User Manual 35234

The document provides an overview of the HH806U/HH806AU Multilogger Thermometer, detailing its specifications, operational modes, and safety information. It supports various thermocouple types and features dual inputs, a USB interface for data upload, and multiple display options. The thermometer operates within specific temperature ranges and includes instructions for its use and maintenance.

Uploaded by

nik98811
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
6 views3 pages

User Manual 35234

The document provides an overview of the HH806U/HH806AU Multilogger Thermometer, detailing its specifications, operational modes, and safety information. It supports various thermocouple types and features dual inputs, a USB interface for data upload, and multiple display options. The thermometer operates within specific temperature ranges and includes instructions for its use and maintenance.

Uploaded by

nik98811
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Input Protection:

INTRODUCTION 24V dc or 24V ac rms maximum input voltage on any


The instrument is a portable digital thermometer that combination of input pins.
OMEGAnet® On-Line Service Internet e-mail measures external thermocouples of type K, J, R, S, T, E, Reading Rate: 2.5 time per second.
HH806U/HH806AU omega.com [email protected] N. The thermocouples types comply with the N.I.S.T. –
MULTILOGGER THERMOMETER ITS 90 standard reference temperature/voltage tables.
The thermometer features a dual thermocouple input, an
ENVIRONMENTAL
Servicing North America: Ambient Operating Ranges:
M4502/0707 adjustable T/C offset and an USB interface for uploading
INSTRUCTION U.S.A: One Omega Drive, Box 4047 0°C to 50°C (32°F to 122°F) <80% R.H.
Stamford, CT 06907-0047 data to a PC using optional software and cable.
SHEET Storage Temperature:
Shop online at: omega.com e-mail: [email protected] ISO 9001 Tel: (203) 359-1660 FAX: (203) 359-7700
For latest product manuals: omegamanual.info -20°C to 60°C (-4°F to 140°F) <70% R.H.
Certified e-mail: [email protected]
Canada: 976 Bergar
SAFETY INFORMATION
Laval (Quebec) H7L 5A1, Canada GENERAL
Tel: (514) 856-6928 FAX: (514) 856-6886
WARNING Display:
e-mail: [email protected] To avoid electrical shock, do not use this instrument There are three display areas on the HH806 series LCD
when working voltages at the measurement surface (liquid crystal display). The Main and Second displays
For immediate technical or application assistance: over 24V AC or DC. are 4 ½ digits with a maximum reading of 19999. These
U.S.A Sales Service: 1-800-826-6342/1-800-TC-OMEGA® are used for displaying the value of T1 or T2. The Third
and Customer Service: 1-800-622-2378/1-800-622-BEST®
Canada: Engineering Service: 1-800-872-9436/1-800-USA-WHEN® WARNING display is for the date, time, or the differential value of
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA To avoid damage or burns, do not make temperature T1 to T2.
Mexico: En Espanol: (001) 203-359-7803 measurement in microwave ovens. Overload: “----.-” or “OL” is display.
e-mail: [email protected] Battery: 1.5V x 4 PCS (SIZE AAA) UM-4 R03.
FAX: (001) 203-359-7807 [email protected] Battery Life: 120 hours typical with carbon zinc battery.
CAUTION
Dimensions: 160mm(H) x 83mm(W) x 38mm(D).
Servicing Europe: Repeated sharp flexing can break the thermocouple
Weight: Approx. 365g including batteries.
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands leads. To prolong lead life, avoid sharp bends in the
Supplied Thermocouple:
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643 leads, especially near the connector.
Toll Free in Benelux: 0800 0993344
1 meter (40”) type K insulated beaded wire thermocouple.
e-mail: [email protected] Maximum insulation temperature is 482°C (900°F).
Czech Frystatska 184, 733 01 Karviná, Czech Republic SPECIFICATIONS Thermocouple accuracy is ±1.1°C or 0.4% of reading
Republic: Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114 (whichever is greater) from 0°C to 1250°C.
Toll Free: 0800-1-66342 e-mail: [email protected] ELECTRICAL Wire Communication Protocol:
Temperature Scale: Celsius or Fahrenheit user-selectable 2400 baud rate. (HH806U)
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Measurement Range: 19200 baud rate. (HH806A)
Toll Free in France: 0800 466 342 K-TYPE (0.1°) -200°C to 1372°C (-328°F to 2501°F)
