Amico Source Equipment Sizing Guide

This document provides a guide for sizing vacuum and air compressor systems for a medical facility based on the NFPA method. It includes tables to input the number and flow requirements of outlets for various locations in the facility, such as operating rooms, ICUs, and patient rooms. Based on the inlet values, it calculates the peak demand for vacuum and medical air and provides recommendations for oil-less scroll or reciprocating compressor systems to meet the demand. Contact information is also provided for project management and technical support.

Uploaded by

Jose Yosh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLS, PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
439 views4 pages

Amico Source Equipment Sizing Guide

This document provides a guide for sizing vacuum and air compressor systems for a medical facility based on the NFPA method. It includes tables to input the number and flow requirements of outlets for various locations in the facility, such as operating rooms, ICUs, and patient rooms. Based on the inlet values, it calculates the peak demand for vacuum and medical air and provides recommendations for oil-less scroll or reciprocating compressor systems to meet the demand. Contact information is also provided for project management and technical support.

Uploaded by

Jose Yosh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLS, PDF, TXT or read online on Scribd

NFPA EQUIPMENT SIZING GUIDE - VACUUM SYSTEM

Name of Facility:
Project Number:
ENTER REQUIRED DATA IN THE BLUE SHADED AREAS BELOW

VACUUM - NFPA Method


LOCATION OF OUTLETS

Units Required

Units

Outlet CFM

Simutanious Use
%

Outlet CFM

Anesthetizing Location
1
2
3

Operating Room
Cystoscopy
Delivery

0
0
0

Room(s)
Room(s)
Room(s)

3.5
2
1

100
100
100

4
5
6
7

Special Proced. (Open heart, Transplant, etc.)


Emergency/ Major Trauma Room
Other Anesthetizing Location
Evacuation Vacuum

0
0
0
0

Room(s)
Room(s)
Room(s)
Room(s)

4
3
1
2

100
100
50
100

1
2
3
4

(Non-Anesthetizing)
Recovery Room
I.C.U (Except Cardiac)
Cardiac I.C.U
Emergency Room

0
0
0
0

Bed(s)
Bed(s)
Bed(s)
Bed(s)

1.5
2
1
1

50
75
50
100

5
6
7
8

Special Procedures (X-Ray, dialysis, etc.)


Catherterization Lab
Surgical Excision Room
Neonatal I.C.U,

0
0
0
0

Room(s)
Room(s)
Room(s)
Bed(s)

1.5
1
1
1

30
10
10
50

1
2
3
4
5

(Non-Anesthetizing)
Patient Room - Surgical
Patient Room - Medical
Exam & Treatment Room
Nursery
Nursery Premature

0
0
0
0
0

Bed(s)
Bed(s)
Room(s)
Bassinet(s)
Bassinet(s)

1.5
1
1
1
1

15
10
10
10
25

1
2
3
4
5
6
7

Respiratory Care Dep't


Teaching
Autopsy
Location1: Vac*
Location2:
Location3:
Location4:

0
0
0
0
0
0
0

Room(s)
Room(s)
Table(s)
Outlet(s)
Outlet(s)
Outlet(s)
Outlet(s)

1.5
1.5
1.5
3

10
10
10
100

Acute Care Location

Sub-Acute Patient Care

Other Patient Rooms

0.00

Future Expansion
PEAK CALCULATED DEMAND IN SCFM @ 19"hg:
Altitude above sea level, ft:
PEAK CALCULATED DEMAND IN SCFM @ 19"hg, ALTITUDE ADJUSTED:

0.00
#N/A

NOTE: The system model below are provided as a reference and variation of the systems are available. Please contact your
local sales rep for details.
* Maybe used alone if you wish to only count outlets needed. You may change that field as you see fit.

Amico Source Contact Information


Project Management - USA
Project Management - Canada
Project Management - International
Parts
Tech Support

Telephone
1-877-462-6426
1-877-462-6426
1-877-462-6426
1-877-462-6426
1-888-770-6035

E-MAIL
as-usa@[Link]
as-canada@[Link]
as-intl@[Link]
as-parts@[Link]
as-techsupport@[Link]

