USER MANUAL
Thank you very much for purchasing our FH (A) series Fume Hood.
MRC Ltd
MRC Ltd
CONTENT
1. Unpacking, Installation and Debugging .............................................................................................
1.5 Installation........................................................................................................................................9
2.1 Functions............................................................................................................................................
4. Warranty ..........................................................................................................................................34
2 http://www.mrclab.com/
MRC Ltd
Please firstly check whether the packing box is in good condition. If the packing box is damaged,
please take photos and contact the freight carrier. MRC and its dealers are not responsible for
shipping damages
Picture 1
3 http://www.mrclab.com/
MRC Ltd
Picture 2
4 http://www.mrclab.com/
MRC Ltd
Picture 3
Use scissor to cut the packing tape, take off the package cover, then move up the box
body.
Picture 4
5 http://www.mrclab.com/
MRC Ltd
Picture 5
Please refer to the main body unpacking method (of carton box) in the previous pages,
use scissor to cut the packing tape, take off the package cover and move up the box
body.
Picture 6
6 http://www.mrclab.com/
MRC Ltd
Accessories are placed inside the working area and base cabinet. Please refer to 1.5.6. to take
out all accessories and check the completeness referring to this packing list.
2 User manual 1 pc
3 Certification of quality 1 pc
4 Inspection report 1 pc
5 UV lamp 1 pc
6 Fuse (5A) 1 pc
7 Fuse (10A) 1 pc
10 Power cord 1 pc
12 Allen wrench 1 pc
2 Exhaust duct 1 pc
3 Duct clamp 1 pc
7 http://www.mrclab.com/
MRC Ltd
To avoid disturbances to the Fume Hood and its operator, please follow the guideline
below, while determining a suitable location for the equipment
a) Fume Hood should not be installed in positions where they are likely to be
affected by other items or equipment. Windows, doorways, fans, room air supply
diffuser or ventilation outlet should be away from the Fume Hood.
b) The distance from the front window to any circulation space or air-handling
equipment should be at least 1000 mm, so as to preserve a zone undisturbed by
anyone other than the operator
c) The position of a Fume Hood should satisfy the spatial requirements (e.g. vision,
lighting and convenience of access) of the operator and personnel working
nearby.
d) When a Fume Hood is installed on a bench top, the leading edge should be flush
with or slightly overhanging the edge of the bench top.
a) Electrical parameters: consistent with the rated voltage of the Fume Hood (See
2.1.4 Technical parameters and 2.1.5 Performance index)
8 http://www.mrclab.com/
MRC Ltd
b) Power supply need to be grounded (Judging method: test the live wire and the
neutral wire of the main socket with multimeter. The voltage between live and
ground should equal to the voltage of local electrical grid, and the voltage
between neutral and ground should equal to 0. Otherwise, the power supply is
not grounded correctly.)
c) Test the voltage stability before using. If the voltage is unstable, use a voltage
regulator to adjust. Otherwise, the control panel and transformer may be easily
damaged
1.5. Installation
1.5.2. Check the surface of the base cabinet and the main body to make sure there is no
scratch, deformation or foreign bodies
1.5.3. Move the whole equipment as close as possible to the final installation place
1.5.4. Position the base cabinet to the final location where an appropriate power supply is
nearby; and brake the caster to stabilize the base cabinet
1.5.5. Make sure the voltage and frequency of power supply is same as the required value
which is shown on the label. Take out the power cord from the packaging box and
connect it with the female power cord connector on the main body. Plug in the other end
to connect a power supply
1.5.6. Press the “ ” button on the control panel to power on the equipment. Afterwards,
press the “ ” button to raise the front window, take out the accessories from the
operating area and remove the work table as shown in the picture below. Lastly, power
off the equipment and prepare for installation.
9 http://www.mrclab.com/
MRC Ltd
Picture 7
a) Lift the main body and place it on the base cabinet. Make sure each side is in
alignment. Besides, please also make sure the mounting holes of bolt (as shown
in Picture 9) at the bottom of the main body are in alignment with the holes on
the top plate of the base cabinet
b) Make sure the main body and base cabinet are stable enough to prevent side-slip.
Open the doors of the base cabinet and take out the components from the
accessory bag
Insert the hexagon socket head cap screws (M10×20) into the holes at the bottom
of the main body (refer to the Picture 8). These screws should pass through the
main body and the base cabinet. Use flat washer 10, spring washer 10 and nut
10 http://www.mrclab.com/
MRC Ltd
M10 to fasten the screws and nuts. Make sure the base cabinet is well connected
with the main body.
