BEL Moisture Anayzer - M5-Thermo163L
BEL Moisture Anayzer - M5-Thermo163L
M5 Thermo
Revisione software: 1_00
1
10 CARE AND MAINTENANCE ...................................................................................................... 70
11 TECHNICAL FUNCTIONS .......................................................................................................... 71
12 TROUBLESHOOTING .................................................................................................................. 72
13 REPLACING ELECTRICAL FUSES .......................................................................................... 72
14 WARRANTY .................................................................................................................................... 73
15 STORAGE CONDITIONS ............................................................................................................. 73
16 DISPOSAL ........................................................................................................................................ 73
2
1 Warnings and Safety Instructions
CAUTION!
Please read these instructions for installation and use before starting your work
with your new balance. One use for the device other than the one shown in this
manual does not guarantee the safety of the product. Keep the instruction manual
thoroughly.
Intended use
The humidity analyzer iThermo is used for quick and precise determination of the humidity of
the liquid, pasty and solid substances using the thermogravimetric method.
Observe the following guidelines for safe and hassle operation of the analyzer of humidity:
Observe the following distance and the free space around the unit to
prevent an accumulation of heat in the device and overheating of the unit
itself:
- 20 cm around the device
- 1 m above the device
Do not place flammable materials on, under or near the equipment as the
heating element heats the surrounding area.
Remove the samples with caution, as the heating element and the plates
sample holder may still be very hot. To avoid burns when parts of the high
temperature oven, wear gloves or use thermally insulated clamps.
3
2 Installation
Gently remove the device from its packaging, check it for visible damage caused by
transportation and that all accessories shown below are present.
4
2.2 Positioning the device
The device was manufactured so that in normal operating conditions they are achieved
reliable weighing results. Choosing the correct placement of the device is therefore important
to ensure optimal and precise operation.
For choosing the place where you can install the device, you must meet the following criteria:
Adjustable Feet
5
2.3 Start-up
The following describes all the steps required to prepare the device for the first start-up.
6
Wait 30 minutes after the start-up and calibrate the
device prior its leveling. For the calibration procedure,
refer to Chapter 7.6. Perform calibration of the device
each time it is moved to another place.
7
3 Serial interface connections
The device is equipped as standard with a RS 232C interface for connection to a serial printer.
To ensure communication between the device and a printer, you must satisfy the following
conditions:
connect the printer via a suitable cable by matching the location of the printer signals with
those of the device.
properly set the printer transmission speed and the type of printer according to the printer
parameters. See section 7.7
N.B. It’s advisable to use one of the printers listed in chapter 7.7, (Mod. Dpp250 or TLP50) to
ensure proper printing operation. (In the case of use of different printers, you cannot perform
the printing of the graph).
The thermobalance is equipped with a keyboard placed in the left side part to perform the
switching off operations.
4.1 Description
The keyboard has a switching on and off key (1), an operating indicator led (2) and a USB
port.
10 11
12
After inserting the power cord, the device will start automatically without having to press the
power key.
Use the power key to switch off the device and restart the system when the power cord is
permanently attached.
N.B. The phase of switching on and off of the device requires about 20 seconds;
during this period, the display will remain switched off.
When powering, wait until you display the home screen before performing other
operations.
While in the switching off phase, wait until the green LED turns off before removing the
power plug if necessary.
8
4.2 Display area functions
The thermobalance is equipped with a color display with tap screen functionality that allows
tapping on the screen in the active areas to quickly perform the access to the various menus
and functionality of the device.
1
2
1
Area showing the weighing values or data related to humidity during and at the end of
the drying cycle
Display of the measurement unit; tapping the symbol, you can access the
g menu of the measurement units and select the ones desired.
2
Area showing the heating operation status
9
3
Bar to display the time and date, to set the weighing parameters, to select the user, to
access to drying functions and to switch off the device.
By tapping this area, you access the menu of the drying mode functions
By tapping this area, you access the weighting menu settings, the viewing
preferences, the devices and user customization setup
By tapping this area, you access the switching off menu or enable the
standby status of the device
By tapping this area, you access the user menu to make your selection
quickly than desired
4
Bar to display the function keys.
By tapping the keys displayed in this area, the relative function will be executed. The
following are all available keys that may vary depending on the selected applications
10
Bars of the keys available at the end of the drying
cycle.
