MACRORADIOGRAPHY
Presented by :
M.Naga Theja
210513003
M.Sc.Radio-Imaging Technology
1st year
Table of Content
● What is macroradiography ?
● Principles of macroradiography
● Magnification factor
● Factors affecting image quality
○ Geometric unsharpness
○ X-ray tube rating
○ Photographic unsharpness
○ Scattered radiation
● Aim & applications of macroradiography
● References
● Questions
What is Macroradiography ?
● Macroradiography is a radiographic
imaging technique used to increase
the size of the image relative to the
object.
● In 1982 an Italian doctor named
Alessandro Vallebone proposed this
technique.
Principles of Macroradiography.
● A technique of producing an image by
direct magnification using an x-ray
tube with very fine focal spot.
● Focal spot-object distance will be
adequatly changed in accordance with
the change in accordance with the
change in magnification.
[Cotd…]
● Magnified image can be produced by increasing the object to film
distance from the point source.
● The required magnification is obtained by moving the object away from
the film.
[Cotd…]
● However, the geometric unsharpness increases with the increase of
magnification.
Magnification Factor
● The ratio of image size to object size is known as magnification factor. Its
value is normally assessed by measuring the focus - film distance (FFD)
and the object - film distance (OFD). the focus - object distance (FOD) is
then obtained by subtraction:
FOD = FFD - OFD
● Thus ,
Magnification factor = FFD/FFD-OFD=FFD/FOD
[Cotd…]
● For a routine examination of the hand (dorsipalmar projection), the
relevant dimensions are typically: FFD = I 00 cm, OFD = I cm. This gives
a magnification factor of 100/99 = 1.01. Thus, a 10 cm long finger gives an
image which is 10.1 cm in length.
● For a lateral projection of the lumbar spine, the relevant dimensions arc
typically: FFD = I 00 cm, OFD = 25 cm (spine-table-top distance + table-
top-film distance). This gives a magnification factor of 100/75 = 1.3. A true
measurement of 10 cm in the spine is reproduced as 13 cm in the image
[Cotd…]
● For macroradiography, it is usual to arrange the FFD and OFD to give a
magnification factor of at least 1.5. Typically, a magnification factor of2.0
is achieved by positioning the structure under examination half-way
between the X-ray tube focus and the film, i.e. FFD = FOD x 2 (sec Fig.
26.2). In practice, the FOD is often maintained at its normal value but the
FFD is increased, i.e. the X-ray tube and patient are positioned normally
hut the film is moved further from the patient.
Factors affecting Image Quality
● The quality of the magnified image is also an important consideration,
which may be affected by modifications to the FFD, FOD and OFD.
● Geometrical factors involves,
○ Geometric Unsharpness
○ X-Ray tube rating
○ Photographic Unsharpness
○ Scattered radiation
Geometric Unsharpness
● Geometric unsharpness refers to the loss of sharpness of feature on
radiograph that correspond to the boundaries from thickness that result
of geometric factor of the radiographic equipment and setup.
● To reduce this the the point source focal spot used.
● In the 1st image, radiation
originates at a very small
source, so very little geometric
unsharpness is produced in the
image.
● In the 2nd image, the source is
large & different path the rays
of radiation can take from their
point of origin in source that
cause the edges of the notch to
be less defined.
Source
Film
Film density
plot
Formulae of Geometric Unsharpness
● When the detector is not placed next to
sample, such as when the geometric
magnification is used then,
Ug = f . b/a
f = Source focal spot
a = Subject - Source distance
b = Subject - detector distance
Source focal spot
Subject
Detector
Penumbra
{Ug}
X_Ray Tube rating
● To limit geometric unsharpness and still achieve substantial
magnification requires the use of an ultrafine focus. Such a small focus
sets serious constraints on the output of the X-ray tube.
● This may well be manifested as a reduction in maximum tube current e.g.
from over 1000 mA on broad focus to less than 400 mA when fine focus
is selected, and consequently, exposure times become longer than
desirable. Unless great care is taken, motional unsharpness is the result.
