Acclarix AX8 User Manual 3371637
Acclarix AX8 User Manual 3371637
Preface
This manual contains necessary and sufficient information to use the Acclarix AX8 system safely
for the purposes for which it is intended. It applies to all 1.3X releases.
Read and understand all instructions in this manual prior to using the Acclarix AX8 ultrasound
system.
Disregarding instructions, particularly warnings and cautions, is considered abnormal use.
Caution: Federal law restricts this device to use by, or on the orders of, a physician.
Contact Information:
For sales or service information please contact your local distributor or the EDAN service
department at: [email protected]
P/N: 01.54.002481
MPN: 01.54.002481014
Release Date: Dec., 2016
© Copyright EDAN INSTRUMENTS, INC. 2015-2016
I
Contents
1 Introduction................................................................................................................................ 1
1.1 Intended Use/ Indications for Use .......................................................................................... 1
1.2 Contra-indications .................................................................................................................. 1
1.3 Device Description ................................................................................................................. 1
1.4 Labeling Symbols .................................................................................................................. 2
2 Safety ......................................................................................................................................... 5
2.1 Warnings................................................................................................................................. 5
2.2 Cautions ................................................................................................................................. 7
3 Getting Started ......................................................................................................................... 10
3.1 System Configuration .......................................................................................................... 10
3.2 System Overview ................................................................................................................. 11
3.3 Battery Use ........................................................................................................................... 13
3.4 Connecting and Disconnecting a Transducer ....................................................................... 15
3.5 Powering on/ off ................................................................................................................... 17
3.6 Screen Layout....................................................................................................................... 18
3.7 Control Panel........................................................................................................................ 21
3.8 Touch Screen ........................................................................................................................ 24
3.9 Trackpad ............................................................................................................................... 25
4 Exam Operation ....................................................................................................................... 27
4.1 How to Start an Exam .......................................................................................................... 27
4.2 How to End an Exam ........................................................................................................... 27
4.3 How to Restart an Exam ...................................................................................................... 27
4.4 The Patient Information Page ............................................................................................... 28
4.5 Modality Worklist ................................................................................................................ 30
5 Imaging .................................................................................................................................... 32
5.1 B-mode ................................................................................................................................. 32
5.1.1. B-mode Touch Screen Controls ................................................................................. 32
5.1.2. Dual Imaging ............................................................................................................. 35
5.1.3. Quad Imaging ............................................................................................................ 35
5.1.4. Panorama ................................................................................................................... 35
5.1.5. Zoom .......................................................................................................................... 36
5.2 PW-mode .............................................................................................................................. 37
5.2.1. PW Touch Screen Controls ........................................................................................ 37
5.2.2. HPRF ......................................................................................................................... 39
5.2.3. Update ........................................................................................................................ 39
5.3 CW-mode ............................................................................................................................. 39
5.3.1. CW Touch Screen Controls........................................................................................ 39
5.4 Color-mode .......................................................................................................................... 41
5.4.1. Color Touch Screen Controls ..................................................................................... 41
5.4.2. Color Mode Variants .................................................................................................. 43
5.5 M-mode ................................................................................................................................ 43
5.5.1. M-mode Touch Screen Controls ................................................................................ 43
5.6 3D/4D Mode* ...................................................................................................................... 44
II
5.6.1. Pre-3D ........................................................................................................................ 45
5.6.2. 3D Volume Sweeping ................................................................................................ 46
5.6.3. 3D Image Review ...................................................................................................... 46
5.6.4. 4D Volume Acquisition .............................................................................................. 51
5.6.5. 4D Live Volume ......................................................................................................... 52
5.6.6. 4D Cine ...................................................................................................................... 52
6 Transducers .............................................................................................................................. 53
6.1 Transducer Model ................................................................................................................ 53
6.2 Name and Function of Each Part of the Transducer ............................................................ 54
6.3 Transducer Cleaning and Disinfecting ................................................................................. 54
7 Features .................................................................................................................................... 58
7.1 Comments ............................................................................................................................ 58
7.2 Body Mark ........................................................................................................................... 59
7.3 Measurements ...................................................................................................................... 60
7.3.1. Generic Measurements .............................................................................................. 63
7.3.2. Application Measurements and Calculations ............................................................ 69
7.4 Patient Information Management ......................................................................................... 73
7.4.1. Storing Images ........................................................................................................... 73
7.4.2. Reviewing Images...................................................................................................... 73
7.4.3. Patient Database ......................................................................................................... 75
7.4.4. Archiving Studies....................................................................................................... 77
7.5 Needle Biopsy Guide ........................................................................................................... 78
7.5.1. Installing Needle Guide Bracket ................................................................................ 78
7.5.2. Activating Needle Guide Function ............................................................................ 82
7.5.3. To Adjust the Needle Guide Line .............................................................................. 83
7.6 Needle Visualization ............................................................................................................ 83
8 Presets ...................................................................................................................................... 85
8.1 Preset Organization .............................................................................................................. 85
8.2 Selecting a Preset ................................................................................................................. 86
8.3 Storing a Preset .................................................................................................................... 86
8.3.1. Presets ........................................................................................................................ 87
8.3.2. Comments Presets ...................................................................................................... 90
8.3.3. Bodymarker Presets ................................................................................................... 91
8.3.4. Measure Presets ......................................................................................................... 93
9 Utilities .................................................................................................................................... 97
9.1 Set-up ................................................................................................................................... 97
9.1.1. General Set-up ........................................................................................................... 97
9.1.2. Patient Set-up ............................................................................................................. 99
9.1.3. Store/Print Set-up ..................................................................................................... 100
9.1.4. Miscellaneous Set-up ............................................................................................... 101
9.1.5. User Set-up .............................................................................................................. 102
9.2 Connectivity ....................................................................................................................... 103
9.2.1. TCP/IP ..................................................................................................................... 103
9.2.2. DICOM .................................................................................................................... 104
9.2.3. Network Store .......................................................................................................... 107
9.3 Maintenance ....................................................................................................................... 108
9.3.1. License ..................................................................................................................... 108
9.3.2. Version ..................................................................................................................... 108
III
9.3.3. Demo ........................................................................................................................ 109
9.3.4. Export/Import........................................................................................................... 109
9.4 Screen Adjust ..................................................................................................................... 111
10 In Between Exams ................................................................................................................. 113
10.1 Unpacking ........................................................................................................................ 113
10.2 Transport .......................................................................................................................... 113
10.3 Storage.............................................................................................................................. 113
11 Order List ............................................................................................................................... 114
12 Troubleshooting and Maintenance ......................................................................................... 116
12.1 Daily Checklist ................................................................................................................. 116
12.2 Troubleshooting ............................................................................................................... 116
12.3 Maintenance ..................................................................................................................... 117
12.4 Cleaning the System......................................................................................................... 117
12.4.1. Cleaning the System Surface ................................................................................... 118
12.4.2. Needle Guide Bracket Cleaning and Sterilization ................................................... 118
13 Specifications ......................................................................................................................... 119
13.1 Electrical Safety Classifications ....................................................................................... 119
13.2 Power Supply ................................................................................................................... 119
13.3 Machine Specifications .................................................................................................... 120
13.4 Display Specifications...................................................................................................... 120
13.5 General Technical Specifications ..................................................................................... 120
13.6 Operating, Storage and Transportation Environment....................................................... 121
13.6.1. Operating Environment ............................................................................................ 121
13.6.2. Storage and Transportation Environment................................................................. 121
14 EMC Information................................................................................................................... 122
IV
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
1 Introduction
1.1 Intended Use/ Indications for Use
For the United States of America
The Edan Acclarix AX8 Ultrasound system is intended for use by a qualified physician or allied
health professional for ultrasound evaluations. Specific clinical applications include:
Abdominal
Gynecology (including endovaginal)
Obstetric
Cardiac
Small parts (Breast, Testes, Thyroid, etc.)
Urology
Musculoskeletal
Peripheral vascular
Intra-operative
1.2 Contra-indications
The Edan Acclarix AX8 Ultrasound system is not intended for ophthalmic use.
1.3 Device Description
The Edan Acclarix AX8 Ultrasound system consists of a main system along with associated
transducers.
The system circuitry generates an electronic voltage pulse, which is transmitted to the
transducer. In the transducer, a piezo electric array converts the electronic pulse into an
ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body
tissues. The waves are then reflected within the body and detected by the transducer, which
-1-
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
then converts back to an electrical signal. The Acclarix AX8 system then analyzes the returned
signal to generate an image or conduct Doppler processing.
The Acclarix AX8 system gives the operator the ability to measure anatomical structures, and
offers analysis packages that provide information used by competent health care professionals
to make a diagnosis.
The system provides both touch screen and hard buttons for the User Interface.
*Descriptions of transducer C5-2MQ, adult cephalic application, 3D/4D function, PW one-key
optimization, acoustic power control, panorama measurement and wireless function are not
approved for use until finishing the related registration.
1 Serial Number
3 Date of Manufacture
4 Manufacturer
5 Operating instructions
Warning
6
(Background: Yellow; Symbol & outline: Black )
8 Caution
9 Biological Risks
The symbol indicates that the device complies with the European Council
10
Directive 93/42/EEC concerning medical devices.
-2-
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
Caution: Federal (U.S.) law restricts this device to sale by or on the order
14
of a physician.