11
4.2.1 Instructions for use of the numeric and alphanumeric keypad for data entry
The functionality of the alphanumeric keyboard described below is valid for all of the data
insertion operations in the various modes of use of the device.
Mode 1
Alphanumeric keyboard.
Mode 3
Key to confirm and store the entry
12
Numerical keypad.
13
5 Standby, logout, switching off functions
After inserting the power cord into the wall outlet, the device will automatically switch on
taking in the basic weighing screen.
Display off.
Logout.
Switching off.
Display off
Selecting the display off mode, the device
will go in the standby state; the display is
turned off, so that only the bottom right
command to turn on the system will be
enabled.
Tapping the symbol, you will return to the
basic weighing screen.
Logout
Selecting the logout mode, you will
disable the current user
Tapping “Start”, it will activate the user
and in the case of password protected user
it will be asked you to enter it.
Switching off
Selecting the switching off command, the
device will be turned off. To turn it on, you
shall press the key on the left of the device
or remove the power plug and insert it
again.
14
6 Simple weighting
After the power unit is connected, the green LED status indicator on the left side of the device
will lit. This phase will last about 20 seconds and, during all the time, the display will remain
off. After the data loading phase, you can display a screen showing the scope and resolution
of the balance. Then, the device will be positioned in a simple weighing screen.
At first power up, the device is started by activating the default user with the factory settings.
Tap the -0- key to set a new zero point and reset
all tare values.
15
Place the material to be weighed on the balance
and wait until the stability symbol turn on before
detecting the value.
16
7 Customized settings
This section describes all the customization functions of the device. Each function described
below can be set differently for each user.
N.B. Some settings can only be modified by users with administrator permissions.
Setup of users.
Setup.
Menu of calibration.
Preferences.
Info
Service
To enter the settings menu, press the key on the top display bar.
You will display the following screens with all the functions relating to the setup menu.
17
7.1 Selecting the language
Select the "Preferences" function in the settings menu.
18
7.2 Setting the date and time
Select the "Setup" function in the settings menu.
Enter the new values of date and time. You can also select the time for the 24 or 12 hour
format.
N.B. Setting the date and time cannot be customized for each user, but, once defined, it
will be changed for all users.
19
7.3 Creating a change and selecting the user profile
The balance can be set up for individual use on multiple users. You can create up to 10
different profiles of use. When you first turn on the device, the system enables the "Default"
user; this user cannot be deleted or renamed.
Each user can then customize in a complete manner all the device settings, database
included. Once you create your profile, tap the "user area" in the basic screen, and select the
desired user. They will be so loaded all the settings for the selected user. From this point on,
all changes made by the current user will be stored and made available every time the user
will be selected.
Creating an administrator user enables the user so created to access all of the setting
functions of the device and to edit or delete the other users.
The creation of a standard user allows the user to set only access to some setup functions of
the device.
The ban indication displayed when you tap a function not available for the standard user is as
follows:
In order to change the protected functions, you must access with an administrator user.
Also after creating an administrator or standard user, you can assign a password that will be
required every time the user will select and the device will be power on.
The following describes all operations relating to the creation, modification and deletion of a
user.
Select the "Setup Users" function in the settings menu.
New user
Edit user
Rename user
Copy users
Delete users
User password
20
New user
By selecting the "New user" function, you
can define a new user. The maximum
number of users is 10.
You will be prompted to select if the user
you want to create must have
administrative powers or be a standard
user.
Administrator
Standard
21
Edit user
In the menu of the "User edit" function, you
can change the user type ("Standard or
Administrator”) and momentarily unlock the
settings locked in the case of standard
user.
Type
If the user is standard, in order to make the
change of the type of user, you will be
prompted to enter the password of a user
with administrator rights.
Settings
Selecting this function, you will be
prompted to enter the password of a user
with administrator rights.
22
Rename users
Selecting the "Rename users" function, you
can modify the user name selected.
Copy users
By selecting the "Copy user" function, you
can create a new user with the same
setting of an existing user .
They will not be copied any databases and
it will not be activated the password
protection if required. To copy another user
databases, you must export and import
them in the user desired.
Delete users
By selecting the "Delete users" function,
you can permanently delete a user and its
databases.
23
Confirm tapping “Yes” or delete the
operation by tapping “No”.
User password
By selecting the "User Password" function,
you can assign an access password for the
"standard" user or "Administrator".
24
7.3.1 Recovering and removing the password
N.B.
