:xxxx, is executed.This error is not generated by front
key operations.
-148 Character data not allowed
A character data element (not violating the standard) has been received at a position where
this instrument does not accept one. Double quotes (“) are omitted where it is necessary to
place a parameter in double quotes (“), for example.
Appendix F 523
Messages
Command error
D
-222 Data out of range
A data element (not violating the standard) outside the range defined by this instrument has
been received. This error occurs when an integer-based command for which the parameter
can be rounded exceeds the range of -65536 to +65536 or when a real-number-based
command for which the parameter can be rounded exceeds the range of -9.9e37 to +9.9e37,
for example.
This error occurs also when a numeric value other than a specified one is entered into a
command in which the “port number” and “CalKit number” are specified as parameters
and hence the parameters are not rounded. Such commands are, for example,
CALC:FSIM:BAL:TOP:BBAL:PPOR, SENS:CORR:COLL:ACQ:OPEN,
SENS:CORR:COLL:ECAL:SOLT3, SENS:CORR:COLL:CKIT:ORD:LOAD, etc.
-104 Data type error
The parser has recognized a data element that must not exist. Block data has been sent
instead of numeric value data or character string data that had been expected, for example.
E
32 ECal module not in appropriate RF path
This error occurs when an ECal command, SENS:CORR:COLL:ECAL:SOLTn or
SENS:CORR:OFFS:COLL:ECAL:xxxx, is executed with the port on the ECal module not
connected correctly to the instrument.
-200 Execution error
An error associated with execution has been generated for which this instrument cannot
specify the error message. This code shows that an error associated with execution defined
in 11.5.1.1.5, IEEE488.2 has occurred. This error occurs also when a calibration
measurement is aborted.
-123 Exponent too large
The absolute value of the exponent exceeds 32,000 (see 7.7.2.4.1, IEEE488.2).
524 Appendix F
Messages
Expression data not allowed
F
31 Failed to configure ECal module
This error occurs when the control of the ECal module fails at time of executing an ECal
command, SENS:CORR:COLL:ECAL:SOLTn or SENS:CORR:OFFS:COLL:ECAL
:xxxx. The failure results from the failure to connect the ECal module to the USB port,
failure of the ECal module, etc.
76 Failed to control signal generator
This error occurs when the external signal source fails to respond during measurement,
though it is enabled (SENS:OFFS:LOC:STAT ON and SENS:OFFS:LOC:CONT ON).
Ensure the connection of the external signal source.
102 Failed to copy file
F. Messages
This error occurs when copying a file (MMEM:COPY command) fails.
104 Failed to create directory
This error occurs when creating a directory (MMEM:MDIR command) fails.
103 Failed to delete file
This error occurs when deleting a file (MMEM:DEL command) fails.
100 Failed to read file
This error occurs when a 2-port touchstone file
(CALC:FSIM:SEND:PMC:PORT:USER:FIL command), the formatted data array
(MMEM:LOAD:FDAT command) and limit table (MMEM:STOR:LIM command) for the
active trace on the active channel, segment sweep table (MMEM:LOAD:SEGM
command) for the active channel, a VBA project file (MMEM:LOAD:PROG command),
etc. cannot be read normally.
101 Failed to write file
This error occurs when the formatted data array (MMEM:STOR:FDATcommand) and
limit table (MMEM:STOR:LIM command) for the active trace on the active channel,
segment sweep table (MMEM:STOR:SEGM command) for the active channel, display
image (MMEM:STOR:IMAG command) for the LCD screen, a VBA project file
(MMEM:STOR:PROG command), etc. cannot be written normally.
-257 File name error
A file name error. This message appears when an error exists in the file name and hence a
command is not executed correctly. This error occurs when you try to copy to an unsuitable
file name, for example.
Appendix F 525
Messages
File name not found
G
-105 GET not allowed
A group execution trigger (GET) has been received in the program message (see 7.7,
IEEE488.2).
H
-114 Header suffix out of range
The unit of the header is outside the range. The header is invalid in the unit for numeric
parameters following a SCPI command.
I
-224 Illegal parameter value
The parameter value is not suitable. This error occurs when the CALC:PAR:DEF
command is used to specify an S-parameter that does not exist in the model (S44 in the
case of a 2-port model), for example.
