AMESA-M

Total Mercury Analyser
Monitoring system for fully-automatic and
continuous mercury analysis in smoke gas
(without wet chemistry)
Already MCERTS and TÜV accredited for continuous dioxin and furan

KEYBENEFITS

monitoring, the AMESA - M has been adapted for mercury monitoring.

•

Maintenance-free (6 months) dry

AMESA - M extracts a sample from the flue gas stream under volume

reactor

proportional conditions.

•
•
•
•

Consisting of a sample probe assembly and a control cabinet, the

High operational safety
Easy maintenance

Internal Data Logging Functionality

Low cross sensitivities

The unit can store up to a four week period of sampling data which is

Thermal mass flow meter to control 	 	

then available to download via a USB flash drive. Analysis can be achieved

mass flow independent of gas &

with an onsite analyser or a certified laboratory.

pressure
Dry gas meter to determine sample 	 	

Measuring Principle

volume

The sample gas is converted into mercury vapour by a combination of
thermal and dry chemical treatment. This is then continuously
measured in a photometer. Probe gas flow is measured after a gas cooler
at 2°C. The concentration is calculated and displayed as dry flue gas.

“

Providing continuous Mercury
monitoring, The AMESA - M is suited
for audit & control purposes

“

•
TECHNICALSPECIFICATION
APPLICATIONS

• Waste incinerations	
• Sewage sludge incineration
• Contaminated soil burning plants
• Mercury mines and refineries
• Fluorescent light bulb recycling plants
MEASUREMENT PRINCIPLE

• Measurements: total mercury
• Measuring ranges 0–45 to 0–500 μg/Nm³
• Measuring principle: UV-absorption

POWER

• Measuring outputs: 2 x 0 / 4–20 mA/500 Ohm
• Digital outputs: 4 relay outputs, permissible load 250 V, 	
100 VA

• Digital inputs: 1 potential free outputs
• Supply voltage: 230 VAC, 50 Hz, 1200 VA
		

sample probe: 650 VA

		

sample line: 100 VA/m

SPAN & DRIFT

• Accuracy: <1% of measuring range	
• Detection limit: <1 μg/Nm³
• Reference point drift: <2% of measuring range/month
• Zero point drift: automatic zero correction

TEMPERATURE & PRESSURE

• Flue gas temperature: 0–250 °C	
• Flue gas pressure: -50 up to +50 hPa
• Ambient temperature: +5 bis +30°C

APPROVALS

• MCERTS 	
• TÜV Certified

PHYSICAL

• Dimensions: cabinet 1600 x 800 x 500 mm (H x W x D)
• Weight: 220 kg
• Duct diameter: >0.5 m

Complete Project Solutions
Our complete project solutions are developed to assist you
in IED compliance. We achieve this through;

•
•
•
•
•

Expert consultancy
Project management
In-house project design & build
Installation & commissioning
After product service & support

11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HT
T: +44(0)151 666 8300 F: +44(0)151 666 8329 E: sales@a1-cbiss.com
W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com

Cert No. 996
ISO 9001

Amesa mercury sorbent trap system datasheet

  • 1.
    AMESA-M Total Mercury Analyser Monitoringsystem for fully-automatic and continuous mercury analysis in smoke gas (without wet chemistry) Already MCERTS and TÜV accredited for continuous dioxin and furan KEYBENEFITS monitoring, the AMESA - M has been adapted for mercury monitoring. • Maintenance-free (6 months) dry AMESA - M extracts a sample from the flue gas stream under volume reactor proportional conditions. • • • • Consisting of a sample probe assembly and a control cabinet, the High operational safety Easy maintenance Internal Data Logging Functionality Low cross sensitivities The unit can store up to a four week period of sampling data which is Thermal mass flow meter to control then available to download via a USB flash drive. Analysis can be achieved mass flow independent of gas & with an onsite analyser or a certified laboratory. pressure Dry gas meter to determine sample Measuring Principle volume The sample gas is converted into mercury vapour by a combination of thermal and dry chemical treatment. This is then continuously measured in a photometer. Probe gas flow is measured after a gas cooler at 2°C. The concentration is calculated and displayed as dry flue gas. “ Providing continuous Mercury monitoring, The AMESA - M is suited for audit & control purposes “ •
  • 2.
    TECHNICALSPECIFICATION APPLICATIONS • Waste incinerations •Sewage sludge incineration • Contaminated soil burning plants • Mercury mines and refineries • Fluorescent light bulb recycling plants MEASUREMENT PRINCIPLE • Measurements: total mercury • Measuring ranges 0–45 to 0–500 μg/Nm³ • Measuring principle: UV-absorption POWER • Measuring outputs: 2 x 0 / 4–20 mA/500 Ohm • Digital outputs: 4 relay outputs, permissible load 250 V, 100 VA • Digital inputs: 1 potential free outputs • Supply voltage: 230 VAC, 50 Hz, 1200 VA sample probe: 650 VA sample line: 100 VA/m SPAN & DRIFT • Accuracy: <1% of measuring range • Detection limit: <1 μg/Nm³ • Reference point drift: <2% of measuring range/month • Zero point drift: automatic zero correction TEMPERATURE & PRESSURE • Flue gas temperature: 0–250 °C • Flue gas pressure: -50 up to +50 hPa • Ambient temperature: +5 bis +30°C APPROVALS • MCERTS • TÜV Certified PHYSICAL • Dimensions: cabinet 1600 x 800 x 500 mm (H x W x D) • Weight: 220 kg • Duct diameter: >0.5 m Complete Project Solutions Our complete project solutions are developed to assist you in IED compliance. We achieve this through; • • • • • Expert consultancy Project management In-house project design & build Installation & commissioning After product service & support 11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HT T: +44(0)151 666 8300 F: +44(0)151 666 8329 E: [email protected] W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com Cert No. 996 ISO 9001