Access 2 IFU
Access 2 IFU
EC REP
Beckman Coulter Eurocenter S.A.
22, rue Juste-Olivier
Case Postale 1044
CH - 1260 Nyon 1, Switzerland
Tel: +41 (0) 22 365 36 11
C02062-AA
May 2017
Printed in U.S.A.
© 2017 Beckman Coulter, Inc.
Beckman Coulter, Inc. • 250 S. Kraemer Blvd. • Brea, CA 92821 • U.S.A.
Publication Notes
This manual, C02062-AA, release date 2017-05, supports:
• Access 2 published system software version 3, including full versions 3.4.2, 3.4.3, and 3.5.0.
• Access 2i published system software version 6, including full versions 6.2.2 and 6.2.3.
Each page of this Access 2 Instructions for Use is identified with its revision. For pages other than the title and
publication pages, revision information is located at the bottom of the page.
All pages of this manual are issued as C02062-AA, release date 2017-05.
Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are
trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other countries.
ProClin™ is a trademark of The Dow Chemical Company ("Dow") or an affiliated company of Dow.
Microsoft, Windows, and the Windows logo are either registered trademarks or trademarks of Microsoft
Corporation in the United States and/or other countries.
All other trademarks are the property of their respective owners.
Access 2 Instructions for Use Table of Contents
Table of Contents
3 Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
• Changing the Wash Buffer Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
• Changing the Liquid Waste Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
• Changing the Substrate Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
• Loading Reaction Vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
• Changing the Reaction Vessel Waste Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
• Loading a Reagent Pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
• Unloading a Reagent Pack. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
• Retrieving Misplaced Reagent Packs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
• Remapping Packs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
• Maintenance Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
• Daily Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
• Weekly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
• System Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
• Maintenance Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
• Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
• Troubleshooting Events. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
• System Check Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
• Instrument Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
• Assay Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
1 System Overview
Intended Use
The Access 2 system is an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative
determination of various analyte concentrations found in human body fluids.
Scope of Manual
The Access 2 Instructions for Use is intended for use after you have become familiar with the Access 2 system.
The Instructions for Use contains short instructions for everyday use and routine maintenance. It also contains
general information about the Access 2 instrument, such as theory of operation, system specifications, safety
labeling, and troubleshooting.
NOTE
Regardless of the frequency with which you perform a procedure, regularly review the complete
procedures, including their cautions for protecting the instrument from damage, and their warnings for
ensuring your personal safety.
Reference Materials
Additional Access 2 system documentation is listed in Appendix B. For more information, contact your
Beckman Coulter representative.
Technical Support
For technical assistance with the Access 2 Immunoassay System:
• In the U.S.A. or Canada, contact Beckman Coulter Technical Support by phone at 1-800-854-3633, or
online at www.beckmancoulter.com. Before using online support the first time, you will need to
register online.
• Outside the U.S.A. and Canada, contact your technical support representative.
NOTE
Be prepared to provide your system ID.
System Description
Processing Functions
Carousel Module Main Pipettor Module Analytical Module
Sample Carousel Precision Pump and Valve Reaction Vessel Loader
Internal Bar Code Reader Pipettor Gantry Incubator Belt
Cup/Tube Detector Ultrasonic Transducer Wash/Read Carousel
Reagent Carousel Primary Probe Luminometer
Fluidic Module
Probe Wash Tower
Vacuum Pump
Vacuum Reservoir
Substrate Pump, Valve, and
Heater
Wash Pump and Valve
Peristaltic Waste Pump
Fluids Tray
Control Functions
Electronics Module
Power Supply
Printed Circuit Boards
Reset Button
Internal Hard Drive
3.5-Inch Disk Drive
Interlock Switch
RV Load Door Alarm
NOTE
Technological advances and
availability may determine the
Instrument Software
specific external computer
components of your
instrument.
Interface Functions For
Workgroup and Server
Peripherals Module
Printer
Touch Screen Monitor
Mouse
Keyboard
Articulated Arm
External Bar Code Reader
External Computer (PC)
0501E.wmf
System Modes
The Access 2 system has four system modes that indicate the operating state of the system. The current mode is
displayed in the upper left corner of each screen. When the system is in the Running mode, the estimated
completion time for the requested function is displayed as a text line above the three system command buttons.
Additional system messages are displayed at various times at this location.
Some instruments are equipped with a status light bar that changes color to indicate the operational status of the
instrument. The color of the light bar matches the color of the system mode area of the user interface.
Status Light
Mode Screen Description
Bar Color
Button Description
Run
Pause
Select to pause the instrument. The system stops pipetting after it finishes
pipetting the current sample. Processing continues on samples already in
progress.
Stop
Select to stop the instrument. The system stops processing and cancels any tests in
progress.
F1 Load Rack /
Get/Edit Rack
F2 Delete Sample
F3 Show/Hide
F1 Load a Rack
Test Menu
F2 Work Pending
F4 Test Results
F3 Test Request
Sample Manager F5 Request QC
F4 Maintenance Request
F1 F6 Request
F6 Get Selected Rack
Calibration
F7 Clear Selected Rack
F7 Clear All
F8 Change Rack ID
Samples
F8 More Options F1 Filter
F2 Review Data
Calibration
F5 Calibrator Setup
F5
F6 Delete
F7 Print
F1 Filter
F2 Find
F3 Rerun Test
F4 Sample Details
F5 Send to LIS
Test Results
F6 Copy to Disk
F2 F1 Report
F7 Print
F2 Completed Samples
F1 Refresh Screen
F2 Configure Screen F2 System Checks
F8 More Options
F3 Delete Data F5 Run Utility Assay
Maintenance Review
F6 Enable/Disable Utility
F6
Assay
F7 Print
F1 System Revisions
F1 Filter F2 Bar Code
F2 Review Chart Reader Setup
and Data F1 System Setup F3 Set Workgroup
Quality Control
F3 Compare Charts Date/Time
F4
F4 Delete Comparison F5 Password Setup
F5 QC Setup F7 Reports Setup
F7 Print
F1 Filter
F2 Edit Units
F3 Ranges
Configure F2 Tests F4 Gray Zone
F8 F5 Special
F7 Print
F8 Update APF
F3 Test Panels
F4 Derived Results
F1 Scheduled Backup
F5 Reflex Tests
F2 Immediate Backup
F6 LIS
F3 Eject Tape
F7 PC Admin
F4 Restore
F8 About Access 2
F5 Auto-Delete Setup
F8 Shut Down PC
0863E.
Interlock Switch
WARNING
Do not defeat the safety interlock switch on the cover.
The Access 2 system is equipped with an interlock switch to protect you from injury. If you open the front panel
of the instrument, the interlock switch stops the movement of the main pipettor. Other mechanical devices will
continue to operate with the front panel open. If you open the front panel while the system is processing
samples, the system may cancel tests.
Safety Statements
The following statements describe general safety concerns.
WARNINGS
• Reagents, calibrators, and controls used with the system may contain small quantities of sodium
azide preservative. Sodium azide preservative may form explosive compounds in metal drain
lines. Refer to National Institute for Occupational Safety and Health Bulletin: Explosive Azide
Hazards (8/18/76).
• The Access 2 instrument has moving parts and uses high voltage in the ultrasonic transducer.
Both present an injury hazard. You should not operate the Access 2 instrument with the covers
open.
CAUTION
Always plug the Access 2 system into a grounded three-conductor outlet. DO NOT bypass the
grounding prong on the plug.
Safety Symbols
Symbol Description
The sharp objects symbol indicates areas of the instrument in which the skin can
be punctured. Do not put your hands in areas marked with this symbol while
the instrument is running.
The moving parts symbol indicates areas of the instrument in which moving parts
can cause injuries. Do not operate the instrument with the covers open.
The electrostatic discharge (ESD) symbol indicates areas of the instrument which
can be damaged by static discharge.
The laser symbol indicates areas of the instrument where laser light is used. Do
not stare into the laser beam.
Symbol Description
WEEE Directive It is important to understand and follow all laws regarding the safe and proper
disposal of electrical instrumentation. The symbol of a crossed-out wheeled bin on
the product is required in accordance with the Waste Electrical and Electronic
Equipment (WEEE) Directive of the European Union. The presence of this
marking on the product indicates that the device:
• was put on the European Market after August 13, 2005.
• is not to be disposed via the municipal waste collection system of any member
state of the European Union.
For products under the requirement of WEEE directive, please contact your dealer
or local Beckman Coulter office for the proper decontamination information and
take-back program, which will facilitate the proper collection, treatment, recovery,
recycling, and safe disposal of the device.
EU Battery Battery Safety and Disposal Instructions - Dispose of all batteries in accordance
Directive with local regulations.
If you have any questions, please contact your local Beckman Coulter
representative for information on the correct disposal or recycle programs for
batteries.
Beckman Coulter products can have two types of batteries.
• User-Replaceable - Consult the product operating manual for battery
replacement instructions.
• Non-User-Replaceable - Please contact your local Beckman Coulter
representative.
Symbol Description
RoHS The following labels and declarations meet the People’s Republic of China
Electronic Industry Standard SJ/T11364-2006 marking requirement for control of
pollution caused by electronic information products.
This logo indicates that this electronic information product contains certain toxic or
hazardous substance or elements, and can be used safely during its environmental
protection use period. The number in the middle of the logo indicates the
environmental protection use period (in years) for the product. The outer circle
indicates that the product can be recycled. The logo also signifies that the product
should be recycled immediately after its environmental protection use period has
expired. The date on the label indicates the date of manufacture.
This logo indicates that the product does not contain any toxic or hazardous
substances or elements. The “e” stands for electrical, electronic and environmental
electronic information products. This logo indicates that this electronic information
product does not contain any toxic or hazardous substances or elements, and is
XXXXXX-XX green and is environmental. The outer circle indicates that the product can be
recycled. The logo also signifies that the product can be recycled after being
discarded, and should not be casually discarded.
Symbol Description
The CSA (Canadian Standards Association) symbol indicates that the Access 2
Immunoassay System meets all U.S. and Canadian requirements for electrical
safety.
C US
A label with the CE mark of conformity signifies that the Access 2 Immunoassay
System complies with applicable EU directives. The instrument Declaration of
Conformity lists the directives with which the Access 2 system complies.
A label with the C-tick mark indicates that the instrument complies with
applicable Australian Communications Authority requirements.
CAUTIONS
• This equipment has been designed and tested to CISPR 11 Class A. In a domestic environment it
could cause radio interference, in which case you may need to take measures to mitigate the
interference.
• Prior to operation of this device, the electromagnetic environment should be evaluated. Do not
use this device in close proximity to sources of strong electromagnetic radiation (for example,
unshielded intentional RF sources) as these could interfere with proper operation.
• If you suspect interference between the Access 2 system and other equipment, you must take
whatever action is required to correct the interference. Beckman Coulter suggests the following
actions:
- Move the equipment so there is a greater distance between the equipment and the
Access 2 system.
- Re-orient the equipment with respect to the Access 2 system.
- Be sure that the equipment is operating from a different power service connector than that of
the Access 2 system.
The handheld bar reader that also operates in a hands-free mode is exempt from an LED safety class, according
to IEC 62471.
Electrical Requirements
The electrical line and any surge suppressors, backup power supplies (UPS units), and line conditioning
transformers you use with the instrument must meet specific requirements.
Electrical Line
The electrical line supplies power to the Access 2 instrument. To avoid damaging the instrument, the electrical
line should meet the following requirements:
NOTE
The Access 2 system may be connected to a line
conditioner. In this case, locate the line
conditioner within 5 ft (1.5 m) of the outlet.
Line protection device Circuit breaker rated:
• 15 A (115–120 V AC line)
• 6 A (220–240 V AC line)
Line voltage fluctuations Not to exceed ±10 V AC per cycle
Electrical Current
Power Consumption Heat Production
Consumption
Surge Suppressors
Beckman Coulter recommends that you do not use a surge suppressor with the Access 2 instrument. The
instrument has built-in protection similar to that provided by a surge suppressor.
Beckman Coulter does recommend that you use a surge suppressor with the external computer, monitor, and
printer. You should connect the surge suppressor to an outlet, not a line conditioner.
Installation
The Access 2 instrument must be installed by a qualified Beckman Coulter technical support representative. Do
not remove the instrument from the shipping crate until a technical support representative is present.
Warranty
The Access 2 Immunoassay System is covered by and subject to the provisions of the warranty included in your
contractual agreement for the system or its reagents.
The customer is responsible for routine preventive maintenance procedures. Repairs arising from the failure to
perform these maintenance procedures at the indicated time intervals will be made at the discretion of
Beckman Coulter, and at the customer’s expense.
During the instrument rebooting process, the instrument system software is reset and a routine brings all devices
to their home, or known, states. This routine is called initialization, and it prepares the system for further
processing.
NOTES
• If you are rebooting both the PC and the instrument, you can reboot them in any order. For simplicity,
the procedures instruct you to reboot the PC first, then the instrument.
• If you are rebooting an Access 2 PC that is a workgroup server, first you must shut down the other PCs
(clients) in the workgroup. Next, restart the server PC. Finally, restart the client PCs.
