06 Chapter 2
06 Chapter 2
EXPERIMENTAL METHODS
2.1 Introduction
The structural, morphological and optical properties of investigated Ag, TiO2, AC,
instruments. Scanning Electron Microscopy (SEM) with EDX was used to detect the surface
Transmission Electron Microscopy (HRTEM) was used to identify the morphology, particle
size, and crystallographic by Selected Area Electron Diffraction (SAED). X-ray Diffraction
(XRD) was used to identify the polymorphism, component composition, crystal size, and
used in this study, as well as the required theory and instrumentation details.
Optical absorption is a more reliable way than photoluminescence to study the bandgap.
not have the problem of exciton recombination. A beam of monochromatic light with a uniform
the transmitted light is detected on the backside of the sample. When the photon energy is
greater than the bandgap of the material, the absorption of the photons can take place, and
electrons can be elevated in energy from the valence band to the conduction band, creating
27
electron-hole pairs. If one plots the square of the absorption coefficient as a function of photon
energy, the interband excitation appears as a linear absorption edge. The bandgap energy can
be determined from the intercept of the absorption edge and the baseline. Since the material
absorbs less and can transmit the photons having energy less than the bandgap of the
absorption coefficient (α), which is measured in units of reciprocal distance. There are two
major intrinsic absorption processes involved in determining whether they are direct and
indirect absorption. As a general rule, the larger the bandgap, the smaller the value of a given
wavelength, but the absorption coefficient depends on the density of the states in the conduction
and valence band (Perkampus et al., 2013). The optical absorption coefficient is related to the
(αhν)2 =A (hν-Eg)
optical bandgap. The optical band gap can be evaluated by plotting α2 versus hυ from the
absorption spectra using the relation (αhν)2 α hν-Eg and extrapolating the straight-line portion
of this plot to the energy axis. The variation of (αhν)2 versus hυ is the fundamental absorption
The direct optical absorption is the first-order process involving only the absorbed
photon and is represented approximately as the vertical line on the energy versus wave vector
plot. The absorption transition conserves energy so that ΔE = hѵ (where hѵ is the photon
energy), there is no change in wave vector (k) between the initial and final states except for the
both the absorbed photon and the simultaneously absorbed or emitted phonon, which occurs at
28
different values of the wave vector k at the minimum of the conduction band and the maximum
of the valence band. The absorption transition conserves energy by requiring ΔE = hѵ + Ep,
Ep is the photon energy, and the change in k between the initial and final states k is
equal to the value of k of the phonon involved. The spectrophotometer consists of continuous
change in wavelength and automatic comparison of light intensities of the sample and reference
material. The automatic operation eliminates many time consuming changes and provides a
rapid spectrogram. In the double beam-in-time arrangement, energy from a dispersed source
passes through the exit slit of the monochromator. It is alternated between reference and sample
compartments at 60 Hz (or other fixed frequency). These two beams alternately strike a single
detector where their optical energy is converted into an electrical signal. The sensors output is
an alternating signal whose amplitude is proportional to the difference in intensities in the two
channels. An automatic slit server system constantly maintains the reference signal that widens
or closes the slits to achieve a 100 % transmittance baseline. The switching between reference
and sample paths is done by rotating half sector mirror system, or by a vibrating mirror
assembly, or by a stack of thin horizontal glass plates that are silvered along their edges and
29
Figure 2.1 Schematic diagram of UV-visible spectroscopy
polymorphism and crystal structure. Crystals are symmetrical arrays of atoms containing rows
and planes of high packing density so that they can be three dimensional diffraction gratings
(R.E. Reed-hill et al., 1994). When X-rays of a given frequency strike an atom, they interact
with its electrons, causing them to vibrate with the frequency of the X-ray beam. When a beam
of X-rays irradiates the regular interval spaced atoms, the scattered radiation undergoes
interference appears in other directions. When an electron strikes a metal target at high speed,
the electrons of the target are excited to raise a shell of higher energy level, and the outer
30
electrons fall to fill in the vacancy X-rays are emitted then. An electron emits k X-rays from
the L to the K shell, and an electron produces k X-rays from M to K shell. The k X-ray is used
for the characterization analysis. Each crystalline plane reflects a part of the X-ray beam and
permits it to pass through. When X-rays strike a crystal, the beam is from the surface layer of
atoms and from atoms underneath the surface to a considerable depth (Skoog et al., 2001,
The beam would be reflected not from just two lattice planes but many parallel planes.
