FMI 12267 Instruction Pristina M3-2 SP1 - Rev3
FMI 12267 Instruction Pristina M3-2 SP1 - Rev3
FMI TYPE
☐Recall/Reportable
☒Business – Mandatory/Not -Reportable
☐Business – On Request/Not -Reportable
Section 1: RESOURCES:
Personnel:
1 Field engineer
Time:
2D system: 5 hours
3D system: 6 hours
None
Page 1 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Section 5: EFFECTIVITY
FMI 12267 is a mandatory FMI for all Pristina IB systems 5582323 (see note below) except the Pristina systems with
Biopsy option.
The Field Action is to upgrade the system Software to “M3-2/SP1” level
FMI 12267 Consignee list: DOC2124548 rev1.
Note: to identify your system, model label is present on the system SIDE LABEL PLATE (for more details, see Pristina
Core Service Manual –Product History and Compatibility - part 2-2 How to Identify Your Model Number and UDI
Labeling)
The current FMI aims to upgrade Pristina Software to “M3-2/SP1” version. It will allow to align all Pristina Systems to
latest features level and software corrections.
Depending on initial versions of the system, the upgrade will bring features and fixes as described in Appendix 2.
Pristina Software and Firmware history is listed in Appendix 1.
The main corrections brought by FMI 12267 software upgrade are related to Quality Check and system boot.
Depending on the initial level of Software on site, the customer will get additional changes. Refer to Appendix 2.
None
Section 8: PREREQUISITE
The FM12267 – Seno Pristina M3-2 SP1 12267 upgrade Kit (PN: 5799566) contains the parts below:
Page 2 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
5797416 RECON v02.8.7 Operating System ref [5]
CDs OM (UPCD)
5721683-1-299 Senographe Pristina™ - User Publication CD ref [6]
5764911-299 Senographe Pristina™ 3D User Publication CD ref [7]
5722122-299 SenoBright™ HD - User Publication CD ref [8]
CDs Service (SPCD)
5762781-2EN Senographe Pristina™ - Service Publication CD Class A ref [9]
5721827-1-2EN Senographe Pristina™ 3D Service Publication CD Class A ref [10]
5721867-2EN Senographe Pristina™ Mobile - Service Publication CD Class A ref [11]
5722124-2EN SenoBright™ HD - Service Publication CD Class A ref [12]
Techpub Paper (OM extract) + CNC
5721681-1-199 Senographe Pristina™ - Operator Manual Extract - Publication Folder Multilingual ref [13]
Senographe Pristina™ 3D Operator Manual Extract - Publication Folder
5759050-1-199
Multilingual ref [14]
5721863-199 Senographe Pristina™ Mobile - Operator Manual Publication Folder ref [15]
5721864-199 Senographe Pristina™ Mobile OM Quick Guide - Publication Folder Multilingual ref [16]
M3-2 SP1 Upgrade instruction and MQSA Release Note - Paper
5803739 M3-2 SP1 upgrade instruction ref [17]
Appendix Techpub Paper - specific country
Senographe Pristina™ Senographe Pristina™ 3D - Registration Information - 2D/3D
5792632-199
For Argentina ref [18]
Senographe Pristina™ - Operator Manual Addendum for Russia Publication Folder
5789145-199
Multilingual - For Russia ref [19]
Labels
5780810 UDI label component ref [20]
5780813 Identification label for upgrade kit ref [21]
Product Locator (PLC)
5780785TRA Upgrade kit with M3-2 software ref [22]
Page 3 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Section 9: PROCEDURE
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
1 Pre-installation Check the system IS NOT a Biopsy system 2
If the system has the Biopsy option, the FMI shall not be
applied: FMI needs to be cancelled.
2D only
3D
3D+2D
3D+2D US
Mobile
SenoBright (CESM)
Page 4 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
2 Pre-installation Brake functional check Refer to core Job Card CHK 15
Calibration suitcase
A001 – Brakes Functional
(5604511)
This check will ensure the brake functionality Torque Test in Core Service
Manual - Ref [A]. 3 cable tie of 40cm
If the test fails go on with the current FMI procedure (no length each
need to consign the system) and plan a corrective job to
apply the required Brake run-in (as described
Completion chapter of Jobcard CHK A001 Completion
chapter). Please note that you will need to get the Brake
replacement kit – 5750487 (loan tool)
M3-1
M3-1/UP1
Page 5 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
M3-1/UP1-2
M3-1/UP1-3
M3-1/UP2
M3-2
M3-2/SP1
Page 6 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
7 Pre-installation Passwords record 5
Clinical:
Admin:
Service:
sdc:
root:
insite:
recon:
8 Pre-Installation Perform a backup SSA Key 5
Page 7 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
3. Name the backup file “FMI1” for the first backup
you want to create.
4. Insert your USB flash memory device into the
Console USB port and enable the USB backup check
box.
Click the START button.