18 Transducer connector
21 Transducer lock
22 Transducer unlock
23 Network port
24 Trademark
25 Direct current
-3-
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
29 Display port
30 Microphone input
31 Headphones
FCC ID:
32 Federal Communication Commission: FCC ID: SMQAX8EDAN
SMQAX8EDAN
NOTE:
The user manual is printed in black and white.
-4-
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
2 Safety
Throughout this document the following terms are used:
Warning: Advises against certain actions or situations that could result in personal injury
or death.
Caution: Advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
Note: Provides useful information regarding a function or a procedure.
Please read all warnings and cautions prior to using the Acclarix AX8 ultrasound system. For
your convenience all warnings and cautions are provided in this section. These may be duplicated
elsewhere in this document in the context of the instructions for use.
2.1 Warnings
Only use an Edan supplied power adapter and power cord.
Only use an Edan supplied battery. Read and understand the battery installation
instructions prior to changing the battery.
Only use Edan supplied transducer. Use of other transducers may result in electric
shock or system malfunction.
Only use a hospital grade, grounded, power outlet and plug. Do not use with an
ungrounded outlet.
The system is ordinary equipment (Sealed equipment without liquid proof). But the
console panel including touch screen, track pad and hard keys is IPX1.
Transducer (not including the transducer connector) is IPX7. Footswitch is IP68.
Do not immerse or expose any of the parts to extended moisture. Splash
resistance does not extend to transducer connectors. Keep connectors dry.
Do not use in a wet environment or when the relative humidity exceeds 95%.
Do not reverse the positive and negative poles when installing a battery.
Do not use the battery near heat sources or when the ambient temperature is over
40oC. Do not heat or dispose of in fire.
Do not destroy the battery; do not pierce or cause a strong impact to the battery.
Do not touch the connector pins on the transducer port.
Parts and accessories used must meet the requirements of the applicable
IEC/EN60601 series safety standards, and/or the system configuration must meet
the requirements of the IEC/EN60601-1.
Use protective barriers (gloves and transducer sheaths) whenever possible.
Follow sterile procedures when appropriate. Thoroughly clean Transducers and
reusable accessories after each patient examination and disinfect or sterilize as
needed. Refer to transducer use and care instructions. Follow all infection control
policies established by your office, department or institution as they apply to
personnel and equipment.
-5-
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
system grounding and cause the system to exceed leakage current limits.
SHOCK HAZARD – Do not connect non-isolated electrical equipment to the same
circuit being used to power the system.
Edan recommends the use of isolated connectors on any electrical equipment
attached to the system, and/or using isolation transformers that complies with
IEC60601-1 to power that electrical equipment.
Always use sterile technique during a biopsy procedure. Sterilize the needle
guide assembly between uses.
Use a sterile needle with each use.
Always confirm the clinical accuracy of the IMT result before entering it into the
report.
Transducer Warnings
To avoid infection, always use protective gloves when cleaning or disinfecting
Read and follow all manufacturer instructions for disinfection agents.
To avoid infection, ensure that expiration date of the disinfecting solution has not
passed.
Disinfect the transducer after each intra-cavity or intra-operative procedure. Use a
new sterile sheath for each such procedure.
Use a pyrogen-free transducer sheath for neurological intra-operative procedures.
Unplug the transducer from the system prior to cleaning or disinfecting.
Do not immerse the transducer beyond the point indicated in Figure 6-2.
Do not allow the transducer connector to get wet.
2.2 Cautions
The system contains no user serviceable components other than the battery. Do
not remove any covers.
Excessive dust and dirt could clog internal airflow and cause overheating. Do not
use in a dusty environment.
Do not use a battery that leaks, emits an odor, appears deformed, or discolored.
Immediately replace it with a new Edan-supplied battery and dispose of the old
battery according to local regulations. Replace a battery that has reached the end
of its service life.
Use care when storing or disposing of batteries. Do not allow the leakage to come
in contact with each other. Do not dispose of them together with household
garbage. At the end of their life hand the batteries over to the applicable collection
points for the recycling of waste batteries.
Inspect the system regularly, at least weekly. Before use ensure there is no visible
-7-
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
-9-
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
3 Getting Started
3.1 System Configuration
The Acclarix AX8 system is shipped with the following components:
Acclarix AX8 main unit
1 rechargeable lithium battery
1 UPS battery
1 AC adapter
1 bottle of coupling gel
1 basic user manual and 1 advanced user manual
The following options are also available (please see the Order List Appendix for details)
Transducers: C5-2XQ, C5-2Q, L10-4Q, L12-5Q, L17-7HQ, E8-4Q, MC8-4Q, P5-1XQ,
L17-7SQ, C5-2MQ*
1 Trolley MT-807
Footswitch
Suitcase
The Acclarix AX8 supports the following external DVD drives: SAMSUNG SE-218GN,
SAMSUNG SE-208GN, LENOVO DB75, ASUS SDRW-08D2S-U and PIONEER
DVR-XU01C.
The Acclarix AX8 supports the following DVD disks: Philips CD-R 700MB 80min 52x,
Sony CD-RW 700MB 1x-4x,Sony DVD-RW 4.7GB 120min 2x/1x,
Verbatim DVD+RW 4.7GB 120min 4x.
Note: Only support DVD disk single burning.
The Acclarix AX8 supports the following printers. Please contact the manufacturer for
details.
- 10 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
6
3
7
8
4
9
3
3
- 11 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
10
- 12 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
The Acclarix AX8 system comes with one lithium-ion battery. When it is fully charged it can run
the system for approximately 60 minutes, depending on use. It is automatically charged when the
system is plugged in.
CAUTION
1. If the system will remain unused for more than one week, charge the battery to at
least 75%capacity, take the battery out and store the system and battery separately.
2. During long term storage the battery should be charged at least once every 3 months
to ensure battery capacity is more than 75%.
- 13 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
- 14 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
- 15 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
To disconnect a transducer:
1. Toggle the locking handle to the top position to unlock the connection of the Transducer.
2. Firmly grasp the transducer connector and carefully remove it from the system port.
3. Store transducer in its protective carrying case prior to transport.
The Multi-Transducer Connector (MTC) is an optional addition to the AX8 that allows up to
three transducers to be connected to the system. The MTC can be mounted to the optional AX8
cart, or lie flat on an adjacent table.
- 16 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
Note:
When both Admin1 and Admin2 password are forgotten, please contact the
serviceman for the system password reset.
To power off:
1. Press the power on/off key on the top left of control panel and the system displays a
confirmation dialog box.
2. Select “Shut Down” from the confirmation dialog box.
If the system is unresponsive, a long press of the power on/off key will shut down the system
directly.
NOTE:
Please unplug the AC adapter from the power socket and disconnect the battery prior to
storage.
Sleep mode
The system will enter a sleep mode that maintains exam information while using minimal battery.
There are three events that can invoke sleep mode:
- 17 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
⑤
Figure 3-11 Main Screen Display
①Information Field:
The top line of this field contains your hospital/institution name. Please see Section 9.1.
General Set-up for information on customizing this area.
The second line of this field contains the patient name, gender, age and ID, as entered
through the Patient Information screen.
This field also contains data fields for:
The currently active transducer
- 18 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
: Battery 60%-80%.
: Battery 40%-60%.
: Battery charging.
Network status icon
The network icons show the connection status of the DICOM or FTP Server. If no
Network Server is defined then the icon does not show at all.
- 19 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
:
Outline in grey color: Successful connection with a Server.
Outline in green color: Data exchange with a Server.
Outline in red color: Failure to connect with a configured Server.
Hard drive icon.
:
Symbol in grey color: USB available
DVD icon
: Printer available.
Wifi icon *
WARNING
The system complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions:
1) this device may not cause harmful interference, and
2) this device must accept any interference received, including interference that may
- 20 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
NOTE:
1. This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
2. Any changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user's authority to operate the equipment.
- 21 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
Enters or exits the Body Mark function. See section 7.2 for
10. Body Mark
details.
- 22 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
- 23 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
- 24 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
3.9 Trackpad
In general the Acclarix AX8 trackpad is used like a trackball or similar pointing device used on
most ultrasound systems. Using a trackpad instead of a trackball helps make the console sealed,
and makes the system easier to clean. In addition, there are several capabilities that the trackpad
supports that would not be possible on a traditional trackball:
The trackpad supports multi-finger gestures. For example:
o While in B-mode a pinch-out gesture can be used to zoom the image.
o While in Color a pinch gesture can re-size the ROI.
Dragging along the Swipe button on top edge of the trackpad as a primary sliding control.
For example:
o In most live imaging modes this is Gain.
The top right of the trackpad is reserved as a „Set‟ button. Pressing this corner is generally
equivalent to a single click on a mouse or trackball. For example:
o In CD it is used to toggle between controlling the size and the position of the pan
box.
o In Measure it is used to toggle between controlling one side or the other side of a
caliper pair.
The top left of the trackpad is reserved as an „Enter‟ button. Pressing this corner is
- 25 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
- 26 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Exam Operation
4 Exam Operation
4.1 How to Start an Exam
To start a new exam, press the Patient key and enter in patient information for the exam, or select
a scheduled patient from the modality worklist.
If there is no previous exam this will bring you directly to the Patient Information Page (see
figure 4-2 below). If a previous exam is still active you will see the following dialog:
- 28 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Exam Operation
- 29 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Exam Operation
- 30 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Exam Operation
- 31 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
5 Imaging
The imaging modes are entered and exited via dedicated hard keys along the right side of the
console.
