0% found this document useful (0 votes)
51 views25 pages

Validation

The document outlines the systematic steps for Analytical Method Validation (AMV) and Analytical Method Transfer (AMT), emphasizing the importance of establishing documented evidence to ensure consistent product quality. It details validation requirements, parameters, and examples of methods needing validation, alongside guidelines from ICH, USP, and WHO. The conclusion highlights the necessity of validation for regulatory compliance and the long-term benefits it offers to companies, despite initial costs.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
51 views25 pages

Validation

The document outlines the systematic steps for Analytical Method Validation (AMV) and Analytical Method Transfer (AMT), emphasizing the importance of establishing documented evidence to ensure consistent product quality. It details validation requirements, parameters, and examples of methods needing validation, alongside guidelines from ICH, USP, and WHO. The conclusion highlights the necessity of validation for regulatory compliance and the long-term benefits it offers to companies, despite initial costs.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 25

Analytical Method Validation (AMV)

&
Analytical Method Transfer (AMT)

M Jagadeesh
AD
Systematic Steps for Validation

Analytical Method
Plant Specifications & STPS
Development
(Q.C)

Test Method

Analytical Method Validation Analytical Method


Protocol (AMVP) Validation Report (AMVR)

Analytical Method Transfer


(AMT)
Validation
 Definition:

Establishing documented evidence that provides a high degree of


assurance that a specific process will consistently produce a product
meeting its pre determined specifications and Quality Attributes.

 Guidelines for Validation:

ICH (Q2) R2: Analytical Method Validation

USP <1225>: Validation of Compendial procedures

WHO: Good Manufacturing Practices: Guidelines on Validations


Purpose of Method Validation

 Identification of sources and Quantification of potential errors

 Establish proof that a method can be used for regular purpose

 Satisfy Regulatory requirements


Examples of Methods that requires Validation
 Spectrophotometric Methods:
UV/Vis , IR ,NMR ,XRD ,MS etc.
 Chromatographic Methods:
HPLC,TLC,GC,GC/MS etc.
 Capillary Electrophoresis Methods
 Particle size analysis Methods:
Laser, Microscopic , Sieving etc.
 Bio Analytical Methods:
PK/PD/Clinical studies
Requirements of Validation
 Impurities Quantities along with COA’s

 API,WRS or RS Quantities along with COA’s

 Sample Quantities along with COA’s

 Placebo Quantities along with COA’s

 Batch Analysis COA’s

 Specifications

 Chemicals and Columns atleast 2 No’s

 Test Methods or STP

 MSDS
Validation Parameters
 System Suitability and System Precision
 Specificity/Selectivity
 Method Precision
 Intermediate Precision
 Accuracy/Recovery
 LOQ & LOD Establishment
 Linearity
 Range
 Solution Stability
 Mobile Phase Stability
 Robustness
System Suitability and System Precision
 System Suitability is to prove that system is working perfectly before
analysis.

 System precision related to reproducibility and repeatability, under same


conditions show the same results.

Acceptance Criteria:
Dissolution by HPLC:%RSD-NMT 2.0%
Dissolution by UV: %RSD-NMT 3.0%
Assay: %RSD-NMT 2.0%
RC:%RSD-NMT10.0%
Specificity/Selectivity
 Specificity is the ability of the analytical method to distinguish between the
analyte (s) and the other components in the sample matrix.

 Blank, Placebo, Standard , impurities, Samples free from interference.

 Specificity plays a major role in

1. Forced Degradation Studies (FDS)

2. Interference Study

3. Filter study
Forced Degradation Studies
 Types of Degradation:
1. Physical conditions: Thermal , Photolytic , Humidity.
2. Chemical conditions: Acid, base, Peroxide, Hydrolysis

 Degradation:5%-20%

 Peak Purity Should be passed for standard ,sample and impurities

 Always PDA HPLC Systems use.

 Normalized Mass Balance: 95% - 105%

 Mass Balance: (%Assay of Stressed sample+% Total imp.)/% Assay of Unstressed


SampleX100

Acceptance Criteria:

Assay and RC: Peak Purity should be passed.

Mass Balance: should be achieved.


Interference Study
 Interference occurs when a substance or process falsely alters an test result.

 Blank , Placebo, Standard,Samples,Impurities plays a major role.

Acceptance Criteria:

Dissolution and Assay: No interference at RT of Active peak

RC: Disregard below LOQ Level


Filter Study
 Also known as Filter Saturation study.
 Blank, Placebo , Standard and Sample solution to be established.
 (Centrifuged or Unfiltered)
 Discarding and collect and inject 1st mL,2nd mL,3rd mL solution.
 Atleast 2 different types of filters required (PVDF, Nylon)
Filter solution-Unfiltered solution
Filter Study = Unfiltered solution X100
Acceptance Criteria:
Dissolution: NMT 3.0%
Specification %Difference
Assay: NMT 2.0% Level Acceptance Criteria
RC: Individual impurities
≤0.5% ±0.05 absolute difference
>0.5% ± 10%of Relative from Initial value
Total impurities
≤1.0% ±0.1% absolute difference
>1.0% ± 10%of Relative from Initial value
Method Precision (Repeatability)
 Definition:
Closeness of Agreement between a series of measurements obtained from
multiple sampling of the same homogenous sample.
 Control and Spiked Samples(Spec. Level) injected (n=6)
Acceptance Criteria:
 Product should meet specification limit (Control Sample)
 Dissolution : %RSD NMT 2.0%
 Assay :%RSD NMT 2.0%
 RC:
%w/w %RSD of
(Mean Value) %w/w
impurities
≤0.05 or LOQ NA
>0.05 To 0.1 30.0%
0.1 To 0.2 15.0%
>0.2 10.0%
Intermediate Precision
 Expresses within laboratories variations: different days,analysts,systems,columns.
 Another set of six samples (Control, Spiked samples) will be performed.
Acceptance Criteria:
 Product should meet specification limit (Control Sample)
 Dissolution : %RSD NMT 2.0%
 Cumulative %RSD(n=12)-NMT 5.0%
 Assay :%RSD NMT 2.0%
 Cumulative %RSD(n=12)-NMT 2.0%
 RC:
Accuracy/Recovery
 Definition:
Closeness of agreement between the value which is accepted as either as
conventional true value or an accepted reference value and the value found.
 Prepare 3 Samples for each level-Total of 9 samples.
 Dissolution: 10%-150% test concentration(NLT 3 Levels)
 Assay:25%-150% test concentration(NLT 3 Levels)
 RC-LOQ to 300% of specification level
 Weighing powder method or by spiking method on placebo or on samples.
Acceptance Criteria:
Dissolution: %Recovery -95-105% and %RSD NMT 5.0%
Assay: %Recovery -98-102% and %RSD NMT 2.0%
RC: Specification Acceptance
Level(%) criteria %RSD
%Recovery
≤0.05 or LOQ 75.0-125% 15.0
>0.05 To 1.0 85.0-115% 10.0
0.1 To 0.2 90.0-110% 5.0
>0.2 90.0-110% 5.0
LOQ and LOD Establishment
 Limit of Quantification:
Lowest amount of analyst in a sample, which can be quantitatively
determined with suitable precision and accuracy.
LOQ=10σ/s Where σ = Standard deviation of Response
S = Slope of Calibration curve

Limit of Detection:
Lowest amount of analyst in a sample, which can be detected but not
necessarily quantitated as an exact value.
LOD=3.3σ/s Where σ = Standard deviation of Response
S = Slope of Calibration curve
Acceptance Criteria:
LOQ: s/n should be 10
LOD: s/n should be 3
Linearity
Definition:
The linearity of an analytical procedure is its ability (within a given range) to obtain
test results which are directly proportional to the concentration (amount) of analyte
in the sample.
Minimum of 6 levels required
Dissolution: 0% to 150% of Target Standard concentration
Assay: 0% to 150% of Target standard concentration
RC: LOQ-300% of shelf life specification
Acceptance criteria:
Dissolution: r2 NLT 0.999,Y-intercept-NMT 2.0%
Assay: r2 NLT 0.99,Y-intercept-NMT 2.0%
RC:r2 NLT 0.98,Y-intercept-NMT 2.0%
Report RRF value
Range
 Range of an analytical method define as interval from the upper to the

lower concentration of the analyte/impurities in the sample with

Accuracy, Linearity and precision experiments.


Solution Stability
 Solution stability plays a major role in sample and standard solution
establishments.
 Intial,24hrs,48hrs-on day basis will be established. If fails-hourly basis.
 Establishes in both RT and Rf conditions.

Acceptance Criteria:
For Standard : Similarity Factor should pass 0.98 to 1.02
Dissolution: Sample from initial to Next time point NMT3.0%
Assay: Sample from initial to Next time point NMT2.0%
RC: Specification Level %Difference
Acceptance Criteria
Individual impurities
≤0.5% ±0.05 absolute difference
>0.5% ± 10%of Relative from Initial value
Total impurities
≤1.0% ±0.1% absolute difference
>1.0% ± 10% of Relative from Initial value
Mobile Phase Stability

 Stability of Mobile phase plays a major role in Validation.


 Mobile phase stability was established for 2 days on BT
 Every day fresh SST will injecting and calculating against aged standard
and sample solutions
 Any Colour change or particles observed .Mobile phase stability will be
aborted.
Acceptance criteria:
Dissolution: %Difference for Sample NMT 2.0%
Assay: %Difference for Sample NMT 2.0%
Specification Level (%w/w) %Relative Variation
RC:
Individual impurities
<0.10 ±30
0.10 To 0.50 ± 20
>0.50 ± 10
Total impurities
0.10 To 0.50 ±25
>0.50 ±15
Robustness
 Definition:
It is a measure of its capacity to remain unaffected by small, but deliberated
variations in method parameters and provides an indication of its reliability
during normal usage.
Dissolution:
Chromatographic Conditions:
Flow Rate, Column Temperature, Buffer pH, Organic Variations, Wave length
Sampling Conditions:
Auto Vs Manual, RPM Variation, Media Volume Variation, Media PH Variation
Media Temperature.
Assay and RC:
Chromatographic Conditions:
Flow Rate, Column Temperature, Buffer pH, Organic Variations, Wave length
Sampling Conditions:
Sonication Time, Shaking Time, Centrifuge Time, Diluent organic composition
Analytical Method Transfer (AMT)
 Analytical Method transfer is the documented process that qualifies a laboratory
(the receiving unit) to use an analytical test procedure that originated in another
laboratory (the transferring unit).
 Guidelines for Analytical Method Transfer :
USP <1224>: Transfer of Analytical Procedures
 Types of Analytical Method Transfer:
 Direct Transfer
 Physical demonstration and joint analysis of sample by Transferring unit, Scientist
(R&D – Analytical) and the Receiving unit Chemist/analyst, (QC/other laboratory).
 Data will be comparable.
 Indirect Transfer
 This type of method transfer shall include only Protocol data will be shared to
Transferring unit
 Data will be comparable.
Flow Chat for AMT
Conclusion
 Summarized the extent, need and necessary of validation.

 Validation may cost initially but it avoids the rise of break down ,
for this reason small companies were concentrating more on
validation their by fulfilling the goals of regulatory markets.
Thank Q

You might also like