• If you reboot only the PC, the instrument continues sample processing if it is in the Running mode.
When the PC reestablishes communication with the instrument, the test data is automatically sent to the
PC.
The instrument does not require periodic shutdowns. However, you should shut down the instrument before
moving it or whenever the power will be turned off for an extended period of time (more than 5 days). Before
shutting down the instrument, contact Technical Support to confirm your decision.
Rebooting the PC
To reboot the PC, you shut it down and then restart it. You do not need to reboot the instrument unless you are
directed to do so. Use this procedure to reboot the PC.
NOTE
If you are shutting down a workgroup server PC, you must shut down each of the client PCs in the
workgroup first.
1. Go to the PC Admin screen. To get to this screen from the Main Menu, select Configure F8 to display the
Configure menu, then select PC Admin F7.
4. Press and hold the PC power switch for at least 15 seconds to turn off the power to the PC.
NOTES
• If the server UI is the only UI that is not accessible, shut down each of the client PCs. Then use this
procedure to shut down the server.
• If a client UI is not accessible, use this procedure to shut down that client PC.
• If the keyboard does not respond during the performance of this procedure, press the PC power switch
on the front of the PC to turn the power off. The power switch is identified by the power switch
symbol.
2379A.bmp
Windows® 8e Start Screen
1. Simultaneously press the [Ctrl], [Alt], and 1. If necessary, press the Windows® key [®] on
[Delete] keys, then select Shutdown. the keyboard to display the Start screen.
The Shut Down Windows window is displayed. 2. Move the mouse cursor to the lower-right corner
of the Start screen.
2. Select Restart or Shut down from the list,
depending upon whether you want to A narrow vertical panel of icons is displayed
immediately restart the PC software. along the right side of the screen.
NOTE
If you are restarting multiple Access 2 PCs in a workgroup, restart the workgroup server first, then restart
the client PCs.
1. Press the PC power switch on the front of the PC to turn the power on. The power switch is identified by the
power switch symbol.
I
1054B.eps
PC Power Switch Symbol
2. Wait until the Main Menu screen appears before you continue normal operation.
If rebooting fails, contact Technical Support.
NOTE
If communication between the PC and the instrument is interrupted for an extended time (more than
30 minutes) while the instrument is processing tests, it may take a few minutes for test results to be sent to
the PC after communication is reestablished. Do not use the system until the PC receives all the test results.
Go directly to the Test Results screen and filter the results by completion time. Watch the Result and
Comp. Time columns. When it appears that the test results that were obtained while communication was
interrupted have been transferred to the PC, you can continue normal operation.
If you have any questions, contact Technical Support.
NOTES
• Do not select any buttons or press any keys until rebooting is complete. After you initiate a reboot,
there is a pause of approximately 2 minutes as the software resets. Then the system enters the
Not Ready mode and system initialization begins.
• During system initialization, the system homes mechanical devices and displays a flashing message in
the system mode area. When most system devices complete initialization, the system changes to the
Ready mode.
• The system continues to initialize the remaining devices and displays a flashing message in the system
mode area. When the message disappears, system initialization is complete.
• If the system does not successfully initialize, contact Technical Support.
WARNING
You will come in contact with potentially infectious
materials during this procedure. Handle and dispose of
biohazard materials according to proper laboratory
procedures. Proper hand, eye, and facial protection is
required.
2. Locate the reset button, just to the right of the pipettor gantry.
3. Push and hold in the reset button for one second, then release.
5. Wait until the system is in the Ready mode and no message appears in the blue system mode area before
you continue normal operation.
WARNINGS
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• If you are moving the instrument, make sure that the new location is properly plumbed.
Reagents, calibrators, and controls used with the system may contain small quantities of sodium
azide preservative. Sodium azide preservative may form explosive compounds in metal drain
lines. Refer to National Institute for Occupational Safety and Health Bulletin: Explosive Azide
Hazards (8/18/76).
Use this procedure to shut down the Access 2 system.
1. Run the Special Clean routine. For more information about performing the Special Clean routine, see the
Help system.
4. Locate the power switch on the lower, right side near the back of the instrument. Press the lower part of the
switch to turn the power off (O position).
WARNINGS
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• Wash buffer contains ProClin™ 300 preservative, which may cause sensitization by skin contact.
After contact with skin, wash immediately with soap and water. Wear suitable gloves.
NOTE
You must equilibrate the substrate used in this procedure to room temperature for the time specified in the
reagent instructions for use before you load it onto the instrument. For detailed information, see the
substrate reagent instructions for use.
2. Locate the power switch on the lower right side near the back of the instrument. Press the upper part of the
switch to turn the power on ( | position).
3. Wait until the system is in the Ready mode and no message appears in the blue system mode area before
you continue the procedure.
If initialization fails, review the Event Log and troubleshoot according to any error event with a similar date
and time to the attempted initialization.
(Optional, except following an extended shutdown) Perform the probe volume checks.
(Optional, except following an extended shutdown) Run the Special Clean routine.
(Optional, except following an extended shutdown) Run the System Check routine.
3 Supplies
Changing the Wash Buffer Bottle
You can change the wash buffer bottle at any time. Because of the internal reservoir, the system continues
processing samples even when you remove the wash buffer bottle. If you wait until the Wash Buffer button is
red, the system schedules no new tests because the internal reservoir is almost empty.
WARNING
Wash buffer contains ProClin™ 300 preservative, which may cause sensitization by skin contact.
After contact with skin, wash immediately with soap and water. Wear suitable gloves.
CAUTION
To avoid contaminating the wash buffer, do not touch any part of the dispense cap assembly that
enters the reservoir. Handle the dispense cap assembly only by the screw cap, not the nozzle.
Wash Buffer
Gently invert new
wash buffer bottle
3 or 4 times
Remove dispense
cap assembly from
empty bottle
0588B.wmf
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazardous materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
Liquid Waste
Loosen cap of full
liquid waste bottle
Stopped
Remove bottle cap and
Processing No
insert in clean bottle
samples?
NOTE Yes
The liquid waste
drain kit may be used Release the quick
disconnects from full
in place of the waste bottle
bottle. The drain kit
directs liquid waste
to a floor drain. Connect quick
Because site disconnects to empty
liquid waste bottle
conditions vary,
Beckman Coulter
requires that you
Remove full liquid
consult with waste bottle and place
Technical Support to clean bottle in fluids tray
determine whether WARNING
your system can be
fitted with the kit. Decontaminate Thoroughly flush any decontaminant
contents of full liquid
waste bottle from the liquid waste bottle before you
store the liquid waste bottle. If you
reinstall the liquid waste bottle before
Dispose of liquid waste
and rinse bottle
you flush it, any remaining
thoroughly decontaminants may react with chemicals
pumped into the liquid waste bottle and
the resulting chemical reactions can
Store clean bottle for
future use create gases harmful to you or the
instrument.
0589C.wmf
Supplies
Change Substrate
F5
Change Substrate
Done
F1
Yes
F1
0505B.wmf
NOTES
• You can load only full cartridges of RVs. If you try to load RVs when the instrument has room for only
a partial cartridge, the system displays a message.
• If the system is processing samples and has one row or fewer RVs left when you try to load more, the
system displays a message. You must wait until the instrument stops before you load the new cartridge.
• RVs can fall between the rake and the wall of the incubator if you do not load RVs properly by
selecting Load RVs F4 from the Supplies screen or Supplies Required screen.
• To be sure that all 98 RVs are firmly seated in the cartridge spine, press down on the spine before you
load the cartridge.
RVs
any screen CAUTION
Remove the empty cartridge
spine to prevent damage to the
RV rake.
Supplies
Load RVs
F4
Load RVs
Remove cartridge
cover, load cartridge,
remove empty spine
Done
F1
0504A.wmf
CAUTIONS
• If the system is in the Running mode when you change the RV waste bag, the system may try to
eject an RV when the plastic collar on the waste bag is blocking the ejection chute. This will cause
a jam.
• To prevent damaging the system, DO NOT push an RV that is sticking out from the ejection
chute all the way back into the chute.
Done
F1
0507A.wmf
WARNING
If you are reloading a partially used reagent pack, it must be returned to the same stand-alone system
or Access 2 workgroup from which it was removed. If a partially used reagent pack is loaded on a
different system or workgroup, it will be inventoried as a full pack and inaccurate results may occur.
CAUTION
To prevent damaging the reagent pack, be sure it is properly seated in the reagent carousel.
Supplies
F3
You can also load a
Supplies Required Supplies reagent pack from the
Reagent Inventory screen
proper laboratory
procedures. Proper hand, IRBR**
** Internal Reagent
eye, and facial protection is Bar Code Reader
present and
enabled?
No
required.
Yes
Manually scan
reagent pack
Load another
pack?
No
Yes
Next Done
F2 F1
0502E.wmf
NOTES
• If the system is using a reagent pack to process tests, you cannot unload it (identified by the in-use
[padlock] icon on the Supplies screen).
• When you unload a partially full reagent pack, the system keeps track of the number of tests left in the
pack until you reload it or manually delete it from inventory.
• Immediately unload any reagent packs that are empty or are rejected as a result of reagent pack
monitoring.
Done
F1
Done
F1
0503A.wmf
NOTE
If your system is equipped with an internal reagent bar code reader, you can use the Remap feature to
automatically identify misplaced packs and add them to the reagent inventory.
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
1. Be sure the system is in the Ready mode or Not Ready mode, and then unload all reagent packs from the
reagent carousel.
2. Go to the Mechanics screen. To get to this screen from the Main Menu, select Diagnostics F7 to display
the Diagnostics menu, then
Select Device Diagnostics F4 to display the Device Diagnostics screen, then
Select Mechanics F1.
4. Slide the carousel door to the left to open it, and then open the reagent carousel door.
5. Gently turn the reagent carousel one rotation. If you find any reagent packs remaining on the carousel,
remove them as they come into position at the reagent carousel door.
6. After verifying that all packs are removed, close the reagent carousel door and then slide the carousel door
to the right to close it.
10. When system initialization is complete, reload all of the packs you removed from the reagent carousel.
Remapping Packs
On systems equipped with an internal reagent bar code reader (instruments with serial numbers ≥570000), you
can perform a Remap operation to scan for bar codes in all slots of the reagent carousel to identify loaded packs.
Packs that were originally manually loaded due to an unreadable bar code may not be identified by this remap
and will be assumed to still be loaded.
2. Go to the Supplies screen. To get to this screen from the Main Menu, select Supplies F3.
The system scans all of the onboard reagent packs and updates the inventory.
NOTE
If the Remap F3 button is unavailable, check the bar code reader configuration to ensure that the internal
reagent bar code reader is enabled. For more information, refer to the Access 2 Reference Manual.
Racks are identified by a bar code label. When the instrument scans the rack bar code label, it identifies the type
of sample containers in the rack and determines the appropriate pipetting depth for aspirating sample.
CAUTION
Only load sample containers on a rack with the appropriate ID.
Sample volume
Number of Total sample
required per x replicates + Dead volume = volume required
replicate
0616A.wmf
NOTES
• To find the sample volume required per replicate, see the reagent instructions for use.
• Dead volume is the amount of sample in the bottom of a sample container that is required to be sure
that enough sample is available for the instrument to complete an assay.
WARNING
If you use tubes with separator gel, be sure that the tube contains sufficient sample volume.
Insufficient sample volume may cause the instrument to attempt to aspirate the separator gel, which
can damage the instrument and compromise the integrity of the test results.
Sample Containers
Place all samples to be tested on the Access 2 system in racks before loading them onto the instrument. The
rack ID on each rack identifies what type of sample containers you may use in that rack. Each rack holds up to
ten sample containers. Place all sample containers except insert cups directly in the rack. Place insert cups into
another sample container appropriate for the rack you are using.
WARNINGS
• You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• Air bubbles in samples can affect level sensing by the Access 2 system and compromise the
integrity of the test results. Avoid creating bubbles when transferring samples to a secondary
container. Always inspect samples before loading on the analyzer, and remove or break any
bubbles.
• Use only recommended sample containers and place them in a rack with a rack ID defined for
that type of sample container. Using containers other than those specified for use with a
particular rack ID may damage the system and compromise the integrity of test results.
The following table lists all of the sample containers you can currently use on the Access 2 instrument, the
corresponding rack ID range and the dead volume for each container.
WARNINGS
• You must have a sufficient volume of sample to process the tests you requested. To be sure you
have enough sample, calculate the total volume needed for the sample container using the
Calculating Sufficient Sample Volume procedure in this section. If you do not follow the
provided procedure, sample processing problems may occur without warning.
• Remove caps from all sample containers before loading on the Access 2 system.
12x75
16x100 GEL
• Border: Lime green • Border: Midnight blue
16x100
• Rack IDs: 1300–1399 • Rack IDs: 1500–1599
• Dead Volume: 500 µL • Dead Volume: 4.5 mL
• Sample rack: 13 mm • Sample rack: 16 mm
NOTE: 12x75 mm tubes may also be used NOTE: 16x100 mm tubes may also be used
with Rack IDs 1–99, at a Dead Volume of with Rack IDs 900–999, at a Dead Volume
3.4 mL. of 7.6 mL.
