Infrared spectroscopy
AYESHA SHAFI
Objectives
Introduction
Principle of IR Spectroscopy
Theory of IR Spectroscopy
Molecular vibrations
Instrumentationof IR Spectroscopy
Applications of IR Spectroscopy
Infrared spectroscopy
It deals with the absorption of electromagnetic radiations in the infra red region.
Infrared region ranges from 10,000-100 cm-1 .This region is categorized as follows;
1. Near IR: 0.8-2.5 µm(12,500-4000 cm-1 )
2. Middle IR: 2.5-15 µm (4000-667 cm-1 )
3. Far IR: 15-200 µm (667-100 cm-1 )
Most of the analytical applications are confined to the middle IR region because absorption
of organic molecules are high in this region.
Principle and Theory of IR spectroscopy
Principle of IR spectroscopy:
The IR radiations causes the excitation of the molecules from ground state to the higher
vibrational energy states on absorption i.e. the absorption is due to change in vibrational state of a
molecule that appear as a band in the IR spectrum.
Theory of IR spectroscopy:
IR spectrum is a graph between the absorbance and the wavelength or the wavenumber. A band is
obtained in the spectrum as changes in the vibrational energy of the molecule are also associated
by changes in the rotational energy . A molecule is not a rigid assembly of atoms. It is said to
resemble a system of balls of varying masses and spring of varying strength, corresponds to
chemical bond of a molecule.
Theory of Infrared Absorption
Spectroscopy
For a molecule to absorb IR, the vibrations or rotations within a molecule must cause a net
change in the dipole moment of the molecule.
If the frequency of the radiation matches the vibrational frequency of the molecule then radiation
will be absorbed, causing a change in the amplitude of molecular vibration. Although the IR
spectrum is characteristics of a whole molecule, certain functional groups give rise to absorption
bands at the same frequency regardless of the structure of the rest of the molecules. Therefore IR
spectroscopy useful for identification of functional groups.
Molecular vibrations
There are two types of molecular vibrations.
1. Fundamental (if vibration are associated with the functional group).
2. Non fundamental
Fundamental is further classified as
 Stretching vibrations
 Bending vibrations
Stretching vibrations
These occur along the bond axis, where interatomic distance increases or decreases but atomic
position remain as such. It require more energy as compare to bending due to shorter wavelength.
Two types of stretching vibrations are possible –
I. Symmetrical stretching
II. Asymmetrical stretching
MOLECULAR
VIBRATIONS
Fundamental
Vibrations
Stretching
Vibration
Symmetric Asymmetric
Bending
Vibration
In-plane
Bending
Scissorin
g
Rocking
Out Of Plane
Bending
Wagging Twisting
Non-fundamental
Vibrations
Over Tones,
Combination
Tones,
Fermi
Resonance
a) Symmetrical stretching:
2 bonds increase or decrease in length simultaneously.
H
H
C
b) Asymmetrical stretching
in this, one bond length is increased and other is decreased.
H
H
C
2. Bending vibrations
It is rhythmical movement of atoms along the bond axis that cause the change in
the bond angle. It require less energy due to longer wavelength.
Occurs at low energy: 1400-666 cm-1
it is of 2 types:
a) In plane bending: scissoring, rocking
b) Out plane bending: wagging, twisting
a) In plane bending
i. Scissoring:
This is an in plane bending. Two atoms approach each other
Bond angles are decrease.
ii. Rocking:
Movement of atoms take place in the same direction.
H
H
CC
H
H
CC
b) Out plane bending
i. Wagging:
2 atoms move to one side of the plane. They move up and down the plane.
ii. Twisting:
One atom moves above the plane and another atom moves below the plane.
H
H
CC
H
H
CC
For linear (3n-5)degree of freedom represent fundamental vibrations
For non linear (3n-6)degree of freedom represent fundamental vibrations
For a molecule containing n number of atom s has 3n degree of freedom
Each atom has 3 degree of freedom depend on x , y ,z
Fundamental vibration of molecule depend on degree of freedom
It gives the relation between frequency of oscillation , atomic mass , force constant of the bond .