Picture 8
Insert the work table and mount the water sink on the reserved sink hole.
Connect the laboratory water pipe and gas pipe (inflow) with the lower end of the taps.
Pass the drainpipe (provided) through the reserved hole at the side of the base cabinet
(refer to Picture 10) and connect it with the sewer pipe in the laboratory room (water
sink and drainpipe have been connected in factory)
Water tap
Water sink
Picture 9
11 http://www.mrclab.com/
MRC Ltd
Drain hose
Water inlet
Gas inlet
Picture 10
Take out the duct clamp and the exhaust duct. Put the clamp on the exhaust duct and
then connect the exhaust duct with the air exhaust outlet on top of the Fume Hood.
Afterwards, use Slotted screwdriver to fasten the duct clamp firmly. The other end of
the exhaust duct should be fixed outside the laboratory room and in the open air.
Picture 11
12 http://www.mrclab.com/
MRC Ltd
Refer to this table and follow the instruction in 2.4.2, check the following items after
powering on the Fume Hood.
13 http://www.mrclab.com/
MRC Ltd
2. User Instructions
2.1. Functions
This product belongs to FH (A) series Fume Hood. Fume Hood is a kind of negative
pressure ventilation system for protecting operator and laboratory environment. Room
air flows inward from the front opening of the Fume Hood and is consistently ventilated
by the extract blower. Therefore, contaminated fumes, vapors, toxic gasses, aerosol and
corrosive substance which are probably generated during experiments could be
prevented from spreading to the operator or into the laboratory room. In addition, the
polluted air could be purified by the active carbon filter (optional) before exhausting
through the duct. Thus, the outdoor environment could also be protected.
Picture 12
14 http://www.mrclab.com/
MRC Ltd
The primary goal of the Fume Hood is to protect operators and laboratory environment
from exposure to infectious aerosol and toxic fumes which may be generated from the
reaction during experiments.
Model
FH-10A FH-12A FH-15A FH-18A
Parameter
External Dimension
1040*800*2200 mm 1240*800*2200 mm 1540*800*2200 mm 1840*800*2200 mm
(W*D*H)
Working Zone
820*670*730 mm 1020*670*730 mm 1320*670*730 mm 1620*670*730 mm
Dimension (W*D*H)
Power Supply
400 W 400 W 500 W 500 W
Consumption
Fluorescent Lamp
T5 8W T5 12W T5 16W T5 16W
Consumption
Noise ≤70dB(A)
a) Vibration amplitude
b) Illumination
The average illumination would not less than 400 lux. Tested real illumination
would not less than 350 lux
c) Electrical performance
Dielectric Withstand Test: the Fume Hood would not breakdown in 5s when the
voltage increases by 1390V (AC) within 5s
Picture 13
16 http://www.mrclab.com/
MRC Ltd
Driving system consists of tubular motor, front window and hauling mechanism
(hauling sash)
b) UV lamp
The entire work zone could be sterilized effectively by the UV lamp located at
the top of work zone. Emission of 253.7 nanometers could ensure the most
efficient decontamination.
c) Fluorescent light
The Fume Hood is equipped with LED lamp tube, which ensures the standard
requirement of average illumination (400 lux) is met. The measured value at any
point inside the working zone should be greater than 350 lux.
MRC Ltd
d) Socket
Socket, located at the two sides beside the front opening, could supply electricity
NOTE: Please make sure the total load of sockets should be ≤ 500W.
e) Fuse protector
Fuse holders are installed on the top right of the equipment (refer to Figure 13).
Fuse of live wire is inside the female power cord connector. Socket (working
zone) fuse holder and neutral wire (power source) fuse holder is beside the
female connector . The specification of each fuse tube complies with the label
right below the fuse holder. Please refer to the label when replace the fuse tube.
f) Structure
ii. Inner wall of work area is fully made of Phenolic Compact Laminate
which provides corrosion resistance as well as attractive appearance;
work table is made of solid chemical resistant laminate which is easy to
clean and wash.
v. The electronic control system could prevent overload of the circuit and
electric shock, stabilize the performance, protect the equipment and
extend the use life of the Fume Hood.
vi. The sockets (at working zone) adopted non-flammable PC material that is
specialized for laboratory use
18 http://www.mrclab.com/
MRC Ltd
Main functions could all be executed by pressing the relevant button. There are totally 8
common buttons on the control panel (refer to Picture 14). The indicator light above
each button shows the working status of the relevant function. The small LED display
shows the speed level of the blower.
