Es MAN DAT400 Utilcell
Es MAN DAT400 Utilcell
Digital transmitter
TECHNICAL MANUAL
For Analog/ProfiBus/DeviceNet versions
INTRODUCTION........................................................................................... Page 3
INSTALLATION............................................................................................... Page 5
ADJUSTMENT........................................................................................ Page 18
USB CONNECTION BETWEEN THE DAT 400 AND THE PC................................................... Page 47
Page 1
PRECAUTIONS
READ this manual BEFORE using or performing maintenance on the instrument.
Follow these instructions carefully.
SAVE this manual for future use.
ATTENTION
Only the installation and maintenance of
this instrument to qualified personnel.
Take extreme precautions when carrying out inspections, tests and
settings with the instrument turned on.
Make the electrical connections without the power supply.
Failure to comply with these instructions can be dangerous.
Do not let untrained personnel work, clean, inspect,
repair or manipulate the instrument.
Page 2
INTRODUCTION
The DAT 400 is a weight transmitter that combines with load cells to detect weight.
in any situation.
The module is easy to install and can be mounted on a 35 mm DIN rail.
The screen allows easy reading of the weight, the status of the instrument, the adjustment parameters and the
errors.
The 4 keys located below the screen allow the operator to use the ZERO functions.
TARE (Tara), GROSS/NET (Gross/net), change them, adjust the consignment weight, the values and the tare.
both theoretical and real.
The DAT 400 instrument uses the RS232 serial port with ASCII and Modbus RTU protocols for
connect to a PC, PLC, and remote units. In parallel with RS232, there is available a
USB port.
There are always 2 programmable weight settings available and control of the maximum weight value.
reached (peak).
The RS422/RS485 serial output allows you to connect up to 32 addressable devices.
The availability of the most common fieldbus protocols, as an alternative to the RS422/RS485 port,
also allows the transmitter to intercommunicate with any monitoring device that is
find in the market.
Versiones disponibles:
DAT 400: weight transmitter with RS232 (USB) serial output, RS485, and peak function. Supports the
Modbus RTU protocols, continuous, slave and on request. Two programmable outputs, 2 inputs.
and a peak function.
DAT 400/A: version with the analog output.
DAT 400/PROFIBUS: weight transmitter with RS232 serial output and PROFIBUS DP.
DAT 400/DEVICENET: weight transmitter with RS232 serial output and DEVICENET.
It is important to communicate this data in case information regarding the instrument is requested, along with
with the number and version of the program indicated on the cover of the manual and that are displayed in
the instrument screen when it is on.
WARNINGS
Only qualified personnel should carry out the following procedures.
All connections must be made when the instrument is turned off.
Page 3
TECHNICAL CHARACTERISTICS
Power supply 24 V DC 15 %
Max absorption 5W
Isolation Class II
Installation category Cat. II
Operating temperature -10°C ÷ +40°C (humedad máx. 85 % sin condensación)
Storage temperature -20°C ÷ +50°C
Weight screen 6-digit red LED numeric and 7-segment (height 14 mm)
Led 4 mm 3 LED
Keyboard 4 mechanical keys
General dimensions 106 mm x 58 mm x 90 mm (l x al x an)
Installation Supports for the DIN profile section or OMEGA bar
Box material Noryl self-extinguishing (UL 94 V1)
Connections Screw terminal plates, depth 5.08 mm
Power supply load cells 5 V CC/120 mA (max. 8 cells of 350 W in
parallel), with short-circuit protection
Input sensitivity 0.02 mV min.
Linealidad 0.01% of the scale range
Temperature deviation 0.001% of the full scale/°C
Internal resolution 24 bits
Resolution of the shown weight Up to 60,000 divisions in net capacity
Measurement range -0.5 mV/V to +3.5 mV/V
Weight capture frequency 5 Hz - 50 Hz
Digital filter To select between 0.2 Hz and 25 Hz
Number of decimal places of the weight 0 ÷ 3 decimal places
Zero calibration and full scale Automatic (theoretical) or executable from the keyboard.
Logical outputs 2 opto-isolated (dry contact), max. 24 VDC / 60 mA
each one
Logical entries 2 opto-isolated at 24 VDC PNP (power supply
external)
Serial port (No. 2) RS232C, RS422/RS485 and USB (in parallel)
Maximum cable length 15 m (RS232) and 1000 m (RS422 and RS485)
Serial protocols ASCII, Modbus RTU
Transmission speed 2400, 9600, 19200, 38400, 115200 for
select
Memory of program codes 64 Kbyte FLASH on reprogrammable RS232 board
Data memory 2 Kbytes
Page 4
INSTALLATION
GENERAL DATA
The DAT 400 instrument consists of a base plate to which available options can be added.
It is located inside a plastic housing on a 35 mm DIN rail.
The DAT 400 should not be submerged in water, subjected to water jets or cleaned or washed with
solvents.
Do not expose it to heat or direct sunlight.
GENERAL DIMENSIONS
90
PRECISE
DAT 400
8. 8. 8. 8. 8. 8.
1 2 NET
46
SET FUN 0 PRG
106 58
ELECTRICAL INSTALLATION
The DAT 400 transmitter uses screw terminals, 5.08 mm deep, for the co-
Electrical connection. The load cell cable must be shielded and placed far away.
from the tension cables to avoid electromagnetic interference.
Page 5
LOAD CELL CONNECTIONS
The cable of the cells should not be placed next to other cables.
The instrument can be connected to a maximum of 8 load cells.
connect the 350 ohms in parallel. The power supply of the
The cells are 5 VDC and are protected against short circuits.
temporaries.
The measurement range of the instrument involves the use of cells of
load with a sensitivity of up to 3.5 mV/V.
The cable from the load cells must be connected to terminals 13-
18. In the case that the load cell cable has 4 wires,
make a bridge between terminals 13 - 16 and 14 - 15.
18 10 9 Connect the shield of the cell cable to terminal 9.
If using two or more load cells, use boxes of
PRG 0 FUN
special splices. Find your connection below.
+EXC 6 14 EXC+ C IN 6
SHD 7
EXC 1 +EXC
-EXC 2 -EXC
+SGN 3 +SGN 10 mA +
-SGN 4 -SGN 11 V +
+24V
+EXC 1 EXC S.GND 25
-EXC 2 -EXC 8 + RXD 26
24
+SGN 3 +SGN 9 - TXD 27
-SGN 4 -SGN
SHD 5 SHD
+EXC 1 +EXC
EXC 2 -EXC
+SGN 3 +SGN
-SGN 4 -SGN
SHD 5 SHD
LOGICAL ENTRIES
The two logic inputs are opto-isolated.
The cable that connects the logic input should not be placed next to
with the power cables.
+ INPUT 1
This is the function of the two inputs:
- INPUT 2
COM. INPUT
ENTRY 1 Reset the displayed value (gross, net, or peak)
ENTRY 2 PRINT
To activate the two functions, the power supply must be connected.
alimentación externa de 24 VCC a los bornes correspondientes,
as shown in the figure.