Thus vibrational frequency is
C = velocity of light
F = force constant
Mx= mass of atom x
My = mass of atom y
NON-
FUNDAMENTAL
OVER TONES:
These are
observed at
twice the
frequency of
strong band.
Ex:
carbonyl group.
COMBINATION
TONES:
Weak bands that
appear
occasionally at
frequencies that
are sum/difference
of 2 or more
fundamental
bands.
FERMI
RESONANCE:
Interaction b/w
fundamental
vibration &
overtones or
combination
tones.
Ex:CO2
Non fundamental vibrations
Coupled interactions
Interactions between vibrations can occur (Coupling) if the vibrating bonds are joined to
a single, central atom.
This is because there is mechanical coupling interaction between the oscillators.
Example:
C=O (both symmetric and asymmetric stretching vibrations).
The main parts of IR spectrometer are as follows:
1. radiation source
2. Sample cell
3. Monochromator
4. Detectors
5. Recorder
INSTRUMENTATION
SOURCE:
Two types of the source is used.
a) Globar b)Nernst glower
Globar: a rod of silicon carbide is heated to approximately 1300 degree centigrade to produce the
radiant energy from 1-40micron.
Nernst glower: a rod of zirconium and yittrium is heated to approximately 1500 degree centigrade
to produce the radiant energy from 0.4-20micron.
INSTRUMENTATION
Monochromator:
it is same like UV spectrophotometer but it is used to isolate the IR radiation of specific wavelength.
Sample cells:
For gas samples:
The spectrum of a gas can be obtained by permitting the sample to expand into an evacuated cell,
also called a cuvette.
For solution sample:
Infrared solution cells consists of two windows
of pressed salt sealed. Samples that are liquid at
room temperature are usually analyzed in pure
form or in solution. The most common solvents are
Carbon Tetrachloride (CCl4) and Carbon Disulfide
(CS2).
SAMPLING OF SOLIDS
Generally 4 techniques are employed for preparing solid samples:
1. Solids run in solution.
2. Solid Films.
3. Mull technique.
4. Pressed pellet technique.
SOLIDS RUN IN SOLUTION
Solids may be dissolved in non-aqueous inert solvent and a drop of
this solution is placed on an alkali metal disc and solvent is allowed
to evaporate, leaving a thin film of solute (or the entire solution is
placed in a liquid sample cell) which is then mounted in
spectrometer.
If the solution of solid can be prepared in a suitable solvent then the
solution is run in concentration of cells for liquids.
Some solvents used are chloroform, carbon tetrachloride, acetone,
Cyclohexane etc.
If a solid is polymer resins & amorphous solids, the sample is dissolved
in any reasonable volatile solvent & this solution is poured on a rock
salt plate (Nacl or KBr) & solvent is evaporated by gentle heating.
If solid is non-crystalline, a thin homogenous film is deposited on the
plate which can be mounted and scanned directly.
Sometimes polymers can be “hot pressed” onto plates.
SOLID FILMS
MULLTECHNIQUE
In this technique a small quantity of sample is thoroughly ground in a clean mortar
until the powder is very fine.
After grinding, the mulling agent (mineral oil or Nujol) is introduced in small
quantities just sufficient to take up the powder (mixture approximates the
consistency of a toothpaste).
The mixture is then transferred to the mull plates & the plates are squeezed
together to adjust the thickness of the sample between IR transmitting windows.
This is then mounted in a path of IR beam and the spectrum is run.
Pressed pellet technique
In this technique a small amount of finely ground solid sample is
intimately mixed with about 100 times its weight of powdered
Potassium bromide, in a vibrating ball mill.
This finely ground mixture is then pressed under very high
pressure in evacuable die or minipress to form a small pellet (about
1-2 mm thick and 1cm in diameter).
The resulting pellet is transparent to IR radiation and is run as such.
Preparing a KBr Disk
Pressed pellet technique (continue…)
The powder (KBr + sample) is introduced in between the
2 bolts and the upper screw A is tightened until the
powder is compressed to a thin disc.
After compressing the sample bolts are removed and a
steel cylinder with pellet inside it is placed in path of the
beam of IR spectrometer and a blank KBr pellet of
identical thickness is kept in the path of reference beam.
