Andor Kymera 328i Specifications
Andor Kymera 328i Specifications
andor.oxinst.com
1 Adaptive Focus*
Adaptive Focus:
Automated scan provides
optimum focus
Automated optimization for the best quality of focus:
2 TruRes TM
3 Quad Turret with RFID
Expand your system’s flexibility
True spectral resolution enhancement option
Combine up to 4 gratings for greater flexibility in one
• Better than 30% spectral resolution improvement single setup:
without the need to change grating or slit width • Spectral resolution: Choice of high, medium or low
• Expands the range of spectral resolutions accessible on a options
single setup at the touch of a button • Blaze: Choice of UV, Visible, NIR or SWIR options
• Superior discrimination of complex spectral features from • Mirror for microspectroscopy
UV to SWIR
eXpressIDTM: RFID –based intelligence ensures
• No mathematical spectral deconvolution required automatic recognition and upload of all important
turret parameters to the spectrograph.
Ne I 754.404 nm
Ne I 754.404 nm
Fully
Resolved
749 751 753 755 749 751 753 755
Wavelength (nm) Wavelength (nm)
Kymera 328i with 300 l/mm grating, iVac 316 with high resolution 15 µm pixels, full vertical binning.
2
Features and Benefits
Feature Benefit
Pre-aligned, pre-calibrated instrument Individually characterized spectrograph-detector systems for out-of-the box operation.
High repetition rate shutter 10 Hz continuous operation and 40 Hz burst mode for ultrafast acquisition.
Simultaneous control of Andor cameras, spectrographs and a wide range of microscopes and
µ-Manager software integration accessories through 1 single software platform. Dedicated, user-friendly spectrum handling
interface.
Integrated in EPICS •20 Integration and operation at EPICS-based large research facilities.
3
Step-by-Step System Configuration
How to customise the Kymera 328i: 1
4
3 2
Resolution Boost
2
Quad Turret
On-axis Rotation
Auto Focus
1 Detector 1
2 Detector 2
1
1 Chassis configuration
a) Select combination of input and output ports (see page 5 for available options).
b) Select type of optics coating required (aluminium + MgF2 is standard, protected silver-coated optics available on
request for NIR detection).
c) Select purge port option (for improved detection down to 180 nm). Shutter for background acquisition and protection
of the detector.
5 Software interface
Select either state-of-the-art Solis software or Software Development Kit (SDK) option – please refer to the appropriate
section for further information.
4
Step 1 - Chassis Configuration
Ordering Information
Side input Direct input Direct output Side output Motorized port
Model
port port port port selection
KYMERA-328i-xx-SIL Protected silver-coated optics options for models shown above (replace x with relevant model number)
75
200 400 600 800 1000 1200 1400 1600 1800 2000
Wavelength (nm)
Chassis Accessories
USB Cable (Standard)
I2C option via camera
Additional Grating Turret •1
(SR-ASZ-10398)
Purge Connector
(SR-ASM-8040)
iStar CCD and sCMOS I2C to BNC
shutter cable (ELC-05323)
TruResTM option*
Direct input: SR-IRIS-DIR
Side input: SR-IRIS-SIDE
Shutters•2 * Not field upgradeable - please
Side input: SR-SHT-9006 contact your local Andor
(or SR-SHT-9006-FIELD for field upgrade only) representative for information
Adjustable feet (standard, set of 4)
Direct input: SR-SHT-9007
(or SR-SHT-9007-FIELD for field upgrade only) 6 mm spacer kit (set of 4 fixed
spacers, SR-ASM-0098)
5
Step 2a - Choosing The Right Platform vs Dispersion Requirements
Resolution calculator
andor.com/calculators
Green
Aberration-free region
Orange
Possible impact on
system resolution
Red
Likely impact on
system resolution
Czerny-Turner spectrographs are designed to provide the best optical performance for a range of grating angles as
reflected on the green parts of the graph above. Outside this range, the spectral lines may exhibit a degree of optical
aberration (such as coma), which will become more prominent at the steeper angles. These configurations are reflected by
the orange to red scales on the graph. In these regions, consideration should be given to higher spectrograph focal length
models with lower groove density gratings to achieve the desired resolution.