Key Description
The B-mode key behaves differently than the other mode keys. B-mode imaging is
the baseline state of the system. Therefore, the B key returns to B-mode imaging at
any time from any display on the system. See section 5.1 for B-mode controls.
Invokes and exits strip Doppler mode. Press D button and press PW/CW on the
touch screen to switch to CW mode (requires P5-1XQ transducer). See section
5.2/5.3 for Doppler controls.
Invokes and exits color modes, including Color, Power Doppler Imaging (PDI) and
Directional Power Doppler Imaging (DPDI). The default color mode is determined
by the preset and can be changed while Color is active. See section 5.4 for Color
controls.
Invokes and exits M-mode. See section 5.5 for M-mode controls
5.1 B-mode
The following touch screen controls impact the B-mode image. Touch screen controls can be
located where you want them to be. See section 3.8 for details.
Name Control Description
TGC The Time Gain Control (TGC) adjusts the gain of the image at
different depths. Each slider can be adjusted separately, or you can
drag your finger vertically to set all of the sliders to a new setting.
Auto The Auto push button controls automatically updates the Gain and
TGC. Each single press of the button renews the automatic
optimization.
Dynamic The Dynamic Range, or log compression, adjusts how echo
Range intensities are converted to brightness. A high dynamic range
presents a flat image, while a low dynamic range present a large
range of intensities.
- 32 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
eSRI eSRI is Speckle Reduction Imaging. There are 4 levels: Off, Low,
Med. and High. High level provides more aggressive speckle
reduction.
Harmonic The Harmonic control invokes and exits harmonic imaging. While in
harmonic imaging the control is highlighted and „H‟ in the B-mode
frequency field. Depending on the transducer there may be multiple
harmonic frequencies.
Spatial Spatial Compounding combines images fired from multiple angles
Compounding in order to reduce speckle, reduce shadow artifacts, and enhance
contrast. Spatial compounding is an on/off control.
Foc. Num Focus Number adjusts the number of focus positions. As the
number of foci increases the frame rate will decrease, but there will
be more image uniformity across depths.
Foc. Pos. Focus Position adjusts the depth of the focus or foci. Upward
presses move the focus shallower, regardless of the U/D invert
status of the image.
Gray Map Gray Map adjusts the post processing map used on the B-mode
image. In general, higher map numbers correspond to more
contrast in the image.
Colorize The Colorize control adds a color tint to the B-mode image.
Tint The Tint control changes the color tint being used. Gold, Sepia,
Blue, Ice and Clear are available. If Colorize had been off,
changing the Color map control will automatically activate it.
- 33 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Steer This control is only available for linear transducers. It steers the
B-mode image left or right, without moving the transducer. This can
be particularly useful when visualizing needles or other objects that
are enhanced by a perpendicular beam.(in Spatial Compounding,
Trapezoid and Panorama states, steer function is not valid)
Trapezoid The Trapezoid control activates the trapezoid imaging on linear
transducer. It is a part of the B-mode function and available in live
imaging.
Panorama The Panorama control invokes the Panorama function. See section
5.1.4 for details.
Image Type B-mode supports presets for Detail, General, and Penetration. See
section 8.3.1.1 for more information.
Frame rate Adjusts the tradeoff between frame rate and spatial resolution.
Higher settings provide more temporal resolution.
Dual B Press to activate dual imaging mode. Dual L activates the live
image on the left side of image field, and Dual R activates the live
image on the right side of image field.
Quad Press to activate quad imaging mode. Refer to section 5.1.3 Quad
Imaging for details.
3D/4D Press to activate 3D/4D imaging mode. Refer to section 5.6 3D/4D
mode for details.
Needle Press to invoke the touch screen menu for Needle Enhancement
Visualization and Needle Biopsy Guide functions. See section 7.5
and 7.6 for more information.
- 34 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Dual Imaging displays images side by side on the screen. During real-time imaging when Dual
imaging is enabled, the active image is displayed in real-time and the other image is frozen.
Press the Dual L or Dual R button on the B-mode touch screen to select the left or right image
for activation. With both images frozen, press the Dual L or Dual R button to toggle between the
images.
Dual Exit
Dual is exited by any of the following:
Pressing the Dual button corresponding to the currently active side.
Pressing the Quad button.
Pressing the B, M, D hard key.
Pressing the 3D/4D button.
Transducer change, preset recall, or new exam.
Color Doppler is available during dual imaging.
- 35 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
The B-mode image will extend to incorporate the newly imaged anatomy.
When the full anatomy has been acquired, press the Enter button on the trackpad or press
the Freeze hard key. System also will enter the panorama review state when exceeding the
default time.
After acquisition the completed Panoramic image is redisplayed to fit on the screen.
Measurements*, Comments and Bodymarkers are then supported.
Note:
1. If a measurement function in panorama mode is activated, the caution information of
“The accuracy of measurements in Panorama is limited and can be lower than
measurements in B-image” in yellow font color appears on the bottom of the image
and a caution symbol appears on the head of the measurement results.
2. IMT measurement is not available on panorama image.
Panorama is available only with linear transducers.
Speed Indicator
During an acquisition, a speed indicator bar on the screen will show the current scan speed. Keep
the indicator in the green center of the graphic in order to achieve best results.
5.1.5. Zoom
Zoom is available in live or frozen B-mode and Color imaging. Press the Zoom hard key to zoom
an image. There are 2 ways to zoom an image: The Zoom paddle key or a pinch-out gesture on
the track pad.
Zoom Paddle key
o Press the upper part of the Zoom paddle to make the image bigger.
o Press the lower part of the Zoom paddle to make the image smaller, down to normal
size.
Pinch Gesture: Use two fingers on the main part of the trackpad, and move the two fingers
outward from each other (“Pinch-out”). The image will enlarge.
When in Zoom mode, use the trackpad to pan the position of zoomed image. The B image
remains zoomed after entering other imaging modes.
- 36 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
5.2 PW-mode
Adjust the Doppler scale to account for the angle between the
Angle Doppler cursor and the blood flow. Upward presses increment
Correct the angle. Downward presses decrement the angle. Angles
above 80 are not allowed.
Gray Map
Adjusts the current postprocessing, either grayscale or tinted
Tint
- 37 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Volume adjusts the audio volume of the Doppler strip. This can
Volume be adjusted in pre-Doppler to set the initial volume upon invoking
Doppler acquisition.
- 38 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Auto The Auto push button automatically updates the PW gain, DR,
Optimization* baseline and PRF scale. Each single press of the button renews
the automatic optimization.
Whether Gain, DR or Scale/Baseline is optimized when pressing
Auto button can be configured in Set-up. See section 9.1.4 for
detail.
Table 5-3 PW Mode Touch Screen Controls
5.2.2. HPRF
Conventional PW Doppler scales are limited by the Nyquist limit. High Pulse Repetition
Frequency (HPRF) allows the system to exceed the Nyquist limit by having multiple Doppler
pulses in the body at the same time. In HPRF imaging multiple Doppler gates are displayed, since
the multiple Doppler pulses could be giving information from different depths. HPRF is
automatically invoked when needed to maintain the requested depth and scale. For example, if
the system is at a high scale and you move the cursor deeper the system may automatically
invoke HPRF. This can also happen if the cursor is deep and you increase scale.
If HPRF is active you will see multiple Doppler gates on the reference image. If you do not want
to be in HPRF, decrease the scale or move the cursor to a more shallow location until only one
gate is displayed.
5.2.3. Update
The Update hard key toggles between acquiring Doppler strip and acquiring the reference image.
5.3 CW-mode
Depending on system configuration, CW may be available on the P5-1XQ transducer. There are
two ways to invoke CW:
Press the CW button on the Doppler touchscreen to switch between PW and CW modes.
Configure the F1 or F2 programmable hardkey to invoke CW directly (see section 9.1 for
details).
5.3.1. CW Touch Screen Controls
The following touch screen controls impact the CW image. Touch screen controls can be located
where you want them to be. See section 3.8 for details.
- 39 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Adjust the Doppler scale to account for the angle between the
Doppler cursor and the blood flow. Upward presses increment
Angle Correct
the angle. Downward presses decrement the angle. Angles
above 80 are not allowed.
Gray Map
Adjusts the current postprocessing, either grayscale or tinted.
Tint
- 40 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Volume adjusts the audio volume of the Doppler strip. This can
Volume be adjusted in pre-Doppler to set the initial volume upon
invoking Doppler acquisition.
5.4 Color-mode
Color A set of radio-buttons display the color modes that are available on the
Mode current transducer, and let you switch between them. See section 5.4.2 for
Variation details.
Scale adjusts the range of velocities that are displayed. Upward presses
Scale increase scale and downward presses decrease it. It is available in
Velocity, PDI, and DPDI. It is not available in Freeze/Cine.
The Baseline control adjusts the Color baseline. Upward presses move
Baseline the baseline up on the scale and downward presses move the baseline
down.Baseline is not available in PDI mode.
- 41 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Map Adjusts the current map for the active color variation.
The smoothness filter determines the spatial filtering that is applied to the
Smooth
Color image. Higher filter levels create a smoother image. Upward
Filter
presses increase the filter. Downward presses decrease the filter.