-282 Illegal program name
This error occurs when a nonexistent VBA program name is specified by the
PROG:SEL:NAME command.
526 Appendix F
Messages
Init ignored
F. Messages
not valid.
-121 Invalid character in number
A character that is invalid for the data type subject to syntactic analysis has been received.
For example, a letter is found in a decimal numeric value or a numeric character “9” in
octal data.
500 Invalid equation expression
This error occurs when the invalid equation is specified in the equation of the equation
editor.
501 Invalid equation label
This error occurs when the invalid equation label is specified in the equation of the
equation editor.
A space is not available for the equation label.
-171 Invalid expression
The expression-data element is invalid (see 7.7.7.2, IEEE488.2). Parentheses are not
paired, or illegal characters are used, for example.
-103 Invalid separator
The parser (a syntactic analysis program) had been expecting a delimiter, but a character
that is not a delimiter has been sent.
-151 Invalid string data
Character string data has been expected, but the character string data that appears is invalid
for some reason (see 7.7.5.2, IEEE488.2). The END message is received before the ending
quotation mark character appears, for example.
Appendix F 527
Messages
Invalid suffix
L
53 Log sweep requires 2 octave minimum span
The span of sweep range is not satisfied the requirement for logarithmic sweep. The sweep
type is automatically changed to linear sweep when this error occurs.
For example, this error occurs when, with the start and stop frequency are set 1 MHz and 2
MHz respectively, the sweep type is changed to logarithmic sweep.
Set the stop frequency to more than four times as many as the start frequency. And then
select logarithmic sweep.
M
-109 Missing parameter
The number of parameters is less than that required for the command, or the parameter has
not been entered. For example, the command SENS{1-6}:SWE:POIN requires one more
parameter.
Therefore, when a message “SENS1:SWE:POIN” is sent to a correct program message
“SENS1:SWE:POIN 201” this instrument receives the former message as an invalid one
because all parameters have not been entered. Enter command parameters correctly.
N
70 No GPIB system controller
The GPIB system controller is not set properly.
This error occurs when USB/GPIB interface is not connected or the interface setting is
improper, during the measurement for power calibration data or for that in scalar-mixer
calibration.
It also occurs when USB/GPIB interface is not connected or the interface setting is
improper, though the control of the external signal source is enabled.
Ensure the connection of USB/GPIB interface and its setting.
52 No valid memory trace
This error occurs when you have executed either DISP:WIND:TRAC:MEM ON command
to display memory trace, or any other command to enable data caqlculation using memory
trace (CALC:MATH:FUNC command with other than NORM is specified), though no
valid data exists in memory trace.This error is not generated by front key operations.
-120 Numeric data error
An error resulting from the numeric value data (including numeric value data having no
decimal point representation) has occurred. A numeric value error other than Errors -121
through -129 has occurred.
-128 Numeric data not allowed
528 Appendix F
Messages
Option not installed
An numeric-value-data element (not violating the standard) has been received at a position
where this instrument does not accept one.
O
200 Option not installed
The command received has been ignored because of the mismatch between the contents of
an option for this instrument and the command.
For example, this error occurs when the source attenuator (power range) is set at a value
other than zero (SOUR:POW:ATT command) in a model not having the extended power
output option.
It also occurs when you try to enable the time domain function in a model not having the
time domain option.This holds true for the frequency offset option.
This error is not generated by front key operations.
-225 Out of memory
Insufficient memory is available in this instrument to perform the required operation.
P
-220 Parameter error
F. Messages
When a parameter-related error other than Errors -221 through -229 occurs, that error is
displayed.
-108 Parameter not allowed
The number of parameters exceeds that required for the command.
For instance, when a program message “:SENS1:SWE:TYPE LIN, SEGM” is sent instead
of a correct program message with a command “:SENS1:SWE:TYPE LIN” which requires
a parameter, the instrument receives the message as the number of parameters is invalid.
See the command reference to confirm the required number of parameters.
41 Peak not found
This error occurs when, after specifying a peak and executing the
CALC:MARK:FUNC:EXEC and CALC:FUNC:EXEC commands, the specified peak is
not found in the marker search analysis.
