Instr METASYS EXCOM Hybrid
Instr METASYS EXCOM Hybrid
1. Index:
Page
1. Index 2
2. Explanation of the pictograms 2
GB 3. General information 3
4. Use 4
5. Type overview 4
6. Construction 5
7. Accessories 6
8. Explanation of the type plate 7
9. Technical data 8
10. Description of function 9
11. Installation guidelines 10
12. Hose connections 11
13. Electrical connections 12
14. Connection COMPACT A8 / A16 with EXCOM hybrid 2/5 14
15. Maintenance, cleaning and disinfection 15
16. Commissioning 16
17. Maintenance 17
18. Disposal of the equipment 17
3. General information:
4. Use:
GB
5. Type overview:
The EXCOM hybrid central suction system can be delivered
in four models:
EXCOM hybrid 2: E
XCOM hybrid 2 is a central suction
system with integrated air/water sepa-
ration for simultaneous operation of
2 treatment units or for up to 3 treatment
2
units with 60% efficiency.
EXCOM hybrid 5: E
XCOM hybrid 5 is a central suction
system with integrated air/water sepa-
ration for simultaneous operation of
3 treatment units or for up to 5 treatment
units with 60% efficiency.
6. Construction:
5 See illustration
5.3
taking EXCOM hybrid 5 as an example
5.4 Prefilter
Coarse solid particles are held back in the prefilter
5.2
1 See illustration
GB 2 See illustration
Spittoon valve:
Order number: 40050002
3 See illustration
Covering hood:
Covering hood for EXCOM hybrid 2/5:
Order number: 40030004
4 See illustration
5 See illustration
The noise reducing item minimizes the sound level of the air
inlet valve.
6 See illustration
Type plate of EXCOM hybrid 2/5:
The type plate can be found at the outside of the suction
motor's silencer 6.1 , above the prefilter 6.2 .
6.2
7 See illustration
Explanation of the type plate, taking EXCOM hybrid 5 - 230 V
as an example:
EXCOM hybrid 5
Un = 230V, 50/60 Hz
Pmax = 1,1/1,3 kw
EA-12345 000
pmax = 180 mbar
Made in Austria
*reguliert
9. Technical data:
1 EXCOM hybrid 2
230 V 400 V
Mains voltage: 230 V AC 400 V AC
Frequency: 50/60 Hz 50/60 Hz
Maximum current consumption: 6,5/7,4 A 2,5/2,8 A
Electrical shaft power: 0,75/0,9 kW 0,75/0,9 kW
Suction motor rotation speed, rpm: 2770/3250 2780/3300
Vacuum, regulated: 180 mbar 180 mbar
Maximum ambient temperature: 40º C 40º C
Operating time: 100% 100%
GB Weight: 31 kg 32 kg
Noise level without covering hood: 63 dB (A) 63 dB (A)
Noise level with covering hood: 58 dB (A) 58 dB (A)
Overall dimensions (L x W x H) cm: 441x342x620 441x342x620
2 EXCOM hybrid 5
1
230 V 400 V
Mains voltage: 230 V AC 400 V AC
Frequency: 50/60 Hz 50/60 Hz
Maximum current consumption: 9,2/7,9 A 3,3/3,7 A
Electrical shaft power: 1,1/1,3 kW 1,1/1,3 kW
Suction motor rotation speed, rpm: 2800/3300 2880/3400
Vacuum, regulated: 180 mbar 180 mbar
Maximum ambient temperature: 40º C 40º C
Operating time: 100% 100%
Weight: 34 kg 35 kg
Noise level without covering hood: 65 dB (A) 65 dB (A)
Noise level with covering hood: 58 dB (A) 58 dB (A)
Overall dimensions (L x W x H) cm: 499x363x626 499x363x626
3 See illustration
15 cm
Its use in areas subject to explosive and fire
hazards is not permitted.
1 See illustration
q When the EXCOM hybrid suction system is installed, the
connection side must be placed at least 15 cm from the
1/1
2 See illustration
EA-54.202/00
I
be switched off at the main switch.