When the system receives both B-mode and CD signal from a region the
threshold determines whether to display that as grayscale or color. Higher
Threshold
values mean a stronger CD signal is needed to display it in color. Upward
presses increase threshold. Downward presses decrease threshold.
This control is only available for linear transducers. It steers the Color ROI
Steer
box angle left or right.
Image Color Doppler supports image presets for Low Flow, Medium Flow, and
Type High Flow.
To activate split screen with simultaneous live B/Color and live B. the live
B image without color and the same live B image with color are
simultaneously displayed on each side of the image field.
Dual Live
Freezing the image will freeze both sides simultaneously.
Cine review will review both sides simultaneously.
- 42 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
The Dynamic Range, or log compression, adjusts how echo intensities are
Dynamic converted to brightness. A high dynamic range presents a flat image, while
Range a low dynamic range present a large range of intensities. Only available in
PDI/DPDI mode.
Adjusts the tradeoff between frame rate and spatial resolution. Higher
Frame rate
settings provide more temporal resolution.
Acoustic Adjusts the acoustic output power of the activated transducer. It is only
power* available in live imaging. Higher acoustic power numbers correspond to
more brightness and contrast in the image and more force of penetration,
but ALARA principle should be followed in actual situations.
5.5 M-mode
The following touch screen controls impact the M-mode image. Touch screen controls can be
located where you want them to be. See section 3.8 for details.
Name Control Description
Gray Map
Adjusts the current postprocessing, either grayscale or tinted.
Tint
- 43 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
- 44 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
and 4D.
NOTE: Measurement function is not available in 3D/4D.
5.6.1. Pre-3D
The Pre-3D mode supports defining the 3D/4D acquisition location and settings.
Figure 5-2 shows an example of the pre-3D image. The ROI (Region of Interest) indicates the
portion of the image that will be used for 3D/4D. The position/size of the ROI can be adjusted
using the trackpad and Set key. Also note that there is a horizontal line going across the ROI near
the top and it can be adjusted to a curved line upwards and downwards by using the trackpad.
This is the clip plane. The initial volume rendering will show everything below this line.
- 45 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
automatically.
3D/4D These two radio buttons determine if the acquisition will be
3D or 4D.
Start Start an acquisition. If 3D was highlighted then a single
volume is acquired. If 4D was highlighted then continuous
acquisitions are started.
Exit Exit 3D/4D
During the process of 3D volume sweeping, the images swept in ROI box are displayed in the
image area.
While scanning, the sector in the right bottom is filled gradually,and no buttons on the
control panel and touch screen can be activated.
- 46 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
The icon at the bottom-right shows the position of each slice with respect to the full 3D
data set.
The following table shows the controls that are available in Volume Imaging Mode
- 47 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
y-axis.
3D
Smooth/ Adjust smoothness of A\B\C slices and 3D images. Higher
filter levels create a smoother image. Upward presses
MPR
increase the filter. Downward presses decrease the filter.
Smooth
- 48 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
Rotation
Rotate the image by the X-axis of the activated window.
X
Rotation
Rotate the image by the Y-axis of the activated window.
Y
- 49 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
3*3, 2*2,
These three radio buttons determine the layout of the display.
3*2
Pan the basis slice along the x-axis or y-axis of the activated
Pan window. Use the trackpad to pan the basis slice along x-axis
y-axis. Parallel slices change with the pan operation.
Reset the operation of pan, rotate and zoom to the initial
Reset
condition.
- 50 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
When there are more slices than can be displayed at once this
Next
control cycles through the subset of slices displayed.
Rotation Rotate the basis slice by the X-axis of the activated window.
X Parallel slices change with the rotation operation.
Rotation Rotate the basis slice by the Y-axis of the activated window.
Y Parallel slices change with the rotation operation.
Rotation Rotate the basis slice by the Z-axis of the activated window.
Z Parallel slices change with the rotation operation.
- 51 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Imaging
4D Live volume continuously sweeps and displays successive 3D volumes. During live 4D
imaging the following controls are available: activate any window of A/B/C/3D,
Single/Dual/Quad,VOI, Threshold, Bright, Contrast, 3D Smooth/ MPR Smooth, 3D Tint/ MPR
Tint, Rotation, Rotation X, Rotation Y, Rotation Z , Zoom and Gain in B mode.
As shown in figure above, image in the left is A slice and image in the right is the live 4D volume
rendering.
5.6.6. 4D Cine
During the process of 4D live volume imaging; press the Freeze button on the control panel to
enter the 4D cine mode. A cine replay scroll bar appears below the images. Use the trackpad to
replay the reset 3D images. For the selected frame, user can execute all operations that are
available 3D review status. Press Freeze again to return to the live image.
Touch screen controls are the same as that of static 3D except for Cine button.
6 Transducers
6.1 Transducer Model
- 53 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Transducer
Transducer head
Transducer cable
Transducer connector
Name Function
Converts electrical signals to sound waves, and then converts received echoes
Transducer head
back to electrical signals. The tip of the transducer is the acoustic lens.
Transmits electrical signals between the transducer head and the transducer
Transducer cable
connector.
- 54 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Transducer
Isopropanol (70%)
OPA (0.55%)
If Ethanol or Isopropanol is used for both cleaning and disinfecting, then a new cloth is required
for the disinfection step.
WARNING
1. Unplug the transducer from the system prior to cleaning or disinfecting.
2. To avoid infection, always use protective gloves when performing cleaning and
disinfecting procedures.
3. To avoid infection, ensure that expiration date of the disinfecting solution has not
passed.
CAUTION
1. Always wear gloves to perform the following steps.
2. To minimize the disease transmission, use of legally marketed, sterile transducer and
sheath is strongly recommended for intra-cavitary and intra-operative procedures.
Legally marketed, sterile, pyrogen free transducer sheath is required to use for
neurological intra-operative procedures.
3. DO NOT use an expired transducer sheath. Check whether the term of validity has
expired prior to using transducer sheaths.
4. The single-use sheath should comply with the local regulations.
To install transducer sheath:
1. Place an adequate amount of sterile coupling gel on the acoustic window of the transducer;
2. Insert the transducer into the sheath;
3. Pull the sheath over the transducer and cable until the sheath is fully extended;
4. Secure the sheath using the bands or clips supplied with the sheath;
5. Check and eliminate bubbles between the face of the transducer and the sheath. Bubbles
between the face of transducer and the sheath may affect the ultrasound imaging.
6. Inspect the sheath to ensure that there are no damages (i.e. holes or tears).
- 55 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Transducer
There are two methods to disinfect the transducer (use the immersion method for the E8-4Q and
L17-7SQ):
Non-immersion:
1. Disconnect the transducer from the system.
2. Prepare the disinfectant solution (75% ethanol or 70% isopropanol).
3. Spray the solution to the transducer interface or wipe it with a soft cloth dampened with the
disinfectant solution.
4. Rinse the transducer according to the disinfectant instructions. Wipe the transducer with a dry
cloth or leave the transducer to air dry.
5. Inspect the transducer to ensure that there is no damage.
Immersion:
1. Prepare the disinfectant solution (Cidex OPA)
2. Place the cleaned and dried transducer in contact with the disinfectant (refers to figure 6-2 for
the contacting area) for the time specified by the disinfectant manufacturer.
3. Rinse the transducer according to the disinfectant instructions. Wipe the transducer with a dry
cloth or leave the transducer to air dry.
4. Inspect the transducer to ensure that there is no damage.
.
C5-2XQ/ L10-4Q/ L17-7HQ L17-7SQ P5-1XQ E8-4Q MC8-4Q
C5-2Q L12-5Q
C5-2MQ*
Disinfectant
WARNING
1. Do not immerse the transducer connector. If the cable connector is immersed, do not
plug the connector into the system. Rinse the connector under running water and dry
it thoroughly. If necessary, contact EDAN for service.
- 56 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Transducer
2. Prohibit infiltration of any type of liquid into the device or the transducer.
3. Do not sterilize the transducer using techniques such as autoclave, ultraviolet, gamma
radiation, gas, steam, or heat. Severe damage May result.
4. Use only ultrasound coupling gel that complies with local regulations.
5. Do not immerse the AC adapter and connector of the transducer into solutions.
Transducers can be submerged to, but not including, the strain relief of the transducer
array. Do not immerse or soak any part of a transducer in any cleaning material not
listed in the recommended list of disinfectant.
6. Use the Cidex OPA immersion method to disinfect the endocavity transducer.
7. Only non-immersion method can be used with solution of ethanol or isopropanol. The
solution concentration should not exceed the value provided above.
8. The immersion time should not exceed the time that is specified by the disinfectant
manufacturer.
9. Patient contact area should be immersed into the solution while using the immersion
method, but should not exceed the depth shown in figure 6-2.
CAUTION
1. Do not disinfect or clean transducers under high temperature, and the temperature
should be below 45°C.
2. In order to avoid damaging the device, the disinfection method is limited to regular
maintenance of devices in hospitals. Disinfecting instruments should be cleaned first.
- 57 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
7 Features
7.1 Comments
The Comments function lets you add annotation to an image. The Comment function is invoked
by pressing the Comment hard key on the console. Figure 7-1 shows an example touch screen
for Comments. The top portion of the screen shows comments that are pre-defined in pre-sets.
See section 8 for details on how to configure these. The bottom portion of the screen shows
controls available while the Comment function is active. Table 7-1 gives an overview of these
controls.