5. Once Back-up successfully ends (the button
changes from In Progress to START),
Page 8 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
9 Pre-Installation Only for USA, 3D+2D license check 2
Note: 3D+2D (For US) will not exist anymore with Pristina
“M3-2/SP1” release. US sites will have same license as in
rest of the world. Systems configured with 3D+2D (For US)
license, will lose the corresponding license at end of
installation.
If it is a 3D system with license 3D+2D (For US), It will be
necessary to configure manually 3D+2D license after
installation.
Go to License pages:
1. Insert your Service Key into the USB port of the AXIS
2. From Tools menu, select:
Service>Service Desktop > Configuration> Option
Management.
3. Note down if 3D+2D (US) option is checked
NO
Page 9 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
Service>Service Desktop>Configuration
>System >Net Routing:
User Preferences>Gantry Preferences>
Motorized compression maximum force:
User Preferences>Gantry Preferences>
Auto-decompression height:
User Preferences>Gantry Preferences>
Rotation default oblique angle:
User Preferences>Gantry Preferences>
Centering light for interventional
duration:
Service>Service Desktop>Configuration
>General>SITE>Language>Español
Antivirus Activated:
No
YES
Page 10 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
If Yes, note down or take a picture:
IP Address:
Port No:
User Name:
Password:
12 Pre-Installation In case of need by customer, make an export of the Refer to 2D QC manual, section CD-ROM 10
Repeat and Reject Analysis database to leave on site. Repeat Analysis - Data- base
Use the AWS Export Data- base feature: backup in TEST RAD TECH 09 - OR
Repeat analysis check
USB flash memory
1. Select the Tools icon menu, select IQTool, and device
then RepeatRejectAnalysis button.
2. Insert a blank CD or a USB key
3. Click on either button:
- Export database on CD-ROM
- Export database on USB
Page 11 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
14 Installation Install AXIS KEEPING AXIS DATA (All systems) AXIS OS – ref [1] 60
Refer to AXIS Booklet AXIS SW – ref [2]
Note: FOR 3D systems only, the Axis
and Recon installation can be
performed in parallel to gain time.
1. Install AXIS OS keeping AXIS data
Using the AXIS OS CD
Follow Axis Booklet instruction but Ignore the
“fresh install” and select “Install and preserve Axis
Data Partition” instead
Page 12 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
16 Installation If applicable, Configure AntiVirus Refer to: 30
Configure the ePO antivirus with the information noted 2D SM, section 6-4 of Job Card
before performing the restore (step 11 of this instruction) PHY A006 - Control Station
Page 13 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
AXIS Computer rear panel connectors:
Page 14 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
For 3D systems:
- replug the ethernet cables between AXIS
computer and Reconstruction Computer, and
external Network as it was originally
-
17 Installation Restore all AXIS parameters from hard drive backup Refer to core Job Card ELE A040 - SSA Key 10
“FMI1date.tar” Backup and Restore Procedure
from core service manual - Ref USB flash memory with
[A]. backup
1. Insert SSA key in USB port of the AXIS Computer
2. From Tools menu, select:
Page 15 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
18 Installation For 3D systems, Reestablish connectivity with the 3D Job card ELE A002 – 15
Reconstruction Computer Reconstruction Computer and/or
AXIS Computer Post Install
Configuration section 7.2.1
1. Power on manually the AXIS Computer by pressing
“Reestablish connectivity with
the power button on the front panel of the AXIS
reconstruction computer in 3D
Computer
Service Manual- Ref [B].
2. Power on manually the Reconstruction Computer
by pressing the power button on the front panel of
the Reconstruction Computer
Page 16 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
Note down the Axis hostname (or take picture)
c. Click the 3D Option node, and click Networking
d. In the Axis Hostname section, in the Axis Name
field enter the exact same parameter that you
have entered as HostName in Step b.
19 Installation Restore the reconstruction Computer Configuration 3D Job Card ELE A002 – 15
Reconstruction Computer and/or
Axis Computer Post Install
1. From Tools menu, select:
Configuration section “Restoring
Utilities > System Tools> Restore the reconstruction Computer
2. Under RESTORE ITEMS select configuration (post LFC)” from
ReconstructionSettings and the 3D Service Manual
ReconstructionAlgorithm
3. Under Media select “FMI1date.tar” from
Internal HDD
4. Click START button
Note: message "This backup was not performed on
the current system. Are you sure you want to
restore it” might display, please ignore and go on
with the process
5. Check restore is successful:
2/2 items restored: SUCCESS
6. Reboot the system
Note: If restore from HDD fails, please use
FMI2date.tar on HDD or, the backup on USB stick
20 Installation User Publications installation Core Job Card ELE A025 - Install Senographe Pristina™ - 15
User Publications in Core Service User Publication CD
manual - Ref [A]. UPCD – ref [6]
1. From Tools menu, select:
Service>Service Desktop > Utilities > If applicable,
eDocumentation > User eDoc. Senographe Pristina™
Page 17 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
2. Insert the User Publication Core CD-ROM: 3D - User Publication
User Publication CD – Senographe PristinaTM CD
UPCD – ref [7]
from FMI kit in the optical drive of the AXIS
Computer.