The symbol d represents the spacing (or distance) between planes. The crystallite size D can
be evaluated using the Scherer equation, using the full-width at half-maximum (FWHM) value
D = kλ/βcosθ,
X-ray wavelength, k is the Scherer constant, which equals 0.9 for spherical particles but varies
depending on particle form (crystallite, domain) and diffraction reflection indices (h k l). It is
the full width at the half limit of the peaks (Rao et al., 2017). When the relationship is satisfied,
the reflected rays will phase, and constructive interference will result. X-ray diffractometer is
(degree of texture), crystal size, crystallinity, phase composition, residual stress, and
orientation relationships. The PXRD system used in this study was an XPert PANalytical
analytical Diffractometer. The source of XRD was Cu Kα1 radiation (λ=1.5460 Å) operated at
40 keV and 30 mA. The X-ray was passed through a 1mm slit, and the signal was passed
through a 1mm slit, a nickel filter, and another 0.1mm slit. The detector type was Xcelerator
type. The scanning rate is 0.05° step size and a collection interval of 2 seconds per step in the
range from 10° to 80°/140°. The quantity of the sample is around 25-50 mg. The analysis time
31
depends on the scanning rate and the total two theta ranges, and the time for a sample is
the excitation of stretching and bending modes of chemical bonds collecting in infrared spectra.
Instead of recording the amount of energy absorbed, when the frequency of the infrared light
is varied (monochromator), the IR light is guided through an interferometer. After passing the
32
spectroscopy. A beam of infrared light is passed through the sample to analyze a sample, and
the amount of energy absorbed at each wavelength is recorded. This may be finished by
scanning through the spectrum with a monochromatic beam, which changes the wavelength
over time, or by using a Fourier transform instrument to measure all wavelengths at once.
wavelength absorbs the IR region. The working principle of FT-IR is almost dependent on the
detection of the covalent bonds, molecular vibration, and bending of organic functional groups.
In addition, including hydrogen bonding, the inter- or intramolecular interactions are also
determined (Pavia et al., 2001, Skoog et al., 2001). FT-IR spectroscopy is sensitive to the short-
range order structure of molecular solids and is used to study fundamental molecular vibrations,
offering a fingerprint of a molecular solid. Nevertheless, the samples have a limit of high
sensitivity for infrared light. FT-IR is an analytical instrument to identify specific molecular
vibrational frequencies according to the changes in dipole moment. In our study, FT-IR can be
characterization of polymorphism are identified (Li et al., 2003). FT-IR spectroscopies are
single spectrum is faster for the FT-IR technique because the information at all frequencies is
collected simultaneously. This allows multiple samples to be collected and averaged together,
improving sensitivity (Mucciniet al., 2004). FT-IR spectrum was obtained by Perkin Elmer
(Spectrum RXI) and used to measure absorbance, reflection, and transmission of samples in
the wavenumber regions between 4000-400 cm-1. In each case, the chamber was allowed to
equilibrate with a flowing nitrogen purge after admitting the sample into the holder. Purge
times were typically at least one minute but two minutes before the background scan and the
33
sample measurement. About 500 mg of KBr was mixed with 1.5-2.0 mg of the sample. Then
the samples were measured in the regions 400-4000 cm-1 with a resolution of 2 cm-1.
providing high resolution, magnification, and depth of field images than optically attained.
SEM images have a characteristic three-dimensional appearance and help judge the surface
structure of the sample. The main applications of SEM are on topography, morphology, and
composition of samples (Skoog et al., 2001). In a typical SEM, electrons are thermionically
emitted from a tungsten or lanthanum hexaboride (LaB6) cathode and are accelerated towards
an anode; alternatively, electrons can be emitted via Field Emission (FE). Tungsten is selected
because of all metals' highest melting point and lowest vapor pressure, thereby allowing it to
be heated for electron emission. The electron beam, which typically has an energy ranging from
100 eV to 50 keV, is focused by one or two condenser lenses into a beam with an excellent
focal spot sized 1 nm to 5 nm. The incident beam passes through pairs of scanning coils in the
objective lens, which deflect the beam in a raster fashion over a rectangular area of the sample
34
surface. The primary electron beam effectively spreads and fills a teardrop-shaped volume
through these scattering events, known as the interaction volume, extending from less than 100
nm to around 5 m into the surface. Interactions in this region lead to the subsequent emission
of electrons which are then detected to produce an image and provide elemental information
The spatial resolution of the SEM depends on the size of the electron spot, which
depends on the magnetic electro-optical system which produces the scanning beam. The size
of the interaction volume also limits the resolution. The spot size and the interaction volume
are both immense compared to the distances between atoms, so the resolution of the SEM is
not high enough to image down to the atomic scale, as is possible in the transmission electron
microscope (TEM).