Advantages of this technique over mull technique
The use of KBr eliminates the problem of bands which appear in IR spectrum due
to the mulling agent as in this case no such bands appear.
KBr pellets can be stored for longer periods of time.
As concentration of the sample can be suitably adjusted in pellets, it can be used
for quantitative analysis.
The resolution of spectrum in KBr is superior to that obtained with mulls.
DETECTORS
The measurement of infrared radiation is difficult as a result of low intensity of available source
and low energy of the infrared photons. As a consequence of these properties, the electrical
signal from the infrared detector is small and its measurement requires large amplification.
Two types of IR detectors are used
•Thermal detector
•Non thermal detector
TYPES OF THERMAL DETECTOR
There are four types of thermal detector.
Bolometers
Thermocouple and thermopile
Pyro electric detector
Golay cell
BOLOMETER
Bolometer is derived from a Greek word (bolometron)
Bolo = for something thrown
Metron = measure
A bolometer consists of an absorptive element, such as a thin layer of metal.
Most bolometers use semiconductor or superconductor absorptive elements rather than
metals.
These material exhibit a relatively large change in resistance as a function of
temperature.
Working
Step No 1
Thin layer of
metal
connected to a
reservoir
Step No 2
Any radiation on the
absorptive element
raises its
temperature above
that of the reservoir.
Step no 3
The temperature
change can be
measured directly
with an attached
thermometer.
Temperature changes Potential difference changes
Thermocouples consist of a pair of junctions of different metals; for example, two pieces
of bismuth fused to either end of a piece of antimony. Whenever IR radiation fall at the
junction, it causes the change in temperature and also a change in the potential
difference that causes the production of electrical signal.
Thermopile
A thermopile is the combination of six thermocouples connected in series. Half
the junction are considered as hot and other half as cold which are thermally
bonded to the substrate and maintains the stable temperature.
The entire assembly is mounted in an evacuated enclosure with an IR
transmitting window made up of KBr so as to minimize the heat loss.
Thermopiles are one of the most simple and direct means of converting radiant
energy into an electrical energy.
Pyro electric Detector
• Single crystalline wafer of a pyro electric material, such as triglycerine sulphate.
• Pyro electric Infrared Detectors (PIR) convert the changes in incoming infrared light
to electric signals.
GOLAYCell
It consist of a small metal cylinder closed by a rigid blackened metal plate at one end and
flexible silvered diaphragm at other end. The whole chamber is filled with xenon gas.
Exposure to IR radiation through a radiation transmitting window causes the gas to expand
and diaphragm to deform. Motion of the diaphragm changes the output of cell and
detected in the form of signal.
Non Thermal Detectors
Photon detectors:
Photoelectric detectors consists of a semiconducting material deposited on a glass surface,
sealed in an evacuated envelop.
Absorption of an IR promotes the nonconducting valence electron to a higher, conducting state.
The electrical resistance of the semiconductor decreases.
SINGLE BEAM
SPCETROPHOTOMETER
DOUBLE BEAM
SPECTROPHOTOMETER
A single beam of light , which can
pass through one solution at a time
(sample or reference).
A single beam of light splits into
two separate beams. One passes
through the sample, another
passes through the reference.
Workingofdoublebeamspectrophotometer
The radiation from the source is divided into two parts.
 Reference beam
 Sample beam
The reference beam is passed through an optical wedge which limits the amount of radiations
that passes through it.
As the chopper is rotated, it causes the sample beam and the reference beam to be reflected
alternatively to the monochromator which sends individual frequencies to detector.
The detector receives an intense beam (the reference and the sample beam). This will lead to an
alternating current which is flow to the amplifier.
Workingofdoublebeamspectrophotometer
APPLICATIONOFIRSPECTROPHOTOMETER
IR spectroscopy is one of the most important and widely used analytical
techniques available to scientists working in a whole range of fields.
The fundamental region in the IR region is termed as the rock salt region. It is
comprised of group frequency region and the finger print region.
GROUP FREQUENCY
REGION
FINGERPRINT REGION
consisting of the absorption bands
of the functional groups.
frequency = 4000-1300cm-¹
wavelength = 2.5-8
IR spectra is called “fingerprints”
because no other chemical
species will have similar IR
spectrum.
