HIOKI 3287 Manual

The HIOKI 3287 CLAMP ON AC/DC HiTESTER is designed for measuring AC and DC currents and voltages with specific safety standards and operational guidelines. It features a maximum rated voltage of 600 V AC/DC, auto power save function, and requires zero adjustment for accurate DC current measurements. Users must follow safety precautions to avoid electric shock and ensure proper operation, including checking lead connections and avoiding measurements beyond specified limits.

Uploaded by

이문곤
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
56 views1 page

HIOKI 3287 Manual

The HIOKI 3287 CLAMP ON AC/DC HiTESTER is designed for measuring AC and DC currents and voltages with specific safety standards and operational guidelines. It features a maximum rated voltage of 600 V AC/DC, auto power save function, and requires zero adjustment for accurate DC current measurements. Users must follow safety precautions to avoid electric shock and ensure proper operation, including checking lead connections and avoiding measurements beyond specified limits.

Uploaded by

이문곤
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Effect of radiated radio- Current measurement: 4 A within NOTE: The auto power save function cannot be canceled.

frequency electromagnetic Voltage measurement: 15 V or less A minute amount of power continues to flow while in the power save state. If you will not
DANGER
field (in 3 V/m) If the end of a test lead short-circuits lines with a voltage between them,
Crest factor 2.5 (current range: 150A max., voltage range: 1,000 V max.)
be using the tester for an extended period of time, set the function switch to OFF or
Location for use Altitude up to 2000 m (6562 feet), indoors
remove the battery. this is very dangerous and can lead to a serious accident. Exercise great
Maximum conductor 35 mm (1.38") „ Zero-adjust Function care when measuring voltages.
diameter for measurement Before measuring DC current ( ), you must perform zero adjustment by simultaneously The maximum rated voltage to ground is 600 V AC/DC.
3287 Operating temperature and 0 to 40 (32 to 104 ), 80%rh or less (no condensation)
humidity
pressing the (Ω ) and HOLD keys while there is no input to the unit.
The zero adjustment function compensates for sensor magnetization and changes in
Do not attempt to measure voltages exceeding 600 V with respect to
ground. This could result in injury or damage to the unit.
Storage temperature -10 to 50 (14 to 122 ) (no condensation)
CLAMP ON AC/DC HiTESTER Temperature
characteristics
In 0 to 40 range: 0.1 x Measurement accuracy /
(In 32 to 104 range: 0.56 x Measurement accuracy / )
current display over time. This function is only effective with measurement of DC current
( ). Check that the test lead plug is firmly connected to the tester before beginning
measurement.
INSTRUCTION MANUAL Power supply Rated supply voltage 3 V DC x 1 CR20 32 x 1 L ithiu m b atte ry
NOTE: Please do not perform zero adjustment while there is any input to the unit. Also
note that the zero-adjust function will not function when the display count is greater than Measuring AC Voltage ( V)
Maximum rated power 15 mVA
1000. 1. Set the function switch to V/ V.
Continuous operating time 25 hours or more (continuous, no load)
June 2001 Revised edition 2 Printed in Japan Dimensions and mass 57(W) x 180(H) x 16(D) mm, approx. 170 g „ Auto-range Function 2. Connect the test leads to the object to be measured. When measuring AC voltage, the
When measuring an AC current ( A), DC current ( A), AC voltage ( V), DC voltage polarity of the leads can be ignored.
3287A980-02 01-06H 2.24"(W) x 7.09 "(H) x 0.63"(D), approx. 6.0 oz.
Accessories Instruction Manual, 9398 CARRYING CASE, 9208 TEST LEADS ( V), or resistance (Ω), the measurement range is automatically set to the most Measuring DC Voltage ( V)
Options 9209 TEST LEADS HOLDER appropriate range. 1. Set the function switch to V/ V.
Introduction Standards accuracy Safety: „ Manual Range Function 2. Press V V key to display .
EN 61010-1:1993+A2:1995 Power on the tester while holding down the Ω key or HOLD key to select a manual 3. Connect the red (+) lead to the + side of the circuit to be measured and the black (-)
Thank you for purchasing this HIOKI 3287 CLAMP ON AC/DC HiTESTER. Current measurement (ACA, DCA) Pollution Degree 2 Overvoltage lead to the - side. A negative reading will result if the leads are reversed.
To get the maximum performance from the unit, please read this manual first, and keep Category (CAT) III (anticipated transient overvoltage 6000 V) range for measuring AC current ( A), DC current ( A), AC voltage ( V), DC voltage
it at hand. Voltage measurement (ACV, DCV) Pollution Degree 2 Overvoltage ( V) or resistance (Ω). Note that this function is not available for continuity testing. Resistance Measurement (Ω)
Category (CAT) II (anticipated transient overvoltage 4000 V) Press the Ω key to step to the next range. To switch between AC voltage ( V)