- 58 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
- 59 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
7.3 Measurements
- 60 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
The Measure function lets you perform measurements on a live or frozen image. The Measure
function is invoked by pressing the Measure hard key on the console. Figure 7-3 shows an
example touch screen for Measure.
There are two types of measurements:
Generic Measurements: These are simple tools like Distance or Area. There is no specific
anatomy associated with these measurements and they do not appear in a report.
Application Measurements: These are measurements for specific anatomy or clinical conditions.
The results can be entered into a report that can be printed later.
On the Measure touch screen the lower part of the touch screen shows Generic Measurements.
These are different for each imaging mode, but are the same for all presets. The upper part of the
touch screen shows Application Measurements. These are different for each imaging mode and
each preset. See section 8 for details on how to configure these.
Tool Options
Some measurements can be done with a choice of different tools, and each tool has a choice of
multiple output result items. The users can configure the default tool and result items for these
measurements by Tool Options. For result options, at least one item should be selected.
- 61 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
- 62 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
The Acclarix AX8 supports the following types of generic B-mode measurements:
Distance
Circ/Area
Angle
Volume
Stenosis
Distance: The distance measurement is always done with a caliper pair.
To measure Distance with a caliper pair:
1. Invoke the measure function on a B-mode image.
2. If Distance is not the default measurement, select Distance from the touch screen.
3. Move the caliper to the desired location.
4. Press the Set key to place that caliper and add the second half of the caliper pair.
- 63 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Further presses of the Set key toggle which side of the caliper can be moved. Pressing the
Enter key completes the caliper and allows you to create another tool.
Circ/Area: Area and Circumference can be measured with an Ellipse or Trace
To measure Area or Circumference with an Ellipse:
1. Invoke the measure function on a B-mode image.
2. If Area is not the default measurement, select Area from the touch screen.
If Ellipse is not the default tool, press Tool Options to switch to Ellipse
3. Move the caliper to the desired location.
4. Press the Set key to place that caliper and add the second half of the caliper pair.
5. Press the Set key to place the second half of the caliper. An ellipse appears and can be
adjusted with the trackpad.
Further presses of the Set key toggle between controlling each side of the ellipse caliper
and the ellipse diameter. Pressing the Enter key completes the caliper and allows you to
create another tool.
To measure Area or Circumference with a Trace:
1. Invoke the measure function on a B-mode image.
2. If Area is not the default measurement, select Area from the touch screen.
If Trace is not the default tool, press Tool Options to switch to Trace. Trace can be done
with a Draw option. Draw will follow your movements.
3. Move the caliper to the desired location.
4. Press the Set key to place the caliper and start drawing the trace.
5. Moving the caliper now leaves behind a trail of dots to mark the trace outline
6. Press the Enter key to complete the trace.
Angle: The Angle measurement is always done with an angle tool.
To measure Angle:
1. Invoke the measure function on a B-mode image.
2. If Angle is not the default measurement, select Angle from the touch screen.
3. Move the caliper to the vertex of the angle you are measuring.
4. Press the Set key to place that point and create a new caliper. Move that to one end of the
angle.
5. Press the Set key to place that end and create a new caliper. Move that to the other end of
the angle. Further presses of the Set key toggle between controlling each of the three
points that make up the angle. Pressing the Enter key completes the angle and allows you
to create another tool.
Volume: Volume can be measured with 3 Distance.
To measure Volume with 3 Distance:
1. Invoke the measure function on a B-mode image.
2. If Volume is not the default measurement, select Volume from the touch screen.
Move the caliper to the desired location.
- 64 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
3. Press the Set key to place that caliper and add the second half of the first caliper pair.
4. Further presses of the Set key toggle which side of the caliper can be moved. Pressing the
Enter key completes the first caliper pair and creates the second caliper pair.
5. Repeat step 3~5 to complete the second caliper pair and create the third caliper pair.
6. Repeat step 3~5 to complete the third caliper pair and allow you to create another tool.
Stenosis: Stenosis can be measured with Caliper Pairs, Ellipses, or Traces
To measure Stenosis with Caliper Pairs:
1. Invoke the Measure Function on a B-mode image.
2. If Stenosis is not the default measurement, select Stenosis from the touch screen.
If Caliper is not the default tool, press Tool Options to switch to Caliper.
3. Move the caliper to the desired location.
4. Press the Set key to place that caliper and add the second half of the caliper pair.
5. Further presses of the Set key toggle which side of the caliper can be moved. Pressing the
Enter key completes the caliper and creates the second caliper pair tool.
6. Repeat step 3~5 to complete the second caliper pair and allow you to create another tool.
To measure Stenosis with Ellipses:
1. Invoke the Measure Function on a B-mode image.
2. If Stenosis is not the default measurement, select Stenosis from the touch screen.
If Ellipse is not the default tool, press Tool Options… to switch to Ellipse.
3. Move the caliper to the desired location.
4. Press the Set key to place that caliper and add the second half of the caliper pair.
5. Press the Set key to place the second half of the caliper. An ellipse appears and can be
adjusted with the trackpad.
6. Further presses of the Set key toggle between controlling each side of the ellipse caliper
and the ellipse diameter. Pressing the Enter key completes the first Ellipse and creates the
second Ellipse tool.
7. Repeat step 3~6 to complete the second Ellipse and allow you to create another tool.
To measure Stenosis with Traces:
1. Invoke the Measure Function on a B-mode image.
2. If Stenosis is not the default measurement, select Stenosis from the touch screen.
If Traces is not the default tool, press Tool Options to switch to Traces.
3. Move the caliper to the desired location.
4. Press the Set key to place the caliper and start drawing the trace.
5. Moving the caliper now leaves behind a trail of dots to mark the trace outline
6. Press the Enter key to complete the trace.
7. Repeat step 3~6 to complete the second trace and allow you to create another tool.
- 65 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
The Acclarix AX8 supports the following types of generic strip Doppler measurements:
Caliper
Trace
Auto Trace
HR (Heart Rate)
RI (Resistive Index)
Caliper: The Caliper measurement can provide a wide range of results.
To use the Caliper measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If Caliper is not the default measurement, select Caliper from the touch screen.
3. Move the caliper to the desired location. At this point the measurement can display the
following results (the actual results shown will depend on the preset):
Vel
PG
4. If you want a caliper pair, press the Set key to place the caliper and add the second half of
the caliper pair. At this point the measurement may display the following results (the
actual results shown will depend on the preset):
V1
V2
RI
S/D
Time
ΔV
Accel
PG1
PG2
PHT
Further presses of the Set key toggle which side of the caliper can be moved.
Pressing the Enter key completes the caliper and allows you to create another tool.
The S/D, PI, and PHT results assume that the first caliper is placed on the peak
systolic point.
Trace: The Trace measurement can provide a wide range of results.
To use the Trace measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If Trace is not the default measurement, select Trace from the touch screen.
3. Trace can be done with a Draw. Draw will follow your movements
4. Move the caliper to the desired start point.
- 66 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
5. Press the Set key to place the start point and start drawing the trace.
6. Moving the caliper now leaves behind a trail of dots to mark the trace outline. Press the
Set key to adjust the positions of the PS or ED points in the trace.
7. Press the Enter key to complete the trace.
8. At this point the trace measurement may display the following results (the actual results
shown will depend on the preset):
PS
ED
MD
TAMax
PGmax
PGmean
PI
RI
S/D
VTI
Time
AT
DT
NOTE: Measure trace can only be active in image frozen status.
Auto Trace: Auto Trace is capable of determining the maximum and/or mean blood flow in
freeze mode. Each result that it calculates is associated with either the maximum or the mean
value. For activate auto trace in real time mode, press Auto Trace button in PW touch screen.
To use the Auto Trace measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If Auto Trace is not the default measurement, select Auto Trace from the touch screen.
The trace waveform(s) automatically appear. If any maximum-related results are enabled then
the maximum value waveform is displayed on the Doppler strip. If any mean-related results
are enabled then the mean value waveform is displayed.
3. The Trace measurement may display the following results (the actual results shown will
depend on the preset):
PS
ED
MD
TAMax
TAMean
PGmax
PGmean
PI
RI
S/D
VTI
Time
AT
DT
- 67 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
HR
NOTE: Measure Auto trace can only be active in image frozen status. Live auto trace is
available as a separate feature from the Doppler touch screen.
HR:The HR measurement can provide the heart rate calculation by measured the time of the
heart beat cycles.
To use the HR measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If HR is not the default measurement, select HR from the touch screen.
3. Move the caliper to position the first heart beat. Press the Set key to activate a second
caliper.
4. Move the second caliper to position the next beat.
5. Press the Enter key to complete or press the Set key to switch moving the first caliper.
6. The HR measurement can display the following result(the actual results shown will
depend on the preset):
HR
RI: The RI measurement can measure the PS and ED to calculate the result of RI and S/D.
To use the RI measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If RI is not the default measurement, select RI from the touch screen.
3. Move the caliper to position the PS point. Press the Set key to activate a second caliper.
4. Move the second caliper to position the ED point.
5. Press the Enter key to complete or press the Set key to switch moving the first caliper.
6. The RI measurement can display the following results (the actual results shown will
depend on the preset):
PS
ED
RI
S/D
7.3.1.3. M-mode Generic Measurements
The Acclarix AX8 supports the following types of generic M-mode measurements:
Caliper
HR (Heart Rate)
Caliper: The Caliper measurement can provide a wide range of results.