220 Phase lock loop unlocked
This error occurs when the PLL circuit of this instrument becomes unlocked while the
measurement is in progress. The measurement value is not correct. This error may occur
when an external reference out of specification is connected to this instrument. Should an
error occur with an external reference not connected, this instrument is faulty. Contact an
Agilent Technology sales office or the company from which you bought the instrument.
221 Port 1 receiver overload
The input to Test Port 1 exceeds the maximum input level. The measurement value is not
correct. When a DUT is an amplifier or the like, this error may occur, damaging the
receiver in the worst case. Should this error occur with a passive part used as the DUT or
with nothing connected to the test port, this instrument is faulty. Contact an Agilent
Technology sales office or the company from which you bought the instrument.
Appendix F 529
Messages
Port 2 receiver overload
If this error occurs, the stimulus signal output is automatically turned to off.
222 Port 2 receiver overload
The input to Test Port 2 exceeds the maximum input level. The measurement value is not
correct. When a DUT is an amplifier or the like, this error may occur, damaging the
receiver in the worst case. Should this error occur with a passive part used as the DUT or
with nothing connected to the test port, this instrument is faulty. Contact an Agilent
Technology sales office or the company from which you bought the instrument.
If this error occurs, the stimulus signal output is automatically turned to off.
223 Port 3 receiver overload (for Options 313, 314, 413, and 414 only)
The input to Test Port 3 exceeds the maximum input level. The measurement value is not
correct. When a DUT is an amplifier or the like, this error may occur, damaging the
receiver in the worst case. Should this error occur with a passive part used as the DUT or
with nothing connected to the test port, this instrument is faulty. Contact an Agilent
Technology sales office or the company from which you bought the instrument.
If this error occurs, the stimulus signal output is automatically turned to off.
224 Port 4 receiver overload (For Options 413 and 414 only)
The input to Test Port 4 exceeds the maximum input level. The measurement value is not
correct. When a DUT is an amplifier or the like, this error may occur, damaging the
receiver in the worst case. Should this error occur with a passive part used as the DUT or
with nothing connected to the test port, this instrument is faulty. Contact an Agilent
Technology sales office or the company from which you bought the instrument.
If this error occurs, the stimulus signal output is automatically turned to off.
73 Power meter not settled
The measurement value of the power meter is unstable, or it goes outside the tolerance.
This error occurs when the power meter’s measurement value does not become stable
within about 10 seconds during measurement for power calibration data or that for
scalar-mixer calibration data.
This error also occurs when the power calibration value exceeds the set tolerance.
In addition, it occurs when the power meter fails to respond for any reason, such as
power-down for example, during measurement under the same conditions.
Ensure a firm connection of the power sensor.
241 Power on test failed
This error occurs when the power-on test fails, indicating a failure of this instrument.
Contact an Agilent Technology sales office or the company from which you bought the
instrument.
61 Power unleveled
The out power level exceeds available range.
For example, if the level after correction exceeds the power level that can be outputted
when correcting the power level with the power calibration and power slope feature, this
error occurs.
Check that the power level is set correctly, the power calibration has been performed
correctly and the correction value of the power slope is set correctly.
530 Appendix F
Messages
Printer error
Q
-430 Query DEADLOCKED
The state that generates a “DEADLOCKED” Query error (see 6.3.1.7, IEEE488.2). This
F. Messages
error occurs when both input and output buffers have become full, preventing the
instrument from continuing processing, for example.
-400 Query error
A comprehensive query error has occurred showing that this instrument cannot detect a
more detailed error. This code simply shows that a query error defined in 11.5.1.1.7 and
6.3, IEEE488.2 has occurred.
-410 Query INTERRUPTED
The state that generates a “INTERRUPTED” Query error (see 6.3.2.3, IEEE488.1). This
error occurs when data bytes (DAB) or GET are received before the transmission of the
response after a query has not been completed, for example.
-420 Query UNTERMINATED
The state that generates an “UNTERMINATED” Query error (see 6.3.2, IEEE488.2). This
error occurs when this instrument is designated as the talker and an incomplete program
message is received, for example.