Plug the test leads into the connector.
Inspection EN 61010-2-031:1994, EN 61010-2-032:1995
UL3111-1:1994, UL3111-2-32:1999, CAN/CSA-C22.2No. 1010.1-92+B-97,
and DC voltage ( V), press and hold the V V key for at least one second.
1. Set the function switch to Ω / .
When the unit is delivered, check that it has not been damaged in transit. If the unit is CAN/CSA-C22.2No. 1010.2.031-94, CAN/CSA-C22.2No. 1010.2.032-96 „ Overflow indication 2. Connect the test leads to the object to be measured.
damaged, or fails to operate according to the specifications, contact your dealer or HIOKI EMC: When the input exceeds the measurement range, "OF" or "-OF" is displayed.
representative. EN 61326-1:1997+A1:1998 Continuity Test ( )
If reshipping the unit, preferably use the original packing. Names and Functions of Parts Plug the test leads into the connector.
„ Before Use
Safety 1. Display 1. Set the function switch to Ω / .
In order to ensure safe operation and to obtain maximum performance from the unit, 2. Clamp sensor 2. 11. 10. 2. Press the Ω key, so that the indication appears.
Before using the unit, inspect it and check the operation to make sure that the 3.
sheathing on the leads is not damaged and that no bare wire is exposed. observe the cautions listed below. 3. Current direction indicator 3. Connect the test leads to the object to be measured. Conductivity is good when the
If there is damage, using the unit could cause electric shock. Contact your dealer or 4. Operation grip buzzer sounds.
HIOKI representative. DANGER 5. Function switch
This equipment is designed to comply with IEC 61010-1 Safety Standards, OFF / AC current ( A) / DC current Maintenance
4. 4.
Accuracy and has been tested for safety prior to shipment. During high voltage ( A) / Voltage ( / V) /
measurement, incorrect measurement procedures could result in injury or Resistance and Continuity check (Ω/ CAUTION
23 5 (73 9 ) 80% rh or less, no condensation ) 1. If the protective functions of the unit are damaged, either remove the unit from
death, as well as damage to the equipment. Please read this manual 6. Ω key Display:
AC current measurement: true RMS value, DC current measurement: average value service or post warnings to prevent others from using the unit inadvertently.
carefully and be sure that you understand its contents before using the V V key Measurement
Accuracy (%rdg. +dgt.) Max. allowable equipment. The manufacturer disclaims all responsibility for any accident or value 6. 7. Gently wipe dirt from the surface of the unit with a soft cloth moistened with a small
Function Range Measuring voltage amount of water or mild detergent. Do not try to clean the unit using cleaners
45 to 66 Hz 10 to 45, 66 to 1 kHz current injury except that resulting due to a defect in its product. Units
Switches between AC voltage ( V) Symbols containing organic solvents such as benzine, alcohol, acetone, ether, ketones,
ACA 10.00 A 100 Arms and DC voltage ( V) Decimal thinners, or gasoline. They may cause discoloration or damage.
( A) 100.0 A (1.5% +5) (2.0% +5)
continuous „ Overvoltage Categories Measuring resistance or continuity point
To ensure safe use of measurement, IEC 60664 establishes safety level When not in use for a long time, to prevent possible corrosion caused by battery
DCA 10.00 A 100 Arms Switches between resistance leakage, remove the battery before storage.
DC (1.5% +5) standards for different locations, classified as CAT I through CAT IV, and
( A) 100.0 A continuous measurement (Ω) and continuity 5.
called overvoltage categories. These are defined as follows. testing ( ). If the unit is not functioning properly, check the battery and test leads. If a problem is
Effect of conductor position: 1.0% (in any direction from sensor center) found, contact your dealer or HIOKI representative. Pack the unit carefully so that it
Voltage in measured circuit: max. 600 V AC rms (insulated conductor) CAT I: Secondary electrical circuits that are connected to a wall outlet Measuring DC current ( A)
Pressing this key together with 8. will not be damaged during transport, and write a detailed description of the problem.
through a transformer or similar device. HIOKI cannot bear any responsibility for damage that occurs during shipment.
AC voltage measurement: true RMS value, DC voltage measurement: average value CAT II: Primary electrical circuits in equipment connected to a wall outlet via HOLD key initiates zero adjustment. Battery cover (rear)
Range Accuracy Max.allowable 7. HOLD key 9.
Function a power cord (portable tools, household appliances, etc.) Replacing Battery
(Accuracy range) (%rdg. +dgt.) Input impedance voltage CAT III: Primary electrical circuits of heavy equipment (fixed installations)
Press this to hold the display value
4.200 V (0.420 to 4.200 V) 11 MΩ 5%
(the indication appears). Press
ACV 42.00 V (4.20 to 42.00 V) (2.3% +8) 10 MΩ 5%