e-mail: [email protected] J-TYPE (0.1°) -210°C to 1200°C (-346°F to 2192°F) External Connections: 1 2
Germany/ Daimlerstrasse 26, D-75392 Deckenpfronn, Germany T-TYPE (0.1°) -200°C to 400°C (-328°F to 752°F) 1. USB Port
Austria: Tel: + 49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 E-TYPE (0.1°) -210°C to 1000°C (-346°F to 1832°F)
Toll Free in Germany: 0800 639 7678 2. DC power JACK(12V)
R-TYPE (1°) 0°C to 1767°C (32°F to 3212°F)
e-mail: [email protected]
S-TYPE (1°) 0°C to 1767°C ( 32°F to 3212°F)
United One Omega Drive, River Bend Technology Centre
Kingdom: Northbank, Irlam, Manchester
N-TYPE (0.1°) -50°C to 1300°C ( -58°F to 2372°F)
ISO 9002 M44 5BD United Kingdom *Based on the ITS-90 temperature standard.
Certified Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 According to temperature standard ITS-90.
Toll Free in United Kingdom: 0800-488-488 Accuracy:
e-mail: [email protected] K/J/T/E-TYPE
±(0.05% rdg + 0.3°C) on -50°C to 1372°C
±(0.05% rdg + 0.7°C) on -50°C to -210°C
±(0.05% rdg + 0.6°F) on -58°F to 2501°F
It is the policy of OMEGA Engineering, Inc. to comply with all ±(0.05% rdg + 1.4°F) on -58°F to -346°F
worldwide safety and EMC/EMI regulations that apply. OMEGA N-TYPE
is constantly pursuing certification of its products to the
European New Approach Directives. OMEGA will add the CE
±(0.05% rdg + 0.8°C) on -50°C to 0°C
mark to every appropriate device upon certification. ±(0.05% rdg + 0.4°C) on 0°C to 1300°C
The information contained in this document is believed to be ±(0.05% rdg + 1.6°F) on -58°F to 32°F
correct, but OMEGA accepts no liability for any errors it ±(0.05% rdg + 0.8°F) on 32°F to 2372°F
contains, and reserves the right to alter specifications without R/S-TYPE
notice.
±(0.05% rdg + 2°C) on 0°C to 1767°C
WARNING: These products are not designed for use in, and
should not be used for, human applications. ±(0.05% rdg + 4°F) on 32°F to 3212°F
Temperature Coefficient:
0.1 times the applicable accuracy specification per °C
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and
82°F to122°F).
or “ ” to step through the data one at a time. Press the (10) “T1/T2” Button (Main display) function.
OPERATING INSTRUCTIONS larger arrows “ ” or “ ” to step through the data The input selection button [T1/T2] selects the input for (6) “CLR ?” Button
1. OPERATIONAL MODE ten at a time. Pressing the overlay “ESC” button deac- the main display, T1 thermocouple or T2 thermocou- The CLR function clears all the saved and logged data
There are three operation modes-Normal, Shift, and tivates the read data function. ple. Press the T1/T2 button to switch between the two in memory. When the CLR button is pressed, indicator
Setup Mode. (5) “LOG/READ” Button inputs. When meter is turned on, it is set to the display “MEM” is displayed and the “CLR” on upper-right of
NORMAL MODE: The read log function works in conjunction with the that was last in use. LCD will blink. Pressing the “ENTER” button printed
This is the default mode, the operating functions for log function. It is used for reading logged data. The log (11) “T1/T2” Button (Second display) on the overlay in white will clear all saved and logged
the normal mode are printed on the top of each button function can be activated in the shift mode. Press the The input selection button [T1/T2] selects the input for data. Press “ESC” button to exit this function without
in white. LOG/READ button to activate the log read function; the second display, T1 thermocouple or T2 thermo- clearing data.
SHIFT MODE: the word “READ” is displayed on the LCD. Press the couple. Press the T1/T2 button to switch between the (7) “REL” Button (Main display)
The operating functions for the shift mode are printed overlay SECOND button to rotate through following two inputs. When meter is turned on, it is set to the The relative value function can be used for comparing
in gray on the buttons. While in the normal mode, display menus: T1, T2, GRP, and #. T1 and T2: Dis- display that was last in use. the saved reference value with other measurements.
press the SHIFT button to switch to shift mode. At the plays the T1 or T2 saved data. GRP: Displays the cur- (12) “T1-T2/TIME” Button (Third display) Press the “REL” button to store the current measure-