35
2.6 High Resolution Transmission Electron Microcopy (HR-TEM)
magnified, and directed to appear either on a fluorescent screen or a layer of photographic film
or detected by a sensor such as a CCD camera. The primary function of TEM is involved in
studying the internal- and microstructure of materials by the diffraction pattern and chemical
composition analysis by Energy Dispersive X-ray Spectroscopy (EDS) and Electron Energy
In TEM, there are two critical systems: (i) the illumination system that controls the
electron source (intensity) and (ii) the imaging system that collects the electrons scattered and
transmitted from the sample. Depending on the mode of operation, the electrons collected by
the imaging system can either form a diffraction pattern or an image. In diffraction mode, the
electrons that form the SAED pattern are elastically scattered by the sample. Thus, if the atomic
arrangement in the sample is periodic, the diffraction pattern consists of well-defined spots. In
contrast, a polycrystalline sample with less atomic order results in diffused spots, arcs, or even
rings. This particular TEM feature is essential and reveals information about the microstructure
of the sample.
Moreover, if the grain size is really fine, the resulting diffraction pattern is in the form
of continuous rings. So, it can be seen that the diffraction technique can also give an idea about
the relative grain sizes of the different samples. The SAED pattern taken from the interface
area provides the diffraction information from the sample, enabling the determination of the
exact relationship. TEM also allows a certain degree of freedom in the rotation of specimens.
If the sample is prepared thoughtfully, the entire crystal structure and the atomic arrangement
36
in more than one direction can be probed using the diffraction technique. A typical diffraction
pattern contains a bright central spot and slightly weaker diffracted spots. The TEM is used
heavily in material science/metallurgy and the biological sciences. The specimens must be
fragile for hundreds of nm and stand in the high vacuum environment and applied voltage
inside the instrument. Because the image formation in the transmission electron microscopy
about the diffraction. The wavelength of an electron is related to its velocity by the following
equation.
ℎ
λ = 𝑚𝑐
Where m is its mass, and h is the Planck constant. In the image mode, the image can be
formed using either the central bright spot (transmitted beam with a small amount of elastically
scattered electrons) or entirely from the scattered electrons. When the image is formed using
the central beam, it is known as a Bright Field (BF), whereas it is known as the Dark Field (DF)
if the diffracted beam or scattered electrons are used to form the image. Both types of images
are based on the amplitude contrast technique. However, DF images usually have better
contrast. Thus, the DF technique is generally used to resolve certain features in the
microstructure (defects, clusters, precipitates, grains, and grain boundaries). The grain size of
the samples has also been calculated using this technique. The imaging techniques are essential
in material structures. The lenses are magnetic, being composed generally of a current-carrying
coil surrounded by a soft iron case. The lenses are energized by direct current. The detail in the
image is formed by the diffraction of electrons from the crystallographic planes of the object
being investigated. Defects in materials influence the electronic and mechanical properties of
materials, so understanding how they behave gives a powerful insight. By carefully selecting
the orientation of the sample, it is possible not just to determine the position of defects and
37
crystal graphics but also to determine the type of defect and crystal graphic present. If the
sample is orientated so that one particular plane is only slightly tilted away from the most vital
diffracting angle (known as the Bragg Angle), any distortion of the crystal plane that locally
tilts the plane to the Bragg angle will produce extreme contrast variations. However, defects
that only displace atoms that do not tilt the crystal to the Bragg angle will not strongly contrast.
𝑛𝜆
θ = 2𝑑
Where θ is the incident angle of the electron beam and is very small, we should expect
the diffracted spots to be deviated through distances that are inversely proportional to the
interplanner spacing d.
that can be performed using TEM. In TEM, an ultra-thin specimen is subjected to a parallel
beam of high energy electrons. As TEM specimens are typically about 100 nm thick and the
electrons usually have high energy of 100-400 keV, the electrons pass through the sample
quickly. In this case, electrons are treated as wave like rather than particle like. Because the
wavelength of high-energy electrons is a fraction of a nanometer and the spacing between the
faces in a solid is only slightly larger, the faces serve as a diffraction grating to the electrons,
which are diffracted. If the sample is moved under the beam, bringing different sections of it
under illumination, the arrangement of the spots in the diffraction pattern will change. If the
sample is tilted, the same crystal will stay under illumination, but different diffraction
conditions will be activated, and other diffraction spots will appear or disappear. SAED is
referred to as "selected" because the user can easily choose the part of the specimen to be
obtained the diffraction pattern. SAED can be used to identify crystal structures and examine
crystal defects. SAED is mainly used in material science and solid-state physics to determine
38
crystallography and polymorphism. The size, composition and atomic structure of the NPs
Dispersive Spectroscopy (EDS) using a 200 keV JEOL JEM 2100 microscope with a lattice
resolution of 0.14 nm and point to point resolution of 0.19 nm. For observation, only a tiny
amount of the sample was necessary. The selected area electron diffraction pattern evaluated
the crystallinity and crystal lattice parameters. The quality and half quantify the elemental
2007).