16x75 mm Elevated Tube Beckman Coulter Access 3.0 mL
• Border: Purple 16x75 Sample Container
13x100
• Rack IDs: 1–99 or 400–499 (only 2.0 mL • Border: Blue
400–499 have the icon) • Rack IDs: 600–699
• Dead Volume: 150 µL • Dead Volume: 400 µL
• Sample rack: 13 mm • Sample rack: 13 mm
Beckman Coulter Access Beckman Coulter 0.5 mL Sample
1.0 mL/13 mm Insert Cup in Cup
13x75 mm Tube • Border: Light pink
13x75
Sarstedt
Tube Adapter • Border: Dark brown
Pediatric
• Border: Pink • Rack IDs: 2100–2199
• Rack IDs: 1800–1899 • Dead Volume: 3 mL
4.9 mL
• Dead Volume: 100 µL • Sample rack: 13 mm
• Sample rack: 13 mm
Sarstedt S-Monovette Tube Sarstedt S-Monovette Tube
75x15 mm 5.5 mL 92x15 mm 7.5 mL
Sarstedt
Sarstedt
• Border: Red-brown • Border: Light brown
• Rack IDs: 2200–2299 • Rack IDs: 2300–2399
• Dead Volume: 3.6 mL • Dead Volume: 5 mL
5.5 mL
• Sample rack: 16x75 mm ELEV • Sample rack: 16x100 mm 7.5 mL
• Border: Gray-brown
• Rack IDs: 2400–2499
• Dead Volume: 6 mL
• Sample rack: 16x100 mm
9.0 mL
5 Sample Manager
Sample management is the process of placing patient, maintenance, quality control, or calibration samples in
racks, entering test requests, and loading the racks onto the Access 2 instrument for processing.
Place samples
on racks
WARNING
To minimize sample evaporation, be sure
Main Menu the sample is pipetted into the RVs within
one hour of loading it onto the instrument.
Be sure sample from 3-mL sample
Sample Manager
containers is pipetted into the RVs within
F1
two hours of loading.
Sample Manager
Are
sample IDs
read by internal Test Request
No Enter a rack ID
F3
bar code
reader?
Test Requests
Yes
Test Request
Using LIS? No F3
All racks
No loaded? All racks
loaded?
No
Yes
Yes
Run
Run
0619D.wmf
Main Menu
Sample Manager
F1
Sample Manager
Load a Rack
F1
All racks
No
loaded?
Yes
System downloads
test requests from LIS
Verify downloaded
sample and test
information and make
any necessary changes
Run
0621C.wmf
Main Menu
Sample Manager
F1
Sample Manager
Enter rack ID
or select rack button
from Off Board list
Test Request
F3
Test Requests
Place samples
in racks
Load Rack X
F1
All racks
loaded? No Back
Yes
Run
0618D.wmf
Sample Manager
Test Requests
Request Calibration
F6
Request Calibration
OK
F1
Test Requests
Run
0510C.wmf
Main Menu
NOTE
You must set up each new
quality control lot in the
Sample Manager Access 2 system before you
F1 can select it during test
request entry.
Sample Manager
Test Request
F3
Test Requests
Request QC
F5
Request QC
Test Requests
Load Rack X
F1
Run
0511C.wmf
Processing Samples
Sample processing begins after you verify supplies, load samples, and enter or verify test request information.
Select Run.
(optional) (optional)
Monitor progress Supplies or
Samples Monitor progress
View results No calibration No
required? View results
Print results required?
Print results
Yes Yes
Override
F8
Yes
(optional)
Override Apply filter
Correct the Load samples
No expired supply/ No condition now?
calibration? Delete samples
Print report
Yes
Load supply,
change waste
container, or
request calibration
0518B.wmf
6 Maintenance
Maintenance Overview
For optimal performance, the Access 2 system requires routine maintenance, including Daily Maintenance and
Weekly Maintenance.
A technical support representative will schedule periodic preventive maintenance procedures on your Access 2
instrument in accordance with the terms of your service agreement, if applicable. For more information about
preventive maintenance, contact technical support.
Daily Maintenance
In order to keep the Access 2 system running properly, perform daily maintenance once every 24 hours.
NOTES
• If you use the Access 2 system to process the Vitamin B12 assay, you should also run the Special Clean
routine at the end of every day or whenever you will not use the instrument to process samples for
8 hours or more. For more information about performing the Special Clean routine, see the Help
system.
• Even if you do not use the system every day to run assays, it is still important to perform daily
maintenance on schedule to ensure that the system is ready when needed.
WARNINGS
• You will come in contact with potentially infectious materials during these procedures. Handle
and dispose of biohazard materials according to proper laboratory procedures. Proper hand,
eye, and facial protection is required.
• Citranox* cleaning solution is acidic and may cause eye or skin irritation. See the manufacturer’s
label for details.
• Contrad† 70 cleaning solution is alkaline and may cause severe eye irritation or mild skin
irritation. See the manufacturer’s label for details.
• Wash buffer contains ProClin™ 300 preservative, which may cause sensitization by skin contact.
After contact with skin, wash immediately with soap and water. Wear suitable gloves.
Required Materials
• Fiber-free polyester swabs (or equivalent fiber-free applicators)
• Maintenance Log (see the Maintenance Log in this section)
• 13-mm rack for 2-mL sample cups; the rack must have a rack ID between 1 and 57 or 400 and 456
• Wash buffer (or deionized water)
• Citranox cleaning solution
• Contrad 70 cleaning solution
• 2.0-mL sample cups (three cups)
NOTE
Use only the 2.0-mL sample cups when performing the maintenance routines. Using any other sample
containers may result in level sensing errors and cancellation of the maintenance routine.
CAUTIONS
• Be sure not to bend or damage the fragile probe tips.
• To avoid contamination, use a new applicator on each type of probe.
• Do not wipe the tip of the probe. Fibers on or inside the probes can clog the probes or valves in
the fluidic module.
Main Menu
Check the
Run the Daily
Zone Temperatures*
Clean System*
OK
F1
Check Diagnostics
System Check the Supplies F7 Maintenance Request
Status and Waste Prime Fluidics
Containers* F2
Prime Fluidics
Main Menu
Sample Manager
F1
Check the
Backup* Run
0787C.wmf
1. Check the System Status
WARNINGS
• You will come in contact with potentially infectious materials during these procedures. Handle
and dispose of biohazard materials according to proper laboratory procedures. Proper hand,
eye, and facial protection is required.
• Unless there is an emergency, let the maintenance routine run to completion. Canceling the
routine may damage the instrument and compromise the integrity of subsequent test results.
• If you cancel the routine, do not initialize the system. Contact Technical Support for assistance.
Sample Manager
Load Rack
F1
F1
Sample Manager
Load rack
Enter the Maintenance Rack
Number in the Rack ID field
Maintenance Request
F4 Run
Request Maintenance
The system cleans
probe interiors
Select Daily Clean (15 minutes)
System
OK
F1
Wait until the status bar
indicates the cleaning
is complete
Place 2.0 mL sample
cups in rack positions
1, 2, and 3
Unload rack
Pipette 2.0 mL
Contrad 70 into cup 1
Initial the
Maintenance Log
Pipette 2.0 mL
1/5 Citranox/deionized
water dilution into cup 2
Complete the remaining
Daily Maintenance
Pipette 2.0 mL routines
deionized water
or wash buffer
into cup 3
0596B.wmf
Weekly Maintenance
In order to keep the Access 2 system running properly, perform weekly maintenance once every seven days.
NOTE
If the system is not used to run assays every day, it is still important to perform weekly maintenance on
schedule to ensure that the system is ready when needed.
WARNINGS
• Methanol is extremely flammable. Do not use near heat or flame. Do not ingest. Avoid contact
with eyes, skin, and clothing. Use with adequate ventilation.
• Contrad 70 cleaning solution is alkaline and may cause severe eye irritation or mild skin
irritation. See the manufacturer’s label for details.
• Wash buffer contains ProClin™ 300 preservative, which may cause sensitization by skin contact.
After contact with skin, wash immediately with soap and water. Wear suitable gloves.
Required Materials
• Lint-free cloth
• Deionized or distilled water
• Maintenance Log (see the Maintenance Log in this section)
• Spare liquid waste bottle (as needed)
• Spare waste filter bottle (as needed)
• Alcohol or alcohol swabs or wipes (methanol can be substituted for cleaning the exterior of the aspirate
probes)
• Proper hand, eye, and facial protection for handling biohazardous materials
• Clean aspirate probes (three)
• Contrad 70 cleaning solution
• Beakers (two)
• Aspirate Probe Cleaning Kit (3.0-mL syringe, disposable aspirate probe brush, aspirate probe syringe
fitting assembly consisting of a fitting and tubing)
• Disposable aspirate probe brushes (as needed)
• Absorbent paper
• Wash buffer
WARNING
You will come in contact with potentially infectious materials during this procedure. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
Clean the
Instrument Exterior*
Discoloration
Inspect the Liquid
Cracking or Crazing
Waste Bottle for Wear*
Bulging
Run Daily
Maintenance*
Run System
Check*
0808C.wmf
Install a clean
aspirate probe
Turn the
instrument on
Yes
0776C.wmf
WARNINGS
• You will come in contact with potentially infectious materials during these procedures. Handle
and dispose of biohazard materials according to proper laboratory procedures. Proper hand,
eye, and facial protection is required.
• Contrad 70 cleaning solution is alkaline and may cause severe eye irritation or mild skin
irritation. See the manufacturer’s label for details.
• Once a disposable aspirate probe brush has been used to clean 1 to 3 aspirate probes, it is
considered a biohazard. Handle and dispose of the brush according to appropriate laboratory
safety procedures. Do not save a used brush for future use.
• Take the necessary precautions when removing and reinserting the probe brush to avoid
scattering droplets of biohazardous material into the air.
CAUTIONS
• Handle the aspirate probes with extreme care. The probes are fragile, and will not function
properly if bent.
• The disposable aspirate probe brush handle bends easily.
0818a.wmf
WARNINGS
• You will come in contact with potentially infectious materials during these procedures. Handle
and dispose of biohazard materials according to proper laboratory procedures. Proper hand,
eye, and facial protection is required.
• Contrad 70 cleaning solution is alkaline and may cause severe eye irritation or mild skin
irritation. See the manufacturer’s label for details.
CAUTIONS
• Do not wipe the tip of the probe. Fibers on or inside the probes can clog the probes or valves in
the fluidic module.
• Handle the aspirate probes with extreme care. The probes are fragile, and will not function
properly if bent.
0819a.wmf
WARNING
You will come in contact with potentially infectious materials during these procedures. Handle and
dispose of biohazard materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
CAUTIONS
Draw the distilled water
into the syringe • Do not wipe the tip of the
probe. Fibers on or inside
the probes can clog the
Attach the probe to the probes or valves in the
syringe fluidic module.
• Handle the aspirate probes
with extreme care. The
Empty the syringe
into a sink
probes are fragile, and will
not function properly if
bent.
Rinse the probe twice
by drawing water into
the syringe; empty it
into a sink
Initial the
Maintenance Log
0778a.wmf
System Check
You perform the System Check routine as part of weekly maintenance to verify system performance. During
weekly maintenance, the three System Checks (washed, unwashed, and substrate) are run together. You can
also run the System Check routine more often, or perform individual checks.
WARNINGS
• Unless there is an emergency, let the System Check routine run to completion. Cancelling the
routine may damage the instrument and compromise the integrity of subsequent test results.
• If you cancel the routine, do not initialize the system. Contact Technical Support for assistance.
• System Check Solution and wash buffer contain ProClin™ 300 preservative, which may cause
sensitization by skin contact. After contact with skin, wash immediately with plenty of soap and
water. Wear suitable gloves.
Required Materials
• 13-mm rack for 2-mL sample cups; the rack must have a rack ID between 1 and 57 or 400 and 456
• 2.0-mL sample cups (four cups)
NOTE
Use only the 2.0-mL sample cups when performing the maintenance routines. Using any other sample
containers may result in level sensing errors and cancellation of the maintenance routine.
Unload rack
OK
F1
Pipette 2.0 mL
undiluted System
Check Solution Results Repeat the
into cup 7 acceptable? No System Check
Enter values
Results
and initial Yes acceptable?
Maintenance Log, as
applicable
Leave cup 9 empty
No
Troubleshoot System
Pipette 1.0 mL Check results, or
1/501 System Check contact Technical
Solution into cup 10 Support
0601B.wmf
DAILY MAINTENANCE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Check Zone
Temperatures
C02062-AA
Check System Supplies
Empty Liquid Waste
Bottle
System Backup
Successful?
Maintenance Log
SYSTEM CHECK RESULTS Week 1 Week 2 Week 3 Week 4 Week 5 Expected Results
Today’s Date *The Substrate :
Washed RLU/%CV 5,000 to 20,000 / ≤12.00 Washed Ratio and the
Unwashed Check RLU
Substrate RLU/%CV 5,000 to 9,000 / ≤5.00
mean result are not
Unwashed RLU/%CV 4 to 10 million* / ≤2.00 system specifications.