-440 Query UNTERMINATED after indefinite response
After a query asking for an indefinite response has been run, another query is received in
the same program message (See 6.5.7.5.7, IEEE488.2).
-350 Queue overflow
The queue contains a specific code in place of the code which caused this error. The code
indicates that the error occurred because of no space available in the queue, but the error is
not recorded.
Appendix F 531
Messages
Recall failed
R
105 Recall failed
This error occurs when reading an instrument status file (State01.sta, etc.)
(MMEM:LOAD:STAT command) fails.
S
106 Save failed
This error occurs when writing an instrument status file (State01.sta, etc.)
(MMEM:STOR:STAT command) fails.
33 Selected parameter not valid for confidence check
This error occurs when the mix mode S parameter has been selected for the S parameter
you want to check, while using the confidence check function for calibration coefficient.
57 SnP request not valid for selected measurement
This error occurs when you try to save data to a Touchstone file but no measurement has
been executed.
This error also occurs when you try to save a Touchstone file with power sweep
measurement specified or with the frequency offset function set to ON.
50 Specified channel hidden
This error occurs when an attempt is made to activate a channel not on display using the
DISP:WIND:ACT command. This error is not generated by front key operations.
23 Specified error term does not exist
The error occurs when the calibration coefficient type of the parameter specified in the
read/write command of the calibration coefficient (SENS:CORR:COEF) is invalid for the
calibration type selected by the calibration type selection command
(SENS:CORR:COEF:METH:xxxx). This error is not generated by front key operations.
21 Specified ports overlapped
This error occurs when a port number is duplicated in a command requiring two or more
port numbers as parameters. Such commands are, for example,
CALC:FSIM:BAL:TOP:SSB:PPOR 1,2,3,3. Specify port setup correctly to avoid
duplication of ports. This error is not generated by front key operations.
51 Specified trace dose not exist
This error occurs when CALC:PAR:SEL command is executed to activate more traces than
specified by CALC:PAR:COUN command.This error is not generated by front key
operations.
26 Standard not selected
This error occurs when you execute the measurement under the condition that the standard
is not selected for the subclass (the standard number is set to 0).
-150 String data error
When a character-string-data element is put to syntactic analysis, an error not
corresponding to one of Error Numbers -151 through -159 occurs.
532 Appendix F
Messages
String data not allowed
F. Messages
-310 System error
One of the errors designated as “system errors” in this instrument has occurred.
T
40 Target value not found
This error occurs when the target is not found during the marker search analysis after
specifying the target and executing the CALC:MARK:FUNC:EXEC and
CALC:FUNC:EXEC commands. This error occurs also when the bandwidth is not found
after executing the bandwidth marker command, CALC:MARK:BWID:DATA?
-124 Too many digits
The number of digits of the argument of the decimal numeric-value-data element exceeds
255 with the preceding 0 removed (see 7.7.2.4.1, IEEE488.2).
-223 Too much data
The block-, expression-, or character-string-type program data that has been received
conforms with the standard. But it exceeds the amount that can be processed under the
condition of the memory or conditions specific to memory-related devices. In this
instrument, this error occurs when the number of characters exceeds 254 in a
character-string parameter.
Appendix F 533
Messages
Transform, Gate not allowed
U
300 Unable to estimate adapter length
Adapter length can not be estimated at zero span. Change the frequency span other than 0.
-113 Undefined header
A command not defined in this instrument, though not illegal in the syntactic structure, has
been received. For example, when a message “:DISP:WIND1:TABL:MEM ON” is sent to
a correct program message “:DISP:WIND1:TRAC1:MEM ON,” the message sent is
received as an undefined command by this instrument. See the command reference and use
correct commands.
This error occurs also when a port not existing on this model is specified in a command
specifying a port number as an index. Such commands are
CALC:FSIM:SEND:DEEM:PORTn:xxxx, CALC:FSIM:SEND:PMC:PORTn:xxxx,
CALC:FSIM:SEND:ZCON:PORTn:Z0:R, and SENS:CORR:EXT:PORTn:TIME; they
include PORTn as a part.
