6.UV Visible Latest
6.UV Visible Latest
ν=(1/λ)c=(c/v)λ=(v/c)
Electro Magnetic Spectrum
Color Wave
length
109 107 105 103 101 10-1 10-3 10-5 10-7 10-9 10-11
(nm)
violet 400-435
gamma
X-rays
indigo 435-480
500 Violet, indigo, blue
Ultra Violet
600 Green, yellow blue 480-500
700 Orange, red
Infra red green 500-560
microwave
yellow 560-595
Radio
waves orange 595-610
red 610-750
• What compounds show UV spectra?
• Generally think of any unsaturated
compounds as good candidates. Conjugated
double bonds are strong absorbers
• Just heteroatoms are not enough but C=O are
reliable
Transitions
• s->s*
– UV photon required, high energy
• Methane at 125 nm
• Ethane at 135 nm
• n-> s*
– Saturated compounds with unshared e-
• Absorption between 150 nm to 250 nm
• e between 100 and 3000 L cm-1 mol-1
• Shifts to shorter wavelengths with polar solvents
– Minimum accessibility
– Halogens, N, O, S
• n->p*, p->p*
– Organic compounds, wavelengths 200 to 700 nm
– Requires unsaturated groups
• n->p* low e (10 to 100)
– Shorter wavelengths
• p->p* higher e (1000 to 10000)
Electron transitions
The UV Absorption process
* and * transitions: high-energy, accessible in
vacuum UV (max <150 nm). Not usually observed in
molecular UV-Vis.
0.0
200 400 800
(-C=C-)5 330 nm
Hyperchromic shift
“hypsochromic” shift
“bathochromic” shift
Hypochromic shift
ii) Hypsochromic shift or blue shift. It involves the
shift of absorption maximum towards shorter
wavelength and may be caused by removal of
conjugation in a system or by change of solvent. For
example: In aniline , absorption maximum takes
place at 280 nm because the pair of electrons on
nitrogen atom is in conjugation with the ᴨ bond
system of the benzene ring. In acidic medium it
absorbs at 200 nm because the electron pair is no
longer present and hence conjugation is removed.
iii) Hyperchromic shift: This effect involves an increase in the
intensity of absorption and is usually brought about by
introduction of an auxochrome. For example. introduction of
CH3 group in position 2 of pyridine increases є from 2750-3560
and λmax from 253 nm to 262 nm
λmax
Solvent effects
A most suitable solvent is one that does not itself absorb in
the region under investigation.
A dilute solution of the sample is always prepared for spectral
analysis.
Most commonly used solvent is 95 % ethanol.
Ethanol is cheap and is transparent down to 210 nm.
Commercial ethanol should not be used because it is having
benzene which absorbs strongly in the ultra violet region.
Some other solvents which are transparent above 210nm are
n-hexane, water, methanol, cyclohexane, acetonitrile, diethyl
ether etc.
Hexane and other hydrocarbons can be used because these
are less polar and have least interactions with the molecule
under investigation.
• The n-ᴨ* (less intense) transition moves to shorter
Wavelength by increasing the polarity of the solvent.
In n-ᴨ* transition, the ground state is more polar than
the excited state
RADIANT WAVELENGTH
SOLVENT
PHOTO- READOUT
SOURCE SELECTOR DETECTOR
Littrow Prism
mirror
Detector
☼
Sample
Chopper
UV Instrumentation
Sources
Hydrogen or Deuterium Discharge Lamps:
Commonly used ultraviolet spectrophotometers are
hydrogen and deuterium lamps under very low pressure
(0.2-0.5 torr) and low voltage ( about 40 V dc) conditions.
Heated cathodes provide the essential function of
maintaining the discharge. At less than 360 nm these
discharge lamps provide a strong continuum that fulfills most
needs in the ultraviolet region.
With fused silica envelopes, work to about 160 nm is
feasible. At higher wavelengths longer than about 380 nm,
the discharge has lower intensity and emissions lines
superimposed on the continuum, thus restricts its use in the
visible region.
Incandescent filament lamps:
• Measurement above 350 nm and into the near infrared to 2.5
μm are usually made with incandescent filament lamps, which
give a continuous spectrum over the range.
• In these lamps a wire filament, generally tungsten, is heated
to incandescence by an electric current.
• The filament is enclosed in a hermetically sealed bulb of glass
filled with an inert gas or a vacuum. Filament s are usually
coiled to increase their emissivity, efficacy, and mean
luminance. Incandescent lamps are rugged, low cost units
sufficiently bright for nearly all absorption work in the visible
and near ultraviolet regions.