Grating (l/mm)
Kymera 193i
Bandpass (nm)•3,•5 902 445 215 98 56 46•6
Kymera 328i
Bandpass (nm)•3,•5 542 268 131 61 41 29•6
Shamrock 500i
Bandpass (nm)•3,•5 357 177 86 40 26 19•6
Shamrock 750
Bandpass (nm)•3,•5 242 120 59 28 18 14•6
Where aberration is a concern for a particular experimental set-up, the table above shows resolution and band-pass
performance for a variety of alternative configurations. This should be used in conjunction with the graph above to assist in
selecting the most appropriate spectrograph platform to meet resolution and band-pass needs, whilst minimising the risk
of potential aberration.
6
Step 2b - Choosing The Right Grating vs Resolution and Band-pass
The Kymera 328i features an on-axis, quadruple grating turret, designed to offer flexibility and control
over your choice and interchange of gratings. The ‘Quad’ grating turret can be easily and speedily
removed, and replaced by an alternative turret with new gratings. The intelligent design of the 328i with
xPressIDTM RFID technology, means that only a simple offset adjustment is required once the new turret
and gratings are added. The 328i is shipped with the grating turret already in place, ensuring your system
is ready for use straight out of the box. Additional grating turrets are available with up to four pre-installed
gratings (see below for details). If the grating you require is not on the list, please contact Andor for further
details. Additional grating turrets (part number SR-ASZ-10398) can also be supplied on request.
Holographic
1200 2.19 61 0.100.07 81 SR-GRT-1200-EH*
(500 nm peak)
Holographic
1800 1.30 36 0.060.04 70 SR-GRT-1800-DH
(250 nm peak)
570
Holographic
1800 1.52 42 0.070.05 62 SR-GRT-1800-FH
(380 nm peak)
Holographic
2400 1.12 31 0.050.04 68 SR-GRT-2400-BH
(220 nm peak) 425
Holographic
2400 0.95 26 0.040.03 73 SR-GRT-2400-GH
(400 nm peak)
7
Step 2c - Selecting The Correct Grating Efficiency Option
All graphs shown below represent efficiency for 45° polarisation
ciency (%)
ciency (%)
60 60
Efficiency
Efficiency
50 50
40 40
30 30
20 20
1000 nm
10 10
500 nm
0 0
200 500 800 1100 1400 1700 2000 2300 200 500 800 1100 1400 1700 2000 2300
Wavelength (nm) Wavelength (nm)
1200 nm 70
ncy (%)
60 60
Efficiency
Efficiency
50 50 1000 nm
40 1000 nm
40
30 30
20 20
10 10
0 0
200 500 800 1100 1400 1700 2000 2300 200 400 600 800 1000 1200 1400 1600 1800
ciency (%)
Efficiency
50 50
40 40
30 30
20 20 300 nm
10 10
0 0
200 400 600 800 1000 1200 200 350 500 650 800 950
Wavelength (nm) Wavelength (nm)
Important Consideration
System throughput is dependent on the grating’s angle of operation and may decrease with higher grating operating angles.
Need to have maximum collection efficiency in the NIR/SWIR? All gratings are also available with protected silver coating.
Please contact your local representative for further information.