Page 6
LOGICAL OUTPUTS
OUTPUT 1
The two opto-isolated relay outputs are normally open contacts. OUTPUTS
OUTPUT 2
24 Vdc
open. The capacity of each contact is 24 V DC, 100 mA max. 100 mA Max COM. OUTPUT
The cable that connects the outputs should not be placed together with the
power cables. The connection should be as short as possible.
COMMUNICATION SERIES
RS232:
The RS232 serial port is always present and works with various
TXD
protocols.
RXD
To use the serial connection, use a shielded cable and make sure S.GND
RS485:
The RS485 serial port (2 wires) is included in the DAT 400 model.
RS485.
To achieve the serial connection, use a suitable shielded cable and
make sure to connect the shielding to one of the two ends: to the pin S.GND
TXD+
23 if connected on the instrument side or to ground if connected TXD-
on the other side. RXD+
RXD-
The cable should not be placed together with the power cables.
Page 7
PROFIBUS DP CONNECTION (OPTIONAL)
Page 8
DEVICENET CONNECTION (OPTIONAL)
Pin Signal Description
1 V- Negative power bus
2 CAN_L Low CAN bus line
3 SHIELDING Screened
2 CAN_L 4 CAN_H High CAN bus line
4 CAN_H 5 V+ Positive power bus
1 V- To connect to the DeviceNet master, use a DeviceNet cable
5 V+ standard or a shielded twisted pair cable, as it moves
straight in the diagram.
3 SHIELD
The cable should not be placed together with the power cables.
For the Fieldbus to operate reliably, it must be used
to a line termination with a value of 121 Ω between the terminal
CAN_L and CAN_H.
To configure the card, there is an available ESD file that
it must be installed on the master.
Page 9
FRONT PANEL OF THE INSTRUMENT
The DAT 400 device has a bright 6-digit display, 4 status LEDs, and four keys.
En este modo de funcionamiento, la pantalla indica el peso y los LED indican el estado del peso y las
instructions.
It is very easy to access the configuration parameters and modify them using the three
front buttons to select, edit, confirm, and save the new settings.
PRECISE
DAT 400
8. 8. 8. 8. 8. 8.
1 2 NET
SET FUN 0 PRG
SCREEN
On the 6-digit screen, the weight of the scale usually appears. According to the diver-
These programming procedures, the screen is used to program the parameters that will be
store in memory or the messages that indicate the type of operation being performed and, for
thus, they assist the operator in the management and programming of the instrument.
LED INDICATORS
Under the screen, there are 4 LED indicators:
1 State of the logical output 1 (ON = contact closed OFF = contact open)
2 Logical output state 2 (ON = contact closed OFF = contact open)
The value displayed is the net weight
0 Indicate the state of the stable weight.
Page 10
USE OF THE KEYBOARD
The instrument is programmed and controlled through the keyboard, which has 4 keys with dual functions.
The instrument automatically sets the selection of one of the key functions.
agreement with the ongoing operation. In general, the management of the programming menus is
use the SET and FUN keys to navigate through the options; the PRG key is used for
enter its submenu or programmable parameter, while button 0 allows you to exit the menu or
return to the upper level.
SET
Access to the menu for programming guidelines
PRG SET
(Press 3 seconds) Access to the settings menu.
+
PRG 0 (Press 3 seconds) Allows access to the lock/unlock menu of the key-
+ turn on the automatic screen off function.
FUN
Allows you to select the previous parameter.
0
Allows you to exit the programming menu or return to the upper level.
0
Allows to pass to the next digit.
PRG
Confirm the displayed value.
Page 11
KEY FUNCTION WHEN DEFINING NUMERIC VALUES
SET
Allows selecting the next value.
FUN
Allows selecting the previous value.
PRG
Confirm and save the displayed value.
0 0
Press the key to return to the main menu. Press the key again .
"StORE?" appears.
PRG
Press the key to return to the main menu.
PRG
To exit without saving any changes, turn off the instrument instead of pressing the key. .
Page 12
INFORMATION SCREEN
When the instrument is turned on, a screen test is performed, and then, in sequence, you can see the
Software identification code and its version. Communicate the codes in case of requesting assistance.
ERROR NOTIFICATION
In operation mode, the following error codes are displayed on the screen.
The weight applied to the load cell exceeds the maximum capacity by more than 9 divisions.
of the weighing system.
O-L There is no signal from the load cells or they are out of the measurement range mV/V.
The Fieldbus network is disconnected.
E-ProF No PROFIBUS interface or not working.
There is no DEVICE NET interface or it is not working.
ϩϩϩϩϩϩScript that appears around the perimeter of the screen: the BLIND function has been activated.
Memory error. Press the PRG key to reset the memory and recover the values.
default parameters. NOTE: the calibration performed is also erased.
Page 13
VISUALIZATION, ZEROING OF WEIGHT AND CALIBRATION
After calibration, when the device is turned on, the current weight appears on the screen.
VISUALIZATION OF NET/GROSS WEIGHT
FUN
Press the key to switch between net weight and gross weight. The displayed value is marked
through the NET LED (when it is on it indicates the net weight). If you have not entered the tare, the weight
neto is equivalent to gross weight.
In case the weight is negative, the minus sign appears in front of the most significant digit.
AUTOMATIC CALIBRATION
The automatic calibration can be carried out under the following conditions:
Instrument in 'net' conditions (the 'NET' LED is on).
Positive gross weight.
The gross weight does not exceed the maximum capacity.
Stable currency.
• Unstable weight. In this condition, we must distinguish two cases:
1. The weight stability control is activated (the parameter "MOTION" (*) must not be equal to
zero): the command that is executed while the weight is unstable only takes effect if the weight takes
to stabilize for 3 seconds from the moment the command is issued.
2. Weight stability control is disabled (the parameter "MOTION" (*) is equal to zero):
The executed command is applied immediately, even with unstable weight.
The operating modes of the "MOTION" parameter are described on page 27.
The automatic calibration is stored in memory even after turning off the ali-switch.
feeding.
RESET
The gross weight reset command is used to correct small zero drifts of the system.
of weighing during normal operation.
Normally, these zero drifts are due to thermal drifts or material residues that are
they accumulate in the weighing system over time.
To execute the command, it is necessary for the instrument to be in 'raw' conditions (with the
LED "NET" off) and that the weight deviation from the scale's zero (the one that is done
with the zero calibration) does not exceed (neither positively nor negatively) the number of divisions defined in
the "0 BAND" parameter (in the PARAM menu).
The command to reset the gross weight does not execute if one of the following occurs
conditions:
Page 14
• Unstable weight (with weight stability control activated). In this case, the command of
Restoration only takes effect if the weight stabilizes in 3 seconds or if the control is deactivated.
of the stability of the peso (the parameter 'MOTION' is equal to zero).
