Laboratory Quality Control
Laboratory Quality Control
Definitions:
Quality Control: the process of detecting errors
to ensure the reliability of the test results to give the best patient care !
Unreliable Performance ?
Potential consequences include: patient misdiagnosis delays in treatment increased costs
avoidable retests cost US 200million USD per year
3% +ve bias
15% increase in patient treatment.
Error Classification..
Pre-analytical: errors before the sample reaches the laboratory
collection timing
24 hour urine
Other Factors..
The sex of the patient
male or female
Dietary effects
low carbohydrate / fat high protein / fat
Patient posture
urinary protein in bed-ridden patients
Other Factors..
Effects of exercise
creatine kinase / CRP
Medical history
heart disease / diabetes / existing medication
Pregnancy
hormonal effects
Analytical Errors..
The sample:
labelling
barcoding / aliquoting
preparation
centrifugation / aspiration
storage temperature
short term refrigeration medium term freezing at 20oC long term freezing at -80oC
Analytical Errors..
The sample: Glassware / pipettes / balances:
used incorrectly contaminated poorly calibrated reuse of pipette tips
Analytical Errors..
The sample: Glassware / pipettes / balances: Reagents / calibrators / controls:
poor quality inappropriate storage
correct temperature badly maintained fridges or freezers
stability
shelf-life / working reagent
incorrect preparation
Analytical Errors..
The sample: Glassware / pipettes / balances: Reagents / calibrators / controls: The application:
incorrect analytical procedures poorly optimised instrument settings
Analytical Errors..
The sample: Glassware / pipettes / balances: Reagents / calibrators / controls: The application: The instrument:
operational limitations
temperature control/read times/mixing/carry-over
lack of maintenance
worn tubing / optics / cuvettes / probes
Other Factors..
Calculation errors:
incorrect factor / wrong calibration values
Accuracy ?
Precision ?
The reproducibility of your results.
Specificity ?
The ability of a method to measure solely the component of interest. A lack of specificity will affect accuracy
falsely elevated values
hormones and drugs
Sensitivity ?
The ability to detect small quantities of a measured component.
will affect both precision and accuracy at the bottom end of the assay range.
Normal Distribution..
Mean value (x)
Frequency
Measured value
Precision ?
How disperse the values are. Quantified by measuring the Standard Deviation (SD) of the set of results.
(xi - x) SD = ( ) n-1
2
The lower the SD the better the Precision.
Example:
Mean result (x) = 100 mmol/L Standard deviation (SD) = 1.0 mmol/L
+1SD
Frequency
68%
99 100 101
+2SD
Frequency
95%
98
100
102
Coefficient of Variation..
Example:
Potassium %CV = (0.1 / 5.0) x 100% = 2.0%
Sodium %CV = (2.0 / 140) x 100% = 1.4%
Sodium has the better CV and in this example is performing better than potassium.
Interpretation..
10 40 41 50 51 70 71 100 101 120 unacceptable performance need for improvement acceptable good excellent
TS Calculations
V = (Result - Mean for Comparison) x 100 Mean for Comparison
The mean for comparison could be either: the all method mean your method mean your instrument mean
TS Calculations
TS = Log10 (3.16 x TCV) V
TCV is Target Coefficient of Variation
x 100
TS Calculations
TS = Log10 (3.16 x TCV) V
3.16 is selected as a constant because:
the log10 of 3.16 is 0.5 so if V = TCV, then the target score will be 50
x 100
TS =
log10 3.16 x TCV x 100 V log10 3.16 x 3.7 3.7 log10 (3.16) x 100 50 x 100
What is Acceptable ?
A sodium control has a target value of 140 mmol/L
139 mmol/L 120 mmol/L 160 mmol/L
140 mmol/L
141 mmol/L
180 mmol/L
What is Acceptable ?
A range of acceptable values is established Sodium Control:- 137 143mmol/L.
Assayed serum:
with predetermined targets and ranges
established by the manufacturer.
Unassayed Serum..
Analysed extensively by the laboratory.
a minimum of 20 sets of data generated a mean +/- 2SD range established
95% of results acceptable
Assayed Serum..
Targets and ranges generated by the manufacturer: abc utilises RIQAS
database of 5,000 laboratories method / instrument / temperature specific values
143
141.5
X
X X X X X X X X
Mean
X X
X X
140
-1SD
-2SD
138.5
137
143
141.5
X X
X X
X
X X
140
-1SD
-2SD
138.5
137
143
X X X X
X X
141.5 140
Mean
X X
X X X X X
-1SD
-2SD
138.5
137
143
141.5 140
X
X
Mean
-1SD
-2SD
138.5
137
143
X X
+1SD
X
141.5
X
Mean
X X X X
140
-1SD
-2SD
X
X X X X X
138.5
137
143
X X
141.5
X
X X X X
Mean
X X
140
-1SD
X X
138.5
137
-2SD
143
X X
141.5
X
X
X X X
Mean
X X X X X
140
-1SD
-2SD
138.5
137
Westgard Rules..
Decision criteria is dependent on the precision of the method or analyser
the less precise the method the more difficult the decision.
Westgard Flowchart..
Control data 1 point outside 2 SD Yes 1 point outside 3 SD No
No
No
4 consecutive control values on one side of the mean and further than 1 SD from the mean No
Yes
EQA Options..
International / National / Regional
International schemes provide: a larger database of results a wider range of analytical methods a global representation of diagnostic manufacturers Compulsory or Voluntary
RIQAS
abc
Management tool
to assess, review and improve performance
RIQAS..
Annual subscription
two six monthly cycles
Weekly samples
one vial reconstituted per week
tested blind as if a patient sample