Single bonds give their absorption
bands in this region.
Frequency=1300-650cm-1
Wavelength=8-15.4
IdentificationofSubstances
The finger print region is mostly used to compare the identity of a compound. Because
small differences in structure & constitution of molecule result in significant changes in
the peaks in this region.
Hence this region helps to identify an unknown compound.
• Criteria: Sample and reference must be tested in identical conditions, like physical state,
temperature, solvent, etc
DeterminationofMolecularStructure
Used along with other spectroscopic techniques.
Identification is done based on position of absorption bands in the spectrum.
Eg: C=O at 1717 cm-1.
Absence of band of a particular group indicates absence of that group in the compd.
DetectionofImpurities
The presence of absorption bands at positions where
compound is not expected to absorb indicates the presence of
impurities.
Determined by comparing sample spectrum with the spectrum
of pure reference compound.
Studyingthe progress of reaction
The progress of a reaction can be followed by examining the IR Spectra.
Observing rate of disappearance of characteristic absorption band in reactants; or
Rate of increasing absorption bands in products of a particular product.
Example: Oxidation of secondary alcohol to ketone can be detected by disappearance of OH band
and appearance of C=O band.
O—H = 3600-3650 cm-1
C=O = 1680-1760 cm-1
Molecularshapeorsymmetrydetermination
The shape or symmetry of a molecule can be determined by the IR
spectroscopy.
For example: If linear --> only 2 bands should be present.
If bent --> 3 bands should be present.
Actual spectrum shows 3 peaks at 750, 1323 and 1616 cm-1.
Dipole Moment determination
IR spectroscopy is also used to evaluate the dipole
moment of the molecule and compounds.
FOURIERTRANSFORMIRSPECTROPHOTOMETER
It is the most advance IR spectrophotometer . In this a laser beam is
used, which is allowed to fall on an interferometer. This photometer
has advantage of fast speed of analysis, so very accurate values of
molecular parameters are found with FTIR. A computer is also linked
to get the store data.

Infrared spectroscopy

  • 1.
  • 2.
    Objectives Introduction Principle of IRSpectroscopy Theory of IR Spectroscopy Molecular vibrations Instrumentationof IR Spectroscopy Applications of IR Spectroscopy
  • 3.
    Infrared spectroscopy It dealswith the absorption of electromagnetic radiations in the infra red region. Infrared region ranges from 10,000-100 cm-1 .This region is categorized as follows; 1. Near IR: 0.8-2.5 µm(12,500-4000 cm-1 ) 2. Middle IR: 2.5-15 µm (4000-667 cm-1 ) 3. Far IR: 15-200 µm (667-100 cm-1 ) Most of the analytical applications are confined to the middle IR region because absorption of organic molecules are high in this region.
  • 4.
    Principle and Theoryof IR spectroscopy Principle of IR spectroscopy: The IR radiations causes the excitation of the molecules from ground state to the higher vibrational energy states on absorption i.e. the absorption is due to change in vibrational state of a molecule that appear as a band in the IR spectrum. Theory of IR spectroscopy: IR spectrum is a graph between the absorbance and the wavelength or the wavenumber. A band is obtained in the spectrum as changes in the vibrational energy of the molecule are also associated by changes in the rotational energy . A molecule is not a rigid assembly of atoms. It is said to resemble a system of balls of varying masses and spring of varying strength, corresponds to chemical bond of a molecule.
  • 5.
    Theory of InfraredAbsorption Spectroscopy For a molecule to absorb IR, the vibrations or rotations within a molecule must cause a net change in the dipole moment of the molecule. If the frequency of the radiation matches the vibrational frequency of the molecule then radiation will be absorbed, causing a change in the amplitude of molecular vibration. Although the IR spectrum is characteristics of a whole molecule, certain functional groups give rise to absorption bands at the same frequency regardless of the structure of the rest of the molecules. Therefore IR spectroscopy useful for identification of functional groups.
  • 6.