To remove the password, when you will
be asked to enter the new password and
confirm the new password, tap the "OK"
button without typing a value.
CAUTION! If the password lost is the one of an administrator, you will need to contact a
service center to recover it.
25
7.4 Preferences: language, display appearance, beeper, brightness
For each user you can set an appearance of the display, the usage language, the activation or
deactivation of the beeper and the brightness of the display. The following describes the
various options available and their activation mode.
Appearance
You can choose between 6 different color
combinations and select the one you want by
tapping the corresponding key.
Brightness
There are 4 levels of display brightness. Select
the one you want by tapping the corresponding
key.
Beeper
The beeper function allows you to enable or
disable the buzzer. When on, an audible signal is
released each time you tap the live parts of the
screen tap.
26
7.5 Setting the weighting parameters
For proper use of the balance, it is important to make the correct settings of the weighing
parameters according to the environment of use.
This section describes the parameters of filter, stability, auto-calibration and measurement
units of weighing.
Select the "Setup" function in the settings menu, and the "Weighing parameters" functions.
27
Autozero: automatic correction of zero
constantly adjusts the value of zero. These
possible changes may be, for example, due to
dirt that may deposit on the pan. You can
disable this function by selecting the "Off"
mode. The Level 1 is the one with minor
corrections until you get to Level 3E where the
correction is maximum.
28
7.6 Calibrating and setting the calibration modes
The electronic balance takes measurements of the mass using gravity (g). Differences in
geographical regions and altitude will vary gravity acceleration (g).
Therefore, for accurate measurements, the balance must be adjusted to the place of use and
environmental conditions. This adjustment is made by the calibration function.
29
Calibration data: in the calibration data
screen, you can check the date of the last
calibration, the manner in which it was
performed, the value of the weight used and
the correction made since the previous
calibration. Tapping "Print", you can print out
the displayed data.
Ext. default
Selecting the default calibration mode when the
calibration command is given, the required weight
will be the default one set at the factory whose
value will vary depending on the balance model.
Ext. custom
Selecting the “ext. custom" calibration mode, you
can set a value of the custom calibration weight.
30
7.7 Setting peripherals
This section describes the functions and the relative RS232 and USB output serial settings of
the device.
Select the "Setup" function and then the "Peripherals" function in the settings menu.
31
File on Usb-Key: controlled print; tapping
the "Print" key, it will be performed the print
on the file. In order to use this mode you will
need a USB stick to be able to perform file
saving. (If possible, use a USB stick
dedicated just for this purpose).
With the mode "File on Usb_Key activated"
before executing the print command, verify
that the USB stick for saving is inserted in
the USB port on the left side of the device.
Image file
32
Baud Rate: parameter to select the speed of
the serial transmission rate. The speeds to be
selected are the following:
1200 Baud.
2400 Baud.
4800 Baud.
9600 Baud.
19200 Baud.
38400 Baud.
33
Paper progress: this function allows you to
set a number of blank lines to be added at the
end of the receipt printing to allow it comes
out from the printer.
The entry numeric keypad of insertion will be
automatically turned on, enter the desired
number of lines (from 0 through 9) and tap the
"OK" key to confirm the entered number.
34
7.8 Information on the software
This section explains how to display information about the software versions of the device.
7.9 Service
This section explains all service-related functions and the ones related to data backup of the
device.
N.B. It’s advisable to use an empty USB stick dedicated to saving and restoring data.
35
System backup: this command allows
you saving all users and their
configurations stored in the system,
including the associated drying
databases and the saved test results.
36
Insert the stick into the USB port on the left
side of the device and press "OK".
Tap the "Abort" key to cancel the system
reset operation.
Tap the "Reset factory" key, confirm by tapping the "Yes" key or cancel the operation by
tapping the "No" key.
N.B. Pay attention as all the data of all users including their database will be lost. If you
want to save them, make a backup before performing the factory reset.
37
7.9.3 Software update
Through the USB port you can make any software updates provided by the manufacturer.
Here are the steps needed to perform the update.
38
8 Menu of the functions
This section describes all the functions available for the analysis of the humidity content of the
substances and the settings of the parameters relating to the mode of operation of the heater.
All the functions described below can be customized for each user.
So every time you change the user, the parameters will be set automatically depending on the
settings selected by the user
This section described the setting parameters and the acceptance tests related to the heater.
To access the heater settings, tap the “Heater setup” icon.