The gross weight is greater (positively or negatively) than the number of divisions defined in the param-
Metro "0 BAND" if the automatic calibration setting has not been programmed.
The zero obtained with the reset of the gross weight is saved in memory.
even when the device is turned off.
The gross weight reset operation can be repeated multiple times, but the reset
from zero divisions accumulates so that when the total exceeds the limit value defined in the parameter
"0 BAND", the zeroing cannot be executed. In this case, it is necessary to calibrate the zero.
When adjusting any automatic calibration parameter upon turning on the instrument (AUTO 0), it reduces
(or cancels, in the case of “AUTO 0”> “0 BAND”) the range of action of the reset command.
PEAK FUNCTION
The instrument continuously records the peak value of the gross weight. The peak value is detected at
the same frequency of weight acquisition (see the table on filters). In addition to visualization,
The peak value can be used in the following functions:
FUNCTION DESCRIPCIÓN
The set points can be adjusted to reference the peak value.
LOGICAL OUTPUT Consult the procedure to adjust the operations of the outputs
logical) .
Acquisition of peak value (peak retention) through the protocols
SERIES PORT
["CONTIN","AUTO","DEMAND","SLAVE","MODBUS"]
EXIT The analog output value can take on the value of the peak (old peak).
ANALOG (Refer to the procedure to adjust the analog output).
Hold down the 'FUN' key for 3 seconds until the left side of the screen
the letter "P" appears.
INPUT/OUTPUT FUNCTIONS
ENTRY
Resetting the displayed value (gross weight, net weight or peak). Closed for 5
1 second -> Zero calibration, which is only performed if the corresponding function is
activated in the PARAM menu (see point "0 ALL").
2 Sending the weight chain in the series line or printing.
EXIT
1 Instruction 1
2 Directive 2
Page 15
SCHEDULING OF WEIGHT MANDATES
Los valores de consigna definidos se comparan con el peso para obtener la salida lógica. El criterio
comparison is established in the process of setting up the I/O logic (see section
corresponding).
To access the setpoint adjustment, press the SET key and follow the instructions in the table below.
SET
Set 1 PRG
123456 PRG
Increment Change
digit selected digit
FUN
Set 2 PRG
123456 PRG
0 SET FUN 0
Increment Change
EXIT selected digit
During the setpoint adjustment phase, both outputs are deactivated. If the setpoint value
in memory is 0, the corresponding output never activates despite the configuration of the
selected consignments. When the weight is not detected or is out of range, all are deactivated.
outputs (open or closed contact depending on the MODE; see the corresponding chapter).
Page 16
KEYBOARD LOCK/UNLOCK FUNCTION
KEYBOARD LOCK/UNLOCK Function that allows you to activate or deactivate the keys of
individually. When the keys are locked, the only way to access these settings
is to keep the PRG and 0 keys pressed for 3 seconds. For more information about the
function, review the above block diagram.
KEYS CORRESPONDING
3 sec.
PRG + 0 SET FUN 0 PRG
Lock PRG
0000 PRG
Change Select
selected digit digit
FUN
blind PRG
off PRG
SET FUN
0
on PRG
EXIT
00 PRG
SET FUN 0
Change Select
selected digit digit
SCREEN OFF This feature allows you to turn off the screen after a programmed time.
Mable. You can turn the BLIND parameter on or off and adjust the time; the time starts to
count from the moment when, after leaving the settings menu, it indicates on the screen the
value of the peso. After the defined time, the screen turns off and only a dash appears. This dash
traverse the perimeter of the screen in a counterclockwise direction. When the screen
it is off, the 4 keys are also deactivated, regardless of how the lock is set
from the keyboard (LOCK). The only way to access the settings is by pressing PRG + 0.
Page 17
ADJUSTMENT
GENERAL DATA
To activate and modify all the functions of the DAT 400, you need to open a simple configuration menu.
action, which is shown on page 21. All selected or activated settings are saved.
even if the transmitter goes off.
The DAT 400 comes preconfigured with a default setting. The following pages indicate
the default values of each parameter.
With the first local installation, it is necessary to change some parameters so that it is displayed the
correct weight (theoretical adjustment).
The settings menu adjustments can be changed using the front keys or through the
software "INOVATION 2" provided.
FUN
Allows you to select the previous menu.
0
Allows you to exit the programming menu or return to the upper level.
FUN
It allows reducing the value of the blinking digit.
0
Allows to proceed to the next digit.
PRG
Confirm and save the displayed value.
Page 18
KEY PR
ES
O
FUNCTION
POSITIONS
LOR
ASV
LOAL
IR N DE
FI
SET
Allows selection of the next value.
FUN
Allows selecting the previous value.
PRG
Allows to confirm and save the displayed value.
The menu parameters can have values that can be adjusted or selected.
0
NOTE To exit and save the modified data, press the key several times until in the
PRG
the screen appearsStorE. Then press the key to return to operational mode.
Page 19
MENU DIAGRAM
Page 20
MENU DIAGRAM
ENTER MENU
5 sec.
PRG
CoverC
SET FUN PRG
Enter Value PRG
Sens1t
SET FUN PRG
Enter Value PRG
nEt
SET FUN PRG
Enter Value PRG
dead L
SET FUN PRG
Enter Value PRG
dspd1J
SET FUN PRG
Select Value PRG
S1GnaL
SET FUN PRG
See Value 0
Caliber
CaL
PRG
SET FUN PRG
AnaLoG
0
CaL
SET
0
Store?
PRG
123456
SET FUN 0
EXIT MENU
60000
PRG
Page 21
CONFIGURATION PARAMETERS
By adjusting the parameters indicated below, theoretical calibration is carried out.
of the scale range of the DAT 400. These steps, along with the zero calibration described in the
Next page, must be carried out mandatorily. The procedure guarantees good accuracy.
of the system (maximum error of <1% FS) if there are no mechanical problems.
Program the known values of total capacity and sensitivity of the load cells and the values
approximately the net capacity and the calibration. If the SENSIT parameter is not programmed, it is used
the value 2.0000 mV/V.
If the CAPAC parameter is set to a value other than 0, according to the CAPAC data,
SENSIT, NET and DEAD L, the instrument automatically performs the following functions:
Restoration of linearization points.
Selection of the division value to be modified up to a maximum of 10,000 divisions.
Calibration of the approximate theoretical calibration of the weight (zero of the scale and full scale).
Automatic programming of the overload setpoint (= NET).
These functions are performed each time one of the 4 parameters displayed is changed.
When changing the DSPDIV selection, the calibration of the scale background is automatically recalculated.
the scale. It indicates whether the selections are incompatible with the calibration parameters or not.
accept the calibration in memory.
The instrument starts with the programmed selection in the operating mode so that the instruction-
It should work that way.
og Conf1G CaL1br
FUN FUN
PRG
load cell and 'N' fixed supports, enter the value of the layer-
SET FUN
Define the corresponding value for the mean sensitivity of the cells
Load in mV/V. The instrument accepts values between 0.5 and 4.