If applicable,
3. Click the install button.
SenoBrightTM HD (CESM
4. After Success install message is displayed, eject the HD) - User Publication
CD-ROM: CD
a. Select WorkStation Tools node>CD-ROM Eject. UPCD – ref[8]
b. click the EJECT button
21 Installation For 3D systems only: 3D Job Card IST C002 - License SSA key 5
activation in 3D Service Manual -
Activate 3D+2D option Ref [B].
Note:
3D+2D mode is now activated for all 3D systems,
it is not an option anymore
for USA, if it was a 3D system configured with
3D+2D (US) before the upgrade (see step 9 of this
instruction), you need to configure the system with
3D+2D.
1. Insert your Service Key into the USB port of the AXIS
2. From Tools menu, select:
Service>Service Desktop >
Configuration>SITE>Option Management
3. Select the 3D+2D check box (if not yet selected)
4. Submit
Page 18 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
22 Installation Backup/restore manual completion Core Job Card ELE A040 Backup 10
and Restore Procedure section 7
1. Reconfigure parameters recorded in previous step Completion in Core Service
10 of this instruction. Manual- Ref [A].
Page 19 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
24 Installation Remove all accessories from the gantry (paddles, bucky) 1
Page 20 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
26 Installation If Poseidon SW is NOT V2.1.11 Gantry and Generator Software POS-Djinn CD -ref [3] 40
Or SW and FW Installation Guide
Booklet
If Poseidon BL is NOT V2.12
(Refer to column “Needs Upgrade” in Appendix 1 to know
if Upgrade is needed)
Update Poseidon
Page 21 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
>Select Restore
>Click Next and Confirm
Reboot Gantry from X-ray Console
Wait end of gantry reboot before going on
Update HC12-MCU
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Positioner Node Download
>Next
2. Select Lift MCU and Install
3. Wait until gantry full automatic reboot
4. After reboot, select Continue
Page 22 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
If Lift DSP version is not V7.552.3
(Refer to column “Needs Upgrade” in Appendix 1 to know
if Upgrade is needed)
Update DSP
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select Lift DSP and Install
3. Wait until gantry fully reboots
Update CPLD
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select PatchLift CPLD and Install
3. Wait until gantry fully reboots
29 Installation If Rotation HC12 version is not V9.0.4 Gantry and Generator Software POS-Djinn CD -ref [3] 10
(Refer to column “Needs Upgrade” in Appendix 1 to know SW and FW Installation Guide
if Upgrade is needed) Booklet
Update HC12-MCU
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select Rotation MCU and Install
Page 23 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
3. Wait until gantry fully reboots
Update DSP
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select Rotation DSP and Install
3. Wait until gantry fully reboots
Update CPLD
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select PatchRotation CPLD and Install
3. Wait until gantry fully reboots
30 Installation If Compression HC12 version is not V7.392.1 Gantry and Generator Software POS-Djinn CD -ref [3] 10
(Refer to column “Needs Upgrade” in Appendix 1 to know SW and FW Installation Guide
if Upgrade is needed) Booklet
Update HC12-MCU
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
Page 24 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
2. Select Compression MCU and Install
3. Wait until gantry fully reboots
Update CPLD
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select PatchCompression CPLD and Install
3. Wait until gantry fully reboots
31 Installation If Tilt HC12 version is not V7.272.0 Gantry and Generator Software 10
POS-Djinn CD -ref [3]
(Refer to column “Needs Upgrade” in Appendix 1 to know SW and FW Installation Guide
if Upgrade is needed) Booklet
Update HC12-MCU
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select Tilt MCU and Install
3. Wait until gantry fully reboots
Page 25 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
If Tilt CPLD version is not V2.7.5
(Refer to column “Needs Upgrade” in Appendix 1 to know
if Upgrade is needed)
Update CPLD
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select PatchTilt CPLD and Install
3. Wait until gantry fully reboots
32 Installation If IRLOCK HC12 version is not V3.282.10 Gantry and Generator Software POS-Djinn CD -ref [3] 10
(Refer to column “Needs Upgrade” in Appendix 1 to know SW and FW Installation Guide
if Upgrade is needed) Booklet
Update HC12-MCU
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select Locking MCU and Install
3. Wait until gantry fully reboots
Update CPLD
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select PatchLocking CPLD and Install
3. Wait until gantry fully reboots
Page 26 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
33 Installation If Collimator FW version is not V1.19 Gantry and Generator Software POS-Djinn CD -ref [3] 10
(Refer to column “Needs Upgrade” in Appendix 1 to know SW and FW Installation Guide
if Upgrade is needed) Booklet
Update FW
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select Collimator FW and Install
3. Wait until gantry fully reboots
Update FW
1. From Tools menu, select:
Service>Service Desktop > Utilities > SW/ FW
Upgrade > Positioner > Next
2. Select Bucky Elixir and Install
3. Wait until gantry fully reboot
36 Installation Check the Djinn Generator Versions, Gantry and Generator Software POS-Djinn CD -ref [3] 10
If Application is NOT D7.3.0 SW and FW Installation Guide
or Booklet
Page 27 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
If FPGA is NOT A2.05
Or
If Rotation Firmware is NOT 2.11
Or
if Rotation Boot is NOT 1.04
(Refer to column “Needs Upgrade” in Appendix 1 to know
if Upgrade is needed)
Page 28 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
39 Check For 3D systems only, check Axis and Reconstruction PC 2
reboot automatically
40 Calibration and 1. Check the Gain Calibration status Core calibration Job cards: SSA Key 35
checks
Note: If initial SW release is prior to M3-1 UP2, some Job Card CAL A007 - Bad Pixel
calibration files are not restored as not compatible to new Calibration
software. Therefore, they will be marked N/A
Job Card CAL A008 - Detector
Gain Calibration
In tools menu select: Service>Service
Desktop>calibration home page. from Core Service Manual- Ref
Check if Gain grid calibration are “N/A” [A].