39
2.7 Energy Dispersive Spectroscopy (EDS)
technique used with transmission electron microscopy (TEM) and SEM. Features as small as
1 m can be analyzed. EDS detects the characteristic X-rays emitted from the sample during
volume. When the electron beam bombards the sample, X-ray photons produced by transitions
between deep atomic levels are ejected from the atoms comprising the sample surface. The X-
ray energy is the characteristic of the element from which it was emitted. X-ray strikes the EDS
detector creates a charge pulse proportional to the X-rays energy. The charge pulse is converted
to a voltage pulse (which remains proportional to the X-ray energy) by a charge sensitive pre
amplifier. The signal is then sent to a multichannel analyzer, where the pulses are sorted by
voltage. The X-ray energy spectrum versus counts is evaluated to determine the relative
abundance of emitted X-rays versus their energy. Hence, the elemental composition of the
sample is considered. Qualitative analysis: X-ray transition energy derived from the EDS
spectrum of the sample can be compared with known characteristic X-ray energy values for
determining the presence of an element in the sample. Elements with atomic numbers ranging
from beryllium to uranium can be detected. The minimum detection limits vary from
approximately 0.1 to a few atomic percent, depending on the element and the sample matrix.
Quantitative analysis: Quantitative results can be obtained from the relative X-ray counts at the
characteristic energy levels for the sample constituents. Semi-quantitative results are readily
available without standards using mathematical corrections based on the analysis parameters
and the sample composition. The accuracy of the non-standard analysis depends on the sample
composition. Greater accuracy is obtained using known standards with structure and
40
Elemental mapping: Characteristic X-ray intensity can be measured relative to the lateral
position on the sample. Variations in X-ray intensity at any characteristic energy value indicate
the relative concentration of the applicable element across the surface. One or more maps are
recorded simultaneously using image brightness intensity to function the local relative
Line profile analysis: The TEM electron beam can be scanned along a preselected line across
the sample while X-rays are detected for discrete positions along the line. Analysis of the X-
ray energy spectrum of each position provides plots of the relative elemental concentration for
each element versus position along the line (Zhang et al., 2008).
41
100 mL aqueous solution of MB dye. Prior to irradiation, the solution was ultra-sonicated and
stirred for 30 min in a darkroom to achieve adsorption/desorption conditions. Then, the solution
was allowed to expose in the direct sunlight under magnetic stirring. At every 10 min time
interval, 5 ml of suspension was sampled and centrifuged at 3000 rpm for 10 min. where the
reaction solution was maintained at 37° C. Then the photodegradation behavior was analyzed
using a UV–Vis spectrophotometer. The effects of photocatalyst dosage and dye concentration
were studied by adopting the same procedure. The degradation efficiency (DE) of MB dye is
Staphylococcus aureus (ATCC 6538), Bacillus subtilis (ATCC 6633), Escherichia coli
(ATCC 8739), and Klebsiella Pneumonia (ATCC 10031) are obtained from the American Type
Culture Collection (ATCC) USA. Using the good diffusion method, the antibacterial activity
of synthesized NPs was tested against gram-positive (S. Aureus and Bacillus) and gram
negative bacteria (E. coli, and Klebsiella). The Muller–Hinton agar was poured into the
sterilized Petri plates and allowed to solidify. After solidification, fresh bacterial culture was
spread over the plates by the spread plate method. Six wells were made in the agar medium
using sterile cork borer and the diameter of each well is 5 mm. About 50 µg/ml of the
nanoparticle solution was loaded in the well and the plates were incubated for 24 h at 36 °C.
After the incubation, the formation of the zone of incubation around the well was observed and
42
scaled in mm. The formation of an inhibition zone indicates the antibacterial activity of
nanoparticles.
2.10 Conclusion
XRD, FT-IR, UV- Vis Spectroscopy, SEM, HR-TEM, and EDS were used to
instruments was fast, economical, and sensible without complex procedures. Besides,
comprehensive technologies are also used for research and measurement of various domains.
The improvement of analytical approaches enhances the progress of science and engineering,
43