0
Counts
SET FUN
PRG
Read Value 0
theory of weight.
0
Upload Receive setup data
Values: from 0.5000 to 4.0000 mV/V SET FUN
PRG
to serial port
Page 22
NetWEIGHING SYSTEM CAPACITY
Programming of the net capacity of the weighing system. Not you
Conf1G CaL1br
FUN FUN
PRG
0
CapaC Values: from 1 to 500000
SET FUN
PRG
Enter Value PRG
Unit: the same as shown
0
SEns1t Default: 10000
SET FUN
PRG
Enter Value PRG
0
nEt DeadlyFIXED WEIGHT CALIBRATION OF THE WEIGHING SYSTEM
PRG
Enter Value PRG
SET FUN
Programming the fixed weight value of the weighing system.
0
dead L Values: from 1 to 500000
SET FUN
PRG
Enter Value PRG
0
S1GnaL dsPdIJDIVISION VALUE
SET FUN
PRG
Read Value 0
The ratio between the system's capacity and the value of the division
0
Counts represents the resolution of the system (number of divisions).
SET FUN
PRG
Read Value 0
divisions.
0
UpLoad Receive setup data
SET FUN
PRG
After changing the value of the division, if the maximum capacity
from serial port
It does not change, the weight calibration is automatically corrected.
0
dnLoad Transmit setup data
SET FUN
PRG
Value to select:
to serial port
0.0001 - 0.0002 - 0.0005
0.001 - 0.002 - 0.005
0.01 - 0.02 - 0.05
0.1 - 0.2 - 0.5
1 -2 - 5
10 - 20 - 50
Predeterminado: 1
Page 23
UploadDATA RECEPTION FUNCTION
Reception function from a serial communication of a file
Conf1G CaL1br
FUN FUN
PRG
EXAMPLE OF ADJUSTMENT/CALIBRATION
0
dead L
A tank with an empty weight of 750 kg will be weighed and with SET FUN
PRG
Enter Value PRG
of 0.2 kg screen.
0
S1GnaL
Before performing the configuration, make sure that the cells of SET FUN
PRG
Read Value 0
sensitivity2,0007
NEt1500
dead0
dsPd1J0.2
Make sure that the value read in the SIGNAL parameter
corresponds to the calibration weight of the system according to
with the following proportion:
3000:2007=750:X
Where X is the value of the signal expressed in mV/V corresponding
the theoretical weight of the empty tank. The value should be 0.5
mV/V approximately.
Carry out the calibration described in the following section or
go to the settings menu while saving the entered data.
The instrument should indicate the value corresponding to the weight of the
empty tank (for example, 756.8).
Reopen the settings menu and enter the value of
weight that has been read in the EadLe parameter, please enter the value
756.8
Exit the settings menu while saving the data.
To increase accuracy, prepare some sample weights.
the material previously weighed in a certified and calibrated system
the instrument as explained in the following section.
Page 24
CALIBRATION
The calibration described in this section must be performed with reference weights and/or products.
previously weighed on a weighing system.
Before performing the scale calibration, always perform the zero calibration.
During the calibration phase, the weight is displayed intermittently on the screen along with the
CaL registration
WARNING: if the instrument is turned off without exiting the settings menu, the program will not be saved.
executed information.
NOTE In case the system indicates linearity errors after calibration, it must
check that the heavy structure has no mechanical restrictions.
1G CaL1br ParaN
FUN FUN
PRG
The displayed weight resets and Cal and 0 appear on the screen. 0
alternative way. You can repeat this operation several more times. PRG
CaL LoadWeight
SET
Before calibration, load a reference weight into the system Enter Load
and I waited for it to stabilize; the value is displayed on the screen 120000 Value
weight. PRG
Press the SET key to adjust the weight. The screen indicates the
PRG
CAL
theoretical weight value with the first digit on the left blinking
I am. Use the arrow keys to enter the value of the actual weight
loaded in the system starting from the first digit that is
blinking. Proceed to the next digit by pressing 0. After confirming
the last digit (the one furthest to the right) with the key
PRG, the weight is corrected. CAL appears on the screen and is modified.
the value of the introduced real peso.
If the defined value is greater than the resolution offered by the instrument
I’m sorry, the weight is not accepted and a message appears on the screen.
of error for a few seconds.
This procedure can be repeated.
Press the PRG key again to go to the CaLIb menu
0
To exit the CaLIb menu, you need to press the key until
que aparezca el mensaje "store?" (¿Guardar?).
To save the new calibration and exit the configuration menu,
press the PRG key.
To cancel the zero and full scale calibration:
FUN +0 allows cancellation of zero calibration.
FUN + SET allows you to cancel the scale calibration.
Page 25
LINEARIZATION PROCESS
ig CaL1br Param
FUN FUN
When entering the reference weight value, please take into account
SET SET
PRG
It states that values greater than the full scale are not accepted.
PRG
Cal With unload
of the scale or lower than the previous point. A
35scale weight that is not stable. If the entered value is accepted, it pro-
0
000000
SET FUN 0
12000VEnatluere LPo1ad
PRG
0
Lin P2 PRG
LoadWeight P2
000000
SET FUN 0
24000VEnatluere LPo2ad
PRG
0
LIn P9 PRG
LoadWeight P9
000000
SET FUN 0
120000VEnatluere LPo9ad
PRG
0
Cal
Page 26
WEIGHING PARAMETERS
The parameters of this menu allow for the adjustment of data acquisition and screen refresh.
automatic or manual reset performed by the transmitter.
F1LtErWEIGHING FILTER
This parameter is used to adjust the refresh rate of the
br ParaNIn-oU
FUN FUN
PRG
0 50 Hz 25 Hz SET FUN
1 50 Hz 16 Hz 0
car 0 PRG
Enter Value PRG
2 25 Hz 8 Hz SET FUN
3 25 Hz 5 Hz 0
0 track
4 25 Hz 2.5 Hz SET FUN
PRG
Select Value PRG
5 10 Hz 1.5 Hz 0
0 band
6 10 Hz 1 Hz SET FUN
PRG
Enter Value PRG
7 10 Hz 0.7 Hz 0
0 ALL
8 5 Hz 0.4 Hz SET FUN
PRG
Select Value PRG
9 5 Hz 0.2 Hz
Default: 5
Auto0AUTO-ZERO ON STARTING
This parameter allows you to define the value of the maximum weight that
you can reset when the instrument is turned on.
This operation corresponds to a zero calibration of the system.
and it only runs if the weight is stable and below the value
defined.
Value from 0 to the value of the parameter CAPAC.
Default: 0
Page 27
0 trackZERO TRACKING
br ParaNIn-oUt
FUN FUN
SET SET This function allows you to perform a momentary zero calibration.
PRG
what compensates for the temperature deviation of weight.
0
f1LtEr When turning off the instrument, it automatically returns to calibration.