3D jobcard
From Tools menu, select:
CHK A003 - Flat Field test 3D
Service>Service Desktop> Calibration > Detector,
from 3D Service Manual- Ref [B
click Detector and execute the calibrations:
• Bad Pixel
• 2D Gain NoGrid calibrations (Mo/Mo & Rh/Ag)
• 2D Gain Grid calibrations (Mo/Mo & Rh/Ag)
Page 29 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
3. If above calibrations were redone, Perform IQ Tools
test
From Tools menu, select:
Service>Service Desktop > Image Quality
Execute the following IQ tests:
• 2D Flat field no grid (Mo Mo and Rh Ag)
• 2D Flat field grid (Mo Mo and Rh Ag)
• Grid line test.
• Grid texture
• For 3D systems only 3D contact flatfield tests
(Mo Mo and Rh Ag)
If applicable:
3D Job Card CHK A003 – Flat Field
test in 3D Service Manual - Ref
[B].
44 Check Do Gantry functional checks of all what was impacted by Core Job Card ELE A013 - Gantry 30
the current procedure. Functional Checks in Core Service
Manual- Ref [A].
Page 30 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
45 Check Connectivity test 60
check all connectivity features used by the customer
1. RIS: Request the creation of a test patient on the
RIS and perform a query
2. Perform exams that correspond to site workflow:
3. 2D standard
4. For 3D systems
a. 3D exams
b. 3D+2D exams
5. For CESM systems
a. CESM exams
6. Close the exam and check all the autopush
transfer are done as configured (check the status
in image browser, but also that images are
present on distant system)
7. Do manual push to all configured PACS and
review stations
8. If customer uses it: perform a Query/Retrieve on
PACS
9. If applicable, print images
10. If site uses an iCad system, make sure all needed
images reach iCad and results goes to review
system (specifically If a SenoIris Connect is used
for routing the images)
11. If site uses a Dicom router system (SenoIris
Connect) make sure the images reach the end
destination, not only to the router system.
12. If site uses Generated 2D (v-preview, enhanced v-
preview) router workflow, make sure that the
Generated 2D are processed and dispatched
correctly to the destinations.
Page 31 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
3D+2D mode have a different value in the
DICOM attribute “Protocol name” (0018,1030).
If a Dicom Router such as the SenoIris Connect
uses this attribute in the routing rules, the
routing will fail; and images will not reach
destination
48 Completion Perform a full backup on Hard-Drive Core Job Card ELE A040 - Backup SSA Key USB Key 10
and Restore Procedure in Core
> Name of backup file: “postFMI1”
Service Manual- Ref [A]
Perform a full second backup on USB and Hard-drive
> Name of backup file: “postFMI2”
49 Completion Aide security diagnostic Core Job Card DIAG C006 – SSA Key 10
Security Diagnostics in Core
Note: the following can be done in parallel
Service Manual - Ref [A].
1. Go in: Service Desktop > Diagnostics >
SUBSYSTEMS > SECURITY > AIDE > START
Page 32 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
2. Run "Generate system integrity database on the
AXIS computer"
3. Run “Check system integrity on the AXIS computer”
4. Run “Update integrity database on the AXIS
computer” (it runs 20min in background)
5. Run "Show last run results on the AXIS computer"
50 Completion Fix the identification and UDI labels on the label plate on Core Job Card PHY A015 – Fix UDI Identification label - 2
the left-hand side of the Gantry. labels in Core Service Manual- ref [20]
Ref [A]. UDI label – ref [21]
Page 33 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
If applicable, RECON
SW – ref [5]
53 Completion Dispose of older user SPCD available on site and replace Core SPCD – ref [9] 10
them with user SPCD supplied in the FMI 12267 kit.
