Digital Mammography

1/
For internal use only DXR 2009 Playbook/
GE Company confidential - Do not copy or distribute 01/20/25
Image Quality- Summary

System Image Quality

Tube Grid DQE


(detector)
AOP/AEC Dose

MTF Noise Contrast


spatial resolution Dose dependent
transfer

Image Quality Measurement 9/

Contrast Detail Phantom


GE Title or job number /
01/20/25
The factors for a good image quality in dense
breasts

Detector Tube

DQE at low dose Bi metal Tube

AOP
Automatic
Measurement of
breast density to set
up right parameters

GE Unique features 10 /
GE Title or job number /
• DQE = Detective quantum efficiency 01/20/25
• AOP = Automatic Optimization of Parameters
Digital Mammography
Image Quality vs. Dose
Control
(AEC/AOP) Tube/Anode

Image Processing
and Display

Detector

11 /
GE Title or job number /
01/20/25
The factors for a good image quality in dense
breasts

Detector Tube

DQE at low dose Bi metal Tube

AOP
Automatic
Measurement of
breast density to
set up right
parameters

GE Unique features 12 /
GE Title or job number /
01/20/25
Revolution Flat panel detector

Complete Production Control

Panel Panel Absorber Electronics System


Production Preparation Deposition Integration Integration

GE GE GE GE GE

Design choices
Technology Improved CsI performance
Pixel size 100 micron
Field of view increased to
24x31 cm
Read out faster and noiseless
Low dose DQE increased by 20 %

Technological choices proven to be 13 /

the best GE Title or job number /


01/20/25
The factors for a good image quality in dense
breasts

Detector Tube

DQE at low dose Bi metal Tube

AOP
Automatic
Measurement of
breast density to
set up right
parameters

GE Unique features 15 /
GE Title or job number /
01/20/25
POLLON unique mammography tube design

Featuring:
• No cooling delays
• Unlimited patient throughput
• Reliable and oil free
• Image quality optimized on all types of breasts
(Rh and Mo dual target)
• Small focal spot (0.1 mm, in addition to 0.3 mm)
• Smallest and lightest tube in mammography

16 /
GE Title or job number /
01/20/25
The factors for a good image quality in dense
breasts

Detector Tube

DQE at low dose Bi metal Tube

AOP
Automatic
Measurement of
breast density to
set up right
parameters

GE Unique features 20 /
GE Title or job number /
01/20/25
Purpose of AOP

The image quality you need


with

Low dose to the patient

As Low sA easonable
R A
chievable

21 /
GE Title or job number /
01/20/25
AOP in mammography
Mechanical Pre-exposure & Optimised
thickness attenuation exposure
measurement measurement
image quality

2 cm

4 cm

6 cm
8 cm

Thickness/composition
Optimum
anode,
anode filter,
Optimisation
kV, mAs
22 /
GE Title or job number /
01/20/25
Exposure Control, AOP (Automatic Optimization of Parameters)
Correct exposure for the densest region in the
breast for all breasts
You need the right exposure parameters (anode, filter, kV, mAs)…
•every
The time
system automatically scans
the whole breast and identifies the
location where it is the densest.
• Thickness + Density + Composites
 AOP (Anode, filter, kV, mAs &
therefore the dose)
High Quality Image

23 /
GE Title or job number /
01/20/25
Senographe Exposure Control
AOP (Automatic Optimization of Parameters)
What if:
•Your exposure was automatically correct for the
densest region in the breast for all breasts?

•Your exposure was extremely consistent all the time


for all breast compositions

•Your exposure achieved optimized image quality at


minimized dose to the patient?
Digital mammography

Image Quality

Systems ergonomics

Digital Workflow

28 /
GE Title or job number /
01/20/25
Seno Essential
System
Ergonomics

29 /
GE Title or job number /
01/20/25
Ergonomics
• Intuitive user interface
• Customization of speed and compression
• Small tube head
• Compact footprint
• Wheelchair access
• Patient comfort

• Isocentric motorized gantry with two speeds


allows instantaneous fine and fast positioning

• Automatic CC and MLO positioning

• Mirror angle for MLO

• One second activation

• Automatic light switch on

• Ergonomic control console and optional control


exposure hand-switch…to reduce repetitive strains
• Immediate access to patient worklist
(with RIS connection) and easy selection
of patient details
• Quick and automatic image display
(<15 sec.) allows better positioning
control and quality assurance decisions
• Capability to push images from imaging
workstation to review workstation, printer
or CAD automatically at the completion
of the study
• Narrow tube head and face shield for more comfortable head and neck positions
• Ergonomic handles, secondary handles, and covers minimize the patient’s muscle contraction

• Ergonomic paddle design and a rounded


bucky shape – to optimize patient comfort

• Compression slows when contacting the breast

• • Maximum desired compression can be pre-set


by the operator

• • Automatic compression release is synchronized


with exposure completion

30 /
GE Title or job number /
01/20/25
Digital Maoography

Image Quality

Systems ergonomics

Digital Workflow

31 /
GE Title or job number /
01/20/25
Optimized workflow
Smart image management
Image file size according to Field of View used

Display optimized. Zoom factor and time


Processing time reduced accordingly
Transfer time and archiving space optimized

Display functions
Fit to screen will automatically display the whole
image on the screen

Printing Functions
Auto print function for diagnostic images.