    Molecular vibrations There aretwo types of molecular vibrations. 1. Fundamental (if vibration are associated with the functional group). 2. Non fundamental Fundamental is further classified as  Stretching vibrations  Bending vibrations Stretching vibrations These occur along the bond axis, where interatomic distance increases or decreases but atomic position remain as such. It require more energy as compare to bending due to shorter wavelength. Two types of stretching vibrations are possible – I. Symmetrical stretching II. Asymmetrical stretching
  • 7.
    MOLECULAR VIBRATIONS Fundamental Vibrations Stretching Vibration Symmetric Asymmetric Bending Vibration In-plane Bending Scissorin g Rocking Out OfPlane Bending Wagging Twisting Non-fundamental Vibrations Over Tones, Combination Tones, Fermi Resonance
  • 8.
    a) Symmetrical stretching: 2bonds increase or decrease in length simultaneously. H H C
  • 9.
    b) Asymmetrical stretching inthis, one bond length is increased and other is decreased. H H C
  • 10.
    2. Bending vibrations Itis rhythmical movement of atoms along the bond axis that cause the change in the bond angle. It require less energy due to longer wavelength. Occurs at low energy: 1400-666 cm-1 it is of 2 types: a) In plane bending: scissoring, rocking b) Out plane bending: wagging, twisting
  • 11.
    a) In planebending i. Scissoring: This is an in plane bending. Two atoms approach each other Bond angles are decrease. ii. Rocking: Movement of atoms take place in the same direction. H H CC H H CC
  • 12.
    b) Out planebending i. Wagging: 2 atoms move to one side of the plane. They move up and down the plane. ii. Twisting: One atom moves above the plane and another atom moves below the plane. H H CC H H CC
  • 13.
    For linear (3n-5)degreeof freedom represent fundamental vibrations For non linear (3n-6)degree of freedom represent fundamental vibrations For a molecule containing n number of atom s has 3n degree of freedom Each atom has 3 degree of freedom depend on x , y ,z Fundamental vibration of molecule depend on degree of freedom
  • 14.
    It gives therelation between frequency of oscillation , atomic mass , force constant of the bond . Thus vibrational frequency is C = velocity of light F = force constant Mx= mass of atom x My = mass of atom y
  • 15.
    NON- FUNDAMENTAL OVER TONES: These are observedat twice the frequency of strong band. Ex: carbonyl group. COMBINATION TONES: Weak bands that appear occasionally at frequencies that are sum/difference of 2 or more fundamental bands. FERMI RESONANCE: Interaction b/w fundamental vibration & overtones or combination tones. Ex:CO2 Non fundamental vibrations
  • 16.
    Coupled interactions Interactions betweenvibrations can occur (Coupling) if the vibrating bonds are joined to a single, central atom. This is because there is mechanical coupling interaction between the oscillators. Example: C=O (both symmetric and asymmetric stretching vibrations).
  • 17.
    The main partsof IR spectrometer are as follows: 1. radiation source 2. Sample cell 3. Monochromator 4. Detectors 5. Recorder
  • 18.
    INSTRUMENTATION SOURCE: Two types ofthe source is used. a) Globar b)Nernst glower Globar: a rod of silicon carbide is heated to approximately 1300 degree centigrade to produce the radiant energy from 1-40micron. Nernst glower: a rod of zirconium and yittrium is heated to approximately 1500 degree centigrade to produce the radiant energy from 0.4-20micron.
  • 20.
    INSTRUMENTATION Monochromator: it is samelike UV spectrophotometer but it is used to isolate the IR radiation of specific wavelength. Sample cells: For gas samples: The spectrum of a gas can be obtained by permitting the sample to expand into an evacuated cell, also called a cuvette. For solution sample: Infrared solution cells consists of two windows of pressed salt sealed. Samples that are liquid at room temperature are usually analyzed in pure form or in solution. The most common solvents are Carbon Tetrachloride (CCl4) and Carbon Disulfide (CS2).
  • 21.
    SAMPLING OF SOLIDS Generally4 techniques are employed for preparing solid samples: 1. Solids run in solution. 2. Solid Films. 3. Mull technique. 4. Pressed pellet technique.
  • 22.