PRG
Select Value PRG
SET FUN
from the previous zero.
0
Not1on The maximum weight that can reset this parameter is 2%
SET FUN
PRG
Select Value PRG
0
0 track Value Change
SET FUN
PRG
Select Value PRG
0 Off control
0
0 band 1 0.5 div/s
SET FUN
PRG
Enter Value PRG
2 1 div/s
0
0 ALL
SET FUN
PRG
Select Value PRG
3 2 div/s
4 3 div/s
Default: 0
0 bandBAND ZERO
This parameter allows you to define the number of divisions that can be
reset it by pressing the front zero button or
through digital input 1.
Values: from 0 to 200
Default: 100
0 ALLZERO CAL
Activate the function that allows for zero calibration while maintaining
press the key 0 or digital input 1 for 5 seconds.
Value:
Off Function disabled
On Function activated
Default: Off
Page 28
INPUT/OUTPUT PARAMETERS
sequence:
SET SET
PRG
0
Node 1 NET The relay output is active in net weight mode
SET FUN
PRG
Select Value GROSS The relay output is active in gross weight mode.
PRG
Select Value
Default: GROSS (Gross)
PRG
Comparison with net weight, gross weight or peak. In this last one
Select Value PRG
In this case, the comparison is made with the last peak value obtained,
0
Hyst-1 even if the peak function is not active.
SET FUN
PRG
Enter Value PRG
0
Hyst-2 Hyst-1HYSTERESIS OF THE CONSIGN 1
SET FUN
PRG
Enter Value PRG
0
t1NEr2 Hysteresis value of the defined setpoint.
SET FUN
PRG
Enter Value PRG
0
tEst1n PRG Read and 0
t1NEr1TIMING OF INSTRUCTION 1
Value of time, in tenths of a second, during which, if one
SET FUN
change status
0
tstoUt Read and
it surpasses the defined value of the weight, the output relative to the instruction
1 remains activated.
PRG 0
SET FUN
change status
After this time, although the value of the peso continues to be
above the command, the output is automatically disabled.
The function does not activate if the scheduled time is equal to zero.
Value: from 0 to 999
Default: 0
dELay1DELAY OF ASSIGNMENT 1
Value of time, in tenths of a second, after which, if
exceeds the defined value of the peso, the output relative to the order
1 remains activated.
The function does not activate if the scheduled time is equal to zero.
Value: from 0 to 999
Default: 0
Page 29
NodE2OPERATION MODE OF MESSAGE 2
NIn-oUt SEr1a
FUN FUN
PRG
sequence:
0
Node 1
NET The relay output is active in net weight mode SET FUN
PRG
Select Value
PRG
GROSS The relay output is active in gross weight mode. Select Value
PEAK The relay output is active in peak mode PRG
Select Value
Default: GROSS PRG
Select Value
STABL Output 2 is active with stable weight PRG
Default: Norml
0
Hyst-2
SET FUN
PRG
Enter Value PRG
0
tEst1n Read and
t1NEr2TIMING OF INSTRUCTION 2 SET FUN
PRG
change status
0
exceeds the defined weight value, the output related to the assignment SET FUN
change status
2 remains active.
After this time, even if the value of the peso continues to be below
above the instructions, the output is automatically deactivated.
The function does not activate if the scheduled time is equal to zero.
Value: from 0 to 999
Default: 0
dELay2DELAY OF TASK 2
Value of time, in tenths of a second, after which, if you
exceeds the defined weight value, the output relative to the task
2 remains activated.
The function does not activate if the scheduled time is equal to zero.
Value: from 0 to 999
Default: 0
Page 30
tEst1nTEST PROCEDURE FOR LOGICAL INPUTS
NIn-oUt SEr1a
FUN FUN
0 = input deactivated
0
Node 1
SET FUN
PRG
Select Value 1 = entrada activada
PRG
Select Value Enable and disable the entries to check the status.
PRG
corresponding to the screen. During this procedure, the function
Select Value PRG
0
t1NEr1 tstoUtLOGICAL OUTPUT TEST PROCEDURE.
SET FUN
PRG
Enter Value PRG
Select Value PRG During this procedure, the normal function of the outputs does not
0
Hyst-2 it is active. Use this procedure only to verify
SET FUN
PRG
Enter Value PRG
the hardware.
0
t1NEr2
SET FUN
PRG
Enter Value PRG
0
dELay2
SET FUN
PRG
Enter Value PRG
0
tEst1n PRG Read and 0
SET FUN
change status
0
tstoUt PRG Read and 0
SET FUN
change status
Page 31
OUTPUT PARAMETERS SERIES
This menu allows you to define the serial ports COM1 and COM2, as well as the communication parameters.
The instrument has two independent serial ports:
COM1 with RS232 or RS422/RS485 interface
COM2 with the optional FIELDBUS interface.
FUN
115200
Default: 9600
Prot-1PROTOCOLO DE COM1
Define how the RS232 serial port will be used.
None: the serial communication is disabled
Continuous: continuous transmission of the weight chain. It can be used,
for example, to communicate with a weight repeater. Consult
the details in the corresponding section.
Demand: when the operator presses the front button or uses the en-
Trade 2, a weight chain is transmitted. The command is accepted.
if the weight is stable. Between two consecutive transmissions, the weight
It must have a variation of at least 20 divisions.
Automatically transfers a weight string when
the peso stabilizes at a value higher than the minimum peso (20 divisions
Between two consecutive transmissions, the weight must have a
variation of at least 20 divisions.
Slave: ASCII protocol. Refer to the details in the corrections section.
corresponding.
Page 32
FUN
Modbus: MODBUS RTU protocol (slave). It is only used if PROT-
SET SeRiaL THERE
2 is configured the same as NONE. Refer to the details in the section
PRG
corresponding. Selectable communication parameters:
0
bas r n-8-1
SET FUN
n-8-2
baud2r E-8-1
SET FUN
o-8-1
Default: n-8-1
Prot-1 data transfer to the printer.
SET FUN
Value to select:
prot-2 None
SET FUN
Continue
Demand
Autom-
Slave
Modbus
Print
Default: None
Prot-2COM2 PROTOCOL:
Define the mode of use of Fieldbus
None: the serial communication is disabled
PROFIB: PROFIBUS fieldbus (if there is an optional card)
DEVNET:DEVICENET fieldbus (if there is an optional board)
Value to select:
None
Profib
Devnet
Default:
None
Page 33
AddrEsDIRECTORATE OF COMMUNICATION SERIES OF COM1 FUN
CENET.
Value: from 0 to 63
Pr-Add
SET FUN
Default: 01
DEVICENET
DelayDELAYED RESPONSE FROM SLAVE PROTOCOLS AND dN-Add
MODBUS RTU OF COM1 SET FUN
This value is expressed in milliseconds and represents the delay with SET FUN
the instrument sends the response to the request received from the
master.