Wash Efficiency PPM -1.0 to +1.0 They are only
6: Maintenance
6-13
Substrate Ratio 0 to 1.40 reference guidelines.
7 Troubleshooting
Event Log
The Event Log is a list of events the Access 2 system generates as it monitors the status of various system
parameters. You can use these events to keep informed of system operations and to assist with troubleshooting.
NOTE
If you have a workgroup with more than one instrument, you can view only the events that are specific for
the instrument attached to the PC displaying the Event Log.
Troubleshooting Events
If a system error or potential problem generates an event in the Event Log, you can view troubleshooting
information for the event. This troubleshooting information includes:
• Possible causes for the event.
• Brief outline of troubleshooting instructions.
• Links to detailed procedures. Always review the detailed procedures if you are not completely familiar
with them.
You can view technical information about an event in the Details window. The event details, particularly the
event code, can be useful for troubleshooting. If you contact Technical Support for assistance, the representative
may request the details for one or more events.
NOTE
Event troubleshooting information is available online within the Access 2 Help system.
Troubleshooting Events
Event Log
Review events
Yes
Event Log
Troubleshoot
F2
Access 2 Help
Yes
Yes
Details
F4
Details
Print
F7
0840B.wmf
NOTE
The corrective actions in the troubleshooting tables contain abbreviated procedures. Regardless of the
frequency with which you perform a procedure, regularly review the complete procedures, including their
cautions for protecting the instrument from damage, and their warnings for ensuring your personal safety.
Links to the procedures are available for the tables in the Help system.
High %CV Insufficient supply of substrate 1. Check the level of substrate in the bottle.
throughout the run 2. If necessary, change the substrate bottle from the Supplies
screen.
3. Repeat the System Check routine.
High %CV Leak in the substrate system 1. With the front panel open, open the top cover.
throughout the run tubing 2. Visually inspect the tubing leading to and away from the
(continued) substrate pump/valve/heater assembly for leaks and/or
crystalline deposits.
3. Deposits may indicate the tubing is damaged and should be
replaced.
• If you find deposits, contact Technical Support for further
instructions.
• If you find no deposits, continue troubleshooting other
possible causes.
Kinked substrate system tubing 1. With the front panel and top cover open, visually inspect the
tubing leading to and away from the substrate pump/valve/heater
assembly for kinks.
• If no kinks are found, continue troubleshooting other possible
causes.
2. Straighten any kinks. Close the top cover and front panel.
• If the kinked tubing is part of the substrate probe, remove and
replace the substrate probe.
3. Close the top cover and front panel.
4. If you straightened kinks, but did not need to replace the
substrate probe:
• Prime the substrate system for 4 cycles.
• Repeat the System Check routine.
Bent substrate probe tip 1. With the front panel open, inspect the substrate probe tip.
• If the substrate probe tip is not bent, continue troubleshooting
other possible causes.
2. If the substrate probe tip is bent, remove and replace the
substrate probe.
High %CV from System Check solution diluted 1. Prepare a new 1/501 dilution using a new vial of System Check
high values in the incorrectly solution.
last one or two 2. Repeat the System Check routine.
replicates
Sample containers out of order in 1. Place the sample containers in the correct order.
the maintenance rack 2. Repeat the System Check routine.
High RLU mean Wash/read carousel or substrate 1. Check the wash carousel and substrate temperatures on the
temperatures too high (results Maintenance Review screen.
may be flagged TRS or TRW) 2. If the temperatures for either device is displayed in red (out of
range), contact Technical Support.
System Check solution diluted 1. Prepare a new 1/501 dilution using a new vial of System Check
incorrectly solution.
2. Repeat the System Check routine.
Sample containers out of order in 1. Place the sample containers in the correct order.
the maintenance rack 2. Repeat the System Check routine.
High RLU mean Substrate pump delivering too Run the Visual Substrate Volume Check procedure.
(continued) much substrate • If the results do not fail, continue troubleshooting other possible
causes.
• If the results fail, contact Technical Support.
Contaminated substrate supply 1. Contact Technical Support to verify the need to decontaminate
the substrate system.
2. If Technical Support agrees, decontaminate the substrate system
and then verify system performance.
Low RLU mean Expired substrate 1. Review the expiration date on the Supplies screen.
2. If necessary, change the substrate bottle from the Supplies
screen.
3. Repeat the System Check routine.
Improperly stored substrate 1. Check the date the bottle was placed at room temperature.
(bottle stored at room 2. If necessary, change the substrate bottle from the Supplies
temperature for too long) screen.
3. Repeat the System Check routine.
Wash/read carousel or substrate 1. Check the wash carousel and substrate temperatures on the
temperatures too low (results Maintenance Review screen.
may be flagged TRS or TRW) 2. If the temperatures for either device are displayed in red (out of
range), contact Technical Support.
Incomplete priming of the 1. Prime the substrate system for 20 cycles again.
substrate after substrate system 2. Repeat the System Check routine.
decontamination
Substrate bottle contaminated 1. Change the substrate bottle from the Supplies screen. Do not
with Citranox cleaning solution prime the substrate system when prompted as you change the
during substrate system bottle.
decontamination 2. Prime the substrate system from the Prime Fluidics window for
20 cycles.
3. Repeat the System Check routine.
Substrate dispense volume too Run the Visual Substrate Volume Check procedure.
low • If the results do not fail, continue troubleshooting other possible
causes.
• If the results fail, contact Technical Support.
No RLU values Substrate bottle contaminated 1. Change the substrate bottle from the Supplies screen. Do not
with Citranox cleaning solution prime the substrate system when prompted as you change the
during substrate system bottle.
decontamination 2. Prime the substrate system from the Prime Fluidics window for
20 cycles.
3. Repeat the System Check routine.
RLUs trending up Substrate delivery inconsistent Troubleshoot according to the High %CV symptoms for substrate
or down over time check.
Substrate bottle contaminated 1. Change the substrate bottle from the Supplies screen. Do not
with Citranox cleaning solution prime the substrate system when prompted as you change the
during substrate system bottle.
decontamination 2. Prime the substrate system from the Prime Fluidics window for
20 cycles.
3. Repeat the System Check routine.
High substrate Dirty or plugged aspirate probes 1. Clean the aspirate probes.
ratio 2. Repeat the System Check routine.
Plugged waste air filter 1. Check the tubing from the liquid waste bottle for constrictions.
• If the tubing is constricted, clear it and repeat the System
Check routine.
2. If the results fail, remove the tubing from the waste filter at the
quick disconnect and place the end of the tubing in one of the
adjacent holes. This may temporarily resolve the error until you
can install a new waste filter assembly.
3. Repeat the System Check routine.
• If the results are acceptable, contact Technical Support for
assistance in ordering and replacing the waste filter assembly.
• If the results fail, contact Technical Support.
High %CV Air in the main pipettor fluid 1. Open the front panel.
lines 2. Visually inspect the main pipettor line for air bubbles. Be sure to
check the tubing leading out of the precision pump and entering
the main pipettor.
3. If bubbles are present, prime the pipettor.
4. Inspect the tubing again.
• If the bubbles are gone, run the System Check routine.
• If you still see bubbles, contact Technical Support for further
instructions.
No wash buffer 1. Check the wash buffer supply in the fluids tray.
• If the wash buffer reservoir is empty, but the Wash Buffer
status button is not red, contact Technical Support.
2. If necessary, change the wash buffer bottle.
3. Repeat the System Check routine.
Precision pump, valve, fittings, 1. Open the front panel and the top cover.
or tubing damaged and leaking 2. Visually inspect the tubing leading from the precision pump
valve, to the pressure monitor if applicable, to the main pipettor
for leaks and/or crystalline deposits.
3. Examine the precision pump and valve for leaks and/or
crystalline deposits.
4. Deposits may indicate that the pipetting system is damaged and
should be replaced.
• If you see deposits, contact Technical Support for further
assistance.
• If you do not see deposits, continue troubleshooting other
possible causes.
High %CV Problems with the substrate 1. If the substrate check %CV result is within the expected range,
(continued) system: the substrate system is not the cause of the high %CV for the
• Insufficient supply of substrate unwashed check. Continue troubleshooting other possible
causes.
• Air in the substrate lines
2. If the substrate check result is not within the expected range,
• Leak in the substrate system
troubleshoot the substrate check result.
tubing
• Kinked substrate system
tubing
• Bent substrate probe
• Substrate pump or valve
failure
Primary probe partially plugged 1. Look at the Pressure Monitor screen to determine if a pressure
sensor is present, and that obstruction detection is enabled.
• If the pressure sensor is not present, or if it is present but
obstruction detection is disabled, continue with step 2.
• If obstruction detection is enabled, review the Event Log, and
troubleshoot according to the events related to main pipettor
obstructions or abnormal pressures.
2. Run the Special Clean routine.
3. Run the System Check routine.
4. If the results fail, contact Technical Support to verify the need to
replace the primary probe.
5. When needed, remove and replace the primary probe.
High RLU mean System Check solution diluted 1. Prepare a new 1/501 dilution using a new vial of System Check
incorrectly solution.
2. Repeat the System Check routine.
Sample containers out of order in 1. Place the sample containers in the correct order.
the maintenance rack 2. Repeat the System Check routine.
Problems with the substrate 1. If the substrate check mean RLU result is within the expected
system: range, the substrate system is not the cause of the high mean
• Wash/read carousel or RLU result for the unwashed check. Continue troubleshooting
substrate temperatures too other possible causes.
high 2. If the substrate check result is not within the expected range,
• Contaminated substrate supply troubleshoot the substrate check result.
• Substrate pump delivering too
much substrate
Low RLU mean System Check solution diluted 1. Prepare a new 1/501 dilution using a new vial of System Check
incorrectly solution.
2. Repeat the System Check routine.
Sample containers out of order in 1. Place the sample containers in the correct order.
the maintenance rack 2. Repeat the System Check routine.
System Check expired or stored 1. Prepare a new 1/501 dilution using a new vial of unexpired
incorrectly System Check solution.
2. Repeat the System Check routine.
Problems with the substrate 1. If the substrate check RLU mean result is within the expected
system: range, the substrate system is not the cause of the low RLU mean
• Expired substrate result for the unwashed check. Continue troubleshooting other
possible causes.
• Insufficient supply of substrate
2. If the substrate check result is not within the expected range,
• Wash/read carousel or
troubleshoot the substrate check result.
substrate temperatures too low
• Incomplete priming after
substrate system
decontamination
• Substrate supply contaminated
with Citranox cleaning
solution during substrate
system decontamination
• Substrate dispense volume too
low
High %CV Dirty or plugged aspirate probes 1. Clean the aspirate probes.
2. Repeat the System Check routine.
High % CV Plugged waste air filter 1. Check the tubing from the liquid waste bottle for constrictions.
(continued) • If the tubing is constricted, clear it and repeat the System Check
routine.
2. If the results fail, remove the tubing from the waste filter at the
quick disconnect and place the end of the tubing in one of the
adjacent holes. This may temporarily resolve the error until you
can install a new waste filter assembly.
3. Repeat the System Check routine.
• If the results are acceptable, order a new waste filter assembly
and replace the current one.
• If the results fail, contact Technical Support.
Damaged peristaltic waste pump 1. Open the front panel and top cover.
tubing 2. Visually inspect the tubing and area under the tubing, from the
peristaltic waste pump to the aspirate probes for leaks and/or
crystalline deposits.
3. Deposits may indicate that the tubing is damaged and should be
replaced.
• If you see deposits, contact Technical Support for further
assistance.
• If you do not see deposits, continue troubleshooting other
possible causes.
One or more aspirate probes 1. With the front panel open, verify that the aspirate probes can move
stuck in the “up” position up and down.
• Grasp each probe gently just below the wash arm.
• The probe should move up and down slightly.
2. Replace any probe that does not move freely.
3. With the front panel closed, repeat the System Check routine.
4. If the problem persists, contact Technical Support.
Air in the wash pump system 1. With the front panel open, visually inspect the tubing from the
tubing wash buffer supply to the dispense probes for air bubbles.
2. If you see bubbles, close the front panel and prime the dispense
probes for 4 cycles.
3. Open the front panel and inspect the tubing again.
• If the bubbles are gone, close the front panel and repeat the
System Check routine.
• If you still see bubbles, contact Technical Support for further
instructions.
Kinked wash pump system 1. With the front panel open, visually inspect the tubing from the
tubing wash buffer supply to the dispense probes for kinks.
• If no kinks are found, continue troubleshooting other possible
causes.
2. Straighten any kinks.
3. Close the front panel and prime the dispense probes for 4 cycles.
4. Repeat the System check routine.
High % CV Problems with the substrate 1. If the substrate check %CV result is within the expected range, the
(continued) system: substrate system is not the cause of the high %CV for the washed
• Insufficient supply of substrate check. Continue troubleshooting other symptoms.
• Air in substrate lines 2. If the substrate check result is not within the expected range,
troubleshoot the substrate check result.