    SOLIDS RUN INSOLUTION Solids may be dissolved in non-aqueous inert solvent and a drop of this solution is placed on an alkali metal disc and solvent is allowed to evaporate, leaving a thin film of solute (or the entire solution is placed in a liquid sample cell) which is then mounted in spectrometer. If the solution of solid can be prepared in a suitable solvent then the solution is run in concentration of cells for liquids. Some solvents used are chloroform, carbon tetrachloride, acetone, Cyclohexane etc.
  • 23.
    If a solidis polymer resins & amorphous solids, the sample is dissolved in any reasonable volatile solvent & this solution is poured on a rock salt plate (Nacl or KBr) & solvent is evaporated by gentle heating. If solid is non-crystalline, a thin homogenous film is deposited on the plate which can be mounted and scanned directly. Sometimes polymers can be “hot pressed” onto plates. SOLID FILMS
  • 24.
    MULLTECHNIQUE In this techniquea small quantity of sample is thoroughly ground in a clean mortar until the powder is very fine. After grinding, the mulling agent (mineral oil or Nujol) is introduced in small quantities just sufficient to take up the powder (mixture approximates the consistency of a toothpaste). The mixture is then transferred to the mull plates & the plates are squeezed together to adjust the thickness of the sample between IR transmitting windows. This is then mounted in a path of IR beam and the spectrum is run.
  • 25.
    Pressed pellet technique Inthis technique a small amount of finely ground solid sample is intimately mixed with about 100 times its weight of powdered Potassium bromide, in a vibrating ball mill. This finely ground mixture is then pressed under very high pressure in evacuable die or minipress to form a small pellet (about 1-2 mm thick and 1cm in diameter). The resulting pellet is transparent to IR radiation and is run as such.
  • 26.
  • 27.
    Pressed pellet technique(continue…) The powder (KBr + sample) is introduced in between the 2 bolts and the upper screw A is tightened until the powder is compressed to a thin disc. After compressing the sample bolts are removed and a steel cylinder with pellet inside it is placed in path of the beam of IR spectrometer and a blank KBr pellet of identical thickness is kept in the path of reference beam.
  • 28.
    Advantages of thistechnique over mull technique The use of KBr eliminates the problem of bands which appear in IR spectrum due to the mulling agent as in this case no such bands appear. KBr pellets can be stored for longer periods of time. As concentration of the sample can be suitably adjusted in pellets, it can be used for quantitative analysis. The resolution of spectrum in KBr is superior to that obtained with mulls.
  • 29.
    DETECTORS The measurement ofinfrared radiation is difficult as a result of low intensity of available source and low energy of the infrared photons. As a consequence of these properties, the electrical signal from the infrared detector is small and its measurement requires large amplification. Two types of IR detectors are used •Thermal detector •Non thermal detector
  • 31.
    TYPES OF THERMALDETECTOR There are four types of thermal detector. Bolometers Thermocouple and thermopile Pyro electric detector Golay cell
  • 32.
    BOLOMETER Bolometer is derivedfrom a Greek word (bolometron) Bolo = for something thrown Metron = measure A bolometer consists of an absorptive element, such as a thin layer of metal. Most bolometers use semiconductor or superconductor absorptive elements rather than metals. These material exhibit a relatively large change in resistance as a function of temperature.
  • 33.
    Working Step No 1 Thinlayer of metal connected to a reservoir Step No 2 Any radiation on the absorptive element raises its temperature above that of the reservoir. Step no 3 The temperature change can be measured directly with an attached thermometer.
  • 34.
    Temperature changes Potentialdifference changes Thermocouples consist of a pair of junctions of different metals; for example, two pieces of bismuth fused to either end of a piece of antimony. Whenever IR radiation fall at the junction, it causes the change in temperature and also a change in the potential difference that causes the production of electrical signal.
  • 35.
    Thermopile A thermopile isthe combination of six thermocouples connected in series. Half the junction are considered as hot and other half as cold which are thermally bonded to the substrate and maintains the stable temperature. The entire assembly is mounted in an evacuated enclosure with an IR transmitting window made up of KBr so as to minimize the heat loss. Thermopiles are one of the most simple and direct means of converting radiant energy into an electrical energy.
  • 36.