8
Step 3 - Selecting The Correct Light Coupling Interfaces
How to customize the Kymera 328i (Side Entrance Port): Side input port (applicable to all models)
Fixed FC Fixed SMA Motorized Slit Assembly•13 Wide Aperture Slit Manual Slit Assembly•13 X-Y Adjustable Fibre
Fibre Adapter Fibre Adapter (SR-ASZ-0035) (SR-ASZ-0086) (Standard) Adapter
(SR-ASM-8011) (ACC-SR-ASM-8003) inc 6 x 4 mm (W x H) inc Ø27 mm Cover Plate inc 6 x 4 mm (W x H) (See page 12 Section B:
Cover Plate Cover Plate Direct X-Y fibre couplers)
SMA Fibre
(50 µm: ACC-ME-OPT-8004) Cover Plate for Motorized Slit Cover Plate for Manual Slit
(100 µm: SR-OPT-8039) (See page 12 Section A: Slit Covers) (See page 12 Section A: Slit Covers)
F/# Matcher for X Adjustable Fibre Adapter, Fixed Fibre Adapter, X-Y Adjustable Optical Cage F-Mount 1.5” Flange Adapter C-Mount Adapter Pen-Ray Lamp Mount
NA = 0.22 Fibre Ferrule Input Ferrule Input Fibre Adapter System Adapter Camera Lens for Newport Oriel (SR-ASM-0021) (SR-ASM-0014)
(SR-ASM-0039)•11 (SR-ASM-8006) (SR-ASM-8001) (See page 11 Section C: (SR-ASM-0065) Adapter Accessories
X-Y fibre couplers (SR-ASM-0013) (SR-ASM-0002)
with slit assembly)
andor.com
9
Step 3 - Selecting The Correct Light Coupling Interfaces
How to customize the Kymera 328i (Direct Entrance Port):
SMA Fibre
(50 µm: ACC-ME-OPT-8004)
Cover Plate for Motorized Slit Cover Plate for Manual Slit
(100 µm: SR-OPT-8039)
(See page 10 Section A: Slit Covers) (See page 12 Section A: Slit Covers)
F/# Matcher for X Adjustable Fibre Fixed Fibre X-Y Adjustable Optical Cage F-Mount 1.5” Flange C-Mount Adapter Pen-Ray Lamp
NA = 0.22 Fibre Adapter, Adapter, Fibre Adapter System Adapter Camera Lens Adapter for (SR-ASM-0021) Mount
(SR-ASM-0038)•11 Ferrule Input Ferrule Input (See page 12 Section (SR-ASM-0065) Adapter Newport Oriel (SR-ASM-0014)
(SR-ASM-8006) (SR-ASM-8001) C: (SR-ASM-0013) Accessories
(SR-ASM-0002)
X-Y fibre couplers
with slit assembly)
andor.com
10
Step 4 - Cameras and Output Port Flanges
How to customize the Kymera 328i: Side Output Port (Applicable to B and D models)
Spacer
(Standard with B1 and D1 models)
Manual Adjustable Slit Output Port X-Y Adjustable Fibre Adapter Multi-channel
Assembly•13 Motorized Slit Assembly•13 (See page 12 Section B: Detector Flange
(Standard) (SR-ASZ-0036) Direct X-Y fibre couplers) (MFL-SR-CCD
inc 6 x 4 mm (W x H) inc 6 x 4 mm (W x H) To be ordered separately,
Multi-channel Detector one per output port as
Cover Plate Cover Plate
Flange required)
to Slit Adapter Plate•14 Suitable for Newton CCD iKon-M Mounting Flange
(ACC-MEC-09741) & EMCCD, iDus CCD & (MFL-SR-IKON-M)
InGaAs, iVac. iXon ULTRA Mounting Flange
Suitable for iStar on side (MFL-SR-IXON)
output port (use MFL-SR- iStar Mounting Flange (direct output
ISTAR-DIRECT for direct port) (MFL-SR-ISTAR-DIRECT)
output port). ZL41 Wave Mounting Flange
(MFL-SR-ZYLA)
Cover Plate for Cover Plate for
Manual Adjustable Marana Mounting Flange
Manual Adjustable Slit Motorized Slit Assembly
Slit Assembly•13 (MFL-KY-MARANA)
Assembly (See page 12 Section A:
(SR-ASZ-0030) inc 6 x 4 mm (Note: Ø 32 mm aperture cover slit
(See page 12 Section A: Slit Covers)
(W x H) cover Plate recommended to avoid vignetting
Slit Covers)
for larger sensor of the Marana)
Optical Cage
Cover Plate for
C-Mount Adapter System Adapter X-Y Adjustable Fibre
Manual Adjustable Slit Assembly
(SR-ASM-0021) (SR-ASM-0065) Adapter,
(See page 10 Section A: Slit Covers)
(See page 12 Section B:
Direct X-Y fibre couplers)
Cage System
(Please refer
to Thorlabs or Linos catalogue)
Note: a flange MUST be ordered separately for any configuration involving a multichannel or InGaAs detector.