Any pipe or hose used must be vacuum tight and
resistant to all chemicals normally used in a dental
practice (e.g. HT discharge pipes made from PP,
3 PVC-C, PVC-U, PE-HD).
4 See illustration
q Only flexible spiral hoses made from PVC or equivalent
materials may be used.
90°
q Connections to the EXCOM hybrid central suction system
must be made by flexible hoses and be as short as
possible.
Mains connection:
The mains connection must only be carried out by a trained
electrician.
The electrical installation must be carried out in accordance
with applicable local regulations.
Before connecting with the mains, the nominal voltage
stated on the type plate on the equipment must be
compared with the mains voltage.
The EXCOM hybrid suction system must only be connected to
the power supply with the supplied power cable. Extension
GB
cables must not be used.
Main switch:
Connection to the mains must be established after the
practice's main switch.
1 Legend:
F1 fuse T2A
K1 motor contactor
K2 timing relay
M1 motor for vac
Q1 protection switch
S1 main switch
T1 transformer
V1 varistor K275
12 Page GB PracticeTechnicians
staff / technicians
Operating components
S1
Follow-up time:
F1
Legend: W3
F1 fuse
K1 motor contactor
K2 timing relay
S1 main switch T1
T1 transformer 3
V1 varistor
W1 power supply
W2 motor for vac T1
W3 sucking contact
S1
W1
F1
W2
W3
K1 K2
44
PracticeTechnicians
staff / technicians GB Page 13
Connection COMPACT A8/A16
with EXCOM hybrid 2/5
GB
max. 0,4 m
In case of high volumes of liquid and sponta-
neously fulminating large quantities of water
an expansion container (Item No. 40040014)
1
needs to be installed between the suction
system (EXCOM hybrid) and the amalgam
separator (METASYS COMPACT A8/A16).
2.2 2.3
2.2 expansion container
min. 2%
max. 0,4 m
3 See illustration 4
3
Following every operation, the rinsing basin must be rinsed
GB
thoroughly!
4 See illustration
Suck off some water with each of the suction tubes after
every treatment!
5 See illustration
Twice a day, after having sucked off some water, use the
disinfectant for amalgam separators Green&Clean M2.
Ideally a disinfection with Green&Clean M2 should be
carried out before longer periods of downtimes of the
dental unit (e.g. lunch break, working day or holidays).
5
6 See illustration
T h e s p i t t o o n b o w l s h o u l d a l s o b e r i ns e d wi t h
GREEN&CLEAN M2 twice a day.
q Cleaning the filter
At least once a week clean and empty the filter. The cleanout
can also be carried out more frequently if necessary.
7 See illustration
Any residues from the prefilter, which might contain
amalgam, must be collected in a suitable container
(e.g. METASYS EcoCenter) and disposed off properly with
EcoTransform.
Please go to
www.metasys.com
to locate a world-wide collection point in your vicinity.
16. Commissioning:
1 See illustration
q Switch on practice and the equipment main switch.
GB
2 See illustration
q Remove the suction hose from its holder.
q Check that all hose connections and other connections
in the suction pipe are air-tight.
4 See illustration
q Suck 3 litres of water and check that the EXCOM hybrid
central suction system is operating correctly.
q Undertake electrical safety checks as required by local
legislation, and record that the checks have been made.
17. Maintenance:
5 See illustration
q The following filters must be checked and cleaned every
week:
Austria
METASYS
Medizintechnik GmbH
Florianistraße 3
A-6063 Rum bei Innsbruck
1 +43/(0)512 /20 54 20-0
5 +43/(0)512/20 54 20-7
Germany
METASYS
Medizintechnik GmbH
Ahornstraße 19
D-85614 Kirchseeon
1+49/(0)2236 / 37 42 42
5+49/(0)2236 / 37 42 41
France
METASYS France S.a.r.l.
9, bd E. Michelet
F-69008 Lyon
1+33/(0)4/37 90 22 15
5 +33/(0)4/37 90 22 47
e-mail: [email protected]
email: [email protected]
www.metasys.com
Italy
ZK-55.064/00
70100031
EBW EXCOM hybrid D/GB/FR/I, 19.03.2012