    Pyro electric Detector •Single crystalline wafer of a pyro electric material, such as triglycerine sulphate. • Pyro electric Infrared Detectors (PIR) convert the changes in incoming infrared light to electric signals.
  • 37.
    GOLAYCell It consist ofa small metal cylinder closed by a rigid blackened metal plate at one end and flexible silvered diaphragm at other end. The whole chamber is filled with xenon gas. Exposure to IR radiation through a radiation transmitting window causes the gas to expand and diaphragm to deform. Motion of the diaphragm changes the output of cell and detected in the form of signal.
  • 38.
    Non Thermal Detectors Photondetectors: Photoelectric detectors consists of a semiconducting material deposited on a glass surface, sealed in an evacuated envelop. Absorption of an IR promotes the nonconducting valence electron to a higher, conducting state. The electrical resistance of the semiconductor decreases.
  • 39.
    SINGLE BEAM SPCETROPHOTOMETER DOUBLE BEAM SPECTROPHOTOMETER Asingle beam of light , which can pass through one solution at a time (sample or reference). A single beam of light splits into two separate beams. One passes through the sample, another passes through the reference.
  • 40.
    Workingofdoublebeamspectrophotometer The radiation fromthe source is divided into two parts.  Reference beam  Sample beam The reference beam is passed through an optical wedge which limits the amount of radiations that passes through it. As the chopper is rotated, it causes the sample beam and the reference beam to be reflected alternatively to the monochromator which sends individual frequencies to detector. The detector receives an intense beam (the reference and the sample beam). This will lead to an alternating current which is flow to the amplifier.
  • 41.
  • 42.
    APPLICATIONOFIRSPECTROPHOTOMETER IR spectroscopy isone of the most important and widely used analytical techniques available to scientists working in a whole range of fields. The fundamental region in the IR region is termed as the rock salt region. It is comprised of group frequency region and the finger print region.
  • 43.
    GROUP FREQUENCY REGION FINGERPRINT REGION consistingof the absorption bands of the functional groups. frequency = 4000-1300cm-¹ wavelength = 2.5-8 IR spectra is called “fingerprints” because no other chemical species will have similar IR spectrum. Single bonds give their absorption bands in this region. Frequency=1300-650cm-1 Wavelength=8-15.4
  • 45.
    IdentificationofSubstances The finger printregion is mostly used to compare the identity of a compound. Because small differences in structure & constitution of molecule result in significant changes in the peaks in this region. Hence this region helps to identify an unknown compound. • Criteria: Sample and reference must be tested in identical conditions, like physical state, temperature, solvent, etc
  • 46.
    DeterminationofMolecularStructure Used along withother spectroscopic techniques. Identification is done based on position of absorption bands in the spectrum. Eg: C=O at 1717 cm-1. Absence of band of a particular group indicates absence of that group in the compd.
  • 47.
    DetectionofImpurities The presence ofabsorption bands at positions where compound is not expected to absorb indicates the presence of impurities. Determined by comparing sample spectrum with the spectrum of pure reference compound.
  • 48.
    Studyingthe progress ofreaction The progress of a reaction can be followed by examining the IR Spectra. Observing rate of disappearance of characteristic absorption band in reactants; or Rate of increasing absorption bands in products of a particular product. Example: Oxidation of secondary alcohol to ketone can be detected by disappearance of OH band and appearance of C=O band. O—H = 3600-3650 cm-1 C=O = 1680-1760 cm-1
  • 49.
    Molecularshapeorsymmetrydetermination The shape orsymmetry of a molecule can be determined by the IR spectroscopy. For example: If linear --> only 2 bands should be present. If bent --> 3 bands should be present. Actual spectrum shows 3 peaks at 750, 1323 and 1616 cm-1.
  • 50.
    Dipole Moment determination IRspectroscopy is also used to evaluate the dipole moment of the molecule and compounds.
  • 51.
    FOURIERTRANSFORMIRSPECTROPHOTOMETER It is themost advance IR spectrophotometer . In this a laser beam is used, which is allowed to fall on an interferometer. This photometer has advantage of fast speed of analysis, so very accurate values of molecular parameters are found with FTIR. A computer is also linked to get the store data.