Standby temperature.
Beeper.
Adjustment data.
Temperature test.
39
Standby temperature: this function
allows you setting a maintaining
temperature value for the heater.
Adjustment data:
you can select this function to display
and print the thermometer calibration
data.
Temperature test:
This function allows you to enter a
temperature value by which you want to
perform a check.
40
After starting the test, wait that the timer set
automatically to 15 minutes is reset. You will
hear a beep, so open the heater and
compare the temperature value displayed
with the one recorded by the thermometer
placed inside
This section describes how to perform the analysis of the humidity content using the "Simple"
method that allows you to perform a quick test by defining only the temperature and the
method of drying end.
To access the function, tap the "Simple" icon in the menu "Functions".
Temperature.
Enter the desired temperature and tap the
“OK" key.
The temperature values that can be set
range from a minimum of 35°C to a
maximum of 160°C) with step of 1°C.
Drying end.
41
Time: setting this method of the end drying,
the cycle will end when the time set will be
elapsed. Also you can end the cycle manually
at any time by tapping the "Stop" key.
Enter the value of time and press the "OK"
button to confirm
42
After closing the heater cover, the cycle will
automatically start and the analysis screen will
appear.
43
8.3 Determination of the humidity content with "Advanced" method
This section describes how to perform the analysis of the humidity content using the
"Advanced" method that allows you to customize all the drying parameters and to define the
initial modes and the temperature profile.
To access the function, tap the "Advanced" icon in the menu "Functions".
Temperature profile
The temperature profile parameter allows
you to set the heating mode for the drying
cycle. Select the mode and the heating
temperature based on the substance to be
analyzed.
For all modes of the heating, the
temperature settable values range from
a minimum of 35°C to a maximum of
160°C.
Standard
Fast
Ramp
Steps
44
Standard: this mode is preset at the
factory and is suitable for most samples.
In this case, after starting, the
temperature is increased with a speed
defined at the factory until reaching the
value set that then is kept fixed until the
end of the measurement.
45
Tap the “Temperature” key to enable the
alphanumeric keypad and enter the value
desired. Then press “Ok”.
46
For the entry of the temperature and time,
the numeric keypad will activate.
Enter the value of the temperature and
time, then press "OK" to confirm the value.
Drying end.
The end drying parameter is the setting that
defines the method by which you want to
determine the end of the humidity
measuring cycle contained in the
substance to be analyzed.
47
Autostop %: setting this method of the
end drying, the cycle will end when the
variation of humidity loss in % is lower
than the value set for the selected time
interval.
It will also be possible to end the cycle
manually at any time by tapping "Stop".
48
Preheating: this function allows you to
preheat the heater before performing a
drying cycle.
49
Unit: this function allows you to select the
parameter with which to display the drying
cycle
Delayed start-up:
the start of the drying cycle can be delayed
with a time interval between 0 and 35
seconds.
Tapping this key, it will activate the numeric
keypad to set the desired time.
50
8.4 Starting a dying cycle in “Advanced” mode
After making the settings described in Section 9.3, tap the "Start" key to start the drying cycle
and to perform the drying cycle with the set parameters.
Step 1
Heater switching on
51
“Automatic start-up”
Step 3
Loading the substance to analyze
52
After reached the minimum value, the bar
changes color and turns green, indicating
that the quantity loaded is within the set
limits. It will then remain in this color range
until it is
exceeded the maximum limit of the
substance set.
53
8.5 Functions available during or at the end of the drying cycle
The following is a description of all functions available during and at the end of the drying
cycle valid for the "Simple" and "Advanced" modes.
Current
temperature of Displaying the
the heater drying
parameter.
Heating mode
Change
54
At the end of drying you will hear a short
beep and display an "End" drying screen
You can tap the unit of measure to change the
display of the drying parameter or tap the
graph to display it in full screen.
55
Print: tapping the “Print” key, you can decide
what to print.
Print all: print of the drying result
including the graph.
Print data: print of the drying result,
data only.
Setup Peripherals: it allows you
accessing the peripherals setup menu
of the device to change its settings.
The function is not active for
"Standard users."
56
8.6 Rapid keys
This function allows you to quickly recall a drying cycle previously stored in the database
To access the function from the menu "Functions", touch the "Rapid keys" icon.
57
8.7 “Preset” and “personal” databases of the drying methods
The “Database” function allows you accessing to the menu of the drying cycles databases;
each user can customize and store its own programs.