• Leak in the substrate system
tubing
• Kinked substrate system
tubing
• Bent substrate probe
• Substrate pump or valve
failure
Problems with the pipetting 1. If the unwashed check %CV result is within the expected range,
system: the pipetting system is not the cause of the high %CV for the
• No wash buffer washed check. Continue troubleshooting other possible causes.
• Worn or damaged precision 2. If the unwashed check result is not within the expected range,
pump seals troubleshoot the unwashed check result.
• Partially plugged primary
probe
• Precision pump, valve,
fittings, or tubing damaged
and leaking
High RLU mean Dirty or plugged aspirate probes 1. Clean the aspirate probes.
2. Repeat the System Check routine.
One or more aspirate probes 1. With the front panel open, verify that the aspirate probes can move
stuck in the “up” position up and down.
• Grasp each probe gently just below the wash arm.
• The probe should move up and down slightly.
2. Replace any probe that does not move freely.
3. With the front panel closed, repeat the System Check routine.
4. If the problem persists, contact Technical Support.
High RLU mean Problems with the substrate 1. If the substrate check RLU mean result is within the expected
(continued) system: range, the substrate system is not the cause of the high RLU mean.
• Wash/read carousel or Continue troubleshooting other possible causes.
substrate temperatures too 2. If the substrate check result is not within the expected range,
high troubleshoot the substrate check result.
• Contaminated substrate supply
• Substrate pump delivering too
much substrate
Low RLU mean Diluted System Check solution 1. Be sure that sample container 1 contains undiluted System Check
used instead of undiluted solution.
2. Repeat the System Check routine.
Containers 1 and 4 switched in 1. Be sure that sample container 1 contains undiluted System Check
the maintenance rack solution and sample container 4 contains 1/501 diluted System
Check.
2. Repeat the System Check routine.
Problems with the pipetting 1. If the unwashed check %CV result is within the expected range,
system the main pipettor is not the cause of the low RLU mean result for
the washed check. Continue troubleshooting other possible causes.
2. If the unwashed check result is not within the expected range,
troubleshoot the unwashed check result.
Problems with the substrate 1. If the substrate check %CV and RLU mean results are within
system: expected ranges, the substrate system is not the cause of the low
• Air in the substrate lines RLU mean result for the washed check. Continue troubleshooting
other possible causes.
• Leak in the substrate system
tubing 2. If the substrate check results are not within the expected ranges,
troubleshoot the substrate check result.
• Kinked substrate system
tubing
• Bent substrate probe
• Substrate pump or valve
failure
• Expired substrate
• Insufficient supply of substrate
• Wash/read carousel or
substrate temperatures too low
• Incomplete priming after
substrate system
decontamination
• Substrate supply contaminated
with Citranox cleaning
solution during substrate
system decontamination
• Substrate dispense volume too
low
Wash Efficiency Problems with the RV wash system (washed Troubleshoot the washed check result.
PPM fails to meet check mean RLUs or %CVs too high)
criteria
Problems with the pipetting system (unwashed Troubleshoot the unwashed check result.
check mean RLUs too low)
Problems with the substrate system Troubleshoot the substrate check result.
(contaminated substrate bottle)
Instrument Troubleshooting
Instrument troubleshooting information can help you identify and correct instrument problems that are not
resolved during System Check troubleshooting. Most instrument troubleshooting is available through the Event
Log. The information in this section addresses problems you notice during operation or inspection that do not
generate a caution or warning event.
NOTE
The corrective actions in the troubleshooting tables contain abbreviated procedures. Regardless of the
frequency with which you perform a procedure, regularly review the complete procedures, including their
cautions for protecting the instrument from damage, and their warnings for ensuring your personal safety.
Links to these procedures are available for these tables in the Help system.
System mode is Not System rebooting or initializing Wait for the instrument to finish rebooting or initializing.
Ready When this occurs, the system mode will change to the
Ready mode and any text message in the system mode area
will disappear.
The Event Log System event occurred Review the event log message and its troubleshooting
button is yellow or information. Take the recommended corrective action.
red
System Database contains too many test and QC 1. Set up the system to automatically delete test and QC
performance is slow results results from the database.
2. If the Auto-Delete function is already set up, consider
reducing the number of days that must elapse before the
test results are deleted.
System fails during No power to instrument Refer to the power supply troubleshooting table.
reboot
CD in CD-ROM or DVD drive, or Remove the CD from the CD-ROM or DVD drive, or remove
data disk in 3.5-inch disk drive. the disk from 3.5-inch disk drive and reboot the instrument.
System fails during Interlock switch detects front panel of Close the front panel of the instrument and initialize the
initialization instrument is open system.
No power to the Faulty power outlet or power 1. Plug the instrument and the peripherals into a different power
instrument or strip outlet or strip.
peripherals 2. If the problem persists, troubleshoot the power system for the
instrument and each peripheral device individually.
3. When the source of the problem has been identified, contact
Technical Support if necessary.
No power to the Instrument power switch not on Turn on the power switch located on the right side of the instrument.
instrument
Power cord connector loose at Be sure that the power cord connections to the outlet and instrument
either end are plugged in tightly.
No power to Peripheral device power switch 1. Turn on the peripheral device power switch.
peripheral(s) not on 2. If you cannot locate the power switch, see the documentation
provided by the manufacturer.
Peripheral device power cord Be sure that the power cord connections to the outlet and peripheral
connector loose at either end device are plugged in tightly.
Keyboard Problems
Use the following table to troubleshoot situations when the computer keyboard is not responding to your key
strokes.
Keyboard Troubleshooting
No response from Loose connector Be sure that the cable connections (located on the back side of the
the keyboard external computer) between the keyboard, the external bar code
reader, and the computer are plugged in tightly.
Invalid key selected Look at the status line above the function buttons near the bottom of
the screen to verify that the key selected is an acceptable keyboard
response.
Printer Problems
Use the following table to troubleshoot problems with the printer.
Printer Troubleshooting
Print job was not Communication interrupted 1. Check the printer paper tray.
printed, or is between system and printer • If empty, load paper into the printer paper tray.
incomplete
2. Try to print the missing or incomplete report once.
• If print service is restored, continue normal operation.
3. If print service is not restored, reboot the PC.
4. If print service is not restored, locate missing report and print
from the appropriate screen. For example, print a missing
Calibration report from the Calibration screen.
5. If the report does not print, be sure the cable connections
between the printer and the computer are plugged in tightly.
6. Reload paper into the printer paper tray.
7. Try to print the report again.
8. If report still does not print, contact Technical Support.
Sample rack bar Bar code label attached Attach the bar code label correctly.
code label was not incorrectly to the rack
read by the internal
bar code reader Incorrect system bar code label If the sample rack is labeled with an Access (not an Access 2)
on rack system label, the rack cannot be used.
Either replace the bar code label with an Access 2 rack label or
move the samples to an Access 2 labeled rack.
Sample bar code Sample bar code reading not Set up the bar code reader to read sample IDs.
label was not read enabled
by the internal bar
code reader Sample container placed Turn the sample container until its bar code label faces the same
incorrectly in the rack direction as the rack bar code label, and is visible through the slot in
the sample rack.
Bar code label applied Apply the bar code label correctly on the sample container.
incorrectly on the sample
container
Bar code reader not set up to read Setup the bar code reader parameters to match the symbology you
the sample bar code symbology are using.
Sample bar code Bar code symbology or 1. Verify that the bar code symbology and parameters used on the
label was not read parameters not supported sample label are supported.
by the internal bar 2. If the symbology or parameters are not supported, contact
code reader Technical Support.
(continued)
Number of bar code characters 1. Check the number of characters entered for the Interleaved 2 of 5
on label does not agree with the symbology in the Bar Code Reader Setup window.
number of characters specified in 2. If necessary, change the number of characters to match the
the Bar Code Reader Setup number of characters on the label.
window (for Interleaved 2 of 5
3. If the number of characters match, contact Technical Support.
only)
Loose connector Be sure that the cable connections (located on the back side of the
external computer) between the keyboard, the external bar code
reader, and the computer are plugged in tightly.
Air in the fluid lines Insufficient priming 1. Prime all fluidic components.
2. Run the System Check routine.
3. If the problem persists, contact Technical Support.
System prime lost due to a leak 1. Be sure the connections between the fittings (located on the left
in the fluidic system side of the instrument) and the fluids tray fit together tightly.
Visually inspect the tubing and fittings for leaks and deposits.
2. Visually inspect the fluidic module tubing and fittings for leaks
and deposits.
3. Inspect the precision pump valve and valve fittings for leaks and
deposits.
4. Leaks and deposits may indicate that the fluidics system is
damaged.
• If you find leaks and deposits, contact Technical Support for
further assistance.
• If you find no leaks or deposits, continue troubleshooting
other possible causes.
Kinked tubing in the fluids tray 1. Visually inspect all tubing leading away from the fluids tray
components for kinks. Pay close attention to the tubing leading
away from the substrate bottle.
• If no kinks are found, continue troubleshooting other possible
causes.
2. Straighten any kinks.
3. Prime all fluidic components.
4. Run the System Check routine.
Plugged waste air filter 1. Check the tubing from the liquid waste bottle for constrictions.
• If the tubing is constricted, clear it and repeat the System
Check routine.
2. If the results fail, remove the tubing from the waste filter at the
quick disconnect and place the end of the tubing in one of the
adjacent holes. This may temporarily resolve the error until you
can install a new waste filter assembly.
3. Repeat the System Check routine.
• If the results are acceptable, order a new waste filter assembly
and replace the current one.
• If the results fail, contact Technical Support.
Main pipettor Primary probe partially plugged 1. Look at the Pressure Monitor screen to determine if a pressure
leaking sensor is present, and that obstruction detection is enabled.
• If the pressure sensor is not present, or if it is present but
obstruction detection is disabled, continue with step 2.
• If obstruction detection is enabled, review the Event Log, and
troubleshoot according to the events related to main pipettor
obstructions or abnormal pressures.
2. Run the Special Clean routine.
3. Run the System Check routine.
4. If the results fail, contact Technical Support to verify the need to
replace the primary probe.
5. When needed, remove and replace the primary probe.
Loose fitting between the main 1. Open the front panel and top cover.
pipettor and precision pump 2. Visually inspect the fittings between the primary probe and
valve precision pump valve for leaks and deposits.
3. Inspect all fluid fittings to the manifold for leaks and deposits.
4. Leaks or deposits may indicate that the tubing or fittings are
damaged.
• If you find leaks or deposits, contact Technical Support for
further assistance.
• If you find no leaks or deposits, continue troubleshooting
other possible causes.
Main pipettor Primary probe partially plugged, 1. Look at the Pressure Monitor screen to determine if a pressure
dispensing fluid causing the main pipettor to sensor is present, and that obstruction detection is enabled.
with excess dispense fluid with too much • If the pressure sensor is not present, or if it is present but
pressure pressure obstruction detection is disabled, continue with step 2.
• If obstruction detection is enabled, review the Event Log, and
troubleshoot according to the events related to main pipettor
obstructions or abnormal pressures.
2. Check RV locations 1 and 2 of the RV Shuttle for crystalline
deposits.
• If you find no deposits, continue troubleshooting other
possible causes.
3. Run the Special Clean routine.
4. Run the System Check routine.
5. If the results fail, contact Technical Support to verify the need to
replace the primary probe.
6. When needed, remove and replace the primary probe.
Communication Problems
Use the following table to troubleshoot communication problems between the external computer (PC) and the
instrument, or other systems in your workgroup.
Communications Troubleshooting
Workgroup Loose connection between PC 1. Be sure that the Access 2 instrument power switch is on.
communication and other systems in the • If the power switch is off, restart the instrument.
error occurs during workgroup
• If the power switch is on, but there is no power to the
rebooting process
instrument, refer to the power supply troubleshooting table.
2. Check both ends of the cable connecting the PC to the other
systems in your workgroup:
• If the cable connection is loose at either end, connect it
securely.
• If the cable is not loose, contact Technical Support.
3. Reboot the PC.
Assay Troubleshooting
Assay troubleshooting information provides troubleshooting tables to help you identify and correct
assay-related problems. Before using assay troubleshooting information to troubleshoot assay issues, run the
System Check routine and troubleshoot unexpected results. Running the System Check routine will help you to
identify and correct, or rule out, some system issues as the cause of your problem.
NOTE
The corrective actions in the troubleshooting tables contain abbreviated procedures. Regardless of the
frequency with which you perform a procedure, regularly review the complete procedures, including their
cautions for protecting the instrument from damage, and their warnings for ensuring your personal safety.
Links to these procedures are available for these tables in the Help system.
Although precision Multi-level calibrators pipetted 1. Be sure that the calibrators are placed in the rack in the proper
is good, the out of order: order.
calibration fails for • Calibration curve does not 2. Repeat the calibration.
any reason other ascend or descend smoothly.
than Insuff Data
• Calibration cutoff is opposite
of the expected result.
Expired calibrator 1. Check the expiration date on the calibrator vial(s) or the
Calibrator Setup screen.
2. If the calibrator is expired, add a new, unexpired calibrator on
the Calibrator Setup screen.