32 /
GE Title or job number /
01/20/25
BIOPSY
Senographe Interventional: Image quality & easy access to lesions

Advantages:
• True Full Field Digital Stereotaxy (FFDS)

• Vertical & Lateral access to lesions

• Biopsy window: 14 x 7 cm

• Bard or Mammotome biopsy gun

• Decubitus Position: patient comfort

• No dedicated room needed !


Strength of 2 in 1 :
Same image quality from diagnosis to digital stereotaxy 34 /
GE Title or job number /
01/20/25
35 /
GE Title or job number /
01/20/25
Both Vertical & Lateral approach available with Senographe
Interventional !!!
36 /
GE Title or job number /
01/20/25
Superior lesion accessibility
Biopsy window 5x4 cm
Scout field of view: 10x8 cm
Stereo field of view: 10x7 cm per image - total max 14x7
cm… the largest on the market
Every lesion in the biopsy window can be accessed

Biopsy
window
opening
Most CCD Senographe DS
stereotaxy devices maximum stereotaxy
have constraints on field of view is 14x7 cm
the volume for superior access to
accessible for lesions
stereotaxy target
designation due to
limited field of view
Biopsy Biopsy
volume volume
accessible accessible

Field of View Field of View


38 /
GE Title or job number /
01/20/25
Advanced Applications are in hands!
1) GE Tomosynthesis
A series of exposures at different angles
(Detector not moving)
Processing as multiple slices or 3D

2) Contrast Enhanced Spectral Mammography

Iodine Contrast Agent

39 /
GE Title or job number /
01/20/25
See more now.
GE Breast Tomosynthesis
“The Difference”
Sensitivity and specificity of mammography
are limited by several factors
1 2

 Dense breast tissue can overlap with


lesions
 3 4
Lesions are not always visible with x-
ray
 Interpretation of images can vary
among radiologists

Kolb et al. Radiology 2002; 225:165-175.


41 /
GE Title or job number /
01/20/25
Volumetric imaging that helps to eliminate

Note: *acquisition time will be dependent of breast thickness and ≤ 10 sec is for an average thick
overlapping tissue with Tomosynthesis
Automated rotational acquisition
 9 projections
 Sweep angle 25° ( +/- 12.5)
 Acquisition time: <10 sec

Source: Digital Breast Tomosynthesis Fundamentals – White


Low dose X-ray sweep – 1,1mGy

Compressed breast

Digital detector
.
.
.
.

paper
.

9
42 /
GE Title or job number /
01/20/25
3D volume imaging is reconstructed to

Source: Digital Breast Tomosynthesis Fundamentals – White paper Micro calcifications: Comparison of clinical cases
obtain a series of tomo-planes
 Thickness of the tomo-plane is equal to 0
 The distance between the tomographic
planes is equal to 0.5mm

0.5mm

reconstructed at 1mm vs. 0.5mm


Tomo-planes
Slabs = 1 cm thick

The 0,5mm distance between planes may improve the visualization


of the microcalcifications…
43 /
GE Title or job number /
01/20/25
Insightful technology with sense

Source: Digital Breast Tomosynthesis Fundamentals – 6-7-8 white


Tube: Beam Quality RH Track Detector: High DQE at low dose I.Q.: 25 angle 9 projections
Low patient dose Allows the visualization of small objects •Preserve small objects and masses
Dense breast penetration such as microcalcifications (no binning) tissue separation

Reconstruction: Iterative
Yields FFDM-like images

paper references
Positive IQ impact for microcalcs *Positioner: Step & Shoot
Artifact free
Target Dose: 3DMLO = 2DMLO
•Avoids image blur

44 /
GE Title or job number /
01/20/25
SenoBright*
Contrast-Enhanced Spectral Mammography

** Trademark
Trademark of
of the
the General
General Electric
Electric Company
Company
Today’s Mammography

49 /
GE Title or job number /
01/20/25
Can we fill this gap?

50 /
GE Title or job number /
01/20/25
SenoBright
Cancers cells create small new
blood vessels. This process is
called Angiogenesis

Iodine injection – Contrast agent will


tend to leak from these new vessels
into cancerous tissues

SenoBright uses multiple x-ray exposures


to reduce background signal, similar to how
cameras reduce red-eye by using a series of
flashes.
By suppression of the signal from the normal
tissue, SenoBright effectively highlights the
iodine absorption in the tissues affected by
angiogenesis
51 /
GE Title or job number /
01/20/25
SenoBright – Case 1
Clarification of mammographically equivocal lesion

79 yo w palpable
mass on left
breast

SenoBright
contrast-enhanced
images clearly
localize the lesion

CC @ 2min MLO @ 4 min

Low Energy Contrast-Enhanced Contrast-Enhanced Low Energy

54 /
GE Title or job number /
01/20/25
Images courtesy of Mikawa Breast Center, Mikawa, Japan
SenoBright – Case 1
Clarification of mammographically equivocal lesion

79 yo w palpable
mass on left
breast

Unsuspected
finding on right
breast with , 2
small
fibroadenomas

CC @ 3min MLO @ 5 min

Low Energy Contrast-Enhanced Contrast-Enhanced Low Energy

55 /
GE Title or job number /
01/20/25
Images courtesy of Mikawa Breast Center, Mikawa, Japan
SenoBright - Know more now.

13

19

56 /
GE Title or job number /
01/20/25
SenoEssential