Picture 14
UV lamp, only works when both the blower and the fluorescent lamp are turned off
Front window up, press and hold to continuously raise the window
19 http://www.mrclab.com/
MRC Ltd
Front window down, press and hold to continuously lower the window
Adjust fan speed, press to adjust the speed of the blower (fan) from F1 to F9
The height of the front window could be adjusted by pressing the and
buttons. The window will be moved continuously when pressing the button and it will
stop moving immediately when the button is released
Please make sure the height of the front window is within the safety height range (520
mm).
The blower (fan) could be turned on by pressing button (only when the UV lamp
is turned off). The speed of the fan could be adjusted by pressing button. 9 speed
levels could be selected. The relevant speed level would be displayed on the LED
screen.
When turn off the blower or power off the Fume Hood, the level of fan speed is
memorized by the equipment and would be resumed when turn on the blower again.
2.3.4. UV sterilization
UV light could be turned on by pressing the UV button. Please make sure the window is
fully closed before starting UV sterilization. Interlock function was adopted between
UV light and the blower/fluorescent light. UV light could be automatically turned off
when either the blower or fluorescent light is turned on. During sterilization, people
should leave the room for safety of eyes and skin.
a) Make sure input voltage is correct and stable. The rated load of main power
socket should be higher than cabinet consumption. Plug must be well grounded.
20 http://www.mrclab.com/
MRC Ltd
b) The equipment should be powered off and unplugged before doing any
replacement of parts, such as UV lamp and fluorescent lamp.
d) The air deflector and other internal accessories should be cleaned according to
the use of the Fume Hood
e) The air duct and the blower of the Fume Hood should be cleaned and maintained
regularly in a proper way
g) Do NOT place any soft or tiny materials (such as soft tissue) on the work table
during the operation to prevent breakdown of the blower causing by sucking
those materials
h) The packed Fume Hood should be stored in a warehouse with relative humidity
no more than 75% and temperature lower than 40℃. The warehouse should have
good ventilation performance without acid, alkali or other corrosive gases
b) Power on the Fume Hood by pressing the power switch under the working zone,
c) Press the POWER button on the control panel to enable all functions
(fluorescent lamp, UV lamp, blower, socket, front window). The LED screen
21 http://www.mrclab.com/
MRC Ltd
would display the accumulated operating time of the blower (if the optional
active carbon filter is ordered and equipped, the LED screen would display the
accumulated operating time of the filter).
d) Press the UP button to raise the front window to a proper height. Please
refer to 2.1.4 for the maximum opening of the front window
e) Press the FAN button to turn on the blower. The LED screen would
display the speed level of the fan memorized from the last time of operation. The
indicator light above the FAN button would be turned on to show the working
status of the blower. Make sure the blower runs at least FIVE minutes before
starting any experiment.
f) Press the LAMP button to turn on the fluorescent light. The indicator light
above the button would be turned on to show the working status of the
fluorescent light. Please refer to the actual condition of illumination in the
laboratory room to decide whether the fluorescent light is needed.
g) After finishing the experiment, turn off the blower and the fluorescent light and
h) Press the UV button to turn on the UV light. The indicator light above the
button would be turned on to show the working status of the UV lamp. Please
make sure the sterilization is at least 30 minutes. Press the UV button again to
turn off the UV lamp. Please refer to the actual situation to decide whether
sterilization is needed.
22 http://www.mrclab.com/
MRC Ltd
i) Press the POWER button to power off the Fume Hood after all functions
have been turned off. Press the power switch to disconnect power before
plugging out.
A detailed daily record of operating time is recommended, as the accumulated using time
will directly affect the plan of maintenance.
NOTE: a) To avoid electric shock, please cut off ALL power before applying
maintenance for the equipment!
b) The blower and the exhaust duct should be inspected and maintained
regularly.
Material needed: soap, hot water or warm water, a piece of soft cotton cloth, a piece of
dry cloth or towel, rubbing alcohol or other disinfectants, 1:100 dilution of household
bleach, abrasive household cleaners, sterile water.
23 http://www.mrclab.com/
MRC Ltd
Wipe the entire surface with a soft cotton cloth which has been soaked with
concentrated liquid soap. Afterwards, wipe off the foam with another cotton
cloth or towel which has been soaked with clean hot/warm water. At the end,
wipe the entire surface with a dry cotton cloth or towel rapidly.