3. Repeat the calibration.
Calibrator unstable or Repeat the calibration. If possible, repeat with a new set of the same
contaminated due to improper calibrator lot.
handling
Reagent pack unstable or 1. Unload the reagent pack and load a new reagent pack.
contaminated due to improper • If the lot number of the new reagent pack is different than the
handling removed pack, recalibrate the assay.
2. For calibration problems, repeat the calibration.
Contaminated substrate supply 1. Contact Technical Support to verify the need to decontaminate
the substrate system.
2. If Technical Support agrees, decontaminate the substrate system
and then verify system performance.
Incorrect calibration information 1. Verify the calibration information on the Calibrator Setup
entered during calibrator setup screen.
2. If necessary, edit the information on the Calibrator Setup screen.
3. Repeat the calibration.
Although precision Wrong calibrator placed in rack 1. Be sure that the correct calibrator is loaded on the rack.
is good, the 2. Repeat the calibration.
quantitative curve
is flat or the Wrong reagent pack loaded, or 1. Verify that the reagent pack was loaded by checking the Supplies
qualitative result is placed in an incorrect position on screen. If it is missing, load a reagent pack.
No Value; the reagent carousel 2. If the reagent pack is listed on the screen, verify its location by
calibration fails for unloading it from the carousel.
any reason other
• If the reagent pack is onboard, but is not presented during this
than Insuff Data
process, contact Technical Support for assistance.
• If Technical Support confirms a misplaced reagent pack,
retrieve the pack.
3. Reload the reagent pack after you have verified its location.
4. Repeat the calibration.
Poor precision, and Daily or weekly maintenance not 1. Perform daily or weekly maintenance.
calibration fails for performed 2. Repeat the calibration.
any reason other
than Insuff Data Problems with the pipetting 1. If the System Check routine results are within the expected
system: ranges, the pipetting system is not the cause of the calibration
• Air in the main pipettor fluid failure. Continue troubleshooting other possible causes.
lines. 2. If the unwashed check result is not within the expected range,
• Wash buffer supply lines troubleshoot the unwashed check result.
kinked.
• Precision pump, valve,
fittings, or tubing damaged
and leaking.
Problems with the substrate 1. If the System Check routine results are within the expected
system: ranges, the substrate system is not the cause of the calibration
• Insufficient supply of failure. Continue troubleshooting other possible causes.
substrate. 2. If the substrate check result is not within the expected range,
• Air in the substrate lines. troubleshoot the substrate check result.
• Kinked substrate system
tubing.
• Leak in the substrate system
tubing.
RLUs are too low at Reagent pack switch occurred, 1. Verify that the second reagent pack was loaded by checking the
one end of the and the second reagent pack is Supplies screen. If the pack is missing, load a new reagent pack.
calibration curve, missing or loaded incorrectly 2. If the second reagent pack is listed on the screen, verify its
and calibration fails (RLUs drop off suddenly at the location by unloading it from the carousel.
for any reason end of the calibration curve)
• If the reagent pack is onboard, but is not presented during this
other than Insuff
process, contact Technical Support for assistance.
Data
• If Technical Support confirms a misplaced reagent pack,
retrieve the pack.
3. Reload the second pack after you have verified its location.
4. Repeat the calibration.
Calibration fails for Quantity of calibrator not 1. Calculate the correct volume of calibrator needed for the number
the reason Insuff sufficient for testing (QNS result of replicates and type of sample container used.
Data flag and event) 2. Pipette the amount of calibrator calculated in step 1 into the
appropriate sample container.
3. Be sure the rack and rack ID are correct.
4. Repeat the calibration.
5. If the calibration fails again, contact Technical Support for
assistance.
Calibration fails for Two or more replicates not 1. Review the Event Log messages for device errors prior to the
the reason Insuff calculated due to instrument calibration failure.
Data error 2. Review the Test Results screen for any error messages.
(continued) 3. Troubleshoot the device errors. If necessary, contact Technical
Support for assistance.
4. After resolving the device errors, repeat the calibration.
Sample container missing or 1. Be sure that the sample containers are in the proper order in the
placed in incorrect position in the rack.
rack 2. Repeat the calibration.
Controls violate Daily or weekly maintenance not 1. Perform daily or weekly maintenance.
applied QC rule performed 2. Repeat the test(s).
Wrong QC control lot number Repeat the test(s) using the correct lot number.
selected for the test
Wrong mean and/or standard 1. Review the QC information using Edit Control F2 from the QC
deviation information entered Setup screen.
when setting up the QC controls 2. If necessary, edit the information.
3. Repeat the test(s).
QC control sample type does not 1. Verify that the sample type used matches what is listed on the
match QC setup Test Request screen.
2. Verify that the sample type used matches the sample type set up
for the QC control.
3. If the QC control sample type entered during setup is incorrect,
you must delete the control and set up the control again using the
correct sample type.
Expired QC control Load new, unexpired QC control, and repeat the test(s).
Controls violate Quantity of QC control not 1. Calculate the correct volume of QC control needed for the
applied QC rule sufficient for testing (QNS result number of replicates and type of sample container used.
(continued) flag and event) 2. Pipette the amount of QC control calculated in step 1 into the
appropriate sample container.
3. Be sure the rack and rack ID are correct.
4. Repeat the test(s).
5. If the test(s) fail again, contact Technical Support for assistance.
Reagent pack unstable or 1. Unload the reagent pack and load a new reagent pack.
contaminated due to improper • If the lot number of the new reagent pack is different than the
handling removed pack, recalibrate the assay.
2. Repeat the test(s).
Wrong reagent pack loaded, or 1. Verify that the reagent pack was loaded by checking the Supplies
placed in an incorrect position on screen. If it is missing, load a reagent pack.
the reagent carousel 2. If the reagent pack is listed on the screen, verify its location by
unloading it from the carousel.
• If the reagent pack is onboard, but is not presented during this
process, contact Technical Support for assistance.
• If Technical Support confirms a misplaced reagent pack,
retrieve the pack.
3. Reload the reagent pack after you have verified its location.
4. Repeat the test(s).
Expired reagent pack 1. Check the reagent pack expiration date on the Reagent Inventory
screen.
2. Unload the expired reagent pack and load a new one.
• If the lot number of the new reagent pack is different than the
expired pack, recalibrate the assay.
3. Repeat the test(s).
Wrong sample container and/or Load the samples using appropriate sample containers and racks. Be
rack sure that the sample containers are in a rack with the correct rack ID.
Sample container missing or 1. Be sure that the sample containers are in the proper order in the
placed in incorrect position in the rack.
rack 2. Repeat the test(s).
Expired assay calibration 1. Check the Reagent Pack button on the Supplies screen. If the lot
number is red, the calibration has expired.
2. View the test results. Any test results calculated with an expired
calibration will be flagged.
3. Recalibrate the assay.
4. Repeat the test(s).
Substrate not equilibrated to The substrate must be at room temperature for specified period of
room temperature when loaded time before it is loaded.
on the system For more information about the substrate, refer to the reagent
instructions for use.
Increasing values QC controls evaporating from 1. Load a fresh aliquot of each QC control.
across the run being on the instrument too long 2. Repeat the test(s). Be sure the controls are pipetted into the RVs
within 1 hour of loading onto the instrument. Be sure controls
from 3-mL sample containers are pipetted into the RVs within
two hours of loading.
Unexpected Incorrect patient sample type Refer to the reagent instructions for use for the appropriate sample
sample results types for each assay.
Patient sample handled Be sure that sample is stored and handled according to proper
incorrectly laboratory procedures. Refer to individual assay reagent instructions
for use for proper sample handling instructions.
Microclots present in the serum 1. Spin the sample again or use a serum filter to separate the serum
sample from any microclots.
2. Repeat the test(s).
Quantity of patient sample not 1. Calculate the correct volume of patient sample needed for the
sufficient for testing (QNS result number of replicates and type of sample container used.
flag and event) 2. Pipette the amount of QC control calculated in step 1 into the
appropriate sample container.
3. Be sure the rack and rack ID are correct.
4. Repeat the test(s).
5. If the test(s) fail again, contact Technical Support for assistance.
Unexpected Patient samples evaporating from 1. Load a fresh aliquot of each patient sample.
sample results being on the instrument too long 2. Repeat the test(s). Be sure the sample is pipetted into the RVs
(continued) within 1 hour of loading it onto the instrument. Be sure sample
from 3 mL sample containers is pipetted into the RVs within two
hours of loading.
Wrong reagent pack loaded, or 1. Verify that the reagent pack was loaded by checking the Supplies
placed in an incorrect position on screen. If it is missing, load a reagent pack.
the reagent carousel 2. If the reagent pack is listed on the screen, verify its location by
unloading it from the carousel.
• If the reagent pack is onboard, but is not presented during this
process, contact Technical Support for assistance.
• If Technical Support confirms a misplaced reagent pack,
retrieve the pack.
3. Reload the reagent pack after you have verified its location.
4. Repeat the test(s).
Expired reagent pack 1. Check the reagent pack expiration date on the Reagent Inventory
screen.
2. Unload the expired reagent pack and load a new one.
• If the lot number of the new reagent pack is different than the
expired pack, recalibrate the assay.
3. Repeat the test(s).
Reagent pack unstable or 1. Unload the reagent pack and load a new reagent pack.
contaminated due to improper • If the lot number of the new reagent pack is different than the
handling removed pack, recalibrate the assay.
2. Repeat the test(s).
Expired assay calibration 1. Check the Reagent Pack button on the Supplies screen. If the lot
number is red, the calibration has expired.
2. View the test results. Any test results calculated with an expired
calibration will be flagged.
3. Recalibrate the assay.
4. Repeat the test(s).
Partial bottles of substrate 1. Change the substrate bottle and prime the substrate.
combined 2. Repeat the test(s).
Wrong sample container and/or Load the samples using appropriate sample containers and racks. Be
rack sure that the sample containers are in a rack with the correct rack ID.
Unexpected Sample container missing or 1. Be sure that the sample containers are in the proper order in the
sample results placed in incorrect position in the rack.
(continued) rack 2. Repeat the test(s).
Sample values are Patient samples evaporating from 1. Load a fresh aliquot of each patient sample.
too high being on the instrument too long 2. Repeat the test(s). Be sure the sample is pipetted into the RVs
within 1 hour of loading it onto the instrument. Be sure sample
from 3-mL sample containers is pipetted into the RVs within two
hours of loading.
Low sample value Incorrect dilution factor 1. If the sample result was generated with an automated dilution
flag (for example, the Diluted ßhCG assay), repeat the sample in the
standard assay format (Total ßhCG assay).
2. If the results are still too high, contact Technical Support.
Patient sample handled Be sure that sample is stored and handled according to proper
incorrectly laboratory procedures. Refer to individual assay reagent instructions
for use for proper sample handling instructions.
Increased assay Wrong reagent pack loaded, or 1. Verify that the reagent pack was loaded by checking the Supplies
variability/poor placed in an incorrect position on screen. If it is missing, load a reagent pack.
precision the reagent carousel 2. If the reagent pack is listed on the screen, verify its location by
unloading it from the carousel.
• If the reagent pack is onboard, but is not presented during this
process, contact Technical Support for assistance.
• If Technical Support confirms a misplaced reagent pack,
retrieve the pack.
3. Reload the reagent pack after you have verified its location.
4. Repeat the test(s).
Unexpected shift in Expired reagent pack 1. Check the reagent pack expiration date on the Reagent Inventory
assay results screen.
2. Unload the expired reagent pack and load a new one.
• If the lot number of the new reagent pack is different than the
expired pack, recalibrate the assay.
3. Repeat the test(s).
Expired assay calibration 1. Check the Reagent Pack button on the Supplies screen. If the lot
number is red, the calibration has expired.
2. View the test results. Any test results calculated with an expired
calibration will be flagged.
3. Recalibrate the assay.
4. Repeat the test(s).
Unexpected shift in Wash buffer contaminated 1. Contact Technical Support to verify the need to rinse the wash
assay results buffer reservoir.
(continued) 2. If Technical Support confirms, rinse the wash buffer reservoir.
3. If the System Check results are acceptable:
• Recalibrate the assay.
• Run controls and repeat the test(s).
4. If the System Check results are not acceptable, contact Technical
Support.
Increasing values Patient samples evaporating from 1. Load a fresh aliquot of each patient sample.
across the run being on the instrument too long 2. Repeat the test(s). Be sure the sample is pipetted into the RVs
within 1 hour of loading it onto the instrument. Be sure sample
from 3-mL sample containers is pipetted into the RVs within two
hours of loading.
NOTE
The system may print the same test result flag twice on a sample report. Ignore the duplicate flag.
Fatal Flags
CCR A result could not be calculated 1. Take one of the following actions:
because: • For results other than derived results, skip to step 2.
• One of the tests included in a • For derived results, review each test result used in the derived
derived result formula did not result formula. If a result failed, troubleshoot according to the
produce a result. flag for that result.