Lubricated Rotary Vane System


System Model #

Drawing Link

Tank Mount Horizontal

#N/A

#N/A

Tank Mount Space Saver

#N/A

#N/A

Duplex Skid Mount

#N/A

#N/A

Triplex Skid Mount

#N/A

#N/A

Quadraplex Skid Mount

#N/A

#N/A

Dry Rotary Vane System


System Model #

Drawing Link

Tank Mount Horizontal

#N/A

#N/A

Tank Mount Space Saver

#N/A

#N/A

Duplex Skid Mount

#N/A

#N/A

Triplex Skid Mount

#N/A

#N/A

Quadraplex Skid Mount

#N/A

#N/A

Dry Contactless Claw System


System Model #

Drawing Link

Tank Mount Horizontal

#N/A

#N/A

Tank Mount Space Saver

#N/A

#N/A

Duplex Skid Mount

#N/A

#N/A

Triplex Skid Mount

#N/A

#N/A

Quadraplex Skid Mount

#N/A

#N/A

NFPA EQUIPMENT SIZING GUIDE - AIR COMPRESSOR SYSTEM


Name of Facility:
Project Number:
ENTER REQUIRED DATA IN THE BLUE SHADED AREAS BELOW

LOCATION OF OUTLETS

MEDICAL AIR - US Typical Method


Simutanious Use
Outlet CFM
%

Units Required

Units

Outlet CFM

Surgical Procedures
1
2
3
4
5
6
7

PreOp Holding
Major Invasive
Minor Invasive
Trauma and Emergency
Catherization & Lab work
Endoscopy
Recovery

1
2
3
4
5
6
7

X-ray, CAT, NMR, PET scans


Dialysis
Exam and Minor Treatment
EEG/EKG
Pulmonary Function
Respiratory Care
Observation

1
2
3
4

Birthing or LDRP
Delivery Room
Nursery
NICU

1
2
3

Adult ICU, CCU, etc.


Pediatric ICU (Except NICU)
Emergency (not surgical)

Workrooms

Medical Lab Uses

1
2
3
4
5
6

Patient rooms
Ventilator
Location1: Critical*
Location2: Non Critical*
Location3: Support Gas*
Location4:

0
0
0
0
0
0
0

Bed(s)
Room(s)
Room(s)
Room(s)
Room(s)
Room(s)
Bed(s)

1
3.5
2
3
1
2
1.5

10
100
100
100
10
100
50

0
0
0
0
0
0
0

Room(s)
Bed(s)
Room(s)
Room(s)
Bed or outlet
Bed or outlet
Bed or outlet

1.5
1.5
1
1
1.5
1.5
1

30
30
10
10
30
30
10

0
0
0
0

Room(s)
Room(s)
Bed(s)
Bed(s)

1
0.5
0.5
1.5

100
100
10
50

0
0
0

Bed(s)
Bed(s)
Bed(s)

2
2
2

50
50
10

Outlet(s)

1.5

10

Outlet(s)

1.5

25

0
0
0
0
0
0

Room(s)
Unit(s)
Outlet(s)
Outlet(s)
Outlet(s)
Outlet(s)

0.5
3
6
3.5
5

10
50
100
100
100

Tests and Outpatient Procedures

Perinatal and Pediatric

Intensive Care

Equipment Maintenance

Laboratory

Other

0.00

Future Expansion
PEAK CALCULATED DEMAND IN SCFM:
NOTES:
1. The SCFM above is based on a peak OUTLET demand of 50 psi. Air compressors operate at 100 psi. When sizing a compressor, you MUST
figure the compressor capacity for operation at 100 psi.

0.00

2. Air outlets in labs used for analysis, research or teaching should be supplied by a seperated compressed air system, not medical air system.
3. * Maybe used alone if you wish to only count outlets needed. You may change that field as you see fit.
Altitude above sea level, ft:
Intake air temperature, F:
Relative Humidity, %:

0%

PEAK CALCULATED DEMAND IN SCFM, ADJUSTED:


NOTE: The system model below are provided as a reference and variation of the systems are available. Please contact your
local sales rep for details.

Amico Source Contact Information


Project Management - USA
Project Management - Canada
Project Management - International
Parts
Tech Support

Telephone
1-877-462-6426
1-877-462-6426
1-877-462-6426
1-877-462-6426
1-888-770-6035

E-MAIL
as-usa@[Link]
as-canada@[Link]
as-intl@[Link]
as-parts@[Link]
as-techsupport@[Link]

Oil-less Scroll System


System Model #

Drawing Link

Tank Mount Horizontal

#N/A

#N/A

Duplex Skid Mount

#N/A

#N/A

Triplex Skid Mount

#N/A

#N/A

Quadraplex Skid Mount

#N/A

#N/A

Oil-less Reciproating System


System Model #

Drawing Link

Tank Mount Horizontal

#N/A

#N/A

Duplex Skid Mount

#N/A

#N/A

Triplex Skid Mount

#N/A

#N/A

Quadraplex Skid Mount

#N/A

#N/A

#N/A