andor.com
11
Step 4A: Slit Covers Step 4B - X-Y Fibre Coupler (with NO slit)
Size Motorised Slit Manual Slit SMA-Ferrule SR-OPT-80XX
Ferrule SR-ASM-8057
Spacer Upgrade to other fibre interface: ACC-yy-
6 x 4 mm (W x H) SR-ASM-0016•14 SR-ASM-0025 DIRECT-APT
Spectrograph
6 x 6 mm (W x H) SR-ASM-0017 SR-ASM-0026 input port SMA-SMA fibre
6 x 8 mm (W x H) SR-ASM-0010 SR-ASM-0027 SMA SMA: SR-ASM-8054
SR-ASM-8054
Filter wheel Upgrade to other fibre interface: ACC-yy-
(if applicable) DIRECT-APT
6 x 14 mm (W x H) SR-ASM-0011 SR-ASM-0029•14
FC/APC-FC/APC fibre
FC/APC SR-ASM-8055
Upgrade to other fibre interface: ACC-yy-
DIRECT-APT
andor.com
12
Step 5 - Selecting A Software Option
The Kymera 328i requires at least one of the following software options:
1 - Solis Spectroscopy A 32-bit and fully 64-bit enabled application for Windows (8, 8.1, 10 and 11) offering rich
functionality for data acquisition and processing, as well as Andor cameras, spectrograph and motorized accessories
simultaneous control. AndorBasic provides macro language control of data acquisition, processing, display and export.
2 - Standalone Solis Spectroscopy GUI for standalone spectrograph operation.
3 - Kymera and Shamrock SDK A software development kit that allows you to control the Andor range of Kymera and
Shamrock spectrographs from your own application. Compatible as 32-bit and 64-bit libraries for Windows (8, 8.1, 10 and
11). Compatible with C/C++, C#, VB.NET and LabVIEW for Windows/Linux.
13
90.0 4 OFF M3 TAPPED
Product Dimensions
13.5 [0.531] DEEP ATTACHMENT
3.543 TO DIRECT EXIT
2.047
52.0
Dimensions in mm [inches]
9.152 ±0.118
232.5 ±3.0
5.665 ±0.118
5.665 ±0.118
143.9 ±3.0
143.9 ±3.0
71.6
326.1 2.820 VIEW A
12.839 20.6°
131.6 20.0
4 OFF M4 TAPPED
ATTACHMENT TO SIDE EXIT 5.181 0.787
1.503
38.2 NITROGEN PURGE
1/8" NPT PLUG
5.665 ±0.118
143.9 ±3.0
38.2
1.503
E
NC
TA
14.863
12.835
DIS
377.5
326.0
0.748
10.0
19.0
13.0
FOC 4
AL
0.394
10.0
0.39
0
TURRET LID 0.512
SLIT BLADE LOCATION
0
17.6
0.416
0.694
10.6
A
5.665 ±0.118
143.9 ±3.0
202.6
7.978
308.6
106.6 12.151
4.198 396.6
15.615
Shutter Specifications
40 Hz - burst; 10 Hz -
Maximum repetition rate
sustained
Magnification 1.1:1
USB Control
Connector type: USB ‘B’ type
Wavelength Drive Performance
I2C Connector
For connection to camera/detector via I2C connection Wavelength accuracy center •17 0.04 nm
(replaces USB cable connection to spectrograph).
Wavelength repeatability •18
Shutter Control Single grating 4 pm
Connector type: BNC Female, 50 Ω Grating-to-grating 10 pm
14
Our Cameras for Spectroscopy
Spectroscopy-based diagnostics in the fields of Material Science, Chemistry, Life Science or Fundamental
Physics & Optics rely on the capture and analysis of optical and chemical signatures with a high degree of
precision.