If applicable,
Dispose of the user SPCD supplied in the FMI 12267 kit not
3D SPCD – ref [10] If
applicable to the site.
applicable,
Mobile SPCD – ref [11]
If applicable,
SenoBright SPCD – ref
[12]
54 Completion Provide the customer the applicable user technical Core OM extract – ref 10
publication delivered with the FMI 12276 kit. [13]
If applicable,
3D OM extract – ref
[14]
If applicable,
Mobile OM – ref [15]
If applicable,
Mobile quick guide –
ref [16]
If applicable,
Argentina registration
– ref [18]
If applicable,
Russia OM addendum
– ref [19]
55 Completion 5
Page 34 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI 12267 FIELD MODIFICATION INSTRUCTION FMI
Step Section Action Job Card/ Instructions Tools/Supplies Time Comments Done
(min)
56 Completion FE debriefing FMI PLC – ref [22] 5
Page 35 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
1. Dispose of the old materials (CD, DVD, and Technical Publications) according to local requirements.
2. Remove trash, installation equipment and tools. Clean-up room, store service manuals and wash hands after
handling FMI material.
3. Ensure that the system is ready for normal customer operation.
4. Following GEHC’s current processes, package and send unused materials to your pole’s renewable
resources center.
5. If printed on paper format, remove the FMI Instruction from the facility.
Page 36 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
FMI Completion Definition Debrief (PAT) Best Practice When / When Not to Use this Code
Code Status
Open Initial Code - Work has not yet
0 N/A N/A
been performed.
Implemented fix/modifications per FMI
This code should ONLY be used when
instructions; include testing performed
the service work is completed per
(if required per the FMI Instructions)
The Field Action has been the FMI instructions.
Attach and/or provide test &
completed on given serial
1 Completed inspection document objective
number per instructions. This code should NOT be used to
evidence as required to your Service
describe that the initial
Record *or submit to Service Admin
contact/inspection is complete when
/OP’s team.
the FMI was not executed.
Document details of why field action is This code should ONLY be used when
not required in the Service Record or criteria a-f applies
The Field Action is not required
submit objective evidence to Service This code should NOT be used when
because:
Admin/ OP’s team. the field action cannot be executed
a) Consignee List was Incorrect
due to:
b) Later Version SW
a) Consignee List was Incorrect • Being unable to schedule the
c) FMI previously installed
b) Later Version SW unit (storage, war,
d) Other Hardware changes
2 Not Required c) FMI previously installed postponement) – See Code 3
invalidating FMI
d) Other Hardware changes • Customer refuses the Recall
e) Customer altered equipment
invalidating FMI – See Code 4
f) Equipment was scrapped
e) Customer altered equipment • Cannot Locate the unit -See
destroyed (Do NOT use the term
f) Equipment was scrapped Code 5
retired for scrapped/destroyed
destroyed (Do NOT use the term
equipment)
retired for scrapped/destroyed
equipment)
The Field Action is not executed a, b, c; Document details of why field This code should ONLY be used when
because: action was not executed and what was criteria a-d applies
a) System is in a storage location done with the kit (left with system or This code should NOT be used when:
and field action kit has been returned to warehouse); in the Service • The customer has provided a
placed with device Record or FMI Completion Certificate specific future date to
b) For consignees (distributors), Report (Distributor) and submit execute on the field action –
3 Not Executed confirmed receipt of kit by objective evidence to Service Admin/ this is not a Postponement
consignee OP’s team. • Customer refuses the Field
c) The device is not reachable due Action – See Code 4
to War, Natural Disasters, d; attach postponement letter to • Cannot Locate the unit - See
Embargo, Quarantine, political Service Record or submit to Service Code 5
unrest, and a field action kit has admin. or Service Manager/OP’s team.
been reserved If customer refuses to sign
Page 37 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Document details of why field action This code should ONLY be used when
was not executed; attach customer criteria a-c applies
signed refusal letter to Service Record This code should NOT be used when:
or send document to your Service • Field action was not required
Consignee/Customer refuses
admin/OP’s team before closure. – See Code 2
delivery or installation of field
If customer refuses to sign, document • Field action was not
action kit and one of the
that attempt in detail on the Refusal executed - See Code 3
following criteria applies:
and/or Location Unknown Form or in • Cannot Locate the unit - See
a) Consignee refuses delivery of
the Service Record. Escalate to Director Code 5
the field action kit.
4 Refused of Service (DOS) or Service Manager.
b) Customer refuses to have field
They will complete second attempt by
action kit installed.
contacting higher authority at site.
c) Consignee/Customer has not
The second attempt should be
responded after 3 attempts
documented by the DOS or Service
(separate days) to contact.
Manager in the Service Record, e-mail
or on the Refusal Form and attach to
the Service Record or send to your
Service admin/ OP’s team before
closure.
Document attempts using the 3 This code should ONLY be used when
Attempts in the “Refusal and/or the device cannot be located.
Location Unknown Form or record the This code should NOT be used when:
required attempt documentation in the • The Field Action is not
Service Record. Documentation” required – See Code 2
included in the FMI instructions. • The Field Action could not
• First attempt must always be a executed due to getting
two-way interaction between access to the system – See
Consignee/Customer or Field GE and the consignee Code 3
Location
5 Engineer cannot find the device. (Distributor or End user). • Customer refusal – See Code
Unknown
• The person contacted by GE 4
must be a credible individual
with accurate knowledge of
medical device locations.
• This may also include physical
searches on site. Note: The
responsibility to track device
locations does ultimately
belong to the consignee.