lower-right corner of the display, the word “Shift” is rent group number. #: Displays the current location of The input selection button [T1-T2] selects the system ment as the reference value, and “REL” should be dis-
displayed to indicate shift mode. To switch back to the read pointer within a selected group. The arrow time and date, or the differential between the two ther- played on the right part of the LCD. The next meas-
normal mode, press the SHIFT button again. buttons on the overlay are used for scrolling through mocouples (T1-T2) for the third display. Press the urement will display the current value compared to the
SETUP MODE: the data and groups. Press the smaller arrows “ ” or T1-T2 button to switch the display options. When me- reference value. Press “REL” button again to clear the
Press the set[ ] button in normal mode to switch to “ ” to step through the logged data or groups one at a ter is turned on, it is set to the display that was last in reference value and deactivate the relative value meas-
setup mode, the indicator “SET” is shown on the left time. Press the larger arrows “ ” or “ ” to step use. urement function.
side of the display. To switch back to normal mode, through the data or groups ten at a time. To navigate (8) “APO” Button
press SET[ ] button. the logged data and groups, press the overlay Press the APO button to turn the “Auto power off”
3. SHIFT MODE
SECOND button until GRP appears in the second dis- The following functions can only be used in the shift function on or off. When this function is enabled, the
2. NORMAL MODE play panel. Then select the group using the arrows. mode. indicator “APO” is shown at the upper left part of the
The following functions can only be used in the normal Press the SECOND button again until the “#” sign is LCD. When APO (Auto power off) is enabled, it will
(2) “°C/°F” Button
mode. displayed. The location of the read pointer in the se- automatically turn the thermometer off no button is
Press the °C/°F button to select the temperature scale,
(1) “ ” Power Button lected group will be displayed. Use the arrows to readings can be displayed in Celsius (°C) or Fahrenheit pressed for a period longer than the set time interval
The “ ” button turns the thermometer on or off. scroll through the data. Pressing the overlay “ESC” (°F). When the thermometer is turned on, it is set to (the default time for APO is 5 minutes). Press power
When the meter is in MAX/MIN record mode, the button deactivates the read data function. the temperature scale that was last in use. button to resume operation.
power off function is disabled. (7) HOLD Mode (only Main display) (4) “SAVE” Button (10) “TYPE” Button (Main display)
(2) “[LIMITS]” Button (only Main display) When HOLD mode is selected, the thermometer holds The save function stores the T1, T2 data in a nonvola- Press this button to change the type of thermocouple in
The limits function will alert the user when a meas- the present readings and stops all further measure- tile memory. Press the SAVE button to save the cur- the main display (K/J/T/E/R/S/N). If the inputs of the
urement exceeds a specified limit. To set the limit val- ments. To activate the data hold mode, press the rent data, the word SAVE is displayed to indicate the main and second display are the same, then pressing
ues, refer to limits function in the setup mode. Press HOLD button, and “HOLD” is displayed on the LCD. data has been saved. The built in memory can store up this button will change the thermocouple type for both
the [LIMITS] button to activate the limits function; the Pressing the HOLD button again cancels the function, to 128 data (HH806U)/256 data (HH806AU). The data displays.
word “LIMIT” should be displayed on the LCD. and the instrument will automatically resume meas- can be read using the read function in the normal (11) “TYPE” Button (Second display)
When the value of the main display exceeds the Hi urements. mode. Press this button to change the type of thermocouple in
limit, the word “Hi” will be displayed and the ther- (8) MIN/MAX with Time record Mode (5) “LOG” Button (HH806U) the second display (K/J/T/E/R/S/N). If the inputs of the
mometer will beep in a pulsed tone. If the value of the (only Main display) The data log function continuously records the data main and second display are the same, then pressing