connected directly to the distribution panel, and feeders between the once more to cancel the hold function. WARNING
600 Vrms distribution panel and outlets. Measuring DC current ( A)
( V) 420.0 V (42.0 to 420.0 V) 30 to 500 Hz 10 MΩ 5% If the unit is connected to a line that is to be measured, dangerous
600 V (420 to 600 V) 10 MΩ 5% CAT IV: The circuit from the service drop to the service entrance, then to the Pressing this key together with Ω key initiates zero adjustment. voltage levels may be applied to the terminals, and removing the case
420.0 mV (42.0 to 420.0 mV) 100 MΩ or more power meter and to the primary overcurrent protection device. 8. Connector 9.Plug
4.200 V (0.420 to 4.200 V) 11 MΩ 5% Connect the test lead plug to the connector for voltage measurement, resistance may expose live components.
DCV 42.00 V (4.20 to 42.00 V) (1.3% +4) 10 MΩ 5% 600 V DC Higher-numbered categories correspond to To avoid electric shock when replacing the battery, first disconnect the
( V) measurement, or continuity testing. Align the slot on the plug with the key in the
420.0 V (42.0 to 420.0 V) 10 MΩ 5% electrical environments with greater connector. unit and the test leads from the object being measured. Also, after
600 V (420 to 600 V) 10 MΩ 5% momentary energy, so a measurement 10. Red test lead (+) 11. Black test lead (-) replacing the battery, always replace the cover and tighten the screw
Resistance device designed for CAT III environments before using the unit.
Range Accuracy Overload can endure greater momentary energy than Measurement Procedures Use only CR2032 lithium battery. Use of any other battery may result in
Function (Accuracy range) (%rdg. +dgt.) Open terminal voltage protection a device designed for CAT II. Use of a
DANGER explosion.
420.0 Ω (42.0 to 420.0 Ω) (2.0% +4) 3.4 V or less lower category product in a higher category
4.200 kΩ (0.420 to 4.200 kΩ) (2.0% +4) 0.7 V (typ.) 3.4 V or less environment could result in a severe accident and must be carefully avoided. Before taking a measurement, check the position of the function switch. When replacing the battery, be sure to insert it with the polarity correct.
Ω 42.00 kΩ (4.20 to 42.00 kΩ) (2.0% +4) 0.47 V (typ.) 3.4 V or less 250 Vrms Setting the function incorrectly may damage the unit or cause an accident Battery may explode if mistreated. Do not short-circuit, recharge,
420.0
4.200
kΩ (42.0 to 420.0 kΩ)
MΩ (0.420 to 4.200 MΩ)
(2.0% +4)
(5.0% +4)
0.47 V (typ.) 3.4 V or less
0.47 V (typ.) 3.4 V or less
Notes on Operation that might result in injury or death. When changing the function, disassemble or dispose of in fire.
42.00 MΩ (4.20 to 42.00 MΩ) (10.0% +4) 0.47 V (typ.) 3.4 V or less disconnect the test leads from the object to be measured. Keep used batteries out of the reach of children. Dispose of used
Continuity
DANGER Do not input a voltage to the resistance measurement function and the batteries according to their type in the prescribed manner and in the
To avoid short circuits and accidents that could result in injury or death, continuity test function. Doing so may damage the unit or cause an proper location.
Accuracy Open terminal Overload
Function Range
(%rdg. +dgt.)
Buzzer
voltage protection use clamp testers only with power lines carrying 600 V AC or less. accident resulting in injury or death.
The maximum rated working voltage is 600 V AC/DC. Do not measure NOTE: When the battery is exhausted, the indication
Continuity 420.0 Ω (2.0% +6) 50 Ω 30 Ω 3.4 V or less 250 Vrms At the time as the continuity test, to avoid electrical accidents, turn off appears in the display.
voltage in excess of these limitations, as doing so may damage the unit the power before measuring a circuit. 1. Remove the unit and the test leads from the test item, Battery cover
Safety Symbols or cause an accident that might result in injury or death.
NOTE: The frequency of a distorted waveform, such as on the secondary side of an and power the unit off.
Always connect the clamp sensor or test leads to the secondary side of a inverter, may not be indicated correctly. 2. Remove the unit from the case, and remove the screws
This symbol is affixed to locations on the equipment where the operator breaker. On the secondary side of a breaker, even if the lines are shorted on the battery cover. Battery
should consult corresponding topics in this manual (which are also the breaker can trip and prevent an accident. On the primary side, Current Measurement 3. Remove the used battery.
marked with the symbol) before using relevant functions of the 4. Being careful about the polarity, insert the new battery of
equipment. however, the current capacity may be large, and in the event of a short-
In the manual, this symbol indicates explanations which it is particularly circuit there may be a serious accident. WARNING the specified type. (CR2032 lithium battery)
The maximum permissible input is 100 A. Current measurements 5. Replace the battery cover and fasten the screws.