Page 38 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Page 39 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Page 40 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI Completion Certification Report
NOTE: Distributors, Dealers or Third Party Service Providers ONLY. FMI #: _____________
Consignee/Customer Name: __________________________________________________________
Street Address: ____________________________________________________________________
City/State/Country: _________________________________________________________________
Country Code: _____________ Phone Number: __________________________________________
Completion Date
System ID Model Number Serial Number Completion Status*
(dd-mmm-yyyy)
☐1-Completed
☐2-Not Required
☐3-Not Executed
☐4 – Refused
☐5 – Location Unknown
☐1-Completed
☐2-Not Required
☐3-Not Executed
☐4 – Refused
☐5 – Location Unknown
☐1-Completed
☐2-Not Required
☐3-Not Executed
☐4 – Refused
☐5 – Location Unknown
☐1-Completed
☐2-Not Required
☐3-Not Executed
☐4 – Refused
☐5 – Location Unknown
*Refer to FMI Completion Guide
Comments:______________________________________________
___________________________________________________________
_____________________________________________
Please provide the signature of the individual submitting this report with responsibility.
Signature: _____________________________________________
Title: __________________________________________________
Page 41 of 51
Doc# DOC2123058 Rev. 1 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
System ID or Device Serial Number and description of Medical Device(s) transferred or sold to
Consignee/Customer:
We, <Consignee/Customer Name>, have elected to postpone the updates required by the Medical Device
Correction referenced above because one of the following conditions exist (Check that which applies).
Device has been un-installed and a return to service date is unknown at this time
Device cannot be taken out of service at this time, future availability is not known
Device has been damaged or modified, such that the updates cannot be made. The schedule for repairs to
the device is not known.
Device believed to be in Consignee/Customer inventory, but cannot be located at this time
OTHER:_________________________________________________________________
Disposition of associated Medical Device Correction Kit. (Check that which applies).
We, < Consignee/Customer Name >, understand the purpose of the correction and associated risk, and will notify
GEHC when the device has been located and is available for updates, or when the updates have been completed.
Please provide the signature of an individual with responsibility for risk and compliance.
Signature: _____________________________________________
Title: __________________________________________________
Page 42 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
We, <Consignee/Customer Name>, have elected not to accept the updates required by the Medical
Device Correction referenced above.
Please provide the signature of an individual with responsibility for risk and compliance.
Signature: ________________________________________________________________
Printed Name: _____________________________________________________________
Title: _____________________________________________________________________
Facility Name (i.e. Clinic or Site Name): ________________________________________
Address: __________________________________________________________________
City/State/Country: ________________________________________________________
Page 43 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
***NOTE: 2nd and 3rd attempts may be more appropriately made by the FMI program team. ***
Attempt 2
Device Name and Serial/ID Number: ______________________________________________________________
Name of the person making the contact (SSO# if applicable): _________________________________________
Date of the Contact (dd/mmm/yyyy): ______________________________________________________________
Contact’s Name: ________________________________________________________________________________
Contact’s Title/Position: __________________________________________________________________________
Contact's Facility Name: _________________________________________________________________________
Response: _______________________ ______________________________________________________________
Contact Method (i.e.; email address, phone) ________________________________________________________
Attempt 3
Device Name and Serial/ID Number: ______________________________________________________________
Name of the person making the contact (SSO# if applicable): _________________________________________
Date of the Contact (dd/mmm/yyyy): ______________________________________________________________
Contact’s Name: ________________________________________________________________________________
Contact’s Title/Position: __________________________________________________________________________
Contact's Facility Name: _________________________________________________________________________
Response: _______________________ ______________________________________________________________
Contact Method (i.e.; email address, phone):________________________________________________________
CONTACT" may include
o The appropriate imaging department, biomedical engineer, and/or asset manager at the last "ship to" address.
o A person with knowledge of medical device locations at the last "billing address.”
o The local Field engineer or Manager that covers "ship to" address o "billing address" area. GEHC employee name
if owned by GEHC.