• A confirmatory test result could not 2. Review the Event Log and troubleshoot according to the error
be calculated. Usually this occurs events with a similar date and time to this event.
because the samples for the
3. Repeat the test.
qualitative and confirmatory tests
were not aspirated at the same time. • For a derived result, repeat all tests included in the derived
result formula.
• Another error prevented the system
from calculating a result. • For a confirmatory result, repeat the confirmatory and
qualitative tests.
4. If the problem persists, contact Technical Support.
CLT An obstruction was detected. 1. Review the Event Log and troubleshoot according to the error
events with a similar date and time to this event.
2. Repeat the test.
3. If the problem persists, contact Technical Support.
IND • For sandwich assays, which use 1. For IND flagged results which meet the following criteria, dilute
positive slope calibration curves, and rerun the sample:
the result is at the low end of the • The result is on a competitive assay, and
concentration curve and cannot be
• Sample dilution is allowed for the assay, and
distinguished from a system failure
because the RLU reading is too • The sample RLU is low.
low. See the reagent instructions for use to determine the assay type
• For competitive assays, which and whether sample dilution is allowed.
use negative slope calibration 2. Rule out a system problem by reviewing the Event Log for error
curves, the result is at either: events with a date and time shortly before this event.
• The high end of the Troubleshoot accordingly.
concentration curve and cannot 3. If you have ruled out a system problem and the IND flag event is
be distinguished from a system the only issue, recalibrating the affected assay may resolve the
failure because the RLU reading problem.
is too low, or • Recalibrate the assay using a new reagent pack and a new set
• The low end of the of calibrators.
concentration curve and cannot • Run QC.
be distinguished from a system
failure because the RLU reading • Repeat the test on the affected sample.
is too high. • If IND flags persist, contact Technical Support.
QNS The sample volume is insufficient. 1. Follow the troubleshooting instructions for the QNS event in the
Additional tests will not be scheduled Event Log.
for this sample. Tests already 2. Repeat the test(s).
scheduled will be completed.
3. If the problem persists, contact Technical Support.
QSD Insufficient reagent volume was 1. Review the Event Log and troubleshoot according to the error
dispensed into an RV. events with a similar date and time to this event.
2. Run canceled tests again.
3. If the problem persists, contact Technical Support.
RLU The relative light units (RLUs) are 1. Review the Event Log and troubleshoot according to the error
outside the acceptable luminometer events with a similar date and time to this event.
measuring range. 2. Repeat the test(s).
3. If the problem persists, contact Technical Support.
SYS A device error occurred during 1. Review the Event Log and troubleshoot according to the error
processing. events with a similar date and time to this event.
2. Repeat the test(s).
3. If the problem persists, contact Technical Support.
TRI The temperature of the incubator was 1. Check the incubator temperature on the Maintenance Review
outside the acceptable limits. All tests screen. If the temperature is outside the acceptable limits,
being incubated when the temperature periodically monitor the incubator temperature on this screen
was outside limits are flagged. until it is within the acceptable limits.
• If the system was rebooted or instrument covers were recently
opened, you may need to wait up to 30 minutes for the
temperature to normalize.
• If the system was recently restarted, you may need to wait up
to 1 hour for the temperature to normalize.
2. When the incubator temperature is within the acceptable limits,
repeat the test(s).
3. If the temperature does not normalize, or if the problem persists,
contact Technical Support.
TRS The temperature of the substrate was 1. Check the substrate temperature on the Maintenance Review
outside the acceptable limits when it screen. If the temperature is outside the acceptable limits,
was dispensed. periodically monitor the substrate temperature on this screen
until it is within the acceptable limits.
• If the system was rebooted or instrument covers were recently
opened, you may need to wait up to 30 minutes for the
temperature to normalize.
• If the system was recently restarted, you may need to wait up
to 1 hour for the temperature to normalize.
2. When the substrate temperature is within the acceptable limits,
repeat the test(s).
3. If the temperature does not normalize, or if the problem persists,
contact Technical Support.
TRW The temperature of the wash/read 1. Check the wash/read carousel temperature on the Maintenance
carousel was outside the acceptable Review screen. If the temperature is outside the acceptable
limits. All tests with RVs in the limits, periodically monitor the wash/read carousel temperature
wash/read carousel when the on this screen until it is within the acceptable limits.
temperature was outside limits are • If the system was rebooted or instrument covers were recently
flagged. opened, you may need to wait up to 30 minutes for the
temperature to normalize.
• If the system was recently restarted, you may need to wait up
to 1 hour for the temperature to normalize.
2. When the wash/read carousel temperature is within the
acceptable limits, repeat the test(s).
3. If the temperature does not normalize, or if the problem persists,
contact Technical Support.
Non-Fatal Flags
CRH The result is above the upper limit of This is a valid test result. No corrective action is necessary.
the critical range.
Not applicable for QC or Calibrations.
CRL The result is below the lower limit of This is a valid test result. No corrective action is necessary.
the critical range.
Not applicable for QC or Calibrations.
EXS The substrate is expired. 1. Change the substrate bottle from the Supplies screen.
2. Repeat the test(s).
GRY For qualitative assays, the result is This is a valid test result. No corrective action is necessary.
within the specified gray zone.
LEX The reagent pack lot is expired. 1. Unload the expired reagent pack and load a new one.
• If the lot number of the new reagent pack is different than the
expired pack, recalibrate the assay.
2. Repeat the test(s).
LRH The result is above the upper limit of 1. Review the result.
the LIS range. The system does not 2. Take one of the following actions:
send results with this flag to the LIS.
• Send the result to the LIS manually.
Notes:
• Delete the result and repeat the test(s).
• This flag is displayed only when the
Auto-Send to LIS option is set to
Verify.
• This flag is for quantitative assays,
semi-quantitative assays, and
derived results only.
LRL The result is below the lower limit of 1. Review the result.
the LIS range. The system does not 2. Take one of the following actions:
send results with this flag to the LIS.
• Send the result to the LIS manually.
Notes:
• Delete the result and repeat the test(s).
• This flag is displayed only when the
Auto-Send to LIS option is set to
Verify.
• This flag is for quantitative assays,
semi-quantitative assays, and
derived results only.
LOW The result is lower than the minimum No corrective action is necessary.
reportable result value defined in the
APF.
ORH The result is above the upper limit of No corrective action is necessary.
the reference range.
Not applicable for QC or calibrations.
ORL The result is below the lower limit of No corrective action is necessary.
the reference range.
Not applicable for QC or calibrations.
OVR The calculated concentration is above 1. Review the Event Log for error events with a similar date and
the highest or most concentrated time to this event.
calibrator. This flag is only used for • If events occurred, troubleshoot.
quantitative and semi-quantitative
2. Take one of the following actions:
assays.
• If events occurred, and you performed troubleshooting
procedures, run controls and then repeat the test.
- If the controls are in range, and the test result is reported as
greater than the value of the highest calibrator (>X), you
may be able to dilute the sample. To identify whether the
assay allows dilutions, see the reagent instructions for use.
- If the controls are not in range, follow the QC
troubleshooting instructions.
• If no events occurred, take one of the following actions:
- If the test result is reported as >X, you may be able to dilute
the sample. To identify whether the assay allows dilutions,
see the reagent instructions for use. If dilutions are allowed,
dilute and repeat the test.
- If dilutions are not allowed, no further action is necessary.
3. If you have questions about the result, or if the problem persists,
contact Technical Support.
PEX The open pack stability time has 1. Unload the expired reagent pack and load a new one.
elapsed for the reagent pack. The • If the lot number of the new reagent pack is different than the
system measures this from the time it expired pack, recalibrate the assay.
first punctures the pack.
2. Repeat the test(s).
QCF A QC control violates one or more QC 1. Display the QC Chart and Data screen to review which criteria is
rules. not met.
2. Follow the QC troubleshooting instructions in Assay
Troubleshooting on page 7-21.
RFX The result is from a reflex test. This is a valid test result. No corrective action is necessary.
8 Theory of Operation
Samples are processed on the Access 2 system by assigning them a position on a sample rack, entering test
requests for those samples, and then loading the sample racks onto the instrument for processing.
Pipetting
RV moves from RV
supply to pipetting
position
Substrate Addition
RV moves from wash
position to substrate
addition position
Luminometer
Reading
RV moves from
substrate to
luminometer position
Ejection
RV moves from
luminometer to inc. belt
and is ejected into RV
waste bag
0506A.wmf
To report a test result, the system requires a current calibration for each requested test. The number of
calibrators required, the math model, and all other calibration parameters are defined in the assay protocol file
(APF) for each assay, and by the information you enter into the system when you set up a new calibrator set.
Quantitative Calibrator test results provide a multi-point calibration curve. The system uses the
calibration curve to convert a measured response in RLUs to an analyte
concentration and then expresses the result in numerical units.
Semi-Quantitative Calibrator test results provide a multi-point calibration curve. The system uses the
calibration curve to convert a measured response in RLUs to an analyte
concentration and then expresses the result in numerical units. These assays may
report their quantitative result as a qualitative interpretation, such as reactive,
non-reactive, or equivocal.
Qualitative Calibrator test results provide a cutoff value based on a predefined formula. The
system compares a test result to the cutoff value and then classifies the result as
reactive or non-reactive for the analyte.
2
1
1 Precision
2 Error Band (the Width of the Error Band is Exaggerated for Illustration
Purposes)
3 X Axis: Analyte Concentration
4 Y Axis: RLUs
If the pressure required to draw a sample exceeds an accepted threshold, the test is cancelled and flagged with
the CLT (obstruction) fatal flag, and an event is logged. If two consecutive failures occur on the same sample,
any remaining tests for that sample are cancelled. If five consecutive failures occur, the system stops scheduling
tests and goes to the Not Ready mode when the tests in progress are completed. The operator must then
initialize the system and inspect, clean, or replace the probe as necessary.
A reference curve is determined when the obstruction detection hardware is installed. Determine a new curve
after the main pipettor is cleared of an obstruction; after the main pipettor, pressure sensor, or associated tubing
is replaced; when following troubleshooting instructions; or as instructed by a technical support representative.
NOTE
The reference curve has no impact on reagent pack monitoring, which is also a function of the pressure
monitor.
Was an
obstruction No Process the sample.
detected?
Yes
Two
consecutive Five
CLT flags for the No consecutive CLT No Continue processing with the next test.
same sample? flags?
Yes Yes
Cancel all remaining tests for the Log a fatal warning event. Stop
sample, wash the probe, log a warning scheduling new tests. Complete any
event, and apply the CLT flag to the tests in progress, then go to the
sample. Not Ready mode.
1030A.wmf
A Temperature-Sensitive Assays
The Access assays listed in the table below are affected by changes in the room temperature of the Access 2
system. The result for a given sample test may shift if the room temperature changes significantly from the
temperature at which the assay was calibrated. Depending on the assay, an increase or decrease in room
temperature causes test results to increase or decrease. The magnitude and direction of the result shift for each
assay is shown in the table.
If the change in room temperature exceeds the restricted calibration temperature range, the assay must be
recalibrated at the new temperature before samples are tested. Recalibrating these assays at the new room
temperature limits the magnitude of the shift on sample results.
For each of the affected assays, Beckman Coulter has established a restricted calibration temperature range in
which the assay should be calibrated and run. Within this range, the change in assay results due to temperature
is expected to be within the allowable performance characteristics of the assay.
Restricted
Allowable If the temperature
* * calibration
Part # Product % Change increases, the
temperature
in result assay result…
range
The Access 2 system does not monitor the room temperature or alert the operator if room temperature changes
from the original calibration temperature for assays identified in the table. You should ensure that your
laboratory has established a procedure to monitor and review temperature during system operation. Quality
control may not detect temperature related change in assay results and cannot be used as a substitute for
temperature monitoring.
If your laboratory is unable to maintain the required temperature ranges, do not report results out of the
laboratory. Please contact Beckman Coulter for additional suggestions regarding temperature monitoring and
control.
B Ordering Information
Assay-Specific Reagents
The availability of these assays in your country depends on the status of submissions to local regulatory
agencies. In addition, an assay may not be available for use on all systems. Contact your Beckman Coulter
representative if you have questions about the availability of particular assays.