To use the Caliper measurement:
1. Invoke the Measure Function on an M-mode strip. A caliper measurement is automatically
invoked. It displays the Depth at the current location.
2. Move the Caliper to the desired location.
- 68 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
3. Press the Set key to place that caliper and add the second half of the caliper pair. Further
presses of the Set key toggle which side of the caliper can be moved. Pressing the Enter
key completes the caliper and allows you to create another tool
4. The Caliper measurement can display the following results:
Distance
Time
Slope
HR:The HR measurement can provide the heart rate calculation by measured the time of the
heart beat cycles.
To use the HR measurement:
1. Invoke the Measure Function on a Doppler strip.
2. If HR is not the default measurement, select HR from the touch screen.
3. Move the caliper to position the first heart beat. Press the Set key to activate a second
caliper.
4. Move the second caliper to position the next beat.
5. Press the Enter key to complete or press the Set key to switch moving the first caliper.
6. The HR measurement can display the following result(the actual results shown will
depend on the preset):
7.3.2. Application Measurements and Calculations
Application Measurements have a pre-defined meaning and can be entered into a report. Acclarix
AX8 supports the following application packages, each with its own set of measurements,
calculations and report:
OB (including multiple fetuses)
Abd
Vascular
Cardiac
Gynecology
Small Parts
Urology
Pediatric(including Neonatal)
Please see the Advanced User Manual for the complete content of each package.
When you select an application measurement from the touch screen it will automatically invoke
the type of measurement it needs. For example, if you select “BPD” from the OB application it
will automatically invoke a distance measurement. These measurements generally behave as
described above for generic measurements. However, a final press of the Enter key after the
measurement is complete will enter the result into the report.
Some application measurements can have multiple variations. For example, in a twin OB exam
the OB measurements can be done on either fetus. As another example, in Vascular exams several
measurements can be done in a Proximal, Mid, or Distal location. When a measurement has
- 69 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
multiple variations you will see buttons on the touch screen that let you pick which variation you
are measuring.
- 70 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
- 71 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
7.3.2.3. PISA
PISA (Proximal Isovelocity Surface Area) is a tool in Cardiac presets used for quantitative
analysis of regurgitation. It is only available in Color and Doppler mode.
To use PISA measurement in Color mode:
1. Adjust the color image until the aliasing appears. Invoke the Measure function.
2. Select PISA measurement on the touch screen and then select a measurement label from MR
Rad, MR Als. Vel, AR Rad, AR Als. Vel, TR Rad, TR Als.Vel, PR Rad and PR Vls. Vel. A
single caliper will be displayed on the image field.
3. Move the caliper and press Set key to position the caliper as the center of the semicircular.
4. Move the trackpad to adjust the radius length and orientation of the semicircular.
5. Press Enter key to fix the semicircular and then a dialog box of “Color Aliasing Velocity”
pops up.
6. Select “Top aliasing velocity”, “Bottom aliasing velocity”, or manually enter the aliasing
velocity from 0 cm/s to 500 cm/s.
7. Click OK and then measurement results are shown.
To use PISA measurement in Doppler mode:
1. Freeze a strip image. Invoke the Measure function.
2. Select PISA measurement on the touch screen and then select a measurement label from MR
Trace, AR Trace, TR Trace and PR Trace.
3. Use the Doppler trace tool to get the measurement results.
When all the measurements described above have been completed, the system automatically
calculates the Flow Rate, Flow and EROA(Effective Regurgitant Orifice Area). See Chapter 6
Cardiac in advanced manual for the detailed PISA measurement and calculation results and
related equations.
- 72 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Storing an image:
1. If a Store key is configured to store a static image it will always capture what is on the
imaging area of the screen. This includes live, frozen, or Cine images. It also includes
reports or other GUI screens and review.
Storing a clip:
1. If a Store key is configured to store a clip image it will capture the moving image in live or
auto cine play. If the image is frozen or a non-image screen is displayed then a static image
is stored.
2. The store starts with the press of the button and continues for the configured length of the
clip or until the clip store is interrupted. The following events can cause the clip store to
stop:
A second press of the store button
Display of a GUI screen or dialog
Mode change
Reaching a pre-defined clip duration (refer to System Setup)
Image parameters change (depth and zoom in B/C mode, M cursor, sweep speed, strip
size and depth in M mode, angle, scale, baseline, invert, strip size and sweep speed in
Strip Doppler mode.)
Cine play pause (press Freeze button, Set button or dragging left along the Swipe button)
7.4.2. Reviewing Images
If Static or Clip images have been stored for the current exam then they can be reviewed by
pressing the Review hard key on the console.
NOTE: If nothing has been stored in the current exam then the Review hard key will
invoke the Patient Database function (see the next section).
Figure 7-9 shows an example touch screen for Review. The following table explains the action of
each button in Review.
- 73 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
- 74 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Select the export path from destination drop list menu and click
OK to confirm. If there is only one removable media then this field
is a read-only field. Static images export as .bmp files. Clip
images export as .avi files.
While export in progress the dialog is displayed as bellow. Cancel
button is used to stop the export. The dialog disappears when
export finishes.
- 75 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
- 76 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
- 77 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
screen, select the exam(s) from the list shown, and select the device from the list of currently
configured destinations.
NOTE: Use proper sterile technique at all times when performing a biopsy.
Always follow these basic precautions:
WARNING
1. Disinfect the needle guide kit before the first use and after each subsequent use.
2. Calibrate the needle guide kit (see section 7.5.3) under any of the following
conditions:
a) The first time that each bracket/transducer combination is used.
b) If the bracket or transducer head is dropped or struck, or has evidence of wear.
c) If previous use has shown some drift of the needle from the center of the
guidelines.
3. The displayed needle guide pathway on the EDAN video monitor is intended for
reference during biopsy procedures. A variety of factors outside EDAN‟s control, such
as changing tissue density, bending of the needle, off-axis pressure by the person
holding the transducer, etc., may cause deflection of a needle outside of the
displayed video pathway even when the transducer, needle guide, and the system
software are all performing as intended and within manufacturing specification. The
specialist performing a biopsy procedure must be aware of potential external factors
when performing an invasive procedure.
4. Do not freeze the system when performing a biopsy.
5. EDAN needle guides are designed and manufactured to attach firmly to designated
transducers and should not require excessive force to assemble or disassemble. Do
not use a needle guide that requires excessive force or manipulation to assemble or
disassemble.
6. A single-use sheath should be used on transducer when performing a biopsy.
WARNING
The needle guide bracket kits are not disinfected or sterilized before delivery. Disinfect
the needle guide kit before the first use and after each subsequent use For detailed
operations, see 12.4.2 Needle Guide Bracket Cleaning and Sterilization.
- 78 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Screw
螺钉
旋钮1
Knob 1
矩形卡槽
Rectangular slot
圆锥台
Cone
旋钮2
Knob 2
Convex
一字形凸起
Notch
凹 槽
- 79 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Figure 7-15 Attaching Bracket Kit to Probe Figure 7-16 Turning Knob 2 Clockwise
Step 4: Insert the needle into the rectangular hole, as shown in Figure 7-17.
Rectangular Hole
矩形孔
WARNING
The needle guide bracket kits are not disinfected or sterilized before delivery. Disinfect
the needle guide kit before the first use and after each subsequent use. For detailed
operations, see 12.4.22 Needle Guide Bracket Cleaning and Sterilization.
- 80 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Knob 1 Lock
Fixing
claws
Notches
- 81 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Round hole
WARNING
1. Calibrate the needle guide under any of the following conditions:
a) The first time a needle guide is used with a given transducer.
b) Any time the needle guide or transducer has been dropped or struck against a
hard surface.
c) After repeated use.
2. Do not use the needle guide bracket if the needle does not track with the guide during
calibration.
- 83 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Features
Angle Line
- 84 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
8 Presets
8.1 Preset Organization
The Acclarix AX8 Presets are divided into two levels: the Exam Presets and Application Presets.
Exam Preset: Each transducer has its own set of Exam Presets. Each Exam Preset contains:
The image parameters that optimize its transducer for an intended use.
The list of data fields displayed on the patient demographic page.
Pointers to Application Presets for Comments, Measurements, and BodyMarkers.
Application Preset: The Application Presets are independent from the transducer. Application
Presets are used to select Comments, BodyMarkers and Measurements. Since they are
independent from the transducer, several Exam Presets can use the same Application Preset.
Figure 8-1 shows an example of how Transducers, Exam Presets, and Application Presets are
related. For simplicity, this shows just a few of the items that are on the real system. In this
example the C5-2XQ and the E8-4Q both have an OB Exam Preset. The C5-2XQ OB Exam
Preset optimizes the C5-2XQ imaging for OB while the E8-4Q OB Exam Preset optimizes the
E8-4Q for the same use. The parameters used for each transducer can be different, and changing
one Exam preset does not change the other.
However, both OB Exam Presets point to the same OB Measurement Preset as the Application
Preset. This means you can, for example, just configure OB measurements once, and both the
C5-2XQ and E8-4Q will have updated OB measurements.
This example just shows a single kind of Application Preset. The Acclarix AX8 actually has
separate Application presets for Comments ,Bodymarkers and Measurements. The same principle
applies to each of these: application presets can be shared across exam presets and transducers.