For the contaminated or dirty work surface and sump, use 70% rubbing alcohol
or other disinfectant to wipe.
NOTE: Disinfectants used for wiping should not damage the 304
stainless steel.
Use a piece of soft cotton cloth or towel with non-abrasive household cleanser to
wipe the surface.
ii. (Not necessary) Use towel with 70% rubbing alcohol or 1:100 dilution of
household bleach to wipe the working table, the inner face of front
window and the inner wall surface of the working area (exclude the top
wind grid). Use another towel with sterile water to wipe those areas to
erase the remains of chlorine
b) Annual maintenance
i. Check the two lifting belt (sash) of the front window tubular motor, make
sure both of them are well connected to the motor with same tightness
24 http://www.mrclab.com/
MRC Ltd
iii. Apply for overall performance test of the cabinet annually to ensure that
the safety performance has met the requirements. User is responsible for
testing costs
Fume Hood should be stored in a warehouse with relative humidity no more than 75%
and temperature lower than 40℃. The warehouse should have good ventilation
performance without acid, alkali or other corrosive gases. Storage period shall not
exceed one year. Fume Hood stored for more than one year needs to be unpacked and
checked before selling and using. Only the tested and qualified safety cabinet could be
sold.
25 http://www.mrclab.com/
MRC Ltd
26 http://www.mrclab.com/
MRC Ltd
JAC07 Main control panel FH(A) series fume hood main control
27 http://www.mrclab.com/
MRC Ltd
28 http://www.mrclab.com/
MRC Ltd
Picture 15
29 http://www.mrclab.com/
MRC Ltd
Please confirm that the power is well connected, the power cord, the circuit and the
fuses are in good condition (without any damage) before trouble shooting the following
problems
Fluorescent lamp and Make sure the fluorescent lamp and the blower are
UV lamp fail to work Lamp holder Connect the tube and lamp holder tightly
then confirm.
is in good condition
Button fail to work Control panel Check if the button is broken
30 http://www.mrclab.com/
MRC Ltd
Signal transmission line Check whether the signal transmission line is well
connected
Display fail to work
Display screen Check whether the screen is in good condition
31 http://www.mrclab.com/
MRC Ltd
Fuse of socket and fuse of neutral wire is located on top of the equipment (refer
to picture 13). For replacing the fuse, turn off the power and disconnect the plug.
Use a Phillips screwdriver and rotate it anticlockwise to unscrew the fuse holder.
Replace the fuse inside the fuse holder and then, use a Phillips screwdriver and
rotate it clockwise to screw back the fuse holder. Live wire fuse is also located
on top of the equipment, inside the female connector. For replacing the live wire
fuse, turn off the power and disconnect the plug. Use a Slotted screwdriver to
lever up the fuse holder to open it. Replace the fuse inside the fuse holder and
then, press the fuse holder back.
The parameter of the fuse tubes in the round fuse holders are required to
conform to labels, that are F5A φ5×20 mm and F10A φ5×20 mm.
Picture 16
For replacing the lamp tube, press UP button to rise up the front window to the
highest position, then turn off the power and disconnect the plug. Remove the
power cord of the lamp tube refer to picture 17. Then remove the lamp tube by
pulling the small slice on the tube base as shown in the Picture 17. Install a new
lamp tube by pushing it in and connecting with the power cord.
32 http://www.mrclab.com/
MRC Ltd
Picture 17
The UV lamp tube should change regularly, for example every 600 hours. For
replacing the lamp tube, press the UP button to rise up the window to the highest
position, then turn off the power and disconnect the plug. Rotate the tube for 90°
to remove it and then install a new UV lamp tube by rotating in reverse
direction.
Picture 18
33 http://www.mrclab.com/
MRC Ltd
Picture 19
a. Operating area 5A socket fuse label, located under the socket fuse holder
b. 10A power fuse label, located under the female power connector
Picture 20
Picture 21
34 http://www.mrclab.com/
MRC Ltd
4. Warranty
4.1. Warranty is 12 months from EX-factory date (excluding consumable accessories, UV and
Fluorescent lamp, fuse
4.2. MRC would not be liable for any repair of damage caused by improper operation
4.3. If the warranty has been expired, MRC would still responsible for repair with relative
charges
4.4. Life time of fume hood is 8 years from production date on the label
4.5. MRC would provide equipment drawings and necessary technical data for maintenance
companies or personnel trained by MRC engineers
35 http://www.mrclab.com/