To access the function from the menu "Functions", touch the “Database” icon.
58
Selection by name, enter the product
name to be called up.
Description
Preheating
Nominal weight
Unit
Start-up mode
Delayed start
Stability test
Rapid key
59
Rapid key: tapping this function, you will
display a list of keyboard Rapid keys.
60
If the operation is successfully performed
you will display a confirmation screen of the
export database.
61
Print: the print function allows you to print
on paper or on a drying program file by
entering the name or by selecting it from a
list.
62
8.7.2 Custom Database
This database is customized for each user.
Every time you create a new user, the
database will be completely empty, so use
the "New" command to add records with
the desired drying programs.
The following describes all the available
functions:
Selection by name: tapping this
key you can call up a program by
entering the relevant name.
Selection from a list: tapping this
key you can call up a program by
selecting it from a list.
New: tapping this key you can
enter a new drying record.
Change: tapping this key you can
access the program parameters
that can be varied.
Delete: tapping this key you can
delete one or more records
previously stored.
Print: tapping this key you can
select a drying program and print it.
Export (Usb): tapping this key you
can export the database onto a
USD stick.
Import (Usb): tapping this key you
can import a Default database from
a USD stick.
The following describes only the “New” and
“Delete” commands; for all functions, refer
to Section 9.7.1 (Default database).
63
Description: tap this command to insert, if
you want, the description of the product to
analyze.
64
8.8 Database of the results
The results database allows you to call up and view the results of the drying previously made
and saved in the device's internal memory. It's also a command to retrieve and display the last
analysis even if not saved.
To access the function from the menu "Functions", tap the "Results" icon.
65
Selection from a list: tapping this
key you can call up a program by
selecting it from a list.
Change: tapping this key you can
access the program parameters
that can be varied.
Delete: tapping this key you can
delete one or more records
previously stored.
Print: tapping this key you can
select a drying program and print it.
Export (Usb): tapping this key you
can export the database onto a
USD stick.
Import (Usb): tapping this key you
can import a Default database from
a USD stick.
66
8.9 Setting the GLP data
This function allows you to customize and create a database with G.l.p parameters ..
To access the function from the "Functions" menu, tap the "GLP" icon.
Balance ID
Project ID
User ID
67
Import (Usb): tapping this key, you
can import a database set before
from the USB stick.
Balance ID
Project ID
User ID
68
9 Tips for the determination of the humidity content of the
substances
In the case of liquids or samples subjected to fusion, we recommend using glass fabric filters
involving the following advantages:
69
9.3 Amount of substance and duration of the drying cycle.
The composition of the substance affects significantly the duration and the accuracy of the
humidity measurement results.
A small amount of the substance will result in a faster drying cycle, but with less precise
results.
While a greater amount of the substance will need higher drying times but with more reliable
and repeatable results.
Therefore, based on the substance, it’s necessary to determine the right compromise
between the drying time and the accuracy of the results you want to obtain.
70
11 Technical functions
The models to which this manual refers are all intended for internal use.
Maximum operating altitude: 4000m
Pollution degree: 2
Overvoltage category: II
71
12 Troubleshooting
The measurement lasts too Switching off criterion set in a bad way.
long.
The measurement cannot be Sample is not homogenous.
repeated. Drying time is too short.
Too high drying temperature (i.e.
oxygenation of the sample material,
exceeded the boiling temperature of the
sample).
Temperature sensor dirty or broken.
72
14 Warranty
The term of the warranty is 24 months from the date of purchase proven by the product bill
or the packing slip.
The warranty covers all parts that may be defective in origin. It does not cover mechanical
and electronic parts damaged due to improper installation, tampering or misuse.
The warranty does not cover damage caused by shocks, falls of the balance or fall of
objects on the weighing pan.
15 Storage conditions
Keep the packaging of the balance in the event of possible deployment for customer
service; unplug all cables and any accessories to prevent unnecessary damage.
Do not expose the balance unnecessarily to extreme temperatures and humidity, and
avoid violent shocks.
16 Disposal
If the packaging is no longer used, it can be delivered to the local waste disposal
center. The packaging is made from environmentally friendly materials and is a
valuable source of secondary raw materials. Exhausted batteries must not be thrown
away with normal household waste. Dispose them of in local collection boxes. In case
of disposal of the product, contact with the local authorities. Prior to disposal of, remove the
batteries.