99T00069

This certificate of calibration summarizes the calibration of a PT100 temperature sensor with a serial number of 99T00069. The sensor was calibrated on February 15, 2021 at Dunlop Dudelange Tires Operations S.A. in Luxembourg. It was calibrated at 5 temperature points between 20 and 120 degrees Celsius and found the sensor measurements to be within 0.2 degrees of the reference values with an uncertainty of 0.12 to 0.16 degrees.

Uploaded by

Hamilton William
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as TXT, PDF, TXT or read online on Scribd

Topics covered

  • Calibration Verification,
  • Calibration Results,
  • Calibration Process,
  • Calibration Recommendations,
  • Ambient Temperature,
  • Instrument Information,
  • Measurement Accuracy,
  • Calibration Technician,
  • Calibration Standards,
  • Calibration Compliance
0% found this document useful (0 votes)
65 views4 pages

99T00069

This certificate of calibration summarizes the calibration of a PT100 temperature sensor with a serial number of 99T00069. The sensor was calibrated on February 15, 2021 at Dunlop Dudelange Tires Operations S.A. in Luxembourg. It was calibrated at 5 temperature points between 20 and 120 degrees Celsius and found the sensor measurements to be within 0.2 degrees of the reference values with an uncertainty of 0.12 to 0.16 degrees.

Uploaded by

Hamilton William
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as TXT, PDF, TXT or read online on Scribd

Topics covered

  • Calibration Verification,
  • Calibration Results,
  • Calibration Process,
  • Calibration Recommendations,
  • Ambient Temperature,
  • Instrument Information,
  • Measurement Accuracy,
  • Calibration Technician,
  • Calibration Standards,
  • Calibration Compliance

CERTIFICATE OF CALIBRATION

2102-13237
Customer information
Client : Goodyear
Dunlop Dudelange Tires Operations S.A.
Contact : Ahmet KOSE
Address : Zone
Industrielle Riedgen
3451,
Dudelange
Luxembourg

Reference Trescal : TP-37823

Instrument information
Make : JUMO
Type : 902050/10
Description : PT100
Serial number :
Identification number : 99T00069
Range : 20 .. 120
°C
Accuracy : Class B (EN
60751): ± (0,3 + 0,005.|t|) °C

Date of calibration : 15 February


2021

Used method
P2-02-S.002 Calibration of temperature sensors (with or without readout) in
calibrators.

Environmental conditions
During this service the following conditions were not exceeded :
Ambient temperature : 23°C ± 5°C
Relative humidity : 45%rh ±
20%rh

Used reference
The equipment used is traceable to National and/or International standards.
R1965 Hygro- / Thermometer
cert n° 200119422
R3615 Dry Block Calibrator & Ref. Probe
cert n° 201012087
R3731 Dry Block / Liquid Bath Calibrator & Ref. Probe
cert n° 200119421
R4400 Dry Block / Liquid Bath Calibrator & Ref. Probe
cert n° 191105454
R4452 Multifunctional calibrator
cert n° 200413969
Note
This service was performed at the clients location.
The instrument is measured but not adjusted, so the results are both 'as found' as
'as left'.
Technician Issue date 15 February 2021
Head of the laboratory
Gregory Gaspart
Luc Van Pelt

This document may not be reproduced other than in full, except with the prior
written approval of the head of the issuing laboratory.
Unless otherwise stated, the calibration was performed at the address mentioned in
the footnote.

Trescal nv | Vosstraat 200 | Berchem (Antwerpen) | Belgium | Telephone + 32 3 542


62 90 | Email: [email protected]

Page 1 to 2
CERTIFICATE OF
CALIBRATION

2102-13237
Initial results:
- Reference Units Instrument Units Calculated
Difference Uncertainty Units
value value value
±
1 20.00 °C 107.75 Ohm 19.88 -
0.13 0.12 °C
2 50.00 °C 119.35 Ohm 49.87 -
0.13 0.12 °C
3 89.71 °C 134.67 Ohm 89.87
0.16 0.16 °C
4 120.03 °C 146.01 Ohm 119.83 -
0.20 0.16 °C
5 20.00 °C 107.75 Ohm 19.88 -
0.12 0.12 °C

The stated uncertainty is that of the entire set-up including the object under
test.
The reported uncertainty is based on a standard uncertainty multiplied by a
coverage factor k=2, providing a
level of confidence of approximately 95% .
The uncertainty is calculated following EA-4/02 in accordance with the requirements
of the ISO/IEC 17025.