Motional unsharpness problems can be alleviated by:
○ Taking positive measures to immobilize the patient;
○ Using fast film-screen systems, if necessary
[Contd…]
● Extremities is often highly successful because the required X-ray exposure
is small and effective immobilization is possible: thus, high-resolution
screens can be used.
● Macroradiography may prove impossible, e.g. for many abdominal
structures the large exposures necessary and the problems of involuntary
movement may exclude macroradiography as a viable technique.
● minimizing the FFD may ease rating problems but the radiographer must
be alert to the increased radiation dose to the patient associated with a
short focus-skin distance.
Photographic Unsharpness
● The resolution of the film or film-screen system is not normally a critical
limiting factor in carrying out macroradiography.
● Photographic unsharpness ( Up) remains constant because it is not the
recorded image that is magnified.
Scattered Radiation
● The large object-film distance used in macroradiography reduces the
amount of forward-scattered radiation reaching the film from the patient.
● It is not necessary to use a secondary-radiation grid, even for regions such
as the skull or spine where a grid would be essential.
● Collimation of the X-ray beam and provision of backscatter protection for
the film help to maintain optimum contrast on the macroradiograph.
Aim of Macroradiography.
● Aimed to improve the study of the bony structures and the
pulmonary designs.
● Allows the small detail obvious in the image.
Application of Macroradiography
● Carpal bones of the wrist (especially the scaphoid bone)
● Bones of hand, e.g., early detection of metabolic bone disease.
● Temporal bone.
● The lacrimal system during Dacryocystography.
● Nodular pattern in chest radiographs.
● Cerebral Angiography.
References
● Macroradiography using conventional radiography.
○ https://doi.org/10.1259/bjr.75.898.750831
● Chesney's Radiographic Imaging 6th edition
Questions ?
1. Whats are the 2 requirements of successful macroradiography ?
2. How is macroradiography is achieved ?
3. What is the main disadvantage of macroradiography ?
4. What is the formulae of geometric unsharpness ?
5. Who proposed macroradiography ?
MACRORADIOGRAPHY.pptx

MACRORADIOGRAPHY.pptx

  • 1.
    MACRORADIOGRAPHY Presented by : M.NagaTheja 210513003 M.Sc.Radio-Imaging Technology 1st year
  • 2.
    Table of Content ●What is macroradiography ? ● Principles of macroradiography ● Magnification factor ● Factors affecting image quality ○ Geometric unsharpness ○ X-ray tube rating ○ Photographic unsharpness ○ Scattered radiation ● Aim & applications of macroradiography ● References ● Questions
  • 3.
    What is Macroradiography? ● Macroradiography is a radiographic imaging technique used to increase the size of the image relative to the object. ● In 1982 an Italian doctor named Alessandro Vallebone proposed this technique.
  • 4.
    Principles of Macroradiography. ●A technique of producing an image by direct magnification using an x-ray tube with very fine focal spot. ● Focal spot-object distance will be adequatly changed in accordance with the change in accordance with the change in magnification.
  • 5.
    [Cotd…] ● Magnified imagecan be produced by increasing the object to film distance from the point source. ● The required magnification is obtained by moving the object away from the film.
  • 6.
    [Cotd…] ● However, thegeometric unsharpness increases with the increase of magnification.
  • 7.
    Magnification Factor ● Theratio of image size to object size is known as magnification factor. Its value is normally assessed by measuring the focus - film distance (FFD) and the object - film distance (OFD). the focus - object distance (FOD) is then obtained by subtraction: FOD = FFD - OFD ● Thus , Magnification factor = FFD/FFD-OFD=FFD/FOD
  • 8.
    [Cotd…] ● For aroutine examination of the hand (dorsipalmar projection), the relevant dimensions are typically: FFD = I 00 cm, OFD = I cm. This gives a magnification factor of 100/99 = 1.01. Thus, a 10 cm long finger gives an image which is 10.1 cm in length. ● For a lateral projection of the lumbar spine, the relevant dimensions arc typically: FFD = I 00 cm, OFD = 25 cm (spine-table-top distance + table- top-film distance). This gives a magnification factor of 100/75 = 1.3. A true measurement of 10 cm in the spine is reproduced as 13 cm in the image
  • 9.
