Optimus 50 60 Fco

This Field Change Order (FCO) from Philips Medical Systems outlines an upgrade for the OPTIMUS 50/65/80 RAD/RF generators to Release 3.3, addressing various software instabilities and interferences. The upgrade requires specific tools and parts, which can be ordered free of charge until March 1999, and the procedure includes firmware updates and hardware modifications. It is applicable to a range of serial numbers and includes a notification report for tracking the completion of the action.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
185 views4 pages

Optimus 50 60 Fco

This Field Change Order (FCO) from Philips Medical Systems outlines an upgrade for the OPTIMUS 50/65/80 RAD/RF generators to Release 3.3, addressing various software instabilities and interferences. The upgrade requires specific tools and parts, which can be ordered free of charge until March 1999, and the procedure includes firmware updates and hardware modifications. It is applicable to a range of serial numbers and includes a notification report for tracking the completion of the action.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Philips Medical Systems

FIELD CHANGE ORDER Service

Issued by : DMC Hamburg Reference No. : 00 135 006


Released : G. Kramm signed Date : February 2000
Publication No. : 4512 980 56521 Product Group : 742

This FCO replaces the issue


(98.0) of September 1998

t ion ce
a c an
fo rm
pe r
f or
OPTIMUS 50 / 65 / 80 RAD / RF

APPLIES TO:

All OPTIMUS 50 / 65 / 80 RAD / RF generators


9890 000 02001
9890 000 00871
9890 000 00661
9890 000 00851/2
9890 000 00881
9890 000 00862
all Optimus generators up to and including
serial numbers
970218 - 970220 970223 - 970340
970361 - 970442 970444 - 970446
970447 - 970763
971002 - 971130
972601 - 972632
980007 - 980445
982001 - 982186
982601 - 982700 982998 + 982999

TITLE:

Upgrade of Release 3.x Optimus generators to Release 3.3

00 135 006 (00.0) 1


Copyright  1998 Philips Medical Systems
ALL RIGHTS RESERVED
LIST OF PAGES & DRAWINGS:

1 ... 3 (00.0)

INTRODUCTION:

Symptom : • 00B5 desk lookup


• 00CQ during system turn on
• no APR backups possible via APRMAN or
XRG-Scope
• release of fuse F3 on mA-control PCB
• error combinations
00CD - 08CF
00CX - 08CY
00CF - 08CY
• erros
00CD, 00CY, 00CQ, 08CY
• errors during the switch - on phase
00XI - 00M3

Cause : Interferences and software instabilities can lead to different problems.

Remedy : Release upgrade to release 3.3 including additional interference suppression activities.

MANPOWER / TIME TO COMPLETE:

1 engineer 3.5 hours

TOOLS & TEST EQUIPMENT:

- 3 + 4 mm Allan keys
- 10 mm open-end wrench
- low watt soldering iron with fine tip
- service PC with hardkey and mouse
- flash light
- PLCC extraction tool AMP 822154-1
2422 487 89772
- standard tools

2 (00.0) 00 000 000


Copyright  1998 Philips Medical Systems
ALL RIGHTS RESERVED
MODIFICATION KIT / PARTS REQUIRED:

For all generators mentioned under APPLIES TO, the following parts can be ordered at SSA Hamburg.

4512 104 39893 Upgrade kit OPTIMUS rel. 3.3

comprising:

4512 113 20524 Eprom Human Interface


4512 104 39901 EZX45W_X52 CAN
Termination plug
2022 333 00024 MKP-KO 33N/10% 1000V
4512 104 90531 Cabel extension
SYSTEM CAN EZ X43
0322 042 00015 Wire for N5:23 - N5:26
4512 116 02203 "Service Disk A"
4512 984 08482 SERV.MAN.UPGR.KIT

Note:
The modification kit will be available Free Of Charge till the end March 1999 at the spare part centre in
Hamburg. Please refer to the generator serial number when ordering the modification kit.

PROCEDURE:

• Exchange of firmware Eprom in Human Interface.


• Software Upload Release 3.3 into CU PCB
• Installation of a short circuit in Rotor Control HS unit.
• Installation of CAN extension cable.

Detailed instruction manual supplied with the upgrade kit.

PARTS DISPOSAL:

All parts are to be disposed of in a safe way in accordance with local Safety Regulations.

DOCUMENTATION:

File this FCO and the Upgrade instruction manual in binder: OPTIMUS under Tab 8

00 135 006 (00.0) 3


Copyright  1998 Philips Medical Systems
ALL RIGHTS RESERVED
FCO ACTION NOTIFICATION REPORT
For local SSD use only; do not return to PMG.
TITLE : Upgrade of Release 3.x Optimus generators to Release 3.3

CLASSIFICATION : Action for Performance FCO REF. NO.: 00 135 006

APPLIES TO : All OPTIMUS 50 / 65 / 80 RAD / RF generators

HOSPITAL / ADDRESS :

LOCATION / FW SITE NO.: SALES ORDER NO. / OA NO.:

PRODUCT NUMBER :

UNIT SERIAL NUMBER :

ACTION ON THIS UNIT WAS: (select one) JOB NO. / SERVICE INCIDENT NO.:

Completed per instruction on


DATE

Completed by the factory prior to delivery.

Not completed as this unit is not affected per instruction because: (state reason)

Not completed because customer has unit in storage.


Required parts & instructions received by the customer.

CUSTOMER ACKNOWLEDGEMENT (Required for MANDATORY ACTIONS only).


The REASON and PURPOSE of this modification have been explained to me.

CUSTOMER NAME (PLEASE PRINT) TITLE

CUSTOMER SIGNATURE DATE

BRANCH
REGION / SERVICE UNIT /
DEALER : SERVICE AREA NO.:

SIGNATURE CUSTOMER SUPPORT ENGINEER DATE

SIGNATURE CUSTOMER SUPPORT MANAGER MAIL TO : SSD Customer Support Manager

00 135 006 (00.0) ANR


Copyright  1998 Philips Medical Systems
ALL RIGHTS RESERVED