# Tests or #
Calibrations
Assay Panel Assay Part # Description Volume
(@ 500 µL/
Calibration)
Adrenal/ Cortisol 33600 Reagent 2 x 50 tests 100
Pituitary 33605 Calibrator 6 x 4.0 mL 8
33606 Diluent (S0) 1 x 4.0 mL
Allergy Total IgE 35000 Reagent 2 x 50 tests 100
35005 Calibrator 7 total; S0 @ 6.0 mL 8
S1–S6 @ 4.0 mL
35006 Diluent (S0) 1 x 6.0 mL
Anemia EPO A16364 Reagent 2 x 50 tests 100
A16365 Calibrator 6 total; S0 @ 10 mL 5
S1–S5 @ 2.5 mL
Ferritin/ 33020 Reagent 2 x 50 tests 100
Dil-Ferritin 33025 Calibrator 6 x 4.0 mL 8
Folate / RBC Folate A98032 Reagent 2 x 50 tests 100
A98033 Calibrator 6 x 4.0 mL 8
A99250 Diluent (S0) 1 x 4.0 mL
A14206 Lysing Agent 2 x 100 mL
Intrinsic Factor Ab 387992 Reagent 2 x 50 tests 100
387993 Calibrator 2 x 4.0 mL 16
387999 QC 2 levels; 3 x 4.0 mL each
sTfR A32493 Reagent 2 x 50 tests 100
Soluble Transferrin A32494 Calibrator 6 total; S0 @ 4 mL 5
Receptor S1–S5 @ 2.5 mL
B11056 QC1 1 level; 2 x 2.5 mL each
B11057 QC2 and QC3 2 levels; 2 x 2.5 mL each
Vitamin B12 33000 Reagent 2 x 50 tests 100
33005 Calibrator 6 x 4.0 mL 8
33006 Diluent (S0) 1 x 4.0 mL
# Tests or #
Calibrations
Assay Panel Assay Part # Description Volume
(@ 500 µL/
Calibration)
Blood Virus HAV Ab* 34200 Reagent 2 x 50 tests 100
34205 Calibrator 5 x 2.0 mL 4
34209 QC 2 levels; 3 x 3.5 mL each
HAV IgM* 34210 Reagent 2 x 50 tests 100
34215 Calibrator 2 x 1.0 mL 2
34219 QC 2 levels; 3 x 2.5 mL each
*
HBc Ab 34240 Reagent 2 x 50 tests 100
34245 Calibrator 2 x 1.0 mL 2
34249 QC 2 levels; 3 x 2.0 mL each
HBc IgM* 34250 Reagent 2 x 50 tests 100
34255 Calibrator 2 x 1.0 mL 2
34259 QC 2 levels; 3 x 2.5 mL each
*
HBs Ab A24296 Reagent 2 x 50 tests 100
A24297 Calibrator 6 x 2.5 mL 5
A24298 QC 2 levels; 3 x 3.5 mL each
HBs Ag* A24291 Reagent 2 x 50 tests 100
A24292 Calibrator 2 x 2.7 mL 5
A24294 QC 2 levels; 3 x 4.0 mL each
HBs Ag A24295 Confirmatory Reagent 2 x 50 tests 100 (50 patient
Confirmatory* samples)
HCV Ab PLUS *, † 34330 Reagent 2 x 50 tests 100
34335 Calibrator 2 x 1.0 mL 2
34339 QC 2 levels; 3 x 2.5 mL each
HCV Ab V3*, † B33458 Reagent 2 x 50 tests 100
B33459 Calibrator 2 levels; 1 x 1.0 mL each 4
B33460 QC 2 levels; 2 x 3.5 mL each
HIV Combo *, † A59428 Reagent 2 x 50 tests 100
A59429 Calibrator 2 x 1.7 mL 3
A59430 QC1, QC2, and QC3 3 levels; 2 x 4.4 mL each
B22822 QC4 and QC5 2 levels; 2 x 4.4 mL each
Bone Intact PTH A16972 Reagent 2 x 50 tests 100
Metabolism (routine and A16953 Calibrator 6 x 1.0 mL 2
intraoperative)
Ostase 37300 Reagent 2 x 50 tests 100
(bone alkaline B83876 Calibrator 6 x 2.5 mL 5
phosphatase)
B83877 QC 2 levels; 4.0 mL each
Ultrasensitive hGH 33580 Reagent 2 x 50 tests 100
33585 Calibrator 6 x 2.0 mL 4
25(OH) Vitamin D B24838 Reagent 2 x 50 tests 100
Total B24839 Calibrator 6 x 1.4 mL 2
# Tests or #
Calibrations
Assay Panel Assay Part # Description Volume
(@ 500 µL/
Calibration)
Cardiovascular AccuTnI+3 A98143 Reagent 2 x 50 tests 100
A98144 Calibrator 6 total; S0 & S1@ 1.5 mL 2
S2–S5 @ 1.0 mL
AccuTnI** A78803 †† Reagent 2 x 50 tests 100
33345 Calibrator 6 x 1.0 mL 2
CK-MB 386371 Reagent 2 x 50 tests 100
386372 Calibrator 6 x 2.0 mL 4
Digoxin 33710 Reagent 2 x 50 tests 100
33715 Calibrator 6 x 4.0 mL 8
33716 Diluent 1 x 4.0 mL
Myoglobin 973243 Reagent 2 x 50 tests 100
973244 Calibrator 6 x 1.0 mL 2
Diabetes Ultrasensitive Insulin 33410 Reagent 2 x 50 tests 100
33415 Calibrator 6 x 2.0 mL 4
Infectious CMV IgG* A40702 Reagent 2 x 50 tests 100
Disease A40703 Calibrator 6 x 1.0 mL 2
A40704 QC 2 levels; 3 x 2.5 mL each
CMV IgM* A40705 Reagent 2 x 50 tests 100
A40706 Calibrator 2 levels; 1.0 mL each 2
A40707 QC 2 levels; 3 x 2.5 mL each
Rubella IgG 34430 Reagent 2 x 50 tests 100
34435 Calibrator 6 x 1.0 mL 2
34439 QC 2 levels; 3 x 2.5 mL each
Rubella IgM* A32937 Reagent 2 x 50 tests 100
34445 Calibrator 4 x 1.0 mL 2
34449 QC 2 levels; 3 x 2.5 mL each
Toxo IgG A31588 Reagent 2 x 50 tests 100
A31589 Calibrator 6 x 1.0 mL 2
A31590 QC 2 levels; 3 x 2.5 mL each
Toxo IgM II 34470 Reagent 2 x 50 tests 100
34475 Calibrator 2 x 1.5 mL 3
34479 QC 2 levels; 3 x 3.5 mL each
Inflammation IL-6* A16369 Reagent 2 x 50 tests 100
A16370 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
A16371 QC 3 levels; 2 x 2.5 mL each
Reproductive AFP / Dil-AFP 33211 300 test kit 6 x 50 tests 300
(Cals included) Cals: 7 x 2.5 mL 5
33210* 100 test kit 2 x 50 tests 100
(outside US only)
33215* Calibrator 7 x 2.5 mL 5
(for 100 test kit)
33216 Diluent 1 x 14.0 mL
# Tests or #
Calibrations
Assay Panel Assay Part # Description Volume
(@ 500 µL/
Calibration)
Reproductive, AMH* B13127 Reagent 2 x 50 tests 100
(continued) B13128 Calibrator 6 x 2.0 mL 4
B13129 QC 3 levels; 2 x 2 mL each
DHEA-S A10826 Reagent 2 x 50 tests 100
A10827 Calibrator 6 x 2.0 mL 4
Estradiol 33540 Reagent 2 x 50 tests 100
33545 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
33546 Diluent (S0) 1 x 4.0 mL
Estriol, 33570 Reagent 2 x 50 tests 100
Unconjugated 33575 Calibrator 7 total; S0 @ 4.0 mL 5
S1–S6 @ 2.5 mL
hFSH 33520 Reagent 2 x 50 tests 100
33525 Calibrator 6 x 4.0 mL 8
hLH 33510 Reagent 2 x 50 tests 100
33515 Calibrator 6 x 4.0 mL 8
Inhibin A A36097 Reagent 2 x 50 tests 100
A36098 Calibrator 7 x 2.5 mL 5
A36100 QC 3 levels; 2 x 2.5 mL each
PAPP-A* A48571 Reagent 2 x 50 tests 100
A48572 Calibrator 6 x 1.0 mL 2
A48573 QC 3 levels; 2 x 2.5 mL each
Progesterone 33550 Reagent 2 x 50 tests 100
33555 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
33556 Diluent (S0) 1 x 4.0 mL
Prolactin 33530 Reagent 2 x 50 tests 100
33535 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
SHBG A48617 Reagent 2 x 50 tests 100
A48618 Calibrator 6 x 1.0 mL 2
A48619 QC 2 levels; 3 x 2 mL each
Testosterone 33560 Reagent 2 x 50 tests 100
33565 Calibrator 6 x 2.5 mL 5
Total ßhCG (5th IS) / A85264 Reagent 2 x 50 tests 100
Diluted Total ßhCG B11754 Calibrator 6 x 4.0 mL 8
(5th IS)
Total ßhCG*, ***/ 33500 Reagent 2 x 50 tests 100
Diluted Total ßhCG 33505 Calibrator 6 x 4.0 mL 8
# Tests or #
Calibrations
Assay Panel Assay Part # Description Volume
(@ 500 µL/
Calibration)
Thyroid Free T3 A13422 Reagent 2 x 50 tests 100
A13430 Calibrator 6 x 2.5 mL 5
Free T4 33880 Reagent 2 x 50 tests 100
33885 Calibrator 6 x 2.5 mL 5
hTSH, Fast / 33820 Reagent 2 x 50 tests 100
HYPERsensitive††† 33825 Calibrator 6 x 4.0 mL 8
Thyroglobulin 33860 Reagent 2 x 50 tests 100
33865 Calibrator 6 x 2.0 mL 4
33866 Diluent 1 x 14.0 mL
Thyroglobulin A32898 Reagent 2 x 50 tests 100
Antibody II A36920 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
Total T3 33830 Reagent 2 x 50 tests 100
33835 Calibrator 6 x 4.0 mL 8
Total T4 33800 Reagent 2 x 50 tests 100
33805 Calibrator 6 x 4.0 mL 8
Thyroid Uptake 33810 Reagent 2 x 50 tests 100
33815 Calibrator 1 level; 6 x 1.0 mL 12
TPO Antibody A12985 Reagent 2 x 50 tests 100
A18227 Calibrator 6 x 2.0 mL 4
TSH (3rd IS)* / B63284 Reagent 2 x 100 tests 200
Diluted TSH (3rd IS) B63285 Calibrator 6 x 2.5 mL 5
Tumor Markers AFP / Dil-AFP 33211 300 test kit 6 x 50 tests 300
(Cals included) Cals: 7 x 2.5 mL 5
33210 * 100 test kit 2 x 50 tests 100
(outside US only)
33215 * Calibrator 7 x 2.5 mL 5
(for 100 test kit)
33216 Diluent 1 x 14.0 mL
BR Monitor 387620 Reagent 2 x 50 tests 100
(cancer antigen 15-3) 387647 Calibrator 6 x 1.5 mL 3
CEA 33200 Reagent 2 x 50 tests 100
33205 Calibrator 6 x 2.5 mL 5
33206 Diluent 1 x 4.0 mL
33209 QC 2 levels; 3 x 2.5 mL each
GI Monitor 387687 Reagent 2 x 50 tests 100
(cancer antigen 19-9) 387688 Calibrator 6 x 2.5 mL 5
Hybritech PSA 37200 Reagent 2 x 50 tests 100
37205 Calibrator 6 x 2.5 mL 5
37206 Diluent 1 x 14.0 mL
37209 QC 3 levels; 5.0 mL each
# Tests or #
Calibrations
Assay Panel Assay Part # Description Volume
(@ 500 µL/
Calibration)
Tumor Hybritech free PSA 37210 Reagent 2 x 50 tests 100
Markers, 37215 Calibrator 6 total; S0 @ 5.0 mL 5
(continued) S1–S5 @ 2.5 mL
37219 QC 2 levels; 5.0 mL each
Hybritech p2PSA A49752 * Reagent 2 x 50 tests 100
A49753 * Calibrator 7 x 2.1 mL 4
B03704 Reagent 2 x 50 tests 100
B03705 Calibrator 7 x 2.1 mL 4
A56934 QC 3 levels; 5.0 mL each
OV Monitor 386357 Reagent 2 x 50 tests 100
(cancer antigen 125) 386358 Calibrator 6 x 2.5 mL 5
Research Use IL-6 (RUO) A30945 Reagent 2 x 50 tests 100
Only (RUO) A30944 Calibrator 6 total; S0 @ 4.0 mL 5
S1–S5 @ 2.5 mL
A30946 QC 3 levels; 2 x 2.5 mL each
AMH (RUO) B69047 Reagent 2 x 50 tests 100
B69048 Calibrator 6 x 2.0 mL 4
B69049 QC 3 levels; 2 x 2.0 mL each
* Not available in the U.S.
† Distributed by Beckman Coulter for Bio-Rad for use on Beckman Coulter immunoassay systems. Bio-Rad is a
trademark of Bio-Rad Laboratories, Inc.
** These products are being replaced world wide with AccuTnI+3 as registrations are obtained. AccuTnI will be
discontinued, please check with your local Beckman Coulter Representative for product availability.
†† The AccuTnI reagent available in your country depends on the status of registration with local regulatory agencies.
*** These products are being replaced world wide with Total ßhCG (5th IS) as registrations are obtained. Total ßhCG
will be discontinued, please check with your local Beckman Coulter Representative for product availability.
††† These products are being replaced worldwide with TSH (3rd IS) as registrations are obtained. HYPERsensitive and
Fast hTSH will be discontinued. Please check with your local Beckman Coulter Representative for product
availability.
System Supplies
System supplies are organized alphabetically in two tables. The first table contains supplies ordered through
Customer Service. The second table contains supplies ordered through Technical Support.