- 85 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 86 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Press System to edit overall system settings unrelated to presets. See section 9 for details.
Press Presets to do any of the following:
o Edit the imaging parameters of an exam preset.
o Configure the demographic fields that appear for that preset
o Configure which Comment, Measure, or Bodymarker presets are associated with
each exam preset.
Press Comments to configure the pre-defined text of each comment preset.
Press Bodymarker to configure which bodymarker graphics appear with each
bodymarker preset.
Press Measure to configure measurement options of each measure preset.
Figure 8-3 Top level Touch Screen for the Setup Function
8.3.1. Presets
Figure 8-4 shows an example of the presets page.
- 87 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Figure 8-5 shows an example imaging set-up screen for imaging parameters. This example shows
the screen for editing the B-mode settings for the C5-2XQ Abdomen exam preset.
Figure 8-6 shows an example of the Patient information configuration page. This page allows you
to configure which demographic fields are shown for each exam preset.
- 88 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Measure: Select one of the pre-defined application measurement packages to be the default
measure for the preset.
8.3.1.4. Selecting Default Comments Preset
Comments: Select one of the pre-defined comments presets to be the default comments for the
preset.
8.3.1.5. Selecting Default Bodymark Preset
Bodymark: Select one of the pre-defined bodymark presets to be the default bodymark for the
preset.
8.3.1.6. Selecting Default TI
- 90 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Suppress Comment on Unfreeze: If this option is enabled the Comment function goes
away when the system is unfrozen, and comes back again with the next freeze.
Manually exiting the function causes it to remain inactive until manually activated.
- 91 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Pre-defined
graphics area
- 92 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Pre-Defined
Measurement
Area
- 93 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 94 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 95 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Logo Position: set the logo position to locating at the left or right side of report header.
Logo Size: set the logo size to Small or Large.
OB Settings
The system supports the following OB measurement settings:
Fetal Age: select the author / measurement options for Fetal age calculation.
Fetal Growth: select the author / measurement options for Fetal Growth calculation.
Estimated Fetus Weight: select an equation for the Estimated Fetus Weight calculation.
Growth Curves Display: set single or quad curves as the default display format for the
Growth Curves.
Default GA display: set the CUA (Composite Ultrasound Age) or AUA (Average
Ultrasound Age) as a default result displayed in the worksheet.
Cardiac Settings
The system supports following Cardiac measurement settings:
HR Cycle: set the number of heartbeats that are assumed in the HR or FHR
measurement calculation. Range: 1-8.
EDV/ESV: set the author of the formula that is used in EDV/ESV calculation. There are
three options: Teichholz, Gibson and Cube.
Suppress Measurement on Unfreeze If this option is enabled the Measure function goes
away when the system is unfrozen, and comes back again with the next freeze. Manually
exiting the function causes it to remain inactive until manually activated.
- 96 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
9 Utilities
The Utilities hard key provides access to setup, screen adjust, connectivity and maintenance
software. Each of these is described in separate sections below.
9.1 Set-up
The Set-up function is divided into several sections, accessed by separate touch screen buttons
System Set-up: Described in the remainder of this section.
Presets: Supports edits of preset organization and image preset. See section 8.3.1
Comments: Supports edits of comment presets. See section 8.3.2
Body Mark: Supports edits of Body Mark presets. See section 8.3.3
Measure: Support edits of Measure presets. See section 8.3.4
System setup is used to configure parameters that are unrelated to presets. Generally, there is one
value of each system setup parameter that is shared across all presets.
Entering/exiting Setup:
Press Utilities on the control panel, the Setup touch screen button, and then the System touch
screen button to access system setup.
Reloading Factory Default Set-up:
1. Press “Restore Factory Settings” on the Set-up screen, you are prompted Restore System
Setting to factory default? This will erase all current system setting.
2. Press Yes to restore to factory default, and press Cancel to cancel.
9.1.1. General Set-up
- 97 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 98 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 99 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 100 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 101 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 102 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
9.2 Connectivity
The Connectivity screens support configuration of network access and services. It is divided
into 3 tabs:
TCP/IP: Configures access to the network from the AX8.
DICOM: Configures network DICOM services.
Network Store: Configures non-DICOM storage to a network via FTP.
9.2.1. TCP/IP
This screen configures access to a network. Access can be either wired or wireless*.
- 103 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
The following controls require that a wireless network has been selected from the network
list. Click on the name of a network to select it.
Join: This joins the selected network. If a password is needed a dialog box will be
displayed to enter it.
Forget Network: This forgets any password of the selected network that had been entered,
but the network is still shown in the list.
Advanced Settings: This brings up a dialog box with additional settings for the selected
wireless network. The dialog box includes the following:
Password: This is useful when the network password has changed.
Method: Either Manually entered (Static) or DHCP. Generally wireless networks
use DHCP, and a Static address is only needed if required by the target PACS system.
IP Address, Subnet Mask, Gateway, DNS: If you chose DHCP then your network
router will set these automatically. If you chose Manually Entered IP then contact
your network IT manager for the appropriate settings.
9.2.2. DICOM
This screen configures DICOM services, including Store and Modality Worklist.
Local Settings:
These settings determine how the AX8 communicates with other DICOM devices. These are
required for both Store and for Modality Worklist.
- 104 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Default: A check mark on the left side indicates which server is the default for
Auto-Transfer.
Name: The name of the server that appears in the drop-down list of the patient
database
AE Title, IP Address, Port, Max PDU: These are the settings of the destination
DICOM server; it‟s how the AX8 finds the DICOM server on your network. The
AE title and IP Address are unique to your network; contact your network IT
manager for these settings. The most common setting of Remote Port for DICOM
servers is 104, although your server may be different.
Testing the server: There are two tests to ensure that the server information is entered
correctly. Click on any field for a given server to make that server selected, then:
Ping: A successful Ping means that the AX8 can communicate with the server at a
low-level; basically that the two computers „see‟ each other. As a security measure,
some servers on the Internet may be configured to not respond to a Ping even if the
connection is successful.
Verify: A successful Verify means that the AX8 can communicate with the server at a
DICOM level; basically that the DICOM on both computers understand each other.
A successful Verify will typically mean that your DICOM configuration is correct.
Other controls:
Add: Adds another line in the list of servers.
Delete: Deletes the selected server.
Default: Makes the selected server the default for Auto-transfer.
Modality Worklist: Modality Worklist will query the configured server for a list of all ultrasound
exams scheduled within the configured date range.
Enable: Turns Modality Worklist on/off.
Update Worklist: Determines when the worklist is updated. Choices are:
On Demand: Updates only when the Update button is pressed from the patient
page.
Every 15 Minutes: Updates on power-up and then regularly every 15 minutes.
Every hour: Updates on power-up and then regularly every hour.
On Start of Exam: Updates when the patient page is displayed at the start of each
exam.
Comment Source: Modality Worklist can fill the ultrasound comments section from the
server. Depending on
Patient Comments: Obtains comments from DICOM tag (0x0010,0x4000);
Scheduled Procedure Description: Obtains comments from DICOM tag
(0x0040,0x0007)
Comments on Schedule Procedure: Obtains comments from DICOM tag
(0x0040,0x0400)
Worklist Server: Selects the server to query for the worklist. Options include any server
configured in Server Settings (see previous section).
Date Range: The range of dates for the worklist query. Note, the option of “Yesterday and
Tomorrow” includes today‟s exams.
- 106 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Testing the server: There are two tests to ensure that the server information is entered
correctly. Click on any field for a given server to make that server selected, then:
Ping: A successful Ping means that the AX8 can communicate with the FTP server at
a low-level; basically that the two computers „see‟ each other. As a security
measure, some servers on the Internet may be configured to not respond to a Ping
even if the connection is successful.
Connect: A successful Connect means that the AX8 can communicate with the FTP
server, and FTP transfers will work. If this fails but Ping works it could mean:
oThe destination computer is not correctly configured as an FTP server.
oThe password on the server has changed. Use Edit to re-enter the password.
oThe Shared Directory does not exist. Use Edit to confirm the directory.
Add: Adds another line in the list of servers. This automatically invokes Edit.
- 107 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
Edit: Available when an FTP Server has been selected. This invokes a dialog that edits the
IP address, name, directory, and password.
Delete: Available when an FTP Server has been selected. This deletes the selected server.
9.3 Maintenance
The Maintenance screens provide access to controls that are not typically needed during normal
operation of the system.
9.3.1. License
The License page displays which features are currently licensed for use on the system. At the
top of the screen is displayed the current license key. Below that is a list of all licensable
features, along with its current status. The system supports temporary trial licenses. If a
feature is so licensed the expiration date is also shown.
Press Import button to import a license file from a USB external storage device.
Press Revert button to restore the previous license if the imported license is not correct.
Please contact your distributor or Edan service engineer to obtain a new license key.
9.3.2. Version
The Version screen shows the current version of software, firmware, and selected hardware for
the system. This information is only needed if requested by an Edan service engineer. This
page also includes an Upgrade, Import Config and Show Config button. Any released
- 108 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
9.3.4. Export/Import
The system supports export/import of user-configured presets and settings to an external storage
device. This same UI supports the export of system log files.
To export user data:
- 109 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 110 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
3. Check the items to be imported: Image Presets, Comment Presets, Body Mark Presets,
User Settings, Findings. The Preset items can be expanded to select a specific preset.