34 User characterization not found in module
This error occurs when the selected user profile is not detected in the ECal memory, while
reading it from the ECal module, written by the user definition ECal.
V
30 Valid Ecal module not found
This error occurs when the number of ports of the ECal module connected is less than the
necessary number of ports. This error occurs, for example, when a 4-port Cal executing
command, SENS:CORR:COLL:ECAL:SOLT4, is executed with a 2-port ECal module
connected. This error is not generated by front key operations.
71 Valid power meter not found
Valid power meter is not connected. (refer to “Power Calibration” as described in
E5070B/E5071B User’s Guide)
This error occurs when the E5070B/5071B and the power meter is not connected with
USB/GPIB interface, the GPIB addresses do not match between the power meter and the
E5070B/5071B’s power meter, or the power meter is turned off, during the measurement
534 Appendix F
Messages
Valid power sensor not found
F. Messages
Appendix F 535
Messages
Auto port extension not allowed
Warning Message
A warning message is displayed in the instrument message/Warning area in the lower left
part of the display against a gray background. Pushing a front panel key or executing
:DISP:CCL command clears the message.
This message simply appears on the display, being not known to a remote environment
such as a GPIB. This message is not displayed when another error (against a red
background) has already been displayed in the instrument message/Warning area.
The warning messages for this instrument are as follows:
Auto port extension not allowed
This message appears when you try to execute the auto port extension when the sweep type
is the power sweep or the frequency offset function is ON.
If this message is displayed, the auto port extension is not available.
Calibration extrapolated
The frequency range that can be calibrated with the ECal module is exceeded. In this case,
the calibration data for the minimum frequency or maximum frequency are used instead of
the data exceeding the allowable frequency range for calibration.
Fixture simulator not allowed
This message appears when enabling the frequency offset, while the fixture simulator
function is turned on.
If this message is displayed, the fixture simulator function is automatically turned off.
Log sweep requires 2 octave minimum span
If you change the sweep range that does not satisfy the necessary condition of the
frequency span (the stop frequency is about 4 times or more of the start frequency) when
the sweep type is set to the log sweep, this message is displayed.
If this message is displayed, the sweep type is automatically set to the linear sweep.
For example, if you attempt to change the log type to the log sweep when the start
frequency is set to 1 MHz and the stop frequency to 2 MHz, this error occurs.
Set the stop frequency to a value of about 4 times or more the start frequency and then set
the low seep.
Peak not found
This warning message is displayed when, with the tracking turned on, the peak specified by
the marker search has not been found by the time the sweep is finished (with the tracking
executed).
Segment table changed
This warning message is displayed when the setting specified segment by segment in the
segment table is automatically changed by a change in the other setting.
For example, this warning message is displayed when, with the power specified segment
by segment in the segment table, the power setting for a segment is adjusted by a change in
the power range setting.
536 Appendix F
Messages
Sweep mode changed to stepped sweep
F. Messages
If this message is displayed, the gating/conversion function is automatically turned off.
User Preset File not found
This warning message is displayed if no user preset file (D:\UserPreset.sta) exists when
executing the SYST:UPR command.
If this message is displayed, a normal preset (SYST:PRES) will be performed.
Appendix F 537
Messages
User Preset File not found
538 Appendix F
REGIONAL SALES AND SUPPORT OFFICES
For more information about Agilent Technologies test and measurement products, applications, services, and
for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one
of the following centers and ask for a test and measurement sales representative. 21/01/2004
Canada:
Test and Measurement Call Center Asia Pacific:
(tel) 1 877 894-4414 (tel) (65) 6375-8100
(fax) 1 888 900-8921 (fax) (65) 6836-0252
Email: [email protected]
China:
(tel) 800 810-0189
(fax) 800 820-2816
Europe:
(tel) (31 20) 547-2323
(fax) (31 20) 547-2390
Japan:
Call Center
(tel) 0120 421-345
(tel) (81) 426 56-7832
(fax) (81) 426 56-7840
Korea:
(tel) (82 2) 2004-5004
(fax) (82 2) 2004-5115
Latin America:
(tel) (305) 269-7500
(fax) (305) 269-7599
Taiwan:
(tel) 0800 047 866
(fax) 0800 286 331