main display is lower then the Lo limit, the word “Lo” The MIN/MAX function records the highest and low- according to a specified time interval. The time inter- this button will change the thermocouple type for both
will be displayed and the thermometer will beep in a est value recorded, and it calculates the average read- val can be set using the interval setup function [INVT] displays.
continuous tone. To exit the limits function, press the ing, and the differences of MAX to MIN. Press in the set up mode. Press the LOG button to activate
[LIMITS] button. MIN/MAX button to enter the MIN/MAX recording the log function; the indicators “LOG” and “MEM” 4. SETUP MODE
(3) “ ” Button mode. The beeper emits a tone when a new minimum will be displayed on the LCD. There are 16 groups that The following functions can only be use in the setup
The backlight function is represented by this button or maximum measurement is recorded. Press the are used for storing the log data, and each group uses mode.
“ ”. Pressing the button will turn on or off the LCD MIN/MAX button again to rotate through the current 64 data slots. If the current log session exceeds 64 data, (2) “[LIMITS]” Button (Hi/Lo limit setting)
backlight. The backlight will turn off automatically af- readings: MAX: The highest measurement recorded. the log function will automatically use the next group Press the LIMITS button to enter the Hi/Lo limit setup
ter. MIN: The lowest measurement recorded. MAX-MIN: to store the following data. A maximum of 1024 data function. The words “LIMIT” and “Hi” will be flash
(4) “SAVE/READ” Button The difference of the highest and the lowest measure- point can be saved in one log session. Press the LOG on the LCD along with the previous value for the Hi
The read data function works in conjunction with the ment. AVG: The average values of the measurements. button again to exit the data log function. limit. Enter the new Hi limit value using the number
save function in the shift mode, it is used for reading This mode works in conjunction with the hold function, “LOG” Button (HH806AU) keys printed in white on the overlay. The resolution of
saved data. The save function can be activated in shift pressing the HOLD button will stop the recording and The data log function continuously records the data Hi/Lo limit setup is 1 count. The “-” button (same
mode. Press the SAVE/READ button to read saved measurements (Previously recorded readings are not according to a specified time interval. The time inter- button as the ESC) can be used to enter negative val-
data; the word “READ” should be displayed on the erased). Press HOLD button again to resume recording val can be set using the interval setup function [INVT] ues. Press the “ENTER” button to confirm the new
LCD. To navigate around the save data table, press the and measurements. To prevent accidental loss of MIN, in the set up mode. Press the LOG button to activate limit. You will then be prompted to enter a new Lo
overlay “SECOND” button until the “#” sign is dis- MAX and AVG data, the MIN/MAX function can only the log function; the indicators “LOG” and “MEM” limit value. Enter the new value and then press enter
played on the second display. The location of the read be cancelled by pressing and holding down the MIN will be displayed on the LCD. There are 16 groups that to finish setup of limits.
pointer within the saved data table will be displayed. MAX key for more then 2 seconds. The automatic are used for storing the log data. A maximum of (3) “[INVT]” Button (Interval time setting)
The arrow buttons on the overlay are used for scrolling power-off feature, and the power, °C/°F, REL, SET, 16,000 data point can be saved in one log session. To setup the time interval for the log function, press
through the saved data. Press the smaller arrows “ ” Hi/Lo Limits, TYPE, T1/T2 buttons are also disabled. Pressing the LOG button again to exit the data log the [INVT] button. The indicator “INV” will blink on
the top-right of the LCD and the previous interval is WARRANTY / DISCLAIMER Where Do I Find Everything I Need
displayed. Press the number buttons printed in white OPERATOR MAINTENANCE
for Process Measurement
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in
on the overlay to change the time interval. Setting is WARNING materials and workmanship for a period of 13 months from date of purchase.