Page 44 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Needs
Upgrade
M3-1 M3-1/UP1 M3-1/UP1.2 M3-1/UP1.3 M3-1/UP2 M3-2 M3-2/SP1
(√/X)
Software Versions
Axis OS NG-HELiOS-6.6.2-1.3 NG-HELiOS-6.6.2-1.3 NG-HELiOS-6.6.2-1.3 NG-HELiOS-6.6.2-1.3 NG-HELiOS-6.6.2-1.3 NG-HELiOS-6.8.1-1.1 NG-HELiOS-6.8.1-1.1
Axis SWPlatform PLATFORM_NGA- PLATFORM_NGA- PLATFORM_NGA- PLATFORM_NGA- PLATFORM_NGA- PLATFORM_NG-2.5-3 PLATFORM_NG-2.6-1
1.20-2 1.20-2 1.20-2 1.20-2 1.20-8
Axis Swacqsvc ACQSVC_NGA-1.20-2 ACQSVC_NGA-1.20-2 ACQSVC_NGA-1.20- ACQSVC_NGA-1.20- ACQSVC_NGA-1.20- ACQSVC_NGB-2.50- ACQSVC_NGB-2.50-17
2B 2B 12 14
Axis SwApps NG_SWAPPS_1.20.4 NG_SWAPPS_1.20.4 NG_SWAPPS_1.20.4 NG_SWAPPS_1.20.4 NG_SWAPPS_1.20.10 NG_SWAPPS_2.2.14 NG_SWAPPS_2.2.21
Axis Axis 1.50 1.50 1.50 1.50 1.50 2.1.11 2.2.55
Recon (3D OS RECON_OS_02.7.0 RECON_OS_02.7.0 RECON_OS_02.7.0 RECON_OS_02.7.0 RECON_OS_02.7.0 RECON_OS_02.8.2 RECON_OS_02.8.7
option)
Recon (3D Application RECON_SW_02.7.0 RECON_SW_02.7.0 RECON_SW_02.7.0 RECON_SW_02.7.0 RECON_SW_02.7.0 RECON_SW_02.8.2 RECON_SW_02.8.7
option)
XFOV Software 4.e 4.e 5.a 5.a 5.a 5.a 5.a
Detector
XFOV Firmware 6.2 6.2 6.5 6.5 6.5 6.5 6.5
Detector
DJINN Application D7.2.10 or D7.3.0 D7.2.10 or D7.3.0 D7.2.10 or D7.3.0 D7.2.10 or D7.3.0 D7.2.10 or D7.3.0 D7.3.0 D7.3.0
Generator
DJINN Boot B7.2.3 B7.2.3 B7.2.3 B7.2.3 B7.2.3 B7.2.3 B7.2.3
Generator
DJINN Firmware D7.2.10 D7.2.10 D7.2.10 D7.2.10 D7.2.10 D7.3.0 D7.3.0
Generator
DJINN FPGA A2.05 A2.05 A2.05 A2.05 A2.05 A2.05 A2.05
Generator
Page 45 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Needs
Upgrade
M3-1 M3-1/UP1 M3-1/UP1.2 M3-1/UP1.3 M3-1/UP2 M3-2 M3-2/SP1
(√/X)
Software Versions
DJINN Rotation 2.11 2.11 2.11 2.11 2.11 2.11 2.11
Generator Firmware
DJINN Rotation Boot 1.04 1.04 1.04 1.04 1.04 1.04 1.04
Generator
POSITIONER BUCKY BL EL_BTL SW 0.6 EL_BTL SW 0.6 EL_BTL SW 0.6 EL_BTL SW 0.6 EL_BTL SW 0.6 EL_BTL SW 0.6 EL_BTL SW 0.6
POSITIONER BUCKY FW ELIXIR V1.3 or ELIXIR ELIXIR V1.3 or ELIXIR ELIXIR V1.3 or ELIXIR ELIXIR V1.3 or ELIXIR ELIXIR V1.3 or ELIXIR ELIXIR V1.4 ELIXIR V1.4
V1.4 V1.4 V1.4 V1.4 V1.4
POSITIONER BIOPSY AXE X N.A. N.A. N.A. N.A. N.A. N.A. N.A.
BL
POSITIONER BIOPSY AXE X N.A. N.A. N.A. N.A. N.A. N.A. N.A.
FW
POSITIONER BIOPSY AXE Y N.A. N.A. N.A. N.A. N.A. N.A. N.A.
BL
POSITIONER BIOPSY AXE Y N.A. N.A. N.A. N.A. N.A. N.A. N.A.
FW
POSITIONER BIOPSY AXE Z N.A. N.A. N.A. N.A. N.A. N.A. N.A.
BL
POSITIONER BIOPSY AXE Z N.A. N.A. N.A. N.A. N.A. N.A. N.A.
FW
POSITIONER BIOPSY N.A. N.A. N.A. N.A. N.A. N.A. N.A.
DISPLAY BL
POSITIONER BIOPSY DISPAY N.A. N.A. N.A. N.A. N.A. N.A. N.A.