Value: from 0 to 999 milliseconds
Default: 000
Page 34
0
rEN-Co rEN-CoREMOTE COMMUNICATION
PRG
Remote PRG
SET FUN
Allows communication with a PC for making adjustments through
connection
0
data F of the PC program
SET FUN
PRG
Enter Value PRG
0
t-NAP dataFREMOTE COMMUNICATION
PRG
trasN-
COM1 series protocol parameters (parity, data bits, stop bits)
SET FUN
End-oh 0
dELNapDATA RESET
This function allows you to retrieve the default mapping of
instrument. While it recovers, ERASE is shown.
Page 35
ANALOG PARAMETERS - ANALOG OUTPUT (ONLY DAT 400/A)
F-SCaLSCALE FUND
aL AnaLoG ConF1
FUN FUN
SET
PRG
corresponding to the net weight, the gross weight or the peak value. 0
tEst PRG
Select Value PRG
NET (Neto) 0
ranGE PRG
Select Value PRG
GROSS
SET FUN
PEAK (Pico) 0
offset PRG
SET FUN
Default: GROSS
-0- PRG
0 SET FUN
Set Offset
AnZEroZERO VALUE OF THE ANALOG OUTPUT -Fs- PRG
Page 36
offsetDRIFT ADJUSTMENT (CALIBRATION)
0
offset
SET FUN
PRG
Measure the output value of the analog signal with a multimeter to perform
-0- PRG
the zero (0) calibration and full scale of the scale (FS).
0 SET FUN
Set Offset
-Fs- PRG
SET FUN
Use the keys y to adjust the analog output.
Hold down the key to change quickly.
0
Press the key to switch between zero drift and the
scale background.
PRG
Press the key to return to the OFFSET submenu. Press the key
0
to return to the ANALOG menu.
Page 37
SERIAL COMMUNICATION PROTOCOLS
Page 38
DESCRIPTION OF THE CONTROL FORMAT:
Double quotes contain constant characters (keep the uppercase and lowercase); the sim-
bolos <y> contain variable numeric fields.
REQUEST FOR NET AND GROSS WEIGHT AND THE CURRENT PEAK
Maestro: <Addr> "N" EOT
DAT 400: “N” <Addr> <status> <net> <gross> <peak> ETX <chksum> EOT
CHANGE IN GROSS WEIGHT
Master: <Addr> “C” “L” EOT
DAT 400: <Addr> "C" "L" ACK EOT
CHANGE IN NET WEIGHT
Master: <Addr> “C” “N” EOT
DAT 400: <Addr> "C" "N" ACK EOT
RESET COMMAND,AUTOMATIC CALIBRATION OR PEAK RESET
Maestro: <Addr> "A" "A" EOT
DAT 400: <Addr> “A” “A” ACK EOT
PROGRAMMING OF TWO WEIGHT ASSIGNMENTS
Master: <Addr> “S” <s1> <s2> ETX <csum> EOT
DAT 400: <Addr> “S” ACK EOT
REQUEST FOR SCHEDULED ASSIGNMENT
Master: <Addr> “R” EOT
DAT 400: <Addr> 'R' <s1> <s2> ETX <csum> EOT
STORAGE OF WEIGHT DATA PERMANENTLY
Master: <Addr> 'M' EOT
DAT 400: <Addr> “M” ACK EOT
In case of a communication error or if a DAT 400 command is not recognized,
will respond with the following string:
DAT 400: <Addr> NAK EOT
DESCRIPTION OF THE FIELDS
Double quotes contain constant characters (keep uppercase and lowercase); the sim-
bolos <y> contain variable numerical fields.
addr = Address of the series communication of the instrument; it is the ASCII character that is obtained
adding 80h to the address number (for example, address 1: <Addr> = 80h + 01h = 81h).
checksum = checksum of the data in the string. It is calculated by performing the exclusive OR
(XOR) of all characters in <Addr> up to ETX, excluding the last one; the result of XOR is
break into 2 characters considering separately the upper 4 bits (first
character) and the 4 lower bits (second character); the 2 characters obtained are encoded to
continuation in ASCII (example: XOR = 5Dh; <csum> = "5Dh", that is 35h and 44h).
ETX (end of text) = 0x03h,
EOT (fin de la transmisión) = 0x04,
ACK (acknowledgment) = 0x06h,
NAK (without confirmation) = 0x15h.
Page 39
<status> = un carácter ASCII que acepta los siguientes valores:
"S" = stable peso
“M” = weight that is not stable (in motion)
"O" = weight greater than the maximum capacity
"E" = weight that cannot be detected.
... = 6 ASCII characters of prompt.
<net weight> = 6 caracteres ASCII del peso neto.
<gross weight> = 6 caracteres ASCII del peso bruto.
<peak> = 6 caracteres ASCII del pico.
If the request is made cyclically, the weight is obtained with a maximum frequency of:
Speed of
Frequency
transmission
200 Hz 115200
50 Hz 38400
35 Hz 19200
25 Hz 9600
8 Hz 2400
Page 40
MODBUS RTU PROTOCOL
The addresses indicated in the following tables use the standard address specified.
in the Modicon PI-MBUS-300 guidelines. Below is an excerpt that assists the user
to communicate with the instrument.
All the data addresses of the Modbus messages have zero as a reference. To the
The first occurrence of a data item is assigned the element number zero. For example:
The coil referred to as 'coil 1' in a programmable controller is coil 0000 in the
data address field of a Modbus message.
Coil 127 decimal is coil 007E hex (126 decimal).
The exploitation registry 40001 is the registry 0000 in the data address field.
from the message. In the function code field, a 'registration' operation is already specified.
exploitation." Therefore, the reference "4XXXX" is implicit.
To confirm a new value in EEPROM, run the MAKE – BACKUP function. If this function does not
By turning off the instrument, the DAT will restore the value it had before the change.
If not specified otherwise, numerical values (such as addresses, codes, and data) are expressed
as decimal values.
For any hardware configuration of the instrument (FIELDBUS or analog), the protocol
MODBUS RTU is always available on COM1 RS232; in case there is no Fieldbus, the pro-
MODBUS RTU protocol is also available on COM2 RS485.
