SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
CW Module installation guide in DC-3
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
REVISION CONTROL
Document Title: CW Module installation guide in DC-3
Author: Mindray Service Department
Date By Action Pages
June 26, 2008 Klaus Hu New document
Review/Approval History
Date By Action Pages
Last Modified: June 17, 2007 Page 2 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Contents
Page
1 BEFORE DISASSEMBLY ................................................................................................................................4
THINGS REQUIRED:.....................................................................................................................................................4
NOTE1: CERTAIN ESD PROTECTION IS NEEDED. ........................................................................................................4
NOTE2: DO NOT CUT OFF THE CABLE HOOPS ............................................................................................................4
NOTE3: MAKE SURE IT SHOULD BE ESPECIALLY CAREFUL WHEN YOU UNPLUG PART FROM OTHER PART AND DO THE
REVERSE. ................................................................................................................................................................... 4
2 OPEN THE MACHINE .....................................................................................................................................5
3 ADDING CW BOARD ..................................................................................................................................... 17
4 CW SOFTWARE INSTALLATION .............................................................................................................. 18
Last Modified: June 17, 2007 Page 3 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
1 Before disassembly
Things required:
a. CW board
b. CW license key disk
Note1: Certain ESD protection is needed.
Electro Static Discharge is very harmful to the boards inside the system. Please adopt certain
measures to protect the boards from ESD especially in dry environment.
Note2: DO NOT cut off the cable hoops
There are many cable hoops attached to the shield covers. Please DO NOT cut them off, otherwise, it may
influence the cable layout when you assemble the system.
Note3: Make sure it should be especially careful when you unplug part from
other part and do the reverse.
Last Modified: June 17, 2007 Page 4 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
2 Open the machine
Before you can see the inner structure of DC-3, please remove the Rear Cover.
Step 1: Remove all the Silicon Rubber Caps on the Rear Cover
Step 2: Remove all the Screws on the Rear Cover
Note: there are three Screws at the bottom of the Rear Cover.
Last Modified: June 17, 2007 Page 5 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 3: Pull out the Rear Cover
Last Modified: June 17, 2007 Page 6 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Now you can see I/O interface and AC Power Input Module.
I/O interface
AC Power
Input Module
Step 4: Remove six screws on both side of the module
Last Modified: June 17, 2007 Page 7 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 5: Rotate the panel in the indicated direction
Step 6: Remove the three screws on the right side of the I/O Interface module.
Last Modified: June 17, 2007 Page 8 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 7: Rotate the panel in the indicated direction
Step 8: Disconnect the soft cable before you take out the panel
Please remove
the fragile cable
carefully, and
there is a Red
Line on the
upper position.
Last Modified: June 17, 2007 Page 9 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 9: Then we can list up the module and pull it out
2) Pull it out
The metal sticks are
1.
used to locate the
module.
1) Lift the panel up.
Step 10: Disconnect all cables and connectors
Last Modified: June 17, 2007 Page 10 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Remove the screws on the main box
Last Modified: June 17, 2007 Page 11 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 11: Remove front bottom cover
Remove three screws
Pull out front bottom cover
Last Modified: June 17, 2007 Page 12 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 12: Remove front cover
Remove plastic caps and screws on both sides of the front cover
Pull out the front cover
Last Modified: June 17, 2007 Page 13 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 13: Remove the screws on the front side of the unit
Step 14: pull out the main box
Last Modified: June 17, 2007 Page 14 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Step 15: Remove screws and pull out the I/O port Board
Step 16: Remove screws and metal cover
Last Modified: June 17, 2007 Page 15 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
Last Modified: June 17, 2007 Page 16 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
3 Adding CW board
Add the CW board to its place like following picture.
Be careful of the slot that they fix themselves.
CW board
Last Modified: June 17, 2007 Page 17 of 18
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
4 CW software installation
(1) Press [Setup] key [System] menu [Option] menu
CW software installed
(2) Select CW
(3) Click ‘Install’ menu
(4) Run the license key file in driver F:
(5) System info ‘ install success’.
PS: the license key file is unique related with Serial number and Mac address.
Congratulations, now you have Continuous Wave Doppler.
In CW mode, kindly use phased array transducer 2P2 to do scanning.
Last Modified: June 17, 2007 Page 18 of 18