from left to right of the following format (HH:MM:SS). To avoid possible electrical shock, disconnect the
OMEGA’s WARRANTY adds an additional one (1) month grace period to the
normal one (1) year product warranty to cover handling and shipping time. and Control?
Press the overlay “ENTER” button to confirm. To exit thermocouple connectors from the thermometer before This ensures that OMEGA’s customers receive maximum coverage on each
product. OMEGA…Of Course!
this function, press the ESC button. removing the cover. Shop online at omega.com
If the unit malfunctions, it must be returned to the factory for evaluation.
HH: interval Hour (0~23) OMEGA’s Customer Service Department will issue an Authorized Return
MM: interval Minute (0~59) Battery Replacement (AR) number immediately upon phone or written request. Upon examination
by OMEGA, if the unit is found to be defective, it will be repaired or replaced
SS: interval Second (0~59) 1. Power is supplied by 4pcs 1.5V (AAA SIZE) UM-4 at no charge. OMEGA’s WARRANTY does not apply to defects resulting
TEMPERATURE
MAX: 23:59:59 R03. from any action of the purchaser, including but not limited to mishandling, Thermocouple, RTD & Thermistor Probes,
improper interfacing, operation outside of design limits, improper repair, or
MIN: 00:00:01 2. The “ ” appears on the LCD display when replace- unauthorized modification. This WARRANTY is VOID if the unit shows
Connectors, Panels & Assemblies
(8) “APO” Button (Auto power off time setting, ment is needed. To replace battery remove screw from evidence of having been tampered with or shows evidence of having been Wire: Thermocouple, RTD & Thermistor