Código Descripción
1 INVALID FUNCTION (the function is not valid or is not supported)
INVALID DATA ADDRESS (the address of the specified data
2
not available)
3 INVALID DATA VALUE (the received data has an invalid value)
Page 41
ADMITTED FUNCTIONS
Function Description
01 READ COIL STATE (Reading the status of the logical outputs)
02 READ INPUT STATE (Reading the state of the logical inputs)
03 READ OPERATIONAL RECORDS (Reading the programmable records)
04 READ ENTRY RECORDS (Reading the records of 'read-only')
05 FORCE SINGLE COIL (Writing the status of each output)
06 PREDEFINED UNIQUE REGISTER (Writing a programmable register)
15 FORCING VARIOUS COILS (Writing several outputs)
16 PREDEFINED MULTIPLE RECORDS (Writing multiple records)
Func + 80h EXCEPTION RESPONSE
Page 42
LIST OF MODBUS PROTOCOL OPERATION RECORDS
Direction. Exploitation registration R/W Note Format
40001 State registry R INT See table A
40002 Gross weight (MSB) R
DINT
40003 Gross weight (LSB) R
40004 Net weight (MSB) R
DINT
40005 Net weight (LSB) R
40006 Peak value (MSB) R
DINT
40007 Peak value (LSB) R
40008 Load cell signal mV/V R INT
40009 Logical inputs R INT LSB = Entrada 1
40010 Exit R/W INT LSB = Salida 1 (solo escribe si la consigna = 0)
40011 Key States R INT See table B,
40012 Firmware code and version R INT See table C
40201 Instruction 1 (MSB) R/W
DINT
40202 Instruction 1 (LSB) R/W
40203 Assignment 2 (MSB) R/W
DINT
40204 Instruction 2 (LSB) R/W
40501 Data Record (MSB) W Data related to command logging
DINT
40502 Data Registration (LSB) W
40503 Command register W INT See table D
41001 Cell capacity (MSB) R/W
DINT
41002 Cell Capacity (LSB) R/W
41003 Cell sensitivity R/W INT
41004 Weight division value R/W INT See table E
41005System Tara (MSB) R/W
DINT
41006 System Side (LSB) R/W
41007 System capacity (MSB) R/W
DINT
41008 System capacity (LSB) R/W
41101 Weight Filter R/W INT 0-9
41102 Stability of the weight R/W INT 0-4
41103 Zero-auto command in % (MSB)R/W 0-100 %.
DINT
41104 Auto-zero assignment in % (LSB)
R/W
Zero Tracking R/W INT 0-4
41106 Zero band in divisions R/W INT
41201 Mode of operation SET 1 R/W INT See table F
41202 Hysteresis defined 1 (MSB) R/W
DINT
41203 Hysteresis defined 1 (LSB) R/W
41204 Timer set 1 R/W INT
41205 Defined delay 1 R/W INT
41206 Operating mode SET 2 R/W INT See table F
41207 Hysteresis defined 2 (MSB) R/W
DINT
41208 Defined Hysteresis 2 (LSB) R/W
41209 Defined timer 2 R/W INT
41210 Defined delay 2 R/W INT
41401 Analog scale. FS (MSB) R/W
DINT
41402 Analog Scale FS (LSB) R/W
41403 Analog mode R/W INT See table G
41404 Analog Range R/W INT See table H
Analog output points (from 0 to 65535). The output
41405 Analog output value R/W INT
analog only uses this value if FS (41402) = 0
The scheduled value is copied automatically to the
42000 Monitor registration W INT
monitor R registration (42100)
Copy of the value entered in the monitor W register
42100 Monitor registration R INT
(42000)
Page 43
TABLE A: RECORD STATUS CODING
BIT 13 12 11 10 9 8 6 5 4 3 2 1 0
Indicator Block Out
Descrip- Exit Exit Entry Entry Load Load Tara Band Central Weight
of of of
action 2 1 2 1 excessive scarce introduced zero zero stable
memory keyboard range
WARNING: Bits 15, 14, and 7 are not managed and are always equal to 0.
Memory indicator (bit 9): when modifying a register that needs to be stored in the E2prom (consult the
table "Data stored in memory with command 0x020" on the next page), this bit is
Define in 1 to remind the user to execute command 0x0020 (to save the data in the
permanent memory). After performing this operation, the bit is automatically reset to zero.
Bit related to a weighing error (bit 6 of the STATUS REGISTER) When it is disconnected
cell or the measured values are out of range, this bit is defined as 1
Bit related to zero band (bit 2 of the STATUS REGISTER) When the gross weight is infe-
Greater than or equal to the parameter 'Zero band in divisions', bit 2 of the status register is set to 1.
Registration value 0 1 2 3 4 5 6 7 8
Division value 0,001 0,002 0,005 0,00 0,02 0,05 0,1 0,2 0,5
Registration value 9 10 11 12 13 14
Division value 1 2 5 10 20 50
Page 44
Modbus registration address Data stored in memory with command 0x0020
41001-41002 Cell capacity
41003 Sensitivity of cells
41004 Weight division value
41005-41006 System Board
41007-41008 System capacity
41101 Weight filter
41102 Stability of the peso
41103-41104 Zero self-insistence
41105 Zero tracking
41106 Zero band in divisions
41201 Defined operating mode 1
41202-41203 Defined hysteresis 1
41204 Timer set to 1
41205 Defined delay 1
41206 Defined operating mode 2
41207-41208 Defined hysteresis 2
41209 Timer set 2
41210 Defined delay 2
41401-41402 Scale background Analog scale
41403 Analog mode
41404 Analog range
Page 45
PRINTING PROTOCOL
This protocol allows communication with a printer. These are the printed report data: NET,
GROSS and TARE (+ PEAK, if activated).
PRG
To use the print command, you must press the key or activate remote input
number 2. (*)
The printed report is only generated if the following conditions are met:
Positive gross weight
Positive net weight
Stable weight (**)
• BLIND function disabled (***)
PRG
(*) In case the key it becomes blocked (see page 17), the print command only
can be used through remote input 2
The stability condition of the weight is only controlled if the 'MOTION' parameter is set (see page)
27) is defined at a value that is not 0.
If MOTION = 0, the printed report is generated even if the weight is unstable.
(*) In case the BLIND function is activated (see page 17), the print command
PRG
it can only be used through remote input 2, not through the key .
Between one print and another, the weight must change by at least 20 numbers.
PRINTER CONNECTION
Page 46
USB CONNECTION BETWEEN THE DAT-400 AND THE PC
The hardware module installed inside the DAT-400 that allows the instrument to in-
The device that communicates with a PC via the USB port is called CP210x and is manufactured by Silicon.
Laboratories.
The CP210x module requires drivers to be installed on the PC.
When installing the INNOVATION 2 utility software on PCs, the drivers will be loaded au-
automatically. For installation, refer to the manual.
Page 47
FIELD BUS PROTOCOL
The FIELDBUS data exchange takes place in two different areas of memory, which is described
I am in the following tables.
WARNING:
The 'input data area' and the 'output data area' are 128 bytes
To transfer the parameters of the output data area to the instrument, it is necessary to activate access.
directly to memory and write the hexadecimal value 0x7FFF in the command register (1 word
for writing) in order to prevent the instrument from resetting all its variables in case of
the output data area has not been initialized
This command must be sent before the connection to inform the instrument that the master has
Initialized the parameters. Now the instrument continuously tests the changes made.
in the parameters and only saves them in case of a real change.
