Mass analyzers separate ionized molecules based on their mass-to-charge ratios. The main types are quadrupole, time-of-flight, magnetic sector, quadrupole ion trap, and ion cyclotron resonance. A quadrupole uses oscillating electric fields to selectively transmit ions through four rods. Time-of-flight separates ions by their time of flight through a field-free region, with lighter ions arriving first. Magnetic sector analyzers use magnetic and electric fields to curve ion trajectories based on m/z.
Mass analyzer isalso called ion separator.
Mass analyzer is the Heart of the mass spectrometer that
takes ionized masses and separates them based on charge to
mass ratios.
Mass analyzer must posses the following characteristics.
I.It should have a high resolution.
II.It must have a high rate of transmission of ions.
4.
TYPES OF MASSANALYZERS
Quadrupole mass analyzer
Time of flight analyzer (TOF)
Magnetic sector mass analyzer
Single focusing
Double focusing
Quadrupole ion trap mass analyzer
Ion cyclotron resonances
It consists of4 voltage carrying rods.
The ions are pass from one end to another end
During this apply the radiofrequency and voltage complex
oscillations will takes place.
Here the single positive charge ions shows the stable
oscillation and the remaining the shows the unstable
oscillations
Mass scanning is carried out by varying each of the rf and
voltage frequencies ratios keeping their ratios constant.
◦Quadrupole ion storage (ion trap)
◦It store the unsorted ions temporarily, they released to the
detector by scanning the electric field.
7.
Advantage:
Classicalmass spectra.
Good reproducibility.
Relatively small and low-cost systems.
Limitations:
Limited resolution.
Peak heights variable as a function of mass.
Not well suited for pulsed ionization methods.
2.TIME OF FLIGHTMASS ANALYZER
(TOF)
In this type of analyzer the sorting of ions is done in absence
of magnetic field.
The ions produced are acquiring different velocities
depending on their masses
Here the particles reach the detector in the order of the
increasing order of their masses
Here electron multiplier detector is used.
The resolution power of this is 500-600
11.
ADVANTAGES
Theoreticallyunlimited mass range.
Moderate to high resolving powers .
Moderate cost .
Relatively high duty cycle.
Instrument is not scanning.
It has horseshoe shaped glass tube which is evacuated,
consists of sample inlet, electron bombarding source,
accelerating plates on one end,& collector slit at other end.
At curvature of tube there is provision to apply
electric/magnetic field
Sample in the form vapour is allowed through inlet and
bombarded with electron beam at 70eV.
It knocks off one electron from every molecule then they
become +vely charged ion.
As these molecules become +ve charged, they are
accelerated by accelerating plates and travel in straight path.
14.
By application ofelectric or magnetic field they travel in
curved path & molecular ions are separated according to
their masses and collected.
Different fragments fall on detector then mass spectrum is
recorded.
It is useddifferentiate the small mass differences of the
fragment.
These provides the high resolution
To achieve better focusing, energy has to be reduced before
ions are allowed to enter the magnetic field and increase
resolving power can be obtained two mass analyzers in series.
In a double-focusing mass analyzer beam is first passes
radial Electrostatic field.
Advantages:
Classical mass spectra.
Very high reproducibility.
High resolution.
High sensitivity.
High dynamic range.
.
17.
Limitations:
Not well-suitedfor pulsed ionization methods (e.g. MALDI).
Usually larger and higher cost than other mass analyzers.
Applications:
All organic MS analysis methods.
Accurate mass measurement.
Isotope ratio measurements.
18.
4.QUADRUPOLE ION TRAPMASS
ANALYZERS
There are two principal ion trapped mass analyzers
Three- dimensional quadrupole ion traps ("dynamic" traps),
and ion cyclotron resonance mass spectrometers
("static" traps).
Both Operate by storing ions by using DC and RF electric
fields.
20.
5.ION CYCLOTRON RESONANCES
Ions in a magnetic field move in circular orbits characteristic
of their m/z values.
If energy is provided at a frequency equal to their precession
frequency, and in a direction perpendicular to their plane of
precession,
Ions will absorb the energy, enabling them to be detected.
21.
Advantage :
The highest recorded mass resolution of all mass
spectrometers.
Powerful capabilities for ion chemistry and MS/MS
experiments.
Well-suited for use with pulsed ionization methods such as
MALDI.
Non-destructive ion détection; ion remesurèrent.
Stable mass calibration in superconducting magnet FTICR
systems.
22.
Limitations:
Limited dynamicrange.
Strict low-pressure requirements mandate an external
source for most analytical applications.
Subject to space charge effects and ion molecule
reactions.
Applications:
Ion chemistry.
High-resolution electrospray experiments for high-mass
analytes.
Laser desorption for materials and surface
characterization.