min. 1 minute) back of meter and lift off the battery cover. damaged as a result of excessive corrosion; or current, heat, moisture or
vibration; improper specification; misapplication; misuse or other operating
Calibrators & Ice Point References

Use this function to change the time for the auto power 3. Remove the battery from battery contacts and replace. conditions outside of OMEGA’s control. Components which wear is not Recorders, Controllers & Process Monitors

off (APO) function. Press the APO button in setup 4. When not in use for long periods remove battery. warranted, include but are not limited to contact points, fuses, and triacs. Infrared Pyrometers

mode, and the indicator “APO” and the current time 5. Do not store in locations with high temperatures, or OMEGA is pleased to offer suggestions on the use of its various
products. However, OMEGA neither assumes responsibility for any
high humidity. PRESSURE, STRAIN AND FORCE
will flash on the LCD. The default time for APO is 5 omissions or errors nor assumes liability for any damages that result
Transducers & Strain Gages
from the use of its products in accordance with information provided
minutes. Press the number button printed in white on by OMEGA, either verbal or written. OMEGA warrants only that the Load Cells & Pressure Gages
the overlay to set the APO time. Press the overlay Cleaning parts manufactured by it will be as specified and free of defects.
Displacement Transducers
“ENTER” button to confirm. To exit this function Periodically wipe the case with a damp cloth and deter- OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF Instrumentation & Accessories
without changing the setting, press the ESC button. gent, do not use abrasives or solvents. TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE
MAX: 19999 minutes ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies FLOW/LEVEL
MIN: 0001 minutes *Software Operation manual is in the Software disk. of purchaser set forth herein are exclusive, and the total liability of Rotameters, Gas Mass Flowmeters & Flow
OMEGA with respect to this order, whether based on contract, warranty,
(9) “[OFS]” Button (Thermocouple offset ad- negligence, indemnification, strict liability or otherwise, shall not Computers
just) exceed the purchase price of the component upon which liability is Air Velocity Indicators
based. In no event shall OMEGA be liable for consequential, incidental
When the main display input is T1 and a thermocouple or special damages.
Turbine/Paddlewheel Systems
is connected, the instrument can adjust the offset of the CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor
Totalizers & Batch Controllers
thermocouple. The same can be done for T2 when on shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used
the main display. In the setup mode, press this button in or with any nuclear installation or activity; or (2) in medical applications or pH/CONDUCTIVITY
used on humans. Should any Product(s) be used in or with any nuclear
to enter the thermocouple Offset Setup Function (OFS) installation or activity, medical application, used on humans, or misused in
pH Electrodes, Testers & Accessories
and the indicator CAL should be displayed on the any way, OMEGA assumes no responsibility as set forth in our basic Benchtop/Laboratory Meters
WARRANTY / DISCLAIMER language, and, additionally, purchaser will Controllers, Calibrators, Simulators & Pumps
top-right of the LCD and the current setting is also indemnify OMEGA and hold OMEGA harmless from any liability or damage
Industrial pH & Conductivity Equipment
shown. Press the number buttons printed in white on whatsoever arising out of the use of the Product(s) in such a manner.

the overlay to change the offset of the thermocouple.


DATA ACQUISITION
The resolution of the setup is 0.1°. Press the “-” button RETURN REQUESTS / INQUIRIES
Data Acquisition & Engineering Software
to set a negative value. Press the overlay “ENTER” Direct all warranty and repair requests / inquiries to the OMEGA Customer
Communications‐Based Acquisition Systems
button to confirm. MAX: ±1999.9 °C/°F. Service Department. BEFORE RETURNING ANY PRODUCT(S) TO
OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) Plug‐in Cards for Apple, IBM & Compatibles
(12) “[TIME]” Button (System time setting) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN
Datalogging Systems
ORDER TO AVOID PROCESSING DELAYS). The assigned AR number
To set the system time, press the [TIME] button in the should then be marked on the outside of the return package and on any Recorders, Printers & Plotters
setup mode. The third display should show the current correspondence.
date and time with the year flashing. Enter the new The purchaser is responsible for shipping charges, freight, insurance and
proper packaging to prevent breakage in transit. HEATERS
value from left to right in the following format Heating Cable
YY:MM:DD and HH:MM:SS. Press the number but- FOR WARRANTY RETURNS, FOR NON-WARRANTY
Cartridge & Strip Heaters
please have the following REPAIRS, consult OMEGA for
tons printed in white on the overlay to set the system information available BEFORE current repair charges. Have the Immersion & Band Heaters
date and time. Press the “ENTER” button to confirm. contacting OMEGA: following information available
Flexible Heaters
1. Purchase Order number under BEFORE contacting OMEGA:
Exit this function by pressing ESC button. which the product was 1. Purchase Order number to cover Laboratory Heaters
PURCHASED, the COST of the repair,
2. Model and serial number of the 2. Model and serial number of the
product under warranty, and product, and ENVIRONMENTAL
3. Repair instructions and/or specific 3. Repair instructions and/or specific MONITORING AND CONTROL
problems relative to the product. problems relative to the product. Metering & Control Instrumentation
OMEGA’s policy is to make running changes, not model changes, whenever Refractometers
an improvement is possible. This affords our customers the latest in tech- Pumps & Tubing
nology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. Air, Soil & Water Monitors
© Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This Industrial Water & Wastewater Treatment
document may not be copied, photocopied, reproduced, translated, or
reduced to any electronic medium or machine-readable form, in whole or in
pH, Conductivity & Dissolved Oxygen
part, without the prior written consent of OMEGA ENGINEERING, INC. Instruments

Made in Taiwan

You might also like