Andor’s range of detectors offer a wide range of sensitivity, time-resolution and sensor formats to best suit
specific experimental conditions from UV to SWIR, nanosecond to hours time resolution, high photon flux to
single photon with super dynamic range and resolution.
iDus CCD & InGaAs | Newton CCD & EM iStar CCD & sCMOS
Newton CCD & EMCCD | iXon EMCCD | iKon-M CCD | iXon EMCCD | ZL41 Wave sCMOS
ZL41 Wave sCMOS | Marana sCMOS | Marana sCMOS | iStar CCD & sCMOS
15
Order Today
Need more information? At Andor we are committed to finding the correct solution
for you. With a dedicated team of technical advisors, we are able to offer you one-
to-one guidance and technical support on all Andor products.
For a full listing of our local sales offices, please see: andor.oxinst.com/contact
Footnotes: Specifications are subject to change without notice Items shipped with your spectrograph:
1. In the case of a multiple grating turret order, please specify desired grating configuration for each turret. 1x 3 m USB 2.0 cable Type A to Type B
2. Shutter operation can be achieved directly through the I2C interface between cameras and spectrograph, or 1x Power supply (+24V, 5A) with 3 m mains cable
through a BNC-to-SMB cable when the spectrograph is operated through USB. 1x I2C to I2C cable
3. Typical values quoted with 27.6 mm wide CCD, e.g. Newton DU940. 1x Andor user guides in electronic format
4. Typical values quoted with 10 μm slit and 13.5 μm pixel CCD, e.g. Newton DU940. Illustrates resolutions 1x Individual system performance booklet
achievable with iris opening range (optional). 1x Solis software or SDK in electronic format
(if requested at time of order)
5. Typical values quoted at 500 nm centre wavelength.
1x hex key set (2 mm, 3 mm and 5 mm)
6. Typical values quoted at 300 nm centre wavelength.
7. Typical values quoted at maximum efficiency wavelength or blaze wavelength unless otherwise stated. Regulatory Compliance
8. Wavelength within the recommended operating spectral region. Compliant with the requirements of the EU EMC and
9. Indicative values; the working range of these gratings is principally in the region where optical aberrations may LVD Directives, compliant with the international EMC
alter the system resolution performance quoted. and safety standards IEC 61326-1 and IEC 61010-1,
10. Useful signal is assumed to be imaged on the entire height of a 6.9 mm sensor (i.e. Newton DU940) and fully and Machinery Directive 2006/42/EC.
vertically binned. Minimum Computer Requirements:
11. Please refer to F/# matcher specification sheet for magnification considerations. • 3.0 GHz single core or 2.4 GHz multi core
12. Please refer to the local sales representative or website for further information on available options and processor
complimentary accessories. • 2 GB RAM
13. Slit widths range from 10 μm to 2.5 mm. • 250 MB free hard disc to install software (at
14. Provided as standard. least 1 GB recommended for data spooling)
• USB 2.0 High Speed Host Controller capable of
15. Recommended for use with fibre-optics and C-mount accessories.
sustained rate of 40 MB/s
16. Measured with a 633 nm laser and a 1200 l/mm grating for Full Vertical Binning (FVB) on a 6.9 mm high sensor, • Windows (8.1, 10 and 11)
and a 1 mm strip vertically centred on the optical axis.
17. Average measurements using > 30 calibration lines, covering the recommended grating angle operating range
with a 1200 l/mm grating.
18. The standard deviation of 20 measurements of a peak’s centre-of-mass position: - each measurement is taken
after switching back and forth between a given centre wavelength and a lower or higher centre wavelength
(single grating) or between two gratings set at the same centre wavelength (grating-to-grating).
19. Side accuracy measured using a 27.6 mm wide sensor, reflecting the dispersion calibration and step-and-glue
accuracy.
20. Only Andor CCD platforms (e.g. Newton, iDus, iKon) can be controlled in conjunction with Kymera and Shamrock
spectrographs in EPICS software.
Operating and Storage Conditions Power Requirements Windows is a registered trademark of Microsoft Corporation.
• Operating Temperature: stable ambient • 100 - 240 VAC 50 - 60 Hz
LabVIEW is a registered trademark of National Instruments.
between 0°C to 30°C • Max. power consumption: 21 W (10 Hz
• Relative Humidity: < 70% (non-condensing) shutter and grating turret operation) MATLAB is a registered trademark of The MathWorks Inc.
• Storage Temperature: -25°C to 50°C
SSKymera328iSS 0425 R1