    [Cotd…] ● For macroradiography,it is usual to arrange the FFD and OFD to give a magnification factor of at least 1.5. Typically, a magnification factor of2.0 is achieved by positioning the structure under examination half-way between the X-ray tube focus and the film, i.e. FFD = FOD x 2 (sec Fig. 26.2). In practice, the FOD is often maintained at its normal value but the FFD is increased, i.e. the X-ray tube and patient are positioned normally hut the film is moved further from the patient.
  • 11.
    Factors affecting ImageQuality ● The quality of the magnified image is also an important consideration, which may be affected by modifications to the FFD, FOD and OFD. ● Geometrical factors involves, ○ Geometric Unsharpness ○ X-Ray tube rating ○ Photographic Unsharpness ○ Scattered radiation
  • 12.
    Geometric Unsharpness ● Geometricunsharpness refers to the loss of sharpness of feature on radiograph that correspond to the boundaries from thickness that result of geometric factor of the radiographic equipment and setup. ● To reduce this the the point source focal spot used.
  • 13.
    ● In the1st image, radiation originates at a very small source, so very little geometric unsharpness is produced in the image. ● In the 2nd image, the source is large & different path the rays of radiation can take from their point of origin in source that cause the edges of the notch to be less defined. Source Film Film density plot
  • 14.
    Formulae of GeometricUnsharpness ● When the detector is not placed next to sample, such as when the geometric magnification is used then, Ug = f . b/a f = Source focal spot a = Subject - Source distance b = Subject - detector distance Source focal spot Subject Detector Penumbra {Ug}
  • 15.
    X_Ray Tube rating ●To limit geometric unsharpness and still achieve substantial magnification requires the use of an ultrafine focus. Such a small focus sets serious constraints on the output of the X-ray tube. ● This may well be manifested as a reduction in maximum tube current e.g. from over 1000 mA on broad focus to less than 400 mA when fine focus is selected, and consequently, exposure times become longer than desirable. Unless great care is taken, motional unsharpness is the result. Motional unsharpness problems can be alleviated by: ○ Taking positive measures to immobilize the patient; ○ Using fast film-screen systems, if necessary
  • 16.
    [Contd…] ● Extremities isoften highly successful because the required X-ray exposure is small and effective immobilization is possible: thus, high-resolution screens can be used. ● Macroradiography may prove impossible, e.g. for many abdominal structures the large exposures necessary and the problems of involuntary movement may exclude macroradiography as a viable technique. ● minimizing the FFD may ease rating problems but the radiographer must be alert to the increased radiation dose to the patient associated with a short focus-skin distance.
  • 17.
    Photographic Unsharpness ● Theresolution of the film or film-screen system is not normally a critical limiting factor in carrying out macroradiography. ● Photographic unsharpness ( Up) remains constant because it is not the recorded image that is magnified.
  • 18.
    Scattered Radiation ● Thelarge object-film distance used in macroradiography reduces the amount of forward-scattered radiation reaching the film from the patient. ● It is not necessary to use a secondary-radiation grid, even for regions such as the skull or spine where a grid would be essential. ● Collimation of the X-ray beam and provision of backscatter protection for the film help to maintain optimum contrast on the macroradiograph.
  • 19.
    Aim of Macroradiography. ●Aimed to improve the study of the bony structures and the pulmonary designs. ● Allows the small detail obvious in the image. Application of Macroradiography ● Carpal bones of the wrist (especially the scaphoid bone) ● Bones of hand, e.g., early detection of metabolic bone disease. ● Temporal bone. ● The lacrimal system during Dacryocystography. ● Nodular pattern in chest radiographs. ● Cerebral Angiography.
  • 20.
    References ● Macroradiography usingconventional radiography. ○ https://doi.org/10.1259/bjr.75.898.750831 ● Chesney's Radiographic Imaging 6th edition
  • 21.
    Questions ? 1. Whatsare the 2 requirements of successful macroradiography ? 2. How is macroradiography is achieved ? 3. What is the main disadvantage of macroradiography ? 4. What is the formulae of geometric unsharpness ? 5. Who proposed macroradiography ?