4. Press Import button to begin the import process.
Note: Import of System Logs is not supported.
- 111 -
Acclarix AX8 Diagnostic Ultrasound System User Manual System Setup
- 112 -
Acclarix AX8 Diagnostic Ultrasound System User Manual In Between Exams
10 In Between Exams
10.1 Unpacking
Visually examine the package prior to unpacking. If any signs of mishandling or damage are
detected, contact the carrier to claim for damage. After unpacking the device, you should follow
the packing list to check the product carefully and to make sure that no damage has occurred
during transportation. Then, install the device according to the installation requirements and
methods.
WARNING
1. Do not use the device if it is found to be damaged or defective.
2. Do not drop or collide with the transducer. Otherwise you shall give up using it.
3. Make sure the AC power supply complies with the following specifications:
100V-240V~, 50Hz/60Hz.
10.2 Transport
The system is designed to be portable and easily transported. Power off the system and secure all
accessories before moving it to another location.
CAUTION
1 Switch off the ultrasound system. Unplug the AC adapter from the power source and
secure the power cable.
2 Remove the transducer and place them in a safe place.
3 Disconnect and secure the connecting cable.
4 Connect optional system accessories.
5 Secure the system and complete the system setup, and then perform all the daily
checking before using it.
To prepare the system for shipment over long distances or rough terrain, repack the system in the
factory packing
To prepare the system for transport over distances: load the system into a vehicle using a lift gate.
To prevent lateral movement of the system, secure the system with cargo straps. To prevent
sudden jarring of the system during transport, provide anti-shock cushions beneath the system.
It is suitable for transportation by air, railway, highway and ship. Protect the system from
inversion, collision, and splashing with rain and snow.
10.3 Storage
Do not place the device near the ground, walls or the roof.
Keep good indoor ventilation. Avoid strong and direct sunlight, and erosive gas.
- 113 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Order List
11 Order List
The following accessories are recommended for use on the Acclarix AX8.
WARNING
Only accessories supplied or recommended by EDAN can be used, the battery and
transducers of EDAN can be only used on EDAN‟s systems. Otherwise, the performance
and electric shock protection cannot be guaranteed. If electrical or mechanical equipment
from other companies need to be connected to the device, please contact EDAN or
authorized representatives before connection.
Adapter 21.21.064243
- 114 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Order List
Suitcase 01.56.466102
NOTE: The part name may vary depending on context, but the part number is constant.
- 115 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Troubleshooting and Maintenance
12.2 Troubleshooting
If any persistent system malfunction is experienced, e.g. an onscreen error message, blank
imaging screen, absent menus, please refer to the following table below. If the failure cannot be
eliminated, please contact EDAN or authorized representatives.
- 116 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Troubleshooting and Maintenance
12.3 Maintenance
The system should be maintained regularly, at least annually, by a qualified technician who has
adequate training, knowledge and experience. That person should be familiar with the AX8
Service Manual (P/N 01.54.002483), available from your Edan representative.
- 117 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Troubleshooting and Maintenance
WARNING
The console (including touch screen, track pad and hard keys) of system is splash resistant (IPX 1) but
not waterproof. Do not immerse or expose to extended moisture. Splash resistance does not
extend to transducer connectors. Keep connectors dry.
For cleaning agents and disinfectants, refer to section 6.3 Transducer Cleaning and Disinfection
12.4.1.Cleaning the System Surface
NOTE:
1. Use proper sterilization technique at all times when performing a biopsy.
2. Ensure that protective gloves are worn.
WARNING
The needle guide bracket kits are not disinfected or sterilized before delivery. The
operators must clean and sterilize the needle guide bracket kits before and after each
use.
Cleaning
Clean the needle guide assembly using a mild detergent prior to sterilizing.
Sterilization
Sterilize the Needle Guide Assembly by autoclaving for at least four minutes at 132oC
- 118 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Specifications
13 Specifications
13.1 Electrical Safety Classifications
According to the type of Internally powered equipment,
protection against electric
shock Class I equipment
- 119 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Specifications
Display TFT-LCD
Diagonal Size 15.6 inches
View Angle Right 80, Left 80, Up 80, Down 80 (min)
Pixel Number 1920x1080
White Luminance 300 cd/m2(typ)
Contrast Ratio 700 (typ)
B Mode: B, 2B,4B
C Mode: B/C(Single, Dual, Quad);
B+B/C(Dual Live);
B/C/PW (triplex mode)
PDI/DPDI Mode:
B/PDI(DPDI) (Single, Dual, Quad);
B+B/PDI(DPDI) (Dual Live);
B/PDI(DPDI)/PW (triplex mode)
Display Modes PW Mode:
B/PW;(Update)
B/PW; (duplex, simultaneous)
B/C/PW(Update)
B/C/PW, B/PDI(DPDI)/PW; (triplex mode)
CW Mode:
B/CW;
B/C/CW, B/PDI(DPDI)/CW;
M Mode: B/M (Display layout: Up/down, Left/right,1:1)
Image Gray Scale 256 levels
In area
Image Magnification Real time: x1.0, x1.3, x1.5, x1.8, x2.0. Frozen: x1.0, x1.3, x1.5, x1.8,
x2.0.
Hard Drive Storage 500 GB
Depth Adjustment Adjustable in real time in all modes
- 120 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Specifications
Language Display Chinese, English, German, French, Italian, Spanish, Russian, Turkish
Patient name, age, sex, time, date, hospital name, doctor name,
Annotations
comment (full-screen character editing)
Body Mark 100 types
USB Port Two USB 3.0 and two USB 2.0 ports.
- 121 -
Acclarix AX8 Diagnostic Ultrasound System User Manual EMC Information
14 EMC Information
Electromagnetic emissions
Electromagnetic emission
The Acclarix AX8 is intended for use in the electromagnetic environment specified below; The
customer or the user of the Acclarix AX8 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions Group 1 The Acclarix AX8 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
CISPR 11
not likely to cause any interference in nearby electronic
equipment.
RF emissions Class A
CISPR 11
Harmonic emissions Class A The Acclarix AX8 is suitable for use in all establishments
IEC 61000-3-2 other than domestic and those directly connected to the
public low-voltage power supply network that supplies
Voltage Complies buildings used for domestic purposes.
fluctuations/flicker
emissions
IEC61000-3-3
- 122 -
Acclarix AX8 Diagnostic Ultrasound System User Manual EMC Information
Electromagnetic immunity
Electromagnetic immunity
The Acclarix AX8 is intended for use in the electromagnetic environment specified below. The customer
or the user of Acclarix AX8 should assure that it is used in such an environment.
Electromagnetic
Immunity test IEC 60601 test level Compliance level
environment -guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood,
discharge (ESD) concrete or ceramic tile. If
±8 kV air ±8 kV air
IEC 61000-4-2 floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast ±2 kV for power supply ±2KV for power supply Mains power quality
transient/burst IEC lines lines should be that of a typical
61000-4-4 commercial or hospital
environment.
Surge IEC ±1 kV line to line ±1 kV line to line Mains power quality
61000-4-5 should be that of a typical
±2 kV line to ground ±2 kV line to ground
commercial or hospital
environment.
Power frequency 3A/m 3A/m Power frequency magnetic
(50/60Hz) magnetic fields should be at levels
field IEC61000-4-8 characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short <5% UT <5% UT Mains power quality
interruptions and should be that of a typical
(>95% dip in UT) (>95% dip in UT) for
voltage variations commercial or hospital
0.5 cycle
on power supply for 0.5 cycle environment. If the user of
input lines IEC the Acclarix AX8 requires
61000-4-11 continued operation during
40% UT (60% dip in UT)
40% UT (60% dip in UT) power mains interruptions,
for 5 cycles
for 5 cycles we recommend that the
Acclarix AX8 be powered
from an uninterruptible
70% UT (30% dip in UT)
70% UT (30% dip in UT) power supply or a battery.
for 25 cycles
for 25 cycles
NOTE UT is the a.c. mains voltage prior to application of the test level.
- 123 -
Acclarix AX8 Diagnostic Ultrasound System User Manual EMC Information
Electromagnetic immunity
Electromagnetic immunity
The Acclarix AX8 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Acclarix AX8 should assure that it is used in such an environment.
IEC
Compliance
Immunity test 60601 Electromagnetic environment -guidance
level
test level
Portable and mobile RF communications equipment
should be used no closer to any part of the Acclarix AX8,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted 3 Vrms 3Vrms
RF IEC d 1.2 P 150kHz ~ 80 MHz
150 kHz
61000-4-6
to 80
MHz
Radiated RF 3 V/m
3 V/m d 1.2 P 80 MHz ~ 800 MHz
IEC 80 MHz d 2.3 P 800 MHz ~ 2.5 GHz
61000-4-3 to 2.5
GHz Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,a should be less than
the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Acclarix AX8 is used exceeds the applicable RF compliance level
above, the Acclarix AX8 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the Acclarix AX8
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
- 124 -
Acclarix AX8 Diagnostic Ultrasound System User Manual EMC Information
The Acclarix AX8 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Acclarix AX8 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Acclarix AX8 as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum 150 kHz ~ 80 MHz 80 MHz ~ 800 MHz 800 MHz ~ 2.5 GHz
output power of
transmitter d 1.2 P d 1.2 P d 2.3 P
(W)
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
- 125 -
- 126 -