Tungsten- halogen lamps :
These are a special class of incandescent lamps with
iodine added to normal filling gases. The envelope is
fabricated of quartz to tolerate higher lamp operating
temperatures of 3500K. The iodine combines chemically
at the bulb wall with sublimed tungsten. The resulting WI 2
migrates back to the hot filament where it decomposes
and tungsten is redeposited. The cycle is repeated ,
continuously cleaning the bulb. These lamps maintain
more than 90 % of their initial light output throughout
their life.
The Xenon discharge lamps and mercury arcs are also can
be used
When the pressure of the gas is low, only line spectra are
emitted. But, if the pressure of the gas is high, band
spectra and continuous spectra will be obtained
Monochromators
The dispersing element may be a prism or grating.
• Detection of conjugation.
• Detection of geometrical isomers
H5C6 H H 5C 6 C6H5
H H H
C6H5
= I0 10-0.4343k’c = I010-K’c
On combining eqs. 1 and 2 called Beer-Lambert’s law
It=I0 10-act or log (I0/It) = act
• The molar absorption coefficient, molar extinction
coefficient, or molar absorptivity, is a measurement of how
strongly a chemical species absorbs light at a given
wavelength. It is an intrinsic property of the species; the
actual absorbance, A, of a sample is dependent on the
pathlength, ℓ, and the concentration, c, of the species via the
Beer–Lambert law, .
• The SI units for ε are m2/mol, but in practice, they are usually
taken as M−1 cm−1 or L mol−1 cm−1.
4 C6H5CH2OH ↔(C6H5CH2OH)4
Dissociation of the polymer increases with dilution. The
monomer absorbs at 2.750-2.765 µ where as the polymer
absorbs at 3.0 µ. Hence absorption at 2.75µ shows negative
deviation whereas at 3.0µ positive deviation
450nm 350 nm
C. Deviation may also occur due to the presence
of impurities that fluoresce or absorb at the
absorption wavelength.
D. Beer’s law cannot be applied to suspension
Instrumental Deviations
A. Polychromatic Radiation:
Beer’s law requires monochromatic radiation because
absorptivity is a constant at a single wavelength and varies with a
change in wavelength. The possibility of error due to the
practical impossibility of obtaining monochromatic radiation
may be minimized by the selection of a spectral region where
the change in absorptivity with a change in wavelength is very
small. This dictates a wavelength selection from a broad band
rather than from a sharply rising or sharply falling section of the
absorption curve.
The energy which passes through the sample and activates the
detector is a function of the weighted average of the
absorptivities of all the wavelengths which are transmitted by
the monochromator system. At the top (point A) of the
absorption band, the absorptivities at λ1 and λ 2 are close to
that of the desired wavelength (λmax). Minimum error will rise
from the use of this wavelength. On the sharply rising (point B)
or falling (point C) positions the weighted average yields low
results, these low results becoming more and more apparent at
increasing concentrations. In summary, because of the
numerous possibilities of an error arising from deviations it is
necessary in practice to prepare a calibration curve for the
absorbance- concentration relationship at the chosen
wavelength. Beer’s law should be verified in each application
on each instrument. It should not be assumed.
Stray light: Stray light is defined as radiation from the
instrument that is outside the nominal wavelength band
chosen for the determination. This stray radiation often is
the result of scattering and refection of the surfaces of
gratings, lenses or mirrors, filters and windows. When
measurements are made in the presence of stray light, the
observed absorbance is given by
A’ = log[(P0 + Ps)/ (P + Ps)]
Where Ps is the radiant power of the stray light.
The deviation due to stray light are most significant at high
absorbance.
Mismatched Cells.
If the cells holding the analyte and blank solutions are
not of equal path length and are not equivalent in optical
characteristics, an intercept will occur in the calibration
curve and A = εbc + k
Will be the actual equation. This error can be avoided by
using either carefully matched cells or a linear regression
procedure to calculate both slope and the intercept of
the calibration curve. Another way to avoid the
mismatched cell problem with single beam instruments
is to use only one cell for the blank and sample by
placing in the same position after careful cleaning.
Problems
1. The absorption spectrum for titanium peroxide complex ion in perchloric
acid showed a maximum of 400nm. The absorbance of a 32.0 µg/ml
solution of titanium gave an absorbance of 0.456. An unknown solution
treated in an identical fashion gave an absorbance of 0.501. Assuming
identical cells, find out the concentration of the unknown.
As= єscsts
Ax = єxcxtx
As/ Ax = єscsts/ єxcxtx
Since the absorptivities of the standard and unknown solutions are the
same and the cell depths ts and tx are equal, the equation can be written as:
As/ Ax = cs/ cx
Cx = ?
= 32.0 x (0.501/ 0.456) = 35.2 µg/ml