important that the user read before using the equipment.
WARNING exceeding 100 A should be of short duration. NOTE: The battery included with this unit was inserted for
Indicates a device which is double-insulated. TESTING PURPOSES ONLY. Battery life will vary. Please
Indicates DC (Direct Current). To prevent electric shock, do not allow the unit to become wet and do During current measurement, to avoid an electric shock accident, do not replace the original battery with a new CR2032 lithium battery as soon as it is depleted.
Indicates AC (Alternating Current). not use the unit when your hands are wet. connect the test leads to the unit. CR2032 lithium batteries can be purchased at electronics and appliance stores where
Indicates both DC (Direct Current) and AC (Alternating Current). To avoid electric shock accidents, when carrying out measurement on NOTE: Clamp the tester on one lead only. specialized batteries are sold.
Indicates a grounding terminal. live lines, wear proper protective gear, including insulating rubber gloves, Measuring AC Current ( A)
The following symbols are used in this Instruction Manual to indicate the relative importance of cautions insulating rubber boots, and safety helmet, and use extreme caution. 1. Set the function switch to A. Using the 9209 Test Lead Holder (Option)
and warnings.
2. Clamp the tester on the conductor, so that 2. 1. Remove the battery cover, and in its
Indicates that incorrect operation presents extreme danger of CAUTION the conductor passes through the center of
DANGER accident resulting in death or serious injury to the user. place fit the test lead holder (option).
Do not use or store the unit where it is exposed to direct sunlight, high temperatures, the clamp core. Fasten the screws securely.
Indicates that incorrect operation presents significant danger of
WARNING accident resulting in death or serious injury to the user.
high humidity, or condensation. If exposed to such conditions, the unit may be
damaged, the insulation may deteriorate, and the unit may no longer satisfy its
OK No 2. Insert the test lead probe into the test
lead holder.
Indicates that incorrect operation presents possibility of injury to specifications. 1. 1.
CAUTION the user or damage to the equipment. Before using the unit, inspect it and check the operation to make sure that the unit Measuring DC Current ( A)
Denotes items of advice related to performance of the equipment was not damaged due to poor storage or transport conditions. If damage is found,
NOTE or to its correct operation. contact your dealer or HIOKI representative.
1. Set the function switch to A.
-
2. After making sure that there is not input to the unit, perform 2.
Before using the unit, make sure that the sheathing on the leads is not damaged zero adjustment by simultaneously pressing the (Ω ) and +
Specification and that no bare wire is exposed. If there is damage, using the unit could cause HOLD keys.
electric shock. Replace the lead with the specified 9208. 3. Clamp the line to be measured so that the arrow on the side of
Zero-adjust Function Before measuring DC current ( ), you must perform zero adjustment by
simultaneously pressing the (Ω ) and HOLD keys. NOTE: Accurate measurement may be impossible in locations subject to strong external the clamp sensor points in the direction of current flow and the
LCD panel 4199 maximum display value magnetic fields, such as transformers and high-current conductors, or in locations subject line is position in the center of the sensor jaws. (A negative
Out of range indication OF or -OF to strong external electric fields, such as radio transmission equipment. reading will result if the arrow points in the opposite direction.)
HEAD OFFICE
Battery low warning Voltage Measurement 81 Koizumi, Ueda, Nagano 386-1192, Japan
Data hold indication Functions and Display TEL +81-268-28-0562 / FAX +81-268-28-0568
Zero suppression 5 count or less (current only) „ Auto Power Save Function DANGER E-mail: [email protected]
Display update rate 400 ms 25 ms This function automatically switches to the power save state when 10 minutes have The maximum overvoltage protection input is 600 V AC/DC. When
HIOKI USA CORPORATION
Range switching Auto range / Manual range elapsed since the last operation. measuring CATIII lines, 300 V rms.
6 Corporate Drive, Cranbury, NJ 08512, USA
Withstand voltage 3.7 kV rms sine wave (for 1 minute ) between case and circuit The auto power save function is activated automatically when the power is turned on. Do not measure voltage in excess of these limitations, as doing so may TEL +1-609-409-9109 / FAX +1-609-409-9108
5.55 kV rms sine wave (for 1 minute ) between clamp sensor and case
5.55 kV rms sine wave (for 1 minute ) between clamp sensor and circuit To restore from the auto power save state, turn the function switch to the OFF position damage the unit or cause an accident that might result in injury or death.
once. URL:http://www.hioki.co.jp