FW
POSITIONER COLLIMATOR BL01.10 BL01.10 BL01.10 BL01.10 BL01.10 BL01.10 BL01.10
BL
POSITIONER COLLIMATOR V1.19 V1.19 V1.19 V1.19 V1.19 V1.19 V1.19
FW
POSITIONER COMPRESSIO V3.2.4 V3.2.4 V3.2.4 V3.2.4 V3.2.4 V3.2.4 V3.2.4
N CPLD
POSITIONER COMPRESSIO V7.392.1 V7.392.1 V7.392.1 V7.392.1 V7.392.1 V7.392.1 V7.392.1
N HC12
POSITIONER COMPRESSIO V2.0.1 V2.0.1 V2.0.1 V2.0.1 V2.0.1 V2.0.1 V2.0.1
N HC12 PBL
POSITIONER COMPRESSIO V1.4.4 V1.4.4 V1.4.4 V1.4.4 V1.4.4 V1.4.4 V1.4.4
N HC12 SBL
POSITIONER IRLOCK CPLD V1.2.10 V1.2.10 V1.2.10 V1.2.10 V1.2.10 V1.2.10 V1.2.10
Page 46 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Needs
Upgrade
M3-1 M3-1/UP1 M3-1/UP1.2 M3-1/UP1.3 M3-1/UP2 M3-2 M3-2/SP1
(√/X)
Software Versions
POSITIONER IRLOCK HC12 V3.282.10 V3.282.10 V3.282.10 V3.282.10 V3.282.10 V3.282.10 V3.282.10
POSITIONER IRLOCK HC12 V2.0.10 V2.0.10 V2.0.10 V2.0.10 V2.0.10 V2.0.10 V2.0.10
PBL
POSITIONER IRLOCK HC12 V1.4.10 V1.4.10 V1.4.10 V1.4.10 V1.4.10 V1.4.10 V1.4.10
SBL
POSITIONER LIFT CPLD V2.5.1 V2.5.1 V2.5.1 V2.5.1 V2.5.1 V2.5.1 V2.5.1
POSITIONER LIFT DSP V7.552.3 V7.552.3 V7.552.3 V7.552.3 V7.552.3 V7.552.3 V7.552.3
POSITIONER LIFT DSP PBL V1.2.3 V1.2.3 V1.2.3 V1.2.3 V1.2.3 V1.2.3 V1.2.3
POSITIONER LIFT HC12 V7.562.2 V7.562.2 V7.562.2 V7.562.2 V7.562.2 V9.0.2 V9.0.2
POSITIONER LIFT HC12 PBL V2.0.2 V2.0.2 V2.0.2 V2.0.2 V2.0.2 V2.0.2 V2.0.2
POSITIONER LIFT HC12 SBL V1.4.1 V1.4.1 V1.4.1 V1.4.1 V1.4.1 V1.4.1 V1.4.1
POSITIONER PDU HC12 V0.172.8 V0.172.8 V0.172.8 V0.172.8 V0.172.8 V0.172.8 V0.172.8
POSITIONER PDU HC12 PBL V2.0.8 V2.0.8 V2.0.8 V2.0.8 V2.0.8 V2.0.8 V2.0.8
POSITIONER POSEIDON BL V2.12 V2.12 V2.12 V2.12 V2.12 V2.12 V2.12
POSITIONER POSEIDON SW V1.1.36 V1.1.38 V1.1.38 V1.1.39 V1.1.39 V2.1.9 V2.1.11
POSITIONER ROTATION V2.2.2 V2.2.2 V2.2.2 V2.2.2 V2.2.2 V2.2.2 V2.2.2
CPLD
POSITIONER ROTATION V7.552.5 V7.552.5 V7.552.5 V7.552.5 V7.552.5 V7.552.5 V7.552.5
DSP
POSITIONER ROTATION V1.2.5 V1.2.5 V1.2.5 V1.2.5 V1.2.5 V1.2.5 V1.2.5
DSP PBL
POSITIONER ROTATION V7.562.4 V7.562.4 V7.562.4 V7.562.4 V7.562.4 V9.0.4 V9.0.4
HC12
POSITIONER ROTATION V2.0.4 V2.0.4 V2.0.4 V2.0.4 V2.0.4 V2.0.4 V2.0.4
HC12 PBL
POSITIONER ROTATION V1.4.2 V1.4.2 V1.4.2 V1.4.2 V1.4.2 V1.4.2 V1.4.2
HC12 SBL
POSITIONER TUBE TILT V2.7.5 V2.7.5 V2.7.5 V2.7.5 V2.7.5 V2.7.5 V2.7.5
CPLD
POSITIONER TUBE TILT V7.272.0 V7.272.0 V7.272.0 V7.272.0 V7.272.0 V7.272.0 V7.272.0
HC12
Page 47 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Needs
Upgrade
M3-1 M3-1/UP1 M3-1/UP1.2 M3-1/UP1.3 M3-1/UP2 M3-2 M3-2/SP1
(√/X)
Software Versions
POSITIONER TUBE TILT V2.0.0 V2.0.0 V2.0.0 V2.0.0 V2.0.0 V2.0.0 V2.0.0
HC12 PBL
POSITIONER TUBE TILT V1.4.5 V1.4.5 V1.4.5 V1.4.5 V1.4.5 V1.4.5 V1.4.5
HC12 SBL
Page 48 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
- Workflow corrections:
Correction of the DICOM tag content not filled with good value
Push was CC then MLO but display on RA1000 was MLO then CC
Correction of the search tool in Worklist
Error in image encoding that degrades image in case of print with film composer
PPS not correctly filled in after change view name
RRA summary report is not updated correctly at exam closure
Increase time out from 15 sec to 30 sec to ease association with PACS and avoid push failure
Configuration files are being removed, causing network issue. Correction prevents the file removal
Page 49 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator
Page 50 of 51
Doc# DOC2123058 Rev. 3 FMI 12267
FMI FIELD MODIFICATION INSTRUCTION FMI
GE Healthcare
Enter required data (For each attempt) into Service System and/or attach documentation into Service System, if feasible or send to local FMI Administrator