ADDRESS
Byte Bit Description Format Note
0 0 Keyboard_Look BOOL See table A, page 44
0 1 Memory_Flag BOOL See table A, page 44
0 2 Input_1 BOOL See table A, page 44
0 3 Input_2 BOOL See table A, page 44
0 4 Output_1 BOOL See table A, page 44
0 5 Output_2 BOOL See table A, page 44
0 6 Spare BOOL See table A, page 44
0 7 Spare_1 BOOL See table A, page 44
1 0 Center of Zero BOOL See table A, page 44
1 1 Stable Weight BOOL See table A, page 44
1 2 Zero_Band BOOL See table A, page 44
1 3 Tare entered BOOL See table A, page 44
1 4 Under_Load BOOL See table A, page 44
1 5 Over_Load BOOL See table A, page 44
1 6 Off_Range BOOL See table A, page 44
2 0 Gross_Weight Dint
6 0 Net_Weight DINT
10 0 Peak Weight DINT
14 0 Load Cells Signal mV/V INT
16 0 Input_Status INT
18 0 Output_Status INT
20 0 Not used BYTE
21 0 SET_Key BOOL See table B, page 44
21 1 FUN_Key BOOL See table B, page 44
21 2 0_Key BOOL See table B, page 44
21 3 PRG_Key BOOL See table B, page 44
Page 48
22 0 Cod_Firmware BYTE See table C, page 44
23 0 Cod_Versione BYTE See table C, page 44
24 0 SP_1 DINT
28 0 SP_2 DINT
32 0 Load_Cell_Capacity DINT
36 0 Load_Cell_Sensitivity INT
38 0 Weight Division Value INT
40 0 Tare_of_the_System DINT
44 0 Capacity_of_the_System DINT
48 0 Filter_Value INT
50 0 Weight_Stability_value INT
52 0 Autozero_Set DINT
56 0 Tracking_Factor INT
58 0 Zero_Band INT
60 0 Not used BYTE
61 0 Set_1_Net_Weight BOOL See table F, page 45.
61 1 Set_1_Gross_Weight BOOL See table F, page 45
61 2 Set_1_NO_NC BOOL See table F, page 45
61 3 Set 1 Positive Negative Values BOOL See table F, page 45
61 4 Set_1_Control BOOL See table F, page 45
62 0 Set 1 Hysteresis DINT
66 0 Set 1 Timer INT
68 0 Set_1_Delay INT
70 0 Not used BYTE
71 0 Set_2_Net_Weight BOOL See table F, page 45
71 1 Set_2_Gross_Weight BOOL See table F, page 45
71 2 Set 2 No NC BOOL See table F, page 45
71 3 Set 2 Positive Negative Values BOOL See table F, page 45
71 4 Set 2 Control BOOL See table F, page 45
72 0 Set 2 Hysteresis DINT
76 0 Set 2 Timer INT
78 0 Set_2_Delay INT
80 0 Monitor_Register INT
READING EXAMPLE
To read the gross weight on the DAT 400, it is necessary to read addresses 2 to 5 of the input area.
To read the net weight, it is necessary to read addresses 6 to 9 of the input area.
When the screen indicates the gross weight value of 12351 in the corresponding bytes, it
you will see the following:
Page 49
OUTPUT DATA AREA
Address
Byte Bit Description Format Note
128 0 Command_Register INT See table D, page 44
130 0 Logic Output INT
132 0 SP_1 DINT
136 0 SP_2 DINT
140 0 Data_Register DINT
144 0 Load_Cell_Capacity DINT
148 0 Load_Cell_Sensitivity INT
150 0 Weight_Division_Value INT See table E, page 44
152 0 Tare of the System DINT
156 0 Capacity of the System DINT
160 0 Filter_Value INT
162 0 Weight_Stability_value INT
164 0 Autozero_Set DENT
168 0 Tracking_Factor INT
170 0 Zero_Band INT
172 0 Not used BYTE
173 0 Set_1_Net_Weight BOOL See table F, page 45
173 1 Set_1_Gross_Weight BOOL See table F, page 45
173 2 Set_1_NO_NC BOOL See table F, page 45
173 3 Set 1 Positive Negative Values BOOL See table F, page 45
173 4 Set_1_Control BOOL See table F, page 45
174 0 Set 1 Hysteresis DINT
178 0 Set_1_Timer INT
180 0 Set_1_Delay INT
182 0 Not used BYTE
183 0 Set_2_Net_Weight BOOL See table F, page 45
183 1 Set_2_Gross_Weight BOOL See table F, page 45
183 2 Set_2_NO_NC BOOL See table F, page 45
183 3 Set 2 Positive Negative Values BOOL See table F, page 45
183 4 Set_2_Control BOOL See table F, page 45
184 0 Set_2_Hysteresis DINT
188 0 Set 2 Timer INT
190 0 Set_2_Delay INT
192 0 Monitor_Register INT
WRITING EXAMPLES
To write the configuration parameters according to the example:
In byte 128 (command register), write the Hex value 7FFF. This value opens the writable area.
from DAT 400.
Page 50
Example: to change the default values of the DAT 400, such as the cell capacity.
de carga, la sensibilidad y la división, el valor debe ser 15000, 2,9965 y 2:
Capacidad Byte 144 Byte 145 Byte 146 Byte 147
Hex 00 00 3A 98
Dec 15000
Page 51
PROBLEM SOLVING
PROBLEM POSSIBLE CAUSE SOLUTION
On the screen The weight cannot be detected because
Check the connections of the
the message appears the cell does not exist or is not well
cells.
O-L connected
The weight cannot be displayed because
On the screen
exceeds the five available digits
the script appears in
or is greater than the capacity of the
the top
cells.
On the screen
The weight cannot be displayed because
the underscore appears it is negative and more than -9999.
at the bottom.
The number of
A value has been selected from Select the division value
decimals is
incorrect division. correct in the main menu.
incorrect.
Check the connections of the form
Communication Incorrect installation.
as indicated in the manual of
series not working The selection of the operation of the installation.
correctly. The serial interface is incorrect.
Select the appropriate settings.
To reset to zero, calibrate the
The zero function The gross weight exceeds the limit of weight.
semi-automatic no semi-automatic zero action.
Wait for the weight to stabilize
it works. The weight is not stabilizing. or adjust the filter parameter of the
weight.
Check the gross weight.
The tare function The gross weight is negative or exceeds the
semi-automatic no maximum capacity. Wait for the weight to stabilize
it works. or adjust the filter parameter of the
The weight is not stabilizing.
weight.
Page 52
GROUND CONNECTIONS OF THE DAT-400
0 Vdc
3
2
Page 53
Page 54
OPTIONAL 24 VOLT POWER SUPPLY
INSTALLATION
Ensure that the installation complies with local regulations and legislation on
electricity.
Connect the AC voltage to the terminals marked as 'L' and 'N'.
When the power supply is turned on, a red LED will light up.
The DC voltage is available at the terminals marked as '+' and '-'. The second
A terminal set is used when two transmitters are activated from a single source of
food (see the notes).
Fuse
L N LED + -
230 VAC 24 VDC
FUSE REPLACEMENT
Page 55